0% found this document useful (0 votes)
12 views5 pages

Tecsis P2114 Mechanical Pressure Gauges Corrosion Resistance DpstarGroup

The document describes all stainless steel safety pressure gauges designed for high corrosion resistance and accuracy in demanding environments, particularly in the chemical industry. It details features such as glycerine filling for wear protection, burstproof construction, and various models with different applications and electrical connections. The gauges are compliant with EN 837-1/S3 standards and are available in a range of measuring capacities from 0.6 bar to 1000 bar.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
12 views5 pages

Tecsis P2114 Mechanical Pressure Gauges Corrosion Resistance DpstarGroup

The document describes all stainless steel safety pressure gauges designed for high corrosion resistance and accuracy in demanding environments, particularly in the chemical industry. It details features such as glycerine filling for wear protection, burstproof construction, and various models with different applications and electrical connections. The gauges are compliant with EN 837-1/S3 standards and are available in a range of measuring capacities from 0.6 bar to 1000 bar.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 5

| |Pressure

Force
Force| |Pressure
Force | Pressure
|Pressure | |Temperature
Temperature | Force
| Temperature | | Switch
| Switch
Temperature |Switch
Switch

All stainless steel safety pressure gauges according to EN 837-1/S3


with or without glycerine filling
New: as multifunctional pressure instrument

Nominal sizes ND 100


Connection position bottom

Description Features

The all stainless steel pressure gauges are ideal for the hard o Measuring system of high corrosion
conditions and the resulting high demands on pressure resistant materials, stainless steel or Monel 400
measurement in production facilities in chemical industry and
o Resistant to chemicals
other comparable branches. Resistance to aggressive media
and environments is achieved by using high-grade materials o Accuracy class 1.0
such as stainless steel both for the measuring system and o Fulfills highest safety requirements
the case. to EN 837-1/S3
The glycerine filling provides wear-protection for the o Solid front between measuring system and dial
measuring system through damping, should pulsating
o Vibration-free display and long service life
pressures and mechanical vibrations occur. The measuring
through glycerine filling
system is of accuracy class 1.0, has overrange protection
amounting to 1.3 times the max. rating and can be loaded up
to the full scale value.

The safety execution of the pressure gauges comprises a Measuring ranges


burstproof solid front between dial and Bourdon tube, a
laminated safety glass as well as a blow-out back (according 0 ... 0,6 bar to 0 ... 1000 bar
to EN 837-1/S3 ). (higher measuring ranges see P1701)
Pressure gauges with glycerine filling are equipped with a
compensation diaphragm. This diaphragm avoids a pressure
rise in the case that is due to temperature bound volume
expansion of the liquid filling, thus avoiding indicated errors.
Applications

If an output signal is expected by the measuring point, Processing technology,


„the multi-functional instrument“ P2114 ND 100 can be Machine and apparatus construction,
used.
It connects the pressure measurement without auxiliary Compressors
energy with the possibility of a sensor signal for the remote
transmission of the upcoming pressure values.
This instrument is particularly suitable for pressure control
rather regulation.
Models: P2112, P2113, P2114, P2143, P2153
tecsis GmbH DE 228 j
Carl-Legien Str. 40 - 44 10/2014
D-63073 Offenbach / Main Sales National Sales International e-Mail: [email protected] p. 1 / 5
Tel.: +49 69 5806-0 Fax: +49 69 5806-7788 Fax: +49 69 5806-7788 Internet: www.tecsis.de
Technical data
Models P2112 P2113 P2143 P2153 P2114 Option
Nominal size 100
Symbol

Contact type without Inductive Multifunctional


)
Number of contacts * without 1 to 3 depending on measuring range
Accuracy 1.0 acc. to EN 837-1
Ranges 0 ... 0.6 bar to 0 ... 1000 bar
negative or positive / negative and positive gauge pressure
Application Constant load: up to full scale value
Alternating load: up to 0,9 x full scale value ( model: P2112 + P2113
short-time: overload capacity 1.3 1.5 to 2 x )
Case Stainless steel, 1.4301 with blow-out back, solid front Rear mounting bracket,
Scale ranges ≤ 0 … 16 bar, with compensating valve to vent case stainless steel
Bezel Stainless steel, 1.4301 bayonet ring Front flange,
VA or VA polished
window Laminated safety glass Polycarbonat
(safety glass)
Dial Al. white, scale and imprint black Dual scale
Pointer Al. black Aluminium, black, adjustable
Movement Stainless steel Brass
Measuring element Stainless steel, 316 L Measuring element:
Monel
Bourdon tube up to 60 bar , helical tube above 100 bar Stainless steel, 1.4571
Connection Stainless steel 316 L
- position Radial bottom
- thread G 1/2 B Other threads on request
Temperature
- Medium Tmin. -20°C…Tmax. 200°C - for unfilled instruments
Tmin. -20°C…Tmax. 100°C - for filled instruments
- Ambient Tmin. -40°C…Tmax. 60°C - for unfilled instruments
Tmin. -20°C…Tmax. 60°C - for filled instruments
Temperature drift 0,4%/10K if deviation from normal temperature 20°C
Liquid filling without Glycerin without Polybuten without Silicone M50
Protection acc. to
IP 65 IP 54 IP 65 IP 65
EN 60 529 / IEC 529
Throttle without ø 0,4 ; ø 0,5; ø 0,8
CE-confirmy Model P2114: acc. to ATEX
Construction acc. to
ATEX: 94/4 Ex ia IIC T4/T5/T6
ATEX Ex II 2 GD c
rather Ex I M2 Ex ia I
Pressure Equipment
97/23/EG
Directive
Weight approx. 0.660 kg 0.80 kg
1)
fsv = full scale value

)
* Number of contacts
Measuring range Magnetic snap-action contact Inductive contact
to 1 bar 1 1
1.6 bar 2 3
above 4 bar 4 3

DE 228 j

p. 2 / 5
Technical data
Models P2112 P2113 P2143 P2153 P2114 Option
Nominal size 100
Symbol

Elektrical Connection Cable connector right hand L-plug connector, 180 °C


side; 6 screw terminals + PE, rotatable, max. 1.5 mm², wire
cross section of the protector,
----- conducting wire 2.5 mm2 cable gland M20 x 1.5,
Screw type conduit fitting external cable diameter 7-13
M20x1.5, outgoing mm,
downwards incl. strain relief
Power supply UB 12 < UB ≤ 30
 Supply voltage effect ----- ----- ≤ 0,1 % fsv/10 V
 Permissible residual ripple ≤ 10 % ss
Output signal 0…20 mA, 0…10 V,
----- ----- 4 … 20 mA, 2-wire 4…20 mA passiv,
acc. to NAMUR NE 43
Permissible max. load RA RA ≤ (UB - 12 V)/0.02 A with
----- ----- RA in Ohm and UB in Volt,
however max. 600 Ω
Effect of load ----- ----- ≤ 0,1 % fsv
Electrical zero point through a jumper across
----- ----- terminals 5 and 6 (see
operating instructions)
- Long-term stability of electronics ----- ----- < 0,3 % fsv / a
- Electrical output signal ----- ----- ≤ 1 % of measuring span
Linearity ≤ 1,0 % of span (limit point
----- -----
calibration)
Conformity specifications ----- ----- Ex - Variant
Power supply ----- ----- 14 … 30 DC V
Short circuit rating ----- ----- 100 mA
Rating ----- ----- 1000 mW
internal capacitance ----- ----- Ci ≤ 12 nF
internal inductance ----- ----- mH - negligible
EMC- directive 2004/108/EG EC Interference
----- ----- emission (Limit Class B) and
immunity to EN 61 326-1

Dimensions

Dimensions in mm
Models
A b D1 D2 e G h ±1 SW
P2112, P2113 24 57.5 101 100 17.5 G1/2 B 87 22

DE 228 j

p. 3 / 5
Dimensions

Model P2143 P2153


Dimension "A" 104 mm 108 mm

Dimension in mm
Model weight in kg
a b c D1 D2 e G h ±1 SW
P2114 25 59.5 94 101 100 17 G1/2 B 87 22 0.80

DE 228 j

p. 4 / 5
Terminal assignment

Terminals 1 and 2 are the terminals for the signal output and for the power supply. The terminal marked
with PE (protective earth) is connected internally to the housing. The connections 3 to 6 or 4 to 6 (for the
3-wire version), must remain free and must not be used as connection points (also see Chapter 10
"Technical data").

2-wire-design 3-wire-design
i.e. 4 ... 20 mA i.e. 0 ... 20 mA / 0 ... 10 V

An unstabilised DC voltage, with a residual ripple of max. 10 % peak-to-peak in the range of the indicated
supply voltage limits, is sufficient as a power supply. Make sure that the supply voltage applied exceeds
the maximum required voltage by at least the value of the voltage drop across the external display or
evaluation devices; i.e. the transmitter can operate using a non-stabilised supply voltage within the given
limits, so long as the voltage available to the transmitter does not fall below 12 V, or below 14 V for the
Ex-version.

Modifications reserved

DE 228 j

p. 5 / 5

You might also like