ABPM Manual 2018-05-v6.3
ABPM Manual 2018-05-v6.3
LABTECH LTD.
Phone +36-52-500-128
Fax +36-52-500-129
Quality certifications
The CE mark indicates that the product meets the essential requirements of European
Council Directive 93/42/EEC concerning medical devices.
Labtech Ltd. also operates a quality system certified to ISO 13485:2016. Made in
Hungary by Labtech Ltd.
If you have any questions, ideas, partnership suggestions please contact us.
LABTECH LTD.
Phone +36-52-500-128
Fax +36-52-500-129
Table of contents
1 INTRODUCTION.................................................................................................... 6
2 PREPARATIONS .................................................................................................... 7
2.1 List of accessories ...........................................................................................7
2.2 Principles of Operation ...................................................................................9
2.3 Warnings and contraindications to using ABPM system .............................. 11
3 MAIN FUNCTIONS ............................................................................................. 13
3.1 Main parts of the ABP Monitor ....................................................................13
3.2 Signs and alarms of ABP ..............................................................................14
3.3 Database management interface ....................................................................17
3.4 Toolbar ..........................................................................................................17
3.4.1 Settings: ....................................................................................................18
3.4.2 Help ..........................................................................................................20
3.4.3 Records toolbar .........................................................................................21
3.4.4 Selected System ........................................................................................23
3.4.5 Filters ........................................................................................................23
3.4.6 Search .......................................................................................................23
3.4.7 Database information ................................................................................24
3.5 How to start a new recording? ......................................................................24
3.5.1 Patient preparation ....................................................................................24
3.5.1.1 Attach the cuff to the patient‟s arm ..................................................24
3.5.1.2 Attach the adhesive cuff anchor .......................................................24
3.5.2 New Recording .........................................................................................25
3.5.2.1 Start recording via Bluetooth ...........................................................26
3.5.2.2 Start recording via USB cable ..........................................................28
3.5.2.3 Start recording without prespecified patient data ............................. 29
3.5.3 Measurement settings ...............................................................................29
3.5.4 Informing patient about batteries and further instructions ........................30
1 Introduction
Dear Customer,
Thank you for purchasing our Ambulatory Blood Pressure Monitor System. We
sincerely hope that our product will help you with your diagnostic and therapeutic work.
To be able to use ABPM System and computer program appropriately, please read the
manuals carefully. A basic knowledge of computers and Windows-based applications is
required. We appreciate any comment you might have on the use of the System. Please,
turn to us with confidence: our colleagues and representatives will be pleased to help
you.
2 Preparations
NOTE: BEFORE STARTING TO USE THE ABPM SYSTEM YOU ARE EXPECTED TO FOLLOW
INSTALLATION AND UPDATE GUIDE OF LABTECH LTD., WHICH PROVIDES GUIDANCE ON
THE INSTALLATION OF THE SYSTEM.
EC-
Code No. Specification Qa. Picture
ABP
ACC-
CD with installation 1 x
0SWCD-01
software
CHA-
Battery charger MAXI 1 x
0004X-01
EC-
Code No. Specification Qa. Picture
ABP
CAB-
Extension cable for 1 x
00USB-01
USB-02
EC-
Code No. Specification Qa. Picture
ABP
BEL-00000-
Waist belt 1 x
01
NOTE: PLEASE REFER TO THE CODE NUMBER ABOVE WHEN PLACING AN ORDER.
Symbols:
Pay special attention to the part, marked with the exclamation mark.
Year of manufacture
Manufacturer
Operating instruction
General warnings:
DO NOT immerse the recorder in any fluid, or attempt to clean the unit with any liquid
detergents or cleaning agents. This may cause an electrical hazard.
DO NOT disassemble the recorder case – doing so automatically invalidates the
warranty. Using a component different from that supplied may result in measurement
error.
DO NOT use the recorder in the presence of flammable anaesthetics and oxygen rich
environment.
DO NOT use recorders in strong heat or in strong electrostatic field.
DO NOT use the recorder if it has failed its diagnostic self-test, or if it displays greater
than zero pressure with no cuff attached. The values displayed by such a unit may be
inaccurate.
DO NOT use on neonates, on patients known to be susceptible to bruising, and do not
attach the cuff to a limb being used for IV infusions.
The ABPM System is designed for external and non-invasive blood-pressure
monitoring on the arm. The cuffs can be placed on intact skin-surface only.
The patient should be instructed on proper and safe removal of the cuff, in case it fails
to deflate within 3 minutes.
Avoid compression or restriction of pressure tubes.
3 Main functions
3.1 Main parts of the ABP Monitor
Cuff
Warning
LED
Event Button
Cuff anchor
Rear view
Symbols
Serial number
Two AA batteries
Battery cover
There is Bluetooth
The Blue LED
connection between the -
seldom flashes
PC and recorder.
The recorders include a buzzer which can operate in different frequencies. By this the
recorders are able to indicate the different statuses and errors.
Start of recording.
One high-pitched long beep
Stop of recording.
Event button is pressed, but there
Two high-pitched short beep
is no function related to pushing
the button in the current state of
the recorder (for example: the
recorder connects to the USB).
Bluetooth connection has been
Two short beep with rising melody
created.
The database management interface enables us to add physicians and patients by name
and ID number (e.g., social security number), which can be sorted and filtered
according to different criteria. The interface also offers a search option by name or ID.
The database management interface displays the list of patients and the pertaining
number of records. Here we can view previous records and start new recordings as well.
3.4 Toolbar
The toolbar contains different buttons which provide access to the most often used
functions in the software.
New patient: Here you can add new patients to the list. Patient‟s first and
last names, date of birth and patient IDs are fields which need to be filled in.
The software allows only unique patient IDs to be entered into the database,
which is the reason why it is recommended to use this field for entering the social
security number of each patient (or any other code which uniquely identifies a person).
Entering address, weight and height is optional when registering a patient.
Modify: Patient data can be edited by clicking on the Modify button. The
interface allows every field to be edited except patient ID (social security
number)
Medical record: This function gives a summary about the medical record.
This summary can be edited and we can also add explanations and
observations.
Delete: Click on the Delete button for deleting patients. When patient is
deleted, all of its corresponding records are deleted as well. After deleting,
there is only one way to restore data. If the “Move deleted records to recycle
bin” option is turned on (Settings/Administration/Settings) you can restore the
deleted record from the “Record recycle bin”. We suggest you to turn this function
before you start using the software to prevent data loss.
3.4.1 Settings:
- Select physician: In this menu the physician, who is
performing the current test, can login to the software
and also select to view the records of other physicians
whose records are not password protected.
- Change password: By selecting this option, the
physician, who is performing the current test, can
change his/her password.
- Administration: Here, we can manage the users and
also the security settings and administrator rights can
be set under this menu point. Further information on
modifying user authorization can be found in
Cardiospy Installation and Update Guide.
- FTP settings: You can find its description in the Cardiospy Installation and Update
Guide.
- Communication: You are able to
communicate through HL7 between Cardiospy
and external software. To setup this, ask your
local distributor and administrator.
- Select systems:
Here, you can enable / disable systems that you
want / do not want to use.
You can activate or deactivate a system by
ticking or un-ticking the checkbox next to the
name of it.
3.4.2 Help
- Installation and update guide: The guide on software
installation and system configuration also contains a
description of the software update process.
- User manual: First of all, choose the system in the Select
systems panel. Once you click on User manual button, the user
manual of the currently selected system will appear.
- What’s new: The novelties of the current software version
can be found in this file.
- Languages: The software offers a choice of several languages, from which you can
select the one you wish to work in.
- Multilanguage translator: With this function, the user is able to translate each term
which appears in the software into the desired language. A password is necessary to
enter this menu point. In case you would like to use this function, ask your distributor
for the password.
- About: Information about the software version, the operating system and the
copyrights can be found here.
- Device Connection Status Indicator: This indicator shows the current connection
status of the devices (the USB-02 interface, the ECG recorder, the blood pressure
monitor and the stress device).
The Device Connection Status Indicator has three types: the first one is used with the
Holter, ABP and Resting Test systems. The second type is used with the Stress Test
systems, and the third type with the Continuous ECG systems.
1 1 1
2 2 3 4 5 2 3 5
.
1. Status of Bluetooth or direct connection
2. Status of connection with Holter ECG recorder /ABP recorder / Resting ECG
recorder / Test ECG recorder
3. Status of connection with blood pressure monitor
4. Status of connection with loading machine
5. Status of connection with Pulse Oximeter (this section is only available when Pulse
Oximeter is activated under System Configuration menu).
Export anonymous records: You can export records without the personal data of
the patient.
- FTP - upload and download:
To set an FTP server, please check Cardiospy Installation and Update Guide.
Upload to FTP: We can select multiple records from
a single patient and upload them to the currently
connected FTP server.
Upload to FTP by date: You can upload records
within a date interval.
Upload to Labtech: In case you have any problem
with the records, then you can upload them to the
Labtech server.
Furthermore, you have the possibility to automatically upload records to FTP. You can
enable it in the “System configuration” / “Settings” menu.
Download from FTP: We can download records from the currently connected FTP
server. You can select multiple records to download at the same time.
- Print rest records by date: You are able to print every Rest record of the selected
physician within a date interval.
- Delete records:
Delete selected record: This buttons deletes every selected
record from the database. You can only restore them from the
record recycle bin (if it is enabled in the settings). You can
enable it in the Administration / Settings by set “Move
deleted records to recycle bin” to “Yes”.
Delete records by date: It deletes every record within a date interval.
Please remember that using the wrong cuff size may result in inaccurate readings.
3.5.1.2 Attach the adhesive cuff anchor
With the patient cable draped over the patient‟s shoulder, attach an
adhesive cuff anchor to the snap on the hose of the cuff bladder.
Do not remove the adhesive backing of the cuff anchor at this time.
7. In the next window, you can set the measuring parameters for the recording.
For further information about the “Holter Monitoring Settings” window, please
check CHAPTER 3.5.3. After you are done
with the settings, click on the “Start
monitoring” button.
6. In the next window, you can set the measuring parameters for the recording.
For further information about the “Holter Monitoring Settings” window, please
check CHAPTER 3.5.3. After you are done
with the settings, click on the “Start
recording” button.
7. After the successful programming,
a new window will pop up in
which you can print the patient
diary.
8. Hold the “Event button” on the
recorder for 3 seconds to start
recording.
9. If the device is started to recording
successfully, then the LED on it
should flash in yellow colour and
you can hear one high-pitched long
beep.
10. Please check CHAPTER 3.6 about how to read a record.
6. Now you can start to examine and edit the record. See CHAPTER 4.
3.6.2 In case you have started the recording without prespecified patient data
1. Take off the recorder-case from the patient, remove the batteries and then
remove the cuff from him/her.
For faster reading process, please place fully charged batteries into the device before
you connect it to the PC.
2. Start the Cardiospy program on the computer. Then select the Holter system.
3. Connect the recorder to the USB reader cable. The reading will start
automatically (by default).
4. Since the recorder was started without prespecified patient data, thus you have
to fill the personal data at the beginning of the reading process in a pop-up
window.
5. Since the recording was started without any PC
connection, you have to verify the date and time
when the recording was started.
6. Click “Next” and then type the ID of the patient.
7. Click “Next” again and then fill the personal data
of the patient.
8. Click “Finish”. A confirmation window will pop
up.
11. Now you can start to examine and edit the record. See CHAPTER 4.
It is always recommended to save your key, so in that case you have to move your
Cardiospy to another PC, you will still be able to read your pending records without any
inconvenience.
However, it is especially necessary to know the function of this key when you are using
the recorders on multiple computers:
In that case where you are using multiple computers in a system, you will have to
set the same protection key for every computer (in Cardiospy) to be able to read the
records with the original patient data.
The data protection key itself only affect recorders where the patient data is actually
stored on the device: EC-ABP, EC-2-3-12H, EC-3H/ABP.
The green line is showing the Heart Rate and the red marks are showing the BP
measurements.
4.1.3 Charts
The software represents systole and diastole values in the form of pie-charts.
4.1.5 Histograms
In this part Systole, Diastole and Map values are represented in the form of histograms.
4.1.7 Scatter
In this function, the cohesive systole-HR, diastole-HR and Systole-Diastole points are
displayed.
You can go back to the patient database more quickly with this button.
4.2.2 Report
By selecting the Report menu a text box appears in which you can write
remarks. When pressing the „Import Automatic Report‟ button both patient‟s
data and the automatic report of the program will be copied to this page. If
you have filled the Indication and the Medication box when you set up the
measurement settings then the content of the symptoms and medication fields will be
imported to the report, as well. The doctor‟s name will be imported to every report‟s
letterhead.
In the Print View, you will see the content of the Report as a new sheet “Report”.
4.2.3 Sequence of tasks
In this menu, you can follow through the whole operation from
reading in to the approving of the evaluated record.
When you have finished examining the Blood Pressure
measurements, you can tick the checkbox next to its name to
mark the record as “Evaluated”.
After you finished with the evaluating, you can mark the record
as “approved”, so it cannot be edited from then, until you untick
the checkbox next to it.
4.2.4 Print
In this menu we can go through all the reports prepared
by the software and print them separately.
4.2.5 Export
In this menu we can go through all the reports prepared by the software and export
them separately (JPG, BMP, PDF, etc).
You can learn more about print / export and its settings in CHAPTER 5.
4.2.6.1 Parameters...
Time:
Here, you can change the “Awake” and the “Special
time”. E.g. you can set when the patient was asleep if
you set the parameters of the “Awake” time to the
beginning and the end of the sleeping period.
Blood Pressure:
Here we can set parameters used for statistical calculations and measurement ranges
used for automatic pre-filtering. Limits for children and adults can be adjusted
separately.
Statistical Limit:
You can set the normal systole and diastole limits in
every interval separately (awake, asleep and special).
If the patient‟s result exceeds this limit the software
indicates that it is in a high range (with red colour).
The program‟s analysis accepts BP values within these ranges. If the results of certain
measurements are outside of this range the software takes the measurements out of the
evaluation and marks them as invalid.
You can play the record with this button next to the scrollbar at the middle of the screen.
4.2.6.7 Service
In this window, you can see information about the
device that was used for the current record. You can see
the device type, recorder version, recorder serial
number, Bluetooth module version, story capacity, total
number of records that was made with the device,
uptime (total time of usage), previously defined
sampling frequency and the current software version (if
you have made the current record in an older version of
Cardiospy, you will see its version number as well).
4.2.6.8 Control panel...
You can also open the Control panel, just like at the main screen (patient database). You
can read more about the control panel in the Cardiospy Installation and Update
Guide.
1. You can select the pages for printing by tick or untick the checkboxes next to
the name of the pages.
2. You are able switch between the pages that you want to see with the radio
buttons next to the name of the pages.
3. It is possible to type in the name of the doctor who is requested the
examination and who signed the report.
4. With this checkbox you can hide the personal data of the examined patient,
making the report anonymous.
5. Here you can select the printer you want to use from the list. First you have to
set the printer in Windows to be able to choose it from here.
6. General printer settings. Number of copies, orientation, etc.
7. Print view settings. More in the next chapter.
“The major difference is that raster image pixels do not retain their appearance as size increases - when you
blow a photograph up, it becomes blurry for this reason. Vector images do retain appearance regardless of
size, since the mathematical formulas dictate how the image is rendered.”
4. Colour mode: You can print in coloured and in black and white mode.
Diurnal Index:
Meaning:
0-10% Non-Dipper
10-20% Dipper
If you have problem with the recorder (you cannot read in for example), then connect
the device to the PC and tick the “Memory content” checkbox.
If you exported the record files that you have problem with, you can attach them to the
bug report by ticking the “Select records or other files” (like screenshots).
8 Technical specifications
Pressure Transducer
± 3 mmHg
Accuracy: