Reference Standards Ppt
Reference Standards Ppt
Presented By
Dr. Pramod Kumar
Assistant Professor
Department of Pharmaceutical Analysis
National Institute of Pharmaceutical Education and Research,
Guwahati-781125
Introduction
Reference Standards (RSs) are the chemical substances or drug
substances with high purity.
Employed as standard in a assay, identification or purity test
Characterized by additional procedures other than used in routine testing
These are not intended to use as drugs.
Reference Standards (RS) plays vital role in all phases of Drug
Development process.
RS serves as basis for evaluation of both Process and Product
performance & serves as bench marks for drug Safety.
2
Sources of Reference Materials
Primary
sources Secondary
sources
United States Pharmacopeia(USP)
European Pharmacopoeia (EP) Custom Manufactures
Japanese Pharmacopoeia (JP) Contract Manufacturers
Swiss Pharmacopoeia (SP) Chemical Suppliers
French Pharmacopoeia (FP)
Mexican Pharmacopoeia (MP)
In India,
Indian Pharmacopoeia Commission,
Ghaziabad
United State Pharmacopoeia, Hyderabad
Brach 3
Contd
…
4
Definition As per USP-NF
6
Indian Pharmacopoeia Reference Substance
(IPRS) & Certied Reference Material (CRM)
I.P. Reference Substance is a primary
Reference substance that has the
appropriate quality within a specied
context and is accepted without
requiring comparison to another
substance
Specific uses
Method validation
Calibration of instruments
9
Qualitative and quantitative analysis includes following aspects,
1) Identification
2) Purity analysis
3) Assay of sample
10 1
Calibration of instruments
When an instrument is giving irrelevant results then it can be calibrated by using
RSs.
If results are close to acceptance criteria then referred as instrument is working
properly.
Some examples are given below,
Instrument RS Used for Calibration
UV-Vis Spectrophotometer Potassium dichromate
IR Spectrophotometer Poly Styrene film
Karl Fischer Titrimeter Di Sodium Tartrate
Flame Photometer Potassium dichromate
Gas Chromatography 1.0% v/v Toluene solution in n-
Hexane
DSC Indium
NMR TMS
11
Method validation & Uncertainty of Method
12
Classificatio
n
Analytical Reference
Standards (ARS)/
Primary Standard
Working
Secondary
Standard (WS)
Authentic Materials
(AM)
13
Analytical Reference Standards (ARS)
Definition:
A material or substance with well established properties, used for the
calibration of apparatus the assessment of a measurement method or for
assigning values to materials.
USES:
✔ Mainly for Research & Drug development purpose
✔ To characterize WS
14
Working Standard (WS)
Definition:
A drug substance of established quality and purity as shown by comparison
to the reference standards material and used as reference working
substance for routine quality control.
USES:
✔ Mainly for Laboratory purpose.
15
Comparison Among ARS, WS & AM
CATEGORY ARS WS AM
Synonym Primary (1˚) / Gold standard Secondary(2˚) standard Authentic Visual
Reference
Purity NLT 99.5% NLT 95% NLT 80%
Quality Highly Characterized Not so as ARS Not so as WS
Characterization Acceptable without comparison Requires comparison with 1˚
standard
Purpose For industrial purpose in ❖ For Laboratory purpose & As visual images to
Drug development process in QC compare certain test
R&D articles to meet
Calibration of apparatus, requirements
methods and materials
Calibration of 2˚ standard
Not Used As drug or cosmetic for For research purpose Not for chemical analysis
consumption
Availability Synthesized independently & Procures from Bulk Available in small quantity
available in very less quantity. manufacturer or prepared in
laboratory itself
Cost Very expensive Low cost Low cost 16
Types of Reference
Standards:
Reference Solution Related Substance System Suitability is
Compounds which include the related a mixture of the
are difficult to impurity, By- compound with a
handle like very products, small amount of the
hygroscopic or degradation impurity is required;
moisture sensitive, a products and used mixtures of
stable solution of to prove the system impurities or a
drug substance can suitability, mixture of impurities
be manufactured qualitative or and the compound
and filled into tight quantitative analysis as reference
glass ampule for standard. SST test
once use only with a based on the
detail of storage unadjusted RRT, RT
condition and shelf and the resolution.
life
17
Classification According To ISO & USP
Certified Reference
Reference Material
Material
• A (CRM)
reference material (RM)
• It is a Working level
whose property values Reference Material
are certified by a used for the calibration
technically valid of an apparatus, the
procedure assessment of a
accompanied by or measurement method,
traceable to a or for assigning values
certificate or other to materials.
documentation which
is issued by a certifying 18
body.
Reference Standards
INCASE OF:
Analytical laboratories facing difficulty to procure ARS from Official
sources
For daily analysis CONDITIONS: Prepared ARS from WS should meet
properties of RS.
19
Production and Planning
Selection of Candidate Material
Measurement/Testing Procedure: (IP/BP/EP/USP/In-house method)
Validation of measurement Procedure (Q2R1)
Verification and calibration of measuring equipment
Assessment of homogeneity study
Assessment of stability study
Assigning property value based on the results of measurement
Uncertainties of assigned value
Ensuring appropriate packaging and labelling
Transport arrangements
Post production stability
2
Selection of Candidate Material
2
Storage and Containerization
▪ Candidate material stored at 2⁰C to 8⁰C
▪ Otherwise specified in the monograph.
▪ Store in a tightly closed amber colored container
▪ If the candidate material is hygroscopic in nature should be stored in
vacuum desiccator.
Vacuum Desiccator
2
Introduction to Homogeneity, Stability and Characterization of a
CRM
Introduction to-
▪ Homogeneity
• Within-unit vs. between-unit
▪ Stability
• Short term vs. long term
▪ Characterization
Homogeneity
▪ Always required (CRMs and RMs) as the definition states “sufficiently homogeneous”.
▪ Reference Material (RM) is sufficiently homogenous and stable with respect to one or more
specified properties, which has been established to be fit for its intended use in a
measurement process.
Dispensin Distribution
g
Bulk Individual
Homogenization units
Stability
▪ Always required (CRMs and RMs) as the definition states “sufficiently stable”.
▪ Covered in ISO Guide 34 (requirements) and ISO Guide 35 (experiments and data
analysis).
▪ All property value(s) to be certified must be assessed.
▪ The stability of the reference material shall be assessed. Testing, calibration,
measurement, sampling and other activities performed for the assessment of stability
shall be carried out in compliance with ISO/IEC 17025. Stability testing can be
performed only if sufficient homogeneity is demonstrated.
▪ Two different types of stability-
Long-term stability
Short-term stability
Transport stability
Long-term stability versus short-term stability
⚫
At relevant temperature (2ºC to 8ºC)
The uncertainty for long-term stability must be quantified (uLTS).
▪ Short-term stability/Accelerated (STS): must predict degradation
⚫
⚫
At accelerated temperature (40ºC/ 75% RH) for 6 months
Performed at 1, 3, and 6 months interval.
▪ Transported stability (TS): Stability of the material under transport conditions
⚫
⚫
Temperature can reach from 25 ºC to 50 °C during dispatch.
Any evidence of degradation at that temperature.
Standardization of IPRS
There can be three different situations depending upon the availability of
reference material
Situation
When any pharmacopoeia reference standard is not available then potency
shall be assigned on the basis of chromatographic purity, LOD/Water,
Residual solvents, and Sulphated ash.
Followed by complete Identification/Molecular evidence (characterization) of
the API as per the protocol .
2
Assigned value determination by mass balance method
Or
Or
Note: Sulphated ash will not be done on salts such as Sodium, Potassium, etc. and in such cases were salts content
such as Sodium, Potassium will be done additionally.
2
Protocol for IPRS standardization
Parameters Experiment
Description Physical examination by naked eye
By Infrared Spectroscopy
Identication By chemical /TLC analysis / UV
By HPLC
Molecular mass conrmation by mass spectrometry
Complete structure elucidation of sample by Nuclear Magnetic
Resonance Spectroscopy
Characterization
Percentage conformation of C,H,N,S & O by CHNSO Analyzer
Other technique
As per IP monograph
Sulphated Ash
Should be done in duplicate
As per IP monograph
Loss on drying Should be done in duplicate
Develop new method for LOD
Water content As per IP monograph
3
Protocol for IPRS standardization
Parameters Experiment Acceptance Criteria
Related substances as per IP monograph by
HPLC
3
Protocol for IPRS standardization
Acceptance
Parameters Experiment
Criteria
As per IP monograph
Difference between
Prepare two sample solutions by injections of test should
duplicate weight. not be more than 1%
3
IPRS- Process Flow
Preparation & Sampling as per
Candidate
approval of test designed protocol
material
protocol
received
Lab-1
(IPL)
Lab -2
Analysis of Issuing of candidate material to
candidate Lab -3
concerned laboratories
material Lab -4
Lab -5
33
Assignment of property value(s) and their
Uncertainty
⚫ Assessment of uncertainties of property value(s) in accordance with
the GUM
⚫ Stability
⚫ Homogeneity
⚫ Value assignment
Term Standard
uncertainty due to
uchar Characterization
ubb Between-unit
(in)homogeneity
ults Long-term
(in)stability
uCRM Property value
3
Certicate of CRM
3
3
Reference Standards available at IPC
3
IP Prednisone Tablet (Dissolution Apparatus) Calibrator)
API: 537
Impurity: 117
(Available)
API: 50
Impurity: 40
(Under Process)
3
Instruments in Reference standard
development Laboratory
Instruments in Reference standard
development Laboratory
4
IPRS Filling Machine
4
Steps involved
1) Bulk material procured from manufacturer will be characterized &
purified (i.e., WS)
2 From above 5-50 grams collected and stored in large glass ampoules
42
WS Preparation
43
Characterization Of Reference Standards
1) Identity
2) Strength
3) Quality
4) Purity
44
Basic requirements for qualifications
❖ Should understand Physico chemical Profile of RS
❖ ARS should have good chemical stability on wide range of storage
❖ Requirements for characterization should be cross-checked to avoid Analytical errors.
❖ The validity (i.e., accuracy, precision, sensitivity, specificity) of the analytical methods used for
purity determination should be demonstrated
❖ Highly pure (i.e., ≥99.5%)
❖ In total the degradation and by-products should not be more then 0.5%
❖ The levels of all impurities found (including those not chemically identified, isolated,
and characterized) should be provided as a summary tabulation
❖ Content of enantiomer in case of chiral compound is not taken into account as by-products
and degradation products by normalized area percent
45
Steps Involved In Characterization
Physical Description Proof of structure
Elemental analysis
Visual Inspection
Optical microscopy U.V. spectroscopy
I.R. Spectroscopy
Mass Spectrometry
NMR Spectroscopy
46
Contd…
Purity
• Loss On Drying (LOD)
• Karl Fisher Titration
• Thermo Gravimetric Analysis (TGA)
• Differential Scanning Calorimetry (DSC)
• Residue On Ignition (ROI)
• Thin Layer Chromatography (TLC)
• High Performance Liquid Chromatography (HPLC)
• Gas Chromatography (GC)
Assay
• Titration
• Phase - Solubility Analysis
47
Proof Of Structure:
Mass spectrometry
IR
spectroscopy NMR 48
Purity Determination
Karl Fischer Titration: Determines free content & water of hydration trapped into the
crystal
Caution must be taken while doing KF Titration. No external moisture should
interrupt.
LOD: Determines amount of volatile matter
The % of material lost on drying should correlate to the total amount of solvent and
moisture obtained by other measurements, when determining mass balance
accountability for reference standard material
TGA: Measures the change in the mass of sample as the temperature is
changed
Gives information regarding moisture and solvent levels, an indication if the material
is a hydrate or solvate, and the rate of reaction
DSC: Measures the difference in energy (heat flow) between reference and
sample.
Used to accurately measure the melting point and purity of the reference
49
material.
Contd…
51
Stability, Expiry & Retest
Expiration and Retest Date:
For any new chemical entity an initial expiry or retest date decided as 6
months if stock material stored at -20⁰C
52
Prerequisite for analytical standard
Essential for bulk drug testing, should be safeguarded by proper storage
56
Records
Only authorized personnel of the laboratory should have an access to reference
standards
Codes should be given to reference standards for easy identification and handling
A log book should be maintained in order to keep the record of the reference
drugs used
The personnel handling reference standards should take utmost care while
maintaining these records
The log book should have compulsorily have the details like date, name of
standard taken, purpose for which it is issued, amount issued, amount remaining,
57
5
Research articles published from NDTL Project : 02