0% found this document useful (0 votes)
461 views25 pages

F2000 Cold Head Service Procedures

Uploaded by

Ivan
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
461 views25 pages

F2000 Cold Head Service Procedures

Uploaded by

Ivan
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd

Gyroscan NT/Intera/Achieva Service Procedure Document

Section F2000 cold head

Contents

1. INTRODUCTION .....................................................................................................................................3

2. PROGRAMMING.....................................................................................................................................3

3. TEST PROCEDURES. ............................................................................................................................3

4. REPLACEMENT PROCEDURES. ..........................................................................................................3


4.1. cold head..............................................................................................................................................3
4.1.1. Introduction ...................................................................................................................................3
4.1.2. Preparation ...................................................................................................................................4
4.1.3. Instructions for cold head rebuild................................................................................................10
4.2. Adsorber replacement........................................................................................................................24

5. EXPLANATIONS...................................................................................................................................25

4522 981 24032 (05.0)(CSIP level 1) 1


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Gyroscan NT/Intera/Achieva Service Procedure Document

SPD
Service Procedure Document
F2000 coldhead
Revision 05.0
Issued: June 2005

Figure 1 - F2000 cold head

VALVE MOTOR BASE VARIABLE BONNET / ORIFICE

VALVE MOTOR

VALVE DISK SLACK PISTON STEM


VALVE STEM
SURGE VOLUME HOUSING

DISPLACER

[Link]

4522 981 24032 (05.0)(CSIP level 1) 2


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Gyroscan NT/Intera/Achieva Service Procedure Document

1. INTRODUCTION
This procedure document describes the cold head of the F2000 magnet.

2. PROGRAMMING
None

3. TEST PROCEDURES.
None

4. REPLACEMENT PROCEDURES.

4.1. COLD HEAD.

4.1.1. INTRODUCTION
The complete cold head must be replaced every two years.

CAUTION

The cold head can be damaged during installation if these operations are not done by trained personnel.
Extreme cold can condense air and produce moisture. This can cause ice to form, which may not be
removable. Install or remove the cold head quickly from the open cryostat to prevent air and moisture from
entering the cryostat.
________________

1. Do not remove the cold head if gas cleanup equipment and helium gas of the required quality
are not available.

WARNING

Only trained personnel must perform this procedure. Failure to properly perform this procedure may
result in death or injury, and/or damage to equipment.

________________

[Link].Symbols and definitions

APD Air Product Division:


CHHC Cold head heater controller
Gauss Magnetic field strength
Hz Hertz
K Kelvin
mm millimetre
Refrigerator cold head gas lines and compressor
SHCS socket head cap screws
4522 981 24032 (05.0)(CSIP level 1) 3
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Gyroscan NT/Intera/Achieva Service Procedure Document

SPS slack piston stem


SVH surge volume housing
VM valve motor
VMB valve motor base

4.1.2. PREPARATION

[Link].General information

Review and understand all instructions before beginning this procedure. See Figure - Cold head main parts ,
for and overview of the cold head.

Figure 2 - Cold head main parts

VALVE MOTOR BASE VARIABLE BONNET / ORIFICE

VALVE MOTOR

VALVE DISK SLACK PISTON STEM


VALVE STEM
SURGE VOLUME HOUSING

DISPLACER

[Link]

[Link].Tools and equipment required

4522 150 21651 Refrigerator tool kit


4522 150 22731 Cold head rebuild kit
4522 130 74033 Adsorber
Helium cylinder 50 litre with a minimum pressure of 1000 psi. (70 Bar of gas that is 99,995 (4.5)
pure helium.
Pressure regulator, capable of delivering 300 psi. (20 Bar) (Locally supplied);
Screwdriver, nbr. 2 Phillips;
Hammer, light;
Mirror, inspection;
Flashlight;
Isopropyl alcohol (solvent). Called alcohol in this documentation;
Adjustable wrench (up to 30mm)
Work bench

4522 981 24032 (05.0)(CSIP level 1) 4


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Gyroscan NT/Intera/Achieva Service Procedure Document

Parts details

Item Qty. 12 nc Description PN


201 1 4522 150 22471 Case, tool kit 43526-01
202 1 4522 150 22561 Cold head heater controller (CHHC) 43047-01
203 1 4522 150 22571 Cable Assy, CHHC, mains 43348-01
204 1 4522 150 22581 Cable Assy, CHHC, thermocouple/heater power 43081-52
205 I 4522 150 22711 Service tool assy (cold head): 43520-51
207 1 4522 150 22611 Wrench, hex key, 5mm, tee-handle 43160-01
208 1 4522 150 22551 Screwdriver, flat, non-magnetic 43161-01
209 1 4522 150 22591 Piston puller 43468-01
210 1 4522 150 22601 High pressure hose assy 43469-01
211 1 4522 150 22701 Cover, cylinder debris prevention 43801-01
212 1 4522 150 22481 Wrench 43810-01
213 1 4522 150 22491 Wrench 43810-02
214 1 4522 150 22501 Screwdriver, flat, 4 inch long 43460
215 1 4522 150 22521 1st stage ring seating tool 43190-01
216 2 4522 150 12601 charge/vent adapter 43331-01
217 1 4522 150 22621 Inner slack piston ring seating tool 43191-01
218 1 4522 150 22631 2nd stage ring seating tool 43263-01
219 1 4522 150 22641 Valve stem alignment tool 43199-51
220 1 4522 150 22651 Wrench, variable bonnet/orifice 43264-01
221 1 4522 150 22511 Screwdriver, quick wedge, flat 22329
222 1 4522 150 22671 Brush, spiral tube, 1/4 inch 22330
223 1 4522 150 22691 Screwdriver, miniature 22308
224 1 4522 150 22681 Punch pin, brass, 0.25 x 6 in. long 22309
225 1 4522 150 22691 Probe, precision straight, SS, 5-1/2 inch long 22310
226 1 Set 4522 150 22531 Vinyl gloves (for handling displacer assembly) 22311
227 1 4522 150 22541 20K Self test cable for CHHC 43813-01
228 1 4522 150 23261 80K Self test cable for CHHC 43813-02
229 1 4522 150 23271 Flask 22295

Table 1: Refrigerator maintenance kit, 12nc 4522 150 21651 (IGC P/N 43038-51)

Item Qty 12nc Description Part Number


APD P/N
Cold head cold end replacement parts

01 1 4522 150 22741 Displacer assembly, qualified SC-R8 F264396A


02 1 4522 150 22751 Slack piston stem assy. 264475A
03 1 4522 150 22761 O-Ring, SVH to warm flange 77266
04 1 4522 150 22771 O-Ring, slack piston stem to warm flange 77036

Cold head warm end replacement parts

05 1 4522 150 22781 Valve disc, 2000 252924C


06 1 4522 150 22791 Valve stem, 2000 263790C
07 2 4522 150 22801 O-Ring, valve motor base to SVH (large) and 77140

4522 981 24032 (05.0)(CSIP level 1) 5


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Gyroscan NT/Intera/Achieva Service Procedure Document

valve motor base to VM


08 1 4522 150 22811 O-Ring, valve motor base to SVH (small) 77167
09 2 4522 150 22821 O-Ring, valve stem and slack piston stem 77037
10 1 4522 150 22831 Equalisation capillary, SC-R8 257119B
11 1 4522 150 22841 Teflon stem washer for variable orifice 52650
12 2 4522 150 22921 Stem washer, SS, for variable orifice 52652
13 1 4522 150 22851 O-ring, variable orifice 77034
14 1 4522 150 22861 O-ring, variable orifice port 77035

Miscellaneous service items

15 1 4522 150 22881 Cotton tipped applicator - 6 inch WH# S-14


16 1 4522 150 12471 Vacuum grease, dow corning 2 Oz. 200741A
17 1 4522 150 22891 Teflon tape, 1/4 inch wide WH# T-3
18 1 4522 150 22901 Antiseize thread lubricant, 3-oz. tube 78337
19 1 4522 150 12481 Liquid leak detector, 2-oz. tube 78338
20 1 4522 150 22911 Pipe cleaners, package of 32 WH#P-20
21 1 4522 150 22871 Kim wipes or teri Reinforced wipes WH# K-1

Spare parts

23 2 4522 150 22931 Bolt, socket head cap screw 74505


24 2 4522 150 22941 Washer, flat 72612
25 2 4522 150 22951 Washer, lock 72455
26 1 4522 150 22961 Screw, valve body to valve motor 70878
27 4 4522 150 12611 Face seal, -8M aeroquip 77002
Table 2: Cold head revision kit 12nc 4522 150 22731 (APD P/N 264214A)

Item Qty 12 nc Description Part Number

Compressor assembly replacement parts

01 1 4522 130 74033 Compressor adsorber, HC-8C/E APD P/N


F257428A
Table 3: Compressor adsorber 12nc 4522 130 74033 (APD P/N F257428A)

4522 981 24032 (05.0)(CSIP level 1) 6


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Gyroscan NT/Intera/Achieva Service Procedure Document

Figure 3 - Refrigerator tool kit 1/2

7
17
15 20

18

22
24
9
25
23 14 21

19

[Link]

Figure 4 - Refrigerator tool kit 2/2

29mm 31mm

26

12 13

26mm 9mm

11
16

[Link]

4522 981 24032 (05.0)(CSIP level 1) 7


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Gyroscan NT/Intera/Achieva Service Procedure Document

Figure 5 - Cold head rebuild parts 1/4

2 3 4

DIAM. 104mm DIAM. 82mm

COLDHEAD COLD END REPLACEMENT PARTS

[Link]

Figure 6 - Cold head rebuild parts 2/4

13

7
DIAM. 56mm DIAM. 33mm (3x)

8
12

DIAM. 102mm (2x)

DIAM. 9mm DIAM. 18mm

6 5
10 11 14

DIAM. 8mm DIAM. 11mm

COLD HEAD WARM END REPLACEMENT KIT


[Link]

4522 981 24032 (05.0)(CSIP level 1) 8


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Gyroscan NT/Intera/Achieva Service Procedure Document

Figure 7 - Cold head rebuild parts 3/4

21

SNOOP 19

15 20 18

17
16

GREASE

MISCELLANOUS SERVICE ITEMS


[Link]

Figure 8 - Cold head rebuild parts 4/4

27 25 24

DIAM. 20mm (4x)

23 26

SPARE PARTS
[Link]

4522 981 24032 (05.0)(CSIP level 1) 9


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Gyroscan NT/Intera/Achieva Service Procedure Document

4.1.3. INSTRUCTIONS FOR COLD HEAD REBUILD

[Link].Preliminary set-up items:

Figure 9 - Cold head heater controller connected to the magnet.

MAGNET

20K HEATER 80K HEATER


CONTROLLER CONTROLLER

COLDHEAD MAINS
0 1 0 1
IGC HEATER CONTROLLER

[Link]

CAUTION

Do not switch on the cold head heater controller before the yellow ball valve is open.
A magnet quench may be the result!
________________

1. Verify that there is positive pressure inside the magnet. If there is no positive pressure inside the
magnet, switch on the B0. Switching on the B0 heater is done via the application software with the
following procedure:
1. Login: Gyroscan
2. Select: Scan control.
3. Select: Scan utilities.
4. Select: Enter service mode.
5. Select: System tuning.
6. Select: System Tuning Tools.
7. Select: Installation procedures
8. Select: Magnet monitor procedures
9. Select: Heater control
10. Select: Activate heater
2. Verify that the heater switch on the cold head heater controller (CHHC) is in the OFF position.
3. Open the yellow ball valve in the vent line near the magnet turret.
4. Connect the thermocouple/heater power cable (item 204) to the CHHC and to the 11 pin Deutsch
connector on the magnet. See Figure - Cold head heater controller connected to the magnet.
5. Connect the CHHC mains cable (item 203). Use the mains connector that is also used for the
magnet power supply tool.
6. Switch ON the mains of the CHHC and check if the temperatures displayed on the ‘read-out’, shown

4522 981 24032 (05.0)(CSIP level 1) 10


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Gyroscan NT/Intera/Achieva Service Procedure Document

in degree Celsius about - 220 (20K) and about -190 (80K). Verify that the temperature set point on
the lower display is set to 15 degree.
7. Switch OFF the compressor in the technical room.
8. Switch ON the heaters switch at the CHHC. The warm-up time is approximately 45 minutes.

NOTE

The following steps may be performed while the magnet is warming:


________________

9. Disconnect the expander cable at the cold head.


10. Loosen, but DO NOT remove, the gas line clamps.

WARNING

Use two wrenches to disconnect the gas line couplings to avoid loosening the cold head coupling.
The gas pressure can project the coupling with enough force to cause injury.
________________

11. Disconnect the gas lines at the ‘supply’ and ‘return’ fittings on the cold head.
12. Use 2 wrenches, (items 212 and 213).

Figure 10 – ‘Breaking’ the connection

TURN HOLD

GAS LINE COLD HEAD


COUPLING COUPLING

[Link]

13. Break the fitting loose at the coupling by holding the fitting at the cold head with one wrench
while loosening the gas line nut with the other, see Figure – ‘Breaking’ the connection
14. Disconnect the gas line from the fitting by holding the body of the gas line coupling with one
wrench while unscrewing the coupling nut from the fitting, see Figure – ‘Breaking’ the
connection
15. Once the gas lines are disconnected, push the clamps retaining the gas lines down until the
gas line connections are below the cold head.

Figure 11 - Removing the gas line.

4522 981 24032 (05.0)(CSIP level 1) 11


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Gyroscan NT/Intera/Achieva Service Procedure Document

HOLD TURN

GAS LINE COLD HEAD


COUPLING COUPLING

[Link]

[Link].Install the service tool assembly:

1. Screw the service plate subassembly with 3 posts/captive socket head cap screws (item 05)
to the magnet turret, hand-tightened ONLY. See Figure - Cold head with charge/vent
adapters
2. Mount 2 charge/vent adapters (item 216) to the ‘supply’ and ‘return’ fittings of the cold head.
See Figure - Cold head with charge/vent adapters

Figure 12 - Cold head with charge/vent adapters

4522 981 24032 (05.0)(CSIP level 1) 12


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Gyroscan NT/Intera/Achieva Service Procedure Document

SERVICE TOOL ASSY

MAGNET

[Link]

3. Mount the motor service tool clamp subassembly (item 205) to the service plate.
4. Securely clamp the valve motor with the valve motor clamp such that the magnetic motor is
restrained and cannot be pulled to the magnet by the magnetic field.
5. Securely tighten the 3 post screws to the magnet turret that were only hand-tightened in the
previous step.
6. STOP, DO NOT proceed further until the preprogrammed ‘set point’ temperatures for BOTH
the 20K and the 80K shields have been reached, as indicated on the CHHC.

[Link].Depressurise the cold head:

Vent the cold head to atmosphere pressure by opening both charge/vent adapter valves.

[Link].Separate the valve motor base from the surge volume


housing:

Figure 13 - Valve motor removal

4522 981 24032 (05.0)(CSIP level 1) 13


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Gyroscan NT/Intera/Achieva Service Procedure Document

VALVE MOTOR

A, B, C

7
5
6

9 VALVE MOTOR BASE

7
8

SURGE VOLUME HOUSING

[Link]

1. STOP, DO NOT proceed further until the preprogrammed ‘set point’ temperatures for BOTH
the 20K and the 80K shields have been reached, as indicated on the CHHC.
2. Use the 5mm tee handle wrench (item 207) to remove the 8 socket head cap screws (SHCS),
lock washers and flat washers (items A,B,C) that secure the base to the housing.
3. Use the swing arm of the service tool assembly to rotate the valve motor and valve motor
base subassembly out of the way, approximately 120 degrees clockwise.
4. Remove and discard the ‘O’-rings (items 07 and 08) from the face of the surge volume
housing that has been exposed as a result of the separating of the base from the housing.

[Link].Remove the surge volume housing:

Figure 14 - Sockets head cap screws (SHCS) in the top of the surge volume housing (SVH).

4522 981 24032 (05.0)(CSIP level 1) 14


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Gyroscan NT/Intera/Achieva Service Procedure Document

SURGE VOLUME
HOUSING

8 SHCS

[Link]

1. Use the 5mm tee handle wrench to fully loosen, but DO NOT remove, the 8 SHCS in the top
of the surge volume housing (SVH). See Figure - Sockets head cap screws (SHCS) in the top
of the surge volume housing (SVH).
2. By hand, raise the surge volume housing until the slack piston stem (item 02 of Figure - Cold
head rebuild parts 1/4) is clear of the magnet.
3. Remove the surge volume housing from the magnet area (DO NOT let the SHCS drop out of
the housing).
4. Remove and discard the 2 ‘O’-rings (items 03 and 04) from the grooves in the cylinder in the
magnet.
5. To prevent unwanted debris from accidentally entering the cold head cylinder, cover the
cylinder with the debris prevention cap (item 11).

[Link].Service the surge volume housing:

Figure 15 - Surge volume housing service

4522 981 24032 (05.0)(CSIP level 1) 15


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Gyroscan NT/Intera/Achieva Service Procedure Document

SURGE VOLUME HOUSING


D, E, F VARIABLE ORIFICE ASSEMBLY
DOWEL PIN 9

11, 14

12

10

[Link]

1. Place the surge volume housing on a flat work surface with the valve motor mounting surface
pointing up.
2. Remove the 8 SHCS, lock washers and flat washers.
3. Remove and discard the slack piston stem (SPS) (item 02) using the piston puller (item 209)
from the surge volume housing. Clean the SPS area of the SVH with alcohol soaked wipes.
4. Thoroughly clean the surge volume void of the SVH with alcohol soaked wipes and cotton
tipped swabs (item 15).
5. Replace the equalisation capillary (item 10) from the SVH by using the wedge screwdriver.
(item 221)
The equalization capillary may be difficult to remove due to screw sealant. If the equalization
capillary is difficult to remove, apply heat with a heat gun then try again,
6. Use the wire brush (item 222) to remove all traces of remaining sealant from the capillary
threads in the SVH.
7. Clean the capillary threads and capillary porting with alcohol and cotton tipped swabs. Clean
the capillary path with alcohol soaked pipe cleaner.
8. Dry these threads and capillary path by purging with helium gas.
9. Wrap teflon tape (item 17) 2-3 turns counter clockwise (when facing slotted end) around the
centre threads of the new capillary (item 10).
10. Mount the new capillary inside the SVH with the wedge screwdriver (item 221), by turning the
capillary into the SVH until it hits the bottom

Service the variable bonnet/orifice assembly

1. Fully open the variable orifice/bonnet assembly by turning counter clockwise.


2. Loosen the variable orifice/bonnet from the SVH with the variable orifice/bonnet wrench
(item 220).

Figure 16 - Variable oriffice/bonnet assembly service

4522 981 24032 (05.0)(CSIP level 1) 16


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Gyroscan NT/Intera/Achieva Service Procedure Document

14

13

11, 12

[Link]

3. Remove and discard the 2 stainless steel washers (item 12).


4. Remove and discard the teflon washer (item 11).
5. Use the dental pick (item 225) to remove the needle valve ‘O’-ring (item 13) and discard the ‘O’
ring.
6. Clean the orifice port by using the wire brush to remove all traces of remaining sealant from the
thread inside the SVH.
7. Clean this thread and orifice porting with alcohol and cotton tipped swaps and pipe cleaners (item
20).
8. Dry the orifice porting by purging with helium gas.
9. Clean the orifice/bonnet assembly with the wire brush to remove all traces of sealant from the
threads on the bonnet.
10. Clean these threads and needle with alcohol and cotton tipped swabs.
11. Dry the thread by purging with helium gas.
12. Reassemble the variable orifice/bonnet assembly by wrapping teflon tape (item 17) 2-3 turns
counter clockwise (when facing the slot of the needle valve screw) on the thread of the variable
orifice/bonnet assembly.
13. Unscrew the needle valve until the slot of the valve is nearly flush with the top of the bonnet;
14. Install a new teflon washer (item 11) over the needle and against the bottom of the bonnet;
15. Install the 2 stainless steel washers (item 12.) over the needle and against the teflon washer;
16. Apply a thin uniform layer of vacuum grease (item 16) on the new variable orifice/bonnet ‘O’-ring
(item 13);
Install the ‘O’-ring over the needle and against the bottom stainless steel washer.
17. Use the variable orifice/bonnet wrench to screw the variable orifice assembly into the SVH until the
bonnet is fully seated and tight.
18. Return the needle valve to its original setting:
Screw the needle valve clockwise until it is fully closed (DO NOT OVERTIGHTEN)
Turn the needle counter clockwise 6 full turns to the original factory setting.
19. Apply a thin uniform layer of vacuum grease (item 16) on the new slack piston stem ‘O’-rings (items
09 and 14).
20. Mount the ‘O’-rings on the new slack piston stem (item 02).

4522 981 24032 (05.0)(CSIP level 1) 17


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Gyroscan NT/Intera/Achieva Service Procedure Document

[Link].Service the valve motor base:

CAUTION

The aeroquip fittings are magnetic.


Do not allow the valve motor base to come into contact with the magnet.
________________

Figure 17 - Valve motor base removal

SCREW HOLDING VALVE MOTOR BASE


TO VALVE MOTOR

9
VALVE MOTOR BASE
6
7 5

VALVE MOTOR CLAMP

VALVE MOTOR

[Link]

1. With the valve motor assembly secured in the VM clamp, rotate the valve motor assembly so
that the base of the assembly is pointed up.
2. Use the flat non-magnetic screwdriver (item 208) to remove the 2 screws from the valve
motor base.
3. Remove the motor base and discard the ‘O’-ring (item 07).
4. Remove and discard the valve disc (item 05) from the valve motor drive shaft.
5. Remove the valve motor base from the magnet area.
6. Place the valve motor base on a work bench with the surface that mates to the surge volume
housing, pointed up:

Figure 18 - Valve Motor Base Service.

4522 981 24032 (05.0)(CSIP level 1) 18


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Gyroscan NT/Intera/Achieva Service Procedure Document

1/4" DRIFT PIN


VALVE MOTOR BASE

6
WORK SURFACE

[Link]

7. Use the 1/4 inch brass punch pin (item 224) to carefully punch the valve stem (item 06) out of
the motor base.
8. Clean the VMB assembly with alcohol soaked wipes and dry the VMB by purging with helium
gas.
9. Remove and discard the valve stem and two ‘O’rings (items 06 and 09).
10. Apply a thin uniform layer of vacuum grease (item 16) to the two new valve stem ‘O’-rings
(item 09).
11. Install the ‘O’-rings on the new valve stem and remove any excess vacuum grease (item 06).
12. Insert the valve stem into the valve motor base.
13. Align the valve stem with the valve stem alignment tool (item 219). Rotate the valve stem to
position the valve stem into the mating hole in the VMB. See Figure - Alignment of the valve
stem.
14. Press the stem down with the valve stem alignment tool until the alignment pin is inserted into
it's mating hole.
15. Remove the valve stem alignment tool and finish installing the valve stem by pushing it down
with your thumbs until it is fully seated.
16. Clean the exposed surface of the valve stem with a lint-free wipe (item 21).

CAUTION

No grease, fingerprints, etc. are allowed on the interfacing surface


between the valve stem and the valve disc.
________________

Reassemble the valve motor base to the valve motor:

1. Clean the ‘O’ ring groove, in the VMB, with alcohol and swabs.
2. Apply a thin uniform layer of vacuum grease (item 16) to the new valve motor to valve motor
base ‘O’-ring (item 07).
3. Mount the ‘O’-ring inside the valve motor base groove.

4522 981 24032 (05.0)(CSIP level 1) 19


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Gyroscan NT/Intera/Achieva Service Procedure Document

Figure 19 - Alignment of the valve stem.

[Link]

4. Return to the valve motor clamped to the top of the magnet.


5. Clean the valve motor with alcohol soaked wipes.
6. Place the new valve disc (item 05) onto the valve motor drive shaft.
7. While the valve motor is in the VM clamp, place the valve motor base onto the valve motor,
align the aeroquip fittings of the motor and the base, then secure it with 2 screws using the
flat screwdriver (item 214).

[Link].Service the displacer assembly:

1. Remove the debris prevention cover.


2. Mount the piston puller (item 209) to the top of the displacer assembly (item 01).
3. Lift the displacer assembly out of the magnet, rotate the assembly as needed to clear the
ceiling or any other obstructions.
4. Remove the used displacer assembly from the magnet area.
5. Blow any dust or debris out of the displacer cylinder with helium gas.
6. Inspect the displacer cylinder inside the magnet for remaining dust or debris.
7. If necessary, wipe the cylinder clean with lint free wipes (item 21).
8. Clean the cylinder ‘O’ ring grooves with alcohol and cotton tipped applicators (item 15).

Install the new displacer assembly:

WARNING

When the new displacer is exposed to the atmosphere, it will absorb moisture causing its dimensions to
change. Therefore: The new displacer must NOT be exposed to the atmosphere for more than 10 minutes.
________________

4522 981 24032 (05.0)(CSIP level 1) 20


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Gyroscan NT/Intera/Achieva Service Procedure Document

1. Using gloves (item 26), remove the new displacer assembly (item 01) from its moisture-proof
shipping bag.
2. Be sure the 2nd stage ring seating tool is clean and not damaged.
3. Seat the 2nd stage seals on the displacer assembly, using the 2nd stage ring seating tool
(item 218), by sliding the seating tool, tapered end first, onto the 2nd stage displacer to seat
the seal rings.
Move the tool over the seal rings carefully to compress the rings. Do not rotate the tool.
4. Leave the 2nd stage seating tool installed for a period of five minutes and then remove the
tool.
5. Place the 1st stage ring seating tool (item 215) onto the cylinder. Tool will act as a guide and
seats the 1st stage seals into the cylinder bore. Attach the piston puller to the new displacer.
remove the 2nd stage ring seating tool, THEN
6. Immediately insert the displacer through the 1st stage ring seating tool and into the cylinder
assembly of the magnet.
7. Use care - DO NOT pinch the displacer seals while inserting the displacer into the cylinder.
8. Apply enough force to bottom the new displacer into the cylinder.
9. Remove the puller and the 1st stage ring seating tool.
10. Install debris prevention cover.

[Link].Reassemble the surge volume housing to the magnet:

1. Remove the debris prevention cover.


2. Use the inner slack cap ring seating tool (item 217) to seat the inner slack rings into the white
slack cap. Jiggle the tool back and forth and to the sides to seat the rings completely.
3. Apply a thin uniform layer of vacuum grease (item 16) to the surge volume housing to warm
flange ‘O’-rings (items 03 and 04).
4. Insert the ‘O’-rings into the cylinder grooves.
5. Remove the slack cap ring seating tool from the white slack cap.
6. Install the new slack piston stem (item 02). Carefully insert it into the white slack cap: DO
NOT PINCH THE INTERNAL SEALS in the slack cap.

4522 981 24032 (05.0)(CSIP level 1) 21


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Gyroscan NT/Intera/Achieva Service Procedure Document

Figure 20 - Slack Piston Stem - Dowel Pin Orientation.

SLACK PISTON STEM (SPS)


ROTATED WITH DOWEL PIN
IN 12 0'CLOCK POSITION

O-RING

[Link]

7. Position the SPS so that the dowel pin is aligned with the VM centre line.
8. By hand, bring the surge volume housing over the slack piston stem, such that the protruding
dowel pin in the slack piston stem aligns with a mating alignment hole in the surge volume
housing.
9. Lower the surge volume housing into the white slack cap.
10. Apply a thin coating of antiseize (item 18) to the threads of the 8 SHCS (item D).
11. Install and loosely thread the 8 SHCS, lock washers and flat washers (items D,E,F) into the
surge volume housing assembly.

[Link].Reassemble the valve motor base to the surge volume


housing:

1. Apply a thin uniform layer of vacuum grease (item 16) to the new valve motor base to surge
volume housing Inner and outer ‘O’-rings (items 08 and 07).
2. Insert the ‘O’-rings into the grooves in the surge volume housing.
3. Clean the valve motor assembly face that mates with the surge volume housing with alcohol.
4. Using the VM clamp, rotate the valve motor assembly to align it with the surge volume
housing assembly.
5. Apply a thin coating of antiseize (item 18) to the threads of the SHCS (item A).
6. Move the valve motor base to the surge volume housing and align the surge volume housing
with the valve motor base.
7. Install 2 of the 8 SHCS (items A,B,C) to secure the valve motor base to surge volume
housing.
8. Securely tighten the 8 SHCS in the surge volume housing assembly that were installed in a
previous step.
9. Install the remaining 6 SHCS, lock washers and flat washers (items A,B,C) into the valve
motor base and evenly tighten them.

4522 981 24032 (05.0)(CSIP level 1) 22


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Gyroscan NT/Intera/Achieva Service Procedure Document

[Link].Depressurise the cold head with helium gas:

1. Use the non-magnetic 9/16 inch wrench (item 212) to connect the high pressure hose
assembly (item 10) to the charge/vent adapter on the ‘supply’ fitting of the cold head. Be sure
to purge the tubing of air with a small flow of helium gas while connecting the hose to the
charge/vent adapter. Set the helium supply pressure regulator to 100 psi (6.9 Bar).
2. Open the valve on the supply fitting to pressurise the cold head to 100 ± 5 psi. (6.9 ± 0.34
Bar).
3. Reconnect the expander cable to the cold head.
4. Set the compressor to ‘stand alone’ operation mode.
5. Switch ON the compressor; verify that the valve motor of the cold head is turning (audible).

[Link].Charge, run and vent the cold head:

1. Open the valve on the ‘return’ fitting, to purge gas through the cold head.
2. Close the valve on the ‘supply’ fitting to vent gas to approximately 0 psi (0 Bar) (audible).
3. Open the valve on the supply fitting to repressurise the cold head to 100 ± 5 psi. (6.9 ± 0.34
Bar) with 99,995 (4.5) helium gas.
4. Repeat steps 1 through 3 9 more times.
5. Close return valve.
6. Pressurise the cold head to 280 +/- 5 psi (19.3 ± 0.34 Bar).
7. Close supply valve.
8. Turn OFF the compressor.
9. Leak check all joints with liquid leak detector (item 19).
10. Correct any leaks found.

[Link].Reassemble the gas lines:

1. Switch OFF the heaters on the CHHC.


2. Close the helium gas supply, then disconnect the high pressure hose assembly (item 210)
from the ‘supply’ charge/vent adapter. Remove the charge/vent adapter from the ‘supply’
fitting on the cold head.
3. Remove the lifting fixture,
4. Remove the motor service tool clamp subassembly,
5. Remove the service plate subassembly,
6. Remove the charge/vent adapter from the ‘return’ fitting on the cold head.
7. Inspect the cold head fittings to verify that the face seals are in place and not marred.
8. If the face seal is in good condition, it may be reused.
9. If the face seal is NOT in good condition, or missing, insert a new -8M aeroquip face seal
(item 27).
10. Reconnect the compressor gas lines to the cold head,
11. Leak checks the reconnected couplings with liquid leak detector. (Item 19).
12. Correct any leaks found.

[Link].Place the cold head system into service:

1. Set the compressor to ‘Normal’ operation mode.


2. Turn ON the compressor, verify that the cold head is operating (audible).
4522 981 24032 (05.0)(CSIP level 1) 23
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Gyroscan NT/Intera/Achieva Service Procedure Document

3. Close the yellow ball valve in the vent line near the magnet turret.
4. Disconnect the CHHC.
5. Disconnect the controller power cable and instrument cable from the magnet.
6. Cold head service is complete.
7. Repack all tools.
8. Check the helium level (> 30%).
9. Check the compressor gas pressure. (310 psi - 350 psi)

4.2. ADSORBER REPLACEMENT

The adsorbers of the compressor have to be replaced every 2 years. Revision of the adsorber is not
possible. Venting is not required when using the following procedure:

REMOVAL
1. Switch off the compressor power.
2. Disconnect the gas supply hose.
3. Remove the compressor top and left cover.
4. Disconnect the self-sealing couplings of the adsorber on the inlet side.

WARNING

Do not permit non-self-sealing fittings in the joint to loosen. The gas charge will vent.
________________

5. Loosen the bolt that holds the adsorber to the base from underneath the bottom panel.
6. Loosen the locking nut on the front panel at the supply fitting.
7. Pull the adsorber back until the supply fitting clears from the front panel.
8. Remove the adsorber from the module.

INSTALLATION
1. Remove the plugs from the gas lines of the new adsorber.
2. Position the adsorber on the base and insert the supply fitting through the front panel. Ensure
that the lock washer is installed on the fitting before inserting it through the front panel.
3. Install the bolt that mounts the adsorber to the base. Tighten the bolt and locknut.
4. Connect the adsorber self-sealing coupling on its inlet side.
5. Connect the gas supply hose.
6. Install the covers.

Used adsorber venting and disposal


1. Mount a venting adapter to one of the self-sealing couplings of the used adsorber.
2. Vent the used adsorber to atmospheric pressure.
3. A used adsorber should be treated as chemical waste.

Datasheet compressor
It's very important to know when it was the last time the adsorber has been replaced.
You can check this on the datasheet attached to the compressor. When no form has been attached, please
insert the next form in the plastic cover to the compressor.

Datasheet compressor
IMPORTANT

4522 981 24032 (05.0)(CSIP level 1) 24


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Gyroscan NT/Intera/Achieva Service Procedure Document

This data sheet belongs to the compressor of the refrigerator system of the MR system.
The service engineer is asked to insert this data sheet in the plastic cover, and attach this to the housing of
the compressor after the system has been installed.
Don't forget to complete the next information:

Site location (country, town): ......................................................................

Magnet number: ....................................................................................…..

Cold head serial number: ............................................................................

Cold head part number: ..............................................................................

Installation date of the refrigerator system: ................................................

Service activities on the refrigerator system:


Notice: Please complete this form after EVERY service action on the refrigerator system!!

Action Date Elapsed time Remarks


meter

5. EXPLANATIONS
None.

Proprietary Notice CSIP Level 1:


This document and the information contained in it is strictly reserved for current Philips Medical Systems (“Philips”) personnel, Philips licensed representatives and Philips customers who have
purchased a valid service agreement for use by the customer’s designated in-house service employee on equipment located at the customer’s designated site. Use of this document by unauthorized
persons is strictly prohibited. This document must be returned to Philips when the user is no longer licensed and in any event upon Philips’ first written request.
Liability / Warranty Disclaimer:
Philips provides this DOCUMENT without warranty of any kind, implied or expressed, including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose.
Philips has taken care to ensure the accuracy of this document. However, Philips assumes no liability for errors or omissions and reserves the right to make changes without further notice to any
products herein to improve reliability, function, or design. Philips may make improvements or changes in the product(s) or program(s) described in this document at any time.

4522 981 24032 (05.0)(CSIP level 1) 25


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED

You might also like