② Load the designated AIA-360 sample cups in positions 1 through 12 and the
sample container Test Tubes in positions 13 through 24 of the turntable
sample holder.
*Make sure to firmly press each one intoplace.
Positions 1 to 12
Positions 13 to 24
3-3 - 1
AIA-360 service Manual Rev.0
(The carousel begins to rotate and the sample containers are automatically read.
The lower half of the screen displays information, including the read OK and read
error counts, together with the information read.)
“UP, DN (S103, 104) both=1:count value”=(recognized as test tube with count of 250 or more)
*(Confirm that the sensor lamp (Green) is on during reading operations. Failure
to light up indicates an electrical fault in the system. Therefore check that
connector is properly installed and there are no disconnected lines, then retry
the reading operation.)
3-3 - 2
AIA-360 service Manual Rev.0
④ When the reading system failures occur, turn off the system power and loosen
the two screws securing the sample sensor. Then adjust sensor position by
changing mounting angle.
*(The sensor is designed to be set at a right angle to the left side.)
3-3 - 3