Adks 4 Gas Detect Alarm Device Manual

The CHNADKS ADKS-4 is a portable, intelligent gas detection device designed for various industries, featuring advanced technology for high sensitivity and reliability. It supports multiple gas types with specific detection ranges and alarm settings, and includes a user-friendly interface for operation and maintenance. The manual outlines the device's structure, technical features, operation procedures, and safety precautions to ensure effective usage.

Uploaded by

Harold
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
70 views14 pages

Adks 4 Gas Detect Alarm Device Manual

The CHNADKS ADKS-4 is a portable, intelligent gas detection device designed for various industries, featuring advanced technology for high sensitivity and reliability. It supports multiple gas types with specific detection ranges and alarm settings, and includes a user-friendly interface for operation and maintenance. The manual outlines the device's structure, technical features, operation procedures, and safety precautions to ensure effective usage.

Uploaded by

Harold
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 14

Manuals+ — User Manuals Simplified.

CHNADKS ADKS-4 Gas Detect Alarm Device Instruction


Manual

Home » CHNADKS » CHNADKS ADKS-4 Gas Detect Alarm Device Instruction Manual

ADKS-4
Portable type four in one gas detector
Operation Manual

Contents
1 Summary
2 Structure feature and working
principle
3 Technical features
4 Function and operation
5 Menu operation
6 Operation notices
7 Common failure and the solve
methods
8 Storage
9 Accessories and others
10 Appendix table
11 Documents / Resources
11.1 References

Summary

This portable type gas detector (hereinafter simply called detector) is completely intelligent gas detector which
adopted the most advanced large scale integrated circuit technology, international standard intelligent technical
level design technology and special digit imitation mix communication technology to design. The detector adopt
the natural disperse method test gas, the sensitivity elements adopt the best quality gas sensor, it has excellent
sensitivity and perfect repetitiveness, convenient to use and maintain, mostly meeting the requirements of
industry field safety monitor to equipment reliability, the outer shell adopt high strength engineering plastics, high
strength, better hand feeling and waterproof, anti dust and explosion proof.
This detector widely applied in the industries like petroleum, chemical industry, environment protection,
metallurgy, gas transporting and distribution, biochemistry, medicine and agriculture, etc.
1.1 The design, manufacture and calibration of this product follow the bellow national standards:
GB3836.1-2010 Explosion environment part 1: equipment common requirements
GB3836. 4-2010 Explosion environment part 4: the
equipment which protected by the essence safety type “i” GB15322. 3-2003 Portable type combustible gas
detector part
3: the portable type combustible gas detector which measure range are (0~100) %LEL
JJG693-2011 Calibrate regulations of combustible gas test alarm device
JJG365-2008 Calibrate regulations of electric-chemical oxygen tester
JJG695-2019 Calibrate regulations of hydrogen sulfide gas tester
JJC915-2008 Calibrate regulations of carbon monoxide test alarm device
GB12358-2006 Common technical requirements of task area environment gas test alarm instrument

Structure feature and working principle

2.1 Structure function contrast table

1. 2. 3 Alarm indicate lamp window 7 Gas response hole

4 Left key 8 Charging port

5 Middle key 9 Liquid crystal screen

6 Right key

2.2 Detector structure: mainly formed with the shell body, circuit board, battery, display screen, sensor, charger
and other components.
2.3 Working principle: electric-chemical type and catalyze burning type

Technical features

Common gas detect range:


Gas types Range Low alarm point High alarm point Resolution ratio

EX (0~100)%LEL 20%LEL 50%LEL 1%LEL

H2S (0~100)PPM 10PPM 20PPM 1PPM

CO (0~1000)PPM 50PPM 200PPM 1PPM

O2 (0~30)%VOL 19.5%V0L 23.5%V0L 0.1%VOL

Please contact our company for other gas assembly.

Indicate error: ±5%FS (combustible), ±10% (carbon


monoxide), ±2μmol/mol or ±10% (hydrogen sulfide).
Responding time: T<45s
Indicate method: LCD indicate the real time data and system status
Light-emitting diode, sound, vibrate indication alarm, failure and lack voltage
Working environment: temperature -20℃~50℃, humidity <95%RH (no condensation)
Working voltage: DC3.7V (lithium battery capacity 2000mAh)
Explosion proofmark: Ex ib IIB T3 Gb
Charging time: 6h~8h
Standby time: 16h~18h under standard situation
Working life of sensor: EX, H2S, CO: 2~3 years, O2 : 5 years
Size: 130*65*45(mm)
Weight: 0.5kg

Function and operation

4. 1 Power on self inspection and preheat process


Press down middle key about 3s when detector under power off status, the backlight of detector display screen
lighting along with two sound “DI DI”, now the detector enter into power on status from power off, the screen
indicate welcome interface at the same time. Check fig 1 and fig 2.

After the welcome interface, the screen will indicate the high report value, low report value and range information
of combustible gas, oxygen, hydrogen sulfide and carbon monoxide in sequence, check fig 3, fig 4, fig 5 and fig 6.
The system enter into outer setting self inspection status and light self inspection status after information page
indication finished: now light will flashing twice; vibrate sheet self inspection: now the vibrate sheet will vibrating
then stop; sound self inspection: now the buzzer will interval sounding twice. The above status normally then
means the outer setting self inspection already normally passed. Check fig 7, fig 8 and fig 9.

4.2 Normally test and alarm status


The screen will normally indicate the current gas value when the detector not detected the gas which
concentration higher than the low limit alarm value. Check fig 10.
The detector will send out interval “DI…..DI….” alarm sound when the detector detected that higher than low limit
alarm value of any one gas, the screen backlight lighting, the vibrate sheet vibrating at the same time. The
detector only stop sounding and vibrating when the detect gas concentration lower than the low limit alarm value,
the screen close the backlight.
4. 3 System status checking
If the users want to check the current system battery use status, date and time then can press down left key under
the normal test status, now the screen will indicate the date, time, battery power and voltage information. Check
fig 11

4.4 Shut Down


Long time press down middle key about 3s when detector under normal test status, the screen will display the
power off interface, press left key then power off, press right key then cancel power off. Check fig 12

4.5 Charging
When the battery voltage of detector lower than 3.5V under the normal test status then will indicate the reminding
“Battrry Low, ShutDown”, now need insert the USB charging wire to charge immediately, otherwise, it maybe
caused system abnormally work because too low voltage. Check fig 13.

Charging under power off status, now charging will indicate “Charging….”. Check fig 14.

It will indicate “Charge Over” after charging finished. Check fig 15.

Menu operation

1. Enter into menu: press middle key and enter into main menu interface.
2. Key performance: press left key in main menu is upward selection, cancel or rightward key position select
function according. Press middle key is confirm, save or enter into sub menu function.
3. If not operate in menu exceed about 5s then system automatically return to normal test interface.

Main menu structure


Operating method:
Press left key then circling upward select menu item, press right key then circling d
ownward select menu item, press middle key then enter into middle menu item, if s
elect withdraw then withdraw the main menu and return to normal test interface.

Zero point micro adjustment


This function is plus or reduce the setting value based on the actually tested gas co
ncentration. Used in that unable to calibrate gas but display that not rest and have
deviation situation.
Operating method: firstly select the zero point micro adjustment in the menu, press
middle key and enter into, the screen display that input password interface, enter in
to gas select interface after input correct password, press left key then circling upw
ard select gas type, press right key then circling downward
select gas type. Press middle key and enter into zero point micro adjustment interfa
ce after selected the gas type.
Press left key then one by one rightward circling select every one byte of the adjust
ment value or symbol. Press middle key then save the current value.
Example: Able to set -0004 here when the combustible gas value in air always displ
ay at 3 and not reset, then save. Withdraw normal test interface. Now the combusti
ble gas display value should be 0.
Note: the setting value a little bigger than the display value at 1 unit then okay, if too
big setting then will
caused too big concentration deviation between the display value and actual gas a
nd not correct.

Gas calibration
To avoid the users error enter into this function and affect the portable type gas det
ector working. This function independently set password, this item function only pro
vide for professional staffs operation, the detail operation steps ignored here.
Alarm setting
This function used to set the alarm limit when test the gas concentration.
Low report means that instrument generate sound-optic alarm when gas
concentration exceed the low report value, high report means that instrument gene
rate sound-optic alarm when gas concentration exceed the high report value.
Operating method:
Firstly select the alarm setting in the menu, press middle key and enter into, the scr
een display that select gas
interface, press left key then circling upward select gas type, press right key then ci
rcling downward select gas type.
Press middle key and enter into alarm setting interface after selected the gas type,
press left key then one by one rightward circling select every one byte of the adjust
ment value. Press right key adjust the current digit byte value. Press middle key the
n save the current value

Alarm record
Alarm record menu include three sub menu items like check alarm, clear alarm and
withdraw.
Operating method: Press left key then upward circling select the sub menu item, pr
ess right key then downward circling select the sub menu item. Press middle after s
elected the gas type and enter into the corresponding menu item.

Alarm View
This function used to check the past alarm records.
Operating method:
Press left key and check the latest one piece record, press right key then check the
last one record. Automatically withdraw after not operate 5s.
Clear the alarm
This function used to clear the alarm record
Operating method: Press left key to select “OK”, used to clear the alarm record; pre
ss right key to select “Quit”, used to cancel that clear alarm record.

Alarm record Exit


This function withdraw the alarm record sub menu.
Operating method: Press middle key to withdraw.

Time setting
This function used to set the current instrument time.
Operating method:
Firstly select the time setting in menu, press middle key and enter into, display the
year setting interface. Press left key and one by one rightward circling select the ba
ck two bytes of value which need be adjusted, press right key to adjust the current
digit value, press middle key and used to save the current input year, then enter int
o month input interface, and so on, press the middle key to save the early input all ti
me information after input minute interface and withdraw the current time setting m
enu item.
Language
This function used to shift system language, include three sub menus like Chinese,
English and withdraw.
Operating method: Firstly select the language shift in menu, press middle key and
enter into, press the left key and upward select language item, press right key and
downward select the language item, press middle key to confirm the current selecti
on. If select the English menu item then system language will shift to
English. Select Chinese in the corresponding menu of English interface then the sy
stem language will shift to
Chinese.

Factory Mode
This function used to set the current instrument engineer setting.
Operating method: Firstly select the engineer setting in menu, press middle key an
d enter into, the engineer setting only limit used to set the instrument by our compa
ny’s professional staffs, this function independently set password to avoid the user
error enter into this function and affect the portable type gas detector work, the det
ail operation steps ignored.

Default Setting
This function used to recover the instrument to leave factory setting status. Detail o
perations: firstly select the
engineer setting in menu, press middle key and enter into, the screen display whet
her recover to setting interface, press “OK” or “Quit” key to confirm whether save it.

Exit
This function used to make instrument withdraw operation and return to main interf
ace.
Detail operations: firstly select the engineer setting in menu, press middle key then i
nstrument display the normal test main interface.

Warning: please don’t charge the portable type gas detector alarm instrument at test field, avoid the spark which
generated by that plug in and plug out charger caused fire or explosion; please do best that not charge the
portable type gas detector under start machine status, avoid it affect the charging speed.

Operation notices
1. Prevent this machine fallen from high space or under serious vibration;
2. When existing high concentration gas or unable to normally use this machine;
3. Please strictly follow the specification to operate and use, otherwise, it maybe caused incorrect test result or
damage this machine.
4. This product can’t be storage or utilize in the environment where contain corrode gas (example more higher
concentration chlorine, etc), also can’t in other serious environment, include too high or too low temperature,
use and storage this machine in more higher humidity, electromagnetism and strong sunshine.
5. Please use clean soft cloth soak water to lightly wipe if long time used and has pollutants on this machine
surface, don’t use the corrode dissolvant and hard objects to wipe surface of this machine, otherwise, it maybe
caused the surface of this machine scratched or damaged.
6. This machine should fix time calibrating to guarantee the test precision, the calibrating period not exceed one
year.
7. Any application or operation failure which exceed that described in this specification then please contact our
company to solve it.
8. Unable to disassemble or renewal the battery group in explode gas environment, also unable to charge the
battery group. Unable to use the outer setting connect equipment which not pass through explosion proof
certification in explode gas environment. Also unable to renewal sensor.

Common failure and the solve methods

Failure appearance Possible failure reason Treatment method

Too low voltage Please charge in time

Please contact the distributor or manufacturer for r


Unable to start machine Dead machine
epair

Please contact the distributor or manufacturer for r


Electric circuit failure
epair

Please contact the distributor or manufacturer for r


No reaction at the test gas Electric circuit failure
epair

Please contact the distributor or manufacturer for r


Sensor exceed the period
Incorrect display enewal sensor

Long time not calibrating Please calibrating in time

Battery power completely


Charging in time and reset the time
exhausted
Time display error
Strong electromagnetis m
Reset the time
disturb

Unable to use the zero point


Too much sensor shifting Calibrating in time or renewal sensor
calibrate function

Light instrument normal dete Please contact the distributor or manufacturer for r
Sensor failure
ct interface display full range enewal sensor

Storage
The detector should be storage in the air venting room where environment temperature is -10℃~55℃ and the
relative humidity not bigger than 85%, and the air not contain the harm gas or varias which act corrode
performance on the detector.

Accessories and others

The assorted detector provide one package box, one set portable type gas detector, one set specification and one
set qualified certificate and warrant card.

Appendix table

(the customized instrument able to select the detect gases in the below table to assemble)

Resolution rati
Test gas Common range Optional range Low alarm High alarm
o

Oxygen 0-30%V0L 0-30%VOL 0.1%VOL 19.5 23.5

Methane 0-4%CH4 0-4%CH4 0.01%CH4 1.00 2.50

Carbon monoxid
0-1000PPM 0-2000/5000PPM 1PPM 50 200
e

Hydrogen sulfide 0-100PPM 0-50/200/1000PP M 1/0.1PPM 10 20

Combustible gas 0-100%LEL 0-100%LEL 1%LEL 20 50

Ammonia 0-100PPM 0-50/500/1000PP M 1/0.1PPM 20 50

Hydrogen 0-1000PPM 0-40000PPM 1/0.1PPM 200 500

Chlorine 0-20PPM 0-100/150PPM 1/0.1PPM 5 10

Hydrogen chlorid
0-20PPM 0-20/150PPM 1/0.1PPM 5 10
e

Sulfur dioxide 0-20PPM 0-50/100PPM 1/0.1PPM 5 10

Nitric oxide 0-250PPM 0-500/1000PPM 1/0.1PPM 20 125


Nitrogen dioxide 0-20PPM 0-50PPM 1/0.1PPM 5 10

Carbon dioxide 0-5000PPM 0-5%/10%vol (Infrared) 1PPM/0.1% vol 1000/0. 2 2000/0.5

Hydrogen phosp
0-20PPM 0-20PPM 1PPM 5 10
hide

Hydrogen cyanid
0-20PPM 0-20PPM 1PPM 10 25
e

Epoxy acetylene 0-100PPM 0-100PPM 1PPM 20 50

Epoxy ethane 0-100PPM 0-100PPM 1PPM 20 50

Ozone 0-100PPM 0-20/100PPM 0.1PPM 20 50

Methanal 0-40PPM 0-50/100PPM 1/0.1PPM 8 20

Benzene 0-1000PPM 0-1000PPM 1PPM 200 500

Methyl benzene 0-1000PPM 0-1000PPM 1PPM 200 500

Dimethyl benzen
0-1000PPM 0-1000PPM 1PPM 200 500
e

VOC 0-1000PPM 0-1000PPM 1PPM 200 500

Chlorethyle ne 0-250PPM 0-250PPM 1PPM 50 125

Methyl alcohol 0-30PPM 0-30PPM 1PPM 6 15

Isobutene 0-90PPM 0-90PPM 1PPM 18 45

Alcohol 0-80PPM 0-80PPM 1PPM 16 40

Note: the test range of customized type instrument subject to the actually leave factory instrument, the gas not
listed here please contact the factory.

Documents / Resources

CHNADKS ADKS-4 Gas Detect Alarm Device [pdf] Instruction Manual


ADKS-4, ADKS-4 Gas Detect Alarm Device, Gas Detect Alarm Device, Detect Alarm Device, Al
arm Device, Device

References
User Manual