Use Manual
Use Manual
Instruction Manual
Translation of the original German instruction manual
CE mark: This is a Class I medical product according to the European Medical Device
Directive (MDD 93/42/EEC) and is compliant with the Directive version currently in force at
the time of product sale.
Conformity: The manufacturer declares conformity of this product with the essential
requirements indicated in Appendix I, MDD, as well as the performance of a conformity
evaluation procedure required for class I products according to Appendix VII, MDD and has
documented such with the CE mark.
The operating table models are identical in function, but they have
different equipment.
Contents
1 Important Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
1.1 Proper use of the equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
1.2 Explanation of symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
1.3 Safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
1.3.1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
1.3.2 Electricity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
1.3.3 High-frequency (HF) surgical devices and defibrillators . . . . . . . . . . . . . . . . . . .7
1.3.4 Pad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
1.3.5 Transport/repositioning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
1.3.6 Maintenance/repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
1.3.7 Protection against infection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
1.3.8 Environmental protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
1.4 Glossary. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2 Receiving Shipment, Unpacking and Setup . . . . . . . . . . . . . . . . . . . . . . . . . 11
3 Overview Operating Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.1 Symbols found on product . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.2 Audible signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4 Pads . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.1 Pad versions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.2 Removing pads . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
4.3 Installing pads . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
5 Control Units. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
5.1 Column keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
5.1.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
5.1.2 Visual indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
5.2 Remote control. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
5.2.1 IR remote control charging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
5.2.2 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
5.2.3 Visual indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
6 Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
6.1 AP safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
6.2 Internal power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
6.3 External (mains) power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
6.4 Charging the operating table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
6.5 Ready for use (power on) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
6.6 Standby mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
6.7 Shutdown . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
7 Coupling Points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
7.1 Definition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
7.2 TRUMPF coupling versions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
7.3 Coupling points on the operating tabletop . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
8 Radiolucence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .37
9 Patient Positioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .38
10 Load Capacity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .39
10.1 Conditions for using the operating table at loads of up
to 450 kg (1000 lb) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .40
10.2 Conditions for using the operating table at loads of up
to 225 kg (500 lb) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .42
10.3 Operating table loads of up to 160 kg (350 lb) . . . . . . . . . . . . . . . . . . . . . . .43
10.4 Standard rails . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .43
11 Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .45
11.1 Selecting the head position . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .46
11.2 Zero position . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .47
11.3 Tilt. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .48
11.4 Anti-/Trendelenburg . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .49
11.5 Lifting mechanism. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .50
11.6 Longitudinal displacement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .51
11.7 Leg section joints . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .53
11.8 Lower back section joints . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .55
11.9 Upper back section joints . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .56
11.10 Flex/Reflex . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .57
11.11 Replacing section segments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .58
11.12 Connector socket for electrical accessories . . . . . . . . . . . . . . . . . . . . . . . . . . .59
12 Travel Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .60
12.1 Brake . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .60
12.1.1 Lock the operating table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .60
12.1.2 Unlock the operating table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .61
12.1.3 Operating table emergency release . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .63
12.2 Directional travel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .64
12.2.1 Extending the wheel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .64
12.2.2 Drive mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .66
12.2.3 Retracting the wheel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .67
13 Care, Cleaning and Disinfection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .68
14 Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .70
14.1 Conditions for operation, storage and transport . . . . . . . . . . . . . . . . . . . . . . . .70
14.2 Operating table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .70
14.3 TITAN remote control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .72
14.4 Chargers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .72
15 Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .74
16 Maintenance and Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .76
17 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .77
18 EMC Data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .78
1 Important Information
TRUMPF Medizin Systeme GmbH + Co. KG products are
manufactured for a long and trouble-free service life. Development,
design, and production by TRUMPF Medizin Systeme GmbH +
Co. KG have been certified in accordance with
DIN EN ISO 13485:2003.
This guarantees:
• highest quality and a long service life,
• easy operation,
• functional design, and
• optimisation for the intended purpose.
The products are manufactured in accordance with the
requirements of the German Medical Device Directive
(Medizinproduktegesetz, MPG) and bear the CE mark. Prior to
using the products, read this instruction manual carefully from the
beginning to the end in order to familiarise yourself with the product
and its diverse functions in a step-by-step manner. Keep the
instruction manual near the product. This operating table can be
used safely. Reference is made to any residual dangers at the
appropriate points in the manual. Pay close attention to these
instructions!
DANGER
Risk of death (for example, electric shock)!
Refers to a directly imminent danger that will result in death or
serious injuries if the appropriate precautionary measures are not
taken.
WARNING
Risk of death!
Refers to a directly imminent danger that can result in death or
serious injuries if the appropriate precautionary measures are not
taken.
CAUTION
Risk of injury (for example, crushing)!
Refers to a possible danger that can lead to slight to moderate
injury or damage to the equipment if the appropriate precautionary
measures are not taken.
CAUTION
Risk of material damage.
Reference to a possible danger that can lead to equipment
damage if the appropriate precautionary measures are not taken.
NOTE
WARNING
Hazard to the patient!
Functional failure of the operating table cannot be completely
excluded given state-of-the-art technology currently available on the
market. Should this rare event occur, motor-driven operating table
functions will no longer be available during the operation.
If the electrical operating table functions fail, the patient can only be
moved with suitable aids (e.g., by placing a pad underneath the
patient) or by manually moving the patient from the operating table.
1.3.1 General
• Risk of material damage to electrical equipment due to
condensation! Do not immediately turn on the operating table.
When moving the operating table from a cold to a warm
environment or vice versa, condensation may occur on the inside
of the operating table. Before turning on the operating table,
place it in a room with the prescribed ambient temperature and
humidity (see chapter 14.1 “Conditions for operation, storage
and transport” on page 70) and allow it to warm up/cool down
for some time (at least 12 hours). The batteries must be
completely recharged after a significant change in temperature
or humidity.
• Check that all electrical and mechanical functions/parts of the
operating table, including accessories, are undamaged and in
good working order before using them.
Defective or damaged products may not be used.
• Hazard to the patient due to improper attachment.
Check that the panel segments and accessories are securely
fastened to the operating table before each use. Spontaneous
movement may occur and loose elements may slip out. Loosen
controls for adjustment purposes only. Clamp them down
immediately thereafter. Check that all fastening elements are
seated properly and tightly.
• Risk of crushing by moving parts! Familiarise yourself with its
operating functions before using the product.
• Risk of injury due to overhanging parts!
1.3.2 Electricity
• Get a qualified electrician to check the electrical safety of the
operating table and the power supply every year.
We recommend an annual general safety check by TRUMPF
customer service.
• The anti-static properties of the operating tabletop comply with
legal requirements if the provided standard/FoamLine pads are
used and properly secured.
• Always keep rechargeable batteries ready for service (i.e.,
charged) so they may be immediately used in the event of a
power loss.
• When charging the operating table during an operation, the
power cable may not be extended, as otherwise in
emergency mode1) the allowable limit of the patient leakage
current for operating on the heart (CF component according to
IEC 60601-1) will be exceeded. Use only the power cable
supplied! Use only the power cable supplied!
• Do not crush or roll over power and equipotential bonding
conductors. Do not use any damaged cables.
Disconnect cables before moving equipment.
If you have any concern about the safety of the power cables
and equipotential bonding conductors, work with the internal
(rechargeable) power supply only until you have replaced the
cable in question.
• Only use the operating table when an equipotential bonding
conductor is connected to an electrically conductive floor!
• Connection points for the equipotential bonding conductor on
the operating table comply with IEC 60601-1
(DIN EN 60601-1/VDE 0750-1).
1.3.4 Pad
• Placing any non-conductive positioning aids on the patient
surgery area eliminates AP safety on all TRUMPF operating
tables.
• The pads and cover retain the patient's body heat, thereby
preventing the body temperature from falling due to heat
diffusion.
• The pads are removable. Use the appropriate pads only
(manufactured by TRUMPF). Do not lay the patient on the
operating table without a pad. The edges of the pad are flush
with the operating table when operating table components are
fitted properly.
• Lift the patient into the desired position on the operating table
and do not pull the patient over the pads. After each
repositioning, lift the patient's affected body parts. Any wrinkles
or shearing forces that have developed will be eliminated.
• Do not place any sharp-edged objects on the pad or stick them
into the pad. Do not attach any adhesive foil.
• Always use only ONE pad version for the patient positioning
area (the tabletop plus the section segments). Note the different
properties of the pad versions (see chapter 4 “Pads” on
page 17).
1.3.5 Transport/repositioning
• Occupational safety for personnel: Given their heavy weight,
when lifting or carrying the panel segments and accessories, or
when manually moving the operating table with or without a
patient, take extra care not to strain your back. Work with an
additional person if necessary. Never couple or uncouple
multiple section segments or heavy, unwieldy accessories at the
same time. Hold the component securely and do not drop it.
• Always place the operating table in the zero position before
placing the patient on the tabletop.
• Secure the patient before adjusting the operating table and
provide active support during adjustments.
• Pay close attention to all electrically supported functions and
movements near the operating table to rule out any risks to the
patient or material damage to equipment, the operating
tabletop or furnishings.
• Perform all manual patient repositioning in a controlled and
responsible manner. Watch the movements of electrically
powered operating table adjustments carefully until the end
position has been reached.
• Pay particular attention to the operating table when lowering,
since the operating tabletop may collide with operating
accessories lying below it, or it may even contact the floor if it is
set in an extreme position. In some circumstances, there is a
possible risk of crushing people as well.
• When the zero position is approached on the operating table,
inclined patient positions may become steeper than they were
before the movement commenced.
Select the ZERO POSITION function only if there is no possibility
of collisions. Closely watch the table as it moves from the zero
position to the end position.
• Risk of tipping over! Before moving the operating table (with
the drive mode):
- Align swivel leg sections parallel to the longitudinal direction
of the operating table.
- Return any tilted position to a horizontal position.
- The tabletop is to be lowered until the appropriate pictogram
on the column cover is obscured (depending on patient
weight).
- Shut or disassemble any accessories.
• Risk of collision! Be especially careful when transporting the
operating table. Be careful to avoid collisions with nearby
persons or furniture to prevent damage to the table.
1.3.6 Maintenance/repair
• Keep a log book for the medical equipment. All services,
inspections, and repairs are to be documented in this book.
• All repairs must be performed in accordance with the safety
regulations as per IEC 60601-1 (DIN EN 60601-1/
VDE 0750-1).
• Within the bounds of the legal requirements, the manufacturer is
only then responsible for the technical safety characteristics of
this apparatus if the maintenance, repairs and modifications to
this apparatus are performed by TRUMPF service technicians or
by personnel authorised and trained by TRUMPF.
• TRUMPF shall not be liable for damage of any kind arising from
the failure to perform inspections or as a result of inadequate
repairs or maintenance, or modifications to the product.
1.4 Glossary
ITALICISED Descriptions of keys, functions,
text indicators and system components
Number in Numbering used in illustrations
brackets [50] (referencing text to graphics)
[a10] Identification of indicators on a control
unit
[i10] Identification of functions (keys) on a
control unit
[10] Identification of parts
+
corresponding
i1 function key
ON
- TITAN remote control
corresponding
+
i49function key
OFF/SHIFT
# Material number
CAUTION
Risk of material damage.
The packaging is not weatherproof! Store in accordance with the
“Technical Data”.
Unpack the operating table in an area with a level floor and with
sufficient free space for the apparatus to be rolled off the pallet.
Unpack in the following sequence:
5m
1. Place the pallet in a position where there is a free space of
about 1.5 m x 5 m (1.6 yd x 5.5 yd) available.
1,5 m
9. CAUTION
Risk of material damage to electrical equipment due
to condensation!
Do not immediately turn on the operating table. When moving
the operating table from a cold to a warm environment or vice
versa, condensation may occur on the inside of the operating
table. Before turning on the operating table, place it in a room
with the prescribed ambient temperature and humidity (see
chapter 14.1 “Conditions for operation, storage and
transport” on page 70) and allow it to warm up/cool down
for some time (at least 12 hours).
Turn on the operating table via the column keypad (press the
green ON button).
10. Hold down both the green ON button and the UNLOCK
button for at least 1 second. The operating table is unlocked.
11. CAUTION
Risk of personal injury. Risk of tipping over!
Do not go over the edge of the pallet!
Run the operating table down off the pallet under the control
of at least two persons.
H
ea
10
d
en
d
12
8 9
24
Fo
25
o
7 A
t
en
d
6 14
2 A 22
13 14
21
1
15 17
16
20
19
18
Key:
[1] Column keypad [2] Remote control
[3] Remote control charger [4] Seat section
[5] Back section, lower [6] Motorised leg section joint plus
[7] Standard rail for seat section Standard rail:
[8] Motorised lower back section joint Coupling point for wide double rail
[9] Standard rail for lower back section [10] Mechanical upper back section joint:
[11] OR seat Coupling point for narrow double rail
[12] Bayonet locking mechanism [13] Connector socket for the CONTROL UNIT
[14] Rechargeable battery compartment cover [15] Running gear
[16] Potential pin connection [17] Connector socket for the POWER SUPPLY
[18] Foot plate [19] Wheel
[20] Wheel for directional travel and drive [21] Crank for EMERGENCY RELEASE
[22] Insertion opening for EMERGENCY [23] Connector socket for ELECTRICAL
RELEASE ACCESSORIES
[24] Safety lever [25] Eccentric handle
IR 3D2 2x 2,5 AT
Connector socket for remote control To the right and left of the column
cables keypad
Rating plate
Key:
[A] Manufacturer
[B] Product name
[C] Material number
[D] Serial number
[E] Voltage 230 V~, Frequency 60 Hz, Current 1.25 A
[F] Mode of operation: Continuous operation with intermittent loading,
2 min ON, 8 min PAUSE
[G] Maximum approved patient weight
[H] Product tare weight
[I] Device is VDE certified as per DIN EN 60601-1
[K] Level of protection against water penetration
[L] Device has an internal power supply
[M] Device conforms with directive 93/42/EEC
[N] Level of protection against electrical shock: User part type B
[O] See instruction manual for information
[P] IR code the IR remote control
[R] Date of manufacture
4 Pads
CAUTION
Hazard to the patient due to improper attachment.
Check that the pad is securely fastened to the operating table. The
pad has to be attached precisely to the pad plate.
Standard pad Align the holes on the underside of the pad with the pad nubs, and
push down on the pad until it locks into place.
FoamLine pad The pads are velcroed to the pad plate. When you attach the pad,
be sure the pad's fleece straps are dry and lined up properly with
the Velcro strap on the pad plate. Position the pad on the pad plate
and press down widely.
5 Control Units
The following control units can be used to adjust the operating
table:
- Column keypad
- TITAN IR remote control
- TITAN cable remote control
- Foot control
The CONTROL UNIT port on the operating table is for connecting
the cable remote control or the foot control only.
Entries keyed into the individual control units are accepted in the
following order of priority:
1. Column keypad
2. Cable remote control
3. Foot control
4. IR remote control
Simultaneously pressing several keys leads to audible error signals
or to functions being executed in order of priority.
NOTE
5.1.1 Overview
The column keypad user interface has keys and displays. The
function of each key is identified by a corresponding pictogram.
i12 i13 i7 i8
i14 i15 i5 i6
i3 i4
Key:
[i1] Function: Turning on, display: READY FOR USE
[i2] Function: Shutdown
[i3] Function: Head position left, display: HEAD POSITION LEFT
[i4] Function: Head position right, display: HEAD POSITION RIGHT
[i5] Function: Raise
[i6] Function: Lower
[i7] Function: Trendelenburg
[i8] Function: Anti-Trendelenburg
[i9] Function: Tilt left
[i10] Function: Tilt right
[i11] Function: Zero position
[i12] Function: Raise leg section joints
[i13] Function: Lower leg section joints
[i14] Function: Raise lower back section joints
[i15] Function: Lower lower back section joints
[i16] Function: Longitudinal shift at head end, display: LD RANGE
[i17] Function: Longitudinal shift at foot end, display: No function
[i45] Function: Lock, display: OR TABLE LOCKED
[i46] Function: Jacking down
LD RANGE
illuminated red The tabletop is at the end of the –
longitudinal displacement range
i16 or in the extended longitudinal
displacement range.
OR TABLE LOCKED
illuminated green The operating table is locked. –
i45
Movement of operating table is
prevented
flashing green The operating table is in the –
directional travel mode.
You cannot push or move the
operating table sideways.
CAUTION
Risk of material damage.
Only use an original TRUMPF battery in the remote control. Please
contact Technical Customer Service at TRUMPF in the case of a
battery fault.
WARNING
Risk of personal injury.
All charger versions must be used outside of explosive areas.
CAUTION
Risk of material damage. Note the mains voltage.
Connect the charger only to the voltage indicated on the rating
plate (on the back side).
5.2.2 Overview
The remote control user interface consists of keys with three displays
at the top. The function of each key is identified by a corresponding
pictogram. For keys with dual functions, the second function is not
depicted as a symbol. The second function can be selected by
pressing the OFF/SHIFT key along with the corresponding function
key (two-key operation).
To adjust the operating table, press the function key until the desired
position is reached. The function stops when the key is released.
When the operating table is turned on, any function selected is
executed immediately.
CAUTION
Hazard to the patient if remote control is used
improperly!
Depending on the model, using the keypad improperly may
damage a key. A mechanically defective key spontaneously
triggers the corresponding operating table function on a continual
basis. If damage to a key is suspected, immediately discontinue use
of the remote control! If damage to a key was not noticed and the
table moves unintentionally, this incorrect movement can be
stopped by doing the following:
• push any function key on the column keypad or
• push the key OFF [i2] on the column keypad or
i2
• press two function keys simultaneously on the faulty remote
control unit or
• press the key OFF/SHIFT [i49] on the remote control
i49
Hold down the selected key while simultaneously removing the
defective remote control from the reception range of the operating
table. Package the defective operating table top securely and send
it back to TRUMPF in Saalfeld, Germany.
Key:
[a1] Display: BATTERY STATUS
a5 a1 a2
[a2] Display: BATTERY CHARGING
[a5] Display: READY FOR USE
i36 i37
i34 i35
[i5] Function: Raise
[i6] Function: Lower
i12 i13 [i7] Function: Trendelenburg
[i8] Function: Anti-Trendelenburg
i14 i15
[i9] Function: Tilt left
i17 i16
[i10] Function: Tilt right
[i11] Function: Zero position
i8 i7 [i12] Function: Raise leg section joints
[i13] Function: Lower leg section joints
i5 i6 [i14] Function: Raise lower back section joints
[i49]+[i14] Function: Raise joint at base holder, motorised
i9 i10
[i15] Function: Lower lower back section joints
i32 i33 [i49]+[i15] Function: Lower joint at base holder, motorised
[i16] Function: Longitudinal shift head end
i47 i48 [i46]+[i16] Function: Extended longitudinal shift, head end
[i17] Function: Longitudinal displacement, foot end
i45 i46 [i32] Function: Reflex (lower back section)
[i49]+[i32] Function: Reflex (leg section)
i11 i49
[i33] Function: Flex (lower back section)
[i49]+[i33] Function: Flex (leg section)
[i34] Function: Raise left leg section joint
[i35] Function: Lower left leg section joint
[i36] Function: Raise right leg section joint
[i37] Function: Lower right leg section joint
[i45] Function: Lock the operating table
[i46] Function: Unlock the operating table
[i47] Function: Drive, foot end
[i48] Function: Drive, head end
[i49] Function: Switch on and off
6 Power Supply
DANGER
Risk of death due to electric shock!
Do not use any electrical cables where it is evident that they have
been crushed or the insulation is damaged.
If you have any reservations about the power supply or the
equipotential bonding conductors, replace these cables
immediately.
The electronics monitor the power supply and the charged state of
the operating table batteries. The following are monitored:
- The charged state (of the rechargeable batteries)
- external power supply
- The charging current for the batteries
CAUTION
Risk of material damage.
Before removing the operating table from use for a considerable
amount of time, you must remove the batteries from the running
gear. Only Technical Customer Service or trained personnel may
remove the batteries.
6.1 AP safety
WARNING
Risk of explosion in the AP area!
Placing any nonconductive positioning aids on the patient surgery
area eliminates AP safety on all TRUMPF operating tables. Use only
anti-static pads on the operating table on anti-static flooring.
Prevent electrostatic charges!
When using the table's own rechargeable batteries (the internal
power supply), the operating table is AP safe according to
IEC 60601-1 and may be used in potentially explosive areas.
When powered (externally) with mains power, the operating table
is not AP safe. In potentially explosive areas, plug in the power
cable in emergencies only and always follow all steps for
connecting to and disconnecting from mains power (see page 30).
CAUTION
Hazard to the patient!
Electrical motorised functions are blocked when batteries are
empty or severely depleted. Connect the external power supply!
WARNING
Risk of explosion in the AP area!
Using the operating table via the mains power in potentially
explosive areas is permissible only when the user adheres to the
following procedure for mains connection and disconnection.
Connecting
1. Flip up the cover of the connector socket [17] on the running
17 gear [15] and insert the power plug into the socket [17].
The connection point is marked with a symbol for mains
15 connection.
2. CAUTION
Hazard to the patient!
Note the mains voltage. Only connect the operating table to
the voltage indicated on the rating plate.
Do not use extension cables. Extensions exceed the
permissible values for the patient leakage current. Use only the
power cable supplied!
Insert the mains power plug into an earthed wall socket
located outside of the room's AP area.
3. If the operating table was switched off, it is automatically
turned on and READY FOR USE [i1] is displayed on the
column keypad. The operating table is ready for use.
i1 4. Check whether there is a connection to the external power
supply. The EXTERNAL POWER SUPPLY [a3] display must be
illuminated on the column keypad.
Disconnecting
a3
CAUTION
Risk of personal injury.
While disconnecting from the mains power supply, do not make
any electrical motorised adjustments on the operating table.
1. Remove the mains power plug from the earthed wall socket.
The EXTERNAL POWER SUPPLY [a3] display on the column
keypad will disappear.
2. Disconnect the mains cable from the connector socket [17]
located on the running gear [15].
NOTE
CAUTION
Risk of personal injury.
Personnel may stumble over the mains connection cables and
equipotential bonding conductors. Do not crush or roll over
with the table.
2. Turn on the operating table
Press the green ON [i1] key on the column keypad. READY
FOR USE [i1] appears on the display, and after a few seconds
i1
an audible signal is heard. The operating table is ready only
after the signal is emitted. READY FOR USE [i1] remains
illuminated while the operating table is on.
When connected to an external power supply, the operating
table is already turned on.
NOTE
6.7 Shutdown
When connected to an external power supply, the operating table
is always turned on. The operating table can be turned off only
when running on battery power (the internal power supply). 5 days
after the last key was pressed on a control unit, the operating table
will shutdown to conserve battery power.
If you wish to shutdown the operating table immediately, hold down
i2 the orange OFF key [i2] on the column keypad for more than two
seconds.
7 Coupling Points
CAUTION
Risk of injury due to improper attachment!
Always ensure that the coupled parts are securely fastened to the
operating table. Check the connection before each use. Never
couple or uncouple more than one component at a time.
The coupling and uncoupling of accessory parts is described in the
respective instruction manuals.
7.1 Definition
All operating tabletops, section segments and accessory parts by
TRUMPF have coupling points from the fastening system developed
by TRUMPF. The coupling point is a separable connection site
between the operating tabletop and additional equipment. The
different coupling versions allow the operating table to be
assembled individually and with mechanical safety.
The coupling points ensure:
• Easy assembly/disassembly
• Reliable and play-free connections
• Different operating table configurations
NOTE
6 10 Head end:
12
9 - Mechanical joint [10] with a narrow double rail coupling point.
7 (Typically, the upper back section is attached to the back section.)
Foot end:
- Supports [11] for OR adapter
The operating table can be converted to an extension table with
the aid of an OR adapter and accessories. The OR adapter is
attached to the supports [11] (at the foot end, beneath the seat
section [4]) and secured with bayonet locks [12]. Read the
relevant instruction manuals for the proper and safe use of the
extension unit.
- Motorised joint [6] with a wide double rail coupling point
(Typically, two-piece leg sections or a pelvis extension is attached
to the seat section).
8 Radiolucence
There is total x-ray capability between the metal arms.
The column area is radiopaque. Avoid any pad creasing for a
minimum artefacts in X rays.
There are rails on the inside of the back section bars to hold an X-ray
cassette (size: 14").1) You can slide the X-ray cassette beneath the
seat section.
1) Optional
9 Patient Positioning
CAUTION
Hazard to the patient!
Extremities must not overhang the ends.
Cover the operating table with sterile material before use.
He
He
ad
ad
en
en
d
d
Fo
Fo
o
o
te
te
nd
NORMAL position: nd
INVERSE position:
The patient's head lies at the For certain positions, it is
head end of the tabletop. possible to design the operating
theatre table inversely. This
means that the patient lies the
other way round on the
operating theatre tabletop (with
the head on the foot end).
NOTE
The head position of the patient must match the display on the
column keypad. See chapter 11.1 “Selecting the head position” on
page 46.
10 Load Capacity
WARNING
Risk of personal injury and material damage when
permissible loads are exceeded.
The operating table may tip over when overloaded. In general,
overloading the operating table can lead to a failure in electrical
functions and cause material damage to mechanical parts. Do not
exceed the limits stated.
The maximum operating table load may only be reached if the
listed conditions are adhered to! The load consists of the weight of
the patient and the weight of the accessories attached.
Irrespective of the maximum tabletop load, do not exceed the
loading capacity of the coupled section segments and of the
accessories.
The instructions can be found in the relevant instruction manuals.
The values stated are valid for an evenly loaded operating tabletop
(surface), including accessories.
Before transporting the operating table, fold up or remove the
accessories.
10.1 Conditions for using the operating table at loads of up to 450 kg (1000 lb)
1. Option
450 kg
E
C A B
F F
G
WARNING
Potential physical injury due to the operating table
tipping over.
Do not enter the extended longitudinal displacement range [G].
The tabletop will stops automatically when the head end has
reached the end of the standard longitudinal displacement range.
2. Option
450 kg
E
C H B
F
G
10.2 Conditions for using the operating table at loads of up to 225 kg (500 lb)
225 kg
E
C A B
225 kg
E
C H B
M = m x l x 10
In the following table the torque is calculated depending on the
weight and distance for some standard situations
General calculation matrix
Weight Distance
10 cm 20 cm 30 cm 40 cm 50 cm
10 kg 10 Nm 20 Nm 30 Nm 40 Nm 50 Nm
20 kg 20 Nm 40 Nm 60 Nm 80 Nm 100 Nm
30 kg 30 Nm 60 Nm 90 Nm 120 Nm 150 Nm
40 kg 40 Nm 80 Nm 120 Nm 160 Nm 200 Nm
50 kg 50 Nm 100 Nm 150 Nm 200 Nm 250 Nm
11 Operation
DANGER
Risk of death due to electric shock!
Check potential connection. Use the operating table only with a
connected potential equalisation cable. The operating table has to
be connected to the equipotential bonding conductor in the
operating theatre.
WARNING
Risk of explosion in the AP area!
Prevent electrostatic charges!
Use only anti-static pads on the operating table on anti-static
flooring.
CAUTION
Hazard to the patient! Risk of collisions and jamming!
Watch all motorised movements of the operating table carefully
until the end position is reached. If the operating table is fully
equipped, not all of the coupled parts are detected by the
electronics. Collisions can thus not be ruled out entirely and must be
prevented by aborting the function.
Closely monitor all accessories attached in the column area when
adjusting the table position.
The seat and back section form a single unit and can be electrically
adjusted by the control units. They have coupling points for
additional components. The system's sensors report the important
settings of the tabletop, as well as its position. These data are
analyzed.
Before use, turn the operating table on and brake (lock into place).
Secure the patient before adjusting the operating table and provide
active support during adjustments. The head position of the patient
must match the display on the column keypad. Set the head position
at the column keypad.
To adjust the operating table, the corresponding function keys on
the control unit have to be pressed until the desired position is
reached. The function stops when the key(s) is released. When
lowering the tabletop segment from a position above the seat
section level, the segment stops automatically in the zero position
(signal). To lower below the zero position, briefly release and then
press the function key(s) again.
NOTE
The head position of the patient must match the display on the
column keypad.
The operating tabletop can be set up in the “reverse” direction for
certain patient positions where the patient needs to be lying at the
other end of the operating tabletop (with head at the foot end). The
column keypad displays do not match the patient's head position.
Some operating table functions (e.g. Trendelenburg, tilting) are then
i3 i4
reversed. In such cases, you must manually switch the head position
using the HEAD POSITION LEFT [i3] key to reflect the actual head
position. The appropriate indicator will then be illuminated. After
you have changed the head position, table movements will be
executed correctly. Switching the operating table off and back on
again, always restores the NORMAL position setting and the HEAD
POSITION RIGHT [i4] display is illuminated.
Adjustments
CAUTION
Hazard to the patient!
Monitor the patient's position. When approaching the zero
position, inclined patient positions may become steeper than they
were before the movement commenced!
1. Adjust the spreadable leg sections so that they are not on top
of one another. Position accessories to prevent collisions when
approaching the zero position.
2. Approach the zero position by pressing the following key(s):
Zero position +
i1 i11 i11
ON
*1
Two-key operation
11.3 Tilt
NOTE
Tilt left +
i1 i9 i9
ON
Tilt right +
i1 i10 i10
ON
*1
Two-key operation
NOTE
11.4 Anti-/Trendelenburg
NOTE
Anti-/ +
Trendelenburg i1 i8 i8
ON
Trendelenburg +
i1 i7 i7
ON
*1 Two-key operation
CAUTION
Hazard to the patient! Risk of collision!
Clear the area beneath the tabletop before lowering the tabletop.
Especially with the LOWER function, collisions may occur with
furnishings or devices beneath the tabletop and possibly covered
by draping.
When leg sections are positioned downward, collision with the
floor or running gear is possible when lowering the tabletop.
Carefully monitor the movement and stop the function before a
collision occurs. If necessary, angle the leg section(s) further
upward and then continue to lower the tabletop.
Press the following key(s) to move the table to the elevated position
you want:
Raise +
i1 i5 i5
ON
Lower +
i1 i6 i6
ON
*1 Two-key operation
Standard range
Adjust the longitudinal displacement by pressing the following
key(s):
Foot-end +
displacement i1 i17 i17
ON
Head-end +
displacement i1 i16 i16
ON
*1
Two-key operation
NOTE
Extended range
WARNING!
Potential physical injury due to the operating table
tipping over.
For NORMAL patient positions, use the extended longitudinal
displacement only for patient weights of up to 225 kg (500 lb).
Head-end
displacement
– +
i46 i16
*1
Two-key operation
When the tabletop is in the extended range, the LONGITUDINAL
DISPLACEMENT RANGE [i16] display next to the LD HEAD
key [i16] will be illuminated red.
i16
CAUTION
6
Hazard to the patient! Risk of collisions and jamming!
Pay close attention to any motorised movements.
When lowering the leg section(s), especially with the Anti-/
Trendelenburg and tilt settings, there is a risk of collision with the
column or the running gear.
6 Before you adjust the leg section joints, make sure that the
spreadable leg sections coupled to the leg section joints are not
above one another.
The two leg section joints [6] are motorised and can be adjusted
individually or jointly. Use the back section functions to tilt any
coupled section segments at the foot end up or down.
Sensors monitor and limit the adjustment range to ±90°.
Sensors check if a closed tabletop segment is connected at the
motorised adjustable coupling point. In this case adjustment of an
individual joint is blocked. In either case do not select individual joint
adjustment.
With head end tabletop positioning, selecting the RIGHT/LEFT LEG
SECTION JOINT DOWN and LEG SECTION JOINTS DOWN
functions limits the downward adjustment range.
*1
Two-key operation
Lower back +
section up i1 i14 i14
ON
Lower back +
section down i1 i15 i15
ON
*1
Two-key operation
When lowering from a position that is above the seat section level,
the process stops as soon as the section segment reaches zero
position. To lower below the zero position, briefly release and then
press the function key(s) again.
CAUTION
Risk of material damage.
If (one-piece) section segments are closed, adjust both joints on the
tabletop to the same angle. If set at different angles, the section
segment on the operating table will be skewed.
11.10 Flex/Reflex
The FLEX and REFLEX function is attained by executing two functions
at once, creating a bend between the seat and lower back sections
of the operating tabletop.
Flex: Function ANTITRENDELENBURG and
LOWER BACK SECTION JOINTS DOWN
Reflex: Function TRENDELENBURG and
LOWER BACK SECTION JOINTS UP
Adjustment is complete when the first final position has been
reached.
The flex/reflex position can be adjusted only via the remote control.
Hold down the respective function key on the remote control until
the operating table is in the position you require.
Reflex –
i32
Flex –
i33
Reflex – +
i49 i32
OFF/SHIFT
Flex – +
i49 i33
OFF/SHIFT
*1
Two-key operation
NOTE
Leg and back section joints without an option stop only at the
zero position point when travelling downward.
CAUTION
Risk of personal injury. Hazard to the patient!
Monitor the patient's position.
When you select the ZERO POSITION function, all electric,
adjustable movements of the operating table are automatically
returned to their defined start (zero) positions, one after the other.
Select the ZERO POSITION function only if there is no possibility of
collisions. Closely watch the table as it moves from the zero position
to the end position.
12 Travel Modes
Use the remote control and the column keypad to set the travel
mode (freewheeling, directional travel, directional travel plus drive)
and the brakes. In the freewheel mode, all four wheels on the
operating table can swivel freely. The operating table can be
pushed in any direction. An additional wheel is extended for
directional travel and directional travel plus drive. Axis rotation is not
possible, nor is transverse motion. Uni-directional steering is
supported.
Use two people to move the operating table longer distances.
NOTE
12.1 Brake
The brake function LOCK prevents the operating table from rolling
away. Foot plates emerge for all four wheels, which secure the
operating table in place.
CAUTION
The brake foot plate can crush cables on the floor.
Do not position the operating table across or on any cabling.
To lock the operating table, push the following key(s) for at least
1 second:
Lock the +
operating table i1 i45 i45
ON
*1
Two-key operation
When locked, you will hear an audible signal and the OR TABLE
LOCKED display on the column keypad next to the LOCK [i45] key
is illuminated.
i45
WARNING
Potential physical injury due to the operating table
tipping over during patient transport.
Unlock and move the operating table with a patient only according
to the following steps.
Selecting the UNLOCK function retracts the foot plate of all four
operating table wheels, and the operating table is once again
mobile.
1. Return any tilted position to a horizontal position.
NOTE
Tilt left +
i1 i9 i9
ON
Tilt right +
i1 i10 i10
ON
*1
Two-key operation
2. The tabletop is to be lowered until the appropriate pictogram
on the column cover is obscured (depending on patient
weight).
Lower +
i1 i6 i6
ON
*1 Two-key operation
Unlock the
operating +
i1 i46 i46
table ON
*1
Two-key operation
You will hear an audible signal after the table is unlocked. The
operating table is mobile and in the freewheel mode (all
wheels swivel). You can move the table to another location,
either manually or with the drive mode.
The operating table can also be unlocked with the following key
(hold down for at least 1 second) and is then in directional travel
mode:
*1
Two-key operation
Proceed as follows
14
1. Remove the battery compartment cover [14] on the running
gear (at the head end).
2. Take out the EMERGENCY RELEASE crank [21] and insert it
in the EMERGENCY RELEASE opening [22] found at the head
21
end of the running gear.
3. Rotate the EMERGENCY RELEASE crank [21] counter-
clockwise up to the stop. The operating table is unlocked.
22 4. Remove the EMERGENCY RELEASE crank [21] and return it
to the battery compartment. Return the cover [14] to the
running gear.
5. You can now move the operating table.
6. Contact Technical Customer Service at TRUMPF in Saalfeld,
Germany.
After manually unlocking the operating table, locking must also be
done manually.
21
Repeat the instructions for the emergency release, but rotate the
EMERGENCY RELEASE crank [21] not counterclockwise, but
clockwise up to the stop.
CAUTION
Risk of material damage.
When moving the operating table, do not tear out, roll over or crush
the power supply cables and equipotential bonding conductors.
NOTE
Tilt left +
i1 i9 i9
ON
Tilt right +
i1 i10 i10
ON
*1 Two-key operation
Lower +
i1 i6 i6
ON
*1 Two-key operation
Drive, foot
–
end i47
Drive, head
–
end i48
*1 Two-key operation
When the wheel is extended, you will hear an audible signal
and the OR table locked display on the column keypad next
to the LOCK key [i45] is blinking.
i45 6. The operating table can be moved manually or using the drive
mode. You cannot push or move the operating table
sideways.
Foot end –
i47
Head end –
i48
CAUTION
Risk of collision!
Drive mode supports only (straight-ahead) travel. The user must
manually steer the operating table.
Unlock the +
operating table i1 i46 i46
ON
*1
Two-key operation
The centre wheel [20] for directional travel and drive modes below
the running gear is retracted and an audible signal is emitted. The
operating table is now in the freewheel mode.
The operating table can also be directly transferred from directional
travel to a braked state via the LOCK button.
20
Adjustments Column keypad *1 Remote control
Lock the +
operating table i1 i45 i45
ON
*1
Two-key operation
CAUTION
Risk of material damage due to incorrect product
maintenance.
- The operating table is not machine-washable.
- Do not use a high-pressure cleaner. With high-pressure water jets,
fluids can penetrate the interior through gaps or openings and
damage the product (e.g. corrosion).
- Incorrect cleaning agents damage the surfaces.
Do not use any cleaners or disinfectants that are abrasive
or contain halogens or peracetic acid when caring for the
product in general. In addition, do not use any cleaners or
disinfectants containing alcohol or (flammable) solvents on
plastic parts and pads.
Cleaning agents and disinfectants have to meet all nationally
applicable regulations for the medical field and/or be in the list of
the DGHM1)/VAH2). Observe the hospital's hygiene regulations.
Clean soiled products immediately. Cleaning/disinfection is limited
to regularly wiping down (not submersing) equipment with suitable
substances.
CAUTION
Risk of personal injury.
Use of incorrect cleaning agents will remove the anti-static
properties of the pads in accordance with IEC 60601-1
(DIN EN 60601-1), and the product will no longer be AP safe.
Use the specified cleaning agents and disinfectants only.
14 Technical Data
NOTE
Adjustable ranges
Lifting mechanism 500 mm (19.7 in)
Trendelenburg ±40°
Tilt ±25°
Longitudinal shift Standard range 230 mm (9.0 in)
Extended range 70 mm (2.8 in)
Leg section joint ±90°
Lower back section joint + 85°/-45°
Upper back section joint + +90°/-55°
Classification
Safety class I
Device with internal power supply, IPS
Level of protection against User part type B
electrical shock Patient leakage current according to CF pursuant to
EN 60601-1
Level of protection against water IP X4
penetration With mains power supply IP X2
Level of protection when using Class AP with battery operation.
flammable mixtures in the area Mains power operation in potentially explosive areas
is permitted only when using electrically conductive
pads and the user follows the procedures for using the
operating table with an external power supply (see
chapter 6.3 “External (mains) power supply” on
page 30).
Mode of operation Continuous running with intermittent loading (DBAB)
2 min ON, 8 min OFF
Classification
Level of protection against water IP X4
penetration
14.4 Chargers
Dimensions including clip #4544122/#1711505/ (125 x 120 x 100) mm
(L x W x H) #1711506/#1711509 (4.9 x 4.7 x 3.9) in
#4544123/#1711507/ (125 x 67 x 85) mm
#1711508/#4544139 (4.9 x 2.6 x 3.3) in
Weight #4544122/#1711505/ approx. 550 g (19.4 oz)
#1711506/#1711509/
#4544123/#1711507/
#1711508/#4544139
Operating voltage #4544122/#1711505/ 230 V~
#1711506/#1711509/
#4544123/#1711507/
#1711508
Supply voltage #4544139 24 V DC/AC
Power consumption #4544122/#1711505/ max. 4 VA
#1711506/#1711509/
#4544123/#1711507/
#1711508
#4544139 max. 3.5 VA
Oscillator frequency 29...31 kHz
Battery charging current 15…22 mA
Frequency #4544122/#1711505/ 50/60 Hz
#1711506/#1711509/
#4544123/#1711507/
#1711508
#4544139 0/50/60 Hz
Classification
Safety class II
Level of protection against water IP X4
penetration
Level of protection when using No AP safety
flammable mixtures in the area
Mode of operation Continuous operation
15 Troubleshooting
Error Possible cause Fix
No response when pushing button Operating table is not switched on Press the green on button on
the column keypad
Operating table cannot be turned Column batteries are out of power Use mains power to operate
on the table and recharge column
batteries
No response when using the The green ON button on the Press the green ON button and
column keypad column keypad was not pressed the required additional function
at the same time (a two-button key on the column keypad at
operation) the same time.
No movement when button is Function is either not possible or
pressed, instead an audible error not permitted at that moment
signal
No response when pushing button, Column batteries are out of power Use mains power to operate
the BATTERY STATE display on the the table and recharge column
column keypad blinks batteries
Operating table cannot be Charging starts 5 minutes after the Contact Technical Customer
recharged, the BATTERY STATE last key has been pressed. If the Service
display on the column keypad is not display is not illuminated, then the
illuminated control electronics, mains cable or
input fuse (2x 2.5 AT) is defective.
Leg sections not adjustable far Longitudinal displacement too far Adjust longitudinal
enough down at head end displacement at foot end
Back sections not adjustable far Longitudinal displacement too far Adjust longitudinal
enough down at foot end displacement at head end
Leg and back sections do not stop Not a problem, sections are
at the zero position when travelling stopped only when travelling
upward downward
An audible signal while moving the Sensor error Contact Technical Customer
tabletop Service
Audible signal while levelling if the Possible risk of collision OK (no error)
leg sections touch one another (warning issued by the signal)
The ZERO POSITION button does OK
not move the leg and back sections
to a horizontal position when they
are in an upper position.
17 Disposal
Operating table, accessories, and packaging must be recycled in
an environmentally friendly fashion. Disposal, including that of
individual parts, must be environmentally friendly, i.e., in
accordance with the legal regulations currently in force. For
information on proper disposal of old equipment, please contact
either Technical Customer Service at TRUMPF, your local sales
representative or the appropriate national agency. TRUMPF will
take back your old equipment or products that are defective or no
longer used. Please contact Technical Customer Service in Saalfeld
for more information.
When the operating table is retired, the lead gel battery/batteries
must be removed from the operating table column by a TRUMPF
service technician or by person trained and authorised by TRUMPF.
Pack removed and unusable batteries in suitable packaging and
send them to TRUMPF Technical Customer Service. TRUMPF will
dispose of the batteries properly.
18 EMC Data
Table 1 according to IEC 60601-1-2:2007
The TITAN operating table is intended for use in the environments specified below. The customer or user of the aforementioned
device should ensure that it is operated in one of the environments as described.
RF emissions as per CISPR 11 Group 1 The TITAN operating table only uses RF energy
exclusively for its internal FUNCTION. Therefore, its RF
emissions levels are very low. It is improbable that
neighbouring devices would be adversely affected.
RF emissions as per CISPR 11 Class A The TITAN operating table is suitable for use in
establishments other than domestic and those connected
Harmonic Emissions as per IEC 61000-3-2 Class A directly to the PUBLIC LOW-VOLTAGE NETWORK that
supplies buildings used for domestic purposes.
Voltage fluctuations/flicker emissions as per In compliance
IEC 61000-3-3
NOTE
The device must not be used in direct proximity to any other devices. If such is
required, the device must be continually monitored to ensure its appropriate use
under these applied conditions.
The TITAN operating table is intended for use in the environments specified below. The customer or user of the aforementioned
device should ensure that it is operated in one of the environments as described.
Shock testing IEC 60601 test level Compliance level Electromagnetic environment -
guidelines
Electro-static discharge ±6 kV contact discharge ±6 kV contact discharge Flooring should be made of wood or
(ESD) as per ±8 kV air discharge ±8 kV air discharge concrete or have ceramic tiles. If the
IEC 61000-4-2 floor is covered with a synthetic
material, the relative air humidity must
be at least 30 %.
Rapid transient electrical ± 2 kV for mains cables ± 2 kV for mains cables The power supply quality should
interference variables/ ± 1 kV for input and output ± 1 kV for input and output correspond to that of a typical
bursts as per cables cables commercial or hospital environment.
IEC 61000-4-4
Surges as per IEC 61000- ±1 kV series mode voltage ±1 kV series mode voltage The power supply quality should
4-5 ±2 kV series mode voltage ±2 kV series mode voltage correspond to that of a typical
commercial or hospital environment.
The TITAN operating table is intended for use in the environments specified below. The customer or user of the aforementioned
device should ensure that it is operated in one of the environments as described.
Shock testing IEC 60601 test level Compliance level Electromagnetic environment -
guidelines
Voltage dips, short <5 % UT <5 % UT Mains power quality should be that of
interruptions and voltage (>95 % dip in UT) for ½ (>95 % dip in UT) for a typical commercial or hospital
variations on power cycle ½ cycle environment. If the user of the TITAN
supply input lines pursuant 40 % UT 40 % UT operating table requires continued
to IEC 61000-4-11 (60 % dip in UT) (60 % dip in UT) operation during power mains
for 5 periods for 5 periods interruptions, it is recommended that
70 % UT 70 % UT the TITAN operating table be
(30 % dip in UT) (30 % dip in UT) powered by an uninterruptible power
for 25 periods for 25 periods source or battery.
<5 % UT <5 % UT
(>95 % dip in UT) (>95 % dip in UT)
for 5 s for 5 s
Magnetic Field at Supply 3 A/m 3 A/m Magnetic fields for the network
Frequency of (50/60 Hz) frequency should comply with values
as per IEC 61000-4-8 commonly found in commercial and
hospital environments.
Recommended separation distances between portable and mobile RF telecommunications equipment and the TITAN
operating table
The TITAN operating table is intended for use in an electromagnetic environment where the RF disturbances are controlled. To
prevent electromagnetic interference, maintain a minimum distance between the portable and mobile RF communications
equipment (transmitters) and the aforementioned device according to the maximum output power of the communications
equipment, as recommended below.
150 kHz until 80 MHz 80 MHz until 800 MHz 800 MHz until 2.5 GHz
W
D=1.2√P D=1.2√P D=2.3√P
100 12 12 23
For transmitters with a maximum nominal power not found in the table above, the recommended safe distance (D) can be
calculated in metres (m) with the equation for the respective column. P is the maximum nominal power of the transmitter in
watts (W) as according to the transmitter manufacturer's data.
Note 1 At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2 These guidelines may not be applicable in all cases. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
Registered Office
Carl–Zeiss–Straße 7–9
07318 Saalfeld
Germany
Phone: +49–3671–5860
Fax: +49–3671–58641105
Austria
TRUMPF Medizin Systeme Österreich GmbH
France
TRUMPF AMSA SAS
United Kingdom
TRUMPF Medical Systems Ltd.
Italy
TRUMPF MED ITALIA s.r.l.
Australia
TRUMPF Med Australia
Singapore
TRUMPF Pte Ltd (Medical Division)
China
TRUMPF Medical Systems Ltd.
Russia
TRUMPF Med OOO
e–mail: [email protected]
www.trumpf.com