The CS 7200 User Guide describes and illustrates the workflow for digital dental radiography using an
imaging plate (storage phosphor screen) for radiographic diagnostic intraoral images. We recommend that
you thoroughly familiarize yourself with this Guide in order to make the most effective use of your system.
Important:
Read and follow the operating instructions in this guide carefully and always
keep it within easy reach.
The room in which this unit is to be installed must comply with all official
applicable regulations.
Improper use and handling can create hazards and cause damage to the unit.
To prevent personal injury or material damage, you must observe all safety
information.
Consult the CS 7200 Safety, Regulatory and the Technical Specifications User
Guide before using the CS 7200.
No part of this Guide may be reproduced without the express permission of Carestream Health, Inc.
U.S. Federal law restricts this device to sale by or on the order of a dentist or physician.
0086
Contents
Chapter 6
Maintenance
Chapter 11
Publications History
ii
1 Conventions in this Guide
Introduction
Note: The CS 7200 is intended for dental professionals. You are required to
have training to use the CS 7200.
The CS 7200 system is used to scan and review intraoral dental X-ray images. When scanning the X-ray
exposed imaging plate a digital image is displayed on the workstation screen. After scanning, the scanner
erases the imaging plate and ejects it. The imaging plate is ready for re-use.
The exam acquisition’s scanned image is sent to the Dental Imaging Software on the workstation. Using the
Dental Imaging Software, the image is processed and reviewed.
Note: The scanner is supplied with a separate viewing and archiving software
package: Dental Imaging Software, approved by Carestream Health.
On/Off switch
Turn scanner On: press for 1 second
Shut down the scanner:
1 press for 2 seconds to reset the scanner, run BIT.
The scanner remains turned on.
1 Base
2 Cables compartment
3 Cover
The imaging plate has two sides: the active side for capturing the X-ray image and the inactive side. The
active side of the imaging plate includes an orientation ‘c’ mark with a dot inside to facilitate positioning
protocols within the patients’ mouth. The inactive side has a printed dot that corresponds to the location of
the orientation mark on the active side. This dot is visible through the transparent side of the hygienic sheath.
Active side
Inactive side
Environmental Requirements
Check the following ambient operating condition requirements of the room, before installing the unit:
Temperature: 15–35 °C
For optimal results it is recommended that the light level in the room is approximately 500 lux (normal light
conditions) to minimize the imaging plates exposure to light.
Important: The scanner should be positioned so that there is always easy access
to the mains power supply socket.
Electrical Requirements
Table 2 Electrical installation specifications
Important:
It is MANDATORY to check that the computer system configuration is
compatible with the computer minimum system requirements specified in the
Safety, Regulatory & Technical Specifications Guide. If necessary you must
update your computer system configuration.
4. Connect the USB cable to the dedicated connector in the cable compartment, and route the cable as
shown on the below image to secure the cable in its place:
6. Connect the free end of the USB cable to the main USB port in the back of the computer.
7. Connect the Power Supply cable plug to the Mains wall socket.
NOTE:
Before installing CS Imaging Software check that:
1 Insert the Carestream Imaging software DVD-ROM in the DVD-ROM drive of the PC.
The installation wizard is automatically loaded.
Note: If the installation wizard is not loaded, run the CSISuiteSetup 1.17 file
located on the installation DVD-ROM.
2 The Installer Language dialog box is displayed. Select the required language and click OK to
continue.
4 The License Agreement screen is displayed. Click I Agree to continue the installation.
Patient file: is selected by default. You can deselect it, if you do not want the patient database.
8 The Installing wizard screen is displayed. Let the installation process run automatically.
9. Select Reboot now to complete the installation and then click Finish.
After the computer is restarted, proceed to “CS 7200 Drivers Installation” on page 18.
Note: If the installation wizard is not loaded, run the wizard file located on the
installation DVD-ROM.
3. Select language from the list on the upper right side of the Installation Wizard window.
5. Minimum system requirements are verified and the status of each checked item is displayed. Click Next
to continue the installation.
6. In the IP2 Image Acquisition window, click Install to install the required plug-ins on your computer.
8. The Destination Folder window is displayed. Click Next, if you accept the default destination folder or
browse to choose another destination folder, and then click Next.
10. The New Scanner Files window is displayed. Click Next to continue without updating the scanner
files.
1. After the installation is complete, the license activation message appears:. Click Next to continue.
2. If the scanner is not connected and powered on, the following message appears:
5. On the Product Activation and Registration page, fill in the sign in details.
6. Select the product type, fill in the product details and click Register Product.
9. In the Online License Activation window, enter the license key received with the scanner, select the
license check box and click Activate.
1. Start the CS Imaging Software and select or create test patient. See “Starting the Imaging Software” on
page 42.
2. Expose a plate size 2 with a metal object (such as coin or key) on it.
3. Load the exposed plate into the scanner and wait until the scan is complete and the image is displayed
on the CS Imaging Software main screen. .
2. The gallery of the available looks is displayed at the bottom of the screen.
3. Click Apply at acquisition to apply this view during the image acquisition process.
4. Hover the mouse pointer over the selected frame in the gallery. The following options are available:
3. Click Create.
4. Adjust brightness and contrast of the new look by using Brightness and Contrast sliders.
Use the zoom slider to zoom in or out certain areas of the image.
5. Click Create to create a new modified look and add it to the frames gallery.
3. Click Modify.
6. Click OK if you really want to exit without saving chnages or click Cancel to return to the CS Adapt
Library screen.
3. Click Delete.
4. In the Delete Filter dialog box that appears, click Yes to confirm the action.
Single images.
Full Mouth Series (FMS).
The Full Mouth Series (FMS) is a static representation of the mouth of the patient using a series of intraoral
images. The images are placed in fixed numbered frames.
Select a Format enables you to select FMS template options to select for acquisition. You can modify existing
templates and create new templates (see the Carestream Imaging Software Online Help).
Note: Only FMS templates for CR images can be used. Using other FMS will
cause images not to be displayed.
Control Panel: enables you to adjust the brightness and contrast of the image. For more information on
Control Panel functions, see the CS Imaging Software Online Help.
Optiview:Maps pure black to blue and pure white to green, leaving other shades
unchanged.
Colors #1: Maps pure black to blue and pure white to red.
Colors #2: Maps middle-gray to orange and leaves pure white and pure black in their
original states.
CS Adapt Library: Opens the CS Adapt Library software that allows to manage your
filters.
Adapt - Unfiltered Endo1: Enhances contrast at the canals and roots (produces brighter
images).
INTRA toolbar: enables you to adjust the image display. For more information on Intraoral toolbar
functions, see the Carestream Dental Imaging Software Online Help.
Display annotations: click to display any annotations on the image. Click again to
hide the anotations.
Relief filter: click to enhabce the outlines of the shapes in an image by displaying the
image in relief mode.
Isodensity dots: click to apply color to pixels of the same bone density.
Caries Detector: click to open the Logicon Caries module for detecting interproximal
caries.
1 Selecting or Creating a Patient Record and Accessing the imaging window. See “Starting the Imaging
Software” on page 42.
2 Inserting the Imaging Plate into a hygienic sheath and sealing it. See “Preparing the Imaging Plates” on
page 43.
3 Positioning, exposing and removing the imaging plate from the patient’s mouth . See “Performing the
X-rays” on page 44.
4 Disinfecting the Hygienic Sheath. Refer to the CS 7200 Safety and Regulatory Guide (6K9678).
5 Scanning the imaging plate and reviewing the image. See “Scanning Imaging Plate” on page 47.
1 On your desktop, double-click the CS Imaging Software icon or click Start > All Programs > ( the
name of your) Dental Imaging Software.
To create a new patient record: Click the Create New Patient Card button alongside the search
field to create a new patient card. For more details, see “Create a New Patient Card” on page 75
The CS 7200 indication LED turns from yellow to green, indicating that the scanner is ready for use and
the Imaging Window appears.
3. Once you have selected or created a patient, in the Patient Browser Toolbar (C) click the arrow button to
open the Imaging application. The Patient Browser window remains open and the Imaging Window is
displayed.
Important: Use a NEW hygienic sheath for each new patient to prevent
cross contamination.
Important: Use a NEW hygienic sheath for each new image to prevent
humidity and contamination of the plate.
Tip: If the imaging plate has not been used for a while, proceed to erasing it
first. See “Manually Erasing the Plate” on page 56..
2 Insert the imaging plate into its protective hygienic sheath with the inactive side facing the adhesive
strip, so the imaging plate’s size number is seen through the transparent side of the hygienic sheath and
the orientation mark visible in the bottom corner (A).
4 Handle the imaging plate by holding the hygienic sheath’s empty edge where the silicone strip is.
5. Select an appropriate positioner for the region of interest and the size of the plate.
It is recommended to continue using X-ray positioning techniques and tools to ensure the resulting image is
complete and includes all the information required for diagnosis.To facilitate matching the orientation of the
image in the software to the clinical reality, we recommend to position the imaging plate in the mouth of the
patient with the orientation mark always towards the bottom.
1 Select the X-ray exposure time according to the region of interest and the patient type. Follow the user
instructions of your X-ray generator. The following tables provide guidelines for exposure times for an
X-ray generator at 70 kV and 7 mA. Add your values for the exposure time in seconds in the column on
the right.
Tip: If images are too dark, reduce the exposure time. If images are too white
or look grainy, increase the exposure time.
3. Move the X-ray generator tube head close to the patient and align it with the tooth of the patient and the
CS 7200 imaging plate.
Important: Make sure you can keep visual contact with the patient
during the X-ray.
6. Trigger the X-ray with the remote switch of the X-ray generator.
7. Remove the CS 7200 imaging plate from the mouth of the patient.
Note: Make sure you insert the plate with the active side facing up. If the plate
is inserted when positioned upside down, an error message is displayed: Plate
loaded upside down.
When the scan is complete, the imaging plate is erased and then ejected. The indication LED stops
blinking and remains steady green.
The ideal image quality is when the indicator of the quality of exposure is a full green bar in the Control
Panel. Avoid under-exposed or over-exposed images indicated by a partial or full red bar.
1 Access the Dental Imaging Software. See “Starting the Imaging Software” on page 42.
2 Click the Use a format button in the Imaging Window toolbar to access the Select a format window.
3. From the list of available formats (templates), select the required template and click Open.
4. Select an appropriate positioner for the region of interest and the size of the plate.
5 Prepare all imaging plates required for completing FMS template. See “Preparing the Imaging Plates”
on page 43.
6. Position the imaging plate in the mouth of the patient depending on the region of interest.
8. Select the X-ray exposure time according to the region of interest and the patient type. Follow the user
instructions of your X-ray generator. The following tables provide guidelines for exposure times for an
X-ray generator at 70 kV and 7 mA. Add your values for the exposure time in seconds in the column on
the right.
1. For each imaging plate you prepared, perform the X-ray. See “Performing the X-rays” on page 44.
2 Scan all X-rayed imaging plates in the required order displayed on the Dental Imaging screen. See
“Scanning Imaging Plate” on page 47.
The next frame to be scanned is always highlighted in green on the displayed FMS format.
Tip: For FMS, it is recommened to use the transportation box that helps to
sort the plates in the correct order.
1. Click the frame for which you want to retake images. The selected frame is highlighted in green.
2. Using the mouse cursor, drag the selected image outside the template window.
5 After the scan is complete, the new image appears in the FMS template.
6 Close the template window to save the performed changes.
1 On the Dental Imaging Software main tool bar, click the CS 7200 Configuration icon.
The CS 7200 window appears:
1 On the Dental Imaging Software main tool bar, click the CS 7200 Configuration icon.
The CS 7200 window appears:
Settings
Tools
About
Settings Tab
The Settings tab enables user to change the scanner preferences:
High resolution
High speed
3 Scanner sound: select the check box to mute the scanner sound.
4 SMPTE: click to adjust monitor brightness and contrast.
5 Default exposure values: change the default exposure values if required.
4 Click Preferences to open the CS 7200 Preferences window and select the Tools tab.
The Retrieve Images button is enabled and the number of pending images is diplayed.
6 Seelect the required images from the list by selecting the check box next to each image.
The Restore and Delete buttons become available.
About Tab
Open the about tab to view scanner information that includes firmware, hardware and software versions,
serial number and other details.
1 On the Dental Imaging Software main tool bar, click the CS 7200 Configuration icon.
The CS 7200 window appears:
2 Select the desired scan resolution and click OK to save the selection.
For all maintenance information, refer to CS 7200 Technical Specifications, Safety and Regulatory Guide.
Quick Troubleshooting
Occasionally, malfunctions can occur during use in the event of an incorrect action. An error
message is displayed.
During the CS 7200 power up, system performs a self test: ejects a plate (if left in the scanner) and reports
whether the system is ready for operation or a malfunction is detected.
The detected malfunction is indicated by the CS 7200 LED status and sound notification and/or by an error
message that appears on screen.
When the system functions properly, the LED is green. If the LED color is yellow or red, it indicates an
existence of the malfunction.
IMPORTANT
If an Error Code message is displayed, the malfunction persists or more
serious conditions occur, contact a qualified technician. When you call
the qualified technician have the following information ready:
Scanner performs a Electrical noise from the peripheral Make sure the scanner is located away from
spontaneous reboot. devices. the suspected noise generators.
Connect the power supply to a separate
mains wall outlet.
The plate loading process is The hygienic sheath is partially sealed. Make sure the hygienic sheath is fully open,
not smooth and requires before loading the plate.
some force to apply.
Scanner returns to home The imaging plate was retrieved before Wait for the scanner to return to its home
position before the loading is it was completely loaded into the position.
completed. scanner. Perform a complete loading process.
Vertical lines appear on the Scanner was not stable or hit during the Make sure the scanner is positioned on a
scanned image. scan. stable surface.
Make sure you do not touch the scanner
during the scan.
During the power up, There is no power supply to the
scanner remains off and the system:
indication LED is off.
Power Supply is not connected to
The CR icon disappears from the mains wall outlet. Make sure the Power Supply is properly
the Dental Imaging Software Power supply is not connected to connected
toolbar. the scanner.
Problems with the facility power
supply.
Power Supply is damaged.
Scanner malfunciton.
LED indicator is steady The Dental Imaging Software is Start the Dental Imaging Software.
yellow. closed. If the problem persists, restart the scanner.
LED indicator remains There is no USB communication Make sure you use the USB cable supplied
blinking yellow for more between the scanner and the with the scanner.
than one minute. computer. Make sure the USB cable is properly
connected on both sides.
Try to connect the USB cable to another
USB port on the computer.
Replace the USB cable.
Call Service.
Scanner malfunction.
66 Chapter 7 Troubleshooting
Symptom Cause Corrective Action
LED indicator blinks red. Restart the Scanner.
The cause is unknown. If problem persists, call Service.
No error message is
displayed.
LED indicator blinks red Scanner hardware malfunction. Restart the Scanner.
during the Scanner power on If the problem persists, repeat for three
and the noise is sounded. times.
If problem persists, call Service.
LED indicator is red. Scanner hardware malfunction. Restart the Scanner.
If problem persists, call Service.
Error code #9011
LED indicator blinks red Scanner hardware malfunction. Restart the Scanner.
Error code #9012 If problem persists, call Service.
LED indicator is red or off. Scanner hardware malfunction. Restart the Scanner.
If problem persists, call Service.
Error code #9014
LED indicator blinks red Insertion panel is detached from the Properly attach the panel to the scanner.
Error code #9015 scanner. Warning: Do not change the position of the
plate selector knob.
Click OK to close the error message and
proceed.
LED indicator blinks red The plate was not scanned and Remove the insertion panel.
remained inside the scanner. While holding the visible part of the plate,
Error code #9016 lift the scanner protection door.
Remove the plate from the scanner.
If the plate fell inside the scanner, remove
the service tray and remove the plate.
Attach the insertion panel.
Click OK to close the error message.
Perform a scan.
LED indicator blinks red Scanner hardware malfunction. Restart the Scanner.
If problem persists, call Service.
Error code #9021, 9022,
9023
LED indicator blinks red Scanner hardware malfunction. Restart the Scanner.
Power Supply malfunction. Replace Scanner Power Supply.
Error code #9024 If problem persists, call Service.
LED indicator blinks red Scanner hardware malfunction. Turn off the scanner.
Wait for five minutes and restart the
Error code #9025 Scanner.
If problem persists, call Service.
LED indicator blinks red Software files version does not match
the scanner files version. Update scanner files.
Error code #9030
LED indicator blinks red Scanner hardware malfunction. Click OK to shut down the scanner.
Restart the scanner. The plate is ejected.
Error code #9031 Rescan the plate.
If the problem persists, call Service.
LED indicator blinks red Scanner hardware malfunction. Click OK to eject the plate.
Rescan the plate.
Error code #9032, 9036 If the problem persists, call Service.
LED indicator blinks red Scanner hardware malfunction. Restart the scanner.
If the problem persists, call Service.
Error code #9039
LED indicator blinks red Plate was not loaded properly. Click OK to eject the plate.
Rescan the plate.
Error code #9043 If the problem persists, restart the Scanner.
Scanner hardware malfunction. If the problem persists, call Service.
LED indicator blinks red Scanner hardware malfunction. Click OK to shut down the scanner.
Restart the scanner.
Error code #9061-69
LED indicator blinks red Plate was not scanned properly. Click OK to eject the plate.
Rescan the plate.
Error code #9046, 9049 If the problem persists, restart the Scanner.
Scanner hardware malfunction.
If the problem persists, call Service.
LED indicator blinks red Scanner hardware malfunction. Click OK to close the error message.
Call Service and continue with regular
Error code #9060 scanner operation.
Note: This error will appear on each power
up, until it is fixed by the Service.
LED indicator blinks red Scanner hardware malfunction. Click OK to close the error message.
Restart the Scanner.
Error code #9080 If the problem persists, call Service.
LED indicator blinks red Scanner hardware malfunction. Click OK to eject the plate.
Plate was not erased properly. Manually erase the plate.
Error code #9081 If the problem persists, call Service.
LED indicator is steady red
Plate was loaded upside down or
Error beeps are sounded Retrieve the plate and load the plate properly.
diagonally.
when loading a plate
LED indicator blinks red Plate was loaded upside down or Click OK to eject the plate.
diagonally. Load the plate properly.
Error code #9090 Plate was not scanned.
LED indicator blinks red Scanner hardware malfunction. Click OK to shut down the scanner.
Restart the Scanner.
Error code #90120 Rescan the plate.
If the problem persists, call Service.
LED indicator blinks red The Scanner files are not up to date.
Update the scanner files.
Error code #90800
LED indicator blinks red Scanner drivers files are not up to date.
Update the Scanner drivers files.
Error code #90801
LED indicator blinks red Scanner hardware malfunction. Click OK to close the error message.
Plate was not scanned or erased Restart the Scanner.
Error code #90852 properly. Reload the plate for scan.
If the problem persists, call Service.
Plate was not ejected Plate was not properly ejected. Remove the insertion panel.
Remove the service tray and remove the
plate.
Attach the insertion panel.
Scanner disconnects during External electronic noise that interferes Verify using the original CS7200 USB cable
the scan/half image appears with communication between the with the ferrites on ot.
after scan is completed/ther scanner and the computer. Verify that devices emitting high electronic
are lines on the image noise (such as Wi Fi routers, UPS and etc.)
are not connected in close proximity to the
scanner and do not use the same power
inlet.
Replace the Controller Card.
68 Chapter 7 Troubleshooting
8 Appendix
1. After the installation is complete, the license activatuion message appears:. Click Next to continue.
70 Chapter 8 Appendix
6. Click Continue to open the Product Activation and Registration - Create Customer Record page.
7. Fill in all required details and click Continue at the bottom of the page.
8. After filling in all details, an account is created and an email is sent to the provided email address.
9. Open the received email and click the provided link to activate your Carestream account.
Serial Number: type in the serial number that appears on the Data Plate label located on the
bottom of the scanner.
License Key: type in the license key that appears on the post card received with the scanner.
User Code: type in the user code that appears in the Offline License Activation window.
13. Click Continue. An email that contains product nem, product license number and activation code is
sent to your email address.
72 Chapter 8 Appendix
14. On the Activation Complete page, copy the displayed activation code..
15. Paste the copied activation code into the Offline License Activation window and click Activate.
74 Chapter 8 Appendix
Create a New Patient Card
To create a new patient card, follow the below steps:
1. In the Patient Browser toolbar, click the Create New Patient Card button.
Note: First Name, Last Name and Gender are mandatory fields. All other
fields are optional. Grayed out fields are not editable.
4. The Patient Card window closes and the new patient appears in the patient list.
Manufacturer’s Address
Authorized Representatives
Authorized Representative in the European Community
EC REP
List of Consumables
Publication Changed
Publication Date File Name Notes
No. Chapters
CS 7200 User and
2015-03 6M0940 First Release
Installation Guide
Added table to the CS 7200 User and
2015-06 6M0940 Revision 2
Consumables List Installation Guide