To use a hemocytometer, it should first be cleaned with 70% ethanol and lens paper. The coverslip is then gently
placed atop the counting chamber. You will know the coverslip is properly positioned if you can observe a
phenomena known as Newton's rings; a concentric-ring pattern of color. A small sample of cell suspension is taken
using a pipette and the pipette is placed near the edge of the chamber, allowing the cell suspension to enter the
counting chamber by capillary action. If cell viability is to determined, trypan blue should be added to the cell
suspension (in a 1:1 ratio) prior to addition to the chamber.
The microscope is then focused on an area of the counting chamber and the cells are counted using a tally counter.
This is typically done using the 1 mm2, 100 nl area of the counting chamber and a 4x or 10x objective, but the precise
area and objective used will depend on the size of your cells and their density in suspension. This process is generally
repeated using four different 1 mm2 areas and the results are averaged. If determining cell viability, separate counts
should be made for live and dead cells, with the dead cells appearing blue due to the permeability of their damaged
membranes to trypan blue.
To use an automated cell counter, the cell suspension (with or without trypan blue) is loaded into a cell counting
chamber that is unique for each counter. The counting chamber is then inserted into the counter, settings are
selected and the run is performed. Most automated counters allow settings to be saved for rapid retrieval. Results
will become available in as little as 7 seconds (if using the Logos Luna™, for example).
Fluorescent cell counters, such as the Luna-FL™ dual fluorescence cell counter, distinguish live and dead cells as well
as perform a cell count by using common stains such as acridine orange and propidium iodide. These types of cell
counters are also superior at counting cells in cultures which may be contaminated with non-cell debris, such as
primary cells, as the dyes will clearly distinguish between cells.
Benefits & Drawbacks: For general purpose cell counting and cell viability applications, automated cell counters are
an affordable and high-throughput solution. The cost of operation is low, they are easy to use, and they greatly reduce
the amount of human effort required to count cells. They are both precise and reliable, but may have difficulty
obtaining accurate measurements of cells that are highly irregularly shaped, are extremely small, or are in cell
suspensions that are extremely dilute or contain a large variety of cells that need to be distinguished. For most cell
types and most applications, automated cell counters provide excellent counting performance at a relatively low cost.
Coulter Counters
Coulter counters are not optical instruments, but rather measure the electrical resistance across one or more
microchannels. Cells, having greater resistance than the electrolyte solution that the cells are suspended in, cause a
brief increase in resistance when passed between the channels. This change, which increases with cell size, is
detected by the Coulter counter.
Use of a Coulter counter is somewhat similar to that of an automated cell counter. The cell suspension is diluted, as
needed, thoroughly mixed to ensure an even cell distribution, added to a vial, and a run is started. Unlike automated
cell counters, however, use of Counter counters requires running a blank first and also flushing the device after use.
Benefits & Drawbacks: Because of their relative speed compared to manual counting and their ability to accurately
count cells of differing size, Coulter counters are frequently used for complete blood counts in hospitals, where red
blood cells and white blood cells need to be quickly and accurately distinguished. However, Coulter counters are not
capable of distinguishing live cells from dead cells, nor do they accurately count cells which form clusters or clumps.
They also require more maintenance.
Flow Cytometers
Flow Cytometers are most frequently used for more detailed cellular analysis, as they are equipped with fluorescence
detection technologies that can detect labelled intracellular components. Not all flow cytometers are capable of
determining cell count, as not all draw specific liquid volumes or measure the volume of liquid drawn. Those that do,
however, are capable of providing highly accurate cell counts, and can discriminate cells based on factors such as
protein expression by using fluorescently labelled antibodies. This makes them able to discern cell types of the same
size within the same sample, or even the same types of cells at different cell stages. However, such complex
experiments also complicate the experimental set up and antibody incubations are often on the scale of hours,
thereby potentially extending the workflow quite significantly.
Use of a flow cytometer is quite simple (load and run), and therefore the overall ease of use depends more heavily on
the experimental set-up.
Benefits & Drawbacks: Flow cytometers are extremely powerful cellular analysis tools but are also extremely
expensive, with costs ranging from $40,000 to over $100,000. Because of this, they are rarely used for general cell
counting applications.
To estimate cell density using a spectrophotometer, place the cell suspension in a cuvette and measure the
absorbance. If you are looking to get a relative density measurement, you may simply compare to another sample.
Otherwise, you must compare to cell suspensions of known density in order to estimate an absolute cell density.
Plating is another method of counting cells, although only in colony-forming cells such as bacteria. To count cells
with plating, the cells are heavily diluted and streaked onto a plate. After given sufficient time for colony growth, the
number of colonies are counted. Based on the dilution and the known volume of suspension that was streaked onto
the plate, the density of the original suspension can be determined.
Plating is only a useful method for microbes, and due to the time required for colony formation it is also the slowest
method.