Creat Enz Labtest
Creat Enz Labtest
: 127
Instruction for use
Intended use . System for determination of creatinine in serum, Methodology . Enzymatic Trinder.
plasma, and urine samples by end-point reaction.
Reagents
Professional use.
Creatinine amidohydrolase The unopened reagents, when stored at indicated temperature, are stable
Creatinine + H2O Creatine up to the expiration date shown on the label. Upon handling, reagents and
the calibrator may be submitted do microbial or chemical contamination,
which may cause a reduction in reagent stability.
Creatine amidinohydrolase
Creatine + H2O Sarcosine + Urea Precautions and warnings
The usual security cares should be applied to the reagent handling. They
Sarcosine oxidase must not be pipetted by mouth aspiration. Avoid ingestion and in case of
Sarcosine + H2O + O2 Glycine + Formaldehyde + H2O2 contact with eyes, wash them with plenty of water and seek medical help.
Disposal of all biological waste material should be in accordance with * Sample and reagent volumes can be modified proportionally without
local guidelines. any loss in test performance, and the calculation procedure remains the
same. In case of volume reduction it is crucial to observe the minimal
necessary volume for photometric reading.
Interference
Concentrations of triglycerides up to 1000 mg/dL, bilirubin (conjugated Calibration
and unconjugated) up to 16 mg/dL, hemoglobin up to 400 mg/dL,
ascorbic acid up to 40 mg/dL, creatine up to 20 mg/dL, cefpiramide up to Automated systems
100 mg/dL, cefotaxime up to 100 mg/dL, and ceftraixone up to 2-point Calibration
Point 0: Reagent blank - deionized water or NaCl 150 mmol/L (0.85%).
100 mg/dL do not interfere in sample testing. Dobutamine and
Point 1: Calibrator - Calibra H Labtest series.
methyldopa interfere negatively in the reaction.
Samples with bilirubin, hemoglobin, and triglycerides levels higher than The creatinine concentration in the Calibra H material is traceable to the
the ones indicated above must be diluted with NaCl 150 mmol/L (0.85%) Standard Reference Material (SRM) 914 from the National Institute of
prior to being tested. Standards and Technology (NIST).
Samples with azide may yield inaccurate results for creatinine caused by Calibration frequency
insufficient creatine conversion. When the internal quality control indicates so;
When using a new reagent lot;
To determine the approximate concentration of hemoglobin in a sample, When using new bottle of reagent from the same lot if a new calibration
dilute 0.05 mL of sample in 2.0 mL of NaCl 150 mmol/L (0.85%) and has been performed for the prior reagent bottle.
measure the absorbance at either 405 or 415 nm, subtracting the zero
absorbance value with deionized water. Calculation . According to recommendations of NKDEP the results
must be reported with two decimal places to avoid systematic errors
caused by rounding of results, which may reach ±6%.
Procedure
DAbs of Test or Calibrator = Abs 2 - Abs 1
To determine the creatinine concentration in urine, dilute the sample 1:5
(0.2 mL of urine + 0.8 mL of NaCl 150 mmol/L). Multiply the result DAbs Test
obtained by 5. Creatinine (mg/dL) = x Calibrator conc. mg/dL
DAbs Calibrator
Urine Creatinine
U
Clearance = x MV (mL/minute) Serum/Plasma (mg/dL)*
S Newborn 0.31 - 0.92
2 weeks - 1 year 0.16 - 0.39
U: urine creatinine (mg/dL)
1 - <3 years 0.17 - 0.35
S: serum creatinine (mg/dL)
MV: minute volume (24-hour urine volume in mL divided by 1440). 3 - <5 years 0.26 - 0.42
5 - <7 years 0.29 - 0.48
Note: The clearance results must be corrected according to the patient's 7 - <9 years 0.34 - 0.55
body surface area, which is obtained via a nomogram that correlates 9 - <11 years 0.32 - 0.64
weight and height, or using the equation below: 11 - <13 years 0.42 - 0.71
0.425 0.725
13 - <15 years 0.46 - 0.81
A=W xH x 0.007184 Adults (women) 18 - 74 years 0.53 - 1.00
2 Adults (men) 18 - 74 years 0.70 - 1.20
A = body surface area (m )
W = weight (kg)
H = height (cm) * Intervals established for results traceable to the IDMS method.
Multiply the clearance value by 1.73 and divide the result by the patient's There are no intervals established for patients between 15 and 18 years
body surface area. old. It is suggested to use the intervals established for adult men and
women.
Glomerular filtration rate . The NKDEP strongly recommends
Conversion of mg/dL to SI units: mmol/L = mg/dL x 88.4
that laboratories report the estimated glomerular filtration rate (eGFR) for
all creatinine results.
When the results for plasma creatinine are traceable to the IDMS method,
Urine (mg/Kg/24 hours)
the following equations are applied, which use creatinine (CREA), age
2 - 3 years 6 - 22
(18 to 70 years) and sex.
> 3 years 12 - 30
Women Adults (women) 16 - 22
eGFR (mL/min/1.73m2) = 175 * (CREA) -1.1154 * (Age) -0.203 * 0.742 Adults (men) 21 - 26
Men
eGFR (mL/min/1.73m2) = 175 * (CREA) -1.1154 * (Age) -0.203 2
Creatinine Clearance (mL/min/1.73m )**
According to recommendations from the NKDEP, eGFR must be Children 70 - 140
reported as calculated value when the result is equal o less than Adults (women) 88 - 128
2
60 mL/min/1.73m . When the calculated value is higher than 60, Adults (men) 97 - 137
2
it must be reported as either higher than 60 mL/min/1.73m or
2 **Intervals established for results traceable to the IDMS method.
>60 mL/min/1.73m .
Operating interval . The reaction is linear between 0.0 mg/dL and The NKDEP recommends the calculation of glomerular filtration rate
150 mg/dL. For higher concentrations, dilute the sample with NaCl (eGFR) instead of creatinine clearance, using the creatinine result
150 mmol/L (0.85%), perform a new test, and multiply the result traceable to the IDMS method.
obtained by the dilution factor used.
Using the regression equation, the following systematic errors (bias) were Methodology sensitivity . A sample containing no creatinine was
found for the Enzymatic Creatinine method: used to evaluate the assay's detection limit. The value found was
0.19 mg/dL, which corresponds to the mean value of 9 assays plus two
standard deviations. Using the standard's absorbance as parameter, the
Decision levels Creatinine estimated Systematic errors
photometric detection limit was equal to 0.04 mg/dL, which corresponds
for creatinine using the regression estimated based on
to a difference in absorbance equal to 0.001.
evaluation equation creatinine decision levels
mg/dL mg/dL mg/dL %
Effect of matrix dilution . A sample with concentration of
1.00 1.00 0.0029 0.29
101.60 mg/dL was used to evaluate the system response to matrix
1.20 1.20 0.0024 0.20
dilution using NaCl 150 mmol/L. Using dilution factors ranging from 1.25
2.00 2.00 0.0003 0.015 to 5, the recovery values found were between 99.5% and 100.7%
Notes
For urine samples:
1. The material cleaning and drying are fundamental factors to the
reagent stability and to obtain correct results.
Comparative Enzymatic
Method Creatinine
2. The clinical laboratory is aimed at providing accurate and precise
Sample Nature Urine
results. Use of inappropriate quality water is a potential cause of analytical
Sample Number 50 50
bias. The water used in the laboratory should have the appropriate quality
Concentration interval for each application. Thus, to prepare reagents, use in the measures and
11.73 - 146.6 11.79 - 146.2
(mg/dL) for use in the final rinsing of the flasks, the water should have resistivity
Estimate mean (mg/dL) 134.87 134.41 ³1 megaohm.cm or conductivity £1 microsiemens/cm and silicate
Enzymatic Creatinine = 0.9956 x concentration <0,1 mg/L. When the deionizing column is with its
Regression equation
Comparative + 0.1070 capacity saturated, occurs release of several ions, silicates and
Correlation coefficient 0.9998 substances with large oxidation or reduction power that deteriorate the
reagents in a few days or even hours, thus changing the results
unpredictably. Consequently, it is essential to establish a quality control
Using the regression equation, the following systematic errors (bias) were program for the water.
found for the Enzymatic Creatinine method:
2. MYERS. G. L. et al. Recommendations for Improving Serum Creatinine Labtest Diagnóstica warrants the performance of this product under the
Measurement: A Report from the Laboratory Working Group of the specifications until the expiration date shown in the label provided that the
National Kidney Disease Education Program. Clinical Chemistry, v. 52, procedures and storage conditions indicated on the label and in this insert
n. 1, p. 5-18, 2006. have been followed correctly.
Enzymatic
Creatinine
127-1/216
()( 1 X 162 mL
1 X 54 mL
2 X 54 mL
127-2/72
Enzymatic Creatinine
Labmax 560/400
127-2/72
)() 2 X 18 mL
2 X 54 mL
2 X 18 mL
Liofilizado Corrosivo
Liofilizado Corrosivo
Lyophilized Corrosive
Instalar até
Instalar hasta
Install before Ref.: 140214