0% found this document useful (0 votes)
8 views34 pages

D-Agent - 1 - work-91-s-Upload-ADRE SXP and 408 DSPi Datasheet - 172179

The ADRE Sxp Software and 408 DSPi form a scalable system for multi-channel signal processing and data acquisition, specifically designed for real-time applications. This system integrates various instrumentation functionalities into a single platform, allowing for remote operation and data storage without additional equipment. It supports a wide range of transducer inputs and offers flexible configurations for data collection and analysis, making it suitable for both rotating machinery and structural analysis applications.

Uploaded by

redwind79
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
8 views34 pages

D-Agent - 1 - work-91-s-Upload-ADRE SXP and 408 DSPi Datasheet - 172179

The ADRE Sxp Software and 408 DSPi form a scalable system for multi-channel signal processing and data acquisition, specifically designed for real-time applications. This system integrates various instrumentation functionalities into a single platform, allowing for remote operation and data storage without additional equipment. It supports a wide range of transducer inputs and offers flexible configurations for data collection and analysis, making it suitable for both rotating machinery and structural analysis applications.

Uploaded by

redwind79
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 34

ADRE Sxp and 408 DSPi

Datasheet
Bently Nevada Machinery Condition Monitoring 172179 Rev. AN

Description
ADRE Sxp Software and the 408 DSPi (Dynamic Signal
Processing Instrument) make up a highly scalable system for
multi-channel signal processing and data acquisition.
Unlike other general-purpose computer-based data
acquisition systems, ADRE Sxp and the 408 DSPi are
specifically designed for real-time highly parallel signal
processing and presentation. This extremely versatile system
incorporates the functionality of many types of
instrumentation, such as oscilloscopes, spectrum analyzers,
filters, signal conditioners, and digital recorders into a single
platform. The system is designed specifically for secure
corporate network environments, allowing it to operate
remotely across a LAN/WAN, or store data in full “stand-
alone” mode without an additional/external computer.
Additional equipment is seldom, if ever, needed. The system’s
real-time display capability permits it to continuously display
data independently of data being stored to permanent
memory. For established ADRE system users, ADRE Sxp also
supports all previous ADRE for Windows databases.
An ADRE Sxp data acquisition system consists of:
l 408 Dynamic Signal Processing Instrument(s).
l ADRE Sxp client software, ADRE Quick Configuration
software.
l A computer system capable of running ADRE Sxp
software.
ADRE Sxp and 408 DSPi
Datasheet 172179 Rev. AN
The 408 DSPi is fully portable or can be rack signal processing and sampling parameters
mounted allowing convenient operation in test can be changed “on-the-fly” without
stands, on-site, or at remote locations. The 408 interrupting data collection.
DSPi’s highly configurable design supports
The 408 DSPi architecture provides flexible
virtually all standard and non-standard input
hardware configuration. Users can insert
types including both dynamic transducer
sampling cards into the chassis as required.
signals (such as those from proximity probes,
Slots 1 through 4 support all standard sampling
velocity transducers, accelerometers, force
cards. Slot 5 is intended specifically for the
hammers, dynamic pressure sensors), and
Digital Replay card as well as future option
static signals (such as process variables from
cards.
transmitters and distributed control systems).
For rotating machinery applications, users can The standard Dynamic Sampling and Speed
provide a Keyphasor or other speed input Input/Trigger cards provide full support for
signal (such as that from a magnetic or optical structural analysis and impact testing. Data
transducer) to drive synchronous sampling can be analyzed natively or exported to 3rd
and order tracking. For structural analysis party applications if desired.
needs, impact testing using force hammers is
supported. The system also supports multiple The 408 DSPi front panel controls and displays
triggering modes for automated data basic functions and data. Users can directly
acquisition, allowing you to use the system as a initiate manual samples and triggering from
data or event logger without an operator the front panel without using ADRE Sxp
present. software. The front panel LEDs indicates
sampling and trigger status and activity. Users
The Client-Server architecture allows multiple can download multiple sampling
software clients to operate and simultaneously configurations to the 408 DSPi and later select
view data from single/multiple 408 DSPi them for use from the front panel.
systems simultaneously, permitting users to
independently view data in the fundamental 8-Channel Dynamic Sampling
measurement units of their choice. Software Card (168905 – AA)
installation and configuration are quick and
easy, allowing mass configuration of multiple The 8-channel dynamic sampling card is an
channels with minimal user intervention. extremely powerful and flexible signal
Configuration templates and software further processing engine. Along with the analog front-
simplifies the process, allowing the user to end conditioning, the user can configure most
install the software, produce a configuration, transducer inputs, with positive or negative
and begin capturing data in minutes. bias, while maintaining maximum signal input
range. An array of DSP processors and 24-bit
408 DSPi Overview ADCs provide maximum resolution. Input
signals can be either AC or DC coupled, and
Each 408 DSPi supports up to 4 sampling cards
users can independently define upper and
for up to 32 channels of data acquisition. The
lower input voltage levels along with full-scale
408 DSPi base system uses internal clocks and
range and transducer scale factor.
simulated speed/Keyphasor signals to support
both asynchronous and synchronous sampling The sampling card can provide a variety of
for all channels. Speed Input/Trigger cards data depending on the configuration and user
support up to 3 independent speed input needs. Each channel can provide multiple
channels for external speed inputs. Each Speed “static” variables including:
Input/Trigger card uses 1 available slot, and the
408 DSPi can use a maximum of 2 Speed l Direct amplitude.
Input/Trigger cards simultaneously providing l Bandpass amplitude.
up to 6 physical speed inputs and 6 simulated
speed inputs. The user can assign any speed
input (KPH) to any channel in a 408 DSPi. Most

2/34
ADRE Sxp and 408 DSPi
Datasheet 172179 Rev. AN
l 1X and 2X Amplitude and Phase, and up Each channel also includes a buffered output,
to 4 additional user defined nX vectors allowing the user to select either raw,
including amplitude and phase. conditioned analog, or TTL outputs. The
buffered outputs are independent of the signal
l Average & instantaneous gap or bias
being used for processing.
voltages.
l Multiple speed values. Digital Replay Card
l Adate/time stamp. (168907 – AA – BB-CC)
In addition to the static data, each channel can The Digital Replay Card provides simultaneous
provide up to 4 user-defined dynamic synchronous and asynchronous internal digital
waveforms. Users can configure waveforms for reprocessing and playback of all channels in
simultaneous Synchronous and/or the 408DSPi. The replay card maintains
Asynchronous sampling, with different exceptional accuracy and precision in the
sampling rates and/or frequency spans. The signal reprocessing that far surpasses the
sampling card supports up to 2 different capabilities of other equipment and
synchronous sampling rates simultaneously. reprocessing techniques. The digital replay
Time-Synchronous Averaging can also be card can play back raw data for all channels
enabled for all synchronous waveforms. In simultaneously including Keyphasor/speed
addition, the card can also sample and store and dynamic sampler inputs. Users can modify
the raw time-continuous data for each all sampling parameters on a KPH channel
channel. All channels within the system are when replayed, and fully manage and re-
always sampled simultaneously, are condition the gain, inversion, clamping, and
synchronized, and are initiated based on a set other characteristics of Keyphasor signals. This
of user defined events. provides the ability to control triggering edges
and thresholds as reprocessing and analysis
3-Ch Speed Input/Trigger Card
requires. On standard dynamic channels, users
(168906 – AA) can modify most sampling parameters. As an
The 3-channel speed input/trigger card example, users may add or modify waveform
supports a variety of transducer inputs and assignments, variable generation, filtering
signal conditioning needs including; proximity, options, frequency span, and Keyphasor
magnetic, optical, and laser pickups. assignments that did not exist in the original
Transducer power for both optical pickup and ± configuration reprocessing the data.
24Vdc proximity is also available if needed. For Additionally, users may add or modify all
impact testing applications, the force hammer sampling criteria and triggering parameters.
output can be connected directly to this card Full scale range, coupling, and transducer type
providing level triggering and full dynamic cannot be altered for standard dynamic
waveform capture. The card integrates a rich channels. The Digital Replay card occupies Slot
set of configurable analog signal conditioning #5 of the 408 DSPi and does not reduce the
tools including: input gain, voltage clamping, number of channels available for data
inversion, rising or falling edge trigger, collection.
auto/manual threshold, and hysteresis. The With the addition of Analog Output Replay Card
user can associate a programmable speed the user may send up to 32 channels of
input (Keyphasor) multiplier/divider for each recorded signals through data ports to auxiliary
channel independently, define discrete values equipment. These signals are analog
for events per revolution, or, a final ratio, recreations of the original data observed by
whichever is more convenient. Each channel ADRE. The Replay Module can be ordered with
can have up to three separate “stages” of or without the Analog Output Replay Card.
multiplier/divider ratios. Trigger/speed input Existing Replay Cards can be modified to
channels provide full dynamic sampling, include the Analog Output feature by sending
complete with static and waveform data, the cards to Product Repair at Bently Nevada.
available for real-time viewing and storage.

3/34
ADRE Sxp and 408 DSPi
Datasheet 172179 Rev. AN

Transducer Power Supply Card Networking Overview


(168908 – AA – BB) The 408 DSPi is a secure network appliance that
The transducer power supply card provides supports DHCP or fixed IP addressing based on
power for a wide variety of displacement, your network environment needs. When
velocity, acceleration including ICP accels, installed on a LAN/WAN in DHCP mode, the
force hammers, and other transducer types DHCP server or router will assign an IP address
used in field and test stand applications. This to the 408 DSPi. When the user makes a direct
card can simultaneously power up to 32 connection between the 408 DSPi and client
transducers in various combinations, and computer in DHCP mode, an IP address will be
provides direct physical connections for up to automatically assigned. The 408 DSPi also
16 transducers, eight ± 24 Vdc transducer supports fixed IP addressing. The user can
systems and eight constant current transducer assign a fixed IP address on one Ethernet port
systems. Field connection cable and adapter while simultaneously running DHCP on the
accessories accommodate additional second Ethernet port. In some instances,
transducer connections. primarily when navigating corporate security
infrastructures, firewalls, or VPNs (Virtual Private
In addition to ± 24 Vdc selections, the card Networks), specific router configuration may be
provides ± positive and negative bias necessary. Contact your local product support
selections for constant current applications, all representative for details specific to your
of which can be used simultaneously, to needs.
provide a highly flexible power source for most
any need. Users can configure transducer
power bias in blocks of four (4) directly from the
card without the need for special tools,
jumpers, or software.
Each output provides individual short-circuit
protection, current regulation, and indicators
for power status and voltage/current selection
(complete card status is provided within ADRE
Sxp client software). Also, this card can occupy
a dedicated option slot and leave all
transducer input slots available.
Field wiring cable accessories allow the user to
conveniently connect both power and signals
to the 408DSPi. These accessories support most
voltage and constant current transducer
applications without the need for additional
bulky equipment. Field wiring accessories and
cables must be ordered separately.

4/34
ADRE Sxp and 408 DSPi
Datasheet 172179 Rev. AN

AC 0.7 to 10 V
Specifications Configurable
Full Scale
408 DSPi Range

DC 0.35 to 50 V
Typical specifications are provided Configurable
for a temperature of +25 °C ± 3 °C Full Scale
(+77 °F ± 5.4 °F ) except where Range
noted.
Status Boot, Selftest, OK/Not OK,
Indication Activity, A/D over range
Data Storage Internal-480GB or 960GB
Capacity Direct Measurement Accuracy
Communication Dual 1000/100Mb RJ45
Ethernet Ports Filter values @ 0 db points unless
Protocol-TCP/IP specified otherwise.
DHCP or Fixed IP
addressing
LAN/WAN compatible 2 Dynamic Sampling Card
Amplitude vs. Frequency
Internal Cloc Battery-Backed Real-Time Cumulative Error
clock (RTC) accuracy is ±
2 seconds / month
Non-RMS, Non-Integrated Amplitude
Signal Conditioning-General AC Coupled – 1.6 Hz to 50 kHz
Hi Mode ± 1.25% of Full Scale Range2
8 Channel Dynamic Sampling Card
AC Coupled- N/A
Slot Position Slots 1 through 4 Low Mode
24 Bit A/D converters DC Coupled – 1 Hz to 50 kHz
Hi Mode ± 1% of Full Scale Range2
Input Impedance
± .011V below 1 V pp
Single Ended 742kΩ

Differential 1.484 MΩ between sig+ and DC Coupled- 0.167 Hz to 20 kHz


sig- 4-20 mA 511Ω Low Mode ± 1% of Full Scale Range2

Inputs
± .011V below 1 V pp
Single-ended 8 Non-Integrated, RMS Amplitude
Differential 4 Acceleration 10 Hz to 50 kHz
AC/DC ± 1% of Full Scale Range2
4-20 mA 8
Coupled Hi
Maximum -25 to 25 V Mode
Signal Input
Range:

5/34
ADRE Sxp and 408 DSPi
Datasheet 172179 Rev. AN

Velocity Bandpass Measurement Accuracy


AC/DC 10 Hz to 50 kHz Specifications are exclusive of filter
Coupled Hi ± 1% of Full Scale Range2 corner settings and transition
Mode regions. Filter values specified @ -3
AC Coupled N/A db points.
Low Mode

Acceleration 3 Hz to 50 kHz Non-RMS, Non- 1 Hz to 50 kHz


DC Coupled ± 1% of Full Scale Range2 Integrated ± 1% of Full Scale Input2
Low Mode Amplitude

Velocity DC 3 Hz to 50 kHz Non-RMS 1 Hz to 20 kHz


Coupled Low (-3db @ 3Hz ) Integrated, RMS ± 1% of Full Scale Input2
Mode ± 1% of Full Scale Range2 Integrated
Amplitude
Non-RMS Integrated, RMS Integrated
Amplitude Bandpass Filter Selections (Typical)

Acceleration 10 Hz to 20 kHz Butterworth 2 Pole ( -40 db/decade )


AC Coupled ± 1% of Full Scale Range2 4 Pole ( -80 db/decade )
Hi Mode 6 Pole ( -120 db/decade )
8 Pole ( -160 db/decade )
Velocity AC 3 Hz to 20 kHz
Coupled Hi ± 1% of Full Scale Range2 Range High Pass 1 Hz to 25.5kHz
Mode Low Pass 10 Hz to 50kHz
Min separation between
Direct Measurement Update Rates
HPF and LPF 2 Pole – 10.24 x HPF
Valid KPH or 0 to Peak / Peak to Peak 4 Pole-3.24 x HPF
Simulated values updated every 4 KPH 6 Pole-2.25 x HPF
KPH present periods. 8 Pole – 1.96 x HPF

Invalid KPH or High Mode – 2 sec sliding Options HPF < 10 Hz ( 2 Pole )
No KPH window HPF ≥ 10 Hz ( 2, 4, 6, 8 Pole)
present Low Mode – 12 sec sliding LPF ≥ 10 Hz ( 2, 4, 6, 8 Pole)
window
100ms update rate -3db Corner HPF & LPF-1 Hz Increments,
Frequencies -3db ± 5%
Integrated High Mode – 2 sec sliding
Values window Bandpass Measurement Update Rates
Low Mode – 2 sec sliding Valid KPH or 0 to Peak / Peak to Peak
window Simulated KPH values updated every 4 KPH
100ms update rate present periods.

Invalid KPH or 100ms update rate


No KPH present

Integrated 2 sec sliding window

Non-Integrated 2 sec sliding window

6/34
ADRE Sxp and 408 DSPi
Datasheet 172179 Rev. AN

Filtered Measurements Average Gap 5.3 / 8.84 sec,


-3db ± 5% @ .09 Hz
Filter Bandwidth Selectable
1.2 cpm, 12 cpm, 120 cpm Process Variable Measurements
(0.02 Hz, 0.2 Hz, 2 Hz)
Voltage Inputs 0 to 10 Vdc ( Typical )
User enabled 120 cpm < to > 12 cpm @ 600 1 to 5 Vdc Typical )
Auto-switching rpm
tracking filters 12 cpm < to > 1.2 cpm @ 60 Measurement -25 to 25 Vdc ( Upper and
transition rpm Range Lower voltage range fully
programmable)
Filter Settling 120 cpm < 0.477 / .796 sec
Time to 95 / 99 12 cpm < 4.77 / 7.96 sec Amplitude ± 0.12% of FSR @ 25V
% of final value 1.2 cpm < 47.7 / 79.6 sec ± 0.30% of FSR @ 10V
± 0.75% of FSR @ 1-5V
nX Amplitude 1 to 120k rpm (FSR = Full Scale Range)
and Phase ± 1% of Full Scale Range2
Accuracy Resolution 152.588 µV ( 0 to 10 Vdc )
± 3˚ of Input (Steady State) 61.035 µV ( 1 to 5 Vdc )

nX Resolution 0.01X Increments Response to 0.95 / 1.59 sec.


and Range 0.01X to ((sample/rev)/2-1)X 95%/99% of
Final Value
Below Minimum Amplitude
Low-pass filter -3db ± 5% @ 0.5 Hz
DC Coupled ≤ 0.015 Vpp
4 – 20 mA Input
AC Coupled ≤ 0.5% of full scale
Input Range 0-41.6 mA max
Gap Voltage Measurements
Amplitude ±1% of Full Scale Input
Measurement 0 V to 24 Vdc
Ranges -24 V to 0 Vdc Resolution 244 nA / bit
-12 to 12 Vdc
-24 to 24 Vdc Response to 5.3 / 8.84 sec
Upper and Lower voltage 95%/99% of
range fully programmable Final Value
between -25 to 25 Vdc
Low-pass filter -3db ± 5% @.09 Hz
Amplitude ± 0.17% of FSR @ -25 to 25 V
Dynamic Filtering associated with
± 0.26% of FSR @ 0 to ± 25 V
Waveform Data asynchronously sampled
± 0.26% of FSR @ -12.5 to 12.5
dynamic waveform data
V
specific to anti-alias filters.
(FSR = Full Scale Range)
Synchronously sampled
Resolution Measured 366.2 µV @ 24 V waveform data is not anti-
FSR alias filtered.

Response to 95%/99% of Final Value Asynchronous 128 to 128kHz ( 2.56 x


Sampling Rates Frequency Span, 50, 100,
Instantaneous 0.95 / 1.59 sec. 250, 500, 1k, 2.5k, 5k, 10k, 25k,
Gap -3db ± 5% @ 0.5 Hz 50k Hz )

Anti-Alias -80 db Minimum

7/34
ADRE Sxp and 408 DSPi
Datasheet 172179 Rev. AN

AC Coupled 1 Hz to 50 kHz Hardware Generated Spectra

Amplitude ±1% of Full Scale Range2 Spectral Lines 6400, 3200, 1600, 800, 400,
200, 100 lines, selectable.
Phase ±3˚ of Input One asynchronous
spectrum per channel.
DC Coupled DC Hz to 50 kHz Windowing provided on
spectrum up to 800 lines,
Amplitude ±1% of Full Scale Range2 Rectangular, Hanning, Flat-
Top, Exponential
Phase ±3˚ of Input
Free Running 1 per channel
Output Up to 4 simultaneous Spectrum
asynchronous waveforms
per channel Zoom 1 asynchronous spectrum
Spectrum per channel.
Synchronous Samples/rev (16, 32, 64, 128,
Sampling Rates 256, 360, 512, 720, 1024, Center Configurable in 1 Hz
2048) Frequency increments
Synch sample rate X rpm ≤
Zoom Factors 2, 5, 10, 20, 50
32 kHz
Spectral Lines 100, 200, 400, 800
Hardware Up to 2048 samples per
Generated waveform, up to 512
2Dynamic Sampling Card
Time Averages
Synchronous i.e. (8 revs @ 256 Amplitude vs. Frequency
Averaging samples/rev), Cumulative Error
(4 revs @ 512 samples/rev)

DC Frequency 937 rpm (15.62Hz) @ 2048x Frequency


Support 1.87k rpm (31.25Hz) @ 1024x dependent
2.66k rpm (44.4Hz) @ 720x amplitude and
3.75k rpm (62.5Hz) @ 512x phase errors
5.3k rpm (88.8Hz) @ 360x added to fixed
7.5k rpm (125Hz) @ 256x range
15k rpm (250Hz) @ 128x specifications.
30k rpm (500Hz) @ 64x
60k rpm (1kHz) @ 32x
120k rpm (2kHz) @ 16x Amplitude error 0 % from 0 to 10kHz
(AC Frequency support vs. frequency (± 4) % from 10kHz to 50kHz
from 1 Hz)
Phase error vs. (-0.5) to (-2.8) deg from 0
Output Up to 2 simultaneous frequency to 5kHz
synchronous waveforms per (-2.8) to (-6.6) deg from
channel 5kHz to 10kHz

Keyphasor /Speed Measurements:


3 Channel Speed Input (KPH) / Trigger
Dynamic Sampling Card

Slot Position Slots 1 through 4

8/34
ADRE Sxp and 408 DSPi
Datasheet 172179 Rev. AN

Accuracy 1 – 120k rpm, (+/- Output Protection Short circuit protected


0.00915)% of Period Input,
(+/- 11) rpm @ 120k rpm Raw Outputs Amplitude and Delay
Input
AC Error -0.91% to 0.42%
Simulated 1 – 120k rpm (+/- 0.02) % of
DC Error ± 60 mV
Keyphasor Period Input
Accuracy Signal Delay 0.66 µS (0.48 deg @ 2kHz)
Keyphasor card
not required to Conditioned Outputs Amplitude and Delay
provide
simulated AC Error -1.05% to 0.39%
Keyphasor (up
DC Error - 0.35 V to +60 mV
to 6 simulated
Keyphasors) Signal Delay 2.0 µS ( 4.0 µS Optical )
(1.4/2.8 deg @ 2kHz)
Inputs Transducer -24 Vdc, 57.6 mA max (-
Power 22.77 Vdc max, -24.48 Vdc
Total Inputs 3 speed inputs per card
min)
(single ended), maximum
2 cards per system.
+24 Vdc, 29 mA max
Supported Proximity, magnetic, or (+24.48Vdc max,
Transducers optical transducers. One +23.13Vdc min)
“powered” optical input
(Channel 3 only) +5 Vdc, 250 mA max (+5.2
Vdc max, +4.25 Vdc min,
Proximity inputs optical transducer power,
Channel 3 only)
Input Range -25 to +25 Vdc
Direct Filter values specified @ 0
Coupling AC or DC Measurement db points
(KPH) Accuracy
Input Impedance 128.9 kΩ
Non-RMS, Non -Integrated Amplitude
Buffered 3 channels, user
Transducer selectable output AC Coupled-High 1.6 Hz to 20 kHz
Outputs Mode ± 1.25% of Full Scale
Range3
Output Types Raw, Analog Conditioned,
TTL AC Coupled-Low N/A
± 22V output maximum Mode
20 µS min duty cycle for
TTL output DC Coupled-High 1 Hz to 20 kHz
Mode ± 1% of Full Scale Range3±
Output 330Ω .011V below 1 V pp
Impedance
DC Coupled-Low 0.167 Hz to 20 kHz
Output Drive 6100 pf (min) Mode ± 1% of Full Scale Range3±
Capacitance .011V below 1 V pp
Load Resistance ≥ 10k Ω

9/34
ADRE Sxp and 408 DSPi
Datasheet 172179 Rev. AN

Bandpass Specifications are Keyphasor Index While the shaft is stopped,


Measurement exclusive of filter corner the Keyphasor Index is
(KPH) Accuracy settings and transition used to assist with
regions. Filter values for positioning a shaft relative
bandpass specified @ -3 to a reference position.
db points Manual threshold must be
selected for speeds below
Non-RMS, Non - 1 Hz to 20 kHz 1 rpm.
Integrated ± 1% of Full Scale Range3
Amplitude Triggering Automatic or Manual
Mode
Gap Voltage (KPH) Measurements Selectable, positive or
negative edge of signal
Measurement 0 to 25 Vdc
input.
Range -25 to 0 Vdc
-12.5 to 12.5 Vdc Speed/Dynamic DC Coupled: DC to 20kHz
-25 to 25 Vdc Frequency Range AC Coupled: 1Hz to 20kHz
Amplitude ± 0.20% of FSR @ -25 to 25 Auto 1 rpm – 120k rpm (0.0167 Hz
V Threshold – 2kHz), min voltage
± 0.28% of FSR @ 0 to ± 25 required at low freq
V
± 0.28% of FSR @ -12.5 to Manual Threshold 1 rpm to 120k rpm (0.0167
12.5 V Hz to 2kHz) -25 Vdc to 25
(FSR = Full Scale Range) Vdc, 0.10 Vdc increments

Resolution Measured 381.47 µV @ 25 Input Clamping -25 to 25 Vdc, 0.01 V


V range increments, positive and
negative
Response to 95%/99% of Final Value
Waveform Inverting or non-inverting
Instantaneous 0.95 / 1.59 sec. Transformation
Gap -3db ± 5% @ 0.5 Hz
Hysteresis 0.2 to 2.0 V, 0.2 V
Average Gap 5.3 / 8.84 sec, increments
-3db ± 5% @ .09 Hz 0.2 to 1.0 V, 0.2 V
increments (Optical )
Status Boot Status
Types Self-test AC Gain 1, 2, 5, 10
Over/Under Speed
Activity Minimum Input 1 μS pulse
Edge Pulse Duty Cycle

Detection Error detection indicated if Maximum Trigger Input ≤ 1kHz: < 0.5 deg,
change in rotative speed Error with Sine 1kHz – 20 kHz: < 1 deg
between consecutive Wave Input
Keyphasor pulses is
greater than 25%, or shaft
rotative speed is less than
1 rpm, or greater than 120k
rpm

10/34
ADRE Sxp and 408 DSPi
Datasheet 172179 Rev. AN

Input Multiplier / 3 stages per input Low Voltage 5 V to 30 V (AC or DC)


Divider channel, 15 mA Maximum current
8 digits pre decimal, 12 2.15 kΩ ± 2 %
digits post decimal per (Low voltage input to
stage, configurable ratio return)
or real number in
software. Sampling Events

∆rpm 1 rpm to 120k rpm, min 1


3Speed Input KPH Card Amplitude
rpm increments
vs. Frequency Cumulative Error:
∆Time 0.1 s to 999 hours, 0.1 s
increments.
Frequency dependent amplitude 16 ms minimum event
and phase errors added to fixed period for all sampling
range specifications. events.

Digital Replay KPH/Speed Input


Amplitude error + 1% to (-1.5) % from 0.1Hz
vs. frequency to 20kHz Measurements
Phase error vs. (-1.5) to (-2.5) deg from 0 Speed Input/KPH cards (168906-02)
frequency to 2kHz delivered prior to November 2006 require
an update to work properly with the
(-2.5) to (-12) deg from
Digital Replay Card.
2kHz to 10kHz

Data Collection Trigger/Event When in replay mode, there is no alteration to


dynamic sampler data. The following apply to
Triggers
KPH measurements, in addition to standard
Amplitude Any variable, “or”, per operational KPH specifications except where
channel. (amplitude, noted:
phase, nX, Direct,
Bandpass, Gap, Process
Variable)

Rpm Upper and lower level, per


speed input

Time User-programmable,
recurring, scheduled

External Contact High or low voltage input ,


“normally open”, “normally
closed” logic selectable in
software

High Voltage 90 V to 240 V (AC or DC)


4 mA Maximum current
62 kΩ ± 2 %
(High voltage input to
return)

11/34
ADRE Sxp and 408 DSPi
Datasheet 172179 Rev. AN
Table 1: Event Rate = Shaft rpm x Events/rev Transducer Power Supply Card
Event Rate Rpm Error Phase Error Slot Slots 1 through 5
Positions
1 to 6k +/- < 1 rpm +/- < 1 deg
Voltage -24Vdc (-23.35Vdc to -24.48
6k to 30k +/- 4 rpm +/- < 1 deg Outputs Vdc)
30k to 60k +/- 7 rpm +/- 1 deg Max 84.5 mA ± 4.5 mA
60k to 90k +/- 20 rpm +/- 1 deg Current

90k to 120k +/- 40 rpm +/- 2 deg Short 55 mA ± 5mA per output, 8
Circuit outputs maximum
120k to 300k N/A +/- 3 deg +24Vdc (+24.48Vdc to +23.34
(typ) Vdc)

300k to 600k N/A +/- 6 deg Current 52 mA ± 5mA per output, 4


(typ) When outputs maximum
Shorted
600k to 900k N/A +/- 8 deg
(typ) Current Outputs

900k to 1200k N/A +/- 9 deg (+) Bias 3.36 mA ± 0.3mA @ Load: 10 Ω to
(typ) 6.5 kΩ per output, 4 outputs max

(-) Bias 3.30 mA ± 0.36mA @ Load: 10 Ω


Amplitude Sine wave input up to 8 Vpp any to 6.5 kΩ per output, 8 outputs
full scale range. max
Raw AC Error Status Boot Status
Outputs4 Indication OK Detection
Current Limit
-1.31% to DC Error Current Source-Bias (+/-)
0.82% Voltage Source-Bias (+/-)
± 60 mV Conditioned Outputs4

AC Error -1.45% to 0.79%

DC Error - 0.35 V to +60 mV

4 Speed Input/KPH Card amplitude


cumulative (total) error

AC Accuracy Speed Input/KPH signal


processing is independent from dynamic data
produced by the standard dynamic sampling
card. Reprocessing of data from the standard
sampling card is therefore not subject to any
additional signal processing errors.

12/34
ADRE Sxp and 408 DSPi
Datasheet 172179 Rev. AN

Physical Vibration

Dimensions Operational 0.25G @ 5-100Hz

Non- 3G @ 5-100Hz
408 DSPi (L x 36.1 x 41 x 10 cm (14.25 x 16 x 3.8 operational
W x H) in)
Shock 10G @ 2mS operational &
Power Supply 18.3 x 13.7 x 9.1 cm (7.2 x 5.4 x non-operational
(L x W x H) 3.6 in)
Environmental Considerations
Weight
The 408 DSPi is designed to meet a broad
408 DSPi 9.5 kg (21 lbs) @ 32 channels range of use cases and environments.
Significant design measures have been
Power Supply 1.3 kg (3 lbs)
implemented to provide a robust electrical and
Construction mechanical package. Shock isolation mounts
are used internally on critical components and
408 DSPi Aluminum Chassis with cast careful attention focused on reliability of the
aluminum front and rear system including extensive shock, vibration,
panels. and temperature exposure. The robust shipping
Black Powder Coating – case is designed for transportation and
textured finish, shipment in the most rigorous environments.
indoor/outdoor use.
Environmentally sealed The 408 DSPi is a measurement “instrument”
momentary tactile switches. and should be treated with the appropriate
Bayonet locking external consideration. Exposure to “extreme”
power connector environments will have constraints. Direct
exposure to condensing liquids, rain, sand, or
Environmental other particulates that could impair ventilation
are however, not appropriate. In scenarios that
Operating 408 DSPi and External may exceed environmental specifications
Temperature PowerSupply custom solutions can be created to meet
0˚ C to +50˚ C (+32˚ F to +122˚ specific needs. If you have any questions
F) regarding an application, please contact your
local Bently Nevada representative.
Rated
specifications at Input Power
fan inlets. Do not
block fans during External Power Supply
operation, un-
obstructed airflow is Input 90/264 Vac, 47/63 Hz auto
required. sensing

Output +32 Vdc ± 5% @ 10 A Max


Storage -30˚ C to +80˚ C (-22˚ F to +5 Vdc ± 3% @ 600 mA Max
Temperature +176˚ F)
Surge Up to 2kV Max
Relative 0% to 95% non-condensing Protection
Humidity
LED AC Power ready
Indicators DC Output enabled
Fault Latch Detection

13/34
ADRE Sxp and 408 DSPi
Datasheet 172179 Rev. AN

Faults AC supply fault


Over/Under output voltage error
Over current detection
Thermal protection
(Faults are latching and require
AC power cycle to reset)

14/34
ADRE Sxp and 408 DSPi
Datasheet 172179 Rev. AN

Compliance and Certifications


FCC
This device complies with part 15 of the
FCC Rules. Operation is subject to the
following two conditions:
l This device may not cause harmful
interference.
l This device must accept any
interference received, including
interference that may cause
undesired operation.

EMC
EN 61000-6-2
EN 61000-6-4
EN 61000-3-2
EN 61000-3-3
EMC Directive 2014/30/EU

RoHS
RoHS Directive 011/65/EU+2015/863

LV
LV Directive 2014/35/EU

India-Battery EPR Marking


GE Oil & Gas India Private Limited
EPR Certificate No.: 1.1595372902047E+20

15/34
ADRE Sxp and 408 DSPi
Datasheet 172179 Rev. AN

03 Transducer Power Supply Card


Ordering Information D: Slot # 4 Option

For the detailed listing of country and 00 Empty Slot


product-specific approvals, refer to
the Approvals Quick Reference Guide 01 8 Ch Dynamic Sampler Card
(108M1756).
02 3 Ch Speed Input (KPH) / Trigger
Card
For additional technical
documentation, please log in to 03 Transducer Power Supply Card
bntechsupport.com and access the
Bently Nevada Media Library. E: Slot # 5 Option

00 Empty Slot
ADRE 408 DSPi
01 Digital Replay Card w/o Analog
Output Card
168679-AA-BB-CC-DD-EE-FF-GG-HH-II-
JJ-KK 02 Digital Replay Card w/ Analog
Output Card

All accessories are ordered as 03 Transducer Power Supply Card


separate line items.
F: Power Supply Option

A: Slot # 1 Option 00 None

00 Empty Slot 01 90/264 Vac 47/63Hz (2US, 1 EU, 1


Br Power Cords Included)
01 8 Ch Dynamic Sampler Card
03 AC Power Supply with US
02 3 Ch Speed Input (KPH) / Trigger 10A/125V Power Cord
Card
04 AC Power Supply with US
03 Transducer Power Supply Card 10A/250V Power Cord

B: Slot # 2 Option 05 AC Power Supply with Euro 10A 2


Wire Power Cord
00 Empty Slot
06 AC Power Supply with Brazil
01 8 Ch Dynamic Sampler Card 10A/250V Power Cord
02 3 Ch Speed Input (KPH) / Trigger G: Carry Case Option
Card
00 None
03 Transducer Power Supply Card
01 Hard Shipping Case
C: Slot # 3 Option
H: Metrology Certification Option
00 Empty Slot
00 None
01 8 Ch Dynamic Sampler Card
01 All Cards
02 3 Ch Speed Input (KPH) / Trigger
Card

16/34
ADRE Sxp and 408 DSPi
Datasheet 172179 Rev. AN

I: Rack Mount Kit Option ADRE Sxp Software


00 None 4080/01-AA-BB-CC-DD-EE-FF-GG
01 with 19 in Rack Mount Kit
A: Version
J: Solid State Drive Option
01 Initial Order
01 480GB Solid State Drive
98 Upgrade
02 960GB Solid State Drive
99 Update
K: Approvals
B: Full Function Client-License Quantity (2)
00 No Approvals

KR Korea (safe area) 00 None

01 One (1) Client License


What is included when ordering the 408 DSPi
(168679): 02 Two (2) Client Licenses
l 1 Ethernet Cable – Cat5e with ferrite, 3 m

(10 ft), required to meet CE certification.
l Signal Input Cables are included with 10 Ten (10) Client Licenses
each Dynamic Sampler-quantity 8, and C: Network Viewer Edition-License Quantity
Speed Input (KPH) card-quantity 6, when (2)
ordered with the 408 DSPi (168679).
00 None
Additional signal input cables (172068)
can be ordered separately if needed. 01 One (1) Client License
l Laser KPH transducer input adapter cable
(169714-01) and Transducer Power Supply 02 Two (2) Client Licenses
cables and accessories are ordered
separately. ⋮
10 Ten (10) Client Licenses

D: Archive Viewer Edition-License Quantity

00 None

01 One (1) Client License

02 Two (2) Client Licenses


10 Ten (10) Client Licenses

E: Full Client-Media Quantity (1) (2)

00 None

01 One (1) CD-ROM

17/34
ADRE Sxp and 408 DSPi
Datasheet 172179 Rev. AN

02 Two (2) CD-ROMs ADRE Quick Configuration Software

⋮ 100M5072-01
10 Ten (10) CD-ROMs

F: Network Viewer Edition-Media Quantity (For use with ADRE Sxp. 2.8 or later.
(1) (2) Please contact your local sales
representative for additional
00 None ordering information)

01 One (1) CD-ROM

02 Two (2) CD-ROMs Setting up an ADRE is now as simple as 1,2 and


3.
⋮ 1. Arrange the cards.
10 Ten (10) CD-ROMs

G: Archive Viewer Edition-Media Quantity (1)

00 None

01 One (1) CD-ROM

02 Two (2) CD-ROMs


10 Ten (10) CD-ROMs 2. Sketch the machine.

(1) Media quantity must be either (1) or equal


to license quantity
(2) Includes ADRE Quick Configuration
Software Media P/N 100M5072-01

For the ADRE Maintenance & Support


Plan 3071/20 details, see System 1
datasheet 108M5214.

3. Plug it in.

18/34
ADRE Sxp and 408 DSPi
Datasheet 172179 Rev. AN
The ADRE Quick Configuration software allows
even the novice user to create an Sxp-ready Speed Input/KPH cards (168906–02)
database to configure ADRE and start receiving delivered prior to November 2006
data within minutes. Bently best practices are require an update to work properly
integrated into the drag and drop menu to with the Digital Replay Card.
automatically create data collection
parameters and a suite of plots to ensure the
full power and productivity of the ADRE system
is realized.
The newest to the most experienced ADRE user
can take advantage of the ease, speed, and
expertise of the ADRE Quick Configuration
software so that more time is spent enhancing
productivity utilizing the data, and less time
worrying about manually configuring the
system.

408 DSPi Accessories

(Accessories are ordered as separate line


items.)

8-Channel Dynamic Sampler Card

168905-AA

A: Approvals

00 No Approvals

KR Korea (safe area)

3-Channel Speed Input/KPH


Trigger Card

168906-AA

A: Approvals

00 No Approvals

KR Korea (safe area)

Requires appropriate adapter cable to


connect optical transducer inputs.

19/34
ADRE Sxp and 408 DSPi
Datasheet 172179 Rev. AN

Digital Replay Card Impact Hammer Kit

168907-AA-BB-CC Impact_Hammer_Kit -AA-BB-CC-DD-


EE-FF
A: Approvals
Transducer Power Supply Card
00 No Approvals
recommended/required
KR Korea (safe area)

B: Output Option A: Small-500lbf pk, 10mV/lbf, 0.3lbm


(285570-01)
00 No output
00 None
01 With analog output
01 Hammer included
C: Cables (only applicable if B option = 0 1)
B: Medium-1000lbf pk, 5mV/lbf, 0.3lbm
00 Not Included (285570-02)

01 One 8 Channel Cable 00 None

02 Two 8 Channel Cables 01 Hammer included

03 Three 8 Channel Cables C: Large-5000 lbf pk, 1 mV/lbf, 2.4 lbm


(285570-02)
04 Four 8 Channel Cables
00 None
Speed Input/KPH cards (168906–02)
delivered prior to November 2006 01 Hammer included
require an update to work properly
with the Digital Replay Card. D: Std Accelerometer Set (200350, 100mV/g
±50g)
00 None
Transducer Power Supply Card
04 Set of 4
168908-AA-BB 08 Set of 8

E: Micro Accelerometers
A: Approvals
00 None
00 No Approvals
F: Triax Accelerometers
KR Korea (safe area)
00 None
B: Output Accessory Option

00 None What is included when ordering the Impact


Hammer Kit:
The Impact Hammer Kit can be configured with
various types of impact hammers and
accelerometers. The kit includes a case
suitable to fit all components.

20/34
ADRE Sxp and 408 DSPi
Datasheet 172179 Rev. AN
Each hammer includes an assortment of tips 169347-01 External Power Supply to 408
and the necessary cable for connection to the DSPi Extension Cable, 1.8 m (6 ft).
408DSPi. Accelerometers can be ordered in sets For use with 169337. Using this
of 4 or 8, and each one will include the cable provides additional cable
necessary cable for connection to the 408 DSPi. length for a total of ~ 3 m (10 ft).
See Figure 4. Convenient for rack mount
Each component of the Impact Hammer Kit applications.
can also be ordered individually if needed, i.e.
169234-01 408 DSPi 19 in Rack Mount Kit
cables, etc. Please refer to the datasheet for the
Impact Hammer Kit for details. 169714-01 Keyphasor input adapter cable
(0.5m). Connects laser
The impact hammers and accels transducer kits to KPH card
require a constant current power 168906 – 02 (or /01). See 408
source. If the end user does not DSPi with Optical KPH and Laser
already have a method to power Tachometer speed input
these devices, a Transducer Power transducers..
Supply Card (168908-AA-BB) will be
required. 166812-01 Laser transducer kit with 2m
cable. Includes transducer and
extension cable. Requires
Additional Accessories adapter cable 169714-01 to
(Accessories are ordered as connect to KPH card 168906 – 02
separate line items) (or /01). See 408 DSPi with
Optical KPH and Laser
Tachometer speed input
172068 Signal Input Cable-SMA to BNC,
transducers.
3 m (10 ft) required for 168905 –
02 (or /01), and 168906 – 02 (or 166813-01 Laser transducer kit with 5m
/01). cable
Includes transducer and
173887 SMA Push-On Adapter, for SMA
extension cable. Requires
to BNC cable 172068. Allows SMA
adapter cable 169714-01 to
cable connector to be pushed
connect to KPH card 168906-02
onto signal input connector for
(or /01). See 408 DSPi with
convenience.
Optical KPH and Laser
286241 BNC male to SMA female Tachometer speed input
Adapter transducers.

174466 Ethernet Cable – Cat5e with 284814- Laser transducer extension


ferrite, 3 m (10 ft). Required to 020 cable 20 meters
meet CE certification. (quantity 1
included with 408DSPi) 284814-100 Laser transducer extension
cable 100 meters
181M6180- 408 DSPi Hard Shipping Case
02290050 Reflective tape roll
01 and Foam Assembly including
end caps 174968 Fan vent cover
update/replacement kit
169337 408 DSPi External Power Supply
90/264 Vac 47/63 Hz 176472 European AC Power Cord 250V
2.5m

21/34
ADRE Sxp and 408 DSPi
Datasheet 172179 Rev. AN

178775-01 408DSPi Transducer Power Answer: 178775-01 and 178897-01, Transducer


Supply Card Standard Field connection A to SIG and B to COM on 178897-
Wiring Harness. See Field Wiring 01; connect 178897-01 to 178775-01; connect
Cables for Transducer Power 178775-01 to input channels 5 through 8 on
and Signal Input and Detail of 168908 which is provided with the Transducer
Standard Wiring Harness Power Supply; set the Constant Current switch
178775-01. in the (pos +) position.

178762-01 408DSPi Transducer Power 2. What do I need to order if I want to


Supply Card Modular Field power 5 to 8 positive biased
Wiring Harness (requires Accelerometers/Velomitor (2-wire)
178763-01) See Field Wiring constant current transducers.
Cables for Transducer Power
and Signal Input5 and Detail of Answer: This would require an additional
Modular Wiring Harness 178762- Transducer Power Supply 168908-02 and 2 set
01 of 178775-01 and 178897-01, Transducer
connection will be the same as Question #1.
178763-01 408DSPi Transducer Power
Supply Card Modular Wiring 3. What do I need to order if I want to
Adapter. See Field Wiring Cables power 1 to 8 negative biased
for Transducer Power and Signal
Accelerometers/Velomitor (2-wire)
Input .
constant current transducers.
178897-01 408DSPi Transducer Power
Supply Card Field Wiring Answer: 178775-01 and 178897-01 (both are 4
Terminal Kit. See Field Wiring channel connectors so a set of 2 would be
Cables for Transducer Power required for 5 to 8), Transducer connection B
and Signal Input to SIG and A to COM on 178897-01; connect
178897-01 to 178775-01; connect 178775-01 to
177067-01 ADRE 408 Protective Cover 168908 which is provided with the Transducer
Assembly Power Supply; set the Constant Current switch
in the (neg -) position.
287743 ADRE 408 Analog Output to 8X
BNC Cable 4. What do I need to order if I want to
power more than 9 to 16 negative
289043-01 ADRE 408 Analog Output Rebuild
biased Accelerometers/Velomitor (2-
Kit (only for existing Replay
Cards) wire) constant current transducers.

173548-01 ADRE 408 Blank Panel Answer: This would require an additional
Transducer Power Supply 168908-02 and up
172070 Panel screw 4-40 Thread, ¾” to 4 set of 178775-01 and 178897-01 (both
Length connectors are capable of having 4
channels), Transducer connection will be the
Transducer Power Supply Cable Frequently same as Question #3.
Asked Questions 5. What do I need to order if I want to
power 1 to 8 negative bias 3-wire
1. What do I need to order if I want to
transducers (Accelerometers,
power 1 to 4 positive biased
Velomitors, or Proximitor Probe).
Accelerometers/Velomitor (2-wire)
constant current transducers.

22/34
ADRE Sxp and 408 DSPi
Datasheet 172179 Rev. AN

Answer: 178762-01 and 178897-01 (2 each 9. What do I need to order if I want to


would be required if channel count was power 13 to 24 positive bias 3-wire
greater than 4) and 178763-01; Transducer transducers (Accelerometers,
connection PWR, COM and SIG on 178897-01; Velomitors, or Proximitor Probe.)
connect 178897-01 to 178762-01; connect
individual channel on 178762-01 to individual Answer: This would require an additional
channels on 178763-01; individual channels TransducerPower Supply 168908-02 and (4 to
can use either the 1st 4 connector positions 6) 178762-01, (4 to 6) 178897-01 01 (each cable
with the Voltage Source switch in the (neg-) can take 4 channels so order accordingly,
position or use the 2nd 4 connector positions. 178763-01, use only the first 4 connector
Daisy chaining is possible up to 3 per channel. positions on 178763-01 and connect in the
same manner in Question #8. You can daisy
6. What do I need to order if I want to chain up to 3 channels per 178763-01 input.
power 9 to 24 negative bias 3-wire
transducers (Accelerometers,
Velomitors, or Proximitor Probe).

Answer: 178762-01 and 178897-01 (each cable


can take 4 channels so order accordingly),
178763-01 use all 8 connector positions and
connect in the same manner in Question #5.
You can daisy chain up to 3 channels per
178763-01 channel.

7. What do I need to order if I want to


power 1 to 4 positive bias 3-wire
transducers (Accelerometers,
Velomitors, or Proximitor Probe).

Answer: 178762-01, 178763-01, 178897-01;


Transducer connection PWR, COM and SIG on
178897-01; connect 178897-01 to 178762-01;
connect individual channel on 178762-01 to
individual channels on 178763-01 only using
channel inputs 1-4; Voltage Source switch in
the (pos +) position.

8. What do I need to order if I want to


power 5 to 12 positive bias 3-wire
transducers (Accelerometers,
Velomitors, or Proximitor Probe.)

Answer: (2 or 3) 178762-01, (2 or 3) 178897-01


01 (each cable can take 4 channels so order
accordingly), 178763-01, use only the first 4
connector positions on 178763-01 and
connect in the same manner in Question #7.
You can daisy chain up to 3 channels per
178763-01 input.

23/34
ADRE Sxp and 408 DSPi
Datasheet 172179 Rev. AN

Minimum Computer Requirements l HD webcam


ADRE Sxp software will run on most computer l Win 10 Pro 64-bit (includes Win 11 Pro
systems, desktop or notebook, provided that
the systems meet minimum specifications. This License)
datasheet provides recommended computer l Notebook Carry Case
requirements; inadequate specification will
impact software operation and system
performance and the ability for the system to 173M4496 High Performance
achieve maximum specifications. You must Notebook Computer
follow the “recommended computer
specifications” to realize the full system l Dell Precision 3580
performance.
4.8 Ghz Intel Core i7 Processor
Minimum Computer Specifications
l

l 32 GB RAM
l 2.4 Ghz or faster, Xeon
l 512GB PCIe x4 NVMe Storage
l 1.8GHz or faster, Dual Core (notebooks)
l USB 2.0 (or higher)
l 2 GB RAM
l 15.6” LED-backlit FHD + 1920x1080
l 160-250 GB HDD
l External HP DVD/RW Drive Included
l 1000/100 MB Ethernet
l HD webcam
l SXGA 128MB VRAM
l Windows 11 Professional 64-bit OS
ADRE Sxp Notebook Computer l Notebook Carry Case
This computer does not have an
external parallel port. You can also Computer Peripherals
use this computer with ADRE for
Users should evaluate the selection of
Windows software but it is not
additional external peripherals on a case-by-
compatible with the 208DAIU.
case basis. User needs can be very different
and therefore we do not recommend generic
169849 High Performance solutions. We can provide peripherals such as
printers, external hard drives, monitors, or other
Notebook Computer devices that best meet your needs. Please
provide your sales representative with specific
l HP ZBook Fury 16 G9 requirements.
l Core i7 12800HX / 2 GHz

l 32 GB RAM

l 1 TB PCIe® NVMe™ TLC SSD

l USB 4.0

l 15.6” LED-backlit FHD + 1920x1080

24/34
ADRE Sxp and 408 DSPi
Datasheet 172179 Rev. AN

Operating System Requirements & Database Access and Storage


Support The 408 DSPi can store 480 or 960GB of data
ADRE Sxp software is designed to run and fully internally. For applications that may need
tested on MicrosoftWindows 7, 8 and 10 significantly more storage capacity, an external
Professional (32bit/64bit) and Windows 2012 rack mounted drive array can also be added.
Server. Installation is allowed on Microsoft Permanent installations, test stands, or very
Windows Vista but is not fully tested for long term storage of raw streamed data can
compatibility.ADRE Sxp software cannot be easily be supported with an external drive
installed on earlier versions of Microsoft array.
Windowsoperating systems such as You can use the ADRE Sxp software to view all
Windows95/98/NT/XP. data on the 408 DSPi, thereby minimizing the
need to move large datasets over the network.
Networking Requirements Users can move entire databases or just the
ADRE Sxp software requires an available portion they need to a client computer.
Ethernet port with TCP/IP protocol support to Because moving large databases over a
communicate with the 408 DSPi. A 1000 MB (GB) network can consume significant bandwidth
Ethernet port is required for the 408 DSPi to and affect network performance we
meet full performance specifications. Under recommend that you use a dedicated or high
typical conditions with only one or two clients bandwidth network or a local (direct)
accessing the 408 DSPi, 100MB Ethernet will connection between the 408 DSPi and the ADRE
provide excellent performance. Sxp client computer.

Bandwidth sharing on any network, depending Data/Database Export


on traffic, is always a consideration and will
affect network performance. Ensure that all All versions of ADRE Sxp software provide the
networking hardware is installed and ability to export all data to a generic delimited
configured per network administrator’s spreadsheet format. Users can select specific
specifications. data types, static (by variable type), and/or
dynamic waveforms. Exported data can then
be imported into many 3rd party applications
for different types of analysis and/or
presentation needs.

Data/Database Import
The latest versions (3.1 greater) of the ADRE Sxp
software provide the ability to import vector
and waveform data from a generic delimited
spreadsheet format. Users can select a direct
static type, and asynchronous waveforms.
Imported data can be presented in the ADRE
Sxp software for different types of analysis
and/or presentation needs.

ADRE for Windows


ADRE Sxp software provides complete support
for existing ADRE for Windows databases. This
allows users to continue using existing ADRE for
Windows systems as needed while sharing
data with those running ADRE Sxp software.
Although ADRE Sxp does not communicate

25/34
ADRE Sxp and 408 DSPi
Datasheet 172179 Rev. AN
directly with the 208 DAIU, ADRE for Windows Display Plot / Formats
software can run on the same computer as
ADRE Sxp. Contact your local technical support The following plots can be selected from stored
representative for details specific to your data. Users can configure the display to show
needs. only the applicable plots for any given
application package.
Network and Archive Viewer Edition
Software l Current Values

There are two "Viewer Edition" options for ADRE l Tabular List
Sxp that are specifically designed to meet use
case scenarios that do not require active l Orbit / Timebase (w/ superposition of
control or configuration of a 408 DSPi. All overlay)
versions of ADRE Sxp can read all ADRE for
Windows databases. l Orbit (w/ superposition of overlay)
Network Viewer Edition-The Network Viewer l Timebase (w/ superposition of overlay)
allows a user to "connect to a 408" and view live
data or any archived data on the 408. Users l Continuous Raw Timebase (w/multi
need this if they wish to view live data, decimation)
participate in a live test, etc. The network viewer
can also display all archived data, either l Bode (w/ Forward & Reverse Vector
locally, or on the 408. This option is intended for transforms, coming soon)
applications that require remote participation
in data collection or viewing of live data, as well l Polar (w/ Forward & Reverse Vector
as local viewing of archive data. transforms)
Archive Viewer Edition – The Archive Viewer l Shaft Centerline (Average and
software is intended for situations when a user
requires support from other personnel, typically Instantaneous position & Orbit Overlay)
from a corporate support team or consultant. l X vs. Y
In such cases, support personnel will not
actually be configuring or collecting live data l Trend / Multivariable Trend
from the 408. Once a user has archived and
moved data from the 408, the Archive Viewer l Spectrum / Full Spectrum
provides all viewing features and data
manipulation capability. The Archive Viewer l Waterfall / Full Waterfall (w/ Campbell
allows users to open databases only after they format)
have been archived and moved to another
computer or placed on storage media. The l Cascade / Full Cascade (w/ Campbell
Archive Viewer is not able to connect to a 408 format)
at all.
l Structural Analysis (Timebase
w/Rectangular and Exponential
Windows)

l Structural Analysis Results (w/Transfer


Function, Coherence, Auto/Cross
Spectrum)

l System Event List

26/34
ADRE Sxp and 408 DSPi
Datasheet 172179 Rev. AN

Full Networ Archiv Full Networ Archiv


ADRE Sxp Client Versions ADRE Sxp Client Versions
Functio k e Functio k e
Functionality Functionality
n Viewer Viewer n Viewer Viewer

View historical data on √ √ √ Modify and Save √


local computer Configuration
Changes
View historical data √ √ √
stored on network On-the-fly √
server/drive Configuration
Changes
Import and view ADRE √ √ √
for Windows data Set Active √
Database/Configurati
Configure, save, and √ √ √ on
view plots and plot
sessions Modify System √
Properties
Export Static and √ √ √
Dynamic Data Update/Service 408 √

Build 408 √ √ √ 408 Diagnostics √


Configurations
Configure Network √
View Configurations √ √ √ Settings

Connect to 408 √ √

Upload Configurations √ √
to 408

View Real-time Data √ √


on 408

Download data from √ √


408 to local computer

View Network Settings √ √


on 408

Quick Configuration √ √
Software

408 DSPi Control and √


Operation

Enter & Exit Store √


Enable

Manual Sample √

Control Triggering √

27/34
ADRE Sxp and 408 DSPi
Datasheet 172179 Rev. AN

Software Licensing Documentation (can be ordered


ADRE Sxp software is available as a single
separately)
“computer” license. A separate license is
required for each installation of this product on 176559 ADRE Quick Start Guide
a different computer. Contact your local sales 172179 ADRE Sxp / 408 DSPi Datasheet
or service representative to purchase or
discuss “site” or “enterprise” licensing
requirements.

Software Technical Support


Agreement
The Software Technical Support Agreement
(TSA) allows you to contact our Product Service
department for assistance at any time during
the selected period of coverage. The support
agreement period begins with your initial
request for assistance, first software update, or
three months after your order, whichever
comes first. In addition to e-mail, fax, and
telephone support, the Support Agreement
provides free software updates as well as
updates via the Internet. Your Technical
Support Agreement ensures that you have
access to the most current version of software
with all the latest enhancements. Technical
support plans are available for single and
enterprise wide installation of our software
products. Contact your local sales
representative for details specific to your
needs.

Registration for ADRE Sxp Technical


Support
Registration of the ADRE Sxp software product is
the only way to activate your Technical Support
Agreement. Please contact your local Technical
Support representative or visit us online for
details specific to product registration.

Training Programs
Training is essential to ensure users are able to
realize full value from their tools in the most
efficient manner. Our technical training group
can provide training to meet your needs.
Training specific to ADRE Sxp and the 408 DSPi
can be provided at your facility or at one of
many Training Centers located around the
world. Contact your local sales representative
for details specific to your needs.

28/34
ADRE Sxp and 408 DSPi
Datasheet 172179 Rev. AN

Graphs and Figures

Figure 1: ADRE Sxp/408 DSPi Front View

Figure 2: ADRE Sxp/408 DSPi Rear View

29/34
ADRE Sxp and 408 DSPi
Datasheet 172179 Rev. AN

Figure 3: 408 DSPi with Optical KPH and Laser Tachometer speed input transducers.

Figure 4: 408 DSPi with Impact Hammer and Accelerometer.

30/34
ADRE Sxp and 408 DSPi
Datasheet 172179 Rev. AN

1. Standard terminal plug (included with 168908)


2. Modular wiring adapter (P/N 178763-01). Required for 178762-01.
3. SMA push-on quick-connect adapter (P/N 173887).
4. 4-connection standard wiring harness for 2- or 3-wire transducers (P/N 178775-01). Detail of
Standard Wiring Harness 178775-01 for detail.
5. 4-connection modular wiring harness for 2- or 3-wire transducers (P/N 178762-01). Detail of
Modular Wiring Harness 178762-01 for detail. Note: Modular wiring harness can be stacked only
with 3-wire voltage powered connections (will not work with constant current transducers).
6. Field wiring terminal kit (one 12-position terminal and jumpers P/N 178897-01).
7. Field wiring for 2-wire constant current transducers with negative bias.
8. Field wiring for 2-wire constant current transducers with positive bias.
9. Field wiring for 3-wire positive or negative voltage powered transducers.
Figure 5: Field Wiring Cables for Transducer Power and Signal Input

31/34
ADRE Sxp and 408 DSPi
Datasheet 172179 Rev. AN

1. Reverse side of connector.


2. Detail of wiring legend label.
Figure 6: Detail of Standard Wiring Harness 178775-01

1. Reverse side of connector.


2. Detail of wiring legend label.
Figure 7: Detail of Modular Wiring Harness 178762-01

32/34
ADRE Sxp and 408 DSPi
Datasheet 172179 Rev. AN

1. ADRE 408 Replay Module with Analog Outputs


2. ADRE 408 Analog Output to 8X BNC Cable
Figure 8: 408 DSPi with Replay Module and Analog Output to 8X BNC Cable

33/34
ADRE Sxp and 408 DSPi
Datasheet 172179 Rev. AN

Copyright 2025 Baker Hughes Company. All rights reserved.

Bently Nevada, ADRE, Keyphasor, M2, Proximitor, System 1, Velomitor and Orbit Logo are registered trademarks of
Bently Nevada, a Baker Hughes business, in the United States and other countries. The Baker Hughes logo is a
trademark of Baker Hughes Company. All other product and company names are trademarks of their respective
holders. Use of the trademarks does not imply any affiliation with or endorsement by the respective holders.
Baker Hughes provides this information on an “as is” basis for general information purposes. Baker Hughes does not
make any representation as to the accuracy or completeness of the information and makes no warranties of any
kind, specific, implied or oral, to the fullest extent permissible by law, including those of merchantability and fitness
for a particular purpose or use. Baker Hughes hereby disclaims any and all liability for any direct, indirect,
consequential or special damages, claims for lost profits, or third party claims arising from the use of the
information, whether a claim is asserted in contract, tort, or otherwise. Baker Hughes reserves the right to make
changes in specifications and features shown herein, or discontinue the product described at any time without
notice or obligation. Contact your Baker Hughes representative for the most current information.
The information contained in this document is the property of Baker Hughes and its affiliates; and is subject to
change without prior notice. It is being supplied as a service to our customers and may not be altered or its content
repackaged without the express written consent of Baker Hughes. This product or associated products may be
covered by one or more patents. See Bently.com/legal.

1631 Bently Parkway South, Minden, Nevada USA 89423


Phone: 1.775.782.3611 (US) or Bently.com/support
Bently.com

34/34

You might also like