SEDECAL X Service Manual
SEDECAL X Service Manual
Technical Publication
SM-0524R4
Service Manual
SEDECAL X
Universal Radiographic System
Este producto ostenta una marca CE de acuerdo con las disposiciones de la Directiva 93/42/CEE del 14 de Junio de 1993 sobre Productos Médicos.
This product bears a CE marking in accordance with the provisions of the 93/42/EEC MDD dated June 14, 1993.
Universal Radiographic System
Service Manual
REVISION HISTORY
This Document is the English original version, edited and supplied by the manufacturer.
The Revision state of this Document is indicated in the code number shown at the bottom of this page.
ADVISORY SYMBOLS
The following advisory symbols will be used throughout this manual. Their
application and meaning are described below.
SM-0524R4
Universal Radiographic System
Service Manual
SAFETY SYMBOLS
Ionizing radiation.
Type B equipment.
Dangerous voltage.
Ground.
SM-0524R4
Universal Radiographic System
Service Manual
TABLE OF CONTENTS
Section Page
1 INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.1 Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
2 UNPACKING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
3 INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
4 ADJUSTMENTS ........................................................ 15
5 MAINTENANCE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
SM-0524R4 i
Universal Radiographic System
Service Manual
Section Page
6 RENEWAL PARTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
7 INTERCONNECTION MAPS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
ii SM-0524R4
Universal Radiographic System
Service Manual
SECTION 1 INTRODUCTION
1.1 TOOLS
• Electricity installation:
SM-0524R4 1
Universal Radiographic System
Service Manual
Illustration 1-1
Dimensions
220
2720
1690
2240
SID: 1000--2000
650
700
1400
1095 1165
228
Reception Area
1181
Reference Axis
1000
2000
2260
2 SM-0524R4
Universal Radiographic System
Service Manual
SECTION 2 UNPACKING
1. Place the shipping pallet near its final site in the room and remove all its
laterals. Do not discard any packing material such as envelopes, boxes,
bags until all parts are accounted for as listed on the packing list.
Illustration 2-1
Unpacking
SM-0524R4 3
Universal Radiographic System
Service Manual
4 SM-0524R4
Universal Radiographic System
Service Manual
SECTION 3 INSTALLATION
2. Install the Control Panel by using the two bolts at its rear side.
4. Place the Column standing-up and position it against the wall on its final
site in the room. While one person is holding up the Column, the other
one should level it vertically on both lateral sides and on the front,
marking its anchoring positions on the floor and on the wall.
Illustration 3-1
Installation of Tube-Collimator Assembly and Upper Wall Support
SM-0524R4 5
Universal Radiographic System
Service Manual
Illustration 3-2
Drill Template
Dimensions in mm
6 SM-0524R4
Universal Radiographic System
Service Manual
6. Position the Column and make it firm to floor and wall. Check that it is
properly levelled on both lateral sides and on the front. Make sure the
Column is firmly anchored. (Refer to Illustration 3-3.)
Illustration 3-3
Column anchoring
CENTRAL CARRIAGE
LOWER ANCHORAGES
SM-0524R4 7
Universal Radiographic System
Service Manual
7. Install the X-ray Tube in the Upper Support of the Column using the
Collimator Adaptation Ring and its four Safety Screws (Allen).
9. Remove the X-ray Tube and Collimator covers and connect the wires
from the upper harness to the X-ray Tube and Collimator accordingly to
their identification (refer to Interconnections Map IM--041 for Manual
Version or A6541-03-S of Motorized Version).
Illustration 3-4
Installation of X-ray Tube and Collimator
COLLIMATOR COVER
CENTERING AND SAFETY SCREWS
COLLIMATOR CABLE
8 SM-0524R4
Universal Radiographic System
Service Manual
10. Remove the Bucky Lock-Lever. Install the Bucky Assembly inserting it
into the lower axis of the Swivel Arm.
11. Re-install the Bucky Lock-Lever keeping the indicator plate between
both nylon washers, and the metallic washers close to the lever.
Illustration 3-5
Bucky Lock-Lever
INDICATOR PLATE
NYLON WASHERS
METALIC WASHER
BUCKY LOCK-LEVER
12. Open the Bucky cover removing its two frontal screws. Release the
plastic lock nut from the Bucky harness of the Swivel Arm (small
harness). Route cables through their entrance into Bucky Assembly and
put back the lock nut into its hole to fasten the harness to the Assembly.
Connect cables inside the Bucky Assembly (Refer to Illustration 3-6).
Illustration 3-6
Bucky Connections
BUCKY LOCK-LEVER
SM-0524R4 9
Universal Radiographic System
Service Manual
13. Install the Control Unit on the wall near the Column (refer to
Illustration 3-2).
a. Install the lower cover of the Control Unit Box. Identify and
separate the cables to be connected to the Control Unit and the
cables to be connected to the Generator Cabinet.
Illustration 3-7
Control Unit Box Connections for Manual Version
TRANSFORMER
LINE FUSES
LOWER COVER
d. Connect directly the ground cables (GND) from the Column and
Control Unit Box to the Main ground stud (GND) of the Generator.
10 SM-0524R4
Universal Radiographic System
Service Manual
f. Verify that the transformer and the fuses of the Control Unit Box
are in accordance to the power line. (Refer to Maps I/M-041 and
A6408-xx).
a. Install the lower cover of the Control Unit Box. Identify and
separate the cables to be connected in the Control Unit and the
cables to be connected to the Generator Cabinet.
Illustration 3-8
Control Unit Connections for Motorized Version
TRANSFORMER
LINE FUSES
SM-0524R4 11
Universal Radiographic System
Service Manual
e. Connect directly the ground cables (GND) from the Column and
Control Unit Box to the Main ground stud (GND) of the Generator.
g. Verify that the transformer and the fuses of the Control Unit are in
accordance to the power line (refer to Interconnections Map
A6541-03-S).
Note . Do not plug the Control Unit until specifically advice in this
Document.
12 SM-0524R4
Universal Radiographic System
Service Manual
Prepare the High Voltage terminals that will be installed in the X-ray Tube
receptacles. Apply Silicone Paste over the entire surface of the Plug
including the Pins.
Carefully connect cables to their related receptors of the Tube and fix
their nuts tightly.
18. Fasten HV cables to lateral sides of Swivel Arm with the clamp on the
carriage of the Tube Collimator Assembly. Give the cable length enough
to enable movements of the of Swivel Arm (refer to Illustration 3-9).
19. Install the Top Cover of the Columm with the supplied screws (refer to
Illustration 3-9).
20. Install the Decorative Base of the Columm (pressing it) (refer to
Illustration 3-9).
21. Remove now the Safety Locking Rod from the Central Carriage (refer to
Illustration 3-9).
SM-0524R4 13
Universal Radiographic System
Service Manual
Illustration 3-9
Covers and HV Cables Clamp
TOP COVER
HV CABLES CLAMP
BASE COVER
22. Plug the Control Unit and turn the system on. Check that all controls and
movements operate correctly.
14 SM-0524R4
Universal Radiographic System
Service Manual
SECTION 4 ADJUSTMENTS
The following special Tools are commonly used for adjustment of the Rad
Room:
• Dynamometer.
Note . These tools are not included with the System. Use the
suggested tools or equivalent.
Note . This section only applies to systems where the X-ray Tube or the
Collimator has to be replaced by a new different X-ray Tube or
Collimator (different weight).
SM-0524R4 15
Universal Radiographic System
Service Manual
1. Turn the unit ON. Place the Swivel Arm in horizontal position and open
SID at maximum meters.
2. Press the “Rotation” button and check that the Swivel Arm is balanced,
that is, the Swivel Arm should not tilt in any direction without force applied
at any point.
If the Swivel Arm is correctly balanced, check the next section. If not,
perform the following steps:
Illustration 4-1
Counterweights Housing
16 SM-0524R4
Universal Radiographic System
Service Manual
1. With the unit turned ON. Place the Swivel Arm in horizontal position and
open SID at 1.5 meters.
4. Disengage the Spring located on the Top of the Column (only one end).
Illustration 4-2
Adjustment of the Central Carriage Balance
CENTRAL CARRIAGE
ADJUSTMENT SCREW
SM-0524R4 17
Universal Radiographic System
Service Manual
6. Carefully, remove the Safety Locking Rod from the Central Carriage.
Illustration 4-3
Adjustment of the Central Carriage Balance
TIGHTENER SCREW
18 SM-0524R4
Universal Radiographic System
Service Manual
1. With the unit turned ON. Place the Swivel Arm in horizontal position and
open SID at 1.5 meters.
4. Loosen slightly the three Safety Screws of the Motor. Then, loosen
completely the Stop Screw of the Motor.
Illustration 4-4
Motor and Chain
CENTRAL CARRIAGE
CHAIN
CHAIN
SM-0524R4 19
Universal Radiographic System
Service Manual
6. Carefully, remove the Safety Locking Rod from the Central Carriage.
Illustration 4-5
Adjustment of the Central Carriage Balance
TIGHTENER SCREW
20 SM-0524R4
Universal Radiographic System
Service Manual
The motorized movements of the Swivel Arm (SID and Vertical movements) are
blocked automatically for safety reasons when an obstacle is found in its travel
applying a strength above 15 ¦ 3 kg.
1. With the unit turned ON. Place the Swivel Arm centered in horizontal
position and open SID at 1.5 meters.
DYNAMOMETER (Pos. 2)
DYNAMOMETER (Pos. 1)
2. Press the respective button to start the vertical down movement of the
Swivel Arm and check with a Dynamometer located in contact with the
lower side of the Swivel Arm (centered) that the vertical down movement
is blocked when a strength of 15 ¦ 3 kg (Peak) is measured on the
Dynamometer.
4. If it is not correct, adjust the Potentiometer R66 of the Control Unit Board
(A3194--XX). Turn the potentiometer screw CW to increase the strength
and turn it CCW to reduce the strength.
SM-0524R4 21
Universal Radiographic System
Service Manual
1. With the unit turned ON. Place the Swivel Arm centered in horizontal
position and open SID at 1.5 meters.
DYNAMOMETER (Pos. 2)
DYNAMOMETER (Pos. 1)
2. Press the respective button to reduce the SID and check with a
Dynamometer located in contact with the inner side of the Receptor
(Table--Top) that the horizontal movement closing the SID is blocked
when a strength of 15 ¦ 3 kg (Peak) is measured on the Dynamometer.
3. Press the respective button to increase the SID and check with a
Dynamometer located in contact with the outer side of the Receptor
(back side) that the horizontal movement opening the SID is blocked
when a strength of 15 ¦ 3 kg (Peak) is measured on the Dynamometer.
4. If it is not correct, adjust the Potentiometer R67 of the Control Unit Board
(A3194--XX). Turn the potentiometer screw CW to increase the strength
and turn it CCW to reduce the strength.
22 SM-0524R4
Universal Radiographic System
Service Manual
Illustration 4-6
Vertical Position of Swivel Arm with Alignment Tools
SM-0524R4 23
Universal Radiographic System
Service Manual
3. Turn on Collimator light and center the Collimator in relation to the Bucky
Table-Top. Horizontal and transversal position of the light axes projected
by the Collimator Lamp must be in line with the axes or film sizes marked
on the Table-Top.
Illustration 4-7
Collimator Screws
4. Position the Collimator Test Tool (RMI model 161B) on the Bucky
Table-Top.
24 SM-0524R4
Universal Radiographic System
Service Manual
6. Adjust the Light Field of the Collimator Lamp to the rectangle drawn
inside the Collimator Test Tool.
7. Place centered the Beam Alignment Test Tool (RMI model 162A) on the
Collimator Test Tool.
Illustration 4-8
Alignment Test Tools
8. Load Cassette film Tray with cassette film 24x30 and insert it.
a. Check that the X-ray Field falls just within the image of the inner
rectangle of the Collimator Test Tool.
If an edge of the X-ray Field falls out of the inner rectangle means
a misalignment of the Light Field respect to the X-ray Field. The
maximum misalignment allowed is 2 % of SID (for SID 1m = 2 cm
tolerance).
Refer to Section 4.4.1 for alignment of Light Field with X-ray Field.
SM-0524R4 25
Universal Radiographic System
Service Manual
-- If the image of the two balls overlap (A) the X-ray Beam is
perpendicular to within 0.5o.
-- If the image of the top ball (larger shadow) intercepts the first
circle (B), the X-ray Beam is about 1.5o away from the
perpendicular.
-- If the image of the top ball (larger shadow) intercepts the second
circle (C), the X-ray Beam is about 3o away from the
perpendicular.
In cases (A) and (B) perpendicularity is within tolerance for SID at
1 meter (top ball is within or intercepting the first circle). The third
case (C) needs readjustment.
A B C
c. Check that the X-ray Beam is properly centered with the Image
Receptor. To determine the center of the Image Receptor, draw
diagonal lines from corner to corner of the X-ray Film. (Alternately,
the film can be folded in half and creased at the center). The two
lines will cross in the center of the Image Receptor (film). Then
draw diagonal lines from the corners of the imaged X-ray Field.
If the center of the X-ray Field and Image Receptor is the same,
the diagonals of both sets of lines should cross at the same point.
The maximum misalignment allowed is 2 % of SID (for SID 1m =
2 cm tolerance).
26 SM-0524R4
Universal Radiographic System
Service Manual
3. Position the Collimator Test Tool (RMI model 161B) on the Bucky
Table-Top.
4. Turn on Collimator light and center the Collimator Test Tool with the light
axes projected by the Collimator Lamp.
5. Adjust the Light Field of the Collimator Lamp to the rectangle drawn
inside the Collimator Test Tool.
SM-0524R4 27
Universal Radiographic System
Service Manual
6. Check on the processed film the adjustment required to meet the Light
Field with the X-ray Field. Identify the deviation on the axes imaged on
the film.
PROCESSED FILM
PROCESSED FILM
28 SM-0524R4
Universal Radiographic System
Service Manual
7. Do not remove the Collimator Test Tool from its original position and
adjust the Light Field by moving the Collimator Unit and/or the Collimator
Lamp.
Illustration 4-9
Collimator Screws
SM-0524R4 29
Universal Radiographic System
Service Manual
In case that perpendicularity is out of tolerance (top ball is out of first circle),
adjust perpendicularity as follows:
3. Position the Collimator Test Tool (RMI model 161B) on the Bucky
Table-Top.
4. Turn on Collimator light and center the Collimator Test Tool with the light
axes projected by the Collimator Lamp.
5. Place centered the Beam Alignment Test Tool (RMI model 162A) on the
Collimator Test Tool and observe if shadow of the Beam Alignment Test
Tool is projected in equal proportion around it.
6. Check on the processed film the adjustment required to center the top
ball mark. Shadow around the Beam Alignment Test Tool can also help
to make a first correction.
7. Loosen slightly the four Safety Screws (Allen M8) of the Tube-Collimator
Assembly. If required, perform the following adjustments:
8. Repeat exposure and procedure until the result is satisfactory (top ball
must be inside of the first circle).
30 SM-0524R4
Universal Radiographic System
Service Manual
Illustration 4-10
Levelling Rods of the Tube-Collimator Assembly Support
SAFETY SCREWS
LOCK TOOTH
SETSCREWS
SAFETY SCREWS
LEVELLING SCREWS
LOCK-LEVER FOR ROTATION OF
TUBE-COLLIMATOR ASSEMBLY
SAFETY SCREWS
SM-0524R4 31
Universal Radiographic System
Service Manual
The error between centers of the X-ray Field and the Image Receptor should
not be greater than 2% of the SID. (for SID 1m = 2 cm tolerance).
DEVIATION
1. Check on the processed film the correction required for centering the
X-ray Field with the Image Receptor. Adjustments will be performed as
the same way described for Perpendicularity correction, so only re-adjust
it if really is necessary.
32 SM-0524R4
Universal Radiographic System
Service Manual
Note . Before starting with the SID Indicator Test, the Alignment of X-Ray
Beam Test and the Alignment of Light Field with X-Ray Field Test
should be performed.
Illustration 4-11
SID Test Stand Tool
3. Place the SID Test Stand Tool on the Bucky Table-Top. (Refer to
Illustration 4-11).
5. Adjust the Light Field to cover the window (circle) on the top SID Stand
Test Tool. (Refer to Illustration 4-12).
SM-0524R4 33
Universal Radiographic System
Service Manual
6. Place the Small Tool included in the SID Stand Test Tool with two straight
narrow wires crossed on the center of circle at top SID Stand Test Tool.
Place a small signal on the circle in order to mark the position.
Illustration 4-12
Light Field over SID Test Stand Tool
a. Measure the height h, it is the height of the SID Test Stand Tool
plus the distance from the Table-Top to the Film.
b. Measure the radius of window (d/2) on the SID Test Stand Tool.
Illustration 4-13
SID Calculation
FOCAL SPOT
X-RAY TUBE
COLLIMATOR
(x) TO CALCULATE
SID ?
WINDOW CIRCLE
d
SID TEST STAND TOOL
h
FILM
34 SM-0524R4
Universal Radiographic System
Service Manual
Illustration 4-14
Radius in Processed Film
SIGNAL
D/2
PROCESSED FILM
10. With the measurements obtained on the step-9., calculate the “X” value,
where:
(d/2) x h
SID X=
d/2 (D/2) - (d/2)
h
D/2
12. The difference between the indicated SID and the calculated SID may
not exceed 1.8 % (rejection limit) of indicated SID. Therefore the
deviation of SID calculated should be not higher than 18 mm for SID at
1 meter.
SM-0524R4 35
Universal Radiographic System
Service Manual
13. For the Manual Version, if the SID value calculated is out acceptance
limits, it will be necessary to readjust the SID Scale position of the Manual
Universal Radiographic System.
14. For the Motorized Version, if the SID value calculated is out
acceptance limits, it will be necessary to readjust the SID Cam for the
Motorized Universal Radiographic System. For that, loose the two
screws that fix the SID Cam and move it to the required position.
Illustration 4-15
SID Cam Adjustment
SID CAM
Note . Before starting with the Field Size Indicator Test, the Alignment of
X-Ray Beam Test, the Alignment of Light Field with X-Ray Field
Test and the SID Indicator Test should be performed.
4. Turn on the Collimator Light and center the Collimator in relation to the
Bucky Table-Top. Horizontal and transversal position of the light axes
projected by the Collimator Lamp must be in line with the axes or Film
Size marked on the Table-Top.
36 SM-0524R4
Universal Radiographic System
Service Manual
Illustration 4-16
X-ray Field Size
FILM 35x43 cm
PROCESSED FILM
9. The difference between the indicated Field Size and the obtained Field
Size may not exceed 1.5 % (rejection limit) of the SID in either direction.
Therefore the deviation in any direction should not be higher than 15 mm
(rejection limit) for SID at 1 meter.
SM-0524R4 37
Universal Radiographic System
Service Manual
Note . Before starting with the Collimator Lamp Brightness Test, the SID
Indicator Test should be performed.
Illustration 4-17
Vertical Position
4. Turn on the Collimator Light and center the Collimator in relation to the
Bucky Table-Top. Horizontal and transversal position of the light axes
projected by the Collimator Lamp must be in line with the axes or Film
Size marked on the Table-Top.
38 SM-0524R4
Universal Radiographic System
Service Manual
5. Place the Light Meter on the Table-Top with sensor directed toward the
Light Source.
6. Place the Light Meter in the center of one quadrant of the light field (refer
to Illustration 4-18).
Illustration 4-18
Light Meter on Table-Top
4th QUADRANT
3rd QUADRANT
LIGHT FIELD
LIGHT METER
7. With Collimator Lamp OFF, measure and keep record of the ambient light
level.
8. Do not move the Light Meter. With the Collimator Lamp ON, measure and
keep record of light level of that quadrant.
11. With the measurements obtained on the step-10., the results of Light
Field Intensity in all quadrants should be higher of 170 lux
(16 foot-candles) and among all the quadrants the intensity light field
shall be not differ more than 10%.
12. The average illumination at a distance of 1 meter from the focal spot shall
not be less than 160 lux (15 foot-candles). Rejection limit is 170 lux
(16 foot-candles).
SM-0524R4 39
Universal Radiographic System
Service Manual
b. If the light level in all quadrants is less than 170 lux, it is necessary
to check the following:
G Check that the Collimator Lamp, the Mirror and the Mylar
window are not dirty or discolored.
3. Open the Collimator blades to set a Field Size of 25 x 25 cm for SID 1m.
4. Turn ON the Collimator Light and center the Collimator in relation to the
Bucky Table-Top. Horizontal and transversal position of the light axes
projected by the Collimator Lamp must be in line with the axes or Film
Size marked on the Table-Top.
5. Place the Light Meter on the Table-Top with sensor directed toward Light
Source.
6. Place the Light Meter in the center of one quadrant of the light field.
(Refer to Illustration 4-19).
7. Minimize the room lighting. With the Collimator Lamp OFF, measure and
record the ambient light level.
40 SM-0524R4
Universal Radiographic System
Service Manual
Illustration 4-19
Light Contrast Calculation
EDGE OF A QUADRANT
LIGHT SENSOR ON THE LEFT EDGE LIGHT SENSOR ON THE RIGHT EDGE
BRH POINT
(LIGHT DROPS 25%)
9. Measure the illumination at a point 3 mm from the edge of the field toward
the center of the field (Light sensor on the left of edge). (Refer to
Illustration 4-19). Record this as I1.
10. Measure the illumination at a point 3 mm from the edge of the field away
from the center of the field (Light sensor on the right of edge). (Refer to
Illustration 4-19). Record this as I2.
11. Correct the values of I1 and I2 by subtracting from each value the ambient
light level measured in step-7. Now divide the corrected value of I1 by
I2. This ratio should be 4 or more.
12. Repeat the process from step-8. for all quadrants of Light Field.
13. If the deviation of Light Field Contrast Ratio is out acceptance limits, it
is necessary check the following:
G Check that the Collimator Lamp, the Mirror and the Mylar window
are not dirty or discolored.
G The ambient light level shall be low as it affects the accuracy of the
measurements.
SM-0524R4 41
Universal Radiographic System
Service Manual
42 SM-0524R4
Universal Radiographic System
Service Manual
SECTION 5 MAINTENANCE
GENERAL CLEANING
EXTERNAL SURFACES Clean the equipment frequently, particularly if corroding chemicals are present. Clean
external covers and surfaces, especially parts in contact with patients, with a cloth
moistened in warm water with mild soap. Wipe with a cloth moistened in clean water.
Do not use cleaners or solvents of any kind. Also check painted surfaces for scratching
and touch up as required.
INTERNAL CLEANING DANGER: NEVER ATTEMPT TO CLEAN INTERNALLY THE UNIT. THE UNIT
MUST BE ALWAYS KEPT SHUT UNLESS A SPECIFIC RENEWAL PART
PROCEDURE IS TO BE UNDERTAKEN BY HIGHLY QUALIFIED MAINTENANCE
PERSONNEL.
SM-0524R4 43
Universal Radiographic System
Service Manual
ELECTRICAL CHECKS
ELECTRICAL CABLES Check that all electrical connections are firm and secure and that all cable clamps and
AND CONNECTIONS strain reliefs are in place. Also check that connectors do not have exposed wires and
check cable sheaths for wear and fraying.
MECHANICAL CHECKS
MOVEMENT COMPONENTS Check correct operation of the controls, detents, limits and movements in all possible
positions.
STEEL CABLE Verify that the steel cable is in good condition and appearance. Replace the Cable if any
steel wire is damaged.
RAILS Clean the Column Rails with a cloth moistened in clean water and lubricate both rails
with moisten cloth with lubricant oil.
In case of noise during movement, clean and lubricate again.
44 SM-0524R4
Universal Radiographic System
Service Manual
SM-0524R4 45
Universal Radiographic System
Service Manual
A0 COLUMN
A6 A3
A1
A4
A5
A2
46 SM-0524R4
Universal Radiographic System
Service Manual
A1
A2 A3
A4 A6
SM-0524R4 47
Universal Radiographic System
Service Manual
48 SM-0524R4
Universal Radiographic System
Service Manual
B8
B20
B27
B6 B21
B2
B7
B23 B14
B4 B3
B2 B13
B5 B26
B1 B24 B25
B11
B17
B18
B10
B22
B12
B33
B15
B19
B16
B32
B9
SM-0524R4 49
Universal Radiographic System
Service Manual
B1 B2 B3 B4
B5 B6 B7 B8
50 SM-0524R4
Universal Radiographic System
Service Manual
SM-0524R4 51
Universal Radiographic System
Service Manual
C0 POWER BOX
C1 Transformer 1 SAT--50509008
C3
C1
C6
C4 C5
C2
C1 C2 C3
C4 C5
52 SM-0524R4
Universal Radiographic System
Service Manual
D0 CABLES
SM-0524R4 53
Universal Radiographic System
Service Manual
54 SM-0524R4
Universal Radiographic System
Service Manual
SM-0524R4 55
Universal Radiographic System
Service Manual
A0 COLUMN
SAT--A8280--01 -- Bucky
A2 Gear 21 Z 1 SAT--12002--01
-- Bucky
SAT--12040--02 -- Canon 50G
A5 Steel Cable Kit 1 -- Trixell
SAT--12040--08 -- Radax
A6 A2
A4
A5
A3
A8
A1
A7
56 SM-0524R4
Universal Radiographic System
Service Manual
A1
A2 A3 A4
A5 A6 A8
SM-0524R4 57
Universal Radiographic System
Service Manual
B7 Microswitch 1 SAT--50605014
58 SM-0524R4
Universal Radiographic System
Service Manual
SM-0524R4 59
Universal Radiographic System
Service Manual
B6 B10 B11
B20
B31
B24
B25
B2
B9
B27 B16
B4 B3
B2 B15
B30
B28
B5
B1 B29
B7
B13
B21 B22
B26
B12 B14
B17
B37
B18
B23
B19 B8
B36
60 SM-0524R4
Universal Radiographic System
Service Manual
B1 B2 B3 B4
B5 B6 B7 B8
B9 B10 B11
SM-0524R4 61
Universal Radiographic System
Service Manual
62 SM-0524R4
Universal Radiographic System
Service Manual
C0 POWER BOX
C1 Transformer 1 SAT--50509033
C3
C1
C6
C4 C5
C2
C1 C2 C3
C4 C5
SM-0524R4 63
Universal Radiographic System
Service Manual
D0 CABLES
64 SM-0524R4
Universal Radiographic System
Service Manual
MANUAL MOTORIZED
SCHEMATIC
VERSION VERSION
SM-0524R4 65
Universal Radiographic System
Service Manual
66 SM-0524R4
A B C D E
OPTIMA
UNIVERSAL Collimator Cable (A3346-02) COLLIMATOR
BOARD TS3
24 VAC
4 (A3127-01) 1 4
0 VAC
2
GND
COLUMN PANEL
3 Control Cable (A3157-02) 3
TS4
VERTICAL
VERTICAL CONTROL
5 1
ROTATION
COLUMN ROTATION CONTROL
6 2
SID
SID CONTROL
4 3
Vertical Lock Vertical Locks Cable (A3154-02) GND
VERTICAL LOCK 7 4
GND
TS2
SID LOCK
4
PIVOTING COLUMN
COLLIMATOR PANEL
VERTICAL
1
SID Lock SID Locks Cable (A3155-02) ROTATION
SID LOCK 2
SID
GND
3
4
Rotation Lock Rot. Locks Cable (A3156-02)
ROTATION LOCK
GND
1 1
8 440 V 2
0 VAC
7 415 V F2
20 V 2 T 10A
400 V TP1
6 CR7 +24 V
Connect according FB1002L
4 x 2200uF
GND
STUD
with the input line
5 240 V -- + 3
+24 VDC
C4 C5 C6 C7 R1
4 230 V R4 160, 2w
3 2K7 K1
2w SID
3 8 3
3 208 V CR4
1N4007 C1
TS1 See Table 1 K1
100 uF
LF1 2 115 V 4 5 7 4
L1 F1 3 SID
SID LOCK
1 3 2 R2 CONTROL
INPUT 6
L2/N F2/N CR1 4 160, 2w
LINE K2
2 4 1N5408
VERT 8
CR5
1 K2
VERTICAL 5 1N4007 C2
5
LOCK 3 100 uF
2 7 5
6
CR2 4 VERTICAL
R3 CONTROL
1N5408
160, 2w
1 K3 K3
Emergency ROTATION 6 5 ROT
3 8
Stop LOCK CR6
Button SW1 2 1N4007 C3
6 100 uF
1 2 CR3 4
2 1N5408 7 6 2
ROTATION
4 5 1 0V 0V CONTROL
7 TP2 7
GND GND GND
8 8
GND GND
1
NAME DATE SHEET / OF
DRAWING F. GARCIA 06/06/01
DWG:
A6408--02
1/1
REVISED A. DIAZ 06/06/01 B A REV
GENERATOR CABINET
4 4
Stator Cable
TS2
Anode Cable
CONTROL Optima URS Motorized Display Cable (A3387-01)
PANEL J1
Cathode Cable
X-RAY
TUBE
HV TANK
Fan Cable (A7306-XX)
(optional)
COLLIMATOR
SID
CARRIAGE
3 3
SID Cable (A3391-01)
J2
ROTATION
Push-Button Cable (A3393-01)
ARM J2
COLUMN ROT LOCK Cable (A3389-01)
TS1
CARRIAGE SID Motor Cable (A3386-02)
TS2
2 2
GND
Stud GND Cable
Power Cable (A3153-02)
from VAC LINE TS1
COLUMN
GND GND
Stud GND Cable Stud
1 1
SW3
SID 200 cm 4
Collimator Cable
(A3388-01) TS1 J1 SW4-1 SW4-2
J2 3
VERT. POS.
Collimator 24 VAC LAMP
3 0 VAC LAMP 1 1 3
GND 2 1
3 2
4 3
5 4
ROTATION LOCK 6 5
7 6
ROT LOCK Cable 8 7 COLUMN CARRIAGE
GND (A3389-01) 9 8
10 9 Lock-up Cable
UPPER LOCK
L1
10 (A3392-01) 6 SW5
11
Rotation Lock 12 LOWER LOCK
13 7 SW6
14
GND TR1 15 UNDERTABLE LOCK
to GND Stud of 16 8 SW7
GND Cable Stud
the Generator 6 shield 9
Optima URS Motorized Board
5 240V 24V (A3194-03)
BUCKY
J3
Tube Assembly Fan Cable 6 HEIGHT UP
3
J1 (A7306-XX) TS1 1
1
2
230 VAC 2 HEIGHT DOWN
1 0 VAC 26
1 1
3 Chassis GND 10
L N 2 14
TUBE
FAN
Generator NAME
F.GARCIA
DATE
06/06/03
SHEET / OF
A6541-03-S
Cabinet
DRAWING
2/2
REVISED A.DIAZ 06/06/03 A REV
R2 Q1 Q2 R4 R12 Q5 Q6 R14
100 100 100 100
H-BHO Q1 TO Q4 S-AHO
ARE MOSFET
STP16NE06FP Q5 TO Q8
HEIGHT MOTOR ARE MOSFET SID MOTOR
TS3 R1 R3 R11 STP16NE06FP R13 TS2
10K 10K 10K 10K
4 1 1 4
2 2
3 H-BHS R18 S-AHS 3
R5 10K
R7 Q3 10K Q4 R10 R17 Q7 Q8 R20
100 100 100 100
H-BLO R8 R15 S-ALO
10K 10K
R6 R9 R16 R19
10K 10K 10K 10K
H-ALO S-BLO
H-AHS S-BHS
C2 C4
H-AHO 0.1 0.1 S-BHO
3 3
TP3
F1 F2 +30V
T 6,25A T 6,25A
CR8
FB1002L R26
1K
- + +30 VDC
R57
R25 160
2K7 4w
2w
TS1
24 VAC COLL. TP5 TP6
1 +12V +5V
0 VAC COLL.
2
24 VAC R60 R61
3 CR6 1K CR7 1K
0 VAC F3 1N4007 +12V 1N4007 +5V
4 T 6A
20 VAC CR9
2 5 2
FB1002L R56
0 VAC 100 REG1 REG2
6 +24V 4w LM7812 LM7805
ROTATION LOCK - +
7 VIN VOUT VIN VOUT TP4 TP7
GND
GND
GND CR10 C51 C52 R59 C33 C32 C35 C34 GND GND
8 1N5408 2200uF 2200uF 2K7 470uF 470uF 470uF 470uF
35V 35V 2w 35V 25V 35V 25V
9
10
K1
ROT
2 R58
14
14
14
14
14
6 160
4 2w 4
8 U5A U6B 6 U8A U9A
7 74HCT08 5 74HCT00 U7A 74HCT08 74HC86
CR11 74HCT14
1N4007
ROTATE
7
1 1
15
+12V R40 C17 UC3524A 10K U7C 10 74HCT08
+5V 4K99 100nF 74HCT14 R39
VIN
C53 R28 16 12 1K +12V
100nF 100R Vref CA C13
2 IN+ CB 13 5 6 4
R63 11 U5B 6 100nF
EA
8
3K32 14 5 74HCT08 CR1
EB R32 MUR120
4 + 5 1 IN- 4
7 9 6K81
Comp U7B
- 6 RT 6
U12B R29 4 74HCT14 1 U1 C5
LM393 2K21 CL+ C8 U5A HIP4082 1uF
5 CL- Osc 3 3
+5V R31 10nF 3 4 2 74HCT08 1 16 H-BHO
+5V Q11 909 C7 BHB BHO H-BHS
10 7 7 15
GND
2N4401 3.3nF S.D. CT AHI BHS H-BLO
3 BLI BLO 14
R62 U12A C18 + f=20KHz 12 2 13 H-ALO
3K32 LM393 R64 100uF U5D BHI ALO
11 4 ALI VDD 12
3 5K62 13 74HCT08 8 11 H-AHS
8
+ DIS AHS
1 5 10 H-AHO
DEL AH0
VSS
2 R43 R22 C37 9 C6
- AHB
R30 2K21 2K21 1nF 1uF
R65 10K 12 R50
C54 909 1 11 U9D 10K C26 12 R27
6
100nF 3 U9A 74HC86 13 10nF U6D 11 10K CR2
74HC86 2 13 74HCT00 MUR120
S-I LIMIT
+5V
C28
100nF +5V R45 C15
U10 4K99 100nF +12V +5V
40
J2 AT89C55 RP11 R67
1 10K 5K
VCC
3 SW COMM 1 21 UP 9 U3 R46 U7D 3
15
1 SID 100 SW P1.0 P2.0 DOWN 8 R44 C19 UC3524A 10K 74HCT14
2 2 P1.1 P2.1 22
SID 180 SW 3 23 SID INH 7 4K99 100nF
VIN
3 SID 200 SW P1.2 P2.2 INCREASE 6
4 4 P1.3 P2.3 24 16 Vref CA 12 9 8
VERT. POS. SW 5 25 DECREASE 5 R33 2 13
5 UPPER LOCK SW INPUTS TO U10 AS P1.4 P2.4 HEIGHT INH 4 100R IN+ CB
6 6 P1.5 P2.5 26 EA 11
LOWER LOCK SW CIRCUITRY ABOVE 7 27 ROTATE 3 14
7 UNDERTABLE LOCK SW P1.6 P2.6 SPARE 2 EB R37
8 (SEE SHEET 3) 8 P1.7 P2.7 28 1 IN-
9 6K81
9 SPARE 1 3 5 7 +5V Comp
10 RT 6
HEIGHT UP RP12 10 39 4
11 HEIGHT DOWN 10K P3.0 P0.0 CL+ C10
12 11 P3.1 P0.1 38 5 CL- Osc 3
INCREASE SID 12 37 R36 10nF
13 DECREASE SID 2 4 6 8 P3.2 P0.2 1K65 C9
13 36 10 7
GND
14 ROTATE ARM P3.3 P0.3 R34 3.3nF S.D. CT
15 14 P3.4 P0.4 35 9
15 34 U6C 8 2K21 f=20KHz
16 P3.5 P0.5 74HCT00
16 P3.6 P0.6 33 10
17 32 Q12
8
P3.7 P0.7 2N4401
J3 19 XTAL1 ALE 30 +5V + C20
XT1 31 +5V 100uF
12MHz EA C27 R51 R35
1 PSEN 29
HEIGHT DOWN 18 10nF 10K 10K
GND
2 XTAL2 R53
3 CS1 RST 9
HEIGHT UP C30 10K SID PULSE
4 DECREASE SID (SEE SHEET 3)
33pF C29 R47 TP2
5 CS2
20
VSS
DECREASE SID RP2 ULN2803 RP1 12 9 C12
5 ROTATE ARM 22 10K 1 U8D AHB 1uF
6 11
HEIGHT UP LAMP 2 1 18 1 2 13 74HCT08
7 HEIGHT DOWN LAMP Q1 I1 3 R38
4 3 17 2
6
8 INCREASE SID LAMP Q2 I2 4 10K CR4
9 6 5 16 Q3 I3 3
DECREASE SID LAMP 8 7 15 4 5 MUR120
10 UNDERTABLE LOCK LAMP Q4 I4 6
11 14 Q5 I5 5
SID 100 LAMP 13 6 7 R24 C38
12 SID 180 LAMP RP3 Q6 I6 8 2K21 1nF
1 12 7 1
13 SID 200 LAMP 22 Q7 I7 9
11 8
GND
14 LAMP COMM Q8 I8
15 8 7
6 5
+12V 4 3 NAME DATE SHEET / OF
2 1 A3194-03
9
+24V
U17
RP10 TLP621-4
2 1 1 16
J2-2
SID 100 SW 2 15
RP10
4 3 3 14
J2-3
SID 180 SW 4 13 RP13
RP10 2 1
4 6 5 5 12 4
J2-4 RP13
SID 200 SW 6 11 +5V 4 3
RP10
8 7 7 10 RP6 RP13
J2-5 1 3K3 6 5
VERT POS SW 8 9 6
7 RP13
8 8 7
U16 9
RP9 TLP621-4 2 RP14
2 1 1 16 3 7 8 U18
J2-6 4 74HCT244
UPPER LOCK SW 2 15 5 RP14 2 18
5 6 1A1 1Y1
RP9 4 1A2 1Y2 16
4 3 3 14 6 14
J2-7 1A3 1Y3
RP14 8 1A4 1Y4 12
LOWER LOCK SW 4 13 3 4 11 9
2A1 2Y1
RP9 13 2A2 2Y2 7
6 5 5 12 RP14 15 5
J2-8 1 2 2A3 2Y3 U10-1
17 2A4 2Y4 3
UNDERTABLE LOCK SW 6 11
RP9 1 U10-2
8 7 CS1 1G
7 10 19 2G
J2-10 C45 C46 C48 C47 C43 C44 C41 C42 U10-3
SPARE 8 9
U10-4
U10-5
3 RP7,RP8, RP9, RP10 are 4K7 resistor arrays 3
U15 U10-6
TLP621-4 RP13, RP14, RP15, RP16 are 470 ohm. resistor arrays
RP8
2 1 1 16 Capacitors are 100 nF, 50V U10-7
J3-10
CWW IN 2 15 U10-8
RP8
4 3 3 14 U19
J3-9 RP15 74HCT244
CW IN 4 13 2 1 2 18
1A1 1Y1
RP8 4 1A2 1Y2 16
6 5 5 12 RP15 6 14
J3-11 +5V 4 3 1A3 1Y3
8 1A4 1Y4 12
H DET SW 6 11 11 9
RP5 2A1 2Y1
RP8 RP15 13 2A2 2Y2 7
8 7 7 10 1 3K3 6 5 15 5
J2-15 6 2A3 2Y3
17 2A4 2Y4 3
8 9 7 RP15
8 8 7 1
CS2 1G
J1-6 9 19
ROTATE ARM 2 2G
RP16
3 7 8
J3-7 4
5 RP16
5 6
J2-13 RP16
3 4 +5V
2 2
J1-4 U14 RP16 C39,C40
INCREASE SID TLP621-4 1 2 20 100nF
RP7
2 1 1 16
J3-6 74HCT244
2 15 U18, U19
RP7 C58 C57 C60 C59 C55 C56 C49 C50
4 3 3 14 10
J2-14
4 13
RP7
J1-5 6 5 5 12
DECREASE SID
6 11
J3-5 RP7
8 7 7 10
8 9
J3-4
J1-2
HEIGHT UP
J2-11
J3-2
1 1
J2-12
HEIGHT DWN
NAME DATE SHEET / OF
J1-3 A3194-03
DRAWING F.GARCIA 07/07/03
D CN 10/512 S.Pérez 27/12/10 REVISED A.DIAZ 07/07/03 3/3 D C B A REV
C CN 09/339 F. Díaz 24/11/09
B CN 08/365 F. García 05/12/08
A board validated F. García 16/01/04 SEDECAL Optima URS Motorized
REV DESCRIPTION ISSUED BY DATE
A B C D E
A B C D E
SW1
HEIGHT UP SW
4 4
SW2
HEIGHT DOWN SW
SW4
INCREASE SID SW
SW5
3 DECREASE SID SW 3
J1
BUTTON COMMON SW3
1 HEIGHT UP ROTATE ARM
2 HEIGHT DOWN
3 INCREASE SID
4 DECREASE SID
5 ROTATE ARM
6 HEIGHT UP LAMP
7 HEIGHT DOWN LAMP RP1
8 INCREASE SID LAMP 1K5
9 DECREASE SID LAMP 1 2
10 UNDERTABLE LOCK LAMP 3 4
11 SID 100 LAMP 5 6
12 SID 180 LAMP 7 8
13 SID 200 LAMP DS1
14 LAMP COMMON UNDERTABLE LOCK LAMP
15
RP2
1K5
2 1 2 2
3 4
5 6
7 8
DS2
SID 100 LAMP
DS3
SID 180 LAMP
DS4
SID 200 LAMP
1 1
+12V
NAME DATE SHEET / OF
DRAWING F.GARCIA 01/01/01
A3195-01
1/1
REVISED A.DIAZ 02/02/01 A REV
Semi--Automatic
Bucky +24 VDC
K3 BUCKY EXP
R2
C2
B8 0 VAC TS1--3
0 VAC
ADD JUMPER
(SEE NOTE 2)
BUCKY SPLY
TS1--2
TS1--1 ADD JUMPER TO
B4 115/220 VAC
{ TS1--26
TS1--27 for 115 VAC
for 220 VAC
GND
Semi--Automatic
Bucky +24 VDC
K3 BUCKY EXP
R2
B8 C2 0 VAC TS1--10
0 VAC
ADD JUMPER
(SEE NOTE 2)
BUCKY SPLY
TS1--9
ADD JUMPER TO
B4 115/220 VAC TS1--8
{ TS1--26
TS1--27 for 115 VAC
for 220 VAC
GND
NOTE 3: In the case of noise due to Bucky, add R2=22 ohm, 1/2w, 5%; and
C2=470 nF, 250 VAC as shown.
Don’t add that R2--C2 for Liebel--Flarsheim 8000 Series Bucky, and
remove resistor R36 and R37 in the INTERFACE CONTROL board.
Bucky
+24 VDC
K3 BUCKY EXP
K5
R2
C2
B8 0 VAC TS1--3
7 0 VAC
ADD JUMPER
BUCKY SPLY
TS1--2
B6 GND
NOTE : In the case of noise due to Bucky, add R2=22 ohm, 1/2w, 5%; and
C2=470 nF, 250 VAC as shown.
NOTA : En caso de ruidos debido al Bucky, añadir R2=22 ohm, 1/2w, 5%, y
C2=470 nF, 250 VAC según se muestra.
Bucky
+24 VDC
K3 BUCKY EXP
K4
R2
C2
B8 0 VAC TS1--10
7 0 VAC
ADD JUMPER
BUCKY SPLY
TS1--9
B6 GND
NOTE : In the case of noise due to Bucky, add R2=22 ohm, 1/2w, 5%; and
C2=470 nF, 250 VAC as shown.
NOTA : En caso de ruidos debido al Bucky, añadir R2=22 ohm, 1/2w, 5%, y
C2=470 nF, 250 VAC según se muestra.
OPTIMA UNIVERSAL
4 4
DETECTOR ASSEMBLY
AEC ADAPTATION Board Note.- In case the AEC ADAPTATION Boad is not required, the ION CHAMBER
CABLE A3234-02 should be used (refer to drawing I/F-003)
ION CHAMBER
URS IC CABLE
ADAPT (A6721-01) J2 IC
GENERATOR CABINET
BUCKY CABLE BUCKY
(A3343-XX) Optima Bucky CABLE
Connections ( ) TS1 J1 (A3342-01)
for Bucky #2
BUCKY 1 MOTION 1
5(12) 1 B1
BUCKY 1 MOTION RTN 2
6(13) 2 B2
BUCKY 1 DR 3
4(11) 3 B3
2 2
115/220 VAC 4
1(8) 4 B4
0 VAC 8
3(10) 5 B8
GND 6
7(14) 6 B6
2(9)
26 26
27 See Note
GND CABLE GND Stud (Column)
GND Stud
1 1
+ FAN 1
4 4
TP3
F2 +30V
T 6A
CR8
FB1002L R26
1K 15 OHM
10 w
- + +32 VDC
3
FANS CABLE FANS HARNESS 3
TS10 TS11 + FAN 3
R25
2K7 +24 VDC +24 VDC
2w 1 1
GND GND GND
2 2
_
+ FAN 4
2 2
+ FAN 5
1 1