Measuring Receiver
Technical Specifications
The Agilent Technologies 8902A measuring receiver receiver, with the 11793A, counts RF signals to
delivers the accuracy and resolution of a high per- 26.5 GHz with 10 Hz resolution and excellent long-
formance power meter at frequencies from 150 kHz term frequency stability. The 8902A measuring
to 1.3 GHz (50 MHz to 26.5 GHz with the Agilent receiver with Option 050 offers increased power
11793A microwave converter) and levels from measurement accuracy. This option specifies Tuned
+30 dBm to –127 dBm. It accurately measures AM, RF Level on the 8902A measuring receiver to an
FM, and fM, including residuals and incidentals, accuracy of ±(0.015 dB + 0.005 dB/10 dB step).
with a single keystroke. The 8902A measuring
AGILENT 8902A MEASURING RECEIVER* AM REJECTION (50 Hz to 3 kHz BW)3:
TECHNICAL SPECIFICATIONS AM Rejection Frequency Range Rates AM Depths
Specifications describe the test set’s warranted performance <20 Hz peak 150 kHz – 1300 MHz 400 Hz or 1 kHz ≤50%
and are valid over the entire operation and environmental deviation
ranges unless otherwise noted. All specifications are valid
<20 Hz peak 150 kHz – 26.5 GHz 400 Hz or 1 kHz ≤50%
after a 30-minute warm-up period of continuous opera- deviation
tion, and within the frequency ranges defined below.
RESIDUAL FM (50 Hz to 3 kHz BW):
Supplemental characteristics are intended to provide <8 Hzrms at 1300 MHz, decreasing linearly with
additional information useful in applying the instru- frequency to <1 Hzrms for 100 MHz and below.
ment by giving typical, but non-warranted performance
parameters. These characteristics are shown in Italics <17 Hzrms' 1300 MHz <fc ≤6.2 GHz.
and labeled as “nominal,” “typical,” or “supplemental.” <33 Hzrms' 6.2 GHz <fc ≤12.4 GHz.
<49 Hzrms' 12.4 GHz <fc ≤18.6 GHz.
* Shaded text signifies measurements made with the <65 Hzrms' 18.6 GHz <fc ≤26.5 GHz.
8902A measuring receiver using the 11793A microwave
Supplemental Characteristics:
converter and 11792A sensor module. With this config-
uration, all standard 8902A specifications apply except MAXIMUM FM DEVIATION, RESOLUTION, AND
where changes are shown as shaded text. MAXIMUM DEMODULATED OUTPUT SENSITIVITY
ACROSS AN OPEN CIRCUIT (600 Ω output impedance)5:
ACCURACY 1, 2, 3:
Resolution is increased one digit with 750 µs de-empha-
FM Accuracy Frequency Range Rates Deviations
sis and pre-display on.
±2% of reading 250 kHz – 10 MHz 20 Hz – 10 kHz ≤40 kHzpeak The demodulated output signal present at the
±1 digit
Modulation Out/Audio In connector is increased in
±1% of reading 10 MHz – 1300 MHz 50 Hz – 100 kHz ≤400 kHzpeak amplitude by a factor of 10 with 750 µs de-emphasis.
±1 digit
±5% of reading 10 MHz – 1300 MHz 20 Hz – 200 kHz ≤400 kHzpeak DEMODULATED OUTPUT DISTORTION 1, 4:
±1 digit THD Frequency Range Rates Deviations
±1% of reading 10 MHz – 26.5 GHz 50 Hz – 100 kHz ≤400 kHzpeak <0.3% 150 kHz – 400 kHz 20 Hz – 10 kHz <10 kHz
±1 digit
<0.1% 400 kHz – 10 MHz 20 Hz – 10 kHz <10 kHz
±5% of reading 10 MHz – 26.5 GHz 20 Hz – 200 kHz ≤400 kHzpeak
±1 digit <0.1% 10 MHz – 26.5 GHz 20 Hz – 100 kHz <100 kHz
For rms detector add ±3% of reading. DETECTORS: +peak, – peak, ±peak/2, peak hold, average (rms
1, 4
sinewave calibrated), rms.
DEMODULATED OUTPUT DISTORTION :
THD Frequency Range Rates Deviations STEREO SEPARATION (50 Hz to 15 kHz): >47 dB.
2
Amplitude Modulation Phase Modulation
RATES: RATES:
20 Hz to 10 kHz, 150 kHz ≤fc<10 MHz. 200 Hz to 10 kHz, 150 kHz ≤fc <10 MHz.
20 Hz to 100 kHz, 10 MHz ≤fc≤1300 MHz. 200 Hz to 20 kHz, 10 MHz ≤fc ≤1300 MHz.
200 Hz to 20 kHz, 10 MHz ≤fc ≤26.5 GHz.
DEPTH: to 99%.
ACCURACY3:
ACCURACY 2, 3, 6:
±4% of reading ±1 digit, 150 kHz ≤fc <10 MHz.
AM Accuracy Frequency Range Rates Depths
±3% of reading ±1 digit, 10 MHz ≤fc ≤1300 MHz.
±3% of reading ±1 digit, 10 MHz ≤fc ≤26.5 GHz.
±2% of reading 150 kHz – 10 MHz 50 Hz – 10 kHz 5% – 99%
±1 digit For rms detector add ±3% of reading.
±3% of reading 150 kHz – 10 MHz 20 Hz – 10 kHz to 99% DEMODULATED OUTPUT DISTORTION: <0.1% THD.
±1 digit AM REJECTION (for 50% AM at 1 kHz rate)3:
±1% of reading 10 MHz – 1300 MHz 50 Hz – 50 kHz 5% – 99% <0.03 radians peak (50 Hz to 3 kHz BW).
±1 digit
MAXIMUM DEVIATION, RESOLUTION, AND MAXIMUM
±3% of reading 10 MHz – 1300 MHz 20 Hz – 100 kHz to 99% DEMODULATED OUTPUT SENSITIVITY ACROSS AN OPEN
±1 digit CIRCUIT (600 Ω output impedance)5:
±1.5% of reading 1300 MHz – 26.5 GHz 50 Hz – 50 kHz 5% – 99%
±1 digit
150 kHz ≤fc <10 MHz
±3% of reading 10 MHz – 26.5 GHz 20 Hz – 100 kHz to 99%
±1 digit
For rms detector add ±3% of reading.
FLATNESS 5, 7:
Flatness Frequency Range Rates Depths
Supplemental Characteristics:
DETECTORS: +peak, –peak, ±peak/2, peak hold, average (rms
sinewave calibrated), rms.
MAXIMUM DEPTH, RESOLUTION, AND MAXIMUM DEMODU-
LATED OUTPUT SENSITIVITY ACROSS AN OPEN CIRCUIT
(600 Ω output impedance)5:
Maximum Maximum Demodulated Depths 2. Not to exceed for stated accuracy: 50 Hz to 40 kHz rates with rms detector.
Resolution Output Sensitivity 3. Peak residuals must be accounted for in peak readings.
5. For optimum flatness, cables should be terminated with their characteristic
0.1% 0.01V / percent AMpeak ≥40.0% impedance.
6. For peak measurements only: AM accuracy may be affected by distortion generat-
0.01% 0.1V / percent AMpeak <40.0% ed by the measuring receiver. In the worst case this distortion can decrease accu-
racy by 0.1% of reading for each 0.1% of distortion.
0.001% 7. Flatness is the variation in indicated AM depth for constant depth on input signal.
(rms detector only) 0.1V / percent AMrms <3.0%
3
Modulation Reference Internal Time Base Reference
AM CALIBRATOR DEPTH AND ACCURACY: FREQUENCY: 10 MHz.
33.33% depth nominal, internally calibrated to an accuracy of ±0.1%.
AGING RATE:
FM CALIBRATOR DEVIATION AND ACCURACY: <1 x 10–6/month.
34 kHz peak deviation nominal, internally calibrated to an accuracy <1 x 10–9/day (Option 002)8.
of ±0.1%.
Supplemental Characteristics:
Supplemental Characteristics:
INTERNAL REFERENCE ACCURACY:
CARRIER FREQUENCY: 10.1 MHz. Overall accuracy is a function of timebase calibration, aging rate,
MODULATION RATE: 10 kHz. temperature effects, line voltage effects, and short-term stability.
Standard Option 002
OUTPUT LEVEL: – 25 dBm.
Aging Rate <1 x 10–6/mo. <1 x 10–9/day
Frequency Counter
Temperature Effects <2 x 10–7/°C <2 x 10–10/°C
RANGE:
150 kHz to 1300 MHz. Line Voltage Effects
150 kHz to 26.5 GHz. (+5%, –10% Line <1 x 10–6 <6 x 10–10
Voltage Change)
SENSITIVITY:
12 mVrms (–25 dBm), 150 kHz ≤fc ≤650 MHz. Short-Term Stability — <1 x 10–9 for
1 second average
22 mVrms (–20 dBm), 650 MHz ≤fc ≤1300 MHz.
40 mVrms (–15 dBm), 150 kHz ≤fc ≤650 MHz.
71 mVrms (–10 dBm), 650 MHz <fc ≤1300 MHz. RF Power
40 mVrms (–15 dBm), 1300 MHz <fc ≤26.5 GHz.
The Agilent 8902A measuring receiver, with
MAXIMUM RESOLUTION: 11722A sensor module, performs RF power meas-
1 Hz. urements from –20 dBm (10 µW) to +30 dBm (1 W)
10 Hz.
at frequencies from 100 kHz to 2.6 GHz.
ACCURACY:
± reference accuracy ± 3 counts of least-significant digit, The 8902A measuring receiver, with 11792A sensor
fc <100 MHz. module, performs RF power measurements from
± reference accuracy ± 3 counts of least-significant digit, or –20 dBm (10 µW) to +30 dBm (1 W) at frequencies
30 Hz, whichever is larger, fc ≥100 MHz.
from 50 MHz to 26.5 GHz.
Supplemental Characteristics:
RF POWER RESOLUTION9:
MODES: Frequency and Frequency Error (displays the difference 0.01% of full scale in watts or volts mode.
between the frequency entered via the keyboard and the actual RF 0.01 dB in dBm or dBrelative mode.
input frequency).
LINEARITY (includes sensor non-linearity):
SENSITIVITY IN MANUAL TUNING MODE: RF range linearity ± RF range-to-range change error.
Approximate frequency must be entered from keyboard.
0.22 mVrms (–60 dBm). RF RANGE LINEARITY (using recorder output)10:
0.71 mVrms (–50 dBm). ±0.02 dB, RF ranges 2 through 5.
±0.03 dB, RF range 1.
Using the RF amplifier and the IF amplifiers, sensitivity can be
Using front-panel display add ±1 count of least-significant digit.
increased to approximately:
–100 dBm. RF RANGE-TO-RANGE CHANGE ERROR (using recorder output):
–90 dBm, fc ≤1300 MHz. ±0.02 dB/RF range change from reference range.
–75 dBm, 1300 MHz <fc ≤26.5 GHz. Using front-panel display add ±1 count of least-significant digit.
INPUT SWR:
Using 11722A sensor module: <1.15.
Using 11792A sensor module:
<1.15, 1300 MHz ≤fc.
<1.25, 1300 MHz <fc ≤18.0 GHz.
8. After 30-day warm-up. <1.40, 18.0 GHz <fc ≤26.5 GHz.
9. The 8902A fundamental RF power measurement units are watts. Further internal
processing is done on this number to display all other units.
10. When using a power sensor, the noise specification may mask the linearity speci-
fication and become the predominant error. When operating on the top RF power
range, add the power sensor's linearity percentages found in the power sensor's
specifications.
4
ZERO SET (digital settability of zero): Tuned RF Level
±0.07% of full scale on lowest range.
Decrease by a factor of 10 for each higher range. POWER RANGE: –127 dBm to 0 dBm, using IF synchronous detector
(200 Hz BW).
Supplemental Characteristics: –100 dBm to 0 dBm, using IF average detector (30 kHz BW).
Power Reference
POWER OUTPUT:
1.00 mW. Factory set to ±0.7%, traceable to the U.S. National
Bureau of Standards.
ACCURACY: ±1.2% worst case (±0.9% rss) for one year (0 °C to
55 °C).
Supplemental Characteristics:
FREQUENCY: 50 MHz nominal.
SWR: 1.05 nominal.
11. The 8902A fundamental Tuned RF Level measurement units are volts. Further
FRONT PANEL CONNECTOR: N-type female. internal processing is done on this number to display all other units.
12. Tuned RF Level accuracy will be affected by residual FM of the source-under-test.
If the residual FMpeak is >50 Hz measured over a 30 second period in a 3 kHz BW.
Tuned RF Level measurements should be made using the IF average detector (30
kHz BW) by using Special Function 4.4. The Tuned RF Level measurement sensi-
tivity when using the IF average detector is –100 dBm.
5
DETECTOR LINEARITY: ACQUISITION TIME:
For IF Synchronous Detector: <4 seconds, ≥–110 dBm.
±0.007 dB/dB change, but not more than <10 seconds, ≥–127 dBm.
±0.02 dB/10 dB change.
<4 seconds, levels ≥–85 dBm.
Typically <±0.004 dB/dB change and
<10 seconds, levels <–85 dBm.
<±0.01 dB/10 dB change.
RESPONSE TIME (responding to changes in level of an acquired
For IF Average Detector (0 °C to +35 °C):
signal):
±0.013 dB/dB change,
<2 seconds, ≥–110 dBm.
but not more than±0.04 dB/10 dB change,
<5 seconds, ≥–127 dBm.
but not more than ±0.06 dB/10 dB change.
Typically <±0.008 dB/dB change and <2 seconds, ≥–85 dBm.
<±0.02 dB/10 dB change. <5 seconds, <–85 dBm.
MIXER LINEARITY: DISPLAYED UNITS: Watts, dBm, dBrelative , %relative , volts, mV, µV,
Negligible, levels ≤–5 dBm. dB V, dB mV, dB µV.
±0.04 dB, levels >–5 dBm and frequencies >1300 MHz.
IF RANGE-TO-RANGE ERROR (see Tuned RF Level range plot)13:
±0.02 dB/IF range change, IF ranges 1 through 5.
±0.05 dB/IF range change, IF ranges 6 through 7.
RF RANGE-TO-RANGE ERROR:
±0.04 dB/IF range change (Tuned RF Level only).
±0.06 dB/IF range change, RF Power to Tuned RF Level.
FREQUENCY DRIFT ERROR: ±0.05 dB/kHz frequency drift from
center of IF (using IF synchronous detector).
NOISE ERROR: ±0.18 dB for levels <–120 dBm, or for levels
<–110 dBm if Special Function 1.9 is selected.
±0.18 dB, levels <–110 dBm, 2.5 MHz ≤fc ≤1300 MHz.
±0.18 dB, levels <–98 dBm, 1300 MHz ≤fc ≤12.4 GHz.
±0.18 dB, levels <–93 dBm, 12.4 GHz ≤fc ≤18.0 GHz. 13
±0.18 dB, levels <–88 dBm, 18.0 GHz ≤fc ≤26.5 GHz.
Negligible elsewhere.
INPUT SWR:
<1.18, at 8902A RF input, RF range 1 and 2.
<1.40, at 8902A RF input, RF range 3.
<1.33, at 11722A RF input, RF range 1 and 2.
<1.50, at 11722A RF input, RF range 3.
<1.33, at 11722A RF input, RF range 3 with Special
Function 1.9.
Using 11792A sensor module:
<1.15, 1300 MHz ≤fc.
<1.25, 1300 MHz ≤fc ≤18.0 GHz.
<1.40, 18.0 GHz ≤fc ≤26.5 GHz.
Supplemental Characteristics:
ABSOLUTE LEVEL MEASUREMENT ACCURACY AT LOW LEVELS
(at constant temperature and after RF range calibration is com-
pleted)12:
Absolute level measurement accuracy is a function of the RF
Power and Tuned RF Level measurement accuracy. For a source
with an output SWR of 1.7 and level of –110 dBm the typical
absolute level measurement accuracy is 0.46 dB rss and 1.02 dB
worst case.
IF FREQUENCY: 455 kHz. 12. Tuned RF Level accuracy will be affected by residual FM of the source-under-test.
If the residual FMpeak is >50 Hz measured over a 30 second period in a 3 kHz
BW, Tuned RF Level measurements should be made using the IF average detector
(30 kHz BW) by using Special Function 4.4. The Tuned RF Level measurement sen-
sitivity when using the IF average detector is –100 dBm.
13. IF Ranges 6 and 7 (see Tuned RF Level range plots) are only used in automatic
operation for Tuned RF Level measurements below approximately –110 dBm for
the IF synchronous detector, and below approximately –85 dBm for the IF average
detector.
6
Audio Frequency Counter
FREQUENCY RANGE:
20 Hz to 250 kHz. (Usable to 600 kHz.)
MAXIMUM EXTERNAL INPUT VOLTAGE: 3Vrms.
ACCURACY (for demodulated signals)14:
Accuracy Frequency Modulation (Peak)
14. With the low-pass and high-pass audio filters used to stabilize frequency readings.
7
Audio Distortion RF Input
FUNDAMENTAL FREQUENCIES: 400 Hz ±5% and 1 kHz ±5%. FREQUENCY RANGE: 150 kHz to 1300 MHz.
MAXIMUM EXTERNAL INPUT VOLTAGE: 3 V. 150 kHz to 26.5 GHz when using the 11793A sensor module.
DISPLAY RANGE: 0.01% to 100.0% (–80.00 dB to 0.00 dB). OPERATING LEVEL:
DISPLAYED RESOLUTION: 0.01% or 0.01 dB. Minimum Maximum Frequency Range
Operating Level Operating Level
ACCURACY: ±1 dB of reading.
12 mVrms (–25 dBm) 7 Vrms (1 Wpeak) 150 kHz – 650 MHz
SENSITIVITY:
Source SWR <4
Modulation: 0.15 kHz peak FM, 1.5% peak AM or
0.6 radian peak fM. 22 mVrms (–20 dBm) 7 Vrms (1 Wpeak) 650 MHz – 1300 MHz
External: 100 mVrms. Source SWR <4
RESIDUAL NOISE AND DISTORTION15:
40 mVrms (–15 dBm) 7Vrms (1 Wpeak) 150 kHz – 650 MHz
0.3% ( –50 dB), temperature <40 °C.
71 mVrms (–10 dBm) 7Vrms (1 Wpeak) 650 MHz – 1300 MHz
Supplemental Characteristics:
MEASUREMENT 3 dB BANDWIDTH: 20 Hz to 50 kHz. 40 mVrms (–15 dBm) 7Vrms (1 Wpeak) 1300 MHz – 26.5 GHz
15. For demodulated signals, the residual noise generated by the 8902A must be
accounted for in distortion measurements (that is residual AM, FM or fM).
16. The >20 kHz low-pass filter is intended for minimum overshoot with squarewave
modulation.
8
OPTION 050 SPECIFICATIONS AGILENT 11793A MICROWAVE CONVERTER
FREQUENCY RANGE: 2.5 MHz to 26.5 GHz. SPECIFICATIONS
TUNED RF LEVEL DYNAMIC RANGE: LO AMPLITUDE RANGE:
–120 dBm to 0 dBm. +8 dBm to +13 dBm, 2 GHz to 18 GHz.
–110 dBm to –15 dBm. +7 dBm to +13 dBm, 18 GHz to 26.5 GHz.
0 dBm to + 5 dBm, 18 GHz to 26.5 GHz with Option 001,
POWER ACCURACY: 011, or 021.
Using an Agilent 8902A Option 050 with 11722A sensor module
(10 to 1300 MHz): TEMPERATURE:
Relative accuracy: Operation: 0 °C to 55 °C.
±0.005 dB/10 dB step (0 to –100 dBm). Storage: –55 °C to 75 °C.
±0.050 dB/10 dB step (–100 to –120 dBm). –25 °C to 75 °C (Options 001, 011, and 021).
±0.015 dB ± 1 digit. POWER: 100, 120, 220, or 240 (+5%, –10%); 48 to 66 Hz;
Absolute accuracy: 20 VA maximum.
±0.005 dB/10 dB step (0 to –100 dBm). WEIGHT: Net 7.5 kg (16.5 lb); shipping 10.9 kg (24 lb).
±0.050 dB/10 dB step (–100 to –120 dBm).
±0.120 dB ± 1 digit. DIMENSIONS: 88 mm H x 425 mm W x 528 mm D.
Using an Agilent 8902A Option 050 with 11722A sensor module Supplemental Characteristics:
and 11793A microwave converter
(1300 to 2600 MHz, –15 to –110 dBm): RF INPUT CONNECTOR: 3.5 mm male.
Relative accuracy, 85 dB dynamic range: LO INPUT CONNECTOR: 3.5 mm male.
±0.005 dB/10 dB step (0 to 60 dB). IF OUTPUT CONNECTOR: N-type female.
±0.050 dB/10 dB step (60 to 85 dB).
±0.015 dB ± 1 digit. REAR PANEL CONTROL CONNECTOR: BNC female.
INCLUDED ACCESSORIES:
Absolute accuracy: Control Cable: 11170A BNC cable.
±0.005 dB/10 dB step (–15 to –100 dBm). LO Output to 11793A LO Input Cable: 3.5 mm female to 3.5 mm
±0.050 dB/10 dB step (–100 to –110 dBm). female flexible cable and 3.5 mm male to N-type male adapter;
±0.120 dB ± 1 digit. Options 001, 011, and 021 delete the 3.5 mm to N-type adapter.
Using an Agilent 8902A Option 050 with 11792A sensor module 8902A RF input to 11793A IF output cable: N-type male to N-type
and 11793A microwave converter male flexible cable.
(1300 MHz to 26.5 GHz, –15 to –100 dBm):
Relative accuracy, 85 dB dynamic range:
±0.005 dB/10 dB step (0 to 60 dB).
±0.050 dB/10 dB step (60 to 85 dB).
±0.015 dB ± 1 digit.
Absolute accuracy:
±0.005 dB/10 dB step (–15 to –100 dBm).
±0.120 dB ± 1 digit.
INPUT SWR:
<1.18, RF range 1 and 2.
<1.40, RF range 3.
TEMPERATURE:
Operating: 15 °C to 30 °C.
Storage: –55 °C to 74 °C.
Supplemental Characteristics:
MEASUREMENT TIME:
10 to 30 seconds.
9
AGILENT 11722A SENSOR MODULE AGILENT 11792A SENSOR MODULE
SPECIFICATIONS SPECIFICATIONS
FREQUENCY RANGE: 100 kHz to 2.6 GHz. FREQUENCY RANGE:
RF Power measurements:
POWER RANGE: +30 dBm (1 watt) to – 20 dBm (10 mW).
50 MHz to 26.5 GHz.
INPUT SWR (connected to an 8902A): 50 MHz to 18.0 GHz, Option 001.
<1.15, for RF Power measurements.
POWER RANGE: +30 dBm (1 watt) to – 20 dBm (10 mW).
<1.33, for Tuned RF Level measurements, RF range 1 and 2.
<1.5, for Tuned RF Level measurements, RF range 3. INPUT SWR (connected to an Agilent 11793A):
<1.33, for Tuned RF Level measurements, RF range 3 with <1.15, 1300 MHz ≤fc.
Special Function 1.9. <1.25, 1300 MHz <fc≤18.0 GHz.
<1.40, 18.0 GHz <fc≤26.5 GHz.
POWER SENSOR LINEARITY:
+2%, – 4%; +30 dBm to +20 dBm. POWER SENSOR LINEARITY:
Negligible deviation, levels <+20 dBm. +2%, – 4%; +30 dBm to +20 dBm.
Negligible deviation, levels <+20 dBm.
CALIBRATION FACTORS:
Each 11722A sensor module is individually calibrated. The calibra- CALIBRATION FACTORS:
tion factors are printed on the 11722A sensor module for easy ref- Each 11792A sensor module is individually calibrated. The calibra-
erence. tion factors are printed on the 11792A sensor module for easy ref-
erence.
CAL FACTOR UNCERTAINTY:
Frequency RSS Uncertainty Worst Case CAL FACTOR UNCERTAINTY:
(MHz) Uncertainty
Frequency RSS Uncertainty Worst Case
Uncertainty
0.1 0.7 % 1.6%
0.3 0.7% 1.6%
2.0 GHz 2.3 4.6%
1.0 0.8% 1.7%
6.0 GHz 2.5 5.0%
3.0 0.8% 1.7%
10.0 GHz 2.9 5.7%
10.0 0.9% 2.0% 14.0 GHz 3.4 6.6%
30.0 0.9% 2.0%
18.0 GHz 3.7 6.9%
50.0 0.0% (ref) 0.0% (ref)
22.0 GHz 3.8 7.8%
100.0 1.1% 2.2%
26.5 GHz 4.1 8.3%
300.0 1.1% 2.2%
1000.0 1.1% 2.2%
2600.0 1.2% 2.3% Supplemental Characteristics:
INPUT CONNECTOR: 3.5 mm male (N-type male, Option 001).
Supplemental Characteristics:
INPUT IMPEDANCE: 50 Ω nominal.
MAXIMUM PEAK POWER: 100 Wpeak or 300 W ms per pulse. SWITCH LIFE: >1,000,000 switchings.
INPUT IMPEDANCE: 50 Ω nominal. WEIGHT: Net 0.8 kg (1.75 lb); Shipping 1.2 kg (2.6 lb).
INPUT CONNECTOR: N-type male. DIMENSIONS: 51.2 mm H x 62.4 mm W x 1935 mm D
SWITCH LIFE: >1,000,000 switchings. (2" x 2.5" x 76.2").
SWITCH ISOLATION: >90 dB.
WEIGHT: Net 0.8 kg (1.75 lb); Shipping 1.2 kg (2.6 lb).
DIMENSIONS: 51.2 mm H x 62.4 mm W x 1935 mm D
(2" x 2.5" x 76.2").
10
AGILENT 11812A VERIFICATION KIT
SPECIFICATIONS
FREQUENCY: 30 MHz.
11812A ACCURACY: ±(0.003 dB + 0.003 dB/10 dB step).
OPTION 050 WORST CASE CUMULATIVE TUNED RF LEVEL
ACCURACY VERIFIED WITH 11812A:
±0.010 dB/10 dB step (0 to –100 dBm).
±0.050 dB/10 dB step (–100 to –120 dBm).
±0.015 dB ± 1 digit.
TEMPERATURE:
Operation: 15 °C to 30 °C.