This document provides an overview of ultraviolet and visible spectroscopy. It discusses key concepts such as wavelength, frequency, energy, electromagnetic radiation, UV-visible spectrum, electronic transitions, chromophores, auxochromes, absorption, Beer-Lambert law, and instrumentation. The instrumentation section describes the main components of a double beam spectrophotometer used in UV-visible spectroscopy.

Uploaded by

Mounik Rout
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PPTX, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
75 views

Ultra-Violet & Visible Spectros

This document provides an overview of ultraviolet and visible spectroscopy. It discusses key concepts such as wavelength, frequency, energy, electromagnetic radiation, UV-visible spectrum, electronic transitions, chromophores, auxochromes, absorption, Beer-Lambert law, and instrumentation. The instrumentation section describes the main components of a double beam spectrophotometer used in UV-visible spectroscopy.

Uploaded by

Mounik Rout
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PPTX, PDF, TXT or read online on Scribd
You are on page 1/ 23

Seminar on

Ultra-Violet & Visible Spectroscopy

Presented by
Mounik Rout
M.Pharm (Pharmaceutical technology)
Guided by
Dr. Sasmita Kumari Acharjya

ROLAND INSTITUTE OF PHARMACEUTICAL


SCIENCES
Contents
SPECTROSCOPY
WAVE NUMBER , FREQUENCY , ENERGY
EMR WAVELENGTH
UV-VISIBLE SPECTRUM
ELECTRONIC TRANSITION
CHROMOPHORE
AUXOCHROME
ABSORPTION
ABSORPTION LAWS
INSTRUMENTATION
Spectroscopy:-
• Study of absorption of Electromagnetic radiation by molecule, atom, or ion of a sample when it
excited from one energy state to another energy state .
• Electromagnetic radiation are used in spectroscopic method as a source of radiation .

Fig2. Detailed wave structure

• 1.Wavelength(λ) - The distance in between both the crests and troughs is wavelength .
• The unit used for wavelength are - Å , nm , μm .
• 2.Wavenumber(̄) – Number of waves passing through the space of 1cm is called wavenumber .
• ̄= 1/λ
• The unit is Cmˉ¹

• 3.Frequency () - Number of waves passing through a point in duration of 1second is


called frequency of the EMR .
• Frequency = C/λ
• Value of C = 3 x10⁸ cm/sec
• Unit is Hz

Fig4. frequency
• 4.Energy(E) :- Electromagnetic radiation have some energy .
• When wavelength increases , energy of EMR decreases , & when wavelength
decreases energy of EMR increases .
• E = h (where h is planck’s constant=6.626 x 10ˉ²⁷ erg sec)
• E= h x C/λ (=C/λ)
Different EMR and their wavelength
Electromagnetic Radiation Wavelength

Gamma Ray <0.001nm


X-Rays 0.01-10nm
Ultra violet Rays 200-400nm
Visible Rays 400-800nm
Infra red Rays 0.8-200μm
Near IR 0.8-2.5μm
Mid IR 2.5-15μm
Far IR 15-200μm
Microwaves 0.01-1m
Radiowaves 1-10⁷m

Table.1
1.30-
1.20-
1.10-
1.00-
0.90-
0.80-
0.70-
0.60-
0.50-
0.40-
0.30-
0.20-
0.10-
0.00-

200nm 300nm 400nm 500nm 600nm 700nm 800nm

Fig5. UV-Visible spectrum


Electronic Transition :-
• When molecule are excited from one energy state to higher energy state , electronic
transition takes place .

• Types of electrons involved in electronic transition ;


Type 1- Bonded electrons (e.g. )
Type 2- Non bonded electrons (e.g. n)
Types of electronic transition ;
1.
2. n
3.
4. n Decreasing order of energy requirement
• Electronic Transition:-
Lowest Unoccupied Molecular Orbital (LUMO)

Highest Occupied Molecular Orbital (HOMO)

Fig6 . Electronic Transition


• Electronic Transition :-

1. * - When electron in get excited to * then this transition is called *transition .


• All the saturated hydrocarbons (e.g. methane , propane) comes under this .
• This transition require high energy than all 4 transition .
• so that we have to go for vacuum UV region which is below 200nm .
• Below 200nm oxygen starts getting absorb , so under vacuum UV region
it will be free from any else compound , and only the transition will occur .

2. n *- When electron in n get excited to * then the transition is called n * transition .


• This transition occur when any heteroatom is present in the saturated compound .
• This transition requires energy less than * transition .
• Heteroatoms used as solvent in UV-Visible spectroscopy because they aren’t
getting absorb in UV-Visible range .
3. * - When electron from molecular orbital get excited in * orbital , the transition
called * transition .
• This transition requires less energy in comparison of n * transition .
• Any double/triple bonded hydrocarbons are present then * transition will occur .

4. n * - When electron from n molecular orbital get excited in * orbital , the transition
called n * transition .
• It requires lowest energy as comparison of all 4 transition .
• Chromophore- These covalently bonded moieties with any compound which are responsible for
absorption of electromagnetic radiation in the UV-Visible region .
Example of chromophore : Aldehyde , Ethylene, Carbonyl .
1. Chromophore with * : ex - >C=C< , -CC-
2. Chromophore with n * : ex - -N=N-
Whenever chromophore compound will added to any compound the absorption wavelength will
increase .

• Auxochrome – It is defined as any moieties which does not shows any absorption when separated
but when it combines with any chromophore it increases the wavelength towards longer wavelength
by
formation of a new chromophore .
ex: -OH , NH₂ ,OR , etc .
• Absorption :-
• It is a quantitative evaluation .
• Spectroscopy is based upon absorption theory.

Incident Radiation(Iₒ)

Transmitted Radiation (Iₜ)

• Various cases of absorption ;


• Case1 – Intensity of Iₜ = Iₒ
• Case2 – Intensity of Iₜ < Iₒ (shows that absorption occurred)
• Case3 – scattering , refraction , reflection occurs .
• To compensate scattering , Nephelo turbidimetre is used .
• Absorption laws :-

 Lambert’s Law :- when a beam of monochromatic radiation is passed through the absorbing medium
then the decrease in the intensity of radiation will directly proportional to the thickness of solution .

• Equation A Log(Iₒ/Iₜ) L
A L L= Thickness of solution / Thickness of cuvette
A= L = Molar absorptivity co-efficient (unit= l/mol.cm)

 Beer’s law :- When a beam of monochromatic radiation is passed through the solution then the
decrease in the intensity of radiation will be directly proportional to the concentration of the solution .

• Equation A Log(Iₒ/Iₜ) C
A C C = Concentration of the solution
A= C
• Beer Lambert’s Law –
• When a beam of monochromatic radiation is passed through the absorbing medium ,
then the decrease in the intensity of radiation is directly proportional to the thickness
as well as the concentration of the solution .

Equation => A = Log (Iₒ/Iₜ)L.C


A L.C L= Thickness of cuvette
A = L.C = Molecular absorptivity coefficient
• .
Instrumentation of UV-Visible spectroscopy
• .
Chopper

Exit slit Reference


Mirror
Detector 1

Detector 2
Sample
Lens

Amplifier

Gratings

Collimating lens
Radiation Recorder
Source Entrance slit

Monochromator
• Instrumentation
• There are 2 type of spectrophotometer used in UV-Visible spectroscopy .
1. Single beam
2. Double beam
• Double beam spectrophotometer – The instruments used in double beam
spectrophotometer are ;
1. Radiation source
2. Monochromator : Entrance slit
Collimating lens
Gratings
Reflecting lens
Exit slit
3. Chopper 5. Cuvette
4. Mirror 6. Detector
7. Amplifier
8. Recorder
• Radiation source – 1. Hydrogen-Deuterium lamp (ranges from 200-375nm)
2. Tungsten lamp (ranges from 375 -800nm for uv-visible spectroscopy)
• Monochromator – used to convert polychromatic radiation beam to monochromatic
radiation .
• Collimating lens – This converts the non-parallel beams to parallel beam .
• Gratings – This is used in place of prism now-a-days because it contains 1200-1400 groves
per mm .
• Exit slit – through this the required wavelength of beam passes to Chopper .
• Chopper – It equally divide the radiating beam in 2 parts in to the mirror .
• Mirror – it reflects the radiation in to both Cuvette containing reference and sample .
• Cuvette – it is made up of Quartz and glass material .
1. Quartz is used because it have very smooth surface and it don’t absorb the UV
radiation .
2. Glass is used in visible spectroscopy & it has UV absorption capacity .
• Detector – Detector used to detect the wavelength coming through the cuvette & it is of 3 types :
1. Photomultiplier tube – It multiplies the photons rays of the radiation .

c at h
ode
Photon Anod
e

Electron
Dynode

fig. Photomultiplier tube

Amplifier connected from the cathode and anode and amplifies the radiation .
2. Photodiode Array – In this detector there is presence of diodes , which converts
radiation into electronic signal .
• It is mainly used in HPLC , and its range is up to 200-800nm .
• 3. Charged coupled diode – Silicon surface is used to catch photon from the radiation of
UV-Visible source & converts radiation in to digital signal .

• Recorder – It records the amplified wavelength and shows the UV-Visible spectrum .

 Single beam spectrophotometer – This is slightly different as comparison to double beam ,


It directly radiates the incident radiation coming from exit slit of monochromator to the
sample cuvette and then detector detects the radiation to the amplifier . Then the amplified
data recorded on the recorder as spectrum .
• Reference :
Text book of Pharmaceutical Analysis by
Dr. S. Ravi Sankar ,
THANK YOU