High Performance Liquid
Chromatography
Kromatografi Cair Kinerja Tinggi
1
HPLC
C
P I
FD
1. Fase gerak dipompa masuk ke tabung
stainless steel yg berisi partikel fase
E diam berdiameter 3-10 m
L
2. Dengan injektor Rheodyne larutan
baku/sampel disuntikan ke dalam
aliran fase gerak
3. Pemisah komponen terjadi ditabung fase diam
4. Komponen dideteksi oleh detektor
5. Data diproses oleh komputer
2
Applications
• The combination of high-pressure liquid chromatography (HPLC) with
monitoring by UV/ visible detection provides an accurate, precise and
robust method for quantitative analysis of pharmaceutical products
and is the industry standard method for this purpose
• Monitoring of the stability of pure drug substances and in drugs in
formulations with quantitation of any degradation products.
• Measurement of drugs and their metabolites in biological fluids.
• Determination of partition coefficients and pKa values of drugs and of
drug protein binding.
3
Strengths
• Easily controlled and precise sample introduction ensures quantitative
precision.
• HPLC is the chromatographic technique which has seen the most
intensive development in recent years leading to improved, columns,
detectors and software control
• The variety of columns and detectors means that the selectivity of the
method can be readily adjusted.
• Compared to GC there is less risk of sample degradation because
heating is not required in the chromatographic process
• Readily automated.
4
Limitations
• There is still a requirement for reliable and
inexpensive detectors which can monitor
compounds that lack a chromophore
• Drugs have to be extracted from their
formulations prior to analysis
• Large amounts of organic solvent waste is
generated, which is expensive to dispose of
5
Modes of HPLC
• Normal Phase mode
• Reverse Phase mode
• Reverse Phase Ion Pairing mode
• Ion Exchange mode
• SEC mode ( GPC / GFC )
• Chiral separation mode
6
Reverse Phase HPLC Columns
• C18 (ODS) type
• C8 (octyl) type
• C4 (butyl) type Non-polar property
• Phenyl type -Si-C18H35
• TMS type
• Cyano type Si
7
Structural Factors Which Govern Rate of
Elution 0f Compounds from HPLC Columns
• Elution of neutral compounds
– For a neutral compound it is the balance between its polarity and Iipophilicity
which will determine the time it takes for it to elute from an HPLC column; the
pH of the mobile phase does not play apart. In the case of a reverse-phase
column, the more lipophilic a compound is the more it will be retained
►
• Ionizable compounds
– Control of elution rate of ionisable compounds by adjustment of pH of mobile
phase. Untuk asam K’app = K’/(1+10pH-pKa) untuk basa K’app = K’/(1+10pKa-pH) ►
– Mobile-phase conditions may be selected in straight-phase chromatography
where the ionisation of the analytes is suppressed, and basic compounds are run
in a basic mobile phase and acidic compounds are run with an acidic mobile
phase
– The pH of the mobile phase can only be set within the range of ca 2-8.5 pH
units because of the tendency for extremes of pH to dissolve silica gel and
break the bonds between silane-coating agents and the silica gel support
8
Hydrophobicity
• If the sample has
– CH3CH2CH2--- : Carbon chain Hydrophobicity
– : Aromatic group becomes strong.
• If the sample has
-COOH : Carboxyl group
-NH2 : Amino group Hydrophobicity
-OH : Hydroxyl becomes weak.
group
9
Retention Time and Hydrophobicity
OH
C18 (ODS)
Weak
Strong
OH
10
Increase of solvent polarity
20 % 30 % 40% / H2O
1 : p-Hydoxymethylbenzoate
2 : p-Hydoxyethylbenzoate Solvent : MeOH
3 : p-Hydoxypropylbenzoate
4 : p-Hydoxybutylbenzoate 11
Effect of stationary phase
C8
Medium
C18 (ODS)
sample
Strong C4
sample
Weak
sample
12
Effect of stationary phase
Analytical Conditions
ODS C8 TMS Column : Shim-pack CLC-ODS
Mobile phase : MeOH : H2O = 7 :3
Flow rate : 1.0 mL/min
Temperature : 40 C
Injection volume : 10 uL
Detection : UV-254 nm
Peaks
1. Methyl benzoate
2. Ethyl benzoate
3. n-Propyl benzoate
4. n- Butyl benzoate
13
Reverse Phase
Ion-Pair Chromatography
Ion-Pair Reagent
14
Ion Paring
Important Considerations
• Type of Ion-Pair reagents
• Concentration of Ion-Pair reagents
• pH of solvent
R-COOH RCOO- + H+
(pKa=4.5)
R-NH2 + H+ R-NH3+
(pKa=6.0)
15
Type of Ion-Pair Reagents
Hexane Sulfonate Pentane Sulfonate
Mobile Phase: H2O/MeOH
1:1,with 0.005M ion pairing
reagent and 1% HOAc
1 Maleic Acid
2 Phenylephrine
3 Phenylpropanolamine
4 Naphazoline
5 Phenacetin
6 Pyrilamine
16
Concentration of
Ion-Pair Reagents
17
Causes of Tailing Peaks
• Built-up of garbage on the column inlet
• Extra column effects (dead volume)
• Sample Overload
• Incorrect solvents for the sample
• Secondary retention effects
– Silanol group
– Residual heavy metal
18
Dead Volume
• Dead volume will cause a broad peak.
– Sample Injector Connection
– Column Connection
– Detector Connection
Good
Dead Volume
19
Incorrect solvent for sample
• Better do not select a high soluble solvent as a sample
solvent.
Methanol as a sample solvent Ethanol as a sample solvent
20 uL Caffeine 20 uL Caffeine
Ethanol as a sample solvent
Better inject small
10 uL Caffeine amount of sample.
20
Solubility effect
Low soluble solvent High soluble solvent
21
Secondary retention effects
• Silanol Group
– Even modified silica gel (e.g. ODS, C8), residual
silanol group are still remained on the surface
area.
– Silanol group will strongly absorb amine
compounds, therefore tailing will be happened.
C18
OH
Silanol group
silica core C18
O- negative charge
C18 22
C18
End capping
• To suppress the silanol group effect, an end
capping type of column will be selected.
C18 C18
OH TMS treatment O-TMS
silica core C18 silica core C18
OH TMS : trimethylsilyl group
O-TMS
C18 C18
C18 C18
[Non-End capping type] [End capping type]
23
Secondary retention effects
• Residual heavy metal
– Normal silica gel has heavy metal as an impurity.
This heavy metal will have an interaction with
chelate compounds, therefore tailing will be
happened.
C18
M+ O-TMS High pure type of
silica core C18 O
+ O-TMS silica gel are available.
M C18 O
C18
24
HPLC system
• Isocratic elution system
– Single solvent of constant composition
• Gradient elution system
– Multi solvents of changeable composition
• High pressure gradient system
• Low pressure gradient system
25
Isocratic Elution System
pump injector oven
detector
column
Single Solvent
26
Gradient Elution System
A
column
injector detector
pump oven
B concentration
B
pump
Time
27
Isocratic Elution Mode
MeOH / H2O = 6 / 4
Long Time Analysis
Bad Separation
MeOH / H2O = 8 / 2
( column : ODS type ) 28
Gradient Elution Mode
MeOH concentration
95%
30%
29
Calibration method
• External calibration method
• Internal calibration method
• Standard additive method
• Correction Normalization method
30
External Standard Calibration
Preparation of Standards
Target Compounds
Dilution Dilution Dilution Dilution
31
External Standard Calibration
200
180
Concentration
160
140
120
100
80
60
40
20
0
0 1000 2000 3000 4000
Peak Area
32
External Standard Calibration
Calculation of Results
Y = bX + a
[Concentration]
b : SLOPE
125 ppm a : Y intercept
2500
2500
[Peak Area]
33
Internal Standard Calibration
Preparation of Standards
Internal
Target Compounds Standard
Dilution Dilution Dilution Dilution
34
Internal Standard Calibration
Analysis of Vanillin
35
Internal Standard Calibration
[Terget Conc. / IS Conc.]
Analysis of Vanillin
4
3.5
3
2.5
2
1.5
1
0.5
0
0 5 10 15
[Tergat Area / IS Area]
36
Internal Standard Calibration
Calculation of Results
[Target Conc. / IS Conc.]
Y = bX + a
b : SLOP
1.67 a : Y intercept
T 2500
5.0
500
[Target Area / IS Area] IS
Y = Target Conc. / IS Conc.
1.67 = Target Conc./ 100 ppm
Target Conc. = 167 ppm
37
Advantage of external standard
calibration method
• Only the target compound separation can be
focused.
Target Target
38
Disadvantage of external standard
calibration method
• Injection error will directly influence the
quantitative result.
10 uL injection 11 uL injection
100 ppm 110 ppm
39
Advantage of internal standard
calibration method
• Injection error can be eliminated.
10 uL injection 11 uL injection
2200
2000 1100
1000 T
IS
IS T
2000 / 1000 = 2 2200 / 1100 = 2
40
Advantage of internal standard
calibration method
• Recovery in the pretreatment process can be
estimated. addition of IS (100 ppm)
standard IS (100 ppm) to actual sample
IS T
IS T
1000 950
Recovery = (950/1000)x100 = 95%
41
Disadvantage of internal standard
calibration method
• Separation is slightly difficult.
IS T IS T
T
IS
42
Disadvantage of internal standard
calibration method
• It is difficult to look for the IS compound.
– The chemical structure of IS compound is
similar with one of target compound.
– IS sample is not existent in the actual sample.
43
Calibration Method
• External standard calibration
– Separation is not difficult
– Injection error will directly influence the
quantitative result
• Internal standard calibration
– Injection error can be eliminated
– Recovery in the pretreatment procedure can be
estimated
– Separation is slightly difficult
– Difficult to look for the IS compound
44
Standard additive
calibration method
Original
Sample T
Target
T
45
Standard additive
calibration method
T x104
Peak Area 100 ppm=10x104
17 ??? ppm= 7x104
T 12
7
??=70 ppm
T
-70 0 50 100 ppm
Added amount 46