Notes tancet
Notes tancet
parallels.com Download
1) Static characteristics
2) Dynamic characteristics
1) Static characteristics of Electrical Measuring
Instruments:
Some applications involve the measurement of quantities that are either constant or
vary slowly with time.Under these circumstances, it is possible to define a set of criteria
that gives a meaningful description of the quality of measurement without interfering with
dynamic descriptions that involve the use of differential equations.These criteria are called
static characteristics.
i) Accuracy:
It is the closeness with which an instrument reading approaches the true value of
the quantity being measured.Thus accuracy of a measurement means conformity to
truth.It the important static characteristic of electrical measuring instruments.
a.Point accuracy:
This is the accuracy of the instrument only at one point on its scale.The specification of
this accuracy does not give any information about the accuracy at other points on the
scale or in the words, does not give any information about the general accuracy of the
instrument.
Let us say that its readings can be taken to 1/100 of an ampere. Now every
time we take a reading, the ammeter is as precise as ever, we can take readings down to
1/100 of an ampere and the readings are consistent and clearly defined. However, the
readings taken with this ammeter are not accurate, since they do not confirm to truth on
account of its faulty zero adjustment.
iii)Stability:
If the input to an instrument increases slowly from some arbitrary non-zero value, it will
be observed that the output of the instrument does not change at all until there is a
certain minimum increment in the input. This minimum increment in what is input is
called resolution of the instrument.Thus, the resolution is defined as the smallest
incremental of the input quantity to which the measuring system responds.This is the
third most important static characteristic of electrical measuring instruments.
If the instrument input is increased very gradually from zero there will be some
minimum value below which no output change can be detected. This minimum value
defines the threshold of the instrument.In specifying threshold, the first detectable output
change is often described as being any noticeable measurable change.
vi)Drift :
Zero drift is a deviation observed in the instrument output with time from the
initial value, all the other measurement conditions are constant. This may be caused by a
change in component values due to variation in ambient conditions or due to
ageing.Typical units by which zero drift is measured are volts per °C in the case of a
voltmeter affected by changes in ambient temperature.This is often called the zero drift
coefficient related to temperature changes.
vii)Repeatability:
It is the closeness with which the same value of the input quantity is measured at
different times and under different conditions of usage of the instrument and by different
instruments. The output signals and indications are checked for consistency over
prolonged periods and at different locations. Perfect reproducibility ensures
interchangeability of instruments and transducers.
ix) Dead Zone:
It is the largest change of input quantity for which there is no output of the
instrument. For instance, the input applied to the instrument may not be sufficient to
overcome the friction and will, in that case not move at all.
The maximum distance or angle through which any part of the mechanical
system may be moved in one direction without applying appreciable force or motion to
the next part in a mechanical sequence.
XI)Hysteresis:
Hysteresis is a phenomenon which depicts different output effects when loading and
unloading whether it is a mechanical system or any electrical system or any other
system.Hysteresis is the difference in the readings of an instrument, which fixed value of
the input signal, which depends on whether that input value is approached from
increasing or decreasing values of input.That is upscale and down scale deflections do not
coincide when the measurement is made of the same value by the method of
symmetry.The non-coincidence between the loading and unloading curves is known as
hysteresis.
Xii)Linearity:
It defines the proportionality between input quantity and output signal. If the
sensitivity is constant for all values from zero to full scale value of the measuring system,
then the calibration characteristic is linear and is a straight line passing through the
origin.If it is an indicating or recording instrument the scale may be made linear.In case
there is a zero error the characteristic assumes the form of the equation given by
y=mx+c where y is output,x is input,m is slope and c is intercept.
Span and range are the two terms that convey the information about the lower
and apa calibration points.The range of indicating instruments is normally from zero to full
scale value and the Span is simply the difference between the full scale and lower scale
value.But same instruments operate under a bias so that they start reading, for example,
voltages from 5V to 25V only. The zero of these instruments is suppressed from indication
by means of a bias. In such case, the scale range is said to be from 5V to 25V and the
scale span is 25-5 i.e.,20V.
Xiv)Bias:
Bias describes a constant error which exits over the full range of measurement of an
instrument. The error is normally removable by calibration.
XV)Tolerance:
It is a term which is closely related to accuracy and defines the maximum error
which is to be expected in some value.While it is not, strictly speaking, a static
characteristic of measuring instruments, it is mentioned here because the accuracy of
some instruments, is sometimes quoted as a tolerance figure. Tolerance, when used
correctly, describe the maximum deviation of a manufactured component from some
specified value.Electric circuit components such as resistors have tolerances of perhaps
5%.
2) Dynamic characteristics of Electrical Measuring
Instruments :
Measurement systems having inputs dynamic in nature, the input varies from instant to
instant, so does the output.The behaviour of the system under such conditions is dealt by
the dynamic response of the system and its dynamic characteristics of electrical
measuring instruments are given below:
1. Dynamic error:
It is the difference of true value of the quantity changing with the time the value indicated
by the instrument provided static error is zero.Total dynamic error is the phase difference
between input and output of the measurement system.
2. Fidelity:
It is the ability of the system to reproduce the output in the same form as the input.In the
definition of fidelity any time lag or phase difference is not included.Ideally a system
should have 100% fidelity and the output should appear in the same form as the input
and there is no distortion produced by the system. Fidelity needs are different for different
applications.
3. Bandwidth:
For other physical systems, electrical filters electronic amplifiers, the above criterion is
relaxed with the result that their bandwidth specification extend to frequencies at which
the dynamic sensitivity is 70.7 % of that at zero or the mid- frequency.
4. Speed of response:
It is associated with the behaviour of a first order system and is defined as the time taken
by the system to reach 0.632 times its final output signal amplitude. System having small
time constant attains its final output amplitude earlier than the one with larger time
constant and therefore, has higher speed of response.
6. Measurement lag:
Measurement lag is of two types. In retardation type, the response of the instrument
begins immediately after a change in the measurand has occurred. In time delay type, the
response of the system begins after a delay time after application of the input.
7. Settling time:
Post a Comment
Post a comment
Most Popular
Categories
Download Ebooks
ABOUT EEI
FOLLOW US
! $