0% found this document useful (0 votes)
9 views10 pages

16-3 Pipettor-B240P-V1.0-20190218

The document provides a guide for disassembling and adjusting components of the B240P series pipettor. It includes detailed steps for disassembly, information on motors and sensors, and instructions for replacing the pipetting needle. Key components such as the 01-E01 board and various cables are also described.

Uploaded by

Arnoldo Felix
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
9 views10 pages

16-3 Pipettor-B240P-V1.0-20190218

The document provides a guide for disassembling and adjusting components of the B240P series pipettor. It includes detailed steps for disassembly, information on motors and sensors, and instructions for replacing the pipetting needle. Key components such as the 01-E01 board and various cables are also described.

Uploaded by

Arnoldo Felix
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 10

Disassembling and adjustment of

B240P series components


--Pipettor

ITSD

FOCUS MAKES PROFESSION


01 Introduction
02 Disassemble steps

03 Motor and sensors

04 01-E01 board

05 Replacement of pipetting needle

FOCUS MAKES PROFESSION


1. Introduction

Sample & Reagent probe

Motion mechanism
2. Disassemble steps

1. disconnect the linking pipeline. 2. remove these four screws.


2. Disassemble steps

3. disconnect these two black cables 4. Take the pipettor out gently and slowly.
3. Motor and sensors

the motor for the


horizontal movement of
Sample & Reagent Probe

the motor for the vertical


movement of Sample &
Reagent Probe

6
3.Motor and sensors

Initialization sensor of
horizontal movement
3.Motor and sensors

Initialization sensor of
vertical movement
4. 01-E01 board

Yellow cable :
Liquid level detection signal cable

cables for motors

cables for Signal and power cable


initialization sensor
5. Replacement of pipetting needle

2
1
3

Replacement of pipetting needle:

1. Loose the screw of the tubing

2. Loose the screw which fixed the needle in the pipetting arm

3. Remove the cable