Mammography
Mammography
Mammography - Chapter 8
1. Introduction
BME HCMUT 1
Mammography – Chapter 8 17-Feb-20
1. Introduction
Continuing refinements in technology have vastly improved mammography
over the last 15 years
1. Introduction
Mammographic features characteristic of breast cancer are:
BME HCMUT 2
Mammography – Chapter 8 17-Feb-20
1. Introduction
Screening Mammography – Identify
Cancer
the AMA, ACS and ACR recommend
a baseline mammogram by age 40,
biannual examinations between ages
40 and 50, and yearly examinations
after age 50
NCI recommends women in their 40s,
50s and older should be screened
every one to two years with
mammography
Require craniocaudal (CC) and
mediolateral oblique (MLO) views of CC MLO
each breast
1. Introduction
Diagnostic Mammography –
Evaluate Abnormalities
may require additional
views, magnification views,
spot compression views,
stereotactic biopsy or other
studies using other
modalities
BME HCMUT 3
Mammography – Chapter 8 17-Feb-20
MRI
has wonderful tissue contrast sensitivity
useful for evaluating silicone implants
accurately assess the stage of breast cancer involvement
UW - Dr. Constance
Lehman
7
1. Modern Mammography
BME HCMUT 4
Mammography – Chapter 8 17-Feb-20
1. Modern Mammography
Mammography equipment
Low contrast sensitivity
high resolution
low dose
1. Modern Mammography
Dedicated Mammography
Equipment
Breast Compression
Devices
Optimized Screen/Film
detector systems
10
BME HCMUT 5
Mammography – Chapter 8 17-Feb-20
11
11
Self-filtration
12
BME HCMUT 6
Mammography – Chapter 8 17-Feb-20
13
13
14
BME HCMUT 7
Mammography – Chapter 8 17-Feb-20
Added tube filters of the same element as the target reduce the
low- and high-energy x-rays in the x-ray spectrum and allow
transmission of characteristic x-ray energies
Rh/Rh
Mo/Rh
15
15
16
16
BME HCMUT 8
Mammography – Chapter 8 17-Feb-20
20 - 23 keV effective
energy leading to increased
penetration of thick and/or
dense breasts
17
17
c.f. Bushberg, et al. The Essential Physics of Medical Imaging, 2nd ed., p. 201.
18
18
BME HCMUT 9
Mammography – Chapter 8 17-Feb-20
19
19
1. Collimation
20
20
BME HCMUT 10
Mammography – Chapter 8 17-Feb-20
2. X-Ray Generator
21
21
c.f. Bushberg,
et al. The
Essential
Physics of
Medical
Imaging, 2nd
ed., p. 205.
22
22
BME HCMUT 11
Mammography – Chapter 8 17-Feb-20
23
23
2. Technique Chart
24
24
BME HCMUT 12
Mammography – Chapter 8 17-Feb-20
3. Compression
25
3. Compression
Use a low attenuating lexan paddle attached to a compression
device
10 to 20 newtons (22 to 44 pounds) of force is typically used
A flat, 90°paddle (not curved) provides a uniform density image
Parallel to the breast support table
Principal drawback of compression is patient discomfort
26
BME HCMUT 13
Mammography – Chapter 8 17-Feb-20
3. Compression
27
3. Scatter Radiation
28
28
BME HCMUT 14
Mammography – Chapter 8 17-Feb-20
3. Scatter Radiation
29
29
3. Magnification
Advantages
Magnification of 1.5x to
2.0x is used
Increased effective
resolution of the image
receptor by the
magnification factor
Small focal spot size used
Reduction of scatter
30
BME HCMUT 15
Mammography – Chapter 8 17-Feb-20
3. Magnification
Disadvantages
Geometric blurring caused
by the finite focal spot size
(more on cathode side)
Breast dose in general
similar to contact
mammography
Long exposure times (small
focal spot, low mA)
patient motion and blur
31
32
BME HCMUT 16
Mammography – Chapter 8 17-Feb-20
33
33
34
34
BME HCMUT 17
Mammography – Chapter 8 17-Feb-20
Bucky factor of 2 to 3
35
35
4. Screen/Film Cassettes
36
36
BME HCMUT 18
Mammography – Chapter 8 17-Feb-20
4. Screen/Film Cassettes
Limiting spatial
resolution is =
15-20 lp/mm
(0.025 - 0.030
mm object size)
37
4. Film Processing
38
38
BME HCMUT 19
Mammography – Chapter 8 17-Feb-20
4. Film Sensitometry
39
39
4. Film Sensitometry
40
BME HCMUT 20
Mammography – Chapter 8 17-Feb-20
4. Film Sensitometry
41
42
42
BME HCMUT 21
Mammography – Chapter 8 17-Feb-20
43
44
44
BME HCMUT 22
Mammography – Chapter 8 17-Feb-20
45
45
Question
A. Mo/Mo
B. Mo/Rh
C. Rh/Rh
D. W/Al
E. W/Rh
46
46
BME HCMUT 23
Mammography – Chapter 8 17-Feb-20
Question
A. 10, 12
B. 15, 16
C. 17, 20
D. 24, 26
E. 59, 69
47
47
Question
48
48
BME HCMUT 24
Mammography – Chapter 8 17-Feb-20
Question
49
49
5. Radiation Dosimetry
50
50
BME HCMUT 25
Mammography – Chapter 8 17-Feb-20
5. Radiation Dosimetry
51
51
5. Radiation Dosimetry
52
BME HCMUT 26
Mammography – Chapter 8 17-Feb-20
5. Radiation Dosimetry
53
5. Radiation Dosimetry
54
54
BME HCMUT 27
Mammography – Chapter 8 17-Feb-20
5. Radiation Dosimetry
The average glandular dose for this phantom is typically 1.5 to 2.2
mGy per view or 3 to 4.4 mGy for two views for a film optical density
of 1.5 to 2.0
55
55
56
BME HCMUT 28
Mammography – Chapter 8 17-Feb-20
57
57
Average glandular breast dose limited to 3 mGy or 300 mrad per film
for a compressed breast thickness of 4.2 cm, 50/50
glandular/adipose breast composition
Increasing kVp reduces dose
Risk of mammogram induced breast cancer is far less than the risk
of developing breast cancer
58
58
BME HCMUT 29
Mammography – Chapter 8 17-Feb-20
59
59
60
BME HCMUT 30
Mammography – Chapter 8 17-Feb-20
c.f. www.gehealthcare.com/rad/xr/education/dig_xray_intro.html
61
61
Advantage
Wide dynamic range
(1000:1) compared with
SFM (40:1)
Dynamic image
manipulation
Ability to post-process
Soft-copy read
accompanied by computer-
aided-diagnosis (CAD)
3D imaging
Radiographics 2004:24,1749 62
62
BME HCMUT 31
Mammography – Chapter 8 17-Feb-20
Advantage
Each component of the mammographic process can be
optimized with digital mammography
Radiographics 2004:24,1750 63
63
Radiographics 2004:24,1750 64
64
BME HCMUT 32
Mammography – Chapter 8 17-Feb-20
Radiographics 2004:24,1751 65
65
Indirect Capture
a scintillator such as cesium
iodide (CsI) absorbs x-rays
and generates a light
scintillation
detected by an array of
photodiodes or charge-
coupled devices (CCDs)
Resolution degradation
http://www.hologic.com/wh/pdf/R-LM-016_Radiology_Management.pdf 66
66
BME HCMUT 33
Mammography – Chapter 8 17-Feb-20
Direct Capture
X-ray photons are directly
captured by a
photoconductor such as
amorphous selenium (a-
Se), which converts the
absorbed x-rays directly to a
digital signal
Spatial resolution limited to
pixel size
http://www.hologic.com/wh/pdf/R-LM-016_Radiology_Management.pdf 67
67
68
BME HCMUT 34
Mammography – Chapter 8 17-Feb-20
Radiographics 2004:24,1753 69
69
Radiographics 2004:24,1753 70
70
BME HCMUT 35
Mammography – Chapter 8 17-Feb-20
71
71
When x-rays
are absorbed,
electronic
charges are
stored
proportionally
in “traps” in the
phosphor
72
BME HCMUT 36
Mammography – Chapter 8 17-Feb-20
http://www.hologic.com/wh/digisel.htm 73
73
Approved recently
24 x 29-cm field of view
Accommodates all breast sizes
http://www.medical.siemens.com 74
74
BME HCMUT 37
Mammography – Chapter 8 17-Feb-20
Fischer Lorad/Hologic
Fuji/Kodak GE
(Hologic) Siemens
http://www.hologic.com/oem/pdf/DROverviewR-007_Nov2000.pdf 75
75
Radiology 2005:234,353
76
76
BME HCMUT 38
Mammography – Chapter 8 17-Feb-20
DS system
Contrast mode - 219 mrad or 2.19 mGy
Film-screen
Standard mode – 151 mrad or 1.51 mGy (28 kV, 65 mAs)
77
77
Radiographics 2004:24,1755
78
78
BME HCMUT 39
Mammography – Chapter 8 17-Feb-20
Radiographics 2004:24,1757
79
79
7. Economics of FFDM
80
80
BME HCMUT 40
Mammography – Chapter 8 17-Feb-20
81
81
82
82
BME HCMUT 41
Mammography – Chapter 8 17-Feb-20
Advantages
Optimize post-processing of images
Disadvantages
Image display and system cost
83
83
84
84
BME HCMUT 42
Mammography – Chapter 8 17-Feb-20
85
85
86
86
BME HCMUT 43
Mammography – Chapter 8 17-Feb-20
8. Radiologist Responsibilities
Responsibilities include
87
87
8. Radiologist Responsibilities
Responsibilities include
88
88
BME HCMUT 44
Mammography – Chapter 8 17-Feb-20
8. Mammography phantom
89
89
8. Mammography phantom
90
90
BME HCMUT 45
Mammography – Chapter 8 17-Feb-20
8. Mammography Phantom
c.f. Bushberg,
et al. The
Essential
Physics of
Medical
Imaging, 2nd
ed., p. 228.
91
91
92
92
BME HCMUT 46
Mammography – Chapter 8 17-Feb-20
93
93
94
94
BME HCMUT 47
Mammography – Chapter 8 17-Feb-20
95
95
8.
Physicist
Quality
Control
c.f. Bushberg, et
al. The Essential
Physics of Medical
Imaging, 2nd ed.,
p. 227.
96
96
BME HCMUT 48
Mammography – Chapter 8 17-Feb-20
Question
97
97
Question
1. Breast doses in mammography are most likely to be reduced by
increasing the:
98
98
BME HCMUT 49
Mammography – Chapter 8 17-Feb-20
Question
1. Mammograms cannot be processed if the weekly phantom does
not pass. A passing score would be:
99
99
Question
100
100
BME HCMUT 50
Mammography – Chapter 8 17-Feb-20
Question
101
101
Question
1. Which grid would be the best choice for use as a stationary grid in
mammography?
102
102
BME HCMUT 51
Mammography – Chapter 8 17-Feb-20
Question
A. Patient dose.
B. Scatter.
C. Motion unsharpness.
D. Subject contrast.
103
103
Question
104
104
BME HCMUT 52