=netACUSON S3000

This document provides detailed instructions for replacing the RM Module in ACUSON S1000, S2000, and S3000 Ultrasound Systems, including steps for removing and installing the Hard Disk Drive (HDD). It outlines the required materials, tools, and safety precautions, as well as specific BIOS and software requirements for different RM Module versions. The process is expected to take approximately one hour and emphasizes the importance of performing leakage tests after replacing electrical components.

Uploaded by

jojotiamo0310
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
67 views33 pages

=netACUSON S3000

This document provides detailed instructions for replacing the RM Module in ACUSON S1000, S2000, and S3000 Ultrasound Systems, including steps for removing and installing the Hard Disk Drive (HDD). It outlines the required materials, tools, and safety precautions, as well as specific BIOS and software requirements for different RM Module versions. The process is expected to take approximately one hour and emphasizes the importance of performing leakage tests after replacing electrical components.

Uploaded by

jojotiamo0310
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 33

21.08.

24, 11:47 ACUSON S3000

ACUSON S Family
System
Replacement Instructions
Parts Replacement

This document is valid for:

ACUSON S1000 Ultrasound Systems

ACUSON S2000 Ultrasound Systems

ACUSON S3000 Ultrasound Systems

US22-101.841.01.19.02 | 04.22

12 Replacing RM Module
12.1 Overview
The replacement RM Module does not include the Hard Disk Drive (HDD). The HDD must be removed from
the existing RM Module and installed into the replacement RM Module.

12.1.1 Workflow

01 Remove the Upper Covers


02 Remove the E Module Cover
03 Remove the RM Module

04 Remove the HDD


05 Install the HDD in the new RM Module
06 Install the RM Module
07 Install the E Module Cover
08 Install the Upper Covers
09 Verification

12.1.2 Required Material


RM Module (Refer to the Service Parts Catalog)
Tie-Wraps

localhost:3006/#/documents/US22-101.841.01.XX.02.json?fragment=%2FUS22-101.841.01.XX.02.d3ff9f51462ebbea0a53dbda7dfa389b.html&s… 1/33
21.08.24, 11:47 ACUSON S3000

12.1.3 Required Tools


Standard CSE tool kit
Service Key

12.1.4 Required Time


Approximately one (1) hour.

12.1.5 System Preparation

01 Press the Power button on the Control Panel and follow the instructions on the screen to shut down
the system.
02 Click Shutdown to shut down the system.
03 Switch the MAIN circuit breaker to the Off position.
04 Disconnect the system power cord from its external power source.
05 Maintain a minimum of 3.5 feet (1 meter) clearance while working on the system.

CAUTION
Leakage tests must be performed after replacing electrical components connected to non-
isolated AC power.

Failure to complete a leakage test may result in damage to the system.

i There are several RM Module versions with different BIOS settings. Replacing RM
Modules will require booting to a new version of software. Refer to Table below.

Minimum
Material Number Name BIOS Version Software
Requirement

10438526 RM300, CPU Module, S2000/VI 5893 or 5893V2 VB10D

10852163 RM300 Module w/o HDD, BIOS 5936 VC20A


5936

10855525 RM300 Mod w/o 5936 VC25C


HD/BIOS5936/GT640/403

10854302 RM300 Module w/o HDD, BIOS 5936 VC25C


5936, GT640

10854750 RM301, CPU Module, w/o HDD, F1.13 VC25D

localhost:3006/#/documents/US22-101.841.01.XX.02.json?fragment=%2FUS22-101.841.01.XX.02.d3ff9f51462ebbea0a53dbda7dfa389b.html&s… 2/33
21.08.24, 11:47 ACUSON S3000

Minimum
Material Number Name BIOS Version Software
Requirement

S Family

Tab. 1: RM300 Module Hardware/Software Matrix

12.2 RM Module and Hard Disk Drive (HDD)


Instructions
12.2.1 Remove the Upper Covers

Remove the covers in the order they are listed in this procedure.

01 Remove the transducers from the system.

02 Remove the upper, left-front cover. Reference 9.2.1.2 Remove the Upper, Left-Front Cover for
instructions.
03 Remove the upper, right-front cover. Reference 9.2.1.3 Remove the Upper, Right-Front Cover for
instructions.
04 Remove the system air filter. Reference 9.2.1.1 Remove the System Air Filter for instructions.
05 Remove the storage bin cover. Reference 9.2.1.4 Remove the Storage Bin Cover for instructions.
06 Remove any OEM cables connected to the system and feed them through the left-rear cover.

07 Remove the upper, right-rear cover. Reference 9.2.1.5 Remove the Upper, Right-Rear Cover for
instructions.
08 Remove the upper, left-rear cover. Reference 9.2.1.6 Remove the Upper, Left-Rear Cover for
instructions.
09 Remove the E-Module OEM Support Bracket.

CAUTION
The system contains components that are sensitive to damage by electrostatic discharge (ESD).

Observe all ESD protective procedures and use ESD protective equipment including an ESD
wrist strap when the protective covers are removed from the system.

12.2.2 Remove the E Module Cover


There are two (2) different E Module Covers:

The older version has a Cable Tray and Captive screws.


The newer version has Cable Clips and non-captive screws.

localhost:3006/#/documents/US22-101.841.01.XX.02.json?fragment=%2FUS22-101.841.01.XX.02.d3ff9f51462ebbea0a53dbda7dfa389b.html&s… 3/33
21.08.24, 11:47 ACUSON S3000

Fig. 177: E Module Cover Versions

01 Disconnect the RM power cable.

02 Loosen the two (2) screws that secure the OEM AC power cable connector to E Module cover.

Fig. 178: Older E Module Cover -- OEM AC Cable Connector and RM Power Cable

Fig. 179: Newer E Module Cover -- OEM AC Cable Connector and RM Power Cable

localhost:3006/#/documents/US22-101.841.01.XX.02.json?fragment=%2FUS22-101.841.01.XX.02.d3ff9f51462ebbea0a53dbda7dfa389b.html&s… 4/33
21.08.24, 11:47 ACUSON S3000

03 Depending upon the E Module cover, remove the cable from the cable clips, or remove the cable tray
cover and then remove the cable.

i Newer E Module covers will have cable clips instead of a cable tray.

04 Depending upon the E Module cover, loosen the four (4) captive screws or remove the four (4) screws.

i Newer E Module covers have non-captive screws.

Fig. 180: E Module Cover -- Screws

05 Loosen the Cable Chute screw. This will make it easier to remove the E-Module cover.

Fig. 181: Cable Chute Screw

06 Carefully pull the E Module cover from the E Module.

i The E Module cover fits tightly on the card cage and requires force to remove.

When removing the E Module cover, reposition any shielding strips that pulled way

i from the E Module.


The older E Module has three (3) shielding strips. The newer E Module has five (5)
shielding strips.

localhost:3006/#/documents/US22-101.841.01.XX.02.json?fragment=%2FUS22-101.841.01.XX.02.d3ff9f51462ebbea0a53dbda7dfa389b.html&s… 5/33
21.08.24, 11:47 ACUSON S3000

Fig. 182: Older E Module with Three (3) Shielding Strips

Fig. 183: Newer E Module with Five (5) Shielding Strips

12.2.3 Remove the RM Module

01 If attached, cut and remove the cable bundle tie-wrap. The HDD power cable should not be bundled
with the other cables. This can cause the HDD power cable to disconnect and damage the cable.

Fig. 184: HDD Cables Bundled with Tie-Wrap

02 Open the upper and lower ejector tabs of the VI board.

localhost:3006/#/documents/US22-101.841.01.XX.02.json?fragment=%2FUS22-101.841.01.XX.02.d3ff9f51462ebbea0a53dbda7dfa389b.html&s… 6/33
21.08.24, 11:47 ACUSON S3000

03 Pull the VI board approximately one-inch out from the card cage to gain access to the VI Board
connectors.
04 Carefully disconnect the VI board connections (P is the RM cable connector, J is the VI board
connector):
P10 - J10 (RM200 Only)
P11 - J11
P13 - J13
P16 - J16
P12 - J12
P17-- J17
P18 – J18
P21 -- J21
RS232 Ribbon Cable (RM 200 Only)
DVI Video Cable

05 Carefully disconnect the following RM connections:


Audio cables, labeled: Line In (Blue); Line Out (Green); Mic (Pink)
Two USB cables: USB 0, USB 1
Assy, PCI-e Cable (J7)
DVI Video Cable

i The Line In (Blue) audio cable is not used with the RM 301 Module.

06 Remove the Ethernet Network cable with Isolator between RM and RJ45 Coupler.

Fig. 185: Ethernet Cable with Isolator

localhost:3006/#/documents/US22-101.841.01.XX.02.json?fragment=%2FUS22-101.841.01.XX.02.d3ff9f51462ebbea0a53dbda7dfa389b.html&s… 7/33
21.08.24, 11:47 ACUSON S3000

i The RM300 requires two (2) Ethernet Cables with the Isolator. The RM301 only uses
one (1) Ethernet Cable and does not require the Isolator.

07 Remove the RJ45 Coupler from the Chassis.

Fig. 186: RJ45 Coupler

08 Remove the upper and lower RM Module bracket screws.

Fig. 187: RM200 Module Bracket Screws

09 Remove the RM Module.

12.2.4 Remove the Hard Disk Drive

01 Disconnect the HDD SATA and Power cables.

02 Remove the HDD module.


a) Loosen the two (2) captive screws or remove the two (2) screws on the HDD mounting bracket.

i Hard drive brackets may use captive or non-captive screws.

localhost:3006/#/documents/US22-101.841.01.XX.02.json?fragment=%2FUS22-101.841.01.XX.02.d3ff9f51462ebbea0a53dbda7dfa389b.html&s… 8/33
21.08.24, 11:47 ACUSON S3000

Fig. 188: HDD Bracket Screws

Fig. 189: HDD SATA and Power Connections

i The non-captive screws are small. Use care when removing them to ensure that
they do not fall into the RM module.

b) Pull the module up and out to disengage it from the bracket.

12.2.5 Install the Hard Disk Drive

CAUTION
The system contains components that are sensitive to damage by electrostatic discharge (ESD).

Observe all ESD protective procedures and use ESD protective equipment including an ESD
wrist strap when the protective covers are removed from the system.

01 Install HDD in its mounting bracket on the new RM Module.


02 Attach the HDD module with bracket to the RM Module.
a) Slide the HDD assembly onto its mounting location in the RM Module, then fasten the screws.
b) Attach the HDD SATA and Power cables to the HDD.

12.2.6 Install the RM Module

i The RM Module replacement requires FPD Optimization. Refer to the appropriate


procedure.

localhost:3006/#/documents/US22-101.841.01.XX.02.json?fragment=%2FUS22-101.841.01.XX.02.d3ff9f51462ebbea0a53dbda7dfa389b.html&s… 9/33
21.08.24, 11:47 ACUSON S3000

RM Module Replacement requires that the field electrical safety test be performed. Refer

i to General Field Electrical Safety Test Procedure / US00-004.860.01 or Field Electrical


Safety Test Procedure / US00-003.860.01 for the procedure, and to Electrical Safety
Limits / US22-101.833.01 for limits.

01 Slide the RM Module into the RM slot in the card cage.

02 Attach the upper and lower RM Module bracket screws.

Fig. 190: RM Module Screws

03 There are three (3) different VI boards that interconnect to the various RM Modules. The table and
figures display the various configurations and wiring connections.

VI Board Material Number RM Module Notes

10039639 RM200 VI includes daughter board

10435338 RM200

10438694 RM300

10438694 RM301

Tab. 2: VI Board to RM Module Configuration

localhost:3006/#/documents/US22-101.841.01.XX.02.json?fragment=%2FUS22-101.841.01.XX.02.d3ff9f51462ebbea0a53dbda7dfa389b.html&… 10/33
21.08.24, 11:47 ACUSON S3000

Fig. 191: VI Board Material Number 10039639 to RM200 Module

Fig. 192: VI Board Material Number 10435338 to RM200 Module

localhost:3006/#/documents/US22-101.841.01.XX.02.json?fragment=%2FUS22-101.841.01.XX.02.d3ff9f51462ebbea0a53dbda7dfa389b.html&… 11/33
21.08.24, 11:47 ACUSON S3000

Fig. 193: VI Board Material Number 10438694 to RM300 Module

Fig. 194: VI Board Material Number 10438694 to RM301 Module

04 Connect the RM Module connectors to the VI board. Ensure that the correct cable is attached to the
proper location. (J is the VI board connector and P is the RM cable). The following tables display the VI
to RM connections.

VI Board Connector RM Cable

J7 PCIe Cable

localhost:3006/#/documents/US22-101.841.01.XX.02.json?fragment=%2FUS22-101.841.01.XX.02.d3ff9f51462ebbea0a53dbda7dfa389b.html&… 12/33
21.08.24, 11:47 ACUSON S3000

VI Board Connector RM Cable

J8 DVI Video Cable

J9 RS232 Ribbon Cable

J10 P10

J11 P11

J12 P12-2 from P12

J13 P13

J14 P14 from J3 Daughter Board

J15 RS232 Ribbon Cable

J16 P16-1 from J2 Daughter Board

J17 P17

J18 P18

J21 P21

J1 Daughter Board P16-2 from P16

J3 Daughter Board P12-1 from P12

Tab. 3: VI Board (10039639) Connector to RM 200 Module

VI Board Connector RM Cable

J7 PCIe Cable

J8 DVI Video Cable

J10 P10

J11 P11

J12 P12

J13 P13

localhost:3006/#/documents/US22-101.841.01.XX.02.json?fragment=%2FUS22-101.841.01.XX.02.d3ff9f51462ebbea0a53dbda7dfa389b.html&… 13/33
21.08.24, 11:47 ACUSON S3000

VI Board Connector RM Cable

J15 RS232 Ribbon Cable

J16 P16

J17 P17

J18 P18

J21 P21

Tab. 4: VI Board (10435338) Connector to RM 200 Module

VI Board Connector RM Cable

J7 PCIe Cable

J8 DVI Video Cable

J11 P11

J12 P12

J13 P13

J16 P16

J17 P17

J18 P18

J21 P21

Tab. 5: VI Board (10438694) Connector to RM 300 Module

VI Board Connector RM Cable

J7 PCIe Cable

J8 DVI Video Cable

J11 P11

J12 P12

localhost:3006/#/documents/US22-101.841.01.XX.02.json?fragment=%2FUS22-101.841.01.XX.02.d3ff9f51462ebbea0a53dbda7dfa389b.html&… 14/33
21.08.24, 11:47 ACUSON S3000

VI Board Connector RM Cable

J13 P13

J16 P16

J17 P17

J18 P18

J21 P21

Tab. 6: VI Board (10438694) Connector to RM 301 Module

Fig. 195: VI Board Connections

05 Connect the following Audio and USB cables between the VI board and RM module:

VI Board Connector RM200 Module Connector

Audio J23 – Line Out (Green) Line Out (Green)

Audio J24 – Line In (Blue) Line In (Blue)

Audio J25 – MIC (Pink) MIC (Pink)

USB0 (Upper) USB0 – Right Side (next to RJ45)

USB1 (Lower) USB1 – Left Side (next to board edge)

localhost:3006/#/documents/US22-101.841.01.XX.02.json?fragment=%2FUS22-101.841.01.XX.02.d3ff9f51462ebbea0a53dbda7dfa389b.html&… 15/33
21.08.24, 11:47 ACUSON S3000

Tab. 7: VI Board 10039639 to RM200 Module

i VI Board 10039639 - USB0 and USB1 cables are reversed from all other VI Boards.

VI Board Connector RM200 Module Connector

Audio J23 – Line Out (Green) Line Out (Green)

Audio J24 – Line In (Blue) Line In (Blue)

Audio J25 – MIC (Pink) MIC (Pink)

USB1 (Upper) USB1 – Right Side (next to RJ45)

USB0 (Lower) USB0 – Left Side (next to board edge)

Tab. 8: VI Board 10435338 to RM200 Module

VI Board Connector RM300 Module Connector

Audio J23 – Line Out (Green) Line Out (Green)

Audio J24 – Line In (Blue) Line In (Blue)

Audio J25 – MIC (Pink) MIC (Pink)

USB1 (Upper) USB1 – Right Side (next to RJ45)

USB0 (Lower) USB0 – Left Side (next to board edge)

Tab. 9: VI Board 10438694 to RM300 Module

VI Board Connector RM301 Module Connector

Audio J23 – Line Out (Green) Line Out (Green)

J25 - MIC (Pink) MIC (Pink)

USB1 (Upper) USB1 – Right Side (next to RJ45)

USB0 (Lower) USB0 – Left Side (next to board edge)

Tab. 10: VI Board 10438694 to RM301 Module

localhost:3006/#/documents/US22-101.841.01.XX.02.json?fragment=%2FUS22-101.841.01.XX.02.d3ff9f51462ebbea0a53dbda7dfa389b.html&… 16/33
21.08.24, 11:47 ACUSON S3000

The USB cable routing is always:

i VI Board USB (upper) – RM Module USB Right connector (next to RJ45)


VI Board USB (lower) – RM Module USB Left connector (next to board edge)

On the RM301, there are two identical USB and RJ45 connector blocks. Only use the top connector
block.

Fig. 196: RM Module USB, RJ45 Network Jack, and Audio Cable Connections

Fig. 197: RM200 Cables and Connections

localhost:3006/#/documents/US22-101.841.01.XX.02.json?fragment=%2FUS22-101.841.01.XX.02.d3ff9f51462ebbea0a53dbda7dfa389b.html&… 17/33
21.08.24, 11:47 ACUSON S3000

Fig. 198: RM300 Cables and Connections

Fig. 199: RM301 Cables and Connections

06 After connections are complete, slide the VI board the rest of the way into the card cage.
07 Engage the tabs on the VI board to secure it.

08 Install the RJ45 Coupler into the chassis.

localhost:3006/#/documents/US22-101.841.01.XX.02.json?fragment=%2FUS22-101.841.01.XX.02.d3ff9f51462ebbea0a53dbda7dfa389b.html&… 18/33
21.08.24, 11:47 ACUSON S3000

Fig. 200: RJ45 Coupler

09 Connect the Ethernet Cable(s) and Isolator (if applicable):


For the RM200 and RM301: Connect the Ethernet cable between the RM Module RJ45 Network Jack
and the RJ45 Coupler.
For the RM300: Connect two Ethernet cables with an Isolator between the RM Module RJ45
Ensure that the Isolator is installed in the correct direction.

i On the RM301, there are two (2) identical USB and RJ45 connector blocks. Only use
the top connector block.

Fig. 201: RM Module USB, RJ45 Network Jack, and Audio Cable Connections

Fig. 202: Isolator Markings - Network Connection and RM300 Connection

i The RM300 requires two (2) Ethernet Cables with the Isolator. The RM200 and
RM301 only uses one (1) Ethernet Cable and does not require an Isolator.

localhost:3006/#/documents/US22-101.841.01.XX.02.json?fragment=%2FUS22-101.841.01.XX.02.d3ff9f51462ebbea0a53dbda7dfa389b.html&… 19/33
21.08.24, 11:47 ACUSON S3000

Fig. 203: RM300 Ethernet Cable with Isolator Properly Connected

Fig. 204: RM301 Ethernet Cable without Isolator

12.2.7 Install the E Module Cover


There are two (2) different E Module Covers.

The older version has a Cable Tray and Captive screws.


The newer version has Cable Clips and non-captive screws.

Fig. 205: E Module Cover Versions

These E Module covers require a different number of E Module shielding strips.

localhost:3006/#/documents/US22-101.841.01.XX.02.json?fragment=%2FUS22-101.841.01.XX.02.d3ff9f51462ebbea0a53dbda7dfa389b.html&… 20/33
21.08.24, 11:47 ACUSON S3000

The older version requires three (3) shielding strips.


The newer version requires five (5) shielding strips.

01 Install the E Module cover on the E Module.

Fig. 206: Older E Module with Three (3) Shielding Strips

Fig. 207: Newer E Module with Five (5) Shielding Strips

02 Tighten the four (4) screws that secure the E Module cover to the system.
03 Route the OEM AC power cable through the cable tray and attach the cable tray cover, or use the cable
clips.
04 Connect the RM power cable to the connector in the E Module cover.

localhost:3006/#/documents/US22-101.841.01.XX.02.json?fragment=%2FUS22-101.841.01.XX.02.d3ff9f51462ebbea0a53dbda7dfa389b.html&… 21/33
21.08.24, 11:47 ACUSON S3000

Fig. 208: Older E Module Cover -- OEM AC Cable Connector and RM Power Cable

Fig. 209: Newer E Module Cover -- OEM AC Cable Connector and RM Power Cable

05 Tighten the Cable Chute screw.

Fig. 210: Cable Chute Screw

12.2.8 Perform Software Installation

i This is only required if the software does not recognize the newer RM Module. Refer to
Table below.

localhost:3006/#/documents/US22-101.841.01.XX.02.json?fragment=%2FUS22-101.841.01.XX.02.d3ff9f51462ebbea0a53dbda7dfa389b.html&… 22/33
21.08.24, 11:47 ACUSON S3000

01 Insert the software disk in the CD/DVD Disk Drive. (Refer to the applicable Software Installation
procedure).

Minimum
Material Number Name BIOS Version Software
Requirement

10438526 RM300, CPU Module, 5893 or 5893V2 VB10D


S2000/VI

10852163 RM300 Module w/o HDD, 5936 VC20A


BIOS 5936

10855525 RM300 Mod w/o 5936 VC25C


HD/BIOS5936/GT640/403

10854302 RM300 Module w/o HDD, 5936 VC25C


BIOS 5936, GT640

10854750 RM301, CPU Module, w/o F1.13 VC25D


HDD, S Family

Tab. 11: RM300 Module Hardware/Software Matrix

12.2.9 Install the Upper Covers

The covers must be installed in the order they are listed in this procedure.

12.2.9.1 Install the E-Module OEM Support Bracket

01 Reinstall the support bracket if removed.

02 Tighten with four (4) captive screws.

Fig. 211: E-Module OEM Support Bracket

localhost:3006/#/documents/US22-101.841.01.XX.02.json?fragment=%2FUS22-101.841.01.XX.02.d3ff9f51462ebbea0a53dbda7dfa389b.html&… 23/33
21.08.24, 11:47 ACUSON S3000

12.2.9.2 Install the Upper, Left-Rear Cover

01 Align the cover on the system.

Fig. 212: Upper Left Rear Cover - Screw Shown

02 Tighten the two (2) captive fasteners.

12.2.9.3 Install the Upper, Right-Rear Cover

01 Connect the existing battery cable between the Upper, Rear Panel and J38 connector located below
the E Module.

Fig. 213: Battery Cable

02 Align the cover to the system.


03 Tighten the three (3) 1/4-turn fasteners.

localhost:3006/#/documents/US22-101.841.01.XX.02.json?fragment=%2FUS22-101.841.01.XX.02.d3ff9f51462ebbea0a53dbda7dfa389b.html&… 24/33
21.08.24, 11:47 ACUSON S3000

Fig. 214: Upper Right Rear Cover Shown

04 Place the battery back into the system battery compartment.


05 Reconnect the OEM cables.
Feed the OEM cables that were previously disconnected back through the opening in the system
platform.
Reconnect the cables according to the labels that were attached to the cables before disconnection.

Fig. 215: OEM Cables

12.2.9.4 Install the Storage Bin Cover

01 Position the storage bin cover on the system.

localhost:3006/#/documents/US22-101.841.01.XX.02.json?fragment=%2FUS22-101.841.01.XX.02.d3ff9f51462ebbea0a53dbda7dfa389b.html&… 25/33
21.08.24, 11:47 ACUSON S3000

Fig. 216: Storage Bin Cover

12.2.9.5 Install the Upper, Right-Front Cover

01 Hold the cover at the transducer ports recess and position the cover on the system.
02 Press the cover to engage the ball snaps.

Fig. 217: Upper Right Front Cover

03 Tighten the 1/4-turn fasteners.

12.2.9.6 Install the Physio Cover

01 Install either the Physio Cover Blank, or Physio Module Cover on the Upper, Left-Front Cover.

i This procedure is only required if a new Upper, Left cover is replaced and requires a
Physio Cover Blank or Physio Module Cover.

localhost:3006/#/documents/US22-101.841.01.XX.02.json?fragment=%2FUS22-101.841.01.XX.02.d3ff9f51462ebbea0a53dbda7dfa389b.html&… 26/33
21.08.24, 11:47 ACUSON S3000

Fig. 218: Upper, Left Cover - Physio Cover Location

12.2.9.7 Install the Upper, Left-Front Cover

01 Hold the cover at the DVD and physio recess and position the cover on the system.
02 Press the cover to engage the ball snaps.

03 Tighten the 1/4-turn fastener.

Fig. 219: Left Front Cover Shown

12.2.9.8 Install the System Air Filter

01 Slide the air filter into the air filter compartment on the system.

localhost:3006/#/documents/US22-101.841.01.XX.02.json?fragment=%2FUS22-101.841.01.XX.02.d3ff9f51462ebbea0a53dbda7dfa389b.html&… 27/33
21.08.24, 11:47 ACUSON S3000

Fig. 220: Air Filter Shown

12.2.10 Test the RM Module


If errors occur during the following steps, use the Boot Sequence table below to identify which modules
are responsible. If necessary, contact the local Customer Care Center for help.

01 Switch on standby power. Make sure the system is connected to an external power source, then turn
on the MAIN power switch on the back of the AC power tray.
02 Power up the system. Press the green STANDBY button on the Control Panel. Verify that the following
occurs within the first four (4) seconds of bootup:
The LEDs on the AC power tray illuminate.
The standby button changes to solid green.
The fans switch on.
The control panel emits a click, indicating that the power-on self-test (POST) is complete.

03 Verify the video initialization.


04 Verify that the BIOS loads.
About 17 seconds after the system has been powered on, the BIOS initiates a POST and the “Siemens”
screen displays. After about 30 seconds, the system clicks.
05 Verify that Windows operating system loads. The Windows operating system screen displays as the
operating system loads. This takes about one (1) minute. Note any error messages that appear during
this time.
06 Verify that syngo loads. syngo takes about two (2) minutes to load. During this time the system
displays the startup window, which displays a number count.
The system emits three (3) “clicks” when the syngo diagnostics become available.
The syngo user interface displays when loading is complete.

07 Verify that the system properly functions by selecting various buttons on the control panel and
viewing expected changes on the system monitor.

The following table summarizes the boot sequence for the system.

localhost:3006/#/documents/US22-101.841.01.XX.02.json?fragment=%2FUS22-101.841.01.XX.02.d3ff9f51462ebbea0a53dbda7dfa389b.html&… 28/33
21.08.24, 11:47 ACUSON S3000

Interval Possible Cause(s) of


Status Indication(s)
(sec) Failure

NA Standby mode AC power tray. See AC


(user initiated) The Standby and monitor LEDs Tray (PSAC) LED Faults /
illuminate amber. US22-101.840.01.
PS cooling module switches
on.

~0 to 17 Powering up (user AC power tray, E module,


initiated) LEDs on the AC power supply or cable.
illuminate.
Cooling module switches on.

Control panel Control panel emits a click. Control panel.


executing a
power-on self-test
(POST)

~17 to 47 BIOS starts “Siemens” screen displays E module. See Boot


executing POST Sequence / US22-
101.840.01.

BIOS finishes
executing POST2 System emits a click.

~47 to Windows Windows screen displays. Various; check error


107 operating system messages and call the
loading Customer Care Center.

~107 to syngo diagnostics System emits three (3) “clicks”. Various; check error
227 available. messages and call
Customer Care Center.
syngo and Syngo and imaging screens
imaging displays.
application
loading..

Tab. 12: Boot Sequence Summary

12.2.11 Perform the Electrical Safety Test

01 Perform the Electrical Safety Procedure. Refer to Refer to Field Electrical Safety Test Procedure / US00-
003.860.01 or General Field Electrical Safety Test Procedure / US00-004.860.01 for the Electrical
Safety Procedure.
For submission of safety reports refer to Electrical Safety Limits, US22-101.833.01 and Electrical
Safety Test Report, US22-101.834.01.

localhost:3006/#/documents/US22-101.841.01.XX.02.json?fragment=%2FUS22-101.841.01.XX.02.d3ff9f51462ebbea0a53dbda7dfa389b.html&… 29/33
21.08.24, 11:47 ACUSON S3000

Siemens Healthineers © 2024 | v3.5.1 | Disclaimer and Copyright | Corporate Information | Privacy Policy
| Cookie Policy | Terms of Use

localhost:3006/#/documents/US22-101.841.01.XX.02.json?fragment=%2FUS22-101.841.01.XX.02.d3ff9f51462ebbea0a53dbda7dfa389b.html&… 30/33
21.08.24, 11:47 ACUSON S3000

localhost:3006/#/documents/US22-101.841.01.XX.02.json?fragment=%2FUS22-101.841.01.XX.02.d3ff9f51462ebbea0a53dbda7dfa389b.html&… 31/33
21.08.24, 11:47 ACUSON S3000

localhost:3006/#/documents/US22-101.841.01.XX.02.json?fragment=%2FUS22-101.841.01.XX.02.d3ff9f51462ebbea0a53dbda7dfa389b.html&… 32/33
21.08.24, 11:47 ACUSON S3000

localhost:3006/#/documents/US22-101.841.01.XX.02.json?fragment=%2FUS22-101.841.01.XX.02.d3ff9f51462ebbea0a53dbda7dfa389b.html&… 33/33

You might also like