Radio Exam 2 Notes
Radio Exam 2 Notes
Dr. Osher
Basics of Attenuation
4 Factors that Affect Attenuation
Kilovoltage
Density
Atomic Number (Z)
Electrons per Gram of Tissue
F-Factor Conversion
F-Factor converts exposure (X) and absorbed dose
(D)
Roentgen to Rad conversion factor
D= f x X
At diagnostic X-ray energies, f-Factor for soft tissues is
close to 1.0
f-Factor for bone is 4 at Low kvP
F-factor for bone is 1 at High kVp
Dose Equivalent
Dose Equivalent (H)
Attempts to quantify biologic damage from deposition of
radiation in the tissues
Dose Equivalent=absorbed dose x quality factor
H=D x QF
X-Rays, Gamma Rays, Electrons, Beta particles have a Quality
Factor= 1.0
Neutrons and Protons= 5.0
Alpha Particles=10.0
Dose Equivalent
Dose Equivalent
Units
Sievert (Sv)SI system
Rem (radiation equivalent man)non SI
1 Sv= 100 Rem
1 Rem= 10 mSV
No Attenuation=Black
On the film, the Emulsion Layer is the active layerdouble-sided film
Lots of RadiationBlack
Little Radiation--Clear
Matter Interactions
PenetratePhotons pass through unaffected
No Deposition of Energy
Exposure to Processing
Processingtransforms the Latent Image into
Manifest Image
Film Development
Converts Silver Halide Crystals with latent image centers into flecks of
silver
Over time, all of the underdeveloped silver halide will be reduced to silver
Time and temperature are critical
Ag+1 must be removed from Gelatin via Fixation prior to viewing the film in light
Fixer=Acetic Acid
FixerAcidic
Provides acid media for hardener and helps to neutralize developer
pH=4.0
Fixing Bath
Removes all unexposed silver from emulsionclears unexposed film
Aluminum chloride hardens the emulsion
Total fix time=2x clearing time
Once cleared, we can view the film in light
Processing
Temperatures
Automatic Processing occurs at temperatures much
higher than manual processing
Manual processing= 68F optimum
Automatic processing= 80-84F
If the temperature is too high, there is Film Fog and
Overdevelopment
Film Fog=Unexposed crystals that turn black by processes
other than X-Ray exposure process
Increase Film fog=Decreases contrast
On Film:
Brown stains=Chromatic Stains
Indicates a washing problemH2S formation
Lines=Pi Lines
Film Processing
Problems
Chromatic Stainsyellow-brown from residual fix on
film
Inadequate wash
H2S Production
Black FilmOverdevelopment
Clear FilmNever Exposed
Development
Catalytic Reduction of Sensitized Silver
Basic pH
Manual5 mins/Automatic22 seconds
Stop Bath
Ends Development
pH7.0
Manual30 seconds/AutomaticN/A
Fixation
Clears unexposed halide and hardens gelatin
Acidic pH
Manual5-15 mins/ Automatic22 seconds
Drying
Removal of water
Manual30 mins/ Automatic26 sec
Sooo
Manual Processing time1 hour
Automatic Processing time90 seconds
Shelf Life
Each film box has an expiration dateTypically not greater than 1 year
Aging of film results in
Loss of speed and contrast
Film age fog
Should store boxes upright to minimize film warping and sticking together
Image ReceptorsFilmScreens
Intensifying Screens
The ability of Crystals of certain inorganic salts (phosphors)
to emit light when excited by X-raysFlourescence (glow)
Emitting Light=Flourescence
Primary Determinants
of Film Density
Primary Determinants of Film Density
Milliamperage x Time (mAs)
kVpprofound effect: to the 4th power
Inverse square Lawtarget film distance
Screen-Film System
Intensifying Screens
The primary function of Intensifying Screens is to Decrease Dose
Reduces patient dosage 20-50X
Shorter Exposure Times
Facilitate radiology of thick body parts which otherwise require high
exposures
Decreases risk of motion blurring
Screen-film Advantages
Exposure without screens is the sharpest..but
Long exposures are unsafe (ALARA)
Increases chance of patient motion
Intensifying Screens
4 basic layers
Base made of plastic or cardboard
Reflective titanium dioxide or absorptive layerchanges the speed
characteristics of the screen
Affects amount of light that hits the screen, which affects the
development of the film
Intensifying Screens
Absorption efficiencypercentage of X-ray Photons
absorbed by the screen
Increase Absorption Efficiency=Increase the thickness
of the screenFluorescent layerphosphor layer
Crystalline Calcium
TungstateOriginal Screen
Phosphor
CaWO4 (Calcium Tungstate) produces light in the
blue spectral region
This can be used with standard X-ray film
CaWO4 is quick to glow and has minimum afterglow
3-5% conversion efficiency
Mechanism of absorption is almost entirely via
photoelectric reactions
Calcium Tungstate
Intensifying Screens
Calcium Tungstate Screen speed increases with
Doubling the number of screensbecause film is double-sided
Increasing the thickness of the Phosphor Layer
More photons are absorbedemitting more light per unit time
**All of these methods increase absorption efficiency of the Calcium Tungstate screen.
Increase screen light diffusion= Decrease image sharpness
Greater speed=faster, but at expense of image sharpness
Speed of the Calcium Tungstate screen is INVERSELY related to its ability to record detail
Higher Speed= LESS detail
Light spread within the screens reduces image sharpnesslateral spreading of light
Intensifying Screens
The Speed of a Calcium Tungstate screen is
INVERSELY proportional to its ability to record detail
Higher speed= Less Detail
Less speed= More Detail
Spectral Matching
Calcium Tungstate screens emit blue and blue-violet
light
Rare earth screens emit UV, blue, green, and red
All silver halide films respond to violet and blue
lightbut not to green, yellow, or red
Radiographic films are either Blue-Sensitive or GreenSensitive
Blue-sensitive requires AMBER safelights
Green-Sensitive requires RED safelight
Film Types
2-Basic Film Types
Screen FilmsThin Emulsion
Have higher contrast
Short scale
Low Latitude
Digital X-Ray
Radiographic Basics
A digital image is an image that has been converted into numerical
values for transmission or processing
A detector must be used initially to acquire the image information
Scan an area line by line
Array detectioninfo received from entire area at once
** Painting by numbers**
Digital Radiography
Like digital cameras
Images can be taken, immediately examined,
deleted, corrected, and croppedcan be sent to a
network of computers
With digital imaging, the main change is the image
receptorit is no longer film based
Has the potential to increase patient dosepresents a
problem
Digital=No film: increased patient dose
Digital Radiography
Digital Radiography replaces film with a Reusable Detector
Cassette-Based Systems
Use photostimulable storage phosphor (PSP) imaging plate (IP)
inside cassette
Once exposed, plate is taken into a reader
Is Indirect Digital (Computed Radiography)
Cassette-Less Systems
Detector and Reader is apart of same unit
Direct digital Radiography (TFT)no light produced
Computed Radiology
Many similarities exist between Computed Radiology
and Screen-Film receptor imaging
Both use image receptors that is an X-ray sensitive
plate in a protective cassette
Both use scintillator plates that emit light with X-ray
exposure
Either can be used interchangebly with any x-ray
imaging system
Both carry a Latent Image that must be made visible
via processing
Digital Computed
Radiology
The response to x-rays in screen film systems is the release of
light by the intensifying screen
The response to X-ray interaction in Digital Computed
Radiography PSPs is both
Immediate Release of Light
TRAPPING of electrons in a Higher Energy metastable state
Barium + Europium is used in Digital Computed Radiography (CR)
Flouresce during exposure
Screens store and trap energy
Computed Radiography
CR Plate
Cassette contains a photostimulable phosphor plate (PSP)
that can be used to store and release image formation in a
usable way
Barium Fluorohalide and Europium Activators
Photostimulable Phosphor
Image Aquisition
PSP stores absorbed X-ray energy in crystal
structuretraps
Energy deposited causes local electrons to be elevated
from a ground state energy level to a stable trap
F-center and Europum +2 and +3
Electronic latent image
Main advantages of
Digital Radiography
Linearity of response of digital IRs over exposure ranges that are orders of
magnitude greater than latitude of any film.
Separation of image acquisition from image processing and image display
Windowing
Edge enhancement
Noise reduction
Computers are used to post process image
CR Tolerance of
Overexposure
In conventional radiography, excessive exposure yields black film
this leads to decreased contrast
Extended CR density range means that imaging plate does not have
a Dmax curvehas a straight line
CR imaging plate continues to record the exposure way beyond the
limits of film
Spatial Resolution
Spatial resolution is the ability of a system to resolve
a small high contrast object
Spatial frequency is expressed in line pair per mm
(lp/mm)
More paired lines=sharper image
Advantages of CR/DR
vs. Film
Range of Exposures leading to acceptable Optical Densities is limited with film
Exposures below toe are clear; above shoulder are opaque
Large dynamic range eliminates retakes for over and under exposure
Windowing
Post-processing abilitybiomechanicals, magnification, sharpening
Long-term cost savings
Space-savingno darkroom, file storage
Digital domainPACS capable and printable
High DQE FPD systems are safer
Higher DQE= better image quality per dose
Disadvantages of CR/DR
vs. Film
Potential for safety abuse
High initial costhigh repair costs
Image degrades rapidly (25% in 8hrs) via Heat/Cosmic
Radiation
CR has a greater sensitivity to scatter
Spatial resolution is less than film
Loss is exponentialthere is nothing after 3 days
DIACOM
Digital Imaging and Communication in Medicine
Allows equipment made by different manufacturers
to communicate in the digital environment
Protocol that allows the exchanging and storage of
medical data, both images, and text
Radiographic Quality
Radiographic QualityThe fidelity which the anatomic structure being
examined is imaged on the radiograph
Resolutionthe ability to image 2 separate objects as being distinct
Characteristic Film
Curve
A Characteristic Film Curve plots relation between
exposure and OD
Y-axis=OD; X-axis=log exposure
Every 0.3 increase doubles the exposure (because the log
of 2 is 0.3)
Base + Fog= OD without exposure
Typical range= 0.1-0.2 OD units
Film Contrast
Film contrast is an inherent property of a film determined by the
slope of the straight line portion of the Characteristic curve
Increased slope= Increased inherent film contrast
Decreased slope= Decreased inherent film contrast
Characteristic Film
Curve
Film GammaMaximum slope of curve
Average Gradientmean slope between 2 specified
Optical Densities
Usually 2.0 and 0.25 above the base + fog density
Windowing
Windowing is a digital concept
An important feature of digital imaging is the
computers ability to mathematically bring density
differences into the visual range
Because range of stored densities is much wider than
the visual range, any digital image is only a small part
of the total data obtained by the computer
Each image on the monitor is only a window of the
total range of data
Windowing=gray scale expansion or compression
Windowing
Window Level (density)the value of each pixel is changed by addition or
subtraction
Increase window level=Increase image density
Window level controls density
Extremely wide window width requires the computer to ignore fine contrast
differences in order to display the entire range of data
Extremely narrow window width requires the computer to ignore a large amount of
data outside the chosen range
Window width controls the visibility of detail
Radiographic Contrast
Film Contrast
Subject Contrast
Fog and Scatter
Subject Contrast
Anatomic Part Thickness
Differing thickness of same material will attenuate Xray beam differentlyBasis of Contrast
Thicker parts=increased attenuation
Thinner parts= decreased attenuation
Magnitude of difference is proportional to degree of
contrast
Subject Contrast
Density Differences
Greater Density of a tissue results in greater beam
attenuation
Water and IceWater is 9% more dense than ice
Atomic Number difference
Increase Atomic Number=Increase in attenuation
Higher atomic numbered tissues attenuate more X-rays
Results from photoelectric attenuating absorptions in
patient
True at Lower kVps
Scatter Radiation
Scattered Radiation that reaches film produces
unwanted density
Increased Scatter Radiation (Compton Scatter)
results in overall darkening of the film areas
Decreases Contrast and Image Detail
Image Detail
Image Detailability of the film to record each point
in the object as a point on the film
With cassette-screen combos, this point for point
reproduction is never perfect
Edge Unsharpness
Randoms
Source Image Distance= Distance between focal spot and
image
Object Image Distance= Distance between object and Image
Umbra=Image you see
Penumbra= Edge sharpness
Increase Penumbra= Increase Contrast and Decrease edge
Sharpness
Is larger toward the Cathode
Object Distortion
Object Thickness
Thick objects are distorted more than thin objects
Object Position
If object plane and image plane are not parallel,
distortion occurs
Object Shape
Object Motion
Movement of the patient can cause image blurring
Short exposure times minimize chance of motion blurring
Utilize film screenspreferably double screens
Types of Mottle
Radiographic Mottle
Screen Mottle
Quantum Mottle
Film Grain
Radiographic Noise
Factors
Film Graininessnot significant
Structure MottleCaused by defects in screen
phosphor layers
Clumping of phosphor or coating variations
Not significant
Quantum Mottle
Occurs with very short exposure times
Result of statistical fluctuation in number of photons
per unit area of the X-ray beam
Percent of fluctuation per square mm increases as the
average number becomes smaller
Grass Seed Analogy
Radiographic Technique
Radiographic Technique
Combination of Settings selected on the X-ray machine control panel that
produce a high-quality radiographic image
Factors influencing Technique
Patient Factors
Image-Quality Factors
Optical Density
Contrast
Image detail and distortion
Exposure-Technique Factors
mAs, kVp, SID are principle factors
Thickness of Part
Body CompositionPrimarily soft tissue, bone, or both
Patient HabitusSthenic, hypo or hyperstenic, asthenic
PathologyRadiolucent or opaque
Controlling Optical
Density
Optical DensityDegree of blackening on the finished
radiograph
mAsDirect proportionality
mAs=Optical Density
SID=via inverse square law
Problematic when varied
Usually fixed at 90cm for mobile, 100cmtable,
chest studies
180 cm
Principal Radiographic
Image Quality Factors
Optical Density
Controlled bymAs
Influenced by
kVp
Distance
Thickness/mass/density
Development time/temp
Image Receptor
Collimation
Grid Ratio
Principal Radiographic
Image Quality Factors
Contrast
Controlled by kVp
Influenced by
mAs
Film factors
Development time/ temp
Image Receptor
Collimation
Grid Ratio
Principal Radiographic
Image Quality Factors
Detail
Controlled by--Focal Spot size
Influenced by
SID
OID
Motion
All factors related to density and contrast
Distortion
Controlled byPatient Positioning
Influenced by
Alignment of tube
Anatomic part
Image receptor
Basics of Subject
Contrast
ChestHigh Subject Contrast
Lunglow mass density
Bonehigh mass density
Mediastinal Structuresintermediate
Can use High kVP and Low mAs combinations
Extremities
Intermediate to high subject contrastbut kVp must be
lowered due to part thickness
Exposure Technique
Pearls
The primary control of Optical Density is mAs
Changes are proportional
Altering Radiographic
Contrast
Given an acceptable Optical Density level, contrast
adjustments are made by varying kVp15% Rule
Variable-kVp Technique
Chart
Uses fixed mAs and variable kVp
kVp generally varies with thickness of anatomy by 2 kVp/cm
Generally uses lower kVp, therefore short scale contrast and higher
patient dose
Fixed kVp technique chart is best
For each anatomic part there is an optimum kVp
Principles of Radiation
Safety
The Overarching Principle
All unnecessary exposure must be avoided and all
absorbed doses be kept As low as is Reasonably
Achievable
ALARA PrincipleICRP 9 and NCRP 22
Irradiation of Macromolecular
Solutions
Main Chain Scission
Produces many smaller molecules
Viscosity decreases
Cross Linking
Side Chains become sticky
Viscosity increases
Point Lesions
Disruption of single chemical bonds
No grossly apparent changes
Considered to be primary mechanism of cellular damage from low
doses of radiation
Accounts for late effects of radiation
RadiationMajor
Effects on DNA
DNA is more sensitive than RNA or cellular proteins
DNA is not as abundant
Radiolysis of Water
80% of the body is water
Free Radical Formation
Very unstableradicals can disrupt molecular bonds
Factors Affecting
Radiosensitivity
Factors Affecting Radiosensitivity
Oxygen Effect
Age
Gender
Male susceptibility > Female
Factors Affecting
Radiosensitivity
Oxygen Effect
Biologic Tissue is more sensitive under Aerobic
conditions
OER= anoxic dose/aerobic dose
OER is LET dependent
Greatest for low-LET, max @3.0
About 1.0 for high-LET radiations
Factors Affecting
Radiosensitivity
Law of Bergonie and Tribondeau
Stem Cells are Radiosensitive
Greater Maturity of cell= Increased Resistance
Decrease age of tissues/organs= Increased
Radiosensitivity
Increased Metabolic Activity= Increased
Radiosensitivity
Increased Cell proliferation rate= Increased
Radiosensitivity
Increased Tissue Growth Rate= Increased
Radiosensitivity
Factors Affecting
Radiosensitivity
Cell Sensitivity to Radiation
Very Sensitive
Lymphocytes
Erythroblasts
Myeloblasts
Spermatogonia/Oocytes
Sensitive
Epithelial Cells
Endothelial Cells
Least Sensitive
Bone
Nerve
Brain
Muscle
Factors Affecting
Radiosensitivity
Chemical Effects
RadiosensitizersHalogenated Pyrimidines
Methotrexate
Gemcitibine
Hydroxyurea
Vitamin K
Radioprotectors
Contain SH group
Compete with Oxygen for free radical binding
Cysteine
Cysteamine
Protective effect
Radiation Effects
4 Types of Serious Radiation Effects
Acute (Non-Stochastic) Effects
Seen after high, brief exposures
Whole body dose of a few Sieverts may lead to death in
several months
Hematologic Depression
Cytogenic Damage
4 stages
Prodrome (N/V)
Latent
Manifest syndrome
Recovery or Death
Stage 3 manifests as
Hematopoetic
GI
Cerebrovascular syndromes
Cytogenic Damage
Damage is usually manifested during the next cellular mitosis
Non-Threshold dose response
2 Typessingle hit and multi hit aberrations
Damage difficult to identify with low doses: less than 5 rads
A hit usually disrupts many molecular bonds and produces visible
chromosomal damage
Almost every type of chromosomal aberration can be radiation induced
Late Effects of
Radiation
Late Risk Estimates
Absolute Risk
Slope of linear, dose response
Cases/106 persons/rad/yr
Excess Risk
Observed casesexpected cases
Relative Risk
Observed/Expected (unexposed) cases
Hematologic Depression
Whole Body25 rads
Skin Erythema
Small Field300 rads
Epilation
Small Field300 rads
Chromosome Aberration
Whole Body5 rads
Gonadal Dysfunction
Local Tissue10 Rads
Genetic Mutations
Solid Tumors
3:1 solid tumor/leukemia ratio
May be higher than this
Linear, non-threshold dose response curve dynamics
Average Latency=20 years or more
4-7 years= 3X increase of getting a solid tumor
Genetic Mutations
Linear non-threshold curves
No-dose rate effect
Genetic effects cumulative
Major Conclusions of
Megamouse Experiments
The doubling dose for genetic mutations is subtantially higher in mice as
compared to Drosophila
Mammalian systems are more sensitive to radiation
Doubling doseradiation needed to double that of spontaneous or natural
population rate
Genetic Risks
Most mutations are harmful
Any dose of radiation entails some genetic risk
# of mutations produced is proportional to dose
Linear extrapolation from high dose=valid estimate of low dose
effects
Risk estimates of mice are not too far off from humans
Exposure Reduction
Timeminimize
Distancemaximize
Shieldingemploy
Primary Beam
Directly comes out of the x-ray tubewhat hits the
patient
Exposure Time
Exposure Time
Exposure= Exposure rate x Time
Factors Affecting
Scatter
Factors Affecting Scatter
Thickness of Body Part
Orientation of Body part and tube
Field Sizeirradiated voxel2
kvP or dose rate
Use of Grid
Minimizing Scatter
Beam LimitationCollimation
Cones and cylinders
Beam Restricting
Devices
Aperture Diaphragm
Film size and SID are constant
Proper design1cm smaller on all sides of the
radiograph
Collimation
Purpose of Collimation
Lowers patient dose by restricting the volume of
irradiated tissues
Improves image contrast by decreasing scatter
Increased Collimation
Patient dose decreases
Scatter radiation decreases
Radiographic contrast increases
Radiographic density decreases
Increased Field Size= **Patient dose increases
Shielding
Barriers, aprons, and the patient
Structural Shieldingtype of radiation must be considered
Primary beam radiationis the actual beam directed at you?
Secondary radiation2 types
Scatter Radiationpatient major source
Leakage radiationemanates from the tube head in all directions
Structural Barriers
Barrier thickness is generally determined for the highest
energy level employed by the x-ray unit
Shielding
Structural Barriers
Primary1/16 lead typical
Secondary1/32 lead typical
Building materials such as concrete, gypsium, or lead
acrylic also used
Minimizing Patient
Radiation Exposure
ALARA principle
All unnecessary exposure must be avoided and all absorbed doses must be kept as low as is
reasonably achievable
Identify patients who are pregnant and possibly pregnantX-ray only if necessary
Filtration
Protecting the
Operator/Staff
Understand what You are doing
Personnel Monitoringmonitor any person likely to receive in one year a dose in excess of 10% of their occupational
dose limits
Time, Distance, and Shielding
Structural Barriers
Minimize Scatter Radiation
Avoid Primary Beam
Never stand in Primary Beam
Avoid holding patient
Use lead gloves
Have mother hold child
Wear apron
Ignorance is greatest enemy
Occupational Dose
Monitoring
Maximum Permissible Dose (MPD)
Assume linear non-threshold model
Largest allowable rad dose that is not expected to
result in significant effects
In any given year, can exceed annual dose limits
provided not over lifetime 5(N-18) in rems
Less than 18 years oldnot employed in radiation
occupation
Max 0.1 rem of whole body radiation
Occupational Dose
Monitoring
Maximum Permissible Dose
Specified for
Whole body radiation exposure
Partial body radiation exposure
Organ Exposure
General Population
Annual Occupational
Dose Limits
Whole Body
NCRP 91/Occupational 5 rems (50 msv)
ICRP 602 rems (20 msv)
Public0.5 rems (5.0msv) infrequent; 0.1 (1.0 msv)
frequent
Lifetime
NCRP/ICRP1 rem (10msv) x age
Annual Occupational
Dose Limits
Skin/Hands/Feet
NCRP/ICRP 50 rem (500msv)
Total Fetal
NCRP0.5 rems (5 msv)
ICRP0.1 rems(1.0 msv)
Film Badges
Provides permanent record
Monthly or quarterly
Control badge for air kerma and or background
radiation
Inexpensive
Latent image fades with timefilm should be
developed within 1-2 months
Heat can affect film and give false recordings
Thermoluminescent
Dosimeters
Thermoluminescent Dosimeters
Use non-metallic crystalline solids
Lithium Fluoride (LIF) is most common
Dose Equivalent
Dose Equivalent
Units
Sievert (Sv)SI system
Rem (radiation equivalent man)non SI
1 Sv= 100 Rem
1 Rem= 10 mSV
Units of Radiation
Deposition and Safety
Absorbed Dose
1 Gray= 1J of energy/kg
1 Rad=100 ergs of energy deposited/gram
1 Gray=100 rads
1 Rad= 10 mGy