HAMILTON-G5/S1 Circuit setup, Coaxial, Adult/Ped
Setup instructions for coaxial breathing circuit for adult and pediatric patients
Installing the expiratory valve Connecting the flow sensor
Refer to the figure below. NOTE. To prevent inaccurate readings, ensure the flow
sensor tubing is not kinked.
`` Position the expiratory valve set in the expira-
1. Insert a flow sensor (1) into the breathing
tory port (1) and twist clockwise until it locks
circuit in front of the patient connection.
into place (2).
2. Attach the blue and clear tubes to the flow sensor
connection ports on the ventilator.
The blue tube attaches
Connecting a mainstream CO2 sensor to the blue connection
port. The clear tube
Refer to the figure below. attaches to the silver
connection port.
1. Connect the sensor cable to the CO2 module
(1) on the ventilator.
3. Calibrate the flow sensor and perform the
2. Attach the CO2 sensor (2) to the airway Tightness test.*
adapter (3), aligning the arrows on both
components.
Press the components together until they click. Connecting a bacterial/viral filter with HME
`` To prevent patient or ventilator contamination,
1
be sure to connect a bacterial/viral filter with
a heat and moisture exchanger (HME) (1)
2 between the patient and the flow sensor.
3 1
3. When connecting a CO2 sensor for the first
time (or when prompted), calibrate the sensor/
adapter.*
4. Connect the sensor/adapter (4) to the * For details, see the HAMILTON-G5/S1 Preoperational Check
Quick Reference Card (ELO2020-112-TW).
breathing circuit in a vertical position as
follows:
4
5. Secure the cable safely out of the way.
Be sure to enable the CO2 sensor before use in the
System > Sensors on/off window.
Positioning the breathing circuit
Follow these guidelines for positioning the assembled breathing
circuit:
• After assembly, position the breathing circuit so that the limbs/
tubing will not be pushed, pulled, or kinked as a result of a
patient‘s movement, transport, or other activities, including
scanner bed operation and nebulization.
• Position the ventilator, including the patient support arm, well
back from the breathing circuit Y-piece.
• Position the flow sensor upright, with the patient end facing
downward (see below).
Ideally, the flow sensor should be at a ≥ 45° angle relative to the
floor.
• Ensure there is no undue stress placed on any tubing or cables.
© 2020 Hamilton Medical AG. All rights reserved
ELO2020-113-TW/00 | 2020-04-07
1 To patient inspiratory port 4 Flow sensor connection ports 7 Flow sensor
2 From patient expiratory port 5 Coaxial inspiratory/expiratory limb 8 Bacterial/viral filter with HME
3 Expiratory valve set 6 CO2 sensor/adapter 9 Expiratory limb extension