BIOBASE Class B Table Top Autoclave BKMZB User Manual

The Class B Table Top Autoclave BKMZB user manual provides essential safety instructions, operational guidelines, and technical specifications for the sterilizer. It includes detailed sections on installation, operation, maintenance, and advanced settings, ensuring safe and effective use in medical environments. Users are advised to follow safety precautions and proper procedures to maintain the equipment's functionality and safety.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
50 views34 pages

BIOBASE Class B Table Top Autoclave BKMZB User Manual

The Class B Table Top Autoclave BKMZB user manual provides essential safety instructions, operational guidelines, and technical specifications for the sterilizer. It includes detailed sections on installation, operation, maintenance, and advanced settings, ensuring safe and effective use in medical environments. Users are advised to follow safety precautions and proper procedures to maintain the equipment's functionality and safety.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 34

Class B Table Top Autoclave

BKMZB
USER MANUAL

BIOBASE DISINFECTION(SHANDONG) CO., LTD.


Version 2022.08
Preface
Thank you very much for purchasing our class B series sterilizer.
Please read the “Operating Instructions” and “Warranty” before operating this unit to assure proper
operation. After reading these documents, be sure to store them securely together with the
“Warranty” at a hand place for future reference.

Warning: Before operating the unit, be sure to read carefully and fully understand important
warnings in the operating instructions.

1
Safety Instructions
Thank you for choosing our autoclave, please read the operation manual carefully before using it.
Safety Tips:

This sign indicates internal earthing protection of the machine.


Scald-proof

This sign can be seen after opening the door.


Important safety tip

This prompts the operator this part is the important safety information.

Security considerations:
Please read the safety warnings and precautions provided in this user guide to ensure that you use the
autoclave safely. If you do not follow the instructions provided in the manual, the protection offered
by the equipment may be damaged.
1. Use a separate protective earthing power socket of three holes 220V/10A, and make sure the
protective earthing terminal to a reliable connection.
2. Insert the autoclave power plug completely into the power socket, and do not use the power other
than those specified.
3. Do not use wet hands to pull the plug.
4. Do not damage, modify, stretch, or excessively bend or twist the power cord, but also do not put
heavy objects on the power cord.
5. Do not place the autoclave on an unstable table, such as a shaky table, ramps or the will-shake
position.
6. Do not block or cover the autoclave door, air vents, or heat-sink window.
7. Do not put anything on the autoclave.
8. If you find autoclave is smelly or unusual noisy in the using process, immediately turn off the
power. Then please contact your local dealer or our sales and service department
9. If the autoclave is not to be used for a period of time, the power cord plug should be power off
and remove the battery of the main board.
10. Do not place the autoclave near the heat source.
11. The LCD screen may work unstable if there is interference in the power grid when the equipment
is in use. The operator can equip it with a voltage regulator in order to solve the above problems.
12. Autoclave package (or product) must be accompanied by chemical indicator.
13. Autoclave is not suitable for sealing liquid, the product that is not resistant to high temperature
and high pressure.
14. Users should confirm the local altitude when using for the first time. If the altitude is more than
200 meters, the highland setting should be modified. Detailed operation please check the 6.3.2 on
page 8.

2
Content
1. Introduction ........................................................................................................................................4
2. Technical parameters ......................................................................................................................... 4
3. Product structure diagram ..................................................................................................................5
4. Installation ......................................................................................................................................... 5
5. Operation ........................................................................................................................................... 6
5.1 Preparation ................................................................................................................................ 6
5.2 Preparation of the equipment ....................................................................................................7
5.3 Sterilization procedure choice .................................................................................................. 7
5.4 Program operation .....................................................................................................................8
5.5 Test Program ........................................................................................................................... 10
6.Advanced Settings ............................................................................................................................ 11
6.1 Enter setup interface ................................................................................................................11
6.2 S1 State ................................................................................................................................... 11
6.3 S2 State ................................................................................................................................... 12
6.4 S3 state ....................................................................................................................................12
6.5 S4 State ................................................................................................................................... 13
6.6 Printer(Optional) .....................................................................................................................13
6.7 USB Port ................................................................................................................................. 14
7. Maintenance ...................................................................................................................................15
7.1 Cleaing the distilled water reservoir ....................................................................................... 15
7.2 Cleaning inner chamber .......................................................................................................... 15
7.3 Changing air filter ...................................................................................................................16
7.4 Cleaning door gasket and the mating surface ......................................................................... 16
7.5 Changing door gasket ............................................................................................................. 16
7.6 Replacing Fuse ........................................................................................................................17
8. Transportation and Storage ............................................................................................................ 18
9. Alarm ........................................................................................................................................... 18
10. Safety Device ...............................................................................................................................19
Appendix I Adding water suggestions ................................................................................................. 20
Appendix II Sterilization Procedure .................................................................................................... 21
Appendix III Main board diagram ....................................................................................................... 27
Appendix IV Piping Diagram .............................................................................................................. 28
Appendix V Packing List For Sterilizer ...............................................................................................29
Appendix VI Printer operating instructions .........................................................................................30
Company Information ..........................................................................................................................33

3
1. Introduction
This sterilizer is composed of control panel, sterilization chamber, sealing door.
In the sterilization process, the temp. error is 0~4℃ between Max temp. and preset temp. The timing
error should be less than 10% of the preset value. The sterilizer has 121 or 134℃set program.
This sterilizer has good sealing effect, the sterilization effect is accord with the standard 4.15 terms.
The electrical safety performance should be accord with the requirements of the registration product
standard appendix H & J.
The pressure volume is accord with the national standard GB 150 requirements.
Accord with the registration standards of BKM series small steam sterilizer.
Classify
By anti-electric shock type: I equipment
Equipment is not AP or APG type
By running pattern: continuous operation
Application range
It is mainly used in medical institution where users need to sterilize various surgical instruments,
dressings, glassware, and injection equipment.
Note: make sure the sterilizer is well grounded when using
Model BKMZB
Capacity(L) 23

2. Technical parameters
(1) Chamber Size: Φ247mmx470mm
(2) Power: AC220V±22V, 50/60Hz
(3) Consumption: 1950VA
(4) Sterilization Temp: 121℃/134℃
(5) Power Fuse: F10A/250V
(6) Water Consumption: 0.5L~2.5L
(7) Ambient Temp.: 5~40℃
(8) External Size (W*H*D): 700x495x410mm
(9) Net Weight: 53Kg
(10) Noise: <60dB
Use Environment
(11) Relative Humidity: Max 80%, no condensation
(12) Atmospheric Pressure: 76Kpa-106kpa

4
3. Product structure diagram

1. Distilled water tank 2. LCD screen 3. Button panel 4. Printer

5. USB data hole 6. Power switch 7. Distilled water outlet 8. Door handle

9. Condensate tank 10. Air filter 11. Safety valve 12.Power socket
drain

13. Insurance tube 14. Condenser 15. Label

4. Installation
* The autoclave must be placed in a horizontal work surface
* Leave at least 10cm around the autoclave, at the top of at least 30cm, around maintain good ventilation
* Don’t cover the sterilizer doors or vents, don’t place near the source of radiation
* Don’t put in the liquid splashing place
* Don’t be near any heat source

5
5. Operation
5.1 Preparation
5.1.1 Open the door; when the first time use, remove the instrument tray, instrument rack and other
accessories from the chamber, then remove the packaging.
5.1.2 Connect the power
Insert one end of the wire in the power socket and connect the other end to
the socket on the back of the autoclave

Note: The power supply outlet must be reliably grounded, connecting wires must be reliable
and secure.
5.1.3 Turn on the power
Press the power switch, after power is connected, LCD will show.
The system will display the position of the door, the water level, the
current program, date, time, etc.
If the system self-test found the problem, the alarm will be issued.
Picture description

Sterilization Procedure
Alarm code

Running curve

Temp.
Set code
Pressurere Time
Date

Flashing indicates water tank lack water

Flashing indicates water tank exceeds the max level

Flashing indicates condensate water tank exceeds the max level

Twinkling shows door unlocked, constant lighting shows door locked

Constant lighting indicates the output function of USB is working

6
Constant lighting indicates the print function is working

5.1.4 Adding distilled water


Removing the water storage cover of autoclave from its top as shown, you
can directly pour water into it as you can see water storage. When reach the
highest water level, Will twinkle with beep. Please stop adding
water.
Attention: water level should not over the vents of water storage.

5.2 Preparation of the equipment


In order to get a better effect of sterilization and protect sterilization items, please operate as follows:
1、make sure that different mental instrument puts on different mechanical disk, or isolate them
totally.
2、You’d better to insert sterilization paper and medical cotton between instruments and
mechanical disk in order to avoid being touch directly, if the material is not stainless steel.
3、Ensure every instrument get sufficient space
4、Containers like glass cup and tube etc should put down or invert for avoiding water
accumulation.
5、Every mechanical disk is not supposed to overload (please find in appendix)
6、Do not fold up mechanical disk for using, or place it directly in the cavity of the sterilizer
7、Please use handheld instrument every time you fetch instrument
8、Every instruments should be packed independently, if several devices must be packed in a
bag
9、Each package bag to be sealed
10、Ensure that sterilization with the paper side up, the plastic side down

Please try to sterilize after using sterilized package for extending storage time

5.3 Sterilization procedure choice


5.3.1 Interface display
Interface displays real-time temperature, pressure
Sterilization status, time, and the alarm code.

5.3.2 TEMP Button


Choose the temperature of sterilization
5.3.3 PROGRAM Button
Choose the program of sterilization
5.3.4 START/STOP Button
Using this button to start sterilization procedure, if push this button in 3second, you can terminate the
program. Pushing this button in 10 seconds after trouble solved can delete the display of wrong
equipment information.
5.3.5 choosing program

7
Pushing TEMP button to choose the temperature of sterilization, and PROGRAM button for
choosing the way the sterilization. The procedure can be seen in appendix 2.

Solid instrument

Packaging instrument

Fabric

Enhanced sterilization

B&D B&D TEST

HELIX HELIX TEST

VACUUM VANCUUM TEST

5.4 Program operation


Put items need to be sterilized into the cavity of sterilizer, please use handheld instrument
5.4.1 Close the door after items in, then revolving door handle clockwise to the maximum position

ATTENTION: Door handle must be to the right of the maximum position, otherwise the
door may be opened and the emergence of the door is forced to open
5.4.2 Start the sterilization process
The sterilizer will start to work automatically when start button pushed.
Total time is about 15-60 minutes (please find in appendix 2)

8
Hereby display total Time for
sterilization, When entering the
sterilization holding time or the
drying process, display the count
down.

Attention: when the door is not closed to its right position, will twinkle, at this time, the
starting program won’t work although starting button is pushed.
5.4.3 Sterilization finished
At the end of a sterilization, program will display END. If you are already connected to the printer,
then the printer will print sterilization information. When the screen displays the pressure to "0", you
can open the sterilizer door and remove sterilized items.

Make sure to use a handheld instrument tray gripping items from being burned.

9
5.4.4 Drain Valve

Waste Water
Tank joints

1 、 Insert the drain plug into 2、After the drain plug is inserted, turn the drain
the drain connector connector counterclockwise and the water is
automatically discharged.

3、After the drainage, turn the drain joint clockwise to seal


the drain.

Note: Ensure that the drainage pipe at the lowest position of the equipment when draining,
otherwise the water tank will not drain the water

5.5 Test Program


5.5.1 Press PROGRAM key,choose “B&D” test program。
5.5.1.1 put “B&D” test paper into the chamber of autoclave
5.5.1.2 Start the program and start working, at the end,
check the test paper to see whether reached sterilization.

5.5.2 Choose HELIX test program


5.5.2.1 put HELIX test paper into HELIX test tube。
5.5.2.2 Put HELIX test tube into the chamber of autoclave,and press
START to start program。
5.5.2.3 After sterilization is complete, check the test paper to see
whether it reach requirements

10
5.5.3 choose VACUUM test program
5.5.3.1 close the door,start program。
5.5.3.2 Results will be displayed after the sterilization.
5.5.3.3 According YY0646 requirements, within 10 minutes,
the air leakage rate of no more than 0.13kpa / min,
if meets the requirements, the interface will show "OK",
Otherwise, it displays "FAIL4", indicates failure

5.5.3.4 If during the test, the temperature fluctuations more than


3 degrees, "FAIL2" will show, representing the void.
You need to wait for sterilization chamber to cool, and then doing
it again.
Note: Vacuum test program must be run when the sterilization
chamber is in cold state. Otherwise there will be running more than
3 degrees temperature fluctuations Situation.

6. Advanced Settings
6.1 Enter setup interface
6.1.1 Press on START,open the main power。
After the buzzer sounds
Enter the setting interface.
6.1.2 press PROGRAM,choose setup state S1,S2,S3,S4
Press START to enter the relevant setup state。

6.2 S1 State
On S1 state,you can adjust time and date.
6.2.1 After entering the state, the default selected hours of high,
the bit display, press the PROGRAM key, the selected flash will
move to the next number.
6.2.2 Press TEMP, number flashing will add 1, and 0-9 cycles.
6.2.3Adjustment is completed, press the START button, the
adjusted value is saved.

11
6.3 S2 State
6.3.1 In this state, you can view Sterilization count
and equipment number, but you can't change it

6.3.2 Set the highland setting


If you use at high altitudes, you need to set the following
highland setting parameters:
1.0>h>0.5km, 0.5; 1.5>h>1.0km, 1.0
2.0>h>1.5km, 1.5; 2.5>h>2.0km, 2.0
3.0>h>2.5km, 2.5; 3.5>h>3.0km, 3.0
4.0>h>3.5km, 3.5; 4.5>h>4.0km, 4.0

Note: Standard atmosphere 100 KP, every rise by 0.5 km above sea level, the pressure drop 5 kpa.

If you set parameter value exceeds 2.0, you need to re-evaluate the sterilizing effect, you can extend
the time to improve sterilization Sterilization.

USB The factory code

6.3.3 Print and USB Settings


1 open the function, “1” open the function, “0” close the function 。 To the left of the parameter
indicates USB。The right of the parameters indicates the printer.

Users must not modify the factory code, the code set up and follow up by manufacturer, for
the maintenance

6.4 S3 state
6.4.1 This state is to adjust the sterilization time and drying time

Press PROGRAM to choose program.


Press TEMP to choose sterilization temperature.
Press START again to adjust the sterilization time and
drying time
6.4.2 Press TEMP to change value
Press PROGRAM to choose the value need to change.
6.4.3 press START to save and out.
6.4.4 sterilization time range 1-60
Drying time range 0-20

12
The adjustment of the drying time is referring to the adjustable part of the drying time.
Drying time points fixed drying time and adjustable drying time. Fixed time of about 7 minutes.

Note: Unless it is necessary, we do not recommend users to easily adjust parameters.

6.5 S4 State
S4 is set up for the manufacturer, the user must not to adjust the Settings.

6.6 Printer(Optional)
6.6.1 The printer is installed in the front of the
autoclave.
6.6.2 The sterilization data will be printed
automatically upon the selected program finishes.

Sample of the printed sterilization data

Program: WRAPPED Program: Vacuum test


Temperature:134 Tp: 1 °C
Pressure: 210.0kpa P1: -70kpa
Vacuum Num: 3 P2: -69.0kpa
Dry Time: 10Min rate of pressure rise:0.1kpa
Ster Time:4.0Min Start Time:08:22
Start Time: 05:38:12/88.1°C End Time:09:01
T1:05:40:03/84.7°C/-70.0kpa Date:2011-06-30
T2:05:42:23/101.8°C、52.0kpa Test Value : Success
T3: 05:44:15、79.4 °C/-70.0kpa S/N: BK000010
T4:05:46:21/108.5°C/50.3kpa Operator
T5:05:49:49/91.6°C/-70.0kpa

13
T6:05:57:34/134.5°C/229.3kpa
TS:134.5°C/225.2kpa ( Average temperature/ Average pressure )
MAX.Temperature:135.0°C ( Maxmum temperature )
MIN.Temperature: 134.0°C ( Minimum temperature )
MAX.Pressure:230.4kpa (Maxmum pressure )
MIN.Pressure:220.4kpa (Minimum pressure )
T7: 06:00:10/134.5°C/223.8kpa
T8:06:03:36/110.7°C/-60.2kpa
T9:06:06:22/102.3°C/-60.2kpa
End Time: 06:10:34/71.3°C
Cyele No: 0015 (Times of Sterilization)
Ster Value: Success (Sterilization Result)
Date: 2011-06-30
S/N: BK000010
Operator

6.7 USB Port


6.7.1 Plug the USB disk to the port on the front of the autoclave
6.7.2 Data transfer is completed when the screen indicates “END”.

Note: The disk must adopt FAT32. Hard disks are not accepted.

14
7. Maintenance
Maintenance Required Operation

Daily Cleaning the door gasket and the mating surface


Cleaning the outside of the door
Weekly Cleaning the water reservoir
Cleaning the inner chamber
Every 3 to 6 months Changing the air filter
(Every 200 operations)
Yearly Changing the door gasket

7.1 Cleaning the distilled water reservoir


Cleaning the distilled water reservoir by using medical disinfectant or distilled water
1.Drain the water reservoir out and wipe out the fouling by using no-shedding cloth dipped with
medical disinfectant or distilled water.

7.2 Cleaning inner chamber


7.2.1 Remove trays and racks
7.2.2 Clean up by using no-shedding cloth dipped with distilled water or medical disinfectant.
7.2.3 Clean the trays and racks using the same method as mentioned in 7.2.2

15
7.3 Changing air filter
7.3.1 Air filter is in the back of the autoclave (as shown in the picture)
7.3.2 Pull out the filter
7.3.3 Change the filter
7.3.4 Connect the new filter to the tube and push it back to the hole.

7.4 Cleaning door gasket and the mating surface


Clean up the door gasket and the mating surface using no-shedding cloth dipped with distilled water.

7.5 Changing door gasket


7.5.1 Open the door of the autoclave.
7.5.2 Remove the gasket
7.5.3 Clean up the mating surface
7.5.4 Wet the new gasket
7.5.5 Install the new one by following steps

1. Embed the 2 spots (up and down) equably into groove 2. Embed the 2 spots (right and left)equably into groove

16
3. Embed the other parts by diagonal way. 4. Press the gasket equably by hand.

Notice: Before changing the gasket, ensure that the sterilizer is cool, in order to
avoid scald.

7.6 Replacing Fuse


1. Turn off the main power supply 2. Unplug the power line, discharge the fuse holder with
a screwdriver

3. Replace a same type fuse 4. Install the fuse holder back

17
8. Transportation and Storage
8.1 Shut off the power switch, unplug the cord, and make the sterilizer been cooled down completely.
8.2 Drain water from reservoir and the condensate collector completely
8.3 Conditions For Transportation And Storage:
Temperature: -20℃~+55℃
Humidity: <85%
Atmospheric pressure: 50kpa-106kpa

9. Alarm

Error
Description Resolve
Code
Check whether the steam generator is
E01 Steam generator overheating
disconnected
Check whether the inner temperature sensor is
E02 Chamber overheating
disconnected
Check whether the exterior temperature sensor
E03 Exterior overheating
is disconnected.
E04 Heating failure Check whether leakage occurs.
E05 Pressure release failure Check whether the exhaust valve is blocked.
Check whether the door handle screw to the
E06 Door is opening during the cycle
maximum.
E07 Working overtime Check whether the exhaust valve is normal.
Check whether there is enough water in the
Constant temperature and pressure
E09 tank.
failure
Check whether leakage occurs.
Check whether vacuum pump and vacuum
E13 vacuum failure
pump valve work.
E20 Process is terminated by manual Turn off the power, and turn on again

18
10. Safety Device
1. Fuse
protect the entire device to prevent overload
Action: Cut off current
2. Thermal Fuse
Prevent short circuit and transformer overheating
Action: Cut off current of the transformer.
3.Safety Valve
Prevent overpressure
Action: Release pressure, when there is overpressure
4. Jiggle Switch
Ensure the door closed completely, avoiding safety risk.
5. Temperature Protector(Exterior)
Prevent overheating of Exterior heater
Action: Cut off current when the temperature is too high.
6. Temperature Protector (Steam Generator)
Cut off current when the steam generator temperature is too high.
7. Door safety Interlocking Device
Protect the user's security
Action: when there is pressure in the sterilizer chamber, the security interlock is in a lockdown
state to make sure the door is not open.

19
Appendix I Adding water suggestions

Item Water supply Condensation residue

Evaporated residue ≤10mg/L ≤1.0mg/kg

Silica ≤1mg/L ≤0.1mg/kg

Iron ≤0.2mg/L ≤0.1mg/kg

Cadmium ≤0.005mg/L ≤0.005mg/kg

Lead ≤0.05mg/L ≤0.05mg/kg

Other heavy metals ≤0.1mg/L ≤0.1mg/kg

Chloride ≤2mg/L ≤0.1mg/kg

Phosphate ≤0.5mg/L ≤0.1mg/kg

Electric conductivity ≤15μS/cm ≤3μS/cm

PH value 5-7.5 5-7


Colorless, clean, no Colorless, clean, no
Appearance
precipitation precipitation
Hardness ≤0.02mmol/L ≤0.02mmol/L

20
Appendix II Sterilization Procedure

Temperature Pressure maintain time Total time Max capacity of Max capacity
Procedure Instrument type
(℃) (kpa) (minute) (min) each tray(kg) (Kg)
Solid 134 210 4 15~20 1.3 4.5
Encapsulate solid load without package
instrument 121 110 20 30~45 1.3 4.5

Packaging 134 210 4 30~45 Hollow equipment without package 1 3


instrument 121 110 20 35~50 Single package solid instrument 1 3
Porous item without package 0.4 0.3
134 210 4 35~45 Porous goods with single packing 0.3 0.9

Fabric Porous goods with double packing 0.25 0.75


Hollow equipment with single packing 1.25 4.5
121 110 20 40~55 Solid or hollow load with double
0.6 1.5
package
Porous goods without packing 0.4 1.2
Porous goods with single packing 0.3 0.9
Strengthen
the 134 210 18 45-60 Porous goods with double packing 0.25 0.75
sterilization
Hollow equipment with single packing 1.25 4.5
Solid and hollow with double package 0.6 1.5
B&D test 134 210 3.5 22~38 — —
HELIX test 134 210 3.5 22~35 — —

21
Vacuum test — — — 15~20 — —

The highest temperature of 134℃ sterilization procedure is 137℃.

The highest temperature of 121℃ sterilization procedure is 124℃.

22
23
24
25
26
Appendix III Main board diagram

TP1: Steam Generator Temperature Sensor V2: Inlet Valve


TP2: Inner Chamber Sensor V3: Water Pump Valve
TP3: External Wall Temperature Sensor V4: Drain Valve
V1: Exhaust Valve V5: Vacuum Pump Valve

27
Appendix IV Piping Diagram

V1: Exhaust Valve


V2: Intake Valve
V3: Water Pump Valve
V4: Drain Valve
V5: Vacuum Pump Valve

28
Appendix V Packing List for Sterilizer
Sequence
Name Quantity Comment
Number
1 Equipment(Sterilizer BKMZB) 1

2 Instrument Tray 3

3 Instrument Rack 1

4 Instrument Clamp 1

5 Drain Pipe(60cm) 1

6 Specification 1

7 Fuse(F10A/AC250V) 1

8 Door Seal 1

Optional accessories

Sequence
Name Quantity Comment
Number
Consumables are 57mm receipt printing
1 Printer 1
paper

2 Filter 1 Consumable

29
Appendix VI Printer operating instructions
Name Diagram instruction
Open Printer Shield:”Push and push” switches
/ Inward push the upper middle position of
Push the printer shield to switch printer
Close
shield.When you open the printer shield,
Printer you can see the printer control panel.
Shield

Printer State Indicating Light:Light on,printer is


Control online; light go out,printer is not online.
Panel Paper Key:When the light on, press down
the paper key and start the paper.Loosen the
paper key and stop the paper.
Printer Inner Shell:”Push and push”
switches
Change First Step: Inward push the middle position
Scroll of the printer operation panel.The inner
Operati shell of the printer pops up automatically.
ng
Sequen
ce
Hand knead position
(sides)

Second Step: When the inner shell of the


printer popped up,pinch both sides and pull
out by hand and pulled in the end.

Third Step: Upwhirl as diagram until the


card is tight.

30
Fourth Step: Put the paper roll in the paper
axis, pinch both sides of the paper shaft, put
Feed Port
the paper roll into the paper warehouse,
loosen the paper shaft and then tighten it.

Paper Warehouse
Scroll
Reel

Note: Please pay special attention to the


direction of the paper roll and must follow
the diagram.

Fifth step: Turn on the power supply,when


the indicating light is bright, press the paper
key, put the paper in the feed port, and when
the paper is out, loosen the paper key and
stop the paper.

Sixth step: Pinch both sides of the inner


shell, downward rotate the printer inner
shell as picture shown, and inward push the
printer inner shell.

Seventh step: When the inner shell of the


printer is fully pushed into and tightly
packed with the outer shell,the preparation
is completed.

Note: The puffiness of the paper cause the


printer inner shell inconvenient loading,it
can even cause a jam when printing.

31
Open the lid from here
Change Ribbon First Step: Inward push the center area of
Tightness the printer control panel,and the inner shell
ribbon
Adjusting of the printer pops up automatically.
Operati Rotary Knob
ng Second Step: Open the lid, and then change
Sequen the ribbon.When finished, fasten the lid and
ce push into the rear.
Upper Cover

ribbon How to change ribbon: Press on the left


side of the ribbon,and the ribbon pops
up.Remove the ribbon and replace the new
ribbon.The new ribbon needs to be adjusted
manually (clockwise) and the knob is on the
right side of the ribbon.

Note: All parts of the equipment needed to be replaced must be supplied by the company.

32
BIOBASE DISINFECTION(SHANDONG) CO., LTD.
Olabo Intelligent Manufacturing Co., Ltd., No. 1 Boke Road, Ancheng Town, Pingyin
County, Jinan City, Shandong Province
Tel: +86-531-81307661
Inquiry: [email protected]
Complaints: [email protected]
After-sales service: [email protected]; [email protected]
Web: www.biobase.cc/www.meihuatrade.com / www.biobase.com