312 - Presentation1
312 - Presentation1
Instrumentation
• Significance of Measurement
Importance of Measurement is eloquently expressed in the statement
of Physicist Lord Kelvin: “I often say that when you can measure what
you are speaking about and can express it in numbers, you know
something about it; when you cannot express it in numbers, your
knowledge is of meagre and unsatisfactory kind.”
• Methods of Measurement
Direct Methods and Indirect Methods
Evolution of Instruments
• MECHANICAL Instrument: These instruments are very reliable for static and
stable conditions. But their disadvantage is that they are unable to respond
rapidly to measurements of dynamic and transient conditions.
• ELECTRICAL Instrument: It is faster than mechanical, indicating the output
are rapid than mechanical methods. But it depends on the mechanical
movement of the meters. The response is 0.5 to 24 seconds.
• ELECTRONIC Instrument: It is more reliable than other system. It uses
semiconductor devices and weak signal can also be detected.
Fundamental Concepts
Choosing Appropriate Measuring Instruments
2. Precision/Repeatability/Reproducibility
Precision is a term that describes an instrument’s degree of freedom
from random errors. Precision is often, although incorrectly, confused
with accuracy. High precision does not imply anything about
measurement accuracy. A high-precision instrument may have a low
accuracy. Low accuracy measurements from a high-precision instrument
are normally caused by a bias in the measurements, which is removable by
recalibration.
Repeatability describes the closeness of output readings when the same
input is applied repetitively over a short period of time, with the same
measurement conditions, same instrument and observer, same location,
and same conditions of use maintained throughout.
Fundamental Concepts
Static Characteristics of Instruments:
Reproducibility describes the closeness of output readings for the same
input when there are changes in the method of measurement, observer,
measuring instrument, location, conditions of use, and time of
measurement. Both terms thus describe the spread of output readings for
the same input. This spread is referred to as repeatability if the
measurement conditions are constant and as reproducibility if the
measurement conditions vary.
4. Range or Span:
The range or span of an instrument defines the minimum and maximum
values of a quantity that the instrument is designed to measure
Fundamental Concepts
Static Characteristics of Instruments:
5. Linearity
It is normally desirable that the output reading of an instrument is linearly
proportional to the quantity being measured. The Xs marked on Figure 8
show a plot of typical output readings of an instrument when a sequence
of input quantities are applied to it.
7. Threshold
If the input to an instrument is increased gradually from zero, the input will
have to reach a certain minimum level before the change in the instrument
output reading is of a large enough magnitude to be detectable. This
minimum level of input is known as the threshold of the instrument.
Fundamental Concepts
Static Characteristics of Instruments:
As an illustration, a car speedometer typically has a threshold of about 15
km/h. This means that, if the vehicle starts from rest and accelerates, no
output reading is observed on the speedometer until the speed reaches 15
km/h.
8. Resolution
When an instrument is showing a particular output reading, there is a lower
limit on the magnitude of the change in the input measured quantity that
produces an observable change in the instrument output. Like threshold,
resolution is sometimes specified as an absolute value and sometimes as a
percentage of f.s. deflection.
Using a car speedometer as an example again, this has subdivisions of
typically 20 km/h. This means that when the needle is between the scale
markings, we cannot estimate speed more accurately than to the nearest 5
km/h. This value of 5 km/h thus represents the resolution of the
instrument.
Fundamental Concepts
Static Characteristics of Instruments:
9. Sensitivity to Disturbance
As variations occur in the ambient temperature, certain static instrument
characteristics change, and the sensitivity to disturbance is a measure of the
magnitude of this change. Such environmental changes affect instruments
in two main ways, known as zero drift (or bias) and sensitivity drift.
Zero Drift/Bias: The mechanical form of a bathroom scale is a
common example of an instrument prone to zero drift. It is quite usual
to find that there is a reading of perhaps 1 kg with no one on the scale. If
someone of known weight 70 kg were to get on the scale, the reading
would be 71 kg, and if someone of known weight 100 kg were to get on
the scale, the reading would be 101 kg. Zero drift is normally removable
by calibration.
Sensitivity drift (also known as scale factor drift): defines the amount
by which an instrument’s sensitivity of measurement varies as ambient
conditions change.
Fundamental
Static Characteristics of Instruments:
Concepts
Example: A spring balance is calibrated in an environment at a temperature of
20℃ and has the following deflection/load characteristic:
Load (kg) 0 1 2 3
Deflection (mm) 0 20 40 60
It is then used in an environment at a temperature of 30℃, and the following
deflection/load characteristic is measured:
Load (kg) 0 1 2 3
Deflection (mm) 5 27 49 71
Determine the zero and sensitivity drifts per ℃ change in ambient temperature.
Solution:
At 20℃, deflection/load characteristic is a straight line. Sensitivity 20 mm/kg.
At 30C, deflection/load characteristic is still a straight line. Sensitivity =22 mm/kg.
Zero drift (bias) = 5 mm (the no-load deflection)
Sensitivity drift = 2 mm/kg
Zero drift/℃ = 5/10 = 0.5 mm/℃
Sensitivity drift/℃ = 2/10 = 0.2 (mm/kg)/℃
Fundamental Concepts
Static Characteristics of Instruments:
10. Hysteresis Effects
Figure 9 illustrates the output characteristic of an instrument that exhibits
hysteresis. If the input measured quantity to the instrument is increased
steadily from a negative value, the output reading varies in the manner
shown in curve A. If the input variable is then decreased steadily, the
output varies in the manner shown in curve B. The non-coincidence
between these loading and unloading curves is known as hysteresis.
Fig.10:Instrument characteristic
with dead space.
Fundamental Concepts
Static Characteristics of Instruments:
Some instruments that do not suffer from any significant hysteresis can still
exhibit a dead space in their output characteristics, however. Backlash in
gears is a typical cause of dead space and results in the sort of instrument
output characteristic shown in Figure 10. Backlash is commonly
experienced in gear sets used to convert between translational and
rotational motion (which is a common technique used to measure
translational velocity).
Fundamental Concepts
Dynamic Characteristics of Instruments:
The static characteristics of measuring instruments are concerned only with
the steady-state reading that the instrument settles down to, such as
accuracy of the reading.
The dynamic characteristics of a measuring instrument describe its
behaviour between the time a measured quantity changes value and the time
when the instrument output attains a steady value in response. As with
static characteristics, any values for dynamic characteristics quoted in
instrument data sheets only apply when the instrument is used under
specified environmental conditions. Outside these calibration conditions,
some variation in the dynamic parameters can be expected.
In any linear, time-invariant measuring system, the following general
relation can be written between input and output for time (t) > 0:
Equation 1.
Fundamental Concepts
Dynamic Characteristics of Instruments:
If we limit consideration to that of step changes in the measured quantity
only, then Equation (1) reduces to
Equation (2)
• Deviation: Deviation is departure of the observed reading from the arithmetic mean of the group of
readings.
• Standard Deviation: The standard deviation of an infinite number of data is defined as the square root
of the sum of the individual deviations squared divided by the number of readings.
• Problem Question:
The following 10 observations were recorded when measuring a voltage: 41.7, 42.0, 41.8, 42.0, 42.1,
41.9, 42.0, 41.9, 42.5, 41.8 volts. Calculate Mean, Standard Deviation, Probable Error and Range.
• Answer
Mean = 41.97 volt
S.D. = 0.22 volt
Probable error = 0.15 volt
Range = 0.8 volt.
Questions
• What is meant by measurement?
Measurement is an act or the result of comparison between the quantity and a Pre-
defined standard.
• Mention the basic requirements of measurement.
The standard used for comparison purpose must be accurately defined and should be
commonly accepted. · The apparatus used and the method adopted must be provable.
• What are the 2 methods for measurement?
Direct method and Indirect method.
• Define Instrument.
Instrument is defined as a device for determining the value or magnitude of a quantity or
variable.
• List the types of instruments.
The 3 types of instruments are · Mechanical Instruments · Electrical Instruments and ·
Electronic Instruments.