Agilent 8902A

Measuring Receiver
Technical Specifications

Agilent 11722A Sensor Module


Agilent 11792A Sensor Module
Agilent 11793A Microwave Converter
Agilent 11812A Verification Kit

The Agilent Technologies 8902A measuring receiver receiver, with the 11793A, counts RF signals to
delivers the accuracy and resolution of a high per- 26.5 GHz with 10 Hz resolution and excellent long-
formance power meter at frequencies from 150 kHz term frequency stability. The 8902A measuring
to 1.3 GHz (50 MHz to 26.5 GHz with the Agilent receiver with Option 050 offers increased power
11793A microwave converter) and levels from measurement accuracy. This option specifies Tuned
+30 dBm to –127 dBm. It accurately measures AM, RF Level on the 8902A measuring receiver to an
FM, and fM, including residuals and incidentals, accuracy of ±(0.015 dB + 0.005 dB/10 dB step).
with a single keystroke. The 8902A measuring
AGILENT 8902A MEASURING RECEIVER* AM REJECTION (50 Hz to 3 kHz BW)3:
TECHNICAL SPECIFICATIONS AM Rejection Frequency Range Rates AM Depths

Specifications describe the test set’s warranted performance <20 Hz peak 150 kHz – 1300 MHz 400 Hz or 1 kHz ≤50%
and are valid over the entire operation and environmental deviation
ranges unless otherwise noted. All specifications are valid
<20 Hz peak 150 kHz – 26.5 GHz 400 Hz or 1 kHz ≤50%
after a 30-minute warm-up period of continuous opera- deviation
tion, and within the frequency ranges defined below.
RESIDUAL FM (50 Hz to 3 kHz BW):
Supplemental characteristics are intended to provide <8 Hzrms at 1300 MHz, decreasing linearly with
additional information useful in applying the instru- frequency to <1 Hzrms for 100 MHz and below.
ment by giving typical, but non-warranted performance
parameters. These characteristics are shown in Italics <17 Hzrms' 1300 MHz <fc ≤6.2 GHz.
and labeled as “nominal,” “typical,” or “supplemental.” <33 Hzrms' 6.2 GHz <fc ≤12.4 GHz.
<49 Hzrms' 12.4 GHz <fc ≤18.6 GHz.
* Shaded text signifies measurements made with the <65 Hzrms' 18.6 GHz <fc ≤26.5 GHz.
8902A measuring receiver using the 11793A microwave
Supplemental Characteristics:
converter and 11792A sensor module. With this config-
uration, all standard 8902A specifications apply except MAXIMUM FM DEVIATION, RESOLUTION, AND
where changes are shown as shaded text. MAXIMUM DEMODULATED OUTPUT SENSITIVITY
ACROSS AN OPEN CIRCUIT (600 Ω output impedance)5:

Frequency Modulation Maximum Maximum Demodulated Deviations


Resolution Output Sensitivity (DF)
1
RATES :
20 Hz to 10 kHz, 150 kHz ≤ fc <10 MHz. 100 Hz 0.01 mV/Hz DFpeak ≥ 40 kHz
20 Hz to 200 kHz, 10 MHz ≤ fc ≤1300 MHz.
10 Hz 0.1 mV/Hz 4.0 kHz ≤
20 Hz to 200 kHz, 10 MHz ≤ fc ≤26.5 GHz.
DFpeak <40 kHz
DEVIATIONS1:
1 Hz 1.0 mV/Hz DFpeak < 4 kHz
40 kHz peak maximum, 150 kHz ≤ fc <10 MHz.
400 kHz peak maximum, 10 MHz ≤ fc ≤1300 MHz. 0.1 Hz 1.0 mV/ Hz DFrms < 0.3 kHz
400 kHz peak maximum, 10 MHz ≤ fc ≤26.5 GHz. (rms detector only)

ACCURACY 1, 2, 3:
Resolution is increased one digit with 750 µs de-empha-
FM Accuracy Frequency Range Rates Deviations
sis and pre-display on.
±2% of reading 250 kHz – 10 MHz 20 Hz – 10 kHz ≤40 kHzpeak The demodulated output signal present at the
±1 digit
Modulation Out/Audio In connector is increased in
±1% of reading 10 MHz – 1300 MHz 50 Hz – 100 kHz ≤400 kHzpeak amplitude by a factor of 10 with 750 µs de-emphasis.
±1 digit

±5% of reading 10 MHz – 1300 MHz 20 Hz – 200 kHz ≤400 kHzpeak DEMODULATED OUTPUT DISTORTION 1, 4:
±1 digit THD Frequency Range Rates Deviations

±1% of reading 10 MHz – 26.5 GHz 50 Hz – 100 kHz ≤400 kHzpeak <0.3% 150 kHz – 400 kHz 20 Hz – 10 kHz <10 kHz
±1 digit
<0.1% 400 kHz – 10 MHz 20 Hz – 10 kHz <10 kHz
±5% of reading 10 MHz – 26.5 GHz 20 Hz – 200 kHz ≤400 kHzpeak
±1 digit <0.1% 10 MHz – 26.5 GHz 20 Hz – 100 kHz <100 kHz

For rms detector add ±3% of reading. DETECTORS: +peak, – peak, ±peak/2, peak hold, average (rms
1, 4
sinewave calibrated), rms.
DEMODULATED OUTPUT DISTORTION :
THD Frequency Range Rates Deviations STEREO SEPARATION (50 Hz to 15 kHz): >47 dB.

<0.1% 400 kHz – 10 MHz 20 Hz – 10 kHz <10 kHz

<0.1% 10 MHz – 1300 MHz 20 Hz – 100 kHz <100 kHz


1. But not to exceed: 20 kHz rates and 40 kHz peak deviations with 750 µs de-
<0.1% 10 MHz – 26.5 GHz 20 Hz – 100 kHz <100 kHz emphasis filter.
2. Not to exceed for stated accuracy: 50 Hz to 40 kHz rates with rms detector.
3. Peak residuals must be accounted for in peak readings.
4. With 750 µs de-emphasis and pre-display "off," distortion is not specified for
modulation outputs >4V peak. This condition can occur near maximum deviation
for a measurement range, at rates <2 kHz.
5. For optimum flatness, cables should be terminated with their characteristic
impedance.

2
Amplitude Modulation Phase Modulation
RATES: RATES:
20 Hz to 10 kHz, 150 kHz ≤fc<10 MHz. 200 Hz to 10 kHz, 150 kHz ≤fc <10 MHz.
20 Hz to 100 kHz, 10 MHz ≤fc≤1300 MHz. 200 Hz to 20 kHz, 10 MHz ≤fc ≤1300 MHz.
200 Hz to 20 kHz, 10 MHz ≤fc ≤26.5 GHz.
DEPTH: to 99%.
ACCURACY3:
ACCURACY 2, 3, 6:
±4% of reading ±1 digit, 150 kHz ≤fc <10 MHz.
AM Accuracy Frequency Range Rates Depths
±3% of reading ±1 digit, 10 MHz ≤fc ≤1300 MHz.
±3% of reading ±1 digit, 10 MHz ≤fc ≤26.5 GHz.
±2% of reading 150 kHz – 10 MHz 50 Hz – 10 kHz 5% – 99%
±1 digit For rms detector add ±3% of reading.

±3% of reading 150 kHz – 10 MHz 20 Hz – 10 kHz to 99% DEMODULATED OUTPUT DISTORTION: <0.1% THD.
±1 digit AM REJECTION (for 50% AM at 1 kHz rate)3:
±1% of reading 10 MHz – 1300 MHz 50 Hz – 50 kHz 5% – 99% <0.03 radians peak (50 Hz to 3 kHz BW).
±1 digit
MAXIMUM DEVIATION, RESOLUTION, AND MAXIMUM
±3% of reading 10 MHz – 1300 MHz 20 Hz – 100 kHz to 99% DEMODULATED OUTPUT SENSITIVITY ACROSS AN OPEN
±1 digit CIRCUIT (600 Ω output impedance)5:
±1.5% of reading 1300 MHz – 26.5 GHz 50 Hz – 50 kHz 5% – 99%
±1 digit
150 kHz ≤fc <10 MHz
±3% of reading 10 MHz – 26.5 GHz 20 Hz – 100 kHz to 99%
±1 digit
For rms detector add ±3% of reading.

FLATNESS 5, 7:
Flatness Frequency Range Rates Depths

±0.3% of reading 10 MHz – 1300 MHz 90 Hz – 10 kHz 20% – 80%


±1 digit

±0.3% of reading 10 MHz – 26.5 GHz 90 Hz – 10 kHz 20% – 80%


±1 digit

DEMODULATED OUTPUT DISTORTION:


<0.3% THD for ≤50% depth.
<0.6% THD for ≤95% depth.
For fc >1300 MHz add 0.4% THD.
FM REJECTION (50 Hz to 3 kHz BW)3: Supplemental Characteristics:
FM Rejection Frequency Range Rates Deviations
MODULATION RATES: usable from 20 Hz to 100 kHz with degraded
<0.2% AM 250 kHz – 10 MHz 400 Hz or 1 kHz <5 kHzpeak performance.
<0.2% AM 10 MHz – 1300 MHz 400 Hz or 1 kHz <50 kHzpeak DETECTORS: +peak, – peak, ±peak/2, peak hold, average (rms
sinewave calibrated), rms.
<0.2% AM 10 MHz – 26.5 GHz 400 Hz or 1 kHz <50 kHzpeak

RESIDUAL AM (50 Hz to 3 kHz BW): <0.01%rms.

Supplemental Characteristics:
DETECTORS: +peak, –peak, ±peak/2, peak hold, average (rms
sinewave calibrated), rms.
MAXIMUM DEPTH, RESOLUTION, AND MAXIMUM DEMODU-
LATED OUTPUT SENSITIVITY ACROSS AN OPEN CIRCUIT
(600 Ω output impedance)5:
Maximum Maximum Demodulated Depths 2. Not to exceed for stated accuracy: 50 Hz to 40 kHz rates with rms detector.
Resolution Output Sensitivity 3. Peak residuals must be accounted for in peak readings.
5. For optimum flatness, cables should be terminated with their characteristic
0.1% 0.01V / percent AMpeak ≥40.0% impedance.
6. For peak measurements only: AM accuracy may be affected by distortion generat-
0.01% 0.1V / percent AMpeak <40.0% ed by the measuring receiver. In the worst case this distortion can decrease accu-
racy by 0.1% of reading for each 0.1% of distortion.
0.001% 7. Flatness is the variation in indicated AM depth for constant depth on input signal.
(rms detector only) 0.1V / percent AMrms <3.0%

3
Modulation Reference Internal Time Base Reference
AM CALIBRATOR DEPTH AND ACCURACY: FREQUENCY: 10 MHz.
33.33% depth nominal, internally calibrated to an accuracy of ±0.1%.
AGING RATE:
FM CALIBRATOR DEVIATION AND ACCURACY: <1 x 10–6/month.
34 kHz peak deviation nominal, internally calibrated to an accuracy <1 x 10–9/day (Option 002)8.
of ±0.1%.
Supplemental Characteristics:
Supplemental Characteristics:
INTERNAL REFERENCE ACCURACY:
CARRIER FREQUENCY: 10.1 MHz. Overall accuracy is a function of timebase calibration, aging rate,
MODULATION RATE: 10 kHz. temperature effects, line voltage effects, and short-term stability.
Standard Option 002
OUTPUT LEVEL: – 25 dBm.
Aging Rate <1 x 10–6/mo. <1 x 10–9/day
Frequency Counter
Temperature Effects <2 x 10–7/°C <2 x 10–10/°C
RANGE:
150 kHz to 1300 MHz. Line Voltage Effects
150 kHz to 26.5 GHz. (+5%, –10% Line <1 x 10–6 <6 x 10–10
Voltage Change)
SENSITIVITY:
12 mVrms (–25 dBm), 150 kHz ≤fc ≤650 MHz. Short-Term Stability — <1 x 10–9 for
1 second average
22 mVrms (–20 dBm), 650 MHz ≤fc ≤1300 MHz.
40 mVrms (–15 dBm), 150 kHz ≤fc ≤650 MHz.
71 mVrms (–10 dBm), 650 MHz <fc ≤1300 MHz. RF Power
40 mVrms (–15 dBm), 1300 MHz <fc ≤26.5 GHz.
The Agilent 8902A measuring receiver, with
MAXIMUM RESOLUTION: 11722A sensor module, performs RF power meas-
1 Hz. urements from –20 dBm (10 µW) to +30 dBm (1 W)
10 Hz.
at frequencies from 100 kHz to 2.6 GHz.
ACCURACY:
± reference accuracy ± 3 counts of least-significant digit, The 8902A measuring receiver, with 11792A sensor
fc <100 MHz. module, performs RF power measurements from
± reference accuracy ± 3 counts of least-significant digit, or –20 dBm (10 µW) to +30 dBm (1 W) at frequencies
30 Hz, whichever is larger, fc ≥100 MHz.
from 50 MHz to 26.5 GHz.
Supplemental Characteristics:
RF POWER RESOLUTION9:
MODES: Frequency and Frequency Error (displays the difference 0.01% of full scale in watts or volts mode.
between the frequency entered via the keyboard and the actual RF 0.01 dB in dBm or dBrelative mode.
input frequency).
LINEARITY (includes sensor non-linearity):
SENSITIVITY IN MANUAL TUNING MODE: RF range linearity ± RF range-to-range change error.
Approximate frequency must be entered from keyboard.
0.22 mVrms (–60 dBm). RF RANGE LINEARITY (using recorder output)10:
0.71 mVrms (–50 dBm). ±0.02 dB, RF ranges 2 through 5.
±0.03 dB, RF range 1.
Using the RF amplifier and the IF amplifiers, sensitivity can be
Using front-panel display add ±1 count of least-significant digit.
increased to approximately:
–100 dBm. RF RANGE-TO-RANGE CHANGE ERROR (using recorder output):
–90 dBm, fc ≤1300 MHz. ±0.02 dB/RF range change from reference range.
–75 dBm, 1300 MHz <fc ≤26.5 GHz. Using front-panel display add ±1 count of least-significant digit.
INPUT SWR:
Using 11722A sensor module: <1.15.
Using 11792A sensor module:
<1.15, 1300 MHz ≤fc.
<1.25, 1300 MHz <fc ≤18.0 GHz.
8. After 30-day warm-up. <1.40, 18.0 GHz <fc ≤26.5 GHz.
9. The 8902A fundamental RF power measurement units are watts. Further internal
processing is done on this number to display all other units.
10. When using a power sensor, the noise specification may mask the linearity speci-
fication and become the predominant error. When operating on the top RF power
range, add the power sensor's linearity percentages found in the power sensor's
specifications.

4
ZERO SET (digital settability of zero): Tuned RF Level
±0.07% of full scale on lowest range.
Decrease by a factor of 10 for each higher range. POWER RANGE: –127 dBm to 0 dBm, using IF synchronous detector
(200 Hz BW).
Supplemental Characteristics: –100 dBm to 0 dBm, using IF average detector (30 kHz BW).

ZERO DRIFT OF METER: POWER RANGE (Using 11792A Sensor Module):


±0.03% of full scale/°C on lowest range. Decrease by a For IF Synchronous Detector:
factor of 10 for each higher range. +10 dBm to –117 dBm, 2.5 MHz ≤fc ≤1300 MHz.
+5 dBm to –105 dBm, 1300 MHz ≤fc ≤12.4 GHz.
NOISE (at constant temperature, peak change over any one-
+5 dBm to –100 dBm, 12.4 GHz ≤fc ≤18.0 GHz.
minute interval for the 11722A or 11792A sensor modules):
+5 dBm to –95 dBm, 18.0 GHz ≤fc ≤26.5 GHz.
0.4% of full scale on range 1 (lowest range).
0.13% of full scale on range 2. For IF Average Detector:
0.013% of full scale on range 3. +10 dBm to –90 dBm, 2.5 MHz ≤fc ≤1300 MHz.
0.0013% of full scale on range 4. +5 dBm to –80 dBm, 1300 MHz ≤fc ≤12.4 GHz.
0.00013% of full scale on range 5. +5 dBm to –75 dBm, 12.4 GHz ≤fc ≤18.0 GHz.
ZERO DRIFT OF SENSORS (1 hour, at constant temperature after +5 dBm to –70 dBm, 18.0 GHz ≤fc ≤26.5 GHz.
24-hour warm-up): 1.9 Special Function degrades Tuned RF Level minimum
±0.1% of full scale on lowest range for 11722A sensitivity by 10 dB.
and 11792A sensor modules.
FREQUENCY RANGE:
RF POWER RANGES OF AGILENT 8902A MEASURING RECEIVER
2.5 MHz to 1300 MHz.
WITH 11722A AND 11792A SENSOR MODULES:
2.5 MHz to 26.5 GHz.
–20 dBm to –10 dBm (10 µW to 100 µW), range 1.
–10 dBm to 0 dBm (100 µW to 1 mW), range 2. DISPLAYED RESOLUTION11:
0 dBm to +10 dBm (1 mW to 10 mW), range 3. 4 digits in watts or volts mode.
+10 dBm to +20 dBm (10 mW to 100 mW), range 4. 0.01 dB or 0.001 dB in dBm or dB relative mode.
+20 dBm to +30 dBm (100 mW to 1 W), range 5.
4 digits in watts or volts mode.
RESPONSE TIME (0 to 99% of reading): 0.01 dB in dBm or dBrelative mode.
<10 seconds, range 1.
<1 second, range 2. RELATIVE MEASUREMENT ACCURACY (at constant temperature
<100 milliseconds, ranges 3 through 5. and after RF range calibration is completed)12:
DISPLAYED UNITS: Detector linearity + IF range-to-range error + RF range-to-range
Watts, dBm, dBrelative , %relative , volts, mV, µV, dB V, dB mV, dB µV. error + frequency drift error + noise error ± 1 digit.
INTERNAL NON-VOLATILE CAL-FACTOR TABLES Detector linearity + mixer linearity + IF range-to-range error + RF
(user-modifiable using special functions): range-to-range error + frequency drift error + noise error ± 1 digit.
Maximum number of cal factor/frequency entries:
Table #1 (primary): 16 pairs plus Reference Cal Factor.
Table #2 (frequency offset): 22 pairs plus Reference Cal Factor.
Maximum Allowed Frequency Entry: 42 GHz.
Frequency Entry Resolution: 50 kHz.
Cal Factor Range: 40 to 120%.
Cal Factor Resolution: 0.1%.

Power Reference
POWER OUTPUT:
1.00 mW. Factory set to ±0.7%, traceable to the U.S. National
Bureau of Standards.
ACCURACY: ±1.2% worst case (±0.9% rss) for one year (0 °C to
55 °C).

Supplemental Characteristics:
FREQUENCY: 50 MHz nominal.
SWR: 1.05 nominal.
11. The 8902A fundamental Tuned RF Level measurement units are volts. Further
FRONT PANEL CONNECTOR: N-type female. internal processing is done on this number to display all other units.
12. Tuned RF Level accuracy will be affected by residual FM of the source-under-test.
If the residual FMpeak is >50 Hz measured over a 30 second period in a 3 kHz BW.
Tuned RF Level measurements should be made using the IF average detector (30
kHz BW) by using Special Function 4.4. The Tuned RF Level measurement sensi-
tivity when using the IF average detector is –100 dBm.

5
DETECTOR LINEARITY: ACQUISITION TIME:
For IF Synchronous Detector: <4 seconds, ≥–110 dBm.
±0.007 dB/dB change, but not more than <10 seconds, ≥–127 dBm.
±0.02 dB/10 dB change.
<4 seconds, levels ≥–85 dBm.
Typically <±0.004 dB/dB change and
<10 seconds, levels <–85 dBm.
<±0.01 dB/10 dB change.
RESPONSE TIME (responding to changes in level of an acquired
For IF Average Detector (0 °C to +35 °C):
signal):
±0.013 dB/dB change,
<2 seconds, ≥–110 dBm.
but not more than±0.04 dB/10 dB change,
<5 seconds, ≥–127 dBm.
but not more than ±0.06 dB/10 dB change.
Typically <±0.008 dB/dB change and <2 seconds, ≥–85 dBm.
<±0.02 dB/10 dB change. <5 seconds, <–85 dBm.
MIXER LINEARITY: DISPLAYED UNITS: Watts, dBm, dBrelative , %relative , volts, mV, µV,
Negligible, levels ≤–5 dBm. dB V, dB mV, dB µV.
±0.04 dB, levels >–5 dBm and frequencies >1300 MHz.
IF RANGE-TO-RANGE ERROR (see Tuned RF Level range plot)13:
±0.02 dB/IF range change, IF ranges 1 through 5.
±0.05 dB/IF range change, IF ranges 6 through 7.
RF RANGE-TO-RANGE ERROR:
±0.04 dB/IF range change (Tuned RF Level only).
±0.06 dB/IF range change, RF Power to Tuned RF Level.
FREQUENCY DRIFT ERROR: ±0.05 dB/kHz frequency drift from
center of IF (using IF synchronous detector).
NOISE ERROR: ±0.18 dB for levels <–120 dBm, or for levels
<–110 dBm if Special Function 1.9 is selected.
±0.18 dB, levels <–110 dBm, 2.5 MHz ≤fc ≤1300 MHz.
±0.18 dB, levels <–98 dBm, 1300 MHz ≤fc ≤12.4 GHz.
±0.18 dB, levels <–93 dBm, 12.4 GHz ≤fc ≤18.0 GHz. 13

±0.18 dB, levels <–88 dBm, 18.0 GHz ≤fc ≤26.5 GHz.
Negligible elsewhere.
INPUT SWR:
<1.18, at 8902A RF input, RF range 1 and 2.
<1.40, at 8902A RF input, RF range 3.
<1.33, at 11722A RF input, RF range 1 and 2.
<1.50, at 11722A RF input, RF range 3.
<1.33, at 11722A RF input, RF range 3 with Special
Function 1.9.
Using 11792A sensor module:
<1.15, 1300 MHz ≤fc.
<1.25, 1300 MHz ≤fc ≤18.0 GHz.
<1.40, 18.0 GHz ≤fc ≤26.5 GHz.

Supplemental Characteristics:
ABSOLUTE LEVEL MEASUREMENT ACCURACY AT LOW LEVELS
(at constant temperature and after RF range calibration is com-
pleted)12:
Absolute level measurement accuracy is a function of the RF
Power and Tuned RF Level measurement accuracy. For a source
with an output SWR of 1.7 and level of –110 dBm the typical
absolute level measurement accuracy is 0.46 dB rss and 1.02 dB
worst case.
IF FREQUENCY: 455 kHz. 12. Tuned RF Level accuracy will be affected by residual FM of the source-under-test.
If the residual FMpeak is >50 Hz measured over a 30 second period in a 3 kHz
BW, Tuned RF Level measurements should be made using the IF average detector
(30 kHz BW) by using Special Function 4.4. The Tuned RF Level measurement sen-
sitivity when using the IF average detector is –100 dBm.
13. IF Ranges 6 and 7 (see Tuned RF Level range plots) are only used in automatic
operation for Tuned RF Level measurements below approximately –110 dBm for
the IF synchronous detector, and below approximately –85 dBm for the IF average
detector.
6
Audio Frequency Counter
FREQUENCY RANGE:
20 Hz to 250 kHz. (Usable to 600 kHz.)
MAXIMUM EXTERNAL INPUT VOLTAGE: 3Vrms.
ACCURACY (for demodulated signals)14:
Accuracy Frequency Modulation (Peak)

±3 counts of least-significant digit >1 kHz AM ≥10%


±Internal Reference Accuracy FM ≥1.0 kHz
f M≥1.5 radians
±0.02 Hz ≤1 kHz AM≥10%
±Internal Reference Accuracy FM≥1.0 kHz
fM≥1.5 radians

±0.2 Hz ≤3 kHz 1.5%≤AM<10%


±Internal Reference Accuracy 0.15 kHz≤FM
(3 kHz low-pass filter inserted) <1.0 kHz
Carrier Noise (Options 030-037) 0.15 radian≤fM
<1.5 radians
FREQUENCY RANGE: 10 MHz to 1300 MHz.
CARRIER POWER RANGE: +30 dBm to –20 dBm;
12.5 kHz, 25 kHz and 30 kHz filters. ACCURACY (for external signals)14:
+30 dBm to –10 dBm; carrier noise filter. Accuracy Frequency Level
DYNAMIC RANGE: 115 dB. ±3 counts of least-significant digit >1 kHz ≥100 mVrms
CARRIER REJECTION (temp. ≤35 °C): >90 dB; for offsets of at ±Internal Reference
least 1 channel spacing or 5 kHz, whichever is greater.
±0.02 Hz ≤1 kHz ≥100 mVrms
RELATIVE MEASUREMENT ACCURACY: ±Internal Reference Accuracy
±0.5 dB; levels ≥–95 dBc; 12.5 kHz, 25 kHz and
30 kHz filters.
Supplemental Characteristics:
±0.5 dB; levels ≥–129 dBc/Hz; carrier noise filter.
DISPLAYED RESOLUTION: 6 digits.
CARRIER NOISE FILTER:
Filter Noise Bandwidth: 2.5 kHz nominal. MEASUREMENT RATE: 2 readings per second.
Noise Bandwidth Correction Accuracy (stored in non-volatile COUNTING TECHNIQUE: Reciprocal with internal 10 MHz timebase.
memory): ±0.2 dB. AUDIO INPUT IMPEDANCE: 100 kΩ nominal.

Supplemental Characteristics: Audio RMS Level


ADJACENT/ALTERNATE CHANNEL FILTERS: FREQUENCY RANGE: 50 Hz to 40 kHz.
6 dB Filter Bandwidth:
8.5 kHz, 12.5 kHz adjacent-channel filter. VOLTAGE RANGE: 100 mV to 3 V.
16.0 kHz, 25 kHz adjacent-channel filter. ACCURACY: ± 4.0% of reading.
30.0 kHz, 30 kHz (cellular radio) alternate-channel filter.
TYPICAL NOISE FLOOR: –150 dBc/Hz, 0 dBm carrier power level. Supplemental Characteristics:
For system noise performance add LO contribution.
FULL RANGE DISPLAY: 0.3000 V, 4.000 V.
AC CONVERTER: True-rms responding for signals with crest factor
of ≤ 3.
MEASUREMENT RATE: 2 readings per second.
AUDIO INPUT IMPEDANCE: 100 kΩ nominal.

14. With the low-pass and high-pass audio filters used to stabilize frequency readings.

7
Audio Distortion RF Input
FUNDAMENTAL FREQUENCIES: 400 Hz ±5% and 1 kHz ±5%. FREQUENCY RANGE: 150 kHz to 1300 MHz.
MAXIMUM EXTERNAL INPUT VOLTAGE: 3 V. 150 kHz to 26.5 GHz when using the 11793A sensor module.
DISPLAY RANGE: 0.01% to 100.0% (–80.00 dB to 0.00 dB). OPERATING LEVEL:
DISPLAYED RESOLUTION: 0.01% or 0.01 dB. Minimum Maximum Frequency Range
Operating Level Operating Level
ACCURACY: ±1 dB of reading.
12 mVrms (–25 dBm) 7 Vrms (1 Wpeak) 150 kHz – 650 MHz
SENSITIVITY:
Source SWR <4
Modulation: 0.15 kHz peak FM, 1.5% peak AM or
0.6 radian peak fM. 22 mVrms (–20 dBm) 7 Vrms (1 Wpeak) 650 MHz – 1300 MHz
External: 100 mVrms. Source SWR <4
RESIDUAL NOISE AND DISTORTION15:
40 mVrms (–15 dBm) 7Vrms (1 Wpeak) 150 kHz – 650 MHz
0.3% ( –50 dB), temperature <40 °C.
71 mVrms (–10 dBm) 7Vrms (1 Wpeak) 650 MHz – 1300 MHz
Supplemental Characteristics:
MEASUREMENT 3 dB BANDWIDTH: 20 Hz to 50 kHz. 40 mVrms (–15 dBm) 7Vrms (1 Wpeak) 1300 MHz – 26.5 GHz

DETECTION: True rms.


Supplemental Characteristics:
MEASUREMENT RATE: 1 reading per second.
AUDIO INPUT IMPEDANCE: 100 kΩ nominal. TUNING:
Normal Mode: Automatic and manual frequency entry.
Audio Filters Track Mode: Automatic and manual frequency entry, fc≥ 10 MHz.
Normal and Track Mode: Manual entry of approximate frequency.
DE-EMPHASIS FILTERS: 25 ms, 50 ms, 75 ms, and 750 ms. De-
emphasis filters are single-pole, low-pass filters with 3 dB frequen- Acquisition Time (automatic operation): ~1.5 seconds.
cies of: 6366 Hz for 25 ms, 3183 Hz for 50 ms, 2122 Hz for 75 ms, INPUT IMPEDANCE: 50 Ω nominal.
and 212 Hz for 750 ms. MAXIMUM SAFE DC INPUT LEVEL: 5 V dc.
50 Hz HIGH-PASS FILTER (2 pole):
Flatness: <1% at rates ≥200 Hz. General Specifications
300 Hz HIGH-PASS FILTER (2 pole): TEMPERATURE:
Flatness: <1% at rates ≥1 kHz. Operating: 0 °C to 55 °C.
Storage: – 55 °C to 75 °C.
3 kHz LOW-PASS FILTER (5 pole):
Flatness: <1% at rates ≤1 kHz. REMOTE OPERATION: GPIB; all functions except the line switch
are remotely controllable.
15 kHz LOW-PASS FILTER (5 pole):
Flatness: <1% at rates ≤10 kHz. DEFINED IN IEEE-488.2 GPIB COMPATIBILITY: SH1, AH1, T5, TE0,
L3, LE0, SR1, RL1, PP0, DC1, DT1, C0, E1.
>20 kHz LOW-PASS FILTER (9 pole bessel)16:
Flatness: <1% at rates ≤10 kHz. EMI: Conducted and radiated interference is within the require-
ments of VDE 0871 (Level B), and CISPR publication 11.
Supplemental Characteristics: POWER: 100, 120, 220, or 240V (+5%, –10%); 48 to 66 Hz; 200 VA
DE-EMPHASIS FILTER TIME CONSTANT ACCURACY: ±3%. maximum.
HIGH PASS AND LOW PASS FILTER 3 dB CUTOFF FREQUENCY WEIGHT: Net 23.4 kg (52 lb); Shipping 31.4 kg (69 lb).
ACCURACY: ±3%.
DIMENSIONS: 190 mm H x 426 mm W x 551 mm D
>20 kHz LOW PASS FILTER 3 dB CUTOFF FREQUENCY: (7.5" x 16.8" x 21.7").
100 kHz nominal.
OVERSHOOT ON SQUARE WAVE MODULATION16: <1%.

15. For demodulated signals, the residual noise generated by the 8902A must be
accounted for in distortion measurements (that is residual AM, FM or fM).
16. The >20 kHz low-pass filter is intended for minimum overshoot with squarewave
modulation.

8
OPTION 050 SPECIFICATIONS AGILENT 11793A MICROWAVE CONVERTER
FREQUENCY RANGE: 2.5 MHz to 26.5 GHz. SPECIFICATIONS
TUNED RF LEVEL DYNAMIC RANGE: LO AMPLITUDE RANGE:
–120 dBm to 0 dBm. +8 dBm to +13 dBm, 2 GHz to 18 GHz.
–110 dBm to –15 dBm. +7 dBm to +13 dBm, 18 GHz to 26.5 GHz.
0 dBm to + 5 dBm, 18 GHz to 26.5 GHz with Option 001,
POWER ACCURACY: 011, or 021.
Using an Agilent 8902A Option 050 with 11722A sensor module
(10 to 1300 MHz): TEMPERATURE:
Relative accuracy: Operation: 0 °C to 55 °C.
±0.005 dB/10 dB step (0 to –100 dBm). Storage: –55 °C to 75 °C.
±0.050 dB/10 dB step (–100 to –120 dBm). –25 °C to 75 °C (Options 001, 011, and 021).
±0.015 dB ± 1 digit. POWER: 100, 120, 220, or 240 (+5%, –10%); 48 to 66 Hz;
Absolute accuracy: 20 VA maximum.
±0.005 dB/10 dB step (0 to –100 dBm). WEIGHT: Net 7.5 kg (16.5 lb); shipping 10.9 kg (24 lb).
±0.050 dB/10 dB step (–100 to –120 dBm).
±0.120 dB ± 1 digit. DIMENSIONS: 88 mm H x 425 mm W x 528 mm D.

Using an Agilent 8902A Option 050 with 11722A sensor module Supplemental Characteristics:
and 11793A microwave converter
(1300 to 2600 MHz, –15 to –110 dBm): RF INPUT CONNECTOR: 3.5 mm male.
Relative accuracy, 85 dB dynamic range: LO INPUT CONNECTOR: 3.5 mm male.
±0.005 dB/10 dB step (0 to 60 dB). IF OUTPUT CONNECTOR: N-type female.
±0.050 dB/10 dB step (60 to 85 dB).
±0.015 dB ± 1 digit. REAR PANEL CONTROL CONNECTOR: BNC female.
INCLUDED ACCESSORIES:
Absolute accuracy: Control Cable: 11170A BNC cable.
±0.005 dB/10 dB step (–15 to –100 dBm). LO Output to 11793A LO Input Cable: 3.5 mm female to 3.5 mm
±0.050 dB/10 dB step (–100 to –110 dBm). female flexible cable and 3.5 mm male to N-type male adapter;
±0.120 dB ± 1 digit. Options 001, 011, and 021 delete the 3.5 mm to N-type adapter.
Using an Agilent 8902A Option 050 with 11792A sensor module 8902A RF input to 11793A IF output cable: N-type male to N-type
and 11793A microwave converter male flexible cable.
(1300 MHz to 26.5 GHz, –15 to –100 dBm):
Relative accuracy, 85 dB dynamic range:
±0.005 dB/10 dB step (0 to 60 dB).
±0.050 dB/10 dB step (60 to 85 dB).
±0.015 dB ± 1 digit.
Absolute accuracy:
±0.005 dB/10 dB step (–15 to –100 dBm).
±0.120 dB ± 1 digit.
INPUT SWR:
<1.18, RF range 1 and 2.
<1.40, RF range 3.
TEMPERATURE:
Operating: 15 °C to 30 °C.
Storage: –55 °C to 74 °C.

Supplemental Characteristics:
MEASUREMENT TIME:
10 to 30 seconds.

9
AGILENT 11722A SENSOR MODULE AGILENT 11792A SENSOR MODULE
SPECIFICATIONS SPECIFICATIONS
FREQUENCY RANGE: 100 kHz to 2.6 GHz. FREQUENCY RANGE:
RF Power measurements:
POWER RANGE: +30 dBm (1 watt) to – 20 dBm (10 mW).
50 MHz to 26.5 GHz.
INPUT SWR (connected to an 8902A): 50 MHz to 18.0 GHz, Option 001.
<1.15, for RF Power measurements.
POWER RANGE: +30 dBm (1 watt) to – 20 dBm (10 mW).
<1.33, for Tuned RF Level measurements, RF range 1 and 2.
<1.5, for Tuned RF Level measurements, RF range 3. INPUT SWR (connected to an Agilent 11793A):
<1.33, for Tuned RF Level measurements, RF range 3 with <1.15, 1300 MHz ≤fc.
Special Function 1.9. <1.25, 1300 MHz <fc≤18.0 GHz.
<1.40, 18.0 GHz <fc≤26.5 GHz.
POWER SENSOR LINEARITY:
+2%, – 4%; +30 dBm to +20 dBm. POWER SENSOR LINEARITY:
Negligible deviation, levels <+20 dBm. +2%, – 4%; +30 dBm to +20 dBm.
Negligible deviation, levels <+20 dBm.
CALIBRATION FACTORS:
Each 11722A sensor module is individually calibrated. The calibra- CALIBRATION FACTORS:
tion factors are printed on the 11722A sensor module for easy ref- Each 11792A sensor module is individually calibrated. The calibra-
erence. tion factors are printed on the 11792A sensor module for easy ref-
erence.
CAL FACTOR UNCERTAINTY:
Frequency RSS Uncertainty Worst Case CAL FACTOR UNCERTAINTY:
(MHz) Uncertainty
Frequency RSS Uncertainty Worst Case
Uncertainty
0.1 0.7 % 1.6%
0.3 0.7% 1.6%
2.0 GHz 2.3 4.6%
1.0 0.8% 1.7%
6.0 GHz 2.5 5.0%
3.0 0.8% 1.7%
10.0 GHz 2.9 5.7%
10.0 0.9% 2.0% 14.0 GHz 3.4 6.6%
30.0 0.9% 2.0%
18.0 GHz 3.7 6.9%
50.0 0.0% (ref) 0.0% (ref)
22.0 GHz 3.8 7.8%
100.0 1.1% 2.2%
26.5 GHz 4.1 8.3%
300.0 1.1% 2.2%
1000.0 1.1% 2.2%
2600.0 1.2% 2.3% Supplemental Characteristics:
INPUT CONNECTOR: 3.5 mm male (N-type male, Option 001).
Supplemental Characteristics:
INPUT IMPEDANCE: 50 Ω nominal.
MAXIMUM PEAK POWER: 100 Wpeak or 300 W ms per pulse. SWITCH LIFE: >1,000,000 switchings.
INPUT IMPEDANCE: 50 Ω nominal. WEIGHT: Net 0.8 kg (1.75 lb); Shipping 1.2 kg (2.6 lb).
INPUT CONNECTOR: N-type male. DIMENSIONS: 51.2 mm H x 62.4 mm W x 1935 mm D
SWITCH LIFE: >1,000,000 switchings. (2" x 2.5" x 76.2").
SWITCH ISOLATION: >90 dB.
WEIGHT: Net 0.8 kg (1.75 lb); Shipping 1.2 kg (2.6 lb).
DIMENSIONS: 51.2 mm H x 62.4 mm W x 1935 mm D
(2" x 2.5" x 76.2").

10
AGILENT 11812A VERIFICATION KIT
SPECIFICATIONS
FREQUENCY: 30 MHz.
11812A ACCURACY: ±(0.003 dB + 0.003 dB/10 dB step).
OPTION 050 WORST CASE CUMULATIVE TUNED RF LEVEL
ACCURACY VERIFIED WITH 11812A:
±0.010 dB/10 dB step (0 to –100 dBm).
±0.050 dB/10 dB step (–100 to –120 dBm).
±0.015 dB ± 1 digit.
TEMPERATURE:
Operation: 15 °C to 30 °C.
Storage: –55 °C to 74 °C.

AGILENT 8902A REAR PANEL


INPUTS/OUTPUTS
Supplemental Characteristics:
FM OUTPUT: 10 kΩ impedance, –9 V to 6 V into an open circuit,
~6 V/MHz, dc coupled, 16 kHz bandwidth (one pole).
AM OUTPUT: 10 kΩ impedance, –4 V to 0 V into an open circuit,
~8 mV/%, dc coupled, 16 kHz bandwidth (one pole).
RECORDER OUTPUT: DC voltage proportional to the measured
results, 1 kΩ impedance, 0 V to 4 V for each resolution range into
an open circuit.
IF OUTPUT: 50 Ω impedance, 150 kHz to 2.5 MHz, –27 dBm to –3
dBm.
10 MHz REFERENCE OUTPUT: 50 Ω impedance, TTL levels (0 V to
>2.2 V into an open circuit). Available only with Option 002
1x10–9/day internal reference.
10 MHz REFERENCE INPUT 17: >500 Ω impedance, 0.5 V peak-to-peak
minimum input level.
LO INPUT (Option 003): 50 Ω impedance, ~1.27 MHz to 1301.5
MHz, 0 dBm nominal.
RF SWITCH REMOTE CONTROL OUTPUT: Provides output signals
necessary to remotely control either an Agilent 33311B,C Option
011 or an 8761A RF switch.
FREQUENCY OFFSET MODE REMOTE CONTROL OUTPUT: TTL
high output if in frequency offset mode (Special Function 27.1 or
27.3) with an external LO frequency >0, TTL low output for all
other cases.

17. External reference accuracy affects accuracy of all measurements.


11
Agilent Technologies’ Test and Measurement By internet, phone, or fax, get assistance with all your
Support, Services, and Assistance test and measurement needs.
Agilent Technologies aims to maximize the value you receive,
Online Assistance
while minimizing your risk and problems. We strive to ensure
that you get the test and measurement capabilities you paid www.agilent.com/find/assist
for and obtain the support you need. Our extensive support Phone or Fax
resources and services can help you choose the right Agilent United States:
products for your applications and apply them successfully. (tel) 1 800 452 4844
Every instrument and system we sell has a global warranty.
Support is available for at least five years beyond the produc- Canada:
tion life of the product. Two concepts underlie Agilent’s (tel) 1 877 894 4414
overall support policy: “Our Promise” and “Your Advantage.” (fax) (905) 206 4120

Our Promise Europe:


“Our Promise” means your Agilent test and measurement equip- (tel) (31 20) 547 2323
ment will meet its advertised performance and functionality. (fax) (31 20) 547 2390
When you are choosing new equipment, we will help you with Japan:
product information, including realistic performance specifica- (tel) (81) 426 56 7832
tions and practical recommendations from experienced test (fax) (81) 426 56 7840
engineers. When you use Agilent equipment, we can verify that
it works properly, help with product operation, and provide Latin America:
basic measurement assistance for the use of specified capabili- (tel) (305) 269 7500
ties, at no extra cost upon request. Many self-help tools are (fax) (305) 269 7599
available. Australia:
Your Advantage (tel) 1 800 629 485
“Your Advantage” means that Agilent offers a wide range of (fax) (61 3) 9210 5947
additional expert test and measurement services, which you New Zealand:
can purchase according to your unique technical and business (tel) 0 800 738 378
needs. Solve problems efficiently and gain a competitive edge (fax) (64 4) 495 8950
by contracting with us for calibration, extra-cost upgrades, out-
of-warranty repairs, and on-site education and training, as well Asia Pacific:
as design, system integration, project management, and other (tel) (852) 3197 7777
professional services. Experienced Agilent engineers and tech- (fax) (852) 2506 9284
nicians worldwide can help you maximize your productivity, Product specifications and descriptions in this
optimize the return on investment of your Agilent instruments document subject to change without notice.
and systems, and obtain dependable measurement accuracy Copyright © 1985, 2000 Agilent Technologies
for the life of those products. Printed in U.S.A. 10/00
5968-5312E

For more information visit our website at:


www.agilent.com/find/wireless

You might also like