Cbc Automation
Cbc Automation
Varun Uppal
Ishwar Chand
TIMELINE…..
Green, Ralph, and Sebastian Wachsmann-Hogiu. "Development, history, and future of automated cell counters." Clinics in laboratory medicine 35.1
(2015): 1-10
AUTOMATION
Greek-automatos: self acting
Definition:
The process of following a pre determined process of operations with
little or no human labor, using specialized equipment and devices that
perform and control processes.
Why automation?
Space Economy
Increase work efficiency of lab
Manpower Economy
Time Economy
Accuracy
Ensures cost-effectiveness
Various Generations Of Coulter Counter
• Semi-automated instruments
• Require some steps, as dilution of blood samples
• Often measure only a small number of variables
Carryover of cells from one sample to the next also is minimized by these
internal cleaning systems.
Coincident passage of more than one cell at a time through the orifice
causes artificially large pulses, which results in falsely increased cell
volumes and falsely decreased cell counts
• Orientation of the cell in the center of
the aperture
• Deformability of the RBC, which may be
altered by decreased hemoglobin
content
Hematology analyzers have some common basic components, including hydraulics, pneumatics, and
electrical systems.
The hydraulics system: an aspirating unit, dispensers, diluters, mixing chambers, aperture baths or
flow cells or both, and a hemoglobinometer.
The pneumatics system: generates the vacuums and pressures required for operating the valves and
moving the sample through the hydraulics system.
The electrical system: controls operational sequences of the total system and includes electronic
analyzers and computing circuitry for processing the data generate. A data
display unit receives information from the analyzer and prints results,
histograms, or cytograms.
BECKMAN COULTER
REAGENTS INVOLVED
Diluent
ISOTON 4 diluent, are isotonic electrolyte solutions that:
• Dilute whole-blood samples.
• Stabilize cell membranes for accurate counting and sizing.
• Conducts aperture current.
• Rinse instrument components between analyses.
• Carry and focus the sample stream in the flow cell to direct the blood cells through the aperture.
Cleaners
• LH Series Cleaner and COULTER CLENZ cleaning agent clean and rinse the internal surfaces of the
instrument components.
• Daily use prevents protein buildup and eliminates the need for routine aperture bleaching and blood
sampling valve maintenance.
A pump draws a maximum of 300 μL (LH 700 Series only) or
550 μL (LH 700 Series SlideMaker) of sample through the
needle and through the Blood Sampling Valve (BSV).
DELIVERY
CBC
After the sample is aspirated:
• The center section of the BSV rotates and segments the sample
into two separate volumes.
• One volume of sample, 1.6 μL, is delivered with 10 mL of diluent to
the RBC bath. This dilution is used for RBC/Plt counting and
MCV/Plt sizing.
• The other volume, 28 μL, is delivered with 6 mL of diluent to the
WBC bath. This dilution is used to count WBC and develop Hgb.
• During delivery to the WBC bath, 1 mL of lytic reagent is added to
the dilution to lyse the red cells and convert Hgb
Conductivity
Alternating current in the radio frequency (RF) range short circuits the bipolar lipid layer of a cell's
membrane, allowing the energy to penetrate the cell. This powerful probe is used to collect
information about cell size and internal structure, including chemical composition and nuclear
volume.
Scatter
When a cell is struck by the coherent light of a LASER beam, the scattered light spreads out in all
directions. Using a proprietary new detector, median angle light scatter (MALS) signals, are
collected to obtain information about cellular granularity, nuclear lobularity and cell
surface structure
Simultaneous Measurements:
VCS is the only single channel analysis that uses 3 independent energy sources to probe
approximately 8,192 cells in their near native state
Reticulocyte Analysis
• A supravital dye, New Methylene Blue, is incubated with whole-blood
samples. The dye precipitates the basophilic RNA network found in
reticulocytes. Hemoglobin and unbound stain are removed by adding a
clearing reagent, leaving clear spherical mature RBCs and darkly stained
reticulocytes.
• Stained reticulocytes are differentiated from mature red cells and other
cell populations by light scatter, direct current measurements, and opacity
characteristics
DATAPLOT DEVELOPMENT
At the Workstation, raw data is analyzed to:
• Position each cell within a three-dimensional space
• Plot the data on a two-dimensional DataPlot
• Display and report results
• The Beckman Coulter AccuGate algorithm uses adaptive contouring methods designed for finding optimal
separation between overlapping clusters of data
• In the DataPlots, different colors represent different memberships (types of cells)
• DIFF:
Lymphocytes Blue
Neutrophils Purple
Eosinophils Orange
Monocytes Green
Basophils White
Non-white debris Red
• RETIC:
RBCs Red
Cell volume (VOL), determined by the low-frequency The Reticulocyte Data Plot shows mature red cells and
impedance measurement, is plotted on Reticulocytes. Cell volume is plotted
the Y-axis; rotated light scatter (RLS) is plotted on the X- on the Y-axis, and linear light scatter (LLS) is plotted on
axis. the X-axis.
The 3D DataPlot view classifies by density, light scatter and opacity.
PARAMETERS
Parameters are either:
• Measured directly. • Derived From RBC or Plt Histogram
• Calculated
• Derived From WBC Histogram
RED CELL INDICES
• MCV :The counter provides MCV which is derived from the histogram (sum
of pulse height / sum of pulse).
• MCH: This is the average weight of the hemoglobin in picograms in a red
cell. It is a calculated value.
MCH =hemoglobin in pg/L / red cell count in milions/L
• MCHC :This indicates the average weight of hemoglobin as compared to the
cell size.
MCHC = (Hemoglobin in g/dL / HCT) x 100
• Hematocrit: Hematocrit (Hct) or (PCV) is the volume of the red cells as
compared to the volume of the whole blood sample
HCT(%) = (RBC X MCV ) /10
RDW: quantitative measurement of cell volume, an equivalent of
microscopic assessment of degree of anisocytosis
Can be expressed as RDW-SD or RDW-CV
RDW-SD is the arithmetic distribution Width RDW-CV is computed as
measured at the 20% Relative Height Level, RDW-CV (%) = 100 X δ/µ
taking Histogram peak as 100%.
The RBC Histogram of normal blood sample
crosses this 20% level twice. RBC-SD is
reported in Femtolitres
PLATELET INDICES
• Mean platelet volume:
The average volume of individual platelets derived from the Plt histogram.
It represents the mean volume of the Plt population under the fitted Plt
curve multiplied by a calibration factor.
This number is expressed in femtoliters.
• Need?
Visual scanning of the histogram gives a good initial sense of the range, size,
shape, and other salient features of the cell morphology.
In addition, it is frequently used, along with the peripheral blood film, as an
aid in monitoring and interpreting abnormal morphological changes,
RBC HISTOGRAM
Distribution of RBC population according to their size and their relative
number.
The Normal red cell distribution curve (Histogram) is Gaussian (bell shape)
and the peak of the curve should be within the normal MCV of 80.0-100.0 fL
The size ranges for RBC histograms are between 24 fL and 360 fL, the
instrument counts only those cells with volume sizes between 36 fL and 360
fL as red cells.
Two discriminators are used upper and lower
Lower Discriminator is placed between the platelets and RBCs
Red Cells Histogram
• (24- 36 fl ) flag may be due
1- RBCs fragments
2- WBC's fragments
3- Giant plts
4- Microcyte
• Shift to right :
- Leukemia
- Macrocytic anemia
- Megaloblastic anemia
• Shift to left :
- Microcytic anemia (IDA)
• Bimodal
- IDA, Megaloblastic anemia with transfusion.
-Sideroblastic anemia.
• Cold agglutination
PLATELET HISTOGRAM
Normal platelet histogram displays cells from (2-20 fl).
• (0-2)
• Air Bubbles
• Dust
• Electronic and Electrical noise
• Over 20 fL
• Microcyte
• Scishtocyte
• WBC's fragments
• Giant Plts
• Clumped plts
WBC HISTOGRAM
• 3-part differential usually counts
• Granulocytes or large cells
• Lymphocytes or small cells
• Monocytes(mononuclear cells) or (middle cells)
• 5-part classify cells to
• Neutrophils
• Eosinophils
• Basophils
• Lymphocytes
• Monocytes
• A sixth category designated “large unstained cells” include cells larger than normal and lacking
the peroxidase activity which include
• Atypical lymphocytes
• Various other abnormal cells.
• Other counters identifies 7 categories including
• Large immature cells(composed of blasts and immature granulocytes)
• Atypical lymphocytes(including blast cells).
Histogram provides Percent and Absolute
Count for these three populations
Small Cell Ratio [SCR]/Small Cell Count [SCC]
Mixed Cell Ratio [MCR]/Middle or Mixed Cell Count [MCC]
Large Cell Ratio [LCR]/Large Cell Count [LCC]
SYSMEX INSTRUMENTATION
• Provide complete RBC, platelet, and WBC analysis with three-part differential; the larger XT-1800i
(SF-3000 and SE-9000) that performs a CBC with five-part differential; and the XE series and the
newest XN series that also provide a fully automated reticulocyte count
• Two unique features enhance the impedance technology:
in the RBC/platelet channel, a sheathed stream with hydrodynamic focusing is used to direct cells
through the aperture, which reduces coincident passage, particle volume distortion, and
recirculation of blood cells around the aperture; and
in the WBC and RBC/platelet channels, floating thresholds are used to discriminate each cell
population
-This floating threshold circuitry allows for discrimination of cell populations on a sample-by-
sample basis
• The platelet analysis on the XN also utilizes a fluorescent count, in addition to the impedance
count and optical count, called the PLT-F, performed by optical measurement.
• In the hemoglobinflow cell, hemoglobin is oxidized and binds to sodium lauryl sulfate (SLS)
forming a stable SLS–hemoglobin complex, which is measured photometrically at 555 nm
• The SE-9000/9500 uses four detection chambers to analyze WBCs and obtain a five-part differential:
the DIFF, IMI (immature myeloid information), EO, and BASO chambers by low-frequency DC and high-
frequency current(DC/RF detection method)
• In the XN-1000, fluorescent flow cytometry is used for the WBC count, WBC differential, and
enumeration of nucleated RBCs.
• The XE-5000, the XT-2000i and the XN series determines the reticulocyte count and IRF by measuring
forward scatter and side fluorescence.
• RET-He (reticulocyte hemoglobin equivalent) that measures the hemoglobin content of the
reticulocytes.It uses a proprietary polymethine dye to fluorescently stain the reticulocyte nucleic acids.
ABBOTT INSTRUMENTATION
• Instruments include the smaller CELL-DYN Emerald, which provides CBC with three-part differential, and the larger CELL-
DYN Sapphire and the midrange CELL-DYN Ruby,provide a CBC with five-part differential and random fully automated
reticulocyte analysis
• A unique von Behrens plate is located in the RBC/platelet counting chamber to minimize the effect of recirculating cells.
• The platelet analysis is based on a two-dimensional optical platelet count using fluorescent technology, the same
technology used for direct nucleated RBC counting by adding a red fluorescence to the sample to stain nucleated red
cells.
• The WBC count and differential are derived from the optical channel using CELL-DYN’s patented multiangle polarized
scatter separation (MAPSS) technology with three-color fluorescent technology
• Scattered light is measured at multiple angles:
- 0-degree forward light scatter measurement is used for determination of cell volume,
- 90-degree orthogonal light scatter measurement is used for determination of cellular lobularity,
- 7-degree narrow-angle scatter measurement is used to correlate with cellular complexity,
-90-degree depolarized light scatter measurement is used for evaluation of cellular granularity
• The CELL-DYN Sapphire uses MAPSS technology to allow fully automated, random access reticulocyte testing.
• The RBCs are stained with a proprietary membrane-permeable fluorescent dye (CD4K530) that binds stoichiometrically
to nucleic acid and emits green light as the cells, in a sheath-stream, pass through a flow cell by an argon ion laser.
SIEMENS HEALTHCARE DIAGNOSTICS INSTRUMENTATION
• Siemens Healthcare Diagnostics Inc. manufactures the ADVIA 2120 and 2120i provide a complete hemogram and
WBC differential, with fully automated reticulocyte count
• Four independent measurement channels are used in determining the hemogram and differential: RBC/platelet
channel, hemoglobin channel, and peroxidase (PEROX) and basophil lobularity(BASO) channels for WBC and
differential data
• RBCs and platelets are isovolumetrically sphered before entering the flow cell to eliminate optical orientation noise.
• Hemoglobin distribution width (HDW), an analogous to RDW index, is calculated as the SD of the RBC hemoglobin
concentration histogram. The reference interval for HDW is 2.2 to 3.2 g/dL.
• Cell hemoglobin concentration mean (CHCM), analogous to MCHC, is derived from cell-by-cell direct measures of
hemoglobin concentration
• The reticulocyte reagent isovolumetrically spheres the RBCs and stains the reticulocytes with oxazine 750, a nucleic
acid–binding dye.
• Unique reticulocyte indices (MCVr, CHCMr, RDWr, HDWr, CHr, and CHDWr) are provided.
• The CHr or reticulocyte hemoglobin content of each cell is calculated as the product of the cell volume and the cell
hemoglobin concentration
QUALITY ASSURANCE
PRE ANALYTICAL
Sample:
ICSH recommends dipotassium EDTA at concentration of 1.50-2.20mg/ml of
blood
EDTA in excess of 2mg/ml:
• Decrease in PCV by centrifugation
• Increase in MCHC
• Platelets swell and disintegrate leading to artificial high counts
• Leuco-agglutination affecting both neutrophils and lymphocytes
PHYSIOLOGICAL VARIATIONS
A. Pregnancy
increased erythropoeitic activity accompanied with increase in plasma volume & progressive decrease in Hb,
Hct & RBC. Moderate leukocytosis is common
B. Elderly
anemia becomes more common in those older than 70 years
C. Posture
Small but significant alteration in plasma volume with increase in Hb & Hct from lying to sitting especially in
women
D. Diurnal & seasonal variations
E. Altitude
Reduced plasma volume with increase in Hb & Hct
ANALYTICAL
Instrument selection
CALIBRATION
• Calibration, or the process of electronically correcting
an instrument for analytical bias (numerical difference
from the “true” value), may be accomplished by
appropriate use of reference methods, reference
materials, or commercially prepared calibrators
• USE OF CALIBRATOR
Installation
Change of reagent lot
Instrument maintenance
Replacement of major component or repair of instrument
Installation of new software
QC shows a shift or trend when no new change has
occurred
The S-CAL calibrator kit is an acceptable alternative to the whole-blood reference method of
calibration.
S-CAL calibrator is traceable to reference methods and materials
CONTROLS
• Substances used in routine practice for checking the performance of an
analytical process (or instrument).
• They may or may not have a pre-assigned value.
• Even though some control preparations have assigned values, they should not
be used as calibrators or standards because:
• Assigned values are only approximations
• Controls are stable for a limited time only
COULTER 5C Cell control monitors the CBC and differential (Diff) parameters.
Range from high, normal and low values for the common CBC parameters
LATRON primer prepares the tubing and instrument components for the LATRON control.
LATRON control monitors the performance of the volume, conductivity and light scatter
measurements.
Reliable
Accurate
Precise
3. Statistical Error is very less as compared to Manual
method because large number of cells are counted.
K- 1000 18
Pentra 23
LH 700 series 25
5.Less amount of blood is required for analysis.
K – 1000 100µL
LH 700 300µL
8.In case of Fully Automated Analyzers the Laboratory Staff members don't
come in contact with specimens and reagents
[Bio hazardous material] and hence working with auto analyzer is very safe.
9. Automation allows the laboratories to process much larger
workload without a comparable increase in number of staff
members.
Flagging of the Abnormal Specimen
Automatic Wash Cycle to reduce carry over
Automatic double/triple counting for checking precision and accuracy
Automatic checking for the Reagents.
Automatic Threshold limit changes for platelets/ microcytes.
Connectivity with the printer reduces the clerical work.
LIMITATIONS
Instrumental specimen
Instrumental
Biohazardous
Laser
Waste
Hazards
Electrical
Hazards
• A common limitation of impedance methods is an instrument’s inability to distinguish
cells reliably from other particles or cell fragments of the same volume
e.g. Cell fragments may be counted as platelets in specimens from chemotherapy-
treated patients with increased WBC fragility
RBCs containing variant hemoglobins such as Hb S or Hb C are often resistant to lysis,
and the unlysed cells can be falsely counted as nucleated RBCs and interfere in the
hemoglobin reaction
• Suppression of automated data, particularly WBC differential data, may occur when
internal instrument checks fail or cast doubt on the validity of the data
NEW PARAMETERS & CLINICAL UTILITY