DM3058 DataSheet en

The RIGOL DM3058 Digital Multimeter offers high-precision measurements with a resolution of 5 ½ digits and a measurement speed of up to 123 readings per second. It features a user-friendly design with a double display, supports various measurement functions including True-RMS AC and DC voltage, and allows for remote control and data management via USB and LAN. The device is suitable for applications in research, education, and production testing, making it a versatile tool for professionals.

Uploaded by

Sebastian Juwa
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
18 views8 pages

DM3058 DataSheet en

The RIGOL DM3058 Digital Multimeter offers high-precision measurements with a resolution of 5 ½ digits and a measurement speed of up to 123 readings per second. It features a user-friendly design with a double display, supports various measurement functions including True-RMS AC and DC voltage, and allows for remote control and data management via USB and LAN. The device is suitable for applications in research, education, and production testing, making it a versatile tool for professionals.

Uploaded by

Sebastian Juwa
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 8

RIGOL

Data Sheet
DM3058 Digital Multimeter

Product Overview

DM3058 is a digital multimeter designed


with 5 ½ digits readings resolution
especially fitting to the needs of
high-precision, multifunction, and
automatic measurement.

Applications
Easy to Use Design
 R&D Laboratory
 Scientific Research and Education  256× 64 LCD
 Detection and Maintenance  Support double display, Chinese and
 Quality Test English menu
 Automatic Production Testing  Push-help makes information acquire
more easier
Main Features  File management (support for U-disc and
local storage)
 Real 5 ½ digits readings resolution
 Up to 123 rdgs/s of measurement speed
Powerful Measurement Functions
 True-RMS AC Voltage Current measuring
 Quickly Save or Recall the 10 groups of
 Basic Measurement Function
Preset Configuration
 Preset 10 groups of Standard Sensor  DC Voltage: 200 mV ~ 1000 V
Configuration, built-in Thermocouple  DC Current: 200 μA ~ 10 A
compensate in Cold Terminal  AC Voltage: True-RMS, 200 mV ~ 750 V
 Clone or backup all the configurations  AC Current: True-RMS, 20 mA ~ 10 A
within instrument into other DM3058 via  2-Wire, 4-Wire Resistance: 200 Ω ~ 100
U-disc MΩ
 The first 5 ½ digit digital multimeter  Capacitance Measurement: 2 nF ~ 10000
passing LXI Criterion in industry, which μF
can achieve system integration easily  Continuity Test: Range is fixed at 2 kΩ
 With easy, convenient and flexible random  Diode Test: Range is fixed at 2.0 V
sensor measurement control software:  Frequency Measurement: 20 Hz ~ 1 MHz
UltraSensor  Period Measurement: 1 μs ~ 0.05 s
 Standard configuration interface: USB  Random Sensor Measurement: Support
Device, USB Host, LAN, RS-232, GPIB, and for 6 types of sensor: DCV, DCI, Freq,
support U-disc storage and Web remote 2WR, 4WR and TC
control  Math Function
 Support remote control via a command
line and enable to compatible with Max, Min, Average, Standard Deviation,
commands from main stream multimeters Pass/Fail, dBm, dB, Relative Measurement
and Vertical chart

Feb. 2009
RIGOL Technologies, Inc.
Double Display Mirror Image Configuration

Double Display function makes your All the configurations (system and sensor
measurements achieve maximum results with configuration) of DM3058 can be cloned into
little effort. It can display two performances of other DM3058 on product line via U-disc to
a signal simultaneously instead of two improve work efficiency.
multimeters or measurements in the past.

Preset Mode Pass/Fail

Based on Preset Mode, the worker operation Pass/Fail test function can prompt for signals
on product line could be greatly simplified. beyond the range based on specified higher
Besides, 10 groups of Preset Configuration and lower limit, and make test result more
are available to be stored and recalled. obvious.

Unique Random Sensor LXI & Web Remote Control


DM3058 is the first 5½ digit digital multimeter
passing LXI Criterion in industry that makes
system integration easier.

Random Sensor measurement is a new Remote Control for DM3058 can be achieved
conception being forward to meet user’s by virtual panel from Web page in the same
requirements. By this particular function, you operation way with front panel of instrument.
can easily connect Pressure Sensor or Flux
Sensor or Temperature Sensor.
DM3058 supports DCV, DCI, and Freq, 2WR,
4WR and Thermocouple TC 6 kinds of Sensor
in all. Meanwhile, there are 10 groups of
basic Sensor preset within the instrument.
Also, the built-in Thermocouple Cold Terminal
temperature sensor can be used for
measuring temperature of thermocouple

Easy, Convenient and Flexible Control Software


UltraSensor is used to control random sensor. The
main functions include:
 Create a project for random sensor measurement
which can be download to DM3058;
 Connect with DM3058 to achieve random sensor
measurement;
 Monitor the sensor data in real time and show
them in the form of figure;
 Enable to save data in the format of CSV and TXT,
and reference curve in the format of BMP.
UltraSensor Software Interface 1
Specifications

DC Characteristics
Accuracy ± (% of reading + % of range)[1]
[2]
Function Range Test current or 1 Year Temperature
Load voltage 23℃± 5℃ Coefficient
0℃ - 18℃
28℃ - 50℃
DC Voltage 200.000 mV 0.015 + 0.004 0.0015 + 0.0005
2.00000 V 0.015 + 0.003 0.0010 + 0.0005
20.0000 V 0.015 + 0.004 0.0020 + 0.0005
200.000 V 0.015 + 0.003 0.0015 + 0.0005
1000.00 V[4] 0.015 + 0.003 0.0015 + 0.0005
DC Current 200.000 μA <8 mV 0.055 + 0.005 0.003 + 0.001
2.00000 mA <80 mV 0.055 + 0.005 0.002 + 0.001
20.0000 mA <0.05 V 0.095 + 0.020 0.008 + 0.001
200.000 mA <0.5 V 0.070 + 0.008 0.005 + 0.001
2.00000 A <0.1 V 0.170 + 0.020 0.013 + 0.001
10.0000 A[5] <0.3 V 0.250 + 0.010 0.008 + 0.001
Resistance[3] 200.000 Ω 1 mA 0.030 + 0.005 0.0030 + 0.0006
2.00000 kΩ 1 mA 0.020 + 0.003 0.0030 + 0.0005
20.0000 kΩ 100 μA 0.020 + 0.003 0.0030 + 0.0005
200.000 kΩ 10 μA 0.020 + 0.003 0.0030 + 0.0005
2.00000 MΩ 1 μA 0.040 + 0.004 0.0040 + 0.0005
10.0000 MΩ 200 nA 0.250 + 0.003 0.0100 + 0.0005
100.000 MΩ 200 nA || 10 MΩ 1.75 + 0.004 0.2000 + 0.0005
Diode Test 2.0000 V[6] 1 mA 0.05 + 0.01 0.0050 + 0.0005
Continuity 2000 Ω 1 mA 0.05 + 0.01 0.0050 + 0.0005
Test
Remarks:
[1] Specifications are for 0.5 hour warm-up, “Slow” measure and calibration temperature 18℃ - 28℃.
[2] 20% over range on all ranges except for DCV 1000 V, ACV 750 V, DCI 10 A and ACI 10 A.
[3] Specifications are for 4-wire measure or 2-wire measure under “REF” operation. ± 0.2 Ω of extra errors will be
generated if perform 2-wire measure without “REF” operation.
[4] Plus 0.02 mV of error per 1 V after the first ± 500 VDC.
[5] 30 seconds OFF after 30 seconds ON is recommend for the continuous current that higher than DC 7 A or AC RMS
7 A.
[6] Accuracy specifications are only for voltage measuring at input terminal. The typical value of current under
measure is 1 mA. Voltage drop at the diode junction may vary with current supply.

DC Voltage
Input Resistance 200 mV and 2 V range 10 MΩ or >10 GΩ selectable
(Input signals that exceed ± 2.5 V in these ranges will pass the 100 kΩ (typical) clamp
resistance.)
Input Bias Current <90 pA, 25℃
Input Protection 1000 V on all ranges
CMRR (common mode 120 dB (For the 1 kΩ unbalanced resistance in LO lead, maximum ± 500 VDC)
rejection ratio)
NMRR (normal mode 60 dB at “slow” measurement rate
rejection ratio) 20 dB are added if open the “AC filter” (Settling time will add 0.35 s while source
impedance nears zero)
Resistance
Testing Method 4-wire resistance or 2-wire resistance optional
For current source refer to LO input
Open-circuit Voltage P/F in <8 V
Maximum Lead 10% of ranges, on the range of 200 Ω, 1 kΩ, for each lead
Resistance 1 kΩ, on all other ranges, for each lead
(4-wire resistance)
Input Protection 1000 V, on all ranges
DC Current

2
Shunt Resistor 200 μA sampling voltage < 8 mV
2 mA sampling voltage < 80 mV
1Ω for 20 mA, 200 mA
0.01 Ω for 2 A, 10 A
Input Protection Rear panel: accessible 10 A, 250 V fast-melt fuse
Internal: 12 A, 250 V slow-melt fuse
Continuity / Diode Test
Measurement Method 1 mA ± 5% constant-current source, < 8 V open-circuit voltage
Response Time 123 samples/sec, with beeper
Continuity Threshold Adjustable between 1 Ω and 2000 Ω
Input Protection 1000 V
Setup time Attentions
The setup time about voltage measurement is influenced by source resistance and media characteristics of cable as
well as input signal

AC Characteristics
Accuracy ± (% of reading + % of range) [1]
Function Range[2] Frequency Range 1 Year Temperature
23℃± 5℃ Coefficient
0℃ - 18℃
28℃ - 50℃
True RMS 200.000 mV 20 Hz – 45 Hz 1.5 + 0.10 0.01 + 0.005
AC Voltage[3] 45 Hz – 20 kHz 0.2 + 0.05 0.01 + 0.005
20 kHz – 50 kHz 1.0 + 0.05 0.01 + 0.005
50 kHz – 100 kHz 3.0 + 0.05 0.05 + 0.010
2.00000 V 20 Hz – 45 Hz 1.5 + 0.10 0.01 + 0.005
45 Hz – 20 kHz 0.2 + 0.05 0.01 + 0.005
20 kHz – 50 kHz 1.0 + 0.05 0.01 + 0.005
50 kHz – 100 kHz 3.0 + 0.05 0.05 + 0.010
20.0000 V 20 Hz – 45 Hz 1.5 + 0.10 0.01 + 0.005
45 Hz – 20 kHz 0.2 + 0.05 0.01 + 0.005
20 kHz – 50 kHz 1.0 + 0.05 0.01 + 0.005
50 kHz – 100 kHz 3.0 + 0.05 0.05 + 0.010
200.000 V 20 Hz – 45 Hz 1.5 + 0.10 0.01 + 0.005
45 Hz – 20 kHz 0.2 + 0.05 0.01 + 0.005
20 kHz – 50 kHz 1.0 + 0.05 0.01 + 0.005
50 kHz – 100 kHz 3.0 + 0.05 0.05 + 0.010
750.000 V 20 Hz – 45 Hz 1.5 + 0.10 0.01 + 0.005
45 Hz – 20 kHz 0.2 + 0.05 0.01 + 0.005
20 kHz – 50 kHz 1.0 + 0.05 0.01 + 0.005
50 kHz – 100 kHz 3.0 + 0.05 0.05 + 0.010
True RMS 20.0000 mA 20Hz – 45 Hz 1.5 + 0.10 0.015 + 0.015
AC Current[5] 45 Hz - 2 kHz 0.50 + 0.10 0.015 + 0.006
2 kHz -10 kHz 2.50 + 0.20 0.015 + 0.006
200.000 mA 20 Hz - 45 Hz 1.50 + 0.10 0.015 + 0.005
45 Hz – 2 kHz 0.30 + 0.10 0.015 + 0.005
2 kHz - 10 kHz 2.50 + 0.20 0.015 + 0.005
2.00000 A 20 Hz – 45 Hz 1.50 + 0.20 0.015 + 0.005
45 Hz - 2 kHz 0.50 + 0.20 0.015 + 0.005
2 kHz – 10 kHz 2.50 + 0.20 0.015 + 0.005
10.0000 A[5] 20 Hz – 45 Hz 1.50 + 0.15 0.015 + 0.005
45 Hz - 2 kHz 0.50 + 0.15 0.015 + 0.005
2 kHz – 5 kHz 2.50 + 0.20 0.015 + 0.005

Additional wave crest factor error (not Sine)[6]


Wave crest coefficient Error (% range)
1-2 0.05
2-3 0.2
Remarks:
[1] Specifications are for 0.5 hour warm-up, “Slow” measure and calibration temperature 18℃ - 28℃.
[2] 20% over range on all ranges except for DCV 1000 V, ACV 750 V, DCI 10 A and ACI 10 A.
[3] Specifications are for amplitude of sine wave input >5% of range. 750 V range limited to 8x107 Volt-Hz. For inputs
3
from 1% to 5% of range and <50 kHz, add 0.1% of range extra error. For 50 kHz to 100 kHz, add 0.13%.
[4] Specifications are for sine wave input >5% of range. 0.1% errors will be added when the range of input sine wave
is 1% ~ 5%.
[5] 30 seconds OFF after 30 seconds ON is recommend for the continuous current that higher than DC 7 A or AC RMS
7 A.
[6] For frequency<100 Hz.

True RMS AC Voltage


Measurement Method AC coupled true RMS measure - up to 1000 V DC bias are permitted on every range
Wave Crest Factor ≤ 3 at full scale
Input Impedance 1 MΩ ± 2% in parallel with <100 pF on all ranges
AC Filter Bandwidth 20 Hz ~ 100 kHz
CMRR (common mode 60 dB (For the 1 kΩ imbalance resistance among Lo lead and <60 Hz, maximum ± 500
rejection ratio) VDC)
True RMS AC Current
Measurement Method DC coupled to the fuse and shunt; AC coupled the True RMS measurement (measures the
AC components only)
Wave Crest Factor ≤ 3 at full scale
Maximum Input The DC + AC current peak value <300% of range. The RMS current including DC current
is <10 A
Shunt Resistor 1 Ω for 20 mA, 200 mA
0.01 Ω for 2 A, 10 A
Input Protection Rear panel: accessible 10 A, 250 V fast-melt fuse
Internal: 12 A, 250 V slow-melt fuse
Setup Time Attentions
Make sure that the RC return at input terminal has been in a stable state completely (higher than 1 s) before accurate
measurement;
Input >300 Vrms (or >5 Arms) will cause the self heating of the signal conditioning component to generate error, this
error is included in the characteristics of the instrument. Internal temperature variation results from the self heating
will cause additional an error on ac range that is lower than 0.03% of readings, and it will disappear after a few
minutes

Frequency and Period Characteristics


Accuracy ± (% of reading + % of range) [1]
Function Range Frequency Range 1 Year Temperature
23℃± 5℃ Coefficient
0℃ - 18℃
28℃ - 50℃
Frequency/Period 200 mV – 750 V[2] 20 Hz - 2 kHz 0.01 + 0.003 0.002 + 0.001
2 kHz - 20 kHz 0.01 + 0.003 0.002 + 0.001
20 kHz - 200 kHz 0.01 + 0.003 0.002 + 0.001
200 kHz - 1 MHz 0.01 + 0.006 0.002 + 0.002

20 mA – 10 A[3] 20 Hz - 2 kHz 0.01 + 0.003 0.002 + 0.001


2 kHz - 10 kHz 0.01 + 0.003 0.002 + 0.001
Remarks:
[1] Specifications are for 0.5 hour warm-up.
[2] Except for special marks, the AC input voltage is 15% to 120% of range when <100 kHz and 40% to 120% of
range when >100 kHz. 750 V range is limited to 750 VRMS. 200 mV ranges is for full scale or higher. For inputs
from 30 mV to 200 mV, multiply total % of reading error by 10.
[3] For AC input current from 15% to 120% of range except where noted. 20 mA range specifications are for full scale.
For inputs from 5 mA to 20 mA, multiply total % of reading error by 10. 10 A range is for AC input current from
25% to 100% of range.

Frequency and Period


Measurement Method: Reciprocal-counting technique, AC-coupled input, AC voltage or AC current measurement
function
Measure Attentions
Generally, errors are leaded into all frequency counters when measuring low voltage or low frequency signal.
Shielding input can extremely help to reduce measuring errors caused by exterior noise
Setup Time Attentions

4
If the variational DC components appeared in signals under measure, errors may be caused while measuring period
or frequency. Please ensure that the RC loop at input terminal must be stable during exact measuring (higher than 1
sec)

Capacitance Characteristics
Accuracy ± (% of reading + % of range) [1,2]
Function Range[2] Maximum Testing 1 Year Temperature Coefficient
Current 23℃± 5℃ 0℃ - 18℃
28℃ - 50℃
Capacitance 2.000 nF 200 nA 3 + 1.0 0.08 + 0.002
20.00 nF 200 nA 1 + 0.5 0.02 + 0.001
200.0 nF 2 μA 1 + 0.5 0.02 + 0.001
2.000 μF 10 μA 1 + 0.5 0.02 + 0.001
200 μF 100 μA 1 + 0.5 0.02 + 0.001
10000 μF 1 mA 2 + 0.5 0.02 + 0.001
Remarks:
[1] Specifications are for 0.5 hour warm-up and “REF” operation. Using of non-film capacitor may generate additional
errors.
[2] Specifications are for from 1% to 120% on 1 nF range and ranges from 10% to 120% on other ranges.

Capacitance Measuring
Measurement Method Measure the rate of change of voltage generated during the current flowing the
capacitance
Connection Type 2-wire
Input Protection 1000 V on all ranges
Measure Attentions:
Small capacitance is easily influenced by external noise and thus causes errors while measuring, shielding input can
extremely help to reduce this kind of errors

Other Measuring Characteristics


Triggering and Memory
Samples/Trigger 1 ~ 2000
Trigger Delay 8 ms ~ 2000 ms optional
TTL compatible (High level when left input
Input Level
terminal is hanging in the air)
External Trigger Input Trigger Condition Selectable
Input Impendence >20 kΩ,
Min Pulse 500 μs
Electric Level TTL compatible ( input >=1 kohm loads)
VMC Output Output Polarity straight polarity and negative polarity optional
Input Impedance 200 ohm, typical

Arbitrary Sensor
Support for multiply types of sensor such as Thermocouple, DC Voltage, DC Current, Resistance (2-wire or 4-wire)
and Frequency output; With thermocouple compensation at cold junction.
Output Polarity: straight polarity and negative polarity optional
Preset ITS-90 transform of B, E, J, K, N, R, S and T thermocouple and transform of platinum Pt100, Pt385 resistance
temperature sensor
Math Functions
Pass/Fail, REL (RELative), Min/Max/Average, dBm, dB, Hold, Historgram, Standard deviation
History Records
Volatile Memory 2000 readings of history records
Nonvolatile Memory 10 gourps of history records (2000 readings/group); 10 groups of sensor records: (1000
readings/group); 10 groups of setting records of instrument; 10 groups of setting records
of arbitrary sensor; support U-disk external storage

5
General Specifications
Power Supply
AC 100 V ~ 120 V 45 Hz ~ 440 Hz
AC 200 V ~ 240 V 45 Hz ~ 66 Hz
Consumption 20 VA peak value
Mechanism
Dimension 107.0 mm× 231.6 mm× 290.5 mm
Weight 2.5 kg
Other Characteristics
LCD display with 256× 64 lattices, support for Double display, Menu display, Operating
Display Screen
help and English/Chinese bilingual.
Full accuracy from 0℃ to 50℃; 80% R.H. and 40℃, non condensing
Storage Temperature: -20℃ ~ 70℃
Operating Environment
Shock and Vibration: conforming to MIL-T-28800E, III, 5 level (only for sine)
height above sea level: up to 3000 meters
Conforming to IEC61010-1: 2001. Measure CAT I 1000 V/CAT II 600 V
Safety
Class of pollution: 2
GPIB, 10/100Mbit LAN, USB2.0 Full Speed Device & Host
Remote Interface
(U disc available), RS-232C
Programmer Language RIGOL 3058 SCPI, FLUKE45, Agilent34401A
LXI Compatibility LXI Class C, Version1.1
Warm Up Time 30 minutes

6
Ordering Information Warranty
Very thank you for choosing RIGOL products!
Name of Product
RIGOL DM3058 Digital Multimeter RIGOL Technologies, Inc. warrants that this
product will be free from defects in materials and
Standard Accessories workmanship from the date of shipment. If a
product proved defective within the respective
 A Power Cord that fits the standard of period, RIGOL will provide repair or
destination country replacement as described in the complete
 Two Test Leads (black and red) warranty statement.
 Two Alligator Clips (black and red)
 An USB Data Cable For the copy of complete warranty statement or
 A Backup Fuse maintenance, please contact with your nearest
 A Quick Guide RIGOL sales and service office.
 User’s Guide and Application software
(CD-ROM) RIGOL do not provide any other warranty items
except the one being provided by this summary
Optional Accessories and the warranty statement. The warranty items
 Kelvin Test Clips include but not being subjected to the hint
 RS232 Cable guarantee items related to tradable
characteristic and any particular purpose.
RIGOL will not take any responsibility in cases
regarding to indirect, particular and ensuing
damage.

Contact Us
If you have any problem or requirement during using our products, please contact RIGOL
Technologies, Inc. or the local distributors.

Domestic: Please call


Tel: (86-10) 8070 6688
Fax: (86-10) 8070 5070
Service & Support Hotline: 800 810 0002
9:00 am –5: 00 pm from Monday to Friday

Or by e-mail:
[email protected]

Or mail to:
RIGOL Technologies, Inc.
156# CaiHe Village, ShaHe Town, ChangPing District, Beijing, China
Post Code: 102206

Overseas: Contact the local RIGOL distributors or sales office.


For the latest product information and service, visit our website: http://www.rigolna.com/