Optimization of Biotinidase Assay: Dr. Faryal Husnain Resident Clinical Chemistry
Optimization of Biotinidase Assay: Dr. Faryal Husnain Resident Clinical Chemistry
Assay
Dr. Faryal Husnain
Resident Clinical Chemistry
Introduction
• Biotinidase deficiency (OMIM #253260) also called Multiple
Carboxylase Deficiency is a rare autosomal recessive disorder of
biotin metabolism.
• Biotinidase is required for the endogenous recycling and release
of biotin from dietary protein.
• Biotin (vitamin H or B8) is an essential co-factor for carboxylase
enzymes involved in fatty-acid synthesis, amino-acid catabolism
and gluconeogenesis.
• Biotinidase deficiency can cause insufficient biotin to be made
available for the maintenance of these processes; which can
lead to life-threatening metabolic derangements eliciting
organic aciduria and neurological derangements.
Biocytin(N-
Biotinyl-L-
Lysine)
Biotinidase Deficiency
• Mayo Clinic reference range=3.5 U/L to 13.8 U/L
• Partial deficiencies and carriers may occur at the low
end of the reference range.
• Profound biotinidase deficiency results when the
activity of biotinidase is reduced to less than 10
percent of normal.
• Partial biotinidase deficiency occurs when biotinidase
activity is reduced to between 10 percent and 30
percent of normal.
• 50% of normal activity is seen in heterozygotes.
Category Mean Biotinidase activity ± SD
(nmol/min/ml serum) (n)
Normal individuals 7.57 ± 1.41 (100)
Obligate heterozygotes 3.49 ± 0.72 (21)
Affected individuals (symptomatic) 0.12 ± 0.18 (23)
Affected individuals (newborn screening) 0.19 ± 0.16 (41)
Individuals with partial biotinidase
1.47 ± 0.41 (23)
deficiency
Biotinidase Deficiency
• Patient's urinary organic acids (GC/MS) analysis shows an excretion of 3-
hydroxyisovaleric acid and 3-hydroxypropionic acid. Their plasma biotin
concentration is decreased, however, it can be normal in early forms
• Ketolactic acidosis and mild hyperammonemia may also be present.
Patients with partial BTD (10% to 30% of mean normal BTD activity) may
be asymptomatic.
• Urinary excretion of 3-hydroxyisovaleric acid can be used for monitoring
response to treatment.
• Other metabolic abnormalities are more variable and may include elevated
excretion of 3-hydroxyisovaleric, lactic and 3-hydroxypropionic acids, and
3-methylcrotonylglycine by urine organic acid analysis, as well as mildly
elevated 3-hydroxyisovalerylcarnitine (C5–OH) by plasma acylcarnitine
analysis.30 These metabolic abnormalities are variable
Objective
• A quantitative colourimetric biotinidase assay
will be validated using a quantitative
validation method and it’s reference range
will be calculated in adults(older than 18
years) and children(younger than 18 years)
SUBJECTS AND METHODS:
•Compare the results from the two methods for the same
samples.it is recommended that a minimum 20 samples are
compared, however, CLSI consider 40 samples to be for a good
method comparison, increasing numbers improve the quality up
to a point, but a good distribution of data is much more important
than a large number.
•Enter the data in Alternate Method Comparison Module of EP
(statistical software) the statistical tool for alternate method
comparison is a linear regression.
•A specimen passes accuracy if all measured results fall within the
target range.
AMR
• Analytical Measurement Range
• Range of analyte values that a method can directly measure
w/o modification (no dilutions, concentrations, other
pretreatments that are not part of the usual assay process)
• When initially introducing a new method, it is necessary to
verify the AMR independently from the calibration process
• For assays that have that have a minimum three calibrators
with low, high & mid-point values covering most of the AMR
and where calibration confirmation also covers most of the
AMR confirmation of linearity is not required.
AMR
• Material for linearity checking needs to have a level
of the analyte at the upper limit of the AMR. Choice
of material may comprise the following:
• Calibrators from a different lot number.
• Material manufactured specially for linearity
checking. Or CAP linearity survey or validation
materials can be used to verify performance claims.
Compare to peer group statistics appearing in the
survey.
Precision
• Measure the same analyte for 20 consecutive days, do not
include the outliers in your calculation (Outliers are results
outside the QC range)
OR
• Twice a day in duplicate for 10 days.
OR
• Once a day for 20 days
OR
• Calculation of Precision assess using EP evaluator software
Reference Range