IN FLOWSIC600 en V5-0 2025.01 EHS
IN FLOWSIC600 en V5-0 2025.01 EHS
FLOWSIC600
Reliable flowmeter for
process applications
Compact – deal for custody transfer applications Standard – ideal for process gas measurement and inter-
nal billing
• The standard design is equipped with a fourpath • A reliable two-path design provides adequate mea-
system in which the measured values of the individual surement accuracy in non-custody applications.
sensor pairs are averaged.
4 FLOWSIC600
Intelligent diagnostics
Gas composition surveillance with permanent speed of Automatic Self-Diagnostics
sound measurement The FLOWSIC600 is equipped with a warning signal that
Thanks to a smart combination of the ultrasonic measure- informs the user of a malfunction before the measured
ment and electronics data interpretation, the flowmeter value is affected. To do this, current diagnostic indicators
provides a lot of information to the user. As the speed of are compared to indicators that were learned by the device
sound depends on media composition, its change indicates under normal measurement conditions. The results of
a deviation. This is permanently measured and reported to the self-diagnostics are permanently logged by an inter-
the user. nal electronic data log to allow a retroactive check of the
measurement process. Important measured variables such
as gas flow and gas velocity (VOG) or diagnostic parameters
such as signal-to-noise ratio (SNR) or speed can be read
directly.
FLOWSIC600
Reliable flowmeter for process applications
Product description
The FLOWSIC600 measurement de- low-maintenance, and has long-term
vice is an ultrasonic gas flowmeter and stability. The FLOWSIC600 features
sets the standard in its market seg- extensive diagnostics options to allow
ment. The design demonstrates that it detection of malfunctions even before
is geared to the tough industry condi- the measurement is affected and pro-
tions. The compact design with inte- vides process surveillance via speed of
grated cable routing means that the sound determination.
measuring system is durable, failsafe,
At a glance
• Highly efficient ultrasonic transduc- • Large measuring range 1 : 120
ers portfolio • Real-time speed of sound
• Bidirectional measurement measurement
• Low power consumption • Intelligent self-diagnostics
• Reduced and full bore design • Integrated log book and data logger
• Compact, robust design
Your benefits
• Long-term stability, reliable • Constant gas composition sur-
measurement veillance with speed of sound
• Wide application range measurement
• Immune to noisy conditions • Low maintenance with intelligent
• Reliable performance in harsh self-diagnostics
environment
Fields of application
• Process gases flow metering • For gases with high H2S component
• Industry gas metering CO2, H2, such as sour gas or biogas
Air, N2, O2, NH3, Ethylene, Helium, • Also for process gases
Argon • For gases with high H2S component
• Internal allocation measurement such as sour gas or biogas
Technical data
The exact device specifications and product performance data may vary and are dependent on the respective application
and customer specifications.
FLOWSIC600
Measured values Volumetric flow, a. c., volume a. c., gas velocity, sound velocity
Number of measuring paths 2, 4
Measurement principle Ultrasonic transit time difference measurement
Measuring medium Natural gas, air, C2H4, steam, process gases (e. g. CO2 up to 100%, N2, O2, Cl2
etc.), gases like sour gas or bio gas with high H2S content
Measuring ranges
Volumetric flow, a. c. 4 ... 360 m³/h / 1,600 ... 36,000 m³/h
(15 ... 1,271 ft³/h / 56,503 ... 1,271,700 ft³/h)
Measuring ranges depend on nominal pipe size
Repeatability < 0.1 % of the measured value
Accuracy
2-path version 1: ≤±1%
4-path version 1: ≤ ± 0.5 %
1
Typical accuracy, see application specific data for detailed information
Diagnostics functions Integrated device diagnosis and extended diagnosis via FLOWGATE software
Gas temperature
ATEX: –40 °C ... +105 °C (–40 °F ... +221 °F) at T1, T2, T3;
–40 °C ... +91 °C (–40 °F ... +196 °F) at T4
Other Ex certifications: –40 °C ... +180 °C (–40 °F ... +356 °F)
Operating pressure 0 bar (g) ... 250 bar (g) (0 psi (g) ... 3,626 psi (g))
On request: Up to 450 bar (g) (6,527 psi (g))
Nominal pipe size
2" ... 48" (DN 50 ... DN 1200)
Ambient temperature
ATEX, CSA: –40 °C ... +60 °C (–40 °F ... +140 °F)
IECEx: –40 °C ... +70 °C (–40 °F ... +158 °F)
Optional IECEx: –50 °C ... +70 °C (–58 °F ... +158 °F)
Storage temperature –40 °C ... +70 °C (–40 °F ... +158 °F)
Ambient humidity ≤ 95 %; relative humidity
Conformities ISO 17089-2
MTBF: 15.4 years
Ex-approvals
IECEx Gb/Ga Ex db eb ib [ia Ga] IIA T4
Gb/Ga Ex db eb ib [ia Ga] IIC T4
Ultrasonic transducers intrinsically safe
ATEX II 1/2 (1) G Ex ia/db eb ia [ia Ga] IIA/IIB/IIC T4...T1 Ga/Gb
III 1/2 (1) G Ex ia/db eb ia [ia Ga] IIC T6 Ga/Gb
Ultrasonic transducers intrinsically safe
NEC/CEC (US/CA) Class I, Division 1, Groups B, C, D T4
Class I, Division 2, Groups A, B, C, D T4
Class I, Division 1, Group D T4
Class I, Division 2, Group D T4
Ultrasonic transducers intrinsically safe
8 FLOWSIC600
Electrical safety CE
Enclosure rating IP 65 / IP 66 / IP 67
Analog outputs 1 output:
4 ... 20 mA, 200 Ω
Active/passive, electrically isolated
Digital outputs 3 outputs:
30 V, 10 mA
Passive, electrically isolated, Open Collector or according to NAMUR
(EN 50227), fmax = 6 kHz (scalable)
Interfaces RS-485 (2x, for configuration, output of measured values and diagnosis)
Bus protocol Modbus ASCII
Modbus RTU
HART
Operation Via meter display and software FLOWGATE
Dimensions (W x H x D) See dimensional drawings
Weight Depending on device version
Material in contact with media Low temperature carbon steel, stainless steel, duplex steel, titanium
Electrical connection
Voltage 12 ... 28.8 V DC
With active current output: 15 ... 28.8 V DC
Power consumption ≤1W
Gas flow meters 9
Measuring ranges
Ordering information
Our regional sales organization will help you to select the optimum device configuration.
10 FLOWSIC600
Dimensional drawings
Versions with nominal sizes from 3" up to and including 14" (cast and forged)
(dimensions in mm (inch))
206 (8.11)
D
A
C
Dimensions
Nomi- Con- Standard Weight Length Height (B) Flange Width of Internal
nal pipe nection (A) diameter measuring diameter
size flange (C) section (E)
(D)
[kg (lbs)] [mm (in)] [mm (in)] [mm (in)] [mm (in)] [mm (in)]
2" Cl. 150 ANSI B16.5 28 (61.73) 250 (9.84) 325 152.4 110 (4.33) 45 (1.77)
(12.80) (6.00)
Cl. 300 31 (68.34) 332 165.1
(13.07) (6.50)
Cl. 600 33 (72.75) 332 165.1
(13.07) (6.50)
DN 50 PN 16 EN 1092-1 30 (66.14) 332 165 (6.50)
(13.07)
PN 40 30 (66.14) 332 165 (6.50)
(13.07)
PN 100 35 (77.16) 347 195 (7.68)
(13.66)
3" Cl. 150 ANSI B16.5 37 (81.6) 240.0 344.0 190.0 180.0 73.0
(9.45) (13.54) (7.48) (7.09) (2.87)
Cl. 300 38 (83.78) 354.0 210.0
(13.94) (8.27)
Cl. 600 42 (92.59) 354.0 210.0
(13.94) (8.27)
Cl. 900 84 400.0 395.0 240.0
(185.19) (15.75) (15.55) (9.45)
DN 80 PN 16 DIN 2633 37 (81.6) 240.0 349.0 200.0
(9.45) (13.74) (7.87)
PN 63 DIN 2636 40 (88.18) 356.0 215.0
(14.02) (8.46)
PN 100 DIN 2637 43 (94.79) 364.0 230.0
(14.33) (9.06)
4" Cl. 150 ANSI B16.5 44 (97.00) 300.0 375.0 230.0 240.0 95.0
(11.81) (14.76) (9.06) (9.45) (3.74)
Cl. 300 55 388.0 255.0
(121.25) (15.28) (10.04)
Cl. 600 66 398.0 275.0
(145.50) (15.67) (10.83)
Cl. 900 99 500.0 408.0 290.0
(218.26) (19.69) (16.06) (11.42)
DN 100 PN 16 DIN 2633 40 (88.18) 300.0 372.0 220.0
(11.81) (14.65) (8.66)
PN 63 DIN 2636 52 405.0 250.0
(114.64) (15.94) (9.84)
PN 100 DIN 2637 61 410.0 265.0
(134.21) (16.14) (10.43)
12 FLOWSIC600
Nomi- Con- Standard Weight Length Height (B) Flange Width of Internal
nal pipe nection (A) diameter measuring diameter
size flange (C) section (E)
(D)
[kg (lbs)] [mm (in)] [mm (in)] [mm (in)] [mm (in)] [mm (in)]
6" Cl. 150 ANSI B16.5 100 450.0 445.0 280.0 300.0 142.0
(220.46) (17.72) (17.52) (11.02) (11.81) (5.59)
Cl. 300 110 465.0 320.0
(242.51) (18.31) (12.6)
Cl. 600 140 483.0 355.0
(308.65) (19.02) (13.98)
Cl. 900 220 750.0 496.0 380.0
(485.01) (29.53) (19.53) (14.96)
DN 150 PN 16 DIN 2633 90 450.0 448.0 285.0
(198.41) (17.72) (17.64) (11.22)
PN 63 DIN 2636 110 478.0 345.0
(242.51) (18.82) (13.58)
PN 100 DIN 2637 130 483.0 355.0
(286.60) (19.02) (13.98)
8” Cl. 150 ANSI B16.5 150 600.0 498.0 345.0 350.0 190.0
(330.69) (23.62) (19.61) (13.58) (13.78) (7.48)
Cl. 300 180 516.0 380.0
(396.83) (20.31) (14.96)
Cl. 600 210 536.0 420.0
(462.97) (21.1) (16.54)
Cl. 900 300 562.0 470.0
(661.39) (22.13) (18.5)
DN 200 PN 16 DIN 2633 140 498.0 340.0
(308.65) (19.61) (13.39)
PN 63 DIN 2636 190 535.0 415.0
(418.88) (21.06) (16.34)
PN 100 DIN 2637 210 543.0 430.0
(462.97) (21.38) (16.93)
10” Cl. 150 ANSI B16.5 240 750.0 548.0 405.0 410.0 235.0
(528.80) (29.53) (21.57) (15.94) (16.14) (9.25)
Cl. 300 250 568.0 445.0
(551.15) (22.36) (17.52)
Cl. 600 330 600.0 510.0
(727.86) (23.62) (20.08)
Cl. 900 470 625.0 545.0
(1,036.88) (24.61) (21.46)
DN 250 PN 16 DIN 2633 220 547.0 405.0
(485.01) (21.54) (15.94)
PN 63 DIN 2636 270 580.0 470.0
(595.25) (22.83) (18.5)
PN 100 DIN 2637 320 597.0 505.0
(705.48) (23.5) (19.88)
Gas flow meters 13
Nomi- Con- Standard Weight Length Height (B) Flange Width of Internal
nal pipe nection (A) diameter measuring diameter
size flange (C) section (E)
(D)
[kg (lbs)] [mm (in)] [mm (in)] [mm (in)] [mm (in)] [mm (in)]
12” Cl. 150 ANSI B16.5 350 900.0 586.0 485.0 470.0 270.0
(770.66) (35.43) (23.07) (19.09) (18.5) (10.63)
Cl. 300 400 605.0 520.0
(881.85) (23.82) (20.47)
Cl. 600 490 625.0 560.0
(1,080.28) (24.61) (22.05)
Cl. 900 720 685.0 610.0
(1,587.30) (26.97) (24.02)
DN 300 PN 16 DIN 2633 325 575.0 460.0
(716.50) (22.64) (18.11)
PN 63 DIN 2636 425 610.0 530.0
(937.0) (24.02) (20.87)
PN 100 DIN 2637 525 638.0 585.0
(1,157.4) (25.12) (23.03)
14” Cl. 150 ANSI B16.5 475 1,050.0 642.0 535.0 540.0 315.0
(1,047.2) (41.34) (25.28) (21.06) (21.26) (12.4)
Cl. 300 600 667.0 585.0
(1,322.8) (26.26) (23.03)
Cl. 600 675 677.0 605.0
(1,488.1) (26.65) (23.82)
Cl. 900 850 700.0 640.0
(1,873.9) (27.56) (25.2)
DN 350 PN 16 DIN 2633 475 635.0 520.0
(1,047.2) (25.0) (20.47)
PN 63 DIN 2636 625 675.0 600.0
(1,377.9) (26.57) (23.62)
PN 100 DIN 2637 750 705.0 655.0
(1,653.5) (27.76) (25.79)
For all meter ≥ 16" an installation length of 3D is optionally available
16” Cl. 150 ANSI B16.5 475 762.0 700.0 595.0 570.0 360.0
(1,047.2) (30.0) (27.56) (23.43) (22.44) (14.17)
Cl. 300 550 728.0 650.0
(1,212.5) (28.66) (25.59)
Cl. 600 640 745.0 685.0
(1,410.9) (29.33) (26.97)
Cl. 900 1,025 800.0 755.0 705.0
(2,250.0) (31.5) (29.72) (27.76)
DN 400 PN 16 DIN 2633 370 762.0 693.0 580.0
(815.7) (30.0) (27.28) (22.83)
PN 63 DIN 2636 600 738.0 670.0
(1,322.8) (29.06) (26.38)
14 FLOWSIC600
Nomi- Con- Standard Weight Length Height (B) Flange Width of Internal
nal pipe nection (A) diameter measuring diameter
size flange (C) section (E)
(D)
[kg (lbs)] [mm (in)] [mm (in)] [mm (in)] [mm (in)] [mm (in)]
18” Cl. 150 ANSI B16.5 660 820.0 754.0 635.0 620.0 405.0
(1,455.4) (32.28) (29.69) (25.0) (24.41) (15.94)
Cl. 300 760 792.0 710.0
(1,675.5) (31.18) (27.95)
Cl. 600 960 820.0 745.0
(2,116.0) (32.28) (29.33)
Cl. 900 1,300 900.0 830.0 785.0
(2,866.0) (35.43) (32.68) (30.91)
DN 450 PN 16 Data on request
20" Cl. 150 ANSI B16.5 750 902.0 815.0 700.0 670.0 450.0
(1,653.5) (35.51) (32.09) (27.56) (26.38) (17.72)
Cl. 300 930 853.0 775.0
(2,050.0) (33.58) (30.51)
Cl. 600 1080 872.0 815.0
(2,381.0) (34.33) (32.09)
Cl. 900 1500 1,000.0 892.0 855.0
(3,306.9) (39.37) (35.12) (33.66)
DN 500 PN 16 DIN 2633 700 902.0 823.0 715.0
(1,543.2) (35.51) (32.4) (28.15)
22" Cl. 150 Data on request
Cl. 300
Cl. 600
Cl. 900
DN 550 PN 16
24" Cl. 150 ANSI B16.5 1,090 991.0 927.0 815.0 760.0 540.0
(2,403.1) (39.02) (36.5) (32.09) (29.92) (21.26)
Cl. 300 1,390 978.0 915.0
(3,064.6) (38.5) (36.02)
Cl. 600 1,615 990.0 940.0
(3,562.1) (38.98) (37.01)
Cl. 900 2,100 1,200.0 1,040.0 1,040.0
(4,629.9) (47.24) (40.94) (40.94)
DN 600 PN 16 DIN 2633 1,015 991.0 940.0 840.0
(2,230.4) (39.02) (37.01) (33.07)
Gas flow meters 15
DN
A 1
≥ 5 DN ≥ 3 DN
≥ 2 DN ≥ 3 DN 1.5 ... 5 DN
B
3 1
1 FLOWSIC600
2 Temperature measuring point
3 Flow conditioner
2 2
DN
A
1
≥ 5 DN ≥ 5 DN
≥ 2 DN ≥ 3 DN ≥ 3 DN ≥ 2 DN
1 ... 5 DN 1 ... 5 DN
2 2
B
1
3 3
1 FLOWSIC600
2 Temperature measuring point
3 Flow conditioner
16 FLOWSIC600
Measured values Volumetric flow, a. c., volume a. c., gas velocity, speed of sound
Number of meas. paths 1+1 paths crossed (DN500 and DN600 2 paths in plane)
Measurement principle Ultrasonic transit time difference measurement
Measurement medium H2 (95 … 100%)
Repeatability ± 0.5% of the measured value (typical)
Accuracy (0.1 Qmax … Qmax) ≤ ± 2.0% (un-calibrated), ≤ ± 4.0% below 0.1 Qmax (typical)
Min. piping requirements Straight inlet section of ≥ 20D or ≥ 10D with flow conditioner
Operating pressure (typical) 5 … 100 bar(a) (72.52 ... 1450.38 psi(a))
(minimum pressure depending on line size and H2 content)
Gas temperature
ATEX: –40 °C ... +105 °C (–40 °F ... +221 °F) at T1, T2, T3;
–40 °C ... +91 °C (–40 °F ... +196 °F) at T4
Other Ex certifications: –40 °C ... +180 °C (–40 °F ... +356 °F)
Ambient temperature
ATEX, CSA: –40 °C ... +60 °C (–40 °F ... +140 °F)
IECEx: –40 °C ... +70 °C (–40 °F ... +158 °F)
Optional IECEx: –50 °C ... +70 °C (–58 °F ... +158 °F)
Ambient humidity ≤ 95% relative humidity; non-condensing
Conformities ISO17089-2
Ex-approvals CSA, ATEX, IECEx, Zone 1
Classification according IIC T4, intrinsically safe transducers
Electrical safety CE
Enclosure rating IP66 / IP67
I/O configuration 1x RS485, 1x Pulse, 2x Status Outputs, 1x Analog Output (4-20mA / HART)
2x RS485, 2x Pulse, 1x Status Outputs
Gas flow meters 17
Measured values Volumetric flow, a. c., volume a. c., gas velocity, sound velocity
Number of meas. paths 2 (line size 3-6 inch / DN80-150), 4 (line size ≥8 inch / DN200)
Measurement principle Ultrasonic transit time difference measurement
Measurement medium CO2 (95 … 100%), gaseous or super critical phase (single phase condition, not mixed)
Repeatability ± 0.1% of the measured value (typical)
Accuracy (0.1 Qmax … Qmax)
4-path meter (8 … 24 inch) ≤ ± 0.5% (un-calibrated), ≤ ± 1.0% below 0.1 Qmax (typical)
2-path meter (3 … 6 inch) ≤ ± 1.0% (un-calibrated), ≤ ± 2.0% below 0.1 Qmax (typical)
Min. piping requirements straight inlet section of ≥ 10D or ≥ 5D with flow conditioner
Operating pressure (typical) 0 … 100 bar (0 ... 1450.38 psi);
(100 … 250 bar (1450.38 ... 3625.95 psi) on request)
Gas temperature
ATEX: –40 °C ... +105 °C (–40 °F ... +221 °F) at T1, T2, T3;
–40 °C ... +91 °C (–40 °F ... +196 °F) at T4
Other Ex certifications: –40 °C ... +180 °C (–40 °F ... +356 °F)
Ambient temperature
ATEX, CSA: –40 °C ... +60 °C (–40 °F ... +140 °F)
IECEx: –40 °C ... +70 °C (–40 °F ... +158 °F)
Optional IECEx: –50 °C ... +70 °C (–58 °F ... +158 °F)
Ambient humidity ≤ 95% relative humidity; non-condensing
Conformities ISO17089-2
Ex-approvals CSA, ATEX, IECEx, Zone 1 and Zone 2 Classification according IIA or IIC T4, intrinsi-
cally safe transducers
Electrical safety CE
Enclosure rating IP66 / IP67
I/O configuration 1x RS485, 1x Pulse, 2x Status Outputs, 1x Analog Output (4-20mA / HART) or
2x RS485, 2x Pulse, 1x Status Outputs
18 FLOWSIC600
Oxygen [O2]
Measured values Volumetric flow, a. c., volume a. c., gas velocity, sound velocity
Number of meas. paths 2 (line size 3-4 inch / DN80-150), 4 (line size ≥6 inch / DN200)
Measurement principle Ultrasonic transit time difference measurement
Measurement medium O2
Repeatability ± 0.1% of the measured value (typical)
Accuracy (0.1 Qmax … Qmax)
4-path meter (8 … 24 inch) ≤ ± 0.5% (un-calibrated), ≤ ± 1.0% below 0.1 Qmax (typical)
2-path meter (3 … 6 inch) ≤ ± 1.0% (un-calibrated), ≤ ± 2.0% below 0.1 Qmax (typical)
Min. piping requirements straight inlet section of ≥ 10D (use of flow conditioner is not allowed in O2)
Operating pressure (typical) 0 … 40 bar(g) (0 ... 580.15 psi(g));
(>40 bar (>580.15 psi) on request with limitation of vmax = 8 m/s (26.25 ft/s))
Gas temperature
ATEX: –40 °C ... +105 °C (–40 °F ... +221 °F) at T1, T2, T3;
–40 °C ... +91 °C (–40 °F ... +196 °F) at T4
Other Ex certifications: –40 °C ... +180 °C (–40 °F ... +356 °F)
Ambient temperature
ATEX, CSA: –40 °C ... +60 °C (–40 °F ... +140 °F)
IECEx: –40 °C ... +70 °C (–40 °F ... +158 °F)
Optional IECEx: –50 °C ... +70 °C (–58 °F ... +158 °F)
Ambient humidity ≤ 95% relative humidity; non-condensing
Conformities ISO17089-2
Ex-approvals CSA, IECEx, Zone 1 and Zone 2
Classification according IIA or IIC T4, intrinsically safe transducers
Electrical safety CE
Enclosure rating IP66 / IP67
I/O configuration 1x RS485, 1x Pulse, 2x Status Outputs, 1x Analog Output (4-20mA / HART)
2x RS485, 2x Pulse, 1x Status Outputs
Gas flow meters 19
Nitrogen [N2]
Measured values Volumetric flow, a. c., volume a. c., gas velocity, sound velocity
Number of meas. paths 4
Measurement principle Ultrasonic transit time difference measurement
Measurement medium N2
Repeatability ± 0.1% of the measured value (typical)
Accuracy (0.1 Qmax … Qmax) ≤ ± 0.5% (un-calibrated), ≤ ± 1.0% below 0.1 Qmax (typical)
Min. piping requirements ISO17089-1&2: with straight inlet section of ≥ 10D or ≥ 5D with flow conditioner
Operating pressure (typical) 0 … 100 bar (0 psi ... 1450.38 psi); (>100 bar (>1450.38 psi) on request)
Gas temperature
ATEX: –40 °C ... +105 °C (–40 °F ... +221 °F) at T1, T2, T3;
–40 °C ... +91 °C (–40 °F ... +196 °F) at T4
Other Ex certifications: –40 °C ... +180 °C (–40 °F ... +356 °F)
Ambient temperature
ATEX, CSA: –40 °C ... +60 °C (–40 °F ... +140 °F)
IECEx: –40 °C ... +70 °C (–40 °F ... +158 °F)
Optional IECEx: –50 °C ... +70 °C (–58 °F ... +158 °F)
Ambient humidity ≤ 95% relative humidity; non-condensing
Conformities ISO17089-2
Ex-approvals CSA, IECEx, Zone 1 and Zone 2
Classification according IIA or IIC T4, intrinsically safe transducers
Electrical safety CE
Enclosure rating IP66 / IP67
I/O configuration 1x RS485, 1x Pulse, 2x Status Outputs, 1x Analog Output (4-20mA / HART)
2x RS485, 2x Pulse, 1x Status Outputs
20 FLOWSIC600
Ammonia [NH3]
Measured values Volumetric flow, a. c., volume a. c., gas velocity, sound velocity
Number of meas. paths 4
Measurement principle Ultrasonic transit time difference measurement
Measurement medium NH3
Repeatability ± 0.1% of the measured value (typical)
Accuracy (0.1 Qmax … Qmax) ≤ ± 0.5% (un-calibrated), ≤ ± 1.0% below 0.1 Qmax (typical)
Min. piping requirements ISO17089-1&2: with straight inlet section of ≥ 10D or ≥ 5D with flow conditioner
Operating pressure (typical) 0 bar … 100 bar (0 psi ... 1450.38 psi) (in gaseous and supercritical phase)
Gas temperature
ATEX: –40 °C ... +105 °C (–40 °F ... +221 °F) at T1, T2, T3;
–40 °C ... +91 °C (–40 °F ... +196 °F) at T4
Other Ex certifications: –40 °C ... +180 °C (–40 °F ... +356 °F)
Ambient temperature
ATEX, CSA: –40 °C ... +60 °C (–40 °F ... +140 °F)
IECEx: –40 °C ... +70 °C (–40 °F ... +158 °F)
Optional IECEx: –50 °C ... +70 °C (–58 °F ... +158 °F)
Ambient humidity ≤ 95% relative humidity; non-condensing
Conformities ISO17089-2
Ex-approvals CSA, IECEx, Zone 1 and Zone 2
Classification according IIA or IIC T4, intrinsically safe transducers
Electrical safety CE
Enclosure rating IP66 / IP67
I/O configuration 1x RS485, 1x Pulse, 2x Status Outputs, 1x Analog Output (4-20mA / HART)
2x RS485, 2x Pulse, 1x Status Outputs
Gas flow meters 21
Ethylene [C2H4]
Measured values Volumetric flow, a. c., volume a. c., gas velocity, sound velocity
Number of meas. paths 4
Measurement principle Ultrasonic transit time difference measurement
Measurement medium C2H4
Repeatability ± 0.1% of the measured value (typical)
Accuracy (0.1 Qmax … Qmax) ≤ ± 0.5% (un-calibrated), ≤ ± 1.0% below 0.1 Qmax (typical)
Min. piping requirements ISO17089-1&2: with straight inlet section of ≥ 10D or ≥ 5D with flow conditioner
Operating pressure (typical) 0 bar … 100 bar (0 psi ... 1450.38 psi) (in gaseous and supercritical phase)
Gas temperature
ATEX: –40 °C ... +105 °C (–40 °F ... +221 °F) at T1, T2, T3;
–40 °C ... +91 °C (–40 °F ... +196 °F) at T4
Other Ex certifications: –40 °C ... +180 °C (–40 °F ... +356 °F)
Ambient temperature
ATEX, CSA: –40 °C ... +60 °C (–40 °F ... +140 °F)
IECEx: –40 °C ... +70 °C (–40 °F ... +158 °F)
Optional IECEx: –50 °C ... +70 °C (–58 °F ... +158 °F)
Ambient humidity ≤ 95% relative humidity; non-condensing
Conformities ISO17089-2
Ex-approvals CSA, IECEx, Zone 1 and Zone 2
Classification according IIA or IIC T4, intrinsically safe transducers
Electrical safety CE
Enclosure rating IP66 / IP67
I/O configuration 1x RS485, 1x Pulse, 2x Status Outputs, 1x Analog Output (4-20mA / HART)
2x RS485, 2x Pulse, 1x Status Outputs
22 FLOWSIC600
Argon [Ar]
Measured values Volumetric flow, a. c., volume a. c., gas velocity, sound velocity
Number of meas. paths 4
Measurement principle Ultrasonic transit time difference measurement
Measurement medium C2H4
Repeatability ± 0.1% of the measured value (typical)
Accuracy (0.1 Qmax … Qmax) ≤ ± 0.5% (un-calibrated), ≤ ± 1.0% below 0.1 Qmax (typical)
Min. piping requirements ISO17089-1&2: with straight inlet section of ≥ 10D or ≥ 5D with flow conditioner
Operating pressure (typical) 0 bar … 100 bar (0 psi ... 1450.38 psi)(in gaseous and supercritical phase)
Gas temperature
ATEX: –40 °C ... +105 °C (–40 °F ... +221 °F) at T1, T2, T3;
–40 °C ... +91 °C (–40 °F ... +196 °F) at T4
Other Ex certifications: –40 °C ... +180 °C (–40 °F ... +356 °F)
Ambient temperature
ATEX, CSA: –40 °C ... +60 °C (–40 °F ... +140 °F)
IECEx: –40 °C ... +70 °C (–40 °F ... +158 °F)
Optional IECEx: –50 °C ... +70 °C (–58 °F ... +158 °F)
Ambient humidity ≤ 95% relative humidity; non-condensing
Conformities ISO17089-2
Ex-approvals CSA, IECEx, Zone 1 and Zone 2
Classification according IIA or IIC T4, intrinsically safe transducers
Electrical safety CE
Enclosure rating IP66 / IP67
I/O configuration 1x RS485, 1x Pulse, 2x Status Outputs, 1x Analog Output (4-20mA / HART)
2x RS485, 2x Pulse, 1x Status Outputs
Gas flow meters 23