Abstract
Significance: Photoacoustic imaging (PAI) is an emerging functional and molecular imaging technology that has
attracted much attention in the past decade. Recently, many researchers have used the Vantage Verasonics® research
system for simultaneous ultrasound (US) and photoacoustic (PA) imaging. This was the motivation to write on the
details of US/PA imaging system implementation and characterization using Verasonics platform.
Aim: We describe the subtle details of US/PA imaging system setup, study the performance parameters of the system,
and explain sequencing of the US/PA signal generation and signal amplification as well as the details required for
efficient use of the hardware of the system and data processing protocols.
Approach: We focused on linear-array based PAI due to its popularity and simple setup, as well as its high potential
for clinical translatability.
Results: We have shown the sequencing of the US/PA signal generation, signal amplification, and related data
processing protocols. A step-by-step guideline to develop and characterize PAI system using Vantage 128 has been
presented. Some of the limitations of the vantage system are also listed.
Conclusions: Photoacoustic imaging is a complement to the already established US imaging technique and may
significantly increase its scope of application in diagnostic imaging and therapeutic monitoring. Combining with
commercial medical US systems, the development of PAI can be accelerated by taking advantage of US image
reconstruction and processing. With the information we presented in the body of this review and the four appendices,
we described most of the experimental considerations one should know when working with the Vantage system for
PAI tests.
Keywords: Photoacoustic imaging, ultrasound, linear array, image reconstruction, Vantage Verasonics®.
1
1 Introduction
imaging modality that works based on the photoacoustic (PA) effect 1. The sample (light
thermoelastic expansion of the absorber followed by emission of acoustic waves. The absorber
could be endogenous such as hemoglobin (oxy- or deoxy-), myoglobin (oxy- or deoxy-), melanin,
lipid, Bilirubin, water, or an exogenous contrast agent such as dyes 2. The absorption spectrum of
some of the endogenous and exogenous absorbers are shown in Figure 1; the absorption spectrum
of two example imaging targets that are commonly used for calibration and performance
evaluation in PAI experiments, i.e., vinyl black tape and carbon pencil lead, are given in Appendix
A.
Fig. 1 Absorption spectrum of (a) endogenous, and (b) exogenous contrast agents. Reproduced with permission from
A = ( ) Cd ( ) C
3-5
. According to Beer-Lambert law, absorbance is defined by where, , , and d are molar
absorptivity, concentration, and cross-section thickness of the contrast agent, respectively. Absorption coefficient is
= ( A d ) log10 e 6
expressed as: .
2
The emitted acoustic waves from the absorber are detected by ultrasound (US) transducers.
The transducer signals are then given to an image reconstruction algorithm to generate the
absorption map of the tissue. The PAI process steps are illustrated in Figure 2.
Fig. 2 Principle of PA signal generation, detection, and image reconstruction. Reprinted with permission from Ref. 7.
A high energy and short pulse laser light illuminates the absorber, leading to a transient temperature rise which results
in a thermoelastic expansion of the absorber, and acoustic (or PA) wave generation. The signals generated from the
waves received by an ultrasound probe are given to a reconstruction algorithm to form a PA image.
Due to strong optical scattering in biological tissues, pure optical imaging modalities have a
1, 8-10
shallow imaging depth . The transport mean free path (i.e., the mean distance after which a
ultrasound imaging can image deep in biological tissues with a high spatial resolution, its acoustic
contrast is incapable of providing certain physiological parameters 13. In PAI, there is no restriction
for photons, thus optical excitation can travel far beyond the diffusion limit and still generate
acoustic waves. Sensitivity of PAI in deep tissues is orders of magnitude higher than that of pure
optical imaging modalities 13, 14; the highest penetration depth reported in PAI is ~12 cm 15-17. PAI
is an ideal modality for measuring or monitoring tissue physiological parameters by imaging the
3
concentration of tissue chromophores 18, which are changed during the course of a disease 19. PAI
has been evaluated in preclinical and recently in clinical applications for disease detection and
monitoring purposes 20-41. For instance, it has been used to study human skin abnormalities 37, 42-
44
, brain disease detection 45, 46, human breast tumor detection 19, 47-50, retina disease diagnosis 51,
52
, and atherosclerosis evaluation of vessel walls 30, 53, 54.
system design (more details are provided in 7). Among all configurations and variations of PAI,
linear-array based PAI is one of the most commonly used due to its straightforward setup, easy
use, and simple clinical translatability. In addition, since linear-array ultrasound imaging systems
have been well-established as clinical tools, slightly modifying them to develop more capable
In recent years, many researchers have used the Vantage Verasonics® system for simultaneous
produce co-registered US and PA images; the PA image represents the optical absorption map of
the tissue while the US image represents the tissue acoustic impedance map. Here, we describe the
subtle details of US/PA imaging system setup, study the performance parameters of the system,
and explain sequencing of the US/PA signal generation and signal amplification as well as the
details required for efficient use of the hardware of the system and data processing protocols.
Vantage hardware is controlled by MATLAB coding. The detailed explanation of the parameters
used in the code for data acquisition and processing, image reconstruction, and display are given
in Appendix B. The Vantage system architecture for US/PA sequencing is shown in Figure 3A.
The structures for US and PA run-time defined in the MATLAB code communicate to runAcq
4
through the activation of the MATLAB loader function VSX. This opens a GUI including imaging
windows of the reconstructed images from RunAcq (US and PA), time gain control, US voltage
transmit, and many other parameters that are used for real-time user modification to the imaging
sequence. These modifications take effect on the hardware between cycles of hand-offs between
VSX and RunAcq. After initial activation of VSX, RunAcq then communicates all of the run-time
transmit and receive structures to the Vantage hardware (silver box) through the PCIe cable.
RunAcq, receives the raw data, stores it in the RcvBuffer variable, and performs image
reconstruction on US and PA data and sends it back to the VSX GUI. The raw data of US and PA
Fig. 3 Vantage system architecture and data acquisition sequencing. (a) Vantage system architecture for US/PA
sequencing and (b) timing diagram of US/PA sequencing in Vantage system. US: Ultrasound, PA: Photoacoustic,
Recon.: Reconstruction. RcvBuffer is the predefined buffer where the US and PA data are stored, Acq: acquisition,
5
TX: transmit, RX: receive, A/D: analog to digital, VSX: Verasonics script execution, GUI: graphical user interface,
The timing sequence in US and PA data acquisition, illustrated in Figure 3B, is as follows. US
acquisition requires a user defined number of steered angles. Between each set of US acquisitions,
the system waits for a laser trigger input to begin one PA event. After PA acquisition, US and PA
data are transferred to the host controller and stored in the RcvBuffer variable. RunAcq then
performs image reconstruction on the averaged data of all of the steered angles and also on the PA
data. The reconstructed US and PA images are then displayed in the MATLAB VSX GUI.
To perform an US/PA experiment, an Nd:YAG laser (PhocusMobil, Opotek Inc., CA, USA) with
a repetition rate of 10 Hz and a pulse width of 8 ns was used. The laser uses an optical parametric
oscillator (OPO) to tune the wavelength of light between 690 nm to 950 nm. For light delivery, we
used a custom fiber bundle (Newport Corporation, Irvine, CA, USA). For data acquisition, the
Verasonics Vantage 128 system was used. The specifications of the vantage system are listed in
Table 1.
Table 1. High frequency Vantage 128 system specifications (Verified by Verasonics’ technical support).
Parameters Specification
Channels 128 Tx / 128 Rx
Power requirements 100V-240V (50-60 Hz)
Programmable pulser voltage 2 to 190 V p-p
Time delay resolution 4.0 ns
HIFU capability Yes
A/D resolution 14 bits
Max sampling rate 62.5MHz
Fastest external trigger 100 KHz
Maximum trigger voltage level 5.5 V
Single transmit event ~ µsec
Data pre-amplification 24 dB
Second stage Amplification 30 dB
6
Noise Figure 1.5 to 3.0 dB
via 8 lanes PCIe 3.0
Data transfer to host computer
sustained data transfer rates up to 6.6 GB/s
The Vantage system is connected to a host computer using a PCIe express cable. A transducer
is connected to the 260 pin Cannon connector mounted on the Vantage system. On the sample end,
the transducer was placed and held perpendicularly to the sample. There is a coupling media
between the imaging target and the transducer. Water and ultrasound gel are the most common
couplant 56. More details about acoustic couplants can be found in Appendix C.
SetUpL11_4vFlashAngPA.m, for the 128-element linear array L11-4v probe (see Appendix B).
This script can be modified to be used with other probes. Two commonly used ultrasound
transducers with Vantage system are ATL Philips L7-4 and Verasonics L22-14v. The
Table 2. Specifications of linear array transducers ATL Philips L7-4, and Verasonics L22-14v
Model (Unit) L7-4 L22-14v
Lens correction 0.8870 0.6804
Central frequency (MHz) 5.2080 15.6250
Number of Elements [4, 7] [14, 22]
Element width (mm) 128 128
Spacing (mm) 0.2500 0.0800
Max high voltage (V) 0.2980 0.1
Focal length (mm) 50 30
Dimension
(11cm × 6.0cm × 3.0cm) (9cm × 3.5cm × 2.5cm)
7
Table 3 lists the modifications on the original code, written for the L11-4v, needed for the L7-
4 and L22-14v. The remainder of the script is compatible as long as the probe is a 128-element
linear array.
Table 3. Modifications to original L11-4 script by Verasonics, for ATL Phillips L7-4 and Verasonics L22-14v probes
Parameters Original Script Modified Script
P(1).startDepth = 2; P(1).startDepth = 0;
P(1).endDepth = 192; P(1).endDepth = 192;
Imaging depth
P(2).startDepth = 0; P(2).startDepth = 0;
P(2).endDepth = 128; P(2).endDepth = 192;
oneway = 0; oneway = 1;
Laser parameters flash2Qdelay = 200; flash2Qdelay = 188;
PA_PRF = 100; PA_PRF = 10;
Trans.name = ‘L11-4’; Trans.name = ‘L7-4’;*
Trans.units = ‘mm’; Trans.units = ‘mm’;
Active transducer Trans = computeTrans(Trans); Trans = computeTrans(Trans);
nElem = Trans.numelements; nElem = Trans.numelements;
Trans.maxHighVoltage = 50; Trans.maxHighVoltage = 50;*
*Trans.name = ‘L22-14v’ and Trans.maxHighVoltage = 30 in the case of the L22-14v.
The Vantage system supports linear array transducers from different companies. The list of
different transducer arrays that can be used with the Vantage system is given in Table 4. If an
independently and input into the computTrans.m script under a new custom transducer case. Key
attributes include the central frequency, bandwidth, number of elements, element width, element
spacing, element position, and connector pinout arrangement. In addition to unrecognized linear
arrays not listed in Table 4, single-element-based arrays, ring arrays, and hemispherical arrays can
also be connected to the Cannon connector mounted on Vantage system via a custom-made
converter.
8
L12-5 50mm P3-2*
CL10-5 P4-1
CL15-7 P4-2
C4-2 P5-3*
C5-2 P6-3
C7-4* P7-4
L10-4v L22-14vX
L11-4v L22-14vX-LF
L11-5v L35-16vX
Verasonics L12-3v L38-22v
L22-8v C5-2v
L22-14v P4-2v
L22-14vLF
GE9LD GEIC5-9D
GE GEC1-6D GEL3-12D
GE4CD GEM5ScD
Siemens L10-5
Vermon 4DL7*
*These transducers are recognizable by the system, but the system does not have the transducer’s
attributes which must be manually entered into the computeTrans.m MATLAB file.
diameter carbon lead phantom in water was imaged at 690 nm. The energy after the fiber bundle
was measured as ~15 mJ/cm2. L22-14v transducer array was used in this experiment. The time
domain signal obtained from the central transducer (i.e., 64th element), is shown in Figure 4A. In
Figure 4B, the magnitude of the single-sided Fast Fourier Transform (FFT) of the time domain
signal is shown. In Figure 4A, we also show that the strength of the US waves reflected back from
9
Fig. 4 Photoacoustic signal in (A) time-domain, and (B) frequency-domain, obtained from a 2 mm diameter carbon
lead phantom, imaged with L22-14v transducer at 690 nm wavelength. The red signal in A shows the overlaid US
We characterized L7-4 and L22-14v for US/PA imaging. We created a resolution phantom of
hair of diameter 54 µm in an open top plastic cubic box filled with deionized water (see Figure
5A). The lateral and axial resolutions were quantified by measuring the full width half maximum
59
(FWHM) of the normalized 1D intensity profile across the hair image . The axial and lateral
resolutions of the L7-4 were measured at depths 0.5 to 5.0 cm with steps of 0.5 cm (see Figure
5B). The axial and lateral resolutions of the L22-14v were measured at depths 0.2 to 2.0 cm with
steps of 0.2 cm (see Figure 5C). In L7-4, the axial resolution remained constant in both US and
PA imaging while the lateral resolution worsened with depth, whereas in L22-14v, the axial and
lateral resolutions stayed almost constant in both US and PA images. This is potentially due to the
focal depth specified by the Vantage system. High frequency transducers have less beam
10
divergence than low frequency transducers, this may have also contributed to worsening the lateral
resolution of the L7-4 and constant lateral resolution of the L22-14v 60.
Fig. 5 US/PA resolution study. (A) Schematic of the experimental setup, including a hair phantom photograph
captured by a 4× objective on a light microscope (SME-F8BH, Amscope, CA, USA). Resolution study when (B) L7-
4 was used, (C) L22-14v was used. (i) US axial resolution versus depth, (ii) US lateral resolution versus depth, (iii)
11
4 Technical Considerations
Photoacoustic imaging relies on careful timing of data acquisition and laser firing to accurately
reconstruct photoacoustic depth information. In Q-switched Nd:YAG lasers, the timing involves
triggering of the flash lamps to stimulate the emission medium (Nd:YAG) followed by time for
optical buildup and subsequent opening of the Q-switch for release of the laser beam. A
straightforward triggering method (see Figure 6A (i)) involves using internal flash lamp and Q-
switch triggers built into the laser being used and using the flash lamp out port to trigger data
acquisition on the Vantage trigger in-1 port. Upon receipt of the flash lamp trigger, the Vantage
waits for a user-specified delay time equal to the known optical build-up time for the laser and
then begins recording approximately when the laser fires. The major drawback to this triggering
method is that the optical build-up time for the laser will “jitter” between tens of nanoseconds and
will cause the peak PA signal to fluctuate between a few sample numbers. This will cause issues
in data averaging between subsequent frames. We have implemented a more optimal triggering
61
method following the method described in (see Figure 6A (ii)). In this method the flash lamp
and Q-switch of the laser are both externally triggered. Using a function generator (ATF20B,
Shenzhen Atten Electronics Co., Ltd., Nanshan, China) a trigger of 10Hz, 5Vpp, 2.5V offset, 50%
duty cycle was simultaneously sent to the flash lamp-in of the Nd:YAG laser (PhocusMobil,
Opotek, CA, USA) and the trigger in-1 of the Vantage system. Within the Vantage MATLAB
script, the flash2Qdelay parameter is set to the external Q-switch delay time specified by the laser
retailer (290 µs). This allows the Vantage to then trigger the Q-switch by connecting the Vantage’s
trigger-out to the Q-switch-in on the laser. We have compared the performance of the two
triggering methods by imaging a 2mm diameter carbon lead, when L7-4 US probe was used for
12
PA signal detection and the fiber bundle we described earlier was used for illumination at 690 nm.
Five sequential photoacoustic frames were recorded and overlaid with both methods (see Figures
6B(i) and 6B(ii)). The results show a greater “jitter” in the first triggering method compared to
nearly no “jitter” in the second method. We then averaged the signals acquired with each of these
triggering methods. Averaging improved the signal-to-noise ratio (SNR) of the PA signal probably
because the signal components being averaged were correlated and the noise components were
not. 64th element of the L7-4 probe data was averaged using various numbers of stored frames: 1,
Fig. 6 Photoacoustic signal “jitter” comparison between two triggering methods when a 2 mm carbon lead was imaged.
64th element signals are plotted. (A) (i) Schematic of straightforward triggering method where laser triggers Vantage
system, (ii) schematic of the function generator driven triggering method where function generator triggers laser flash
lamp and Vantage system, which then triggers laser Q-switch, (B) (i) five sequential PA frames showing significant
13
“jitter” between frames, (ii) five sequential PA frames showing nearly no “jitter” between frames, and (iii) averaged
62
Light illumination is one of the main components of a photoacoustic imaging system . The
orientation of the illumination of light can influence the amplitude of the PA signal received by
12
the transducer based on how the light energy is deposited on the object . To investigate this
matter, we imaged a 2 mm diameter carbon lead in a scattering media, i.e., 2% milk, at 690 nm.
We acquired PA data at different illumination angles. The angles were measured as the angle
between the transducer and the illumination plane (see the experimental setup in Figure 7A). In
these experiments, the transducer plane was held constant (always perpendicular to the sample)
and the illumination plane was changed while the illumination spot was at the same location on
the surface of the phantom. Figure 7B (i) shows the data collected from the 64th element of the L7-
4 probe at an angle of = 40o where the PA signal was weakest. Figure 7B (ii) shows the data
collected from the 64th element of the L7-4 probe at an angle of 48o where the PA signal was
strongest. The difference between the amplitudes of the PA signals is probably related to: (i) the
light illumination path at each angle which depends on the scattering map of the phantom, (ii) the
mechanical property of the tissue as to which directions the thermoelastically generated pressure
waves have the maximum amplitude, and (iii) at greater angles, larger area of the target phantom
is illuminated, hence, more numbers of point sources are excited and collectively a larger pressure
wave is generated. It can be seen that the shape of the two induced PA signals are consistent while
the PA signal amplitude is higher at one angle over another. This demonstrates the importance of
14
Fig. 7 Impact of illumination angle. (A) Experimental setup of illumination angle investigation. (B) PA signal profile
from 64th element of L7-4 probe taken from a 2 mm carbon lead phantom with different illumination angles. (i) 40
The Vantage system automatically samples signals at a rate that is 4 times the central frequency of
the transducer being used. This is sufficient to meet the Nyquist limit, however it may be necessary
to analyze PA signals at a higher sampling rate. This requires modifications to the script as well
as the use of an in-built filter tool to change the spectrum of the bandpass filter on the receive end
of the transducer so that the system does not mistake the higher sampled signals as high frequency
the sampling rate will be multiplied. This is to allow the data buffer to hold the increased number
Finally, the parameter Receive.inputFilter is modified by entering filterTool into the MATLAB
command line. This brings up a GUI which outputs the Receive.inputFilter value based on the user
input parameters. Inside the GUI, sampleMode are set to custom with the decimSampleRate set to
the rate of choice. The bandpass filter central frequency and relative bandwidth are then set to the
15
standard values of the probe being used. The modifications to the script can be seen in Table 5 for
using the L7-4 probe and changing the sampling rate from 20.8MHz to 62.4MHz.
Table 5. Script modifications to change the sampling rate of L7-4 from 20.8 MHz to 62.4 MHz. The modifications
are indicated in red.
Modification Original Modified
no.
1 Resource.RcvBuffer(1).rows Resource.RcvBuffer(1).rowsPerFra
PerFrame = 2048*(na + ne); me = 3*2048*(na + ne);
We imaged a black tape of 18 mm width at 3 sampling rates, 20.8, 41.6, and 62.4MHz. The
tape was imaged in a water bath with the L7-4 probe at the illumination wavelength of 690 nm.
The results for these sampling rates are given in Figure 8A, B, and C, respectively. To demonstrate
the improvement in the reading of the signals with increasing the sampling frequency, we
calculated the difference between the top peaks and also the bottom peaks in the signal. The values
16
are indicated in each figure. One drawback of using higher sampling frequency is slower available
frame rate, therefore, an alternative solution to increase the number of samples without increasing
the sampling frequency is interpolation. In Figure 8D, we showed the interpolated signal (3 times
highest peak value normalized by respective sample numbers ( xnor = x ( ymax ) sample no. ) and the ratio
Fig. 8 PA signal profile from 64th element of L7-4 probe taken from a black tape phantom of 18mm width with varying
sampling rates: (A) 20.8 MHz, (B) 41.6 MHz, (C) 62.4 MHz, and (D) interpolated signal using ‘spline’ algorithm.
xnor = x ( ymax ) sample no. y r x = Ratio of y values , where y = positive or negative peaks .
Here, and
The default MATLAB script provided by Verasonics® utilizes a single imaging window for both
US and PA image display, with the user being able to modify parameters to view one or the other.
With real-time imaging, it is more important to simultaneously view US and PA images side by
side. The following describes how to modify the original example script to simultaneously display
US and PA images in real-time. With a single imaging window, all parameters for the Resource
structure have arguments of 1. Adding a second window requires duplication of all of the
17
pdelta, Position, ReferencePt, numFrames, AxesUnits, Colormap, and splitPalette are copied from
to hot. The other change necessary to keep both windows real-time is modification of the
Process(n).Parameters structure where n is 1 for US and 2 for PA. In the original structure,
mappingMethod is set to lowerHalf for US and upperHalf for PA, but both should now be set to
full. Also, displayWindow was originally set to 1 in both processes and now it should be 1 for US
and 2 for PA. Modifications to the original script to simultaneously display US and PA images, in
Vantage offers beam steering as a method of improving US image quality. This technique modifies
transmit delay times in the beam apodization to steer the beam at different directions into the
imaging field. This provides received echoes from different angles of reflection off of the imaging
targets. The received echoes from multiple steered angles are then averaged to form an improved
image. The variable na in the MATLAB script can be set to any positive integer to determine the
𝑛𝜋
number of steered angles. The transmit waveforms will be emitted at angles 𝑁+1, where N is equal
𝜋
to the variable na. n iterates with each transmit angle from 1 to N, where is normal to the
2
transducer elements. Upon definition of na, these angles and corresponding time delays in the
apodization for each element are calculated, and the time occurs between each US steered angle is
determined. Taking this time into account with the pulse repetition rate of the laser, the number of
the angles occur between each laser pulse is determined. Further, increasing the number of steered
18
angles and the number of stored frames in each buffer are balanced to avoid slowdown of the
system due to memory build up, which can potentially freeze MATLAB if the system RAM is
overflowed. To investigate the effect of the number of steered US transmit angles on the image
quality, we imaged a phantom, with 2 mm, 0.9 mm, and 0.2 mm diameter carbon leads in water
with varying numbers of steered angles (i.e., 8, 16, 32, 64, 128), and quantified the contrast to
12
noise ratio (CNR) of resultant images following the method explained in . Figure 9A
demonstrates the experimental setup. Figures 9(B-F) show the US image for 8, 16, 32, 64, and 128
steered angles, respectively. Figure 9G shows the improvement of CNR with greater number of
steered angles.
Fig. 9 US image quality improvement. (A) Experimental setup showing an ultrasound probe and a 3-carbon lead
phantom. (L1) 2 mm diameter carbon lead, (L2) 0.9 mm diameter carbon lead, (L3) 0.2 mm diameter carbon lead, (B-
F) US images of the 3 carbon lead phantom for 8, 16, 32, 64, and 128 steered angles, respectively. Green dashed circle
encloses object pixels and red dashed box encloses both object and background pixels, (G) CNR versus number of
steered angles
19
4.6 Photoacoustic Signal Amplification
The PA pressure waves received by the linear array transducer are weak (i.e., their SNR is low) 64,
65
. In cases where the imaging media is highly scattering, only weak pressure waves reach the
transducer and produce low level voltage signals. These low level signals may not be accurately
digitized using the default Analog Front End (AFE) settings of the Vantage data acquisition (DAQ)
system. The AFE gain can be increased by up to 12 dB by modifying the settings in the RcvProfile
structure, however in some cases, additional gain may be required. To increase the amplitude of
the received signal, we used a 128-channel amplifier (AMP128-18, Photosound, Houston, TX,
USA). The amplifier has 40 dB gain with -6 dB cut off frequencies of 25 kHz and 35 MHz. The
amplifier was attached to the 260 pin Cannon connector on Vantage system (see Figure 10). The
US probe was then plugged into the front end of the amplifier. In the MATLAB script, i.e.,
computeTrans.m, a custom transducer ID was added. Details of the ultrasound transducer were
copied to the custom ID so that the amplifier can be recognized. This is similar to the addition of
any unrecognized transducer to the system, only the transducer attributes (e.g., central frequency,
bandwidth, number of elements, element width, element spacing, and element position) are
determined independently and input into the computTrans.m script under a new custom transducer
case. To allow for identical wiring path lengths through the amplifier from the transducer end to
Vantage system end, the pins were not connected one to one and their organization were corrected
in the MATLAB script in the Trans.Connector variable. Since the amplifier was programmed to
multiply any signal going in or out, to avoid causing damage to the transducer elements during US
transmit, the transmit beam was disabled by setting TX.Apodization to zeros. This allowed the
Vantage system to only work in the receiving mode with amplification. To demonstrate the utility
of the amplifier, a two-wire phantom (each with 600 µm diameter) was imaged with and without
20
the amplifier at varying concentrations of intralipid, i.e., 0%, 25%, 50% (Sigma Aldrich, USA).
The experimental setup is shown in Figures 10A and B. The setup consists of a 50 Hz, 532 nm
Nd:YAG laser (NL231-50-SH, EKSPLA, Vilnius, Lithuania) with 7 ns pulses, and an L7-4 US
transducer. The laser illuminated the 2-wire phantom through a large fiber optic bundle (1 cm
addition to the signal peak increase in all concentrations of intralipid (see Figures 10B, C, and D).
SNR improvement was probably due to the high pre-amplifier input impedance which shifted the
transducer noise spectrum to low frequencies, which was then filtered out using a high-pass filter
66
.
21
Fig. 10 PA signal amplification. (A) System setup to test the Photosound AMP128-18 amplifier: (i) display monitor,
(ii) Nd:YAG laser head, (iii) fiber optic bundle, (iv) laser power supply, (v) laser chiller. (B) Wire imaging results:
wire phantom suspended in 0% intralipid imaged (i) without and (iv) with amplifier, wire phantom suspended in 25%
intralipid imaged (ii) without and (v) with amplifier, and wire phantom suspended in 50% intralipid image (iii) without
22
and (vi) with amplifier. (C) Raw data from the 64 th channel of the linear array when imaging a single wire in 50%
intralipid concentration to demonstrate SNR improvement from (i) without amplifier to (ii) with amplifier, (D) Bar
chart demonstrating signal amplitude increase corresponding to phantom imaging in B with and without amplifier, at
5 Fluence Compensation
The initial PA pressure is proportional to the product of local optical absorption coefficient and
the optical fluence deposited at the same location. However, the optical fluence could vary
significantly in living tissue which necessitates the use of optical fluence compensation for
67
quantitative PA imaging . If the optical parameters of the tissue, e.g., absorption coefficient,
scattering coefficient, and anisotropy factor, are known, the optical fluence can be calculated by
solving the photon diffusion equation using Monte Carlo (MC) or finite element method 68, 69. With
the known optical fluence distribution, a fluence-compensated absorption map can be calculated
and ink mixture (see Figure 11A and B). We used 0.5 mm diameter carbon lead in three layers of
gelatin phantom with thicknesses of 5, 8, and 7 mm, respectively. We added intralipid (Sigma
Aldrich, USA) as scatterers and used ink to represent absorption. The concentration of the
intralipid solutions were 4%, 1%, and 6%, and those of ink were 0.1%, 0.4%, and 0.2%,
respectively; these values were chosen to represent various biological tissues. Using Mie calculator
70
, the scattering coefficient, absorption coefficient, and anisotropy factor of the layers were
calculated at 532 nm. The scattering coefficients were 15.4 cm-1, 3.1 cm-1 and 22.3 cm-1,
respectively. The absorption coefficients were 0.3 cm-1, 0.9 cm-1, and 0.4 cm-1, respectively. The
absorption coefficient of water, i.e., 0.11 cm-1, was added to these numbers. We used another layer
23
of pure gelatin with the thickness of 5 mm. The scattering coefficient of both pure gelatin and US
gel were considered 0.05 cm-1. The absorption coefficient of US gel (3mm layer) was considered
0.11 cm-1. The experimental protocol was as follows: (1) we imaged the phantom using the L7-4
probe and generated an US image (see Figure 11C) and a PA image (see Figure 11D); (2) we then
71
segmented the US image using the segmentation method given in ; (3) the abovementioned
optical properties were assigned to each segment, creating the phantom optical model; (4) fluence
map (see Figure 11E) of the phantom optical model was generated by an MC simulation (using
MCX software 72); (5) finally, the PA image was divided by the MC simulated image. The resultant
image was a fluence-compensated PA image (see Figure 11F). The PA intensity variation between
the peaks (at the location of imaging targets) in the fluence-compensated PA image was less than
5%. In this experiment we assumed that the field of view of each element of the transducer is
cylindrical.
Fig. 11 Fluence Compensation algorithm validation. (A) Schematic of a phantom with 3 layers of gelatin phantom
mixed with intralipid/ink solution and 3 carbon lead imaging targets, (B) photograph of the phantom made in a cube
24
box and magnified to show the layers, (C) US image of the phantom, (D) PA image before fluence compensation, (E)
Monte Carlo simulation of light propagation in the phantom, (F) PA image after fluence compensation.
6 PA Image Reconstruction
After detecting PA signals by the linear array, the channel data are given to an image reconstruction
algorithm and an optical absorption distribution map, i.e., PA image, is reconstructed 73-77. Several
78-81
image reconstruction algorithms have been studied for linear-array PA imaging . The most
basic algorithm is delay-and-sum (DAS). DAS follows the dynamic focusing protocol, in which
the focus is adjusted for each pixel of the imaging target; the detected signals are delayed
proportional to the distance between the focal point and the position of the element in the imaging
array. Finally, the delayed signals are summed up, and an image is formed. As a result of this
summation, the on-axis signals will overlap on each other while off-axis signals will be suppressed.
In the remaining of this section, we used two signal quality metrics to compare the performance
of different image reconstruction algorithms: (i) resolution, which is defined as the full width half
maximum (FWHM) of the lateral profile of the image of a point source, and (ii) sidelobe, defined
as the power of the first lobe next to the main lobe. Although DAS is simple to implement and
fast, it treats all the detected signals the same way which results in low resolution images with high
sidelobes; this is mainly due to the constructive overlap of the off-axis signals 82. Delay-Multiply-
and-Sum (DMAS) can generate images with a finer resolution and lower sidelobes compared to
DAS 83. Double stage DMAS (DS-DMAS) algorithm offers finer resolution, and lower sidelobes
80, 82, 84
than DMAS; DS-DMAS uses two stages of correlation process to suppress off-axis signals.
Minimum variance (MV) significantly improves the image resolution compared to DAS, DMAS
and DS-DMAS, however the produced images have high sidelobes; this problem has been
78, 85
addressed in MV-DMAS . The Eigen-space version of MV-DMAS, i.e., EIMV-DMAS,
25
79
provides similar resolution to MV-DMAS with lower sidelobes . Methods that use MV
algorithm, are sensitive to the quality of the received data; this issue limits the application of MV-
based algorithms for PA image reconstruction, because PA signals are usually weak. Among the
image reconstruction algorithms described above (see Table 6), DS-DMAS is the most suitable
one for linear-array based PA image reconstruction, because it reconstructs images with fine
Four wire phantoms were prepared for this study (Figure 12A). Transducers were securely held
by a clamp which was attached to an x-y translation stage. The phantom container was fixed to the
optical table and filled with distilled water. Each phantom was imaged 50 times. Figures 12(B-E)
demonstrate images of the four wire phantoms taken with the L7-4 probe where (i) is the US image,
(ii) is the PA image reconstructed in the Vantage system, and (iii) is the PA image reconstructed
with DS-DMAS. Figures 12(F-I) demonstrate images of the four wire phantom taken with the L22-
14v probe where (i) is the US image, (ii) is the PA image reconstructed in the Vantage system, and
26
Fig. 12 Comparison of the performance of DS-DMAS to Vantage PA reconstruction method. (A) Wire phantoms for
resolution study, (i) one-wire, (ii) two-wire, (iii) two-wire cross, and (iv) three-wire. (B-I) US and PA images produced
by L7-4 and L22-14v probes where the image reconstruction is: (i) Vantage default US, (ii) Vantage default PA, and
(iii) photoacoustic DS-DMAS. L7-4 probe: (B) single wire cross section, (C) 3-wire cross section, (D) 2-wire cross
section, and (E) 2-wire cross section. L22-14v probe: (F) single wire cross section, (G) 2-wire cross section, (H) 2-
wire cross section, and (I) 3-wire cross section. Images of the targets are shown in white dotted circles or straight
lines.
27
7 Conclusions
Photoacoustic imaging is a complement to the already established US imaging technique and may
significantly increase its scope of application in diagnostic imaging and therapeutic monitoring.
Combining with commercial medical US systems, the development of PAI can be accelerated by
linear-array PAI is becoming popular, mainly because linear-array US transducers can easily be
manufactured, hence the production cost is lower as compared to custom-made curved or ring
arrays 88, 89. Moreover, these transducers are commonly used in clinical applications, implying the
We discussed technical considerations for US/PA imaging implementation using the Vantage
research system. With the information we presented in the body of this technical note and the four
appendices, we described most of the experimental considerations one should know when working
with the Vantage system for PAI tests. Although the Verasonics Vantage System has many
capabilities and advantages, it has some limitations. The transducer connector limits which
transducers can directly connect to the system. Verasonics offers some connector adapters to
alleviate this problem, but it can make switching between transducers cumbersome. Another
limitation is that the signal pre-amplification occurs far (~1-2 meters) from the probe which
increases the noise level for weak PA signals. Further, the maximum sampling rate of the system
Measuring the absorption spectrum of the imaging target, i.e., amplitude of the optical energy
versus wavelength, is essential for the analysis of results in PA experiments. Here we explain a
method of how to extract the absorption spectra of two example imaging targets, vinyl black tape
28
and carbon pencil lead. The internal energy readings are not proportional to the energy deposited
on the sample due to dispersion, i.e., variable attenuation of wavelengths in the optical fiber. Using
the PhocusMobil (Opotek, CA, USA) laser, that has a built-in energy meter, we first obtained the
City, Canada) to record the energy after the illumination fiber and dividing it by the internal energy
reading inside the PhocusMobil. This yields the relation between the internal Opotek energy
reading to the energy deposited on the sample. We imaged a 2 mm carbon lead phantom and a
black vinyl tape phantom, and simultaneously recorded the internal laser energies. The PA data
was compensated by dividing the raw PA signals by the product of the energies recorded with each
phantom and the energy ratio. The absorption spectrum of the black tape and carbon pencil lead
after performing a fourth order polynomial fit are shown in Figure 13A and B, respectively.
Fig. 13 Measurement of absorption spectrum of two commonly used imaging targets: (A) black tape, and (B) carbon
pencil lead.
29
Appendix B: MATLAB Codes for Data Acquisition, Processing, Image Reconstruction
MATLAB code for the Vantage system for real-time data acquisition, data processing, image
reconstruction, and display in order of variable appearance in the script is provided as a link in
“Code, Data, and Materials Availability” section 90. Explanation of the code is below:
Start and end depth of imaging are set for US and PA imaging with P(1).startDepth and
P(1).endDepth and P(2).startDepth and P(2).endDepth, respectively. na sets the number of steered
US angles that are transmitted for a single US frame. The default number is 7. oneway, sets whether
or not the system runs in simulation mode: 0 for simulation mode is off or 1 for simulation mode
is on. Simulation mode ignores hardware and mimics transducer elements to create virtual signals
and test if the MATLAB loader program VSX is communicating correctly with RunAcq. RunAcq
communicates receive and transmit commands to the ultrasound hardware. Flash2Qdelay is the
time between trigger input and start of acquisition in microseconds and must equal the time delay
between the flash lamp output and Q-switch output from the laser. We used an oscilloscope to find
the exact delay time and set the time value to the flash2Qdelay variable (see Table 3 for the values
we used for our system). PA_PRF is set to the pulse repetition rate of the laser. The next section
of the code involves setting up system parameters, such as data buffers, and the transducer array
specifications. For Vantage 128 system and a 128-element linear array transducer,
the two linear array transducers we used, Trans.name is set to either ‘L7-4’ or ‘L22-14v’.
ComputeTrans(Trans) populates all the attributes of the specified transducer. PData defines the
pixel grid to be manipulated by the image reconstruction software. PData(n).PDelta defines the
spacing between pixels in all dimensions, PData(n).Size defines rows, columns, and sections of
the data, and PData(n).Origin defines the x,y,z coordinates of the left corner of the reconstructed
30
image, where n is 1 for the US image and n is 2 for the PA image. RcvBuffer, InterBuffer and
ImageBuffer are used to store the US and PA raw data and reconstructed images. Transmit
waveform (TW), transmit (TX), and transmit power controller (TPC) are transmit objects. TW
structure array defines specification of a transmit waveform (type, frequency, duty cycle, duration,
and polarity), where TW(1) is for US and TW(2) for PA. The TX structure array defines the
specification of the beam characteristics of each transmit action, including which transmitters are
active in the aperture (apodization), and the delay time before transmit for each active transmitter.
With 7 US transmits, TX(1:7) is defined for US transmit events. For PA, only one TX structure is
needed, so we define TX(8) for this event. In TX(8),all transmitters are turned off for the receive-
only beamformer. TPC(1) for US and TPC(2) for PA are defined, where TPC sets the transmit
power level for each specific transmit event. Receive objects are defined next and populate all the
characteristics of the receive phase of an acquisition event. The Transmit and Receive periods both
start with the data acquisition. Next, the time-gain-control (TGC) object defines the time-gain-
compensation curve for the receive portion of the acquisition event. To define a TGC waveform,
the user specifies the TGC.CntrlPts array and TGC.rangeMax. TGC.Waveform is then synthesized
and applied to the received data. The next section describes the reconstruction protocol. Recon
structure provides the general attributes of the reconstruction, including the source and destination
buffers to use. Recon.senscutoff is a value from 0.0 to 1.0 that sets the threshold of sensitivity
below which an element is excluded from the reconstruction summation for a given pixel.
Recon.pdatanum(n) specifies the number of the PData structure that defines the pixel locations of
the reconstructed data, where n is 1 for the US image or is 2 for the PA image. Recon.RcvBufFrame
is an optional attribute that when provided, overrides the frame number specified by the ReconInfo
structures. Setting Recon.RcvBufFrame to -1 allows the last acquisition frame transferred into the
31
RcvBuffer to be used in the reconstruction and be displayed for real-time imaging.
Recon.IntbufDest and Recon.ImgBufDest specify the destination buffer and frame that will receive
the reconstructed output, respectively. Recon.RINums is a row vector that specifies the ReconInfo
structure indices associated with the most recent reconstruction. For each Recon, there is an
associated set of ReconInfo objects which contain information on how to perform the
reconstruction. This information includes which data in the data buffer to reconstruct along with
where within PData. Further, ReconInfo chooses which type of reconstruction is performed
between replace, add, or multiply intensity, where each successive reconstructed frame replaces
the previous, is added to the previous, or is multiplied by the previous, respectively. Replace
intensity is chosen for our purpose. Process objects are used to describe the type of processing to
apply to the acquired data. After defining the sequence control objects, event objects, and some
graphical user interface (GUI) controls, the script is complete for data acquisition and displaying
US and PA data.
In a PA imaging experiment, there are ultrasonic transducers and an acoustic coupling layer
between the transducers and the imaging target 91. Propagation of generated ultrasound waves from
the imaging target is least attenuated when received by the ultrasound sensor through a coupling
agent that exhibits minimal acoustic impedance mismatch with the imaging target. The coupling
92
agent helps minimize degradation and signal loss . Acoustic couplants can be characterized as
liquid, gel, semi-dry, and dry. Liquids and gels generally have lower acoustic impedance than dry
couplant materials. More details about different acoustic couplants can be found in the literature.
Here we describe details on the most widely used acoustic couplant, water. Water has a low
acoustic attenuation and impedance (1.5 MPa.s/m) which makes it suitable as an acoustic coupling
32
material. In terms of water optical properties, it absorbs over a wide range of the electromagnetic
radiation spectrum with rotational transitions and intermolecular vibrations responsible for
intramolecular vibrational transitions responsible for absorption in the infrared (≈ 200nm- 10 µm),
93-95
and electronic transitions responsible for absorption in the ultraviolet region (< 200 nm) . In
regular water (H2O), the first large absorption band occurs at around 980 nm 93, 96, 97, followed by
another band at ~1450 nm. The absorption spectrum of heavy water (D2O) is different from that
of H2O, mainly due to the heavier deuterium nucleus; absorption peaks occur at around 1000 nm,
98, 99
1330 nm, and 1600 nm . The spectral features of water absorption also depend upon the
temperature 93, 98, 100; as temperature decreases, the fraction of hydrogen-bound water molecules is
increased, causing absorption peaks to reduce in intensity, broaden in bandwidth, and shift to lower
96
energy . The optical absorption spectrum of regular water and heavy water at various
Fig. 14 Optical absorption spectrum of regular water (H2O) and heavy water (D2O) at various temperatures in NIR
region. Data for this figure has been extracted from 93, 98 and reproduced.
33
Appendix D: MATLAB Code for Simultaneous Diplay of US and PA Image
The modifications to the original script to simultaneously display US and PA images, in real-time
Table 7. Modifications to the original script to simultaneously display US and PA images, in real-time. The
modifications are indicated in red.
Original Modified
Resource.DisplayWindow(1).Title = Resource.DisplayWindow(1).Title =
mfilename; mfilename;
Resource.DisplayWindow(1).pdelta = 0.4; Resource.DisplayWindow(1).pdelta = 0.4;
ScrnSize = get(0,'ScreenSize'); ScrnSize = get(0,'ScreenSize');
DwWidth = DwWidth =
ceil(PData(1).Size(2)*PData(1).PDelta(1)/ ceil(PData(1).Size(2)*PData(1).PDelta(1)
Resource.DisplayWindow(1).pdelta); /Resource.DisplayWindow(1).pdelta);
DwHeight = DwHeight =
ceil(PData(1).Size(1)*PData(1).PDelta(3)/ ceil(PData(1).Size(1)*PData(1).PDelta(3)
Resource.DisplayWindow(1).pdelta); /Resource.DisplayWindow(1).pdelta);
Resource.DisplayWindow(1).Position = Resource.DisplayWindow(1).Position =
[250,(ScrnSize(4)-(DwHeight+150))/2, ... [250,(ScrnSize(4)-(DwHeight+150))/2, ...
DwWidth, DwHeight]; DwWidth, DwHeight];
Resource.DisplayWindow(1).ReferencePt = Resource.DisplayWindow(1).ReferencePt =
[PData(1).Origin(1),0,PData(1).Origin(3)] [PData(1).Origin(1),0,PData(1).Origin(3)
; ];
Resource.DisplayWindow(1).numFrames = 2; Resource.DisplayWindow(1).numFrames = 2;
Resource.DisplayWindow(1).AxesUnits = Resource.DisplayWindow(1).AxesUnits =
'mm'; 'mm';
Resource.DisplayWindow(1).Colormap = Resource.DisplayWindow(1).Colormap =
gray(256); gray(256);
Resource.DisplayWindow(1).splitPalette = Resource.DisplayWindow(1).splitPalette =
1; 1;
Resource.DisplayWindow(2).Title =
mfilename;
Resource.DisplayWindow(2).pdelta = 0.4;
Resource.DisplayWindow(2).Position =
[250,(ScrnSize(4)-(DwHeight+150))/2, ...
DwWidth, DwHeight];
Resource.DisplayWindow(2).ReferencePt =
[PData(1).Origin(1),0,PData(1).Origin(3)
];
Resource.DisplayWindow(2).numFrames = 2;
Resource.DisplayWindow(2).AxesUnits =
'mm';
Resource.DisplayWindow(2).Colormap =
hot;
Resource.DisplayWindow(2).splitPalette =
1;
34
Process(1).classname = 'Image'; Process(1).classname = 'Image';
Process(1).method = 'imageDisplay'; Process(1).method = 'imageDisplay';
Process(1).Parameters = Process(1).Parameters =
{'imgbufnum',1,... {'imgbufnum',1,...
'framenum',-1,... 'framenum',-1,...
'pdatanum',1,... 'pdatanum',1,...
'pgain',1.0,... 'pgain',1.0,...
'reject',10,... 'reject',10,...
'persistMethod','simple',... 'persistMethod','simple',...
'persistLevel',50,... 'persistLevel',50,...
'interpMethod','4pt',... 'interpMethod','4pt',...
'grainRemoval','none',... 'grainRemoval','none',...
'processMethod','none',... 'processMethod','none',...
'averageMethod','none',... 'averageMethod','none',...
'compressMethod','power',... 'compressMethod','power',...
'compressFactor',40,... 'compressFactor',40,...
'mappingMethod','lowerHalf',... 'mappingMethod','full',...
'display',1,... 'display',1,...
'displayWindow',1}; 'displayWindow',1};
Disclosures
Acknowledgments
We would like to thank Verasonics Inc., Kirkland, WA technical support team, especially Mike
Vega, for their suggestions and comments. We are also grateful to the support team from
35
PhotoSound Technologies Inc., especially Dr. Peter Brecht, Houston, TX, for their prompt
response and help to integrate the amplifier with the Vantage system
https://codeocean.com/capsule/5228184/tree /code/code_appendixB.txt
References
1. Y. Zhou, J. Yao and L. V. Wang, Journal of biomedical optics 21 (6), 061007 (2016).
2. D. Pan, B. Kim, L. V. Wang and G. M. Lanza, Wiley Interdisciplinary Reviews: Nanomedicine
and Nanobiotechnology 5 (6), 517-543 (2013).
3. J. Yao and L. V. Wang, Photoacoustics 2 (2), 87-101 (2014).
4. B. Herzog and F. Sengün, Photochemical & Photobiological Sciences 14 (11), 2054-2063 (2015).
5. V. Dogra, B. Chinni, S. Singh, et al., Journal of biomedical optics 21 (6), 066019 (2016).
6. J. Gong and S. Krishnan, in Dye-Sensitized Solar Cells (Elsevier, pp. 51-81, 2019.
7. A. Fatima, K. Kratkiewicz, R. Manwar, et al., Photoacoustics, 100137 (2019).
8. C. Kim, C. Favazza and L. V. Wang, Chemical reviews 110 (5), 2756-2782 (2010).
9. M. Zafar, K. Kratkiewicz, R. Manwar and M. Avanaki, presented at the Photons Plus Ultrasound:
Imaging and Sensing 2019, 2019 (unpublished).
10. Z. Turani, E. Fatemizadeh, T. Blumetti, et al., Cancer research 79 (8), 2021-2030 (2019).
11. C. Li and L. V. Wang, Physics in medicine and biology 54 (19), R59-R97 (2009).
12. K. Kratkiewicz, R. Manwar, M. Zafar, et al., Applied Sciences 9 (21), 4505 (2019).
13. L. V. Wang, IEEE Journal of Selected Topics in Quantum Electronics 14 (1), 171-179 (2008).
14. A. Hariri, A. Fatima, N. Mohammadian, et al., Journal of biomedical optics 22 (7), 075001
(2017).
15. R. A. Kruger, C. M. Kuzmiak, R. B. Lam, D. R. Reinecke, S. P. Del Rio and D. Steed, Medical
physics 40 (11) (2013).
16. G. Ku and L. V. Wang, Optics letters 30 (5), 507-509 (2005).
17. U. Chitgupi, N. Nyayapathi, J. Kim, et al., Advanced Materials, 1902279 (2019).
18. H. F. Zhang, K. Maslov, G. Stoica and L. V. Wang, Nature biotechnology 24 (7), 848 (2006).
19. S. Mallidi, G. P. Luke and S. Emelianov, Trends in biotechnology 29 (5), 213-221 (2011).
20. S. Yang, D. Xing, Q. Zhou, L. Xiang and Y. Lao, Medical physics 34 (8), 3294-3301 (2007).
21. J. Gamelin, A. Maurudis, A. Aguirre, et al., Optics express 17 (13), 10489-10498 (2009).
22. E. Zhang, J. Laufer, R. Pedley and P. Beard, Physics in medicine and biology 54 (4), 1035
(2009).
23. H.-P. Brecht, R. Su, M. Fronheiser, S. A. Ermilov, A. Conjusteau and A. A. Oraevsky, Journal of
biomedical optics 14 (6), 064007-064007-064008 (2009).
24. L. V. Wang and S. Hu, Science 335 (6075), 1458-1462 (2012).
25. E. Z. Zhang, J. Laufer and P. Beard, presented at the Biomedical Optics (BiOS) 2007, 2007
(unpublished).
26. X. Wang, Y. Pang, G. Ku, X. Xie, G. Stoica and L. V. Wang, Nature biotechnology 21 (7), 803-
806 (2003).
36
27. C. Li, A. Aguirre, J. Gamelin, A. Maurudis, Q. Zhu and L. V. Wang, Journal of biomedical optics
15 (1), 010509-010509-010503 (2010).
28. W. Lu, Q. Huang, G. Ku, et al., Biomaterials 31 (9), 2617-2626 (2010).
29. J. Laufer, P. Johnson, E. Zhang, et al., Journal of biomedical optics 17 (5), 0560161-0560168
(2012).
30. S. Sethuraman, J. H. Amirian, S. H. Litovsky, R. W. Smalling and S. Y. Emelianov, Optics
express 16 (5), 3362-3367 (2008).
31. B. Wang, J. L. Su, A. B. Karpiouk, K. V. Sokolov, R. W. Smalling and S. Y. Emelianov, IEEE
Journal of selected topics in Quantum Electronics 16 (3), 588-599 (2010).
32. S. Sethuraman, S. R. Aglyamov, J. H. Amirian, R. W. Smalling and S. Y. Emelianov, IEEE
transactions on ultrasonics, ferroelectrics, and frequency control 54 (5), 978-986 (2007).
33. S. Hu, K. Maslov, V. Tsytsarev and L. V. Wang, Journal of biomedical optics 14 (4), 040503-
040503-040503 (2009).
34. X. Wang, X. Xie, G. Ku, L. V. Wang and G. Stoica, Journal of biomedical optics 11 (2), 024015
(2006).
35. G. Ku, X. Wang, X. Xie, G. Stoica and L. V. Wang, Applied Optics 44 (5), 770-775 (2005).
36. M. Zafar, K. Kratkiewicz, R. Manwar and M. Avanaki, Applied Sciences 9 (3), 374 (2019).
37. K. Kratkiewicz, R. Manwar, A. Rajabi-Estarabadi, et al., Sensors 19 (12), 2815 (2019).
38. M. Nasiriavanaki, J. Xia, H. Wan, A. Q. Bauer, J. P. Culver and L. V. Wang, Proceedings of the
National Academy of Sciences 111 (1), 21-26 (2014).
39. J. Yao, J. Xia, K. I. Maslov, et al., Neuroimage 64, 257-266 (2013).
40. M. Nasiriavanaki, J. Xia, H. Wan, A. Q. Bauer, J. P. Culver and L. V. Wang, presented at the
Photons Plus Ultrasound: Imaging and Sensing 2014, 2014 (unpublished).
41. A.-R. Mohammadi-Nejad, M. Mahmoudzadeh, M. S. Hassanpour, et al., Photoacoustics (2018).
42. E. Z. Zhang, B. Povazay, J. Laufer, et al., Biomedical optics express 2 (8), 2202-2215 (2011).
43. Z. Chen, E. Rank, K. M. Meiburger, et al., Scientific Reports 7 (1), 17975 (2017).
44. D. Xu, S. Yang, Y. Wang, Y. Gu and D. Xing, Biomedical optics express 7 (6), 2095-2102
(2016).
45. M. Li, J. Oh, X. Xie, et al., Proceedings of the IEEE 96 (3), 481-489 (2008).
46. L. Nie, X. Cai, K. Maslov, A. Garcia-Uribe, M. A. Anastasio and L. V. Wang, Journal of
biomedical optics 17 (11), 110506-110506 (2012).
47. S. A. Ermilov, T. Khamapirad, A. Conjusteau, et al., Journal of biomedical optics 14 (2), 024007-
024007-024014 (2009).
48. J. A. Copland, M. Eghtedari, V. L. Popov, et al., Molecular Imaging & Biology 6 (5), 341-349
(2004).
49. S. Manohar, S. E. Vaartjes, J. C. van Hespen, et al., Optics express 15 (19), 12277-12285 (2007).
50. A. A. Oraevsky, A. A. Karabutov, S. V. Solomatin, et al., presented at the BiOS 2001 The
International Symposium on Biomedical Optics, 2001 (unpublished).
51. S. Hu, B. Rao, K. Maslov and L. V. Wang, Optics letters 35 (1), 1-3 (2010).
52. S. Jiao, M. Jiang, J. Hu, et al., Optics express 18 (4), 3967-3972 (2010).
53. B. Wang, J. L. Su, J. Amirian, S. H. Litovsky, R. Smalling and S. Emelianov, Optics express 18
(5), 4889-4897 (2010).
54. K. Jansen, M. Wu, A. F. W. van der Steen and G. van Soest, Optics Express 21 (18), 21472-
21484 (2013).
55. Y. Deng, N. C. Rouze, M. L. Palmeri and K. R. Nightingale, IEEE transactions on ultrasonics,
ferroelectrics, and frequency control 64 (1), 164-176 (2017).
56. V. Bindal, (2000).
57. H. Pennes, Int. J. Radiation Oncology Biol. Phys. 25, 289-297 (1948).
58. M. Breazeale and O. Leroy, Physical Acoustics: Fundamentals and Applications. (Springer
Science & Business Media, 2012).
59. R. W. Cole, T. Jinadasa and C. M. Brown, Nature protocols 6 (12), 1929 (2011).
37
60. W. H. Blanton, (East Tennessee State University), 2009.
61. J. Xia, (Verasonics Inc., Verasonics Inc.), 2019.
62. S. Mahmoodkalayeh, M. A. Ansari and M. Nasiriavanaki, presented at the Photons Plus
Ultrasound: Imaging and Sensing 2018, 2018 (unpublished).
63. H. Akima, Journal of the ACM (JACM) 17 (4), 589-602 (1970).
64. M. H. Rayyan Manwar, Ali Hariri, Karl Kratkiewicz, Shahryar Noei, Mohammad R. N. Avanaki
Photoacoustic Sensing and Imaging in Biomedicine, MDPI Sensors (2018).
65. M. Zafar, R. Manwar, K. Kratkiewicz, et al., presented at the Photons Plus Ultrasound: Imaging
and Sensing 2019, 2019 (unpublished).
66. V. Ivanov, edited by K. Kratkiewicz (PhotoSound Technologies Inc, PhotoSound Technologies
Inc), pp. 20, 2018.
67. Z. Yuan and H. Jiang, Applied physics letters 88 (23), 231101 (2006).
68. S. R. Arridge, Inverse problems 15 (2), R41 (1999).
69. K. D. Paulsen and H. Jiang, Medical Physics 22 (6), 691-701 (1995).
70. S. Prahl, (OMLC), 2018.
71. L. Zhao, M. Yang, Y. Jiang and C. Li, Journal of Innovative Optical Health Sciences 10 (04),
1740002 (2017).
72. Q. Fang and D. A. Boas, Optics express 17 (22), 20178-20190 (2009).
73. P. Omidi, M. Diop, J. Carson and M. Nasiriavanaki, presented at the Photons Plus Ultrasound:
Imaging and Sensing 2017, 2017 (unpublished).
74. M. Mozaffarzadeh, A. Mahloojifar, M. Nasiriavanaki and M. Orooji, presented at the Photonics
in Dermatology and Plastic Surgery 2018, 2018 (unpublished).
75. P. Omidi, M. Zafar, M. Mozaffarzadeh, et al., Applied Sciences 8 (9), 1570 (2018).
76. M. H. Heidari, M. Mozaffarzadeh, R. Manwar and M. Nasiriavanaki, presented at the Photons
Plus Ultrasound: Imaging and Sensing 2018, 2018 (unpublished).
77. M. Mozaffarzadeh, A. Mahloojifar, M. Nasiriavanaki and M. Orooji, presented at the Photons
Plus Ultrasound: Imaging and Sensing 2018, 2018 (unpublished).
78. M. Mozaffarzadeh, A. Mahloojifar, M. Orooji, K. Kratkiewicz, S. Adabi and M. Nasiriavanaki,
Journal of biomedical optics 23 (2), 026002 (2018).
79. M. Mozaffarzadeh, A. Mahloojifar, V. Periyasamy, M. Pramanik and M. Orooji, IEEE Journal of
Selected Topics in Quantum Electronics 25 (1), 1-8 (2019).
80. M. Mozaffarzadeh, A. Hariri, C. Moore and J. V. Jokerst, Photoacoustics 12, 22-29 (2018).
81. M. Mozaffarzadeh, V. Periyasamy, M. Pramanik and B. Makkiabadi, Journal of biomedical
optics 23 (12), 121604 (2018).
82. M. Mozaffarzadeh, A. Mahloojifar, M. Orooji, S. Adabi and M. Nasiriavanaki, IEEE
Transactions on Biomedical Engineering 65 (1), 31-42 (2018).
83. G. Matrone, A. S. Savoia, G. Caliano and G. Magenes, IEEE transactions on medical imaging 34
(4), 940-949 (2015).
84. M. Mozaffarzadeh, M. Sadeghi, A. Mahloojifar and M. Orooji, Ultrasound in medicine & biology
44 (3), 677-686 (2018).
85. M. Mozaffarzadeh, A. Mahloojifar and M. Orooji, presented at the Digital Optical Technologies
2017, 2017 (unpublished).
86. S. Park, A. B. Karpiouk, S. R. Aglyamov and S. Y. Emelianov, Optics letters 33 (12), 1291-1293
(2008).
87. J. Park, S. Jeon, J. Meng, L. Song, J. S. Lee and C. Kim, Journal of biomedical optics 21 (3),
036010 (2016).
88. T. L. Szabo and P. A. Lewin, Journal of Ultrasound in Medicine 32 (4), 573-582 (2013).
89. B. A. Angelsen, H. Torp, S. Holm, K. Kristoffersen and T. Whittingham, European Journal of
Ultrasound 2 (2), 151-164 (1995).
90. K. K. M. R. Z. M. M. A. Kamran, (Code Ocean Inc., Code Ocean), 2020.
91. N. Meimani, N. Abani, J. Gelovani and M. R. Avanaki, Photoacoustics 7, 27-35 (2017).
38
92. J. Nyholt and G. N. Langlois, (Google Patents, 2013.
93. J. A. Curcio and C. C. Petty, JOSA 41 (5), 302-304 (1951).
94. S. Matcher, M. Cope and D. Delpy, Physics in Medicine & Biology 39 (1), 177 (1994).
95. F. M. Sogandares and E. S. Fry, Applied Optics 36 (33), 8699-8709 (1997).
96. S. Chung, A. Cerussi, S. Merritt, J. Ruth and B. Tromberg, Physics in Medicine & Biology 55
(13), 3753 (2010).
97. M. F. Bakhsheshi and T.-Y. Lee, (Taylor & Francis, 2015.
98. A. Tam and C. Patel, Applied Optics 18 (19), 3348-3358 (1979).
99. S. A. Sullivan, JOSA 53 (8), 962-968 (1963).
100. S. Mahmoodkalayeh, H. Z. Jooya, A. Hariri, et al., Scientific Reports 8 (1), 4873 (2018).
imaging. Karl received his Bachelor's in Biomedical Physics Summa Cum Laude and Master's in
University of Illinois at Chicago, Chicago, IL, USA. He worked as Post Doctoral Research Scholar
at Wayne State University, Detroit MI, USA from 2017-2020. He received his bachelor’s is in
Electrical and Electronic Engineering from Islamic University of Technology (IUT), Bangladesh
in 2011 and PhD from University of Windsor, Windsor, Ontario in 2017. His research interests
include MEMS based design, fabrication and characterization, photoacoustic and ultrasound
imaging.
Yang Zhou received his Master of Engineering (ME) degree in Mechanical Engineering at Wayne
State University, Detroit Michigan and currently enrolled as a PhD student in Biomedical
Engineering department. He received his Bachelor of Engineering degree from Xi’an Jiaotong
Moein Mozaffarzadeh received his BSc degree in electrical engineering from Babol Noshirvani
University of Technology, Mazandaran, Iran, in 2015, and his MSc degree in biomedical-
bioelectric engineering from Tarbiat Modares University, Tehran, Iran, in 2017. He is currently a
39
PhD student in the Department of Imaging Physics, Delft University of Technology. His research
at University of Illinois at Chicago, Chicago, IL, USA. He received a Ph.D. with Outstanding
Achievement in Medical Optical Imaging and Computing from the University of Kent in the
United Kingdom. His bachelor's and master's degrees with honors are in electronics engineering.
in the OILab.
Caption List
Fig. 1 Absorption spectrum of (a) endogenous, and (b) exogenous contrast agents. Reproduced
3-5 A = ( ) Cd
with permission from . According to Beer-Lambert law, absorbance is defined by
where, ( ) , C , and d are molar absorptivity, concentration, and cross-section thickness of the
Fig. 2 Principle of PA signal generation, detection, and image reconstruction. Reprinted with
permission from Ref. 7. A high energy and short pulse laser light illuminates the absorber, leading
to a transient temperature rise which results in a thermoelastic expansion of the absorber, and
acoustic (or PA) wave generation. The signals generated from the waves received by an ultrasound
Fig. 3 Vantage system architecture and data acquisition sequencing. (a) Vantage system
architecture for US/PA sequencing and (b) timing diagram of US/PA sequencing in Vantage
system. US: Ultrasound, PA: Photoacoustic, Recon.: Reconstruction. RcvBuffer is the predefined
40
buffer where the US and PA data are stored, Acq: acquisition, TX: transmit, RX: receive, A/D:
analog to digital, VSX: Verasonics script execution, GUI: graphical user interface, RcvBuffer:
Fig. 4 Photoacoustic signal in (A) time-domain, and (B) frequency-domain, obtained from a 2 mm
diameter carbon lead phantom, imaged with L22-14v transducer at 690 nm wavelength. The red
signal in A shows the overlaid US signal acquired from the same sample.
Fig. 5 US/PA resolution study. (A) Schematic of the experimental setup, including a hair phantom
Resolution study when (B) L7-4 was used, (C) L22-14v was used. (i) US axial resolution versus
depth, (ii) US lateral resolution versus depth, (iii) PA axial resolution versus depth, (iv) PA lateral
Fig. 6 Photoacoustic signal “jitter” comparison between two triggering methods when a 2 mm
carbon lead was imaged. 64th element signals are plotted. (A) (i) Schematic of straightforward
triggering method where laser triggers Vantage system, (ii) schematic of the function generator
driven triggering method where function generator triggers laser flash lamp and Vantage system,
which then triggers laser Q-switch, (B) (i) five sequential PA frames showing significant “jitter”
between frames, (ii) five sequential PA frames showing nearly no “jitter” between frames, and (iii)
Fig. 7 Impact of illumination angle. (A) Experimental setup of illumination angle investigation.
(B) PA signal profile from 64th element of L7-4 probe taken from a 2 mm carbon lead phantom
Fig. 8 PA signal profile from 64th element of L7-4 probe taken from a black tape phantom of 18mm
width with varying sampling rates: (A) 20.8 MHz, (B) 41.6 MHz, (C) 62.4 MHz, and (D)
41
xnor = x ( ymax ) sample no.
interpolated signal using ‘spline’ algorithm. Here, and
Fig. 9 US image quality improvement. (A) Experimental setup showing an ultrasound probe and
a 3-carbon lead phantom. (L1) 2 mm diameter carbon lead, (L2) 0.9 mm diameter carbon lead,
(L3) 0.2 mm diameter carbon lead, (B-F) US images of the 3 carbon lead phantom for 8, 16, 32,
64, and 128 steered angles, respectively. Green dashed circle encloses object pixels and red dashed
box encloses both object and background pixels, (G) CNR versus number of steered angles.
Fig. 10 PA signal amplification. (A) System setup to test the Photosound AMP128-18 amplifier:
(i) display monitor, (ii) Nd:YAG laser head, (iii) fiber optic bundle, (iv) laser power supply, (v)
laser chiller. (B) Wire imaging results: wire phantom suspended in 0% intralipid imaged (i) without
and (iv) with amplifier, wire phantom suspended in 25% intralipid imaged (ii) without and (v) with
amplifier, and wire phantom suspended in 50% intralipid image (iii) without and (vi) with
amplifier. (C) Raw data from the 64th channel of the linear array when imaging a single wire in
50% intralipid concentration to demonstrate SNR improvement from (i) without amplifier to (ii)
with amplifier, (D) Bar chart demonstrating signal amplitude increase corresponding to phantom
Fig. 11 Fluence Compensation algorithm validation. (A) Schematic of a phantom with 3 layers of
gelatin phantom mixed with intralipid/ink solution and 3 carbon lead imaging targets, (B)
photograph of the phantom made in a cube box and magnified to show the layers, (C) US image
of the phantom, (D) PA image before fluence compensation, (E) Monte Carlo simulation of light
Wire phantoms for resolution study, (i) one-wire, (ii) two-wire, (iii) two-wire cross, and (iv) three-
42
wire. (B-I) US and PA images produced by L7-4 and L22-14v probes where the image
reconstruction is: (i) Vantage default US, (ii) Vantage default PA, and (iii) photoacoustic DS-
DMAS. L7-4 probe: (B) single wire cross section, (C) 3-wire cross section, (D) 2-wire cross
section, and (E) 2-wire cross section. L22-14v probe: (F) single wire cross section, (G) 2-wire
cross section, (H) 2-wire cross section, and (I) 3-wire cross section. Images of the targets are shown
Fig. 13 Measurement of absorption spectrum of two commonly used imaging targets: (A) black
Fig. 14 Optical absorption spectrum of regular water (H2O) and heavy water (D2O) at various
temperatures in NIR region. Data for this figure has been extracted from 93, 98 and reproduced.
Table 1 High frequency Vantage 128 system specifications (Verified by Verasonics’ technical
support).
Table 2 Specifications of linear array transducers ATL Philips L7-4, and Verasonics L22-14v.
Table 3 Modifications to original L11-4 script by Verasonics, for ATL Phillips L7-4 and
Table 5 Script modifications to change the sampling rate of L7-4 from 20.8 MHz to 62.4 MHz.
Table 7 Modifications to the original script to simultaneously display US and PA images, in real-
43