Bioanalytical Aspect in BE Study: Yantirta Indra Kurniawan
Bioanalytical Aspect in BE Study: Yantirta Indra Kurniawan
in BE Study
1
Competencies & Compliance in BA/BE
Competent Competent
Clinical Analytical
Resources Resources
3
Bioanalytical Method Validation
• Selectivity
• Lower Limit of Quantification (LLOQ)
• Calibration Curve
• Accuracy, precision, and recovery
• Carry over
• Stability
• Short term
• Long term
• Freeze and Thaw
• Post preparative
• Reinjection (if necessary)
• Matrix effect (for MS detector only)
• Cross over of Plasma anti coagulant (if necessary)
4
Selectivity
• Each of 6 blank should be tested for interference, and selectivity
should be ensured at the lower limit of quantification (LLOQ).
%Interference of Analyte and IS should be < 20% and < 5% to
LLOQ, respectively
• If method of simultaneous determination of multiple analytes,
must demonstrate that analytes do not interfere with each other
Interfering analytes
Equilab Presentation 5
Lower Limit of Quantification (LLOQ)
• The lower limit of quantification (LLOQ) is the lowest
concentration of analyte in a sample which can be
quantified reliably, with acceptable accuracy and
precision
• < 1/20 Cmax (less than 5% of Cmax)
• Normally 5-fold blank noise
• within- run accuracy and between-run accuracy
• within- run precision and between-run precision
• N = 5 or N > 5 for between
• < 20% nominal value
6
Lower Limit of Quantification
(LLOQ)
LLOQ
Noise
7
The calibration curve (1)
• The relationship between instrument response and known
concentrations of the analyte.
• The calibration curve should be prepared in the same
biological matrix with known concentrations of the analyte
8
The calibration curve (2)
criteria not met: reject calibration standard and re-evaluated
standard curve
recommended using freshly prepared curves during validation
continuous &
reproducible
Equilab Presentation 9
Accuracy
1 run/within-accuracy between-accuracy
N=5 3 runs/minimum 2 days
<15%,
except LLQC (<20%)
Precision
1 run/within-precision between-precision
N=5 3 runs/minimum 2 days
<15%,
except LLQC (<20%)
Recovery
Recovery need not be 100%, but recovery of an
analyte and Internal standard (IS) should be
consistent and reproducible
13
Stability
Stock/working solutions:
17
QC Samples in Method Validation
QC Samples :
LLOQ QC
Low QC : 3x LLOQ
Mid QC : 30-50% of ULOQ
High QC : > 75% of LLOQ
For Accuracy and precision Runs :
Four QCs, including LLOQ, low, mid, and high from at least
five replicates in at least three runs
For Other Validation Runs (US-FDA 2018):
Low, Mid, and High QCs in duplicates
18
Bioanalysis
19
Challenges in Bioanalysis
Reference Standard
QC Samples
Audit Trail & Data Integrity
Incurred Sample Reanalysis (ISR)
Reanalysis of Sample
20
Reference Standard
• Reference standards used during the validation and study
sample analysis should be obtained from an authentic and
traceable source :
• Compendial standard
• Commercial available standard
• Characterised standards prepared in-house or by an external
non-commercial organisation
• A certificate of analysis (CoA) is required to ensure purity and
provide information on storage conditions, expiration date and
batch number of the reference standard
• When mass-spectrometry (MS) detection is used in the
bioanalytical method, a stable isotope-labelled IS is
recommended to be used whenever possible
21
QC Samples (1)
• QC Solution should be prepared separately from stock
solution
• Distribution and position of QC Samples in analytical
run should be divided over the run
• At least 2 QC sample levels should fall within the range
of concentrations measured in study samples
Additional QC should be applied if samples only
disperse at one area of calibration range
• The overall (mean) accuracy and precision of the QC
samples of all accepted runs should be calculated at each
concentration level and reported in the analytical report. In
case the overall mean accuracy and precision exceeds
15%, this should lead to additional investigations justifying
this deviation. In the case of bioequivalence trials it may
result in the rejection of the data
22
QC Samples (2)
23
QC Samples (3)
24
Audit Trail & Data Integrity (1)
gap should be
explained
25
Audit Trail & Data Integrity (2)
• Electronic Audit Trail should be provided for every
batch of Analytical Run
26
Audit Trail & Data Integrity (3)
• Manual integration can be captured in Electronic Audit Trail
automatically Reason should be explained
27
Incurred Sample Reanalysis (1)
Used to measure the reproducibility of study
sample concentrations
The total number of ISR samples should be
depending on the total number of samples
(EMA : 10% for total samples <1000 or 5% for
total samples > 1000, US-FDA : The first 1000
samples = 10%, 5% of the remaining samples)
Samples should cover the entire PK profile
(Cmax concentrations and elimination
concentration)
Incurred Sample Reanalysis (2)
• ISR should be performed in separate runs at
different days (after all study samples
completed or throughout the conduct of study)
• Two-thirds (67%) of the repeated sample results
should be within 20% of the mean of 2 values
29
Reanalysis of Sample
• Rejection of an analytical run because the run did not fulfil the
acceptance criteria with regard to accuracy of the calibration
standards and/or the QC samples
• Internal standard response significantly different from the
response for the calibration standard and QC samples, if
such criteria have been pre-defined in a SOP, e.g. -50% to
+50%
• Improper sample injection or malfunction of equipment,
• The obtained concentration is above the ULOQ or below the
run’s LLOQ, in runs where the lowest standard sample has been
rejected from a calibration curve, resulting in a higher LLOQ
compared with other runs
• Identification of quantifiable analyte levels in pre-dose samples
or placebo sample
• Poor chromatography
30
Thank you
yantirta.kurniawan@
equilab-int.com
31