Potassium in Water by Atomic Absorption Spectrophotometry: Standard Test Method For
Potassium in Water by Atomic Absorption Spectrophotometry: Standard Test Method For
Copyright by ASTM Int'l (all rights reserved); Fri Jul 10 10:52:33 EDT 2015
2
Downloaded/printed by
Universidad Nacional De Colombia (Universidad Nacional De Colombia) pursuant to License Agreement. No further reproductions authorized.
D4192 − 15
operators analyzing each sample on three consecutive days, 14.3.3 Calculate the mean and standard deviation of the
within its range for reagent water varies with the quantity being seven values and compare to the acceptable ranges of bias in
measured according to Table 1. Table 1. This study should be repeated until the recoveries are
NOTE 3—Only reagent water was used to obtain the precision statement
within the limits given in Table 1. If a concentration other than
since this test method is designated for the determination of trace amounts the recommended concentration is used, refer to Practice
of potassium in low-solids water. D5847 for information on applying the F test and t test in
13.2 It is the user’s responsibility to ensure the validity of evaluating the acceptability of the mean and standard devia-
this test method for waters of untested matrices. tion.
13.3 Recoveries of known amounts of potassium (from 14.4 Laboratory Control Sample (LCS):
potassium chloride) in the series of prepared standards for the 14.4.1 To ensure that the test method is in control, analyze
same laboratories and operators were as given in Table 1. a LCS containing a known concentration of potassium with
each batch (laboratory-defined or twenty samples). The labo-
13.4 Precision and bias for this test method conforms to
ratory control samples for a large batch should cover the
Practice D2777 – 77, which was in place at the time of
analytical range when possible. The LCS must be taken
collaborative testing. Under the allowances made in 1.4 of
through all of the steps of the analytical method including
Practice D2777 – 13, these precision and bias data do meet
sample preservation and pretreatment. The result obtained for
existing requirements for interlaboratory studies of Committee
a mid-range LCS shall fall within 615 % of the known
D19 test methods.
concentration.
14. Quality Control 14.4.2 If the result is not within these limits, analysis of
samples is halted until the problem is corrected, and either all
14.1 In order to be certain that analytical values obtained the samples in the batch must be reanalyzed, or the results must
using these test methods are valid and accurate within the be qualified with an indication that they do not fall within the
confidence limits of the test, the following QC procedures must performance criteria of the test method.
be followed when analyzing potassium .
14.5 Method Blank:
14.2 Calibration and Calibration Verification:
14.5.1 Analyze a reagent water test blank with each
14.2.1 Analyze at least four working standards containing
laboratory-defined batch. The concentration of potassium
concentrations of potassium that bracket the expected sample
found in the blank should be less than 0.5 times the lowest
concentration, prior to analysis of samples, to calibrate the
calibration standard. If the concentration of potassium is found
instrument (see 10.2). The calibration correlation coefficient
above this level, analysis of samples is halted until the
shall be equal to or greater than 0.990. In addition to the initial
contamination is eliminated, and a blank shows no contamina-
calibration blank, a calibration blank shall be analyzed at the
tion at or above this level, or the results must be qualified with
end of the batch run to ensure contamination was not a problem
an indication that they do not fall within the performance
during the batch analysis.
criteria of the test method.
14.2.2 Verify instrument calibration after standardization by
analyzing a standard at the concentration of one of the 14.6 Matrix Spike (MS):
calibration standards. The concentration of a mid-range stan- 14.6.1 To check for interferences in the specific matrix
dard should fall within 615 % of the known concentration. being tested, perform a MS on at least one sample from each
Analyze a calibration blank to verify system cleanliness. laboratory-defined batch by spiking an aliquot of the sample
14.2.3 If calibration cannot be verified, recalibrate the with a known concentration of potassium and taking it through
instrument. the analytical method.
14.2.4 It is recommended to analyze a continuing calibra- 14.6.2 The spike concentration plus the background concen-
tion blank (CCB) and continuing calibration verification tration of potassium must not exceed the high calibration
(CCV) at a 10 % frequency. The results should fall within the standard. The spike must produce a concentration in the spiked
expected precision of the method or 615 % of the known sample that is 2 to 5 times the analyte concentration in the
concentration. unspiked sample, or 10 to 50 times the detection limit of the
14.3 Initial Demonstration of Laboratory Capability: test method, whichever is greater.
14.3.1 If a laboratory has not performed the test before, or if 14.6.3 Calculate the percent recovery of the spike (P) using
there has been a major change in the measurement system, for the following formula:
example, new analyst, new instrument, etc., a precision and P 5 100 @ A ~ V s 1V ! 2 B V s # /C V (1)
bias study must be performed to demonstrate laboratory
capability. where:
14.3.2 Analyze seven replicates of a standard solution A = analyte known concentration (mg/L) in spiked sample,
prepared from an Independent Reference Material containing a B = analyte known concentration (mg/L) in unspiked
mid-range concentration of potassium. The matrix and chem- sample,
istry of the solution should be equivalent to the solution used C = known concentration (mg/L) of analyte in spiking
in the collaborative study. Each replicate must be taken through solution,
the complete analytical test method including any sample V s = volume (mL) of sample used, and
preservation and pretreatment steps. V = volume (mL) of spiking solution added.
Copyright by ASTM Int'l (all rights reserved); Fri Jul 10 10:52:33 EDT 2015
3
Downloaded/printed by
Universidad Nacional De Colombia (Universidad Nacional De Colombia) pursuant to License Agreement. No further reproductions authorized.
D4192 − 15
14.6.4 The percent recovery of the spike shall fall within the 14.7.2 Calculate the standard deviation of the duplicate
limits, based on the analyte concentration, listed in Guide values and compare to the precision in the collaborative study
D5810, Table 1. If the percent recovery is not within these using an F test. Refer to 6.4.4 of Practice D5847 for informa-
limits, a matrix interference may be present in the sample tion on applying the F test.
selected for spiking. Under these circumstances, one of the 14.7.3 If the result exceeds the precision limit, the batch
following remedies must be employed: the matrix interference must be reanalyzed or the results must be qualified with an
must be removed, all samples in the batch must be analyzed by indication that they do not fall within the performance criteria
a test method not affected by the matrix interference, or the of the test method.
results must be qualified with an indication that they do not fall 14.8 Independent Reference Material (IRM):
within the performance criteria of the test method. 14.8.1 In order to verify the quantitative value produced by
NOTE 4—Acceptable spike recoveries are dependent on the concentra- the test method, analyze an Independent Reference Material
tion of the component of interest. See Guide D5810 for additional (IRM) submitted as a regular sample (if practical) to the
information. laboratory at least once per quarter. The concentration of the
14.7 Duplicate: IRM should be in the concentration mid-range for the method
14.7.1 To check the precision of sample analyses, analyze a chosen. The value obtained must fall within the control limits
sample in duplicate with each laboratory-defined batch. If the established by the laboratory.
concentration of the analyte is less than five times the detection
limit for the analyte, a matrix spike duplicate (MSD) should be 15. Keywords
used. 15.1 analysis; atomic absorption; potassium; water
SUMMARY OF CHANGES
Committee D19 has identified the location of selected changes to this standard since the last issue
(D4192 – 08) that may impact the use of this standard. (Approved Feb. 1, 2015.)
(1) 8.3 was modified to clarify the purity of the commercial (3) Section 13 was modified to allow for direct reading
standards. instruments.
(2) Section 10 was modified with standard and calibration (4) 14.2.4 and 14.6.3 were modified.
information and to use a computer.
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or [email protected] (e-mail); or through the ASTM website
(www.astm.org). Permission rights to photocopy the standard may also be secured from the Copyright Clearance Center, 222
Rosewood Drive, Danvers, MA 01923, Tel: (978) 646-2600; http://www.copyright.com/
Copyright by ASTM Int'l (all rights reserved); Fri Jul 10 10:52:33 EDT 2015
4
Downloaded/printed by
Universidad Nacional De Colombia (Universidad Nacional De Colombia) pursuant to License Agreement. No further reproductions authorized.