Unit 4_Microwave Measurements
Unit 4_Microwave Measurements
Microwave Measurements
Learning Outcome 1
Measure the characteristics of Microwave
LO2
devices
= ( R + jL)(G + jC ) = + j
( R + jL)
Z0 =
(G + jC )
Characteristics of transmission line
A: Terminated in Zo
Zs
Vs
Zo Zo r = Zo - Zo = 0
Zo + Zo
B: Short Circuit
Zs
- Zo
Vs
Zo r= 0 = -1
0- + Zo
C: Open Circuit
Zs - Zo
Vs R= =1
+ Zo
The Reflection and Transmission
Losses
• When the resistive load termination is not equal to the
characteristic impedance, part of the power is reflected back and
the remainder is absorbed by the load
• . The amount of voltage reflected back is called voltage reflection
coefficient.
or alternatively:
• Zmin = Zo/VSWR
• Relationship between VSWR and Reflection Coefficient:
VSWR = (1 + | Γ |)/(1 - | Γ |)
Γ = (VSWR – 1)/(VSWR + 1)
TYPES OF MEASUREMENT
EQUIPMENTS
TYPES OF MEASUREMENT
➢ Wavemeter s (absorption, transmission or reaction).
➢ Slotted lines.
FREQUENCY-DOMAIN
➢ Spectrum analyzer, frequency sweepers and frequency counters.
➢ Sampling oscilloscope.
DISPLAY OF TIME-DOMAIN
➢ Oscilloscope.
MICROWAVE
SOURCE
POWER VSWR
METER INDICATOR
ISOLATOR
• When reflection occured, the incident and the reflected waves will
reinforce each other in some places, and in others they will tend to
cancel each other out.
DOUBLE MINIMUM METHOD
MEASUREMENT ( VSWR > 10)
• ‘Double Minimum’ method is usually employed
for VSWR values greater than about 10.
E2MAX d
E2MIN λ/2
d/2
distance along the
line
• The detector output (proportional to field strength squared) is
plotted against position. The probe is moved aling the line to
find the minimum value of signal.
DC VOLTAGE
R1
DETECTORS
• With no power to the detector that contains the sensor element, the sensor-line R1 is
adjusted to zero reading through the meter M1 and the bridge circuit is balanced.
• When signal is applied to the sensor element, causing its temperature to change, the
sensor resistance changes, causing the bridge to become unbalanced.
• Resistor R1 is adjusted to balance meter M1. The change in the reading of meter M2
in the sensor element leg is a direct measure of the microwave power.
THERMOCOUPLES
• Are used as power monitors in the low-to-medium power regions and
are very sensitve.
• Is a thin wire made of two disimilar metals. Hence there will be two
junctions (hot & cold).
• When the temperature at two junctions are different, a voltage is
developed across the thermocouple (i.e across both junctions). This
developed voltage is proportional to the difference between the two
junction temperatures.
• When the temperature at both junctions are the same, the difference
in voltage = 0.
Thermocouple
MICROWAVE CRYSTALS
• Are non-linear detectors that provide current in proportion
to the power. It is limited to making low-power
measurements.
• The current is proportional to the power due to the square-
law characteristic of the crystal. This square-law
characteristic only occurs for small signal levels.
• At larger signal levels the relationship is linear, as with any
diode. Therefore the proportional relationship between
power and current output is only true at power levels below
10mW.
Microwave Crystal
CALORIMETERS
• The calorimeters are the most accurate of all instruments
for measuring high power.
• Calorimeters depend on the complete conversion of the
input electromagnetic energy into heat.
• Direct heating requires the measurement of the heating
effect on the medium, or load, terminating the line.
CALORIMETERS
• Indirect heating requires the measurement of the heating
effect on a medium or body other than the original power-
absorbing material.
• Power measurement with true calorimeter methods is based
solely on temperature, mass, and time.
• Substitution methods use a known, low-frequency power to
produce the same physical effect as an unknown of power
being measured.
• Calorimeters are classified as STATIC (non flow) types and
CIRCULATING (flow) types.
CALORIMETER
References:
❖ Liao, S.Y., “Microwave Devices and Circuits”, Prentice Hall of India,
2006