SCOPE OF ACCREDITATION TO ISO/IEC 17025:2017
& ANSI/NCSL Z540-1-1994
CALIBRATION
Valid To: Certificate Number:
In recognition of the successful completion of the A2LA evaluation process, accreditation is granted to this
laboratory to perform the following calibrations1, 7:
I. Electrical – DC/Low Frequency
Parameter/Equipment Range CMC2, 3 (±) Comments
DC Voltage – Generate (0 to 330) mV 0.0016 % + 0.81 µV Fluke 5522A
(0.33 to 3.3) V 0.000 86 % + 2.8 µV
(3.3 to 33) V 0.000 96 % + 22 µV
(33 to 330) V 0.0014 % + 0.13 mV
(330 to 1020) V 0.0014 % + 1.2 mV
DC Voltage – Measure (0 to 200) mV 0.00082 % + 0.15 µV Fluke 8508A
(0.2 to 2) V 0.00057 % + 0.64 µV
(2 to 20) V 0.00035 % + 3.9 µV
(20 to 200) V 0.00054 % + 39 µV
(200 to 1000) V 0.00054 % + 0.46 mV
DC Current – Generate (0 to 330) μA 0.012 % + 16 nA Fluke 5522A
(0.33 to 3.3) mA 0.0077 % + 43 nA
(3.3 to 33) mA 0.0076 % + 0.26 uA
(33 to 330) mA 0.0076 % + 3.3 µA
(0.33 to 1.1) A 0.015 % + 35 µA
(1.1 to 3) A 0.028 % + 95 µA
(3 to 11) A 0.038 % + 0.55 mA
(11 to 20.5) A 0.077 % + 0.85 mA
Non-Toroidal Clamp (10 to 16.5) A 0.38 % + 17 mA Fluke 5522A & 50
(16.5 to 150) A 0.39 % + 0.11 A turn coil
(150 to 1025) A 0.38 % + 0.4 A
Parameter/Equipment Range CMC2, 3 (±) Comments
DC Current – Measure (0 to 200) μA 0.0020 % + 0.51 nA Fluke 8508A
(0.2 to 2) mA 0.0023 % + 4.6 nA
(2 to 20) mA 0.0023 % + 51 nA
(20 to 200) mA 0.0076 % + 1 µA
(0.2 to 2) A 0.017 % + 15 µA
(2 to 20) A 0.039 % + 0.32 mA
Parameter/Range Frequency CMC2, 3 (±) Comments
AC Voltage – Generate
(1 to 33) mV (10 to 45) Hz 0.062 % + 4.9 µV Fluke 5522A
(0.045 to 10) kHz 0.012 % + 4.9 µV
(10 to 20) kHz 0.016 % + 4.9 µV
(20 to 50) kHz 0.079 % + 4.9 µV
(50 to 100) kHz 0.27 % + 9.4 µV
(100 to 500) kHz 0.62 % + 39 µV
(33 to 330) mV (10 to 45) Hz 0.023 % + 6.6 µV
(0.045 to 10) kHz 0.011 % + 6.7 µV
(10 to 20) kHz 0.013 % + 6.7 µV
(20 to 50) kHz 0.027 % + 6.5 µV
(50 to 100) kHz 0.062 % + 25 µV
(100 to 500) kHz 0.16 % + 54 µV
(0.33 to 3.3) V (10 to 45) Hz 0.023 % + 39 µV
(0.045 to 10) kHz 0.012 % + 46 µV
(10 to 20) kHz 0.015 % + 47 µV
(20 to 50) kHz 0.023 % + 39 µV
(50 to 100) kHz 0.055 % + 96 µV
(100 to 500) kHz 0.19 % + 0.46 mV
(3.3 to 33) V (10 to 45) Hz 0.023 % + 0.50 mV
(0.045 to 10) kHz 0.012 % + 0.46 mV
(10 to 20) kHz 0.019 % + 0.47 mV
(20 to 50) kHz 0.027 % + 0.46 mV
(50 to 100) kHz 0.070 % + 1.2 mV
(33 to 330) V (0.045 to 1) kHz 0.015 % + 1.5 mV
(1 to 10) kHz 0.016 % + 4.6 mV
(10 to 20) kHz 0.021 % + 4 mV
(20 to 50) kHz 0.025 % + 4.1 mV
(50 to 100) kHz 0.16 % + 39 mV
Parameter/Range Frequency CMC2, 3 (±) Comments
AC Voltage – Generate
(cont)
(330 to 1020) V (0.045 to 1) kHz 0.023 % + 8.8 mV Fluke 5522A
(1 to 5) kHz 0.019 % + 9.4 mV
(5 to 10) kHz 0.023 % + 9.2 mV
AC Voltage – Measure
(0 to 200) mV (1 to 10) Hz 0.027 % + 20 µV Fluke 8508A
(10 to 40) Hz 0.020 % + 6.3 µV
(40 to 100) Hz 0.017 % + 6.3 µV
(0.1 to 2) kHz 0.017 % + 3 µV
(2 to 10) kHz 0.021 % + 6.3 µV
(10 to 30) kHz 0.050 % + 13 µV
(30 to 100) kHz 0.11 % + 30 µV
(0.2 to 2) V (1 to 10) Hz 0.024 % + 0.18 mV
(10 to 40) Hz 0.016 % + 52 µV
(40 to 100) Hz 0.014 % + 30 µV
(0.1 to 2) kHz 0.0068 % + 27 µV
(2 to 10) kHz 0.010 % + 25 µV
(10 to 30) kHz 0.020 % + 43 µV
(30 to 100) kHz 0.050 % + 0.19 mV
(100 to 300) kHz 0.23 % + 1.9 mV
(0.3 to 1) MHz 0.78 % + 19 mV
(2 to 20) V (1 to 10) Hz 0.015 % + 1.1 mV
(10 to 40) Hz 0.010 % + 0.19 mV
(40 to 100) Hz 0.0085 % + 0.19 mV
(0.1 to 2) kHz 0.0070 % + 0.19 mV
(2 to 10) kHz 0.010 % + 0.19 mV
(10 to 30) kHz 0.020 % + 0.39 mV
(30 to 100) kHz 0.050 % + 1.9 mV
(100 to 300) kHz 0.23 % + 19 mV
(0.3 to 1) MHz 0.78 % + 0.19 V
(20 to 200) V (1 to 10) Hz 0.015 % + 11 mV
(10 to 40) Hz 0.010 % + 1.9 mV
(40 to 100) Hz 0.0085 % + 1.9 mV
(0.1 to 2) kHz 0.0070 % + 1.9 mV
(2 to 10) kHz 0.010 % + 1.9 mV
(10 to 30) kHz 0.020 % + 3.9 mV
(30 to 100) kHz 0.050 % + 19 mV
(100 to 300) kHz 0.23 % + 0.19 V
(0.3 to 1) MHz 0.78 % + 1.9 V
Parameter/Range Frequency CMC2, 3 (±) Comments
AC Voltage – Measure
(cont)
(200 to 1050) V (1 to 10) Hz 0.015 % + 62 mV Fluke 8508A
(10 to 40) Hz 0.011 % + 19 mV
(0.040 to 10) kHz 0.011 % + 19 mV
(10 to 30) kHz 0.021 % + 39 mV
(30 to 100) kHz 0.054 % + 0.19 V
AC Current – Generate
(29 to 330) μA (10 to 20) Hz 0.16 % + 78 nA Fluke 5522A
(20 to 45) Hz 0.12 % + 77 nA
(0.045 to 1) kHz 0.098 % + 78 nA
(1 to 5) kHz 0.23 % + 0.12 µA
(5 to 10) kHz 0.62 % + 0.16 µA
(10 to 30) kHz 1.2 % + 0.31 µA
(0.33 to 3.3) mA (10 to 20) Hz 0.16 % + 0.12 µA
(20 to 45) Hz 0.098 % + 0.11 µA
(0.045 to 1) kHz 0.078 % + 0.12 µA
(1 to 5) kHz 0.16 % + 0.15 µA
(5 to 10) kHz 0.40 % + 0.19 µA
(10 to 30) kHz 0.78 % + 0.46 µA
(3.3 to 33) mA (10 to 20) Hz 0.14 % + 1.5 µA
(20 to 45) Hz 0.071 % + 1.5 µA
(0.045 to 1) kHz 0.034 % + 1.5 µA
(1 to 5) kHz 0.064 % + 1.5 µA
(5 to 10) kHz 0.16 % + 2.3 µA
(10 to 30) kHz 0.32 % + 3 µA
(33 to 330) mA (10 to 20) Hz 0.14 % + 16 µA
(20 to 45) Hz 0.071 % + 16 µA
(0.045 to 1) kHz 0.033 % + 16 µA
(1 to 5) kHz 0.079 % + 39 µA
(5 to 10) kHz 0.16 % + 77 µA
(10 to 30) kHz 0.32 % + 0.15 mA
(0.33 to 1.1) A (10 to 45) Hz 0.14 % + 80 µA
(0.045 to 1) kHz 0.04 % + 82 µA
(1 to 5) kHz 0.47 % + 0.77 mA
(5 to 10) kHz 1.9 % + 3.9 mA
Parameter/Range Frequency CMC2, 3 (±) Comments
AC Current – Generate
(cont)
(1.1 to 3) A (10 to 45) Hz 0.14 % + 72 µA Fluke 5522A
(0.045 to 1) kHz 0.058 % + 35 µA
(1 to 5) kHz 0.47 % + 0.75 mA
(5 to 10) kHz 1.9 % + 3.9 mA
(3 to 11) A (45 to 100) Hz 0.051 % + 1.6 mA
(0.1 to 1) kHz 0.08 % + 1.6 mA
(1 to 5) kHz 2.3 % + 1.5 mA
(11 to 20.5) A (45 to 100) Hz 0.095 % + 3.9 mA
(0.1 to 1) kHz 0.12 % + 3.9 mA
(1 to 5) kHz 2.3 % + 3.9 mA
Clamp Ammeters –
Toroidal
(10 to 16.5) A (45 to 65) Hz 0.012 % + 0.6 A Fluke 5522A & 50
(65 to 440) Hz 0.10 % + 0.59 A turn coil
(16.5 to 150) A (45 to 65) Hz 0.080 % + 0.59 A
(65 to 440) Hz 0.42 % + 0.54 A
(150 to 1025) A (45 to 65) Hz 0.19 % + 0.45 A
(65 to 440) Hz 0.52 % + 1.3 A
Clamp Ammeters –
Non-Toroidal
(10 to 16.5) A (45 to 65) Hz 0.057 % + 0.6 A Fluke 5522A & 50
(65 to 440) Hz 0.17 % + 0.59 A turn coil
(16.5 to 150) A (45 to 65) Hz 0.29% + 0.61 A
(65 to 440) Hz 0.64 % + 0.58 A
(150 to 1025) A (45 to 65) Hz 0.41 % + 0.87 A
(65 to 440) Hz 0.68 % + 1.6 A
Parameter/Range Frequency CMC2, 3 (±) Comments
AC Current – Measure
(0 to 200) μA (1 to 10) Hz 0.078 % + 30 nA Fluke 8508A
(0.01 to 10) kHz 0.078 % + 30 nA
(10 to 30) kHz 0.10 % + 30 nA
(30 to 100) kHz 0.51 % + 30 nA
(0.2 to 2) mA (1 to 10) Hz 0.051 % + 0.30 µA
(0.01 to 10) kHz 0.047 % + 0.30 µA
(10 to 30) kHz 0.10 % + 0.30 µA
(30 to 100) kHz 0.51 % + 0.30 µA
(2 to 20) mA (1 to 10) Hz 0.051 % + 3 µA
(0.01 to 10) kHz 0.047 % + 3 µA
(10 to 30) kHz 0.10 % + 3 µA
(30 to 100) kHz 0.51 % + 3 µA
(20 to 200) mA (1 to 10) Hz 0.051 % + 30 µA
(0.01 to 10) kHz 0.046 % + 30 µA
(10 to 30) kHz 0.094 % + 30 µA
(0.2 to 2) A (0.01 to 2) kHz 0.056 % + 0.20 mA
(2 to 10) kHz 0.068 % + 0.25 mA
(10 to 30) kHz 0.22 % + 0.39 mA
(2 to 20) A (0.01 to 2) kHz 0.071 % + 1.9 mA
(2 to 10) kHz 0.19 % + 1.9 mA
Parameter/Equipment Range CMC2, 3 (±) Comments
Resistance – Generate (0 to 11) Ω 0.0031 % + 7.8 mΩ Fluke 5522A
(11 to 33) Ω 0.0023 % + 12 mΩ
(33 to 110) Ω 0.0022 % + 12 mΩ
(110 to 330) Ω 0.0022 % + 15 mΩ
(0.33 to 1.1) kΩ 0.0022 % + 15 mΩ
(1.1 to 3.3) kΩ 0.0022 % + 0.15 Ω
(3.3 to 11) kΩ 0.0023 % + 0.77 Ω
(11 to 33) kΩ 0.0022 % + 0.78 Ω
(33 to 110) kΩ 0.0023 % + 0.77 Ω
(110 to 330) kΩ 0.0025 % + 7.7 Ω
(0.33 to 1.1) MΩ 0.0025 % + 8.0 Ω
(1.1 to 3.3) MΩ 0.0048 % + 0.12 kΩ
(3.3 to 11) MΩ 0.010 % + 0.19 kΩ
(11 to 33) MΩ 0.020 % + 1.9 kΩ
(33 to 110) MΩ 0.04 % + 2.3 kΩ
(110 to 330) MΩ 0.23 % + 78 kΩ
(330 to 1100) MΩ 1.2 % + 0.39 MΩ
Resistance – Measure (0 to 2) Ω 0.0015 % + 9.3 µΩ Fluke 8508A
(2 to 20) Ω 0.000 85 % + 31 µΩ
(20 to 200) Ω 0.000 77 % + 48 µΩ
(0.2 to 2) kΩ 0.000 78 % + 0.47 mΩ
(2 to 20) kΩ 0.000 78 % + 4.7 mΩ
(20 to 200) kΩ 0.000 78 % + 47 mΩ
(0.2 to 2) MΩ 0.000 94 % + 0.91 Ω
(2 to 20) MΩ 0.0019 % + 93 Ω
(20 to 200) MΩ 0.012 % + 9.3 kΩ
Capacitance – Generate (220 to 399.9) pF 0.37 % + 8 pF Fluke 5522A
(0.4 to 1.1) nF 0.37 % + 8.1 pF
(1.1 to 3.3) nF 0.39 % + 7.8 pF
(3.3 to 11) nF 0.20 % + 8.3 pF
(11 to 33) nF 0.20 % + 8.0 pF
(33 to 110) nF 0.18 % + 31 pF
(110 to 330) nF 0.19 % + 56 pF
(0.33 to 1.1) μF 0.19 % + 0.9 nF
(1.1 to 3.3) μF 0.20 % + 2.5 nF
(3.3 to 11) μF 0.20 % + 8.8 nF
(11 to 33) μF 0.31 % + 25 nF
(33 to 110) μF 0.35 % + 91 nF
(110 to 330) μF 0.35 % + 0.27 µF
(0.33 to 1.1) mF 0.35 % + 0.89 µF
(1.1 to 3.3) mF 0.45 % + 3 µF
(3.3 to 11) mF 0.45 % + 10 µF
(11 to 33) mF 0.75 % + 30 µF
(33 to 110) mF 1.1 % + 0.10 mF
Parameter/Equipment Range CMC2 (±) Comments
Electrical Simulation of
Thermocouple &
Thermocouple Indicating
Devices – Measure &
Generate
Type B (600 to 800) °C 0.34 °C Fluke 5522A
(800 to 1000) °C 0.27 °C
(1000 to 1550) °C 0.24 °C
(1550 to 1820) °C 0.25 °C
Type C (0 to 150) °C 0.24 °C
(150 to 650) °C 0.21 °C
(650 to 1000) °C 0.25 °C
(1000 to 1800) °C 0.39 °C
(1800 to 2316) °C 0.65 °C
Type E (-250 to -100) °C 0.39 °C
(-100 to -25) °C 0.13 °C
(-25 to 350) °C 0.12 °C
(250 to 650) °C 0.13 °C
(650 to 1000) °C 0.17 °C
Type J (-210 to -100) °C 0.21 °C
(-100 to -30) °C 0.13 °C
(-30 to 150) °C 0.12 °C
(150 to 760) °C 0.14 °C
(760 to 1200) °C 0.18 °C
Type K (-200 to -100) °C 0.26 °C
(-100 to -25) °C 0.15 °C
(-25 to 120) °C 0.13 °C
(120 to 1000) °C 0.21 °C
(1000 to 1372) °C 0.31 °C
Type N (-200 to -100) °C 0.31 °C
(-100 to -25) °C 0.18 °C
(-25 to 120) °C 0.15 °C
(120 to 410) °C 0.15 °C
(410 to 1300) °C 0.21 °C
Parameter/Equipment Range CMC2 (±) Comments
Electrical Simulation of
Thermocouple &
Thermocouple Indicating
Devices –
Measure & Generate
(cont)
Type R (0 to 250) °C 0.45 °C Fluke 5522A
(250 to 400) °C 0.27 °C
(400 to 1000) °C 0.26 °C
(1000 to 1767) °C 0.31 °C
Type S (0 to 250) °C 0.38 °C
(250 to 1000) °C 0.29 °C
(1000 to 1400) °C 0.29 °C
(1000 to 1767) °C 0.35 °C
Type T (-250 to -150) °C 0.47 °C
(-150 to 0) °C 0.19 °C
(0 to 120) °C 0.13 °C
(120 to 400) °C 0.12 °C
Type U (-200 to 0) °C 0.44 °C
(0 to 600) °C 0.21 °C
Electrical Simulation of
RTD Indicating Devices
– Generate
Pt 385, 100 Ω (-200 to -80) °C 0.039 °C Fluke 5522A
(-80 to 0) °C 0.039 °C
(0 to 100) °C 0.054 °C
(100 to 300) °C 0.070 °C
(300 to 400) °C 0.078 °C
(400 to 630) °C 0.093 °C
(630 to 800) °C 0.18 °C
Pt 3926, 100 Ω (-200 to -80) °C 0.039 °C
(-80 to 0) °C 0.039 °C
(0 to 100) °C 0.054 °C
(100 to 300) °C 0.070 °C
(300 to 400) °C 0.078 °C
(400 to 630) °C 0.093 °C
Parameter/Equipment Range CMC2 (±) Comments
Electrical Simulation of
RTD Indicating Devices
– Generate (cont)
Pt 3916, 100 Ω (-200 to -190) °C 0.19 °C Fluke 5522A
(-190 to -80) °C 0.031 °C
(-80 to 0) °C 0.039 °C
(0 to 100) °C 0.047 °C
(100 to 260) °C 0.054 °C
(260 to 300) °C 0.062 °C
(300 to 400) °C 0.070 °C
(400 to 600) °C 0.078 °C
(600 to 630) °C 0.18 °C
Pt 385, 200 Ω (-200 to -80) °C 0.031 °C
(-80 to 0) °C 0.031 °C
(0 to 100) °C 0.031 °C
(100 to 260) °C 0.037 °C
(260 to 300) °C 0.094 °C
(300 to 400) °C 0.10 °C
(400 to 600) °C 0.11 °C
(600 to 630) °C 0.12 °C
Pt 385, 500 Ω (-200 to -80) °C 0.037 °C
(-80 to 0) °C 0.042 °C
(0 to 100) °C 0.042 °C
(100 to 260) °C 0.049 °C
(260 to 300) °C 0.064 °C
(300 to 400) °C 0.043 °C
(400 to 600) °C 0.078 °C
(600 to 630) °C 0.09 °C
Pt 385, 1000 Ω (-200 to -80) °C 0.024 °C
(-80 to 0) °C 0.024 °C
(0 to 100) °C 0.032 °C
(100 to 260) °C 0.038 °C
(260 to 300) °C 0.053 °C
(300 to 400) °C 0.06 °C
(400 to 600) °C 0.06 °C
(600 to 630) °C 0.18 °C
PtNi 385, 120 Ω (-80 to 0) °C 0.062 °C
(0 to 100) °C 0.062 °C
(100 to 260) °C 0.11 °C
Cu 427, 10 Ω (-100 to 260) °C 0.23 °C
II. Time & Frequency
Parameter/Equipment Range CMC2, 4 (±) Comments
Frequency – Measuring (0.01 to 119.99) Hz 5.8 mHz Fluke 5522A
Equipment (120 to 1199.9) Hz 58 mHz
(1.2 to 11.999) kHz 0.58 Hz
(12 to 119.99) kHz 5.8 Hz
(120 to 1199.9) kHz 58 Hz
(1.2 to 2) MHz 580 Hz
____________________________________________
1
This laboratory offers commercial calibration service.
2
Calibration and Measurement Capability Uncertainty (CMC) is the smallest uncertainty of measurement
that a laboratory can achieve within its scope of accreditation when performing more or less routine
calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. CMCs represent
expanded uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage
factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory
may be greater than the CMC due to the behavior of the customer’s device and to influences from the
circumstances of the specific calibration.
3
The stated measured values are determined using the indicated instrument (see Comments). This
capability is suitable for the calibration of the devices intended to measure or generate the measured value
in the ranges indicated. CMC’s are expressed as either a specific value that covers the full range or as a
percent or fraction of the reading plus a fixed floor specification.
4
The type of instrument or material being calibrated is defined by the parameter. This indicates the
laboratory is capable of calibrating instruments that measure or generate the values in the ranges indicated
for the listed measurement parameter.
5
This scope meets A2LA’s P112 Flexible Scope Policy.