Sensitometry
Sensitometry
Without good contact between the film and screens of the cassette, detail will be
lost on the radiograph. Therefore, it is necessary to routinely perform a screen
contact test on all of your cassettes. Screen contact tests are required at intervals
not to exceed 24 months, per Minnesota Rule Chapter 4732.1100. (Installation
Calibration Tests and Equipment Perfomance Tests for Quality Assurance
Program).
Frequency of test
Procedure
Ensure that cassettes to be tested have been allowed to sit for at least 15 minutes
after loading. This will allow for any trapped air to dissipate.
Place a loaded cassette on top of the x-ray table and place the wire mesh on top
of the cassette.
Select a standard focal film distance (usually 40") and center the x-ray beam to
the cassette. (Dental pan cassettes may need a greater distance to cover the full
cassette.)
"Cone down" or restrict the primary beam to the outside edges of the cassette.
Identify the cassette with lead numbers or letters corresponding to the I.D. on
the screen inside. (All screens need identifying marks; e.g., numbers or letters
that show up on each radiograph. This enables you to identify which cassette
has a problem when one is observed.)
Select a technique that will produce an optical density of approximately 2.0 on
the developed film. For a 400 speed film-screen combination a technique of 65
kVp and 5 mAs is a starting point. Finding a technique to produce the 2.0 (± 0.5
O.D.) may take some trial and error. Remember that cassettes of a different
speed (e.g., extremity) will require a different exposure.
Record for future use the technique that produced the appropriate optical
density on the film, the room used, and focal-film-distance used.
Label, expose and process the film from each cassette.
View each film on a view box in a dimly lit room from a distance of
approximately six feet or more.
Look for areas that are darker and/or more blurry than the rest of the film. This
indicates poor contact. Do not try to identify poor contact by closely looking for
fuzzy wires on the film.
Areas of poor contact need to be evaluated as to their land size. An area of poor
contact extending less than an inch from the edge into the film may not be a
problem depending on the type of radiographs you take. If the area of poor
contact could obscure important information on a radiograph, remove the
cassette from service
SCREEN-FILM –CASSETTE SPEED MATCH: The minimum test interval is
is not to exceed 24 months. Densities within ± 0.10 OD for all cassettes of the
same speed used for imaging.
CR IMAGING PLATES: Test at intervals not to exceed three months or upon
observation of image artefacts. Follow manufacturer’s recommendation