Manual_P26
Manual_P26
Differential Pressure
Transmitter
Stegener Straße 10
D-79199 Kirchzarten
Email: [email protected]
Internet: www.halstrup-walcher.de
Document 7100.004084
P26 Instruction Manual
Table of Content:
Hysteresis ............................................................................................................... 19
Tv delay time .......................................................................................................... 19
Tn stopping time (from Rev2.11) ............................................................................. 20
Warning signal ........................................................................................................ 20
Filter(from Rev2.14) ................................................................................................ 20
7.5 Setting .......................................................................................................................... 20
Language ................................................................................................................ 21
Output ..................................................................................................................... 21
Filter........................................................................................................................ 21
Touchtone ............................................................................................................... 21
Resolution ............................................................................................................... 21
Zeroing ................................................................................................................... 21
SMU (creep suppression) ....................................................................................... 21
Illumination.............................................................................................................. 22
Colour change ........................................................................................................ 22
Brightness ............................................................................................................. 22
Contrast ................................................................................................................ 22
Read factory settings ............................................................................................ 22
Code? ................................................................................................................... 22
7.6 Operation ..................................................................................................................... 22
Interfaces ......................................................................................................................... 23
8.1 USB port (optional) ...................................................................................................... 23
8.2 Serial interface (via internal connector) ........................................................................ 23
8.3 List of commands ......................................................................................................... 24
Technical data ................................................................................................................. 27
Troubleshooting ............................................................................................................ 28
Dimension drawing ....................................................................................................... 29
Menu tree ....................................................................................................................... 30
Certificate of Conformity............................................................................................... 33
3
P26 Instruction Manual
The manufacturer reserves the right to continue developing this instrument model without
documenting such development in each individual case. The manufacturer will be happy to
determine whether this manual is up-to-date.
Conformity
This device is state of the art. It complies with the legal requirements of EC
directives. This is shown by the CE mark.
4
P26 Instruction Manual
Safety precautions
2.1 Appropriate use
The P26 is used to measure pressure, volumetric flow, mass flow and flow rate.
Always observe the operating requirements – particularly the permissible supply voltage –
indicated on the rating plate and in the “Technical data” section of this manual.
The instrument may only be handled as indicated in this manual. Modifications to the
instrument are prohibited. The manufacturer is not liable for damages caused by improper use
or failure to follow these instructions. Violations of this type render all warranty claims null and
void.
Do not close the pressure inlets during shipping. Changes in barometric pressure may
damage devices with low measuring ranges.
Assembly and the electrical connections should only be handled by professionals. They should
be given proper training and be authorised by the operator of the facility.
Although the P26 pressure transducer is highly robust, it is nevertheless a precision instrument
and should be handled with care. Avoid mounting the P26 in the direct vicinity of any sources
of heat or radiation. Ideally, the instrument should be mounted vertically on a wall not subject
to vibration.
The instrument may only be operated by appropriately trained individuals who have been
authorized by the operator of the facility.
Do not carry out a function test with compressed or breathable air. This would damage
instruments with low measuring ranges.
The pressure ports may only be connected and disconnected when the device is switched off.
Measurement errors may occur if the instrument is not kept protected from sunlight.
Specific safety precautions are given in individual sections of this manual.
5
P26 Instruction Manual
The individual responsible for the electrical connections must be notified if the instrument is
damaged or if errors occur that cannot be corrected as indicated in Section 10.
This individual must take the instrument out of service until the error has been corrected and
ensure that it cannot be used unintentionally.
Always unplug the power cord resp. shutdown all supply voltages before opening the
instrument!
The electronic components of the instrument contain environmentally hazardous materials and
materials that can be reused. The instrument must therefore be sent to a recycling plant when
you no longer wish to use it. The environment codes of your particular country must be
complied with.
2.4 Symbols
The symbols given below are used throughout this manual to indicate instances when
improper operation could result in the following hazards:
WARNING! This warns you of a potential hazard that could lead to bodily injury up to
and including death if the corresponding instructions are not followed.
WARNING: This warns you of a potential hazard that could lead to significant
property damage if corresponding instructions are not followed.
6
P26 Instruction Manual
Instrument description
3.1 Functions
Keyboard
µC
Valves
Serial
Display Relay inter-
face
7
P26 Instruction Manual
When the air consumption function is activated, the keys perform the following functions:
8
P26 Instruction Manual
Standard version
9
P26 Instruction Manual
Standard version
Output signal Relay1 Relay2 Power supply
10
P26 Instruction Manual
Output signal:
Relay 1:
Relay 2:
Power supply:
o 24V DC
Left contact => Ground
Central contact => +24V DC
o 24V AC
Left contact => Ground
Central contact => +24V AC
with transformer
o 115V AC
Central contact and right contact
o 230V AC
Left contact and right contact
o 24 VAC
left contact and right contact
Reset:
11
P26 Instruction Manual
USB (optional) C a b l e g l a n d s:
Signal Relay Power supply
Pressure ports: + -
RS-232:
USB:
INFORMATION!
12
P26 Instruction Manual
External zero-point calibration is also available (optional). This function is only possible in
connection with a supply voltage of 24 VDC. In this version, there is an additional contact on the
output connector. To initiate a zero-point calibration, this connection must be connected briefly
with +24 V.
INFORMATION!
If a zero-point calibration is initiated via the serial interface or the external zero-point
calibration input, this will still be carried out. In this case, it doesn’t matter if the zero-point
calibration is deactivated or the P26 is in the menu.
Overpressure protection
The P26 has an internal overpressure safeguard that protects the precision pressure
measurement capsule from damage. (Overload range: 200 x [max. 600 kPa])
Display (optional)
The measured values and the menu are shown on the display. The top line is the information
line. This line displays the type of measured value or the status of a zero-point calibration. The
measured value is displayed in the middle line. If the permitted measuring range is exceeded or
not reached, the figures are replaced by '-' characters. The bottom line displays the unit, with the
first two digits showing the status of the two relays.
13
P26 Instruction Manual
Menu (optional)
If the P26 includes a display, you can use the menu to adjust various settings. Press the Menu
key to bring up the menu.
In Menu mode, the parent menu item always appears in the top display line. The middle line
always shows the current sub-menu item or value to be changed. The bottom line displays units
or other help texts.
Start the Menu mode by pressing the "Menu" key. It can be protected with a four-character
password. At the main menu level, “Menu” is displayed in the top line and “Display” in the
middle line. Select your desired sub-menu by pressing the “▲” and “►” keys. Press “Enter” to
go to a sub-menu or enter a value. Press the “Menu” key to go to the next menu level or to stop
entering values.
7.1 Display
The display unit can be selected with this menu item. Press the “▲” and “►” keys to select the
unit you want and press “Enter” to confirm. The type of display (pressure, flow rate etc.) is
selected automatically.
When the air consumption function is activated, the selection of units that can
be displayed is limited to the following: m³/s, m³/h, ft³/h, kg/s, kg/min or kg/h.
7.2 Scale
The scaling function is not available when the air consumption function is
activated. Entering the scaling values defines the density of the medium.
Changing these values when the air consumption function is activated will result
in the calculation of incorrect totals.
The P26 is usually supplied with standard measuring ranges. You can use the scale to adjust
the measuring range for your own use. The output voltage or currents are then copied to this
scaled range. The scaled range should always be more than 0.1 times of that of the P26’s
measuring range, because otherwise the resolution of the outputs and the accuracy will be
poorer.
With this menu item, there are 4 sub-menu items:
Pressure
Volume flow
Mass flow
Flow rate
14
P26 Instruction Manual
Pressure
This is an important menu item. It is used to specify the pressure range, which indicates the
output value. This pressure range forms the basis for many other settings such as the limits for
the relays (warnings) and factors for the flow display.
There are two default settings for the pressure scale:
Top = Pressure at which the output has its maximum value (e.g. 5V, 10V or 20 mA)
Bottom – Pressure at which the output has its minimum value (e.g. -5V, 0V, 0mA or
4mA)
Because the values can be freely assigned, settings such as 0V at 0Pa or 10V at -250Pa are
also possible.
7.2.1.1 Top
With this menu item, you can set the pressure at which the output will achieve its maximum
value. Any value within the measuring range can be selected. It can also be less than the
lower scale value.
7.2.1.2 Bottom
With this menu item, you can set the pressure at which the output will achieve its minimum
value. Any value within the P26’s measuring range can be selected.
7.2.1.3 Units
Use this menu item to select the pressure unit. The following pressure units can be selected:
Pa
hPa
kPa
mbar
mmH2O
mmHg
Psi
inH2O
inHg
Volume flow
This menu item is used to adjust the volume flow display values.
The following sub-menu items are available:
Value
Units
15
P26 Instruction Manual
7.2.2.1 Value
This is used to set the volume flow value displayed at the maximum scaled pressure. 0 is
always used as the lower scale value on the volumetric flow display (root-extracted measured
value). The maximum pressure is the larger of the two scale values, which is used as a basic
calculation.
7.2.2.2 Units
You can use this menu item to select the scaling unit of the volume flow value. This unit don’t
have to be identical to the selected display unit. The following volume flow units can be
selected: m³/s, m³/h, ft³/h.
Mass flow
This menu item is used to adjust the mass flow display values.
The following sub-menu items are available:
Value
Units
7.2.3.1 Value
This is used to set the mass flow value displayed at the maximum scaled pressure. 0 is always
used as the lower scale value on the mass flow display (root-extracted measured value). The
maximum pressure is the larger of the two scale values, which is used as a basic calculation.
7.2.3.2 Units
You can use this menu item to select the scaling mass flow unit. This unit don’t have to be
identical to the selected display unit. The following mass flow units can be selected: kg/s,
kg/min, kg/h
Flow rate
This menu item is used to adjust the flow rate display values.
The following sub-menu items are available:
Value
Units
7.2.4.1 Value
This is used to set the flow rate value displayed at the maximum scaled pressure. 0 is always
used as the lower scale value on the flow rate display (root-extracted measured value). The
maximum pressure is the larger of the two scale values, which is used as a basic calculation.
7.2.4.2 Units
Use this menu item to select the scaling flow rate unit. This unit don’t have to be identical to
the selected display unit. The following flow rate units can be selected: m/s, mph, f/s, f/min,
km/h
16
P26 Instruction Manual
7.3 Air consumption function (option available from firmware version 3.0)
The air consumption function is only available if you have selected one of the following
units as the display unit:
m³/s, m³/h, ft³/h, kg/s, kg/min or kg/h.
This function records the consumed volume or mass of the medium being measured. The
settings are protected by a separate code. You can change the parameters after entering the
correct code. The code remains active until you leave the menu. After switching to the display
mode, you must re-enter the code if you wish to change the parameters.
Mode
7.3.1.1 Off
Recording of the quantity consumed only for pressures > 0. When creep suppression is
activated, only pressures greater than the amount of creep will be recorded.
7.3.1.3 Σ(ΔP)
Recording of the consumption values also takes negative differential pressures into account.
The consumption or mass values calculated from the negative pressures are deducted from
the total.
In this mode, the sign (+/-) of the differential pressures is not taken into account. In both cases,
the consumption or mass value calculated is added to the total.
17
P26 Instruction Manual
Mass/Pulse or Vol./Pulse
The recorded consumption values are signalled to the connected control module by a relay.
Every pulse corresponds to a defined volume or a definable mass. This menu item can be
used to set this volume or mass. The parameter displayed depends on the display units
selected. Adjusting the scaling feature adjusts the associated volume or mass value. This
enables you to switch between displaying mass and volume readings even while the unit is
operating. Values are adjustable between 0.1 and 10000 m³ or kg.
The pulse output can generate a pulse rate of max. 5 pulses/s. Adjust
volume/pulse or mass/pulse so that this pulse rate is not exceeded even at the
max. flow rate. If the pulse rate is exceeded, the output will continue to pulse
even when the section is not sending a signal as it attempts to transmit all the
accumulated pulses. Longer periods of activation or inactivity also reduce this
pulse rate.
Pulse output
This menu item is used to adjust the pulse length and the min. break for the volume or mass
pulses.
This parameter is used to defined the duration of the relay pulse. Values are adjustable
between 20 and 1999ms.
This parameter is used to define the min. break after a relay pulse. Values are adjustable
between 50 and 1999ms.
Filter
Either the filtered or unfiltered value can be used to record the consumption values. Use this
setting to select either the filtered or unfiltered value.
Reset sum
This function is used to reset the sum for the consumption function. The last value is saved as
backup. (Yes --> Sum = 0, No --> Sum is retained)
Old sum
This menu item is used to display the consumption values sum from before the last reset.
Recall Sum
This function can be used to restore the sum for the consumption function from before the last
sum reset. (Yes --> Recall, No -- Current value is retained.)
Security code
This function can be use to set the code for accessing the menu. You should set a security
code when the air consumption function is activated in order to prevent the device from being
manipulated. When the security code is > 0000, it is only possible to access the menu after
entering this security code.
18
P26 Instruction Manual
7.4 Warning
Use this menu item to influence the behaviour of the relays. The first menu item that appears
here is Select Relay. The settings for this relay should then be changed. Use the “▲” and “►”
keys to select the number of the relay and press Enter to move to that relay's menu.
The following sub-menu items are available:
Mode
Value
Hysteresis
Tv
Tn
Warning signal
Filter
Mode
During normal operation, the relay outputs are used to signal when values rise above or fall
below definable pressure limits. This parameter can be used to influence the behaviour of the
relays.
The following settings are available:
Off Relay not active
Increasing Relay picks up when the warning pressure is exceeded
(depends on air pressure)
Decreasing Relay picks up when the warning pressure is underrun
(depends on air pressure)
Air consumption Relay is used as a pulse output for the air consumption function
if this is activated, otherwise the relay is deactivated
Special version Software Rev, 4.01: Here the switching direction of the relays is inverted, that
means the relays drop out when the warning pressure is exceeded or underrun.
In "air consumption" mode, the other relay parameters are not used and
therefore not displayed in the menu.
Value
This item adjusts the pressure value at which the particular relay should switch. The pressure
range defined by the scale can be used as the setting range. Select the pressure unit set for
the display as the unit.
Hysteresis
This item can be used to select the hysteresis for the particular relay. If the volume flow, mass
flow or flow rate (root-extracted display) is displayed, the pressure unit used will be the unit
that was active at the time when the root-extracted display was activated. The hysteresis is
always positive and can be set between 0% and 120% of nominal pressure value.
Tv delay time
The entry value of the relays is derived from the unfiltered pressure value. With this parameter,
you can now specify how long the pressure value can be exceeded or not met until the relay
switches. The adjustment is done in ms.
19
P26 Instruction Manual
The entry value of the relays is derived from the unfiltered pressure value. This parameter can
now be used to specify how long the relay should remain active after the pressure value has
returned to the range in which the relay is not activated. The adjustment is done in ms.
Warning signal
A warning signal sounds once every second when a threshold value is exceeded. With this
parameter, you can set the duration of this warning signal for each warning. The maximum
value here is 1000 ms (continuous tone). If both threshold parameters have been exceeded,
both warning tones will sound. In this case, the duration of the warning tone will be equal to
the sum of the durations of the two separate tones but max. 1000 ms.
With firmware Rev 2.21 and later, the warning signal can be switched of by pressing any of the
front key of the P26. It will be reactivated if the pressure will pass any of the the warning levels
again.
Filter(from Rev2.14)
This item can be used to adjust whether the filtered or unfiltered pressure value is used as
source for the relay. Depending on the filter time constants, the filtered pressure values can be
used to filter out short-lived fluctuations in the pressure level.
7.5 Setting
With this menu item, you can adjust various parameters that affect the instrument’s behaviour.
The following sub-menu items are available:
Language
Output
Filter
Touchtone
Resolution
Zeroing
CS (creep suppression)
Lighting
Colour change
Brightness
Contrast
Read factory settings
Code?
20
P26 Instruction Manual
Language
The menu language can be selected with this menu item. You can choose from the following
languages:
English
German
Italian
French
Make your choice with the “▲” or "►" keys. Press Enter to confirm your selection.
Output
With this parameter, you can select the output and output range. You can choose from the
following options:
4…20 mA
0…20 mA
-5V…+5V
0V…10V
Filter
The measured pressure values can be smoothened by a filter before they reach the display or
analogue output. This parameter lets you set a time constant for this filter.
Touchtone
Resolution
This parameter affects the display resolution. The display is changed according to the adjusted
values. For example, if the setting is 1%, the display changes to showing the measuring range
in 1% jumps. This can be used with severely fluctuating pressures to steady the display.
However, this does not have any effect on the relays.
Zeroing
This parameter specifies the time interval for the automatic zero-point calibration. The entry is
in min. The parameter “AutoNull” can be asingn to values from 0 to 2999. Setting the
parameter to 0 deactivates automatic zero-point calibration.
This parameter specifies the value for creep suppression as a percentage. If this value falls
short of the measured pressure value, the display is set to zero. Creep suppression only works
with volume flow, mass flow and flow rate (root-extracted measurement values).
21
P26 Instruction Manual
Illumination
This parameter can be used to select the backlighting colour for the LCD display in the menu or
when the colour change function has been switched off. You can choose from the following
three options:
Off The background lighting is shut-off
White The background lighting is white
Blue The background lighting is blue
Colour change
This parameter can be used to specify whether or not the display background should change
colour when a threshold parameter has been exceeded. You can choose from the following two
options:
• off The colour change function is deactivated
• on The colour change function is activated
When a value specified in the list of warnings is out of range, the display will turn yellow for the
time Tv or Tn and then be backlit in red after the relay is triggered.
Brightness
This parameter lets you adjust the brightness of the background lighting. You can adjust the
values from 0 (dark) to 16 (bright).
Contrast
This parameter is used to adjust the contrast of the display. This enables you to compensate for
different temperatures or viewing angles. Values from 0 (low contrast) and
32 (high contrast) are possible.
The factory settings are established in the instrument before delivery. With this menu item,
they can be reactivated if required.
Code?
You can use this menu item to specify an access code for the menu. If this code is <>0, the
access code will only be queried after the Menu key has been pressed. Only when the correct
code has been entered can you proceed to the menu.
If the code = 0, this query will not be made.
7.6 Operation
The operating hours meter displays the recording period for the current air consumption
function. When the air consumption sum is reset to zero, the operating hours meter is also
reset to 0 h.
22
P26 Instruction Manual
Interfaces
8.1 USB port (optional)
The USB interface incorporates a USB socket with an integrated USB-to-serial converter. For
this, the driver from the manufacturer FTDI must be installed on the PC.
Please visit the manufacturer's website: www.ftdichip.com
Under DRIVERS / VCP DRIVERS, select the appropriate driver for your operating system.
Once installed, the driver will set up an additional virtual COM port.
9600 baud
8 data bits
no parity
one stop bit
23
P26 Instruction Manual
24
P26 Instruction Manual
25
P26 Instruction Manual
26
P26 Instruction Manual
Technical data
Measurement data
Measurement range 10 / 50 / 100 / 250 / 500 Pa
(also ± measurement 1 / 2.5 / 5 / 10 / 20 / 50 / 100 kPa
ranges) (freely scalable from 10 to 100% within a measurement
range)
Margin of error ± 0.2 % FS (for measurement ranges ≤ 50 kPa) or
(at 20°C; ±2°C) ± 0.5 % FS
(0.3 Pa margin of error for the reference)
Resolution dependent upon measurement range (max. 5 significant
digits)
Ambient conditions
Medium Air, non-aggressive gases
Operating temperature +10 °C to +50 °C
Storage temperature -10 °C to +70 °C
Protection class IP65/ USB-version IP40
Certifications CE, CSA
Electrical data
Power consumption approx. 6 VA
Supply voltage see rating plate
24VAC/DC +/-10%
24VAC(with galvanic separation) +6%/-15%
230/115VAC +6%/-15%
Output signal 0 … +10 V (RL ≥ 2 k Ω) or
-5 … +5 V (RL ≥ 2 k Ω) or
0 … 20 mA (RL ≤ 500 Ω) or
4 … 20 mA (RL ≤ 500 Ω)
Relay
2 x changeover contacts 6 A / 230 VAC each
Interfaces (optional)
RS-232 9600 baud, 8 data bits, no parity, one stop bit
USB 12 Mbps USB 2.0 Full-Speed
27
P26 Instruction Manual
Troubleshooting
28
P26 Instruction Manual
Dimension drawing
29
P26 Instruction Manual
Menu tree
30
P26 Instruction Manual
31
P26 Instruction Manual
For additional specifications and dimension drawings, please visit our website at
http://www.halstrup-walcher.de/en/produkte/druckmesstechnik/index.php
32
P26 Instruction Manual
Certificate of Conformity
33