Short Notes
Short Notes
Gross Error:
Gross Errors mainly covers the human mistakes in reading instruments and recording and calculating measurement results.
Example: Due to oversight, the read of temperature as 31.5° while the actual reading may be 21.5°.
Gross Errors may be of any amount and then their mathematical analysis is impossible. Then these are avoided
by adopting two means:-
1. Great care is must in reading and recording the data.
2. Two, Three or even more reading should be taken for the quantity under measurement.
Systematic Error:
Systematic Errors are classified into three categories:
1. Instrumental Errors
2. Environmental Errors
3. Observational Errors
Instrumental Error:
These errors arise due to three main reasons.
1. Due to inherent shortcoming in the instrument.
Example: If the spring used in permanent magnet instrument becomes weak then instrument will always read high. Errors
may be caused because of friction, hysteresis, or even gear backlash.
2. Due to misuse of the instruments.
3. Due to Loading effects of instruments.
4. Calibration error
5. Ageing effect error
Environmental Error:
• These errors are due to conditions external to the measuring Device including conditions in the surrounding
instruments.
• These may be effects of temperature, pressure, humidity, dust, vibrations or of external magnetic or electrostatic
fields.
Observational Error:
There are many sources of observational errors:-
• Parallax, i.e., Apparent displacement when the line of vision is not normal to the scale.
• Inaccurate estimate of average reading.
• Wrong scale reading and wrong recording the data.
2
Static Correction
It is a value which is added or subtracted from measured value to get actual value.
Static correction = Atrue – Ameasured
Characteristic of measuring instrument
Accuracy
It is the closeness to the true value variable. Accuracy can be improved by calibration.
Note: Accuracy is maximum at full-scale deflection and as we move to initial scale limiting error increases and accuracy
decreases.
Repeatability
It is the degree of closeness when the same input is measured multiple times by same instrument with observer under same
condition.
Reproducibility
It is a degree of closeness when same input is measured multiple time by same instrument with different observer under
different conditions.
Sensitivity
Sensitivity is the ratio of change in output to the change in input.
Change in output
Sensitivity =
Change in input
It is a figure of merit of the instrument.
1
S=
𝐼FSD
Note: • If instrument has high sensitivity, then it has lesser loading effect.
• Sensitivity is also called as scale factor
Resolution
3
For the smallest change in input if there is definite change in output is called resolution.
Full scale deflection
Resolution =
No of division
Note: Number of divisions increases then resolution increases for same range
Hysteresis Error
If the instrument shows two different readings for the same input value then it is called as hysteresis error.
Drift
It is an error which is constant through the range of scale there are two types of drift.
(i) Zero Drift (ii) Span Drift
Zero Drift
Entire calibration shifts gradually due to permanent set.
Span Drift
It is proportional change in indication all along the upward scale is called span drift.
Arithmetic Mean:
x1 + x2 + x3 + x4 + ....xn xi
x= =
n n
Where x = arithmetic mean x1, x2, … xn = reading taken n = number of readings
Average range of error:
• Max error = Max- Mean
• Min error = Mean-Min
(Max error + Min error)
• Avg Range of error =
2
Deviation from the mean:
• Deviation is the departure of a given reading from the arithmetic mean of the group of readings. If the deviation
of the first reading, x1, is called d1, and that of the second reading, x2, is called d2, and so on,
• Then the deviations from the mean can be expressed as
d1 = x 1 – x
d2 = x 2 – x
dn = x n – x
Note that the deviation from the mean may have a positive or a negative value and that the algebraic sum of all
the deviations must be zero.
Average Deviation:
• The average deviation is an indication of the precision of the instruments used in making the measurements.
• Average deviation is the sum of the absolute values of the deviation divided by the number of readings.
d + d 2 + d3 + ... + d n di
D= 1 = (For n > 20)
n n
Standard Deviation:
By definition, the standard deviation of an infinite number of data is the square root of the sum of all the individual deviations
squared, divided by the number of readings.
Mathematically:
4
Variance:
• Variance or mean square deviation, which is the same as the standard deviation except that the square root is
not extracted.
• Variance, V = mean square deviation = (𝜎)2
Note:
0.4769
• Probable error = 0.6745 𝜎 or (where h is precision index)
h
• Standard deviation mean, m =
n
• Standard deviation of Standard deviation = =
2
absolute error Am − At A
r = = =
true value At At
Percentage static is given by the equation shown below:
% ∈r = ∈r × 100
At = Am – ∈r At
Am
At =
1+ r
Where, At = true value
Am = Measured value
Limiting Error:
Limiting Error is defined as the maximum deviation either in positive side or negative side in the measurement by an
instrument from the nominal value or true value.
Let us assume that the true value or nominal value of a quantity is At and the measured value by the instrument is Am, then.
Limiting Error ΔA = (Am – At).
For example, the nominal value of a resistor is 100 Ω with a limiting error of ±10 Ω. The magnitude of measurement will
be between the limits, (100 ±10)Ω i.e., 90 and 110Ω. Thus, the measurement of resistor by the instrument is guaranteed to
be between 90 & 110 Ω.
% Limiting error:
( Am − At )
% Limiting error = 100
At
% limiting error is a variable error
5
Composite Error
% r1
x1 = a ±
% r2
x2 = b ±
% r3
x3 = c ±
(a) Additional/Subtraction
x = x1 ± x2 ± x3
a b c
% Limiting Errors (x) = r1 + r2 + r3 %
a + b+ c a + b+ c a + b+ c
(b) Product/Division Terms
x1 x2
x = or x = x1 x2 x3
x3
% x = ( r1 + r2 + r3 ) %
(c) Power Terms
x1m x2n
x = p
or x = x1m x2n x3p
x3
% x = (m r1 + n r2 + p r3 ) %
Uncertainty Errors
Z = f ( x1 , x2 , xn )
u1, u2, … un are uncertainty errors
2 2 2
z 2 z 2 z 2
uz = x u1 + x u2 + + x un
1 2 n
uz = uncertainty of z
Absolute Instrument:
• These instruments measured the measurand in term of time constant or deflection of the instrument
• Absolute instrument is indirect reading instrument and th4se are used for standardization of instrument in laboratories
example Tangent Galvanometer
C (s)
H (s ) = = constant
R (s )
6
Example: Potentiometer
All indicating instruments are second order instruments. Ideally indicating instrument are critically damped (𝜁= 1) but
practically these are underdamped (𝜁 = 0.6 to 0.8)
I (Linear Scale)
d I (Works only for D.C)
SPMMC = (20Ω/v - 30Ω/v) SPMMC = sensitivity of PMMC
I
Td Td
S
A weight
"Torque"
ratio represent sensitivity ( unity )
weight
weight
Frictional error
Torque
Advantage:
Used in aeroplane, aerospace industry. ‘0’Hz frequency, uniform scale.
Disadvantage:
Only for D.C, not A.C.
Swamping Resistance
Swamping resistance has extremely low-temperature coefficient. Swamping resistance is made of manganin or constantan.
It is connected in series with the instrument. It is used to compensate the error due to temperature variation.
PMMC Ammeter:
Extend the range of PMMC ammeter. We connect low value resistance in parallel with meter.
Rsh = shunt resistance used for enhancement.
Rm = PMMC resistance
Rm Rm
R sh = R sh = Ω (m > 1)
Iext m −1
I − 1
Shunt: Manganin (86% Cu + 12% Mn + 2% Ni)
• Ammeter is always connected in series.
• Internal Resistance of Ideal Ammeter is always ‘Zero’.
PMMC Voltmeter:
Rse = series resistance need to add for enhancement.
Rm = meter Resistance
Rse = Rm (m–1) kΩ
Rse
m =1+
Rm
1 Rm
Sensitivity of voltmeter (SV) = = / volt
I FSD VFSD
8
Rm + Rse
Sensitivity of enhanced voltmeter, Sv = / volt
VFSD
If Rm increases, then sensitivity increases. Due to increase of sensitivity loading effect decreases therefore error decreases.
Rectifier type instruments can be used for both a.c. and d.c. voltage and current measurement. This instrument consists of a
rectifier and a PMMC instrument.
Reading of rectifier type instrument = form factor × reading of PMMC
2 Vrms 2 2 Vrms
I avg = I avg =
(
Rs + R f + Rm ) (
Rs + 2 R f + Rm )
I avg = 0.45 I rms I avg = 0.9 I rms
Advantage:
• Works on AC & DC both.
• It has lowest hysteresis error used only for power frequency range.
• M.I type, wattmeter does not exist.
• M.I type Ammeter: Condition to eliminate error due to frequency.
Lm Lsh
=
R m R sh m = sh
• M.I type voltmeter: to make voltmeter frequency independent a compensating capacitor of suitable value is
connected across series multiplier resistance (Rse)
Lm
C = 0.41 2
RSe
dL
= constant
d
Disadvantage of MI instrument.
(i) Scale is non-uniform
(ii) Due to electromagnet, it has a weak operating field.
(iii) Torque to weight ratio is less. Therefore, it has lesser sensitivity than PMMC.
(iv) It has hysteresis error due to iron disc frequency error is present in MI instrument.
(v) High stray magnetic field error due to weak magnetic field. To minimize this error shielding is required.
Note: In Moving iron instrument eddy current damping technique is not used because the produced eddy current will
distort weak operating field therefore, we use air friction damping technique.
Advantages
(i) EDM type instrument can be used with high precision grade accuracy up to 10 kHz
(ii) No eddy current and hysteresis error.
(iii) It can work on AC as well as DC.
(iv) Calibration for AC and DC are same for EDM type instrument. Therefore it is a transfer instrument.
Disadvantage
• EDM has non-uniform scale.
• Due to low operating magnetic field, it has low torque to weight ratio therefore it has lowest sensitivity.
Note:
• Induction instrument works for AC only and PMMC instrument works for DC only Remaining all the instruments
work for both AC as well as DC.
• In PMMC, power consumption is very low because it draws current only at the time of measurement therefore
there will be less temperature raise and in consequence less temperature error.
Loading Effect
The loading effect can be defined as the effect of the load impedance on the source. Loading effect radius the voltage effect
of voltage source.
If the sensitivity of the instrument is high i.e., with the high resistance, the loading effect will be less.
Ideally, an ammeter has a zero resistance and practically it has very low resistance. There will be no loading effect due to
the ammeter, but a small amount of voltage drop occurs due to the internal resistance.
l Heat
l I 2 Rt
Seeback Effect:
When two dissimilar materials having different temperature coefficient are connected to form a junction then due to
temperature difference a thermal emf produced across the metal. This effect is called seeback effect.
Reverse of seeback effect is Peltier effect.
(2) Thermocouple:
It works on the principle of seeback effect
(3) Temperature:
• Change in Resistance.
• Example: RTD, Thermistor
Electrostatic Voltmeter (ESV)
(1) Linear Motion Type:
• It works on the change in capacitor.
1 dc 1 V 2 dc
Fd = V 2 x=
2 dx 2 K c dx
• Suitable for High Voltage measurement.
(2) Circular Motion Type:
1 dc
d = V 2
2 d c = K c
Advantage:
• Works on A.C + D.C
• No Hysteresis error
• No eddy current
• Low power consumption.
• Best suitable for High Voltage
Disadvantage:
Non-uniform, no electrostatic Ammeter, Largest in size, not suitable for L.V.
• To extend the range of ESV – Potentiometer Method.
L
Vext = Vm = m Vm
l
• To using series capacitor:
Cm
Cse = (m > 1)
m −1
12
R 1
V V V R
Rm = = = =
I I R + IV V / Rv 1 + R / Rv
1
True value of resistance, R = R m
1 − Rm / Rv
If Rv ≫ Rm then Rm/Rv is very small.
R
R = Rm 1 + m
Rv
Rm − R Rm2
Relative error, r = =
R Rv R
R
If Rm R then r = −
Rv
Ra R
Note: The relative error of above two cases are equal when =
R Rv
R = Ra Rv
PS qr P P
R= + −
Q p + q + r Q q
Note: We short link resistance (r = 0) to convert kelvin bridge into wheat stone bridge.
Measurement of Medium Resistance
1 R 100k
Circuit diagram
PS
R=
Q
VS (R)
Deflection of galvanometer (for small unbalance), = Sv where Sv is voltage sensitivity.
( R + S )2
Bridge sensitivity
SvV
SB =
P Q
+2+
Q P
For a bridge with equal arms, R = S = P = Q
SvV
SB =
4
Ohmmeter
We use Ohmmeter to determine approximate resistance of circuit components. Ohmmeter is aid to precision bridge
laboratories because it is convenient to measure resistance directly.
15
Series Ohmmeter
It consists of d’Arsonval movement connected in parallel with shunt resistance R2. This parallel circuit is in series with
resistance R1 and a battery of emf E. The series circuit is connected to terminal A and B of the unknown resistance Rx.
When Rx = 0Ω then position of the pointer is marked 0Ω (short circuit condition).
When Rx = ∞Ω then the current in the meter drop to zero and position of the pointer is marked 0Ω (open side condition).
Shunt Ohmmeter
It can consist of a battery in series with an adjustable resistor R1 and a basic d’Arsonval movement (meter). The unknown
resistance is connected across terminals A and B, parallel with the meter.
When Rx = 0Ω then meter current is zero. (Short Circuit).
When Rx = ∞Ω then current will through the meter.
This ohmmeter therefore has ‘zero’ mark on the left hand side of the scale and infinite mark on the right hand side of the
scale.
R 100k
Meggar
Megger is working on the principle of ratio meter which is similar to PMMC. It is used for measurement of high resistance
in the range of M. It consisting of two moving coils (C1 and C2) which are perpendicular to each other, placed in a
permanent magnetic field. A hand device generator is used to provide supply to the coil C1 and C2 . Flux due to the current
flowing in the coil and permanent magnetic flux interact and produced deflecting torque which is calculated in term of test
resistance.
Application
It is most practically used in the substation for measurement of insulation resistance of cables, transformers, generators and
motors. Tt is basically design for high resistance measurement. In practical it is also used for measurement of low resistance
like earth conductor resistance, continuity of circuits and circuits breaker continuity.
AC BRIDGES
AC Brdiges:
1. Maxwell Inductance Bridge
2. Maxwell Inductance capacitance bridge
3. Hay's Bridge
4. Owen's bridge
5. Anderson's Bridge
6. Desauty's Bridge
7. Schering Bridge
8. Wiens Bridge
AC Bridge Sources:
• For low frequency applications – power line
• For high frequency applications – high frequency electronic oscillator
AC Bridge Detectors:
Detectors Frequency
D C Galvanometer 0 Hz dc
Vibration Galvanometer 5Hz-1kHz
Head phone Galvanometer 250 Hz – 4kHz
Tuneable Amplifier Galvanometer detector 10 Hz to 100k Hz
CRO In MHz Range
Z1, Z2, Z3 and Z4 are the impedance of bridge arms at balance point:
E BA = E BC or I1Z1=I2Z2
R2 R3
R1 =
R4
L1 = R2 R3C4
L1
Q = C4 R4 =
R1
Maxwell: Inductance Bridge:
R1 + jL1 R4 = R2 + jL2 R3
Under balance condition R1R4 = R2 R3
19
R2 R3
R1 =
R4
L1R4 = L2 R3
L R
L1 = 2 3
R4
Hay’s bridge
Hay’s bridge is suitable to measure inductance of high value of quality factor coils. Q > 10
2 R2 R3 R4C42
R1 =
1 + 2 R42C42
R2 R3C4
L1 =
1 + 2 R42C42
1
Q=
C4 R4
Owen’s bridge
L1 = R2 R3 C4
− R1 − R3
=
C4 C2
C
R1 = R3 4
C2
• This bride has two capacitors
• This bridge is suitable for medium Q-factor measurement.
• Owen’s bridge is used for incremental inductance measurement from known value of the fixed capacitance (C4).
• Hay’s bridge is suitable to measure inductance of high volve of quality factor coils. Q > 10
20
5. Anderson Bridge
• Anderson bridge uses a fixed capacitance ‘C’ and for the bridge balance the resistance r and r1 are selected as variable.
• These are used for measurement of low Q coil. (Q > 1)
• This bridge is used for accurate measurement of inductance over ‘mH’ to wide range.
• Bridge will give two balance equations because of ‘fixed point’
Note: It is more complex bridge and shielding of the bridge is complicated.
R3
L1 = C r ( R4 + R2 ) + R2 R4
R4
R R
R1 = 2 3 r1
R4
De-Sauty’s Bridge
• This bridge is suitable for perfect capacitor value measurement only.
• It is not used to measure the dielectric losses.
R3 ( R2 + r2 ) − R1
r1 =
R4
R4C2
C1 =
R3
Schering Bridge
Schering Bridge is used for capacitance and dissipation factor measurements.
D-factor = tan δ = ωRC
21
= Derating factor
= Dissipation factor
Dielectric losses, relative permittivity and properties of insulating material is also measured by this bridge.
R4C2
C1 =
R3
RC
r1 = 3 4
C2
Note:
(i) Pure capacitor (No-dielectric loss)
Wein’s Bridge
• Measurement of frequency is done by Wien’s bridge.
• This bridge is not suitable for measurement of frequency of signals having harmonics, because bridge balance is
difficult.
• It is used for measurement of frequency from 100 Hz to 100 kHz.
22
1
=
R1R2C1C2
1
f =
2 R1R2C1C2
R4 R2 C1
= +
R3 R1 C2
If R1 = R2 = R and C1 = C2 = C. Then R4/R3 = 2 and f = 1/2 πRC
Measurement of self- Measurement of Mutual Measurement
Measurement of Capacitance
inductance inductance of frequency
Maxwell’s Inductance Heaviside Mutual Inductance
D Sauty’s Bridge Wien’s Bridge
Bridge Bridge
Maxwell’s Inductance Schering Bridge (It is also used for Campbell’s Modification of
Capacitance Bridge measurement of relative permittivity) Heaviside Bridge
Carey Foster Bridge (use for
Heaviside Campbell equal
Hay’s Bridge measurement of capacitance in terms
ratio bridge
mutual inductance)
Anderson’s Bridge Campbell Bridge
Owen’s Bridge
Note:
1
(Q) Quality factor D= = tan
Q
𝑋𝐿 R
𝑅 XL
R XL
XL R
XC R
R XC
R XC
XC R
23
MEASUREMENT OF ENERGY
Single Phase Energy Meter (Induction Type)
• AC induction energy meter working on principle of induction is used to cumulatively record energy consumed by the
load and hence it is an integrating type meter.
• Potential coil of an energy meter must be highly inductive so that the error in measurement of energy is minimized.
Equivalent Circuit
Where is angle between voltage across potential coil and current in potential coil.
PT = True power = VI cos
Error in power = Pm − PT
= VI sin( − ) − VI cos
= VI sin( − ) − cos
For error in power to be zero should be equal to 90°, therefore we need highly inductive potential coil.
Lag coil or shading ring are used to maintain 90° phase displacement between potential coil voltage and flux passing
through the disc.
(vi) Braking torque is directly proportional to speed of disc.
At constant load speed is constant, therefore the torque of disc is equal to braking torque.
Energy = (speed) dt
E = N dt
1
(vii) Speed of disc
distance of permanant magnet from center of disc
(viii) Creeping: If the friction is over compensated then the disc starts rotating slowly without any current in the circuit
coil. Creeping can be reduced by proving holes or slots on the opposite sides of disc or by placing a small iron piece
on disc.
(ix) For over voltage compensation we use saturable shunt magnet between the limbs of shunt magnet.
(x) During overloading current coil flux increases which can damage the meter. For diversion of excess flux saturable
series magnet is kept between the limb of series magnet.
Phantom Loading
Phantom loading is use in the calibration of wattmeter and energy meter. The pressure coil and current coil are separately
exited by supply. In this method rated voltage is applied to potential coil through small voltage source and the current
energises the current coil at very small voltage. The total current supplied for the phantom loading is the sum of the pressure
coil current which is supplied at normal voltage and the current of the current coil supply at low voltages.
Measurement of Power
EDM Type Wattmeter
dM dM V dM Pavg . dM
(i) Td = I1I 2 cos = I cc .I pc cos = I . cos = .
d d RS d RS d
(ii) RS = potential coil resistance
Pavg dM
(iii) = Pavg
kc RP d
• Wattmeter measures average active power.
• Wattmeter works on the principle of electro-dynamo meter (EDM).
• It is consisting of fixed and moving coil.
• The thickness of current coil is larger compare to potential coil because CC carries large load current.
• Moving coil is connected with high resistance in series for measurement of load and supply voltage hence called
potential coil.
• PC of wattmeter must be highly resistive so that error in the measurement of power is minimized whereas PC of
energy meter must be of highly inductive so that error in the measurement of energy is reduced.
25
• In EDM wattmeter spring provides controlling torque and air friction damping technique is used.
Blondel’s theorem:
• ‘n’-phase, ( n + 1) wire –
Balance Unbalance
1-watt-meter required ‘n’ wattmeter required
‘n’-phase, ‘n’ - wire –
Balance Unbalance
(n – 1) watt-meter required (n – 1) watt-meter required
3-phase Active Power
0° 90° 3 3 W1 = W2
VL I L VL I L 3VL I L
2 2
VL I L 3
30° 0.866 VLIL VL I L W1 = 2W2
2 2
3 3 W1 = W
60° 0.5 VL I L 0 VL I L
2 2 W2 = 0
VL I L VL I L W1 = +ve
90° 0 − 0
2 2 W2 = –ve
Error in Wattmeter
(a) Potential coil on source side
(This connection is preferred for small load current)
26
Pm = Measured Power = PT + I 2 rc
Pm − PT
% r = 100
PT
I 2 rc
= 100
PT
(b) PC on load side
I1 = I
dM dM
Defection torque = I1I 2 cos = I1I 2 cos( − )
d d
Pmeasured = (V cos ) ( I ) cos( − )
= VI cos cos( − )
= measured power
True Power cos
= =
(i) Correction factor Measured Power cos cos( − )
(ii) Percentage error ( % r )
Pmeasured Ptrue
27
Pmeasured Ptrue
• Multiplication factor of wattmeter (Mf)
Voltage setting Current setting Power factor
Mf =
Maximum range of scale
Note: In case of laboratory wattmeter iron shielding is used around the meter which will divert the external magnetic field
so that error is reduced.
• Different phosphorous coating is used on the screen which converts heat energy into light energy depends on
application.
• Bandwidth (Hz) × Rise time (sec) = 0.35
• Probe is a connecting channel between two devices. Probe must have high input resistance and low output resistance
so that there will be no loading effect.
• Time constant of probe and time constant of CRO must be equal to make probe independent of frequency.
( R p C p ) = ( RCRO CCRO )
Vd .x.
y=
2d .Va
y xl
Sensitivity, S = =
Vd 2dVa
1 Vd 2dVa
Deflection factor = = =
S y xl
(i)
30
y x
= sin −1 internal = sin −1 internal
yexternal xexternal
(ii) LP in 2nd and 4th quadrant
x
= 180 − sin −1 1
x2
y
= 180 − sin −1 1 second possibility = (360 − )
y2
Measurement of frequency by CRO:
• Frequency ratio.
fy = Maximum no. of horizontal Intersection drawn to LP
fy = Maximum no. of vertical Intersection drawn to LP
fy = Maximum no. of horizontal tangent drawn to LP
fx = Maximum no. of vertical tangent drawn to LP
fy Number of Horizontal Tangencies
=
fx Number of Vertical Tangencies
f high Total no. of Bright spot in LP
=
flow 1
f high Total no. of Corner in LP
=
flow 1
Example:
31
Digital Voltmeter
Specifications of DVM
Resolution (R):
Resolution is the term used to describe the smallest input value that a digital meter can measure.
1
R= in fraction
10 N
V
R = maxN in volts
10
N = Number of full digits
Sensitivity:
Sensitivity is the lowest input value that a digital meter can measure within the specified range.
Number of digits in display = N + 1 where N is the number of full digits which may display any digit from 0 to 9 and one
left most digit is called half digit which may display only 0 or 1.
1
3 Digits display:
2
3V 30 V 300 V
Resolution of minimum voltage 0.001 0.01 0.1
Maximum voltage 2.999 29.99 299.9
3
3 Digits display:
4
3V 30 V 300 V
Resolution of minimum voltage 0.001 0.01 0.1
Maximum voltage 2.999 29.99 299.9
4-Digits display:
1V 10 V 100 V
Resolution of minimum voltage 0.0001 0.001 0.01
Maximum voltage 0.999 9.999 99.99
Over Ranging
1
The extra digit is switched ON is called over ranging.
2
1
3-Digit Display 3 Digits display
2
Percentage error due to extra count is given by,
Count Full scale value
Error = Percentage error due to reading +
Number of counts
32
Q-Meter
Q-Meter works on the principle of series.
Applications of Q-Meter
1. Measurement of unknown capacitance
−Cd
• Error due to distributed capacitance %error = 100%
Cd + C
• The magnitude of % error introduced in measurement of Q by inserting a resistor r.
−r
% error due to insertion resistance r = 100%
Rcoil + r
33
Current Transformer
N2
• Turns ratios = (n) = ,
N1
I P actual
• Actual/Transformation ratio = (R) =
I S actual
X + Xs
• Secondary burden angle → (S) = tan −1 e
Re + Rs
Ip
• Nominal ratio K =
Is
I w cos + I m sin
R= +n
IS
•
not valid for = 0 because some approximation is made
• Iw = Io sin α
• Im = Io cos α
•
Ip = ( nI S )2 + I m 2
when S = 0 for R burden
R
• Ratio corelation factor =
K
K−R
• Ratio error = 100%
R
Ip 180
• Phase angle error (degree) = (𝜙) =
nI s
• Phase angle between IP & IS = (180º – 𝜙)
• Magnetising MMF = Im × primary turns
• Iron loss = EP × Iw
E
• EP = 2
n
• ES = I S Z S
• Bar primary = 1 turn
• VA rating of a current transformer expressed the burden on it.
Burden current transformer
The external load applied to the secondary of a current transformer is called the burden.
The burden can be expressed as
(i) Total impedance in ohms of the circuit or the total volt-amperes (VA)
(ii) Power factor at aa specified value of current or voltage and frequency.
34
Potential Transformer
N1
• Turns ratio = (n) =
N2
V V
• Nominal ratio (k) = P = 1
VS V2
VP V1
• Actual transformation ratio = (R) = =
VS V2
X
• Secondary burden angle = (Δ) = tan −1 1
R1
35
Ri li
R0 l0
R0
V0 = Vi
Ri
l0
V0 = Vi
li
V0 R0 x
= = (x is change in displacement due to which there is change in voltage)
Vi Ri l0
Vi V0 V V
S = Senstivity = = .....V/ S = Senstivity = i = 0 .....V/
Ri R0 Ri R0
Stroke: Maximum displacement measured by the potentiometer is called stroke.
Stroke = li
36
V0 = Vs1 – Vs2
V Volt
Sensitivity S = 0
x mm
Hall-effect Transducer
If a current carrying conductor placed in magnetic field experiences force so that the internal charges within the material
are displaced which produces electrical field. Electrical field is perpendicular to the direction of current and magnetic field.
Due to electric field an emf is develop across surface of the metal which is call hall voltage.
Hall voltage:
BI
VH =
nqW
B = Magnetic flux density
I = Current flowing through sensor
n = No of charge carrier per unit volume
q = charge
W = thickness of sensor
Hall Coefficient RH:
E
RH =
J B
J is current density
E is induced electric field
B is magnetic strength
Applications:
1. VH BE : Therefore, it is used as multiplier of 2-signals.
37
Stress
Note: Young’s modulus =
Strain
q C
Charge sensitivity = d =
F N
E V V-m
Voltage sensitivity = g = = 0
P t p N
38
V0 = g P t
V0 P (P is pressure, E is electric field)
RTD
• RTD has positive temperature coefficient of resistance, and this is 1st order device.
• It uses mercury as a liquid thermal expansion material which is used to measures temperature upto 400 °C.
• Solid thermometer used platinum, copper, nickel measures up to 900°C.
• Platinum has more stability & it has higher sensitivity compared to other materials. These are used in electrical
heaters, micro-ovens and small size boilers.
• Wheatstone bridge is used for converting non-electrical into electrical i.e., changes in the temperature will cause
change in the resistance and hence bridge became unbalanced this unbalanced voltage is calibrated in terms of
temperature change.
Thermocouple
If two different metals farming a junction operating at different temperature an emf is produced across the junction which
is proportional to the temperature difference.
Thermistor
• It is a combination of Nickel, Chromium and Copper and metal alloys it has highest sensitivity and small in size in
the range of mm. Used to measure both positive and negative temperature.
• It is used in multiple application like temperature control in electrical heaters and electronic circuits for Q point
(Quotient point in BJT) stabilization.
• It has negative temperature coefficient of resistance.
• By using self-heating property, it is used to measure other than temperature like vacuum, liquid and gases flow rate,
thermal conductivity etc.
• In case of temperature measurements, it is corrected with a fixed resistor across the thermistor for compensation of
self-heating property.
1 1
−
e
T To
RT = Ro
β = constant = (3000 – 4000)
T, T0 = temperature in Kelvin
R0 = Resistance at T0
39
dRT
Sensitivity = S = Ω/K
dT
Radiation Pyrometers
• It is working on the principle of black body radiation.
• If a high temperature metal surface is able to radiate thermal waves the thermal energy flow between 2 radiated
metals of A and B is given by relation.
q = (TA4 − TB4 )
= Stefen Boltzmann Constant
q = Thermal energy flow
TA, TB ⇒ Temperature of metals A, B
Bimetallic Strip
It consists of two metals normally made of nickel and chromium which has different thermal expansion coefficient when
the temperature is change the metal will expand and produces bending moment, this principle is used in the bimetallic strips.
Application
Used in MCB
Note:
Parameter Applications
Resistance
(i) Potentiometer device Pressure, displacement
(ii) Resistance strain gauge Force, torque, displacement.
(iii) Resistance thermometer Temperature, radiant heat
(iv) Thermistor Temperature, flow
(v) Photoconductive cell Photosensitive relay
Capacitance
(i) Variable Displacement, pressure
(ii) Capacitor microphone Speech, music, noise
(iii) Dielectric gauge Liquid level, thickness
Inductance
(i) Magnetic circuit Pressure, force, displacement
(ii) Differential transformer Pressure, displacement position
(iii) Eddy current gauge Displacement, thickness
40