Patient Lifter - Hill Rom - Viking - M-L-XL
Patient Lifter - Hill Rom - Viking - M-L-XL
Instruction Guide
English
7EN137105-02
2013-12-16
(Viking M)
Product Description
Viking M, L and XL are three lift models intended for The wireless hand control gives the caregiver the
most common transfers, for instance, transfers between freedom to work closer to the patient, which provides
bed and wheel chair, to and from toilet and bath-tub, extra security in the lifting situation. For the caregiver
or for lifting from and to the floor. A Viking equipped this also enables an ergonomically enhanced work
with Armrests can also be used for patients’ gait training. posture.
All three lift models are equipped with Liko Diagnostic Individual fitting of the sling and other lifting accessories
System™, a monitoring system which continually is of utmost importance for function and safety
registers operating data. when using Liko’s lifts.
On the battery of the lift the battery and charging status Viking M, L, and XL have manual as well as electrical
can be read, even when the battery is not attached emergency lowering.
to the lift. The battery is of environmentally friendly
NiMH type.
In this document, the person being lifted is referred to as the patient, and the person helping them is referred to
as the caregiver.
is a warning triangle used for situations which require extra care and attention.
IMPORTANT!
Read the instruction guide for both the patient lift and lifting accessories before use. Lifting and transferring
a person always involves a certain level of risk. It is important to completely understand the contents of the
instruction guides. The equipment should only be used by trained personnel. Please contact your Liko/Hill-Rom
representative in the event of any uncertainties or questions.
Table of Contents
Safety Instructions.........................................................................2
Definitions......................................................................................3
Technical Data...............................................................................3
Dimensions....................................................................................4
Assembly.......................................................................................5
Operation.......................................................................................6
HandControl IR .............................................................................8
Operation Panel ...........................................................................9
Charging the Batteries.................................................................10
Maximum Load............................................................................11
Recommended Lifting Accessories.............................................11
Simple Troubleshooting...............................................................13
Care and Maintenance................................................................14
NOTE!
This instruction guide contains important information for the user of the product.
All who use the product should review and completely understand the contents of the instruction guide.
Remember to keep the instruction guide in a place where it is always available to those using the product.
Safety Instructions
Before using, make sure that:
• the lift is assembled in accordance with the assembly instructions;
• the lifting accessory is properly attached to the lift;
• the batteries have been charged for at least 5 hours;
• you have read the instruction guides for the lift and lifting accessories;
• personnel using the lift are informed of the correct operation and use of the lift.
Viking M, L, and XL have been tested by an accredited testing institute and comply with the requirements specified
in the Medical Device Directive 93/42/EEC, for Class 1 products.
Viking M, L, and XL comply with the requirements in EN ISO 10535, IEC 60601-1, IEC 60601-1-2, ANSI/AAMI
ES60601-1, and CAN/CSA C22.2 no 60601-1.
The lift must not be modified under any circumstance. If you have any questions, please contact Liko/
Hill-Rom.
Particular care must be taken when using powerful sources of electromagnetic disturbance, such as diathermy, etc,
so that, for example, diathermy cables are not positioned on or near the lift. If you have any questions, please consult
the responsible assistive-device technician or the supplier.
The lift may not be used in areas where flammable mixtures may occur, for example in areas where flammable
goods are stored.
Technical Data
Maximum load: Viking M: 205 kg (450 lbs) Sound Level: Viking M: 46 dB(A)
Viking L: 250 kg (550 lbs) Viking L: 51 dB(A)
Viking XL: 300 kg (660 lbs) Viking XL: 51 dB(A)
Material: Aluminium Protection Viking M: IP X4
Class: Viking L: IP X4
Weight: Total weight: Heaviest part:
Viking XL: IP X4
Viking M 34.0 kg (75 lbs) 15.3 kg (34 lbs)
Viking L 36.5 kg (80 lbs) 20.5 kg (45 lbs) Operating HandControl: 4 N
Viking XL 38.3 kg (84 lbs) 21.3 kg (47 lbs) Forces Operation Panel: 4 N
Wheels: Viking M of Controls:
Front: 75 mm (3 inch.) twin wheels. HandControl Batteries: 2 AAA LR03 1.5 V.
Rear: 75 mm (3 inch.) twin wheels with brakes. IR Reach: 2 m (80 in)
Viking L, XL Electrical Data: 24 V
Front: 100 mm (4 in) twin wheels.
Intermittent Int. Op 10/90, active operation max 2 min.
(Or 75 mm* (3 in) twin wheels).
operation: Only 10% of a given length of time may
Rear: 125 mm (5 in) twin wheels. be active, yet no more than 2 min.
(Or 100 mm (4 in) single castors with brakes).
Batteries: NiMH cells, 2.2 Ah. New batteries provided
Turning Viking M: 1400 mm (55.2 in) by the supplier.
diameter: Viking L: 1460 mm (57.4 in)
Battery CBL20001, built-in, 100-240 V AC,
Viking XL: 1570 mm (61.8 in)
Charger: 50-60 Hz, max 600 mA.
Emergency Manual and electrical Lift Motor: Permanent magnet motor with mechanical
Lowering: safety mechanism
Lifting interval: Viking M: 1270 mm (50.0 in) Viking M: 24 V, 8.5 A
Viking L: 1330 mm (52.4 in) Viking L: 24 V, 6.4 A
Viking XL: 1370 mm (54.0 in) Viking XL: 24 V, 8.0 A
Lifting Speed Viking M: 42 mm/s and 38 mm/s, respectively Motor for
(no load): Viking L: 21 mm/s and 13 mm/s, respectively base-width Permanent magnet motor
Viking XL: 21 mm/s and 13 mm/s, respectively adjustment: 24 V, 6 A
(Viking M: 1.7 in/s and 1.5 in/s, respectively)
(Viking L: 0.8 in/s and 0.5 in/s, respectively) The device is intended for use indoors
(Viking XL: 0.8 in/s and 0.5 in/s, respectively)
Type B, in accordance with the electrical shock protection class.
Class II equipment.
Amax.
O F F1
Lmin D2
Viking M Measurements in mm
C D
Amax.* Amin.* B B1 B2 B3*** D2** E* F F1 Lmax* Lmin* M N O
max. min. max. min.
2020 1440 650 1100 1790 520
1970 1390 1230 870 660 650 1110 690 970 560 195 1050 105 30 1740 470 385 220 495
1920 1340 595 1000 1690 420
Viking L Measurements in mm
C D
Amax Amin B B1 B2 B3** D2** E F F1 Lmax Lmin M N O
max min max min
2030 1390 1360 900 630 560 1070 695 950 625 230 1060 165 43 1790 460 500 280 550
Viking XL Measurements in mm
C D
Amax. Amin. B B1 B2 B3** D2** E F F1 Lmax Lmin M N O
max. min. max. min.
2110 1500 1420 960 600 560 1130 815 1015 605 145 1010 170 40 1910 540 557 365 665
Note: The measurements are based on the lift being equipped with standard sling bar and standard wheels. When changing to other lifting
accessories,or other wheels, check that the lift still achieves desired lifting height.
* Different measurements apply to Viking M depending on the height-setting position, see ”Assembly”, page 5.
** Reference measurement according to Standard EN ISO 10535:2006.
1. Lock both rear wheels. 2. Secure the lift mast in the locking handles provided. Adjust the position of the locking
Place the lift mast handles so that they point downwards. NOTE! Viking M has an adjustable lift mast
in the foot of the base. which can be set to three different levels. The middle hole is recommended in most
cases. The lowest hole of the lift mast is recommended for extra high lifting height. The
highest hole is recommended for lower lifting heights. The distance between two holes
is 50 mm (2 in). See the measurement table, page 4.
”click”
1
2 3
3. Connect the cables as follows: 4. Thread the cables through 5. Place the battery in its bracket
- Charger cable to socket 1 the opening in the cable lid. in the control box. Check that
- Cable from lift motor to socket 2. Push the lid up and secure with the the battery is secured
- Cable from motor for base-width provided Allen key and screws (2). in its position.
adjustment to socket 3.
A C
B D
6. Hang the hand control on the green ”button” fixed 7. Attachment of spiral cord for HandControl IR
to the lift mast. Two extra adhesive ”buttons”
are provided for hanging up the hand control.
These can be placed anywhere on the lift.
10. B
efore the first use, charge the battery by plugging 11. Place the charger cable on the hook provided
the charger cable into the connector cable. Then plug on the mast after completed charging.
the charger cable into an electric outlet (100-240 VAC).
The lamps on the control box front flash alternatively
(green/orange/orange) during charging. The battery is
fully charged after approx. 5 hours.
Operation
Unlock
Unlock Lock
Lock
Figure 1. Figure 2.
Installation of Latches Lift correctly!
After installation, ensure that Before each lift, make sure that:
the spring loaded latches is taut – the Sling loops at opposite sides of the Sling are at the same height
against the sling bar and moves – all the Sling loops are fastened securely in to the Sling bar hooks
freely in the sling bar hook. – the Sling bar is level during the lift, see Figure 1.
If Sling bar is not level (see Figure 2), lower the user to a firm
surface and adjust according to the Sling in use Instruction Guide.
An improper lift can be uncomfortable for the user and cause
damage to the lift equipment! (see Figure 2).
Before Use
HandControl IR
Always enable the HandControl IR with the lift in question, see Enable IR control.
The hand control has 6 buttons:
Up
Down
nable IR control:
E
Up (lower speed)
Hold the HandControl IR approx. 20 cm (8 in) from the
Down (lower speed) control box’s IR receiver and press the hand control’s
large Up and Down buttons simultaneously for approx.
Base widening 2 sec. When a sound signal is emitted, the function is
enabled.
Base narrowing
The lift is now ready to be used with the HandControl IR.
If the lift has not been used for more than 24 hours, the system must be enabled by pressing any button
on the control box operation panel before the HandControl IR can be used.
One HandControl IR can control different patient lifts from Liko, equipped with the same IR system.
For more information, read the section ”Communication Mode” on page 9.
Up
ELT, Expected Life Time (red light)
Down - when the light flashes during lifting, the Liko Diagnostic
System™ indicates that the expected life time for the lift
Up (lower speed) is approaching its end.
Contact the service technician authorized by Liko/Hill-Rom.
Down (lower speed)
If the ELT proceeds to shine continuously when lifting,
Base widening
the lift should be taken out of service.
Base narrowing Contact the service technician authorized by Liko/Hill-Rom.
If the lift has not been used for more than 24 hours, the system must be enabled by pressing any button
on the control box operation panel before the HandControl IR can be used.
2
Battery Capacity
Check the battery capacity by pressing the i-button (1). The following
3
information can be read:
2 - Green light: More than 50% of the battery capacity remains.
4 3 - Orange light: 25–50 % of the battery capacity remains.
4 - Orange light: 0–25 % of the battery capacity remains.
The indication lamps light up for about half a minute; and are then turned off.
1 The battery capacity is also shown when the lift is in use.
To obtain max. service life, it is important to charge the batteries regularly. We recommend charging after
use of the lift or every night. Maximum charge is obtained after about 5 hours. When the batteries are fully
charged, the charger switches to “trickle charging” automatically.
During charging the lights shine alternatively (orange, orange, green). For deeply discharged batteries,
the indication for initiated charging can be delayed a few minutes. Completed charging is indicated
by a flashing green light. When the operation time is noticeably shortened, the batteries probably
need changing. Stop charging and switch batteries.
The lift cannot be charged with the emergency stop engaged.
Never charge batteries in a wet area.
NOTE! If the lift is not to be used for a long period, the battery should be connected to charging.
Charging
NOTE! The lift cannot be used when the charger cable is plugged into an electrical outlet.
Stretchers
Viking M, L, and XL can be used in combination with some Liko
Stretchers for horizontal lifting.
Contact Liko/Hill-Rom for further information.
Scale
For weighing patients in combination with Viking, we recommend using
LikoScale 350 (adapter 12 mm is required). Max 350 kg (770 lbs).
LikoScale 350 is certified according to the European Directive
NAWI 90/384 (Non-Automatic Weighing Instruments).
Contact Liko/Hill-Rom for further information. LikoScale 350 Adapter 12 mm
Prod. No. 3156228 Prod. No. 2016504
The lift does not work up/down 1. Make sure that the emergency stop is not engaged (page 6).
with the HandControl IR. 2. If the lift has not been used for 24 hours, enable the system
The base-width adjustment (page 8).
doesn’t work (in/out) with 3. Check that the HandControl IR is enabled (page 8).
the HandControl IR.
4. Check that the charger cable is not connected to an electric
outlet.
5. HandControl IR indicates battery change (page 7).
6. If the lift works via the operation panel,
change the HandControl IR.
7. If the problem persists, please contact Liko/Hill-Rom.
The lift does not work up/down 1. Make sure that the emergency stop is not engaged (page 6).
with the operation panel. 2. Make sure that the cables to the control box are connected
The base-width adjustment correctly (page 5).
doesn’t work (in/out) with 3. Check that the charger cable is not connected to an electric
the operation panel. outlet.
4. Check the battery capacity (page 10).
5. Check that the battery’s contact plates are not defective or
broken.
6. If the problem persists, please contact Liko/Hill-Rom.
The charger doesn’t work. 1. Make sure that the emergency stop is not engaged (page 6).
2. Check that the charging cable is correctly connected
(page 6 & 10)
3. Make sure that the battery is properly attached.
4. Check that the battery’s contact plates are not defective or
broken.
5. If the problem persists, please contact Liko/Hill-Rom.
The lift is stuck in the high 1. Make sure that the emergency stop is not engaged (page 6).
position. 2. Check the battery capacity (page 10).
3. HandControl IR indicates battery change (page 7).
4. Use the electrical emergency lowering device to lower
the patient onto a firm surface (page 6).
5. Use the manual emergency lowering device to lower
the patient onto a firm surface (page 7).
6. If the problem persists, please contact Liko/Hill-Rom.
When necessary, clean the lift with a moist cloth and check that the wheels are free from dirt. Find more
detailed information regarding cleaning and disinfection of your Liko/Hill-Rom product in the document Care and
Maintenance at our website: www.liko.com.
The lift should not be exposed to running water.
Service
A periodic inspection of the lift should be carried out at least once a year.
Periodic inspection, repair and maintenance should be performed only in accordance with the Liko
Service Manual and by personnel authorized by Liko and using original Liko spare parts.
Service Agreement
Liko offers the opportunity to enter into service contracts for the maintenance and regular inspection of your Liko
product.
Recycling
For instructions on how to recycle your Liko product, please visit our website: www.liko.com.
Product Changes
Liko’s products undergo continuous development, which is why we reserve the right to make product changes
without prior notice. Contact your Liko/Hill-Rom representative for advice and information about product upgrades.
Hill-Rom´s Management system are certified in accordance with ISO 9001 and its equivalent for the medical device
industry, ISO 13485. Hill-Rom´s Management system is also certified in accordance with environmental standard ISO
14001.
w w w . l i k o . com
Manufacturer:
Liko AB
SE-975 92 Luleå
Sweden
[email protected]