0% found this document useful (0 votes)
35 views8 pages

UOP 988-11 Low Trace Sulfur in LPG and Gaseous Hydrocarbons by Oxidative Combustion With Ultravi

The document outlines UOP Method 988-11 for determining low trace sulfur in liquefied petroleum gas (LPG) and gaseous hydrocarbons using oxidative combustion and ultraviolet fluorescence detection. It details the method's scope, apparatus, reagents, procedure, and necessary precautions, emphasizing the importance of calibration and proper sampling techniques. The method is applicable for sulfur concentrations ranging from 10 to 1400 ng/g, with specific instructions for sample preparation and analysis.

Uploaded by

allstructure1157
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
35 views8 pages

UOP 988-11 Low Trace Sulfur in LPG and Gaseous Hydrocarbons by Oxidative Combustion With Ultravi

The document outlines UOP Method 988-11 for determining low trace sulfur in liquefied petroleum gas (LPG) and gaseous hydrocarbons using oxidative combustion and ultraviolet fluorescence detection. It details the method's scope, apparatus, reagents, procedure, and necessary precautions, emphasizing the importance of calibration and proper sampling techniques. The method is applicable for sulfur concentrations ranging from 10 to 1400 ng/g, with specific instructions for sample preparation and analysis.

Uploaded by

allstructure1157
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 8

Low Trace Sulfur in LPG and Gaseous Hydrocarbons

by Oxidative Combustion
with Ultraviolet Fluorescence Detection
UOP Method 988-11
Scope
This method is for determining sulfur in liquefied petroleum gas (LPG) and gaseous hydrocarbons
at concentrations ranging from 10 to 1400 ng/g (0.01 to 1.40 mass-ppm). A trap & release procedure
is used to increase the sensitivity of the analysis.
Higher concentrations can be determined by ASTM Method D6667, “Standard Test Method for
Determination of Total Volatile Sulfur in Gaseous Hydrocarbons and Liquefied Petroleum Gases by
Ultraviolet Fluorescence,” or ASTM Method D7551, “Total Volatile Sulfur in Gaseous Hydrocarbons
and Liquefied Petroleum Gases and Natural Gas by Ultraviolet Fluorescence,” by using the
instrument without the trap & release module.
LPG samples are expanded into the gas phase before analysis. If any heavy sulfur compounds are
present in the LPG that do not volatilize quantitatively with the LPG, they may be underreported.
Halogens interfere at concentrations greater than approximately 0.3%. The method can also be
applied to the analysis of hydrogen and other gas samples.

References
ASTM Method D6667, “Total Volatile Sulfur in Gaseous Hydrocarbons and Liquefied Petroleum
Gases by Ultraviolet Fluorescence” www.astm.org
ASTM Method D7551, “Total Volatile Sulfur in Gaseous Hydrocarbons and Liquefied Petroleum
Gases and Natural Gas by Ultraviolet Fluorescence,” www.astm.org
UOP Method 373, “Composition of C2 through C5 Hydrocarbon Mixtures by GC,” www.astm.org
UOP Method 516, “Sampling and Handling of Gasolines, Distillate Fuels and C3-C4 Fractions,”
www.astm.org
UOP Method 539, “Refinery Gas Analysis by Gas Chromatography,” www.astm.org
UOP Method 999, “Precision Statements in UOP Methods,” www.astm.org

Outline of Method
A commercial instrument is set up and calibrated with different volumes of LPG standards. The
IT IS THE USER'S RESPONSIBILITY TO ESTABLISH APPROPRIATE PRECAUTIONARY PRACTICES AND TO
DETERMINE THE APPLICABILITY OF REGULATORY LIMITATIONS PRIOR TO USE. EFFECTIVE HEALTH AND
SAFETY PRACTICES ARE TO BE FOLLOWED WHEN UTILIZING THIS PROCEDURE. FAILURE TO UTILIZE THIS
PROCEDURE IN THE MANNER PRESCRIBED HEREIN CAN BE HAZARDOUS. MATERIAL SAFETY DATA SHEETS
(MSDS) OR EXPERIMENTAL MATERIAL SAFETY DATA SHEETS (EMSDS) FOR ALL OF THE MATERIALS USED IN
THIS PROCEDURE SHOULD BE REVIEWED FOR SELECTION OF THE APPROPRIATE PERSONAL PROTECTION
EQUIPMENT (PPE).
© COPYRIGHT 2011 UOP LLC. All rights reserved.
Nonconfidential UOP Methods are available from ASTM International, 100 Barr Harbor Drive, P.O. Box C700, West
Conshohocken, PA 19428-2959, USA. The UOP Methods may be obtained through the ASTM website, www.astm.org, or by
contacting Customer Service at [email protected], 610.832.9555 FAX, or 610.832.9585 PHONE.
2 of 8

gas or LPG materials are dispensed into fluoropolymer gas sampling bags and, if LPG, are allowed to
evaporate. The atmospheric pressure gas sample is injected into an argon carrier flow using a
syringe-based gas injector. The gas mixture is combined with oxygen at high temperature. The
organic material is converted to carbon dioxide and water. The sulfur in the sample is converted to
sulfur dioxide and preconcentrated using a trap & release unit prior to measurement. The signal is
proportional to the total volatile sulfur in the sample.

Apparatus
References to catalog numbers and suppliers are included as a convenience to the method user.
Other suppliers may be used.
Regulator, argon, single-stage, with stainless steel diaphragm, delivery pressure range 30-700 kPa
(4-100 psi), Matheson Tri-Gas, Cat. No. 3231. This regulator is installed downstream of the
two-stage regulator to provide better flow control.
Regulator, argon, two-stage, with stainless steel diaphragm, delivery pressure range 30-700 kPa (4-
100 psi), Matheson Tri-Gas, Cat. No. 3122-580
Regulator, oxygen, single-stage, with stainless steel diaphragm, delivery pressure range 30-700 kPa
(4-100 psi), Matheson Tri-Gas, Cat. No. 3231. This regulator is installed downstream of the two-
stage regulator to provide better flow control.
Regulator, oxygen, two-stage, with stainless steel diaphragm, delivery pressure range 30-700 kPa
(4-100 psi), Matheson Tri-Gas, Cat. No. 3122-540
Sulfur analyzer, with attached furnace, autosampler, controls and computer, Model TS-100V, with
SD-100 Sulfur Detector, TRU-100 Trap & Release Unit and GI-220 Gas Injector, Mitsubishi
Chemical Analytech, available from COSA Instrument. This method was developed and
validated using the Mitsubishi analyzer. The procedure for analysis may be different for other
instruments and other instruments need to be validated before using for this method. Not all
combustion/UV fluorescence instruments are capable of running this analysis. The Mitsubishi
analyzer must be equipped with the following accessories:
Gas Injector, Mitsubishi GI-220, COSA Instruments
Membrane drier, Perma Pure MD-110-24F-4 or Tube Dryer, Mitsubishi, Cat. No. TN6RPC,
COSA Instruments (see Note 1)
Trap & Release Unit, Mitsubishi TRU-100, COSA Instruments

Reagents and Materials


References to catalog numbers and suppliers are included as a convenience to the method user.
Other suppliers may be used.
The following items are required to perform the analysis. Additional reagents and materials may be
required depending on the specific instrument.
Air, compressed, dry, oil-free, for membrane drier (if instrument does not purge the drier with
argon), local supply
Alumina balls, Mitsubishi, Cat. No. TS3CAT, COSA Instruments
Argon, compressed gas, 99.99% minimum purity. UHP, Matheson Tri-Gas or local supply
Gas sampling bags, 1.6-L or 3.8-L, Cat. No. 10-923-10 or -11, Fisher Scientific, See Note 2

988-11
3 of 8

LPG standard, 1.0 mass-ppm methyl mercaptan (methanethiol) in n-butane, Matheson Tri-Gas
Oxygen, compressed gas, 99.98% minimum purity, UHP, Matheson Tri-Gas or local supply
Quartz wool, Mitsubishi Cat. No. TNQWL, COSA Instruments

Procedure
The analyst is expected to be familiar with general laboratory practices, the technique of sulfur
analysis, and the equipment being used.
Preparation of Standards
The calibration standard of methylmercaptan in butane is purchased as a certified standard from gas
suppliers. A 1.0 mass-ppm methylmercaptan in n-butane standard is 0.667 mass-ppm S. When
expanded to atmospheric pressure in the gas sampling bag, it contains 1.728 ng/mL (as S) at 0°C or
1.583 ng/mL (as S) at 25°C. Calculate the gaseous sulfur concentration in the vaporized standard
from the concentration of the methylmercaptan in n-butane LPG standard using Equation 1.
 273.15 
A = 1.728 C  (1)
 T 
where:
A = gaseous sulfur concentration of the vaporized standard, ng/mL
C = concentration of methylmercaptan in LPG blend, µg methylmercaptan/g of LPG
(mass-ppm)
T = absolute temperature, K
 32.064   58.124 
1.728 =    (1a)
 48.107   22.414 
where:
32.064 = atomic mass of S, g/mol
48.107 = molecular mass of methylmercaptan, g/mol
58.124 = molecular mass of butane, g/mol
22.414 = volume of ideal gas at STP, L/mol
Calculate the mass of sulfur injected from the injection volume and the gaseous sulfur
concentration using Equation 2.
B=AV (2)
where:
A = gaseous sulfur concentration in bag, at specific temperature, ng/mL
B = mass of injected sulfur, ng
V = gaseous volume of expanded calibration blend injected, mL
Sampling
The sample should be collected according to UOP Method 516, “Sampling and Handling of
Gasolines, Distillate Fuels and C3-C4 Fractions.” The use of a passivated cylinder (e.g. Silcosteel™,
Sulfinert™, etc) is required (see Note 3). The transfer of the LPG or gas sample to the gas sampling
bag should be performed in a fume hood. A new bag should be used for each sample to prevent
contamination.
1. Flush the sample cylinder outlet with sample for a few seconds before connecting to the gas
sampling bag.
2. Open the valve on the gas sampling bag.

988-11
4 of 8

3. Open the cylinder valve briefly, allowing sample to flow into the bag. Do not overfill the bag.
• For LPG samples, transfer a small volume of LPG (2 - 4 mL) and allow to vaporize (this is for a 1.6-
L bag; approximately twice that amount for a 3.2-L bag).
• For gas samples, fill until the bag expands, but is not completely full.
4. Close the valve on the gas sampling bag and remove from the cylinder.
5. Analyze the sample within 3 hours of transferring to the gas sampling bag.
Preparation of Apparatus
1. Set up the instrument according to the manufacturer's instructions for the trap and release unit.
Connect the membrane drier in series between the combustion tube and the detector. Allow the
instrument to warm up and the baseline to stabilize before injecting samples. Suggested
Operating Conditions for the Mitsubishi TS-100V analyzer are listed in Table 1. The same trap
time must be used for the calibration and the samples – use the longest trap time if calibrating
for multiple ranges. A 25-mL syringe is installed in the GI-220 gas injector.
Table 1
Operating Conditions for Mitsubishi TS-100V/SD-100/TRU-100
a
Upper temperature 900ºC
b
Lower temperature 1000ºC
Argon main 170 mL/min
Oxygen main 150 mL/min
Argon auxiliary 50 mL/min
Oxygen auxiliary 400 mL/min
GI-220 carrier argon 80 mL/min
Gain Ultra
Normal end Off
Minimum area 40000
Base line 75%
GI-220 Syringe 25 mL
Absorption rate 50 mL/min
Injection rate 20 mL/min
a
Combustion tube, upper portion
b
Combustion tube, lower portion

Table 2
Recommended Sample Size and Instrument Timing
Concentration Sample
Range (ng/g) Volume (mL) Trap Time (s) Timer Start (s) Timer End (s)
10 - 200 80 532 535 835
200 - 500 40 296 310 610
500 - 1400 10 250 270 570

Calibration
Calibrate weekly when in use. Check the calibration daily when in use, by analyzing one of the
calibration standards or a reference material.
1. Transfer the calibration standard to a gas sampling bag, and then connect to the GI-220 gas
injector module.
• The same gas sampling bag can be re-used for the standard. Empty the bag before refilling.

988-11
5 of 8

2. Set up the sample table for multiple sample volumes and replicate injections of the sulfur
standard. Three or four injections are recommended for volumes below 10 mL and two
injections for volumes above 10 mL.
• Use the trap time and timer start/end from Table 2. If calibrating for multiple ranges, use the trap
and integration times for the longest method that will be used for all standards and samples.
• For a low level calibration, use 1-, 2-, 3-, 5-, and 10-mL volumes. For a mid or high level
calibration, use 5-, 10-, 20-, and 30-mL volumes.
• Do not exceed 30 mL of the standard, or the trap may be saturated. See Note 4.
• If run as a “calibration” method, the “START” button will need to be pressed for each new sample
volume. If run as a “sample” method, the instrument will run all measurements on the same
material automatically, and the S content and response areas can be entered in manually. See
Note 5.
• Relative standard deviation (RSD), as calculated by the instrument for the standards, should be
within 15% for the 1- or 2-mL injections and within 10% for 3-mL or larger injections.

3. Analyze the standard according to the instrument manufacturer’s instructions.


4. Create a regression line using the instrument software using the appropriate volumes of the
standards. Set the regression line to “y=bx+c” for this calibration.
Sample Analysis
1. Determine the average molecular weight of the sample, see the Appendix.
2. Attach the gas sampling bag with the first sample to the GI-220 gas injector module.
3. Add the sample to the sample table. If unsure of concentration, start with a small volume such
as 5 mL; otherwise, select the volume based on Table 2.
• The amount of sulfur injected should not exceed 50 ng per measurement.
• Analyze samples above 100 ng/g in duplicate. For samples below 100 ng/g, 3 or more
measurements should be made.
• Enter the sample average molecular weight to calculate the results as mass-ppm.
4. Analyze the sample replicates according to the instrument manufacturer’s instructions.
5. Repeat Steps 1 through 4 for each additional sample.
• If the detector becomes contaminated (trace off scale), disconnect the gas sampling bag and
analyze air blanks until the response stabilizes. Then, confirm that the sensitivity has not changed
by analyzing a calibration standard or a control standard.
• Integration and calculations are done automatically. The average of the replicate injections is
calculated by the instrument software.

Calculations
All calculations are performed by the software, and results are displayed and printed in mass-ppm
(mg/kg) or mass-ppb (ng/g). The molecular weight of the sample is input during sample data entry
and is used by the instrument to convert results from mass/volume to mass/mass.
• Multiple injections of the same sample are averaged by the instrument software.

Report
Report results as mg/kg (mass-ppm) to two decimal places.

988-11
6 of 8

Notes and Precautions


1. The membrane drier is used to remove the water produced during combustion. If not removed,
the water vapor would enter the detector and result in a reduction of the chemiluminescence
intensity. The membrane drier consists of a thin walled Nafion™ tube within a larger plastic
tube. The combustion product gas flows through the Nafion™ tube. Dry air or other dry gas
flows counter-current through the outer plastic tube. The Nafion™ membrane allows water to
pass through, and be carried away by the air stream on the other side.
2. Gas sampling bags are made of Tedlar™ or other fluoropolymers. The fitting on the gas sampling
bags may not match the connector on the GI-220. In that case, an adapter, such as a short section
of silicone tubing, should be used to make the connection. Samples should be analyzed within 3
hours of transfer into the gas sampling bag.
3. The use of passivated cylinders (e.g. Silcosteel, Sulfinert, etc) is required to prevent the loss of S
compounds to the cylinder wall. The internal components of the cylinder valves should be
similarly treated.
4. Do not exceed 50 ng of S for the TRU-100 trap and release unit. If a single measurement
contains more than 50 ng S, it may be necessary to bake out the trap and release unit to remove
excess sulfur. Follow the manufacturer’s instructions if the bake out is needed.
5. When running the calibration as “samples,” type in the calibration data manually to create a
calibration. Run the sample report to print the individual measurements. Calculate the sulfur
content for each injection volume using the equations in “Preparation of Standards.” Under the
“System” menu, select “Default Calibration Curve.” Click “Edit” and enter all of the calibration
data points. Save the calibration.

Precision
Precision statements were determined using UOP Method 999, “Precision Statements in UOP
Methods.” As described in the Procedure and Calculations, two to four replicate injections were
averaged for each analysis.
Repeatability and Site Precision
A nested design was carried out for determining impurities in three LPG samples and one gas
sample by two analysts on two separate days, performing two analyses each day for a total of 32
analyses. Using a stepwise analysis of variance procedure, the within-day estimated standard
deviations (esd) were calculated at the concentration means listed in Table 3. Two analyses
performed in one laboratory by the same analyst on the same day should not differ by more than the
repeatability allowable differences shown in Table 3 with 95% confidence. Two analyses performed
in one laboratory by different analysts on different days should not differ by more than the site
precision allowable differences shown in Table 3 with 95% confidence.
Table 3
Repeatability and Site Precision for Procedure, mg/kg
Repeatability Site Precision
Within- Allowable Within- Allowable
Sample Mean Day esd Difference Lab esd Difference
LPG 1 0.68 0.013 0.05 0.015 0.06
LPG 2 0.26 0.002 0.01 0.002 0.01
LPG 3 0.01 0.001 0.01 0.001 0.01
Gas 0.06 0.001 0.01 0.001 0.01

988-11
7 of 8

The data in Table 3 represent short-term estimates of the repeatability of the method. When the test
is run routinely, use of a control standard and a control chart is recommended to generate an estimate
of long-term repeatability.
Reproducibility
There is insufficient data to calculate the reproducibility of the test at this time.

Time for Analysis


The elapsed time for the calibration is approximately 4 hours. When performed as in “calibration”
mode, then the labor requirement for calibration is identical to the elapsed time, 4 hours. In “sample”
mode, the labor requirement is approximately 1 hour. Each sample analysis requires 0.5 hour for
duplicate measurement.

Suggested Suppliers
COSA Instrument Corp., 55 Oak St., Norwood, NJ 07648, USA (201-767-6600), distributor for
Mitsubishi Chemical Analytech, 370 Enzo, Chigasaki, Kanagawa Pref., 253-0084, Japan (+81-
467-86-3864) www.cosainstrument.com
Fisher Scientific, 711 Forbes Ave., Pittsburgh, PA 15219, USA (412-562-8300) www.fishersci.com
Matheson Tri-Gas, 166 Keystone Dr., Montgomeryville, PA 18936, USA (215-641-2700)
www.mathesontrigas.com
Perma Pure Inc., 8 Executive Dr., Toms River, NJ 08754, USA (732-244-0010)
www.permapure.com

988-11
8 of 8

Appendix
Calculation of the Average Molecular Weight of a Gas
Scope
This appendix may be used to calculate the average molecular weight of a gaseous material from its
known composition.

References
ASTM Method D2163, Hydrocarbons in Liquefied Petroleum (LP) Gases and Propane/Propene
Mixtures by Gas Chromatography,” www.astm.org
UOP Method 373, “Composition of C2 through C5 Hydrocarbon Mixtures by GC,” www.astm.org
UOP Method 539, “Refinery Gas Analysis by Gas Chromatography,” www.astm.org

Calculations
If the composition of the material is not known, analyze it using the following test methods. Other
similar methods may be used.
For gas samples: UOP Method 539, “Refinery Gas Analysis by Gas Chromatography”
For LPG samples: UOP Method 373, “Composition of C2 through C5 Hydrocarbon Mixtures by
GC,” or ASTM Method D2163, Hydrocarbons in Liquefied Petroleum (LP) Gases and
Propane/Propene Mixtures by Gas Chromatography”
Calculate the average molecular weight of the gas sample (after expansion if the sample was an
LPG), from its known composition, using Equation A1.
n
Pi Wi
M= ∑
i =1 100
(A1)

where:
M= average molecular weight of the gas
Pi = concentration of each component in the sample, vol-%
Wi = molecular weight of that component, Table A1
100 = correction for %
Table A1
Molecular Weights of Gas Sample Components
Molecular
Component Weight
Hydrogen 2.02
Nitrogen 28.01
Methane 16.04
Ethane 30.07
Ethylene 28.05
Propane 44.10
Propylene 42.08
C4 Paraffins 58.12
C4 Olefins 56.11

988-11

You might also like