Module 2
Module 2
Heart and cardiovascular system (brief discussion), electro conduction system of the heart.
Electrocardiography, ECG machine block diagram, ECG lead configurations, ECG recording
system, Einthoven triangle, analysis of ECG signals. Measurement of blood pressure: Direct,
indirect and relative methods of blood pressure measurement, auscultatory method,
oscillometric and ultrasonic non- invasive pressure measurements. Measurement of blood
flow: Electromagnetic blood flow meters and ultrasonic blood flow meters.
Cardio vascular system can be viewed as closed hydraulic system with 4 chamber pump.
It is mainly used for transportation of oxygen, Carbon dioxide, numerous chemical
compounds and the blood cells.
In some part of the system, diameter of the arteries are changed to control pressure.Heart is an
isolated two stage synchronized chamber
1. The first stage is to collect blood from the system and pump it in to 2nd stage.
2. The second stage then pump these blood to the system
Structure of Heart
One of the two stage pump (Right side) collect fluid from the system and pump it
through oxygenation system (Lungs).
Other side pump receives blood from oxygenation system (Lungs) and pump blood to
main hydraulic system.
Blood act as communication and supply network for all parts of the body
The blood is carried out to the various parts of the body through blood vessels. There are
three types of blood vessels
1) Arteries--- Thick, Carries oxygenated blood
2) Veins--- Thin, De-oxygenated blood
3) Capillaries---Smallest, Last level of blood vessels, 800000 km of capillaries
Heart pumps blood through the pulmonary circulation to the lungs and through the
systemic circulation to the other parts of the body.
1) Pulmonary circulation
2) Systemic circulation
In pulmonary circulation, venous blood (de-oxygenated) flows from right ventricle
through pulmonary artery to lungs.
The arterial (oxygenated) blood flows to left atrium through pulmonary veins.
In systemic circulation blood flows from left auricle to left ventricle and it is pumped
to aorta and its branches
The heart is able to create it's own electrical impulses and control the route the
impulses take via a specialised conduction pathway.
The conducting system of the heart consists of cardiac muscle cells and
conducting fibers that are specialized for initiating impulses and conducting them rapidly
through the heart.
They initiate the normal cardiac cycle and coordinate the contractions of cardiac
chambers. Both atria contract together, as do the ventricles, but atrial contraction occurs first.
The conducting system provides the heart its automatic rhythmic beat. For the
heart to pump efficiently and the systemic and pulmonary circulations to operate in
synchrony, the events in the cardiac cycle must be coordinated.
III. Electrocardiography
⚫ Bio electric potentials generated by heart muscles are called Electro Cardio Gram.
⚫ It is sometimes called EKG(Electro Kardio Gram)
⚫ Electrocardiography (ECG) is an interpretation of the electrical activity of the heart over a
period of time.
⚫ The recording produced by this noninvasive procedure is termed as electrocardiogram (also
ECG or EKG).
To record ECG 12 electrodes connected to the body of the patient. Electrodes connected to
ECG machine using wires called leads. Leads are electrodes which measure the difference in
electrical potential between either:
1. Two different points on the body (bipolar leads)
2. One point on the body and a virtual reference point with zero electrical potential, located in
the center of the heart (unipolar leads).
Classification
The Standard Limb Leads are used to display a graph of the potential difference recorded
between two limbs at a time. In these leads, one limb carries a positive electrode and the
other limb, a negative one.
The three limb electrodes, I, II and III form a triangle (Einthoven’s Equilateral Triangle), at
the right arm (RA), left arm (LA) and left leg (LL).
Lead I has a positive electrode on the left arm and a negative electrode on the right arm.
Lead I is a bipolar, indirect lead.
Lead II has a positive electrode on the left leg and a negative electrode on the Right arm.
Lead II is a bipolar, indirect lead.
Lead III has a positive pole on the left foot and a negative pole on the left hand. Lead III is a
bipolar, indirect lead.
The three limb electrodes, I, II and III form a triangle (Einthoven’s Equilateral Triangle), at
the right arm (RA), left arm (LA) and left leg (LL).
2. Augmented Limb Leads (Unipolar): aVR, aVL & aVF
3. Precordial Leads: V1- V6
The recording of electrical activity associated with the functioning of the heart is known as
electrocardiogram.
It‟s a quasi periodical, rhythmically repeating signal synchronized by the function of the
heart, which act as a generator of bioelectric events.
This generated signal can be described by means of an electric dipole. i.e. pole consisting of
positive and negative pair of charge.
The dipole generates a field vector changing periodically in time and space and its effects are
measured on the surface.
The waveform thus recorded standardized in terms of amplitude and phase relationship and
any deviation from this reflect the presence of abnormality.
1. PR Interval
Represents the time taken by the heart impulse to travel first through the intro-ventricular
system and then through the free walls of the ventricles. ventricular contraction
Measured from the onset of Q wave to end of S wave.
Duration is between 0.05 and 0.10 secs.
Amplitude is 1 mv
Since the ventricles contain greater muscle mass than the atria, the QRS complex is larger
than the P wave.
⚫ The systolic pressure is the maximum pressure in an artery at the moment when the heart is
beating and pumping blood through the body.
⚫ The diastolic pressure is the lowest pressure in an artery in the moments between beats when
the heart is resting.
⚫ Both the systolic and diastolic pressure measurements are important
⚫ If either one is raised, it means you have high blood pressure (hypertension).
1. Indirect
2. Direct
Indirect Method
1. AUSCULTATORY METHOD
⚫
2. Oscillometric Measurement Method
3. Ultrasonic Doppler Shift Method
Direct Method
There are used when the highest degree of absolute accuracy, dynamic response and Continuous monitoring are
required.
For direct measurement, a catheter or a needle type probe is inserted thorough a veine or artery to the area of
interest.
3. Catheter tip probe in which sensor is mounted on the tip of the probe and the pressure exerted on it
converted to the proportional electrical signals.
4. Fluid filled catheter type, which transmits the pressure exerted on its fluid filled column to an external
transducer. This transducer converts the exerted pressure to corresponding electrical signals.
• Blood flow is the one of the important physiological parameter and the most difficult to
measure accurately.
• The average velocities of blood flow vary over a wide range depending on diameter of blood
vessel.
• The Blood flow at any point in the circulatory system is the volume of blood that passes a
point during a unit of time.
• Its unit is L/min
• Blood flow is highest in the pulmonary artery and Aorta, where these blood vessels leave the
heart.
• The flow at these points, called cardiac output is between 3.5 and 5 L/min for a normal adult
at rest.
• There are many techniques for measuring the blood flow and velocity.
• They are categorized into
1. Invasive (surgical).
Structure
• A permanent magnet or electromagnet positioned around the blood vessel generates a
magnetic field perpendicular to the direction of the flow of the blood.
• The blood stream which is a conductor cuts the magnetic field and voltage is induced in the
blood stream.
• Voltage induced in the moving blood column is measured with stationary electrodes located
on opposite sides of the blood vessel and perpendicular to the direction of the magnetic field.
• This method requires that the blood vessel be exposed so that the flow head or the measuring
probe can be put across it.
• The magnitude of the voltage picked up is directly proportional to the strength of the
magnetic field, diameter of the blood vessel and the velocity of blood flow.
Types of Electromagnetic Flow Meters
Basically, all modern flowmeters consist of a generator of AC, a probe assembly, a series of
capacitance coupled amplifiers, a demodulator, a DC amplifier and a suitable recording
device.
Basing shape of the energizing current waveform for the electromagnet 2 types of EM
Flowmeters are :
Probe magnet is energized with a sine wave and the induced voltage will also be sinusoidal.
Since the flow of blood acts as a secondary terminal of a transformer w.r.t probe magnet, an
additional artifact voltage induced is called transformer voltage.
This voltage is 90° out of phase with the original signal corresponding to flow of blood .
A method for eliminating transformer voltage by using a gated amplifier (amplify signal
onlyduring flow induced voltage is maximum).
Electromagnetic blood flow meters are based on the principle of electromagnetic induction.
According to Fleming's right hand rule, the EMF induced in the conductor under the influence
of magnetic field is directly proportional to the velocity of motion of conductor.
Here we consider the blood vessel carrying blood as the conductor.
So here the velocity of motion is simply the blood flow through the blood vessel.
PRINCIPLE
BLOCK DIAGRAM
uses an oscillator of low frequency (up to 400 Hz) to drive the electromagnet which is
placed in such a way that the magnetic field is perpendicular to the direction of blood flow.
set of electrodes are placed across the blood vessel mutually
perpendicular to both magnetic field and direction of blood flow.
The EMF induced across the blood vessel will be proportional to the velocity of blood, Lumen
probes with varying diameters are used for the accuracy of measurement.
the amplifier.
1. Doppler-shift type.
2. Transit time velocity meter
ULTRASONIC BLOOD FLOW METER - DOPLER TYPE
• It is a non-invasivc technique to measure blood velocity in a particular vessel from the
surface of the body.
• lt is based on the analysis of echo signals from the erythrocytes in the vascular structures.
Because of the Doppler effect, the frequency of these echo signals changes relative to the frequency
which the probe transmits.
• The Doppler frequency shift is a measure of the size and direction of the flow velocity.
• The principle is illustrated in Fig.
The incident ultrasound is scattered by the blood cells and the scattered wave is received by
the second transducer. The frequency shift due to the moving scatterers is proportional to the
velocity of the scatterers. Alteration in frequency occurs first as the ultrasound arrives at the
„scattered and second as it leaves the scattcrer.
The transit-time ultrasonic flow meter is an instrument designed for measuring the volume flow
rate in clean liquids or gases. It consists of a pair of ultrasonic transducers mounted along an axis
aligned at an angle _ with respect to the fluid-flow axis, as shown in Figure.
Each transducer consists of a transmitter–receiver pair, with the transmitter emitting ultrasonic
energy which travels across to the receiver on the opposite side of the pipe. These ultrasonic
elements are normally piezoelectric oscillators of the same type as used in Doppler shift flow
meters. Fluid flowing in the pipe causes a time difference between the transit times of the beams
travelling upstream and downstream, and measurement of this difference allows the flow velocity
to be calculated. The typical magnitude of this time difference is 100 ns in a total transit time of
100 μs, and high-precision electronics are therefore needed to measure it.