Hologic 2015
Hologic 2015
Learning Objectives
• Tomosynthesis
• Theory, QC
• C-View™ Synthetic 2D Imaging
• Theory, QC
• I-View™ Dual Energy Mammography
• Theory, QC
• Tomo-Guided Biopsy
• Theory, QC
Hologic 2016. 2
Hologic 2016. 3
1
8/2/2016
X-Ray Generation
• FFDM: 50 μm Rh; 50 μm Ag
X-Ray Filtration • Tomo: 0.7 mm Al
• I-View: 0.3 mm Cu
Hologic 2016.
• 15 projections
• Moving detector
• Reconstruction
• ~100 μm pixel size
• 1 mm slice spacing Hologic 2016.
System Configurations
Selenia Dimensions 2D
AWS 5000/6000/8000/9000
Configuration
Hologic 2016.
2
8/2/2016
2D
Rapid review of calcification clusters
Comparison to 2D priors and left/right asymmetry
CAD operates on the 2D images
Clinical Results
2D+3D superior to 2D in cancer detection and false positive rate (Friedewald, …)
Hologic 2016.
C-View™ Image
Hologic 2016. 8
C-View 2D Image
Hologic 2016.
3
8/2/2016
2D C-View™ image
2D C-View™ image
4
8/2/2016
2D C-View™ image
2D C-View™ image
RCC LCC
5
8/2/2016
RMLO LMLO
RMLO LMLO
2D Tomosynthesis slices
6
8/2/2016
2D Tomosynthesis Slice
Invasive Ductal Carcinoma
2D C-View™ image
7
8/2/2016
Dual-Energy Subtraction
• You cannot see the iodine contrast drug in high or low kV images
• But you can when you subtract the images:
Subtraction = high kV - low kV
MLO CC CC MLO
H H H H
E ne rgy
L L L L
inject
8
8/2/2016
H H H H
Energy
L L L L
inject
Example Image
Courtesy of Andrea Woodroof, Kentucky Breast Care
9
8/2/2016
10
8/2/2016
Hologic 2016.
11
8/2/2016
• Tomo
• Acquires tomosynthesis images only
• TomoHD
• Acquires tomosynthesis images only
• Produces C-View images
• Combo
• Acquires 2D images
• Acquires tomosynthesis images
• ComboHD
• Acquires 2D images
• Acquires tomosynthesis images
• Produces C-View images Hologic 2016.
• 2D Contrast Acquisition
• 2D low-energy image
• 2D high-energy image
Hologic 2016.
QC Modes
Hologic 2016.
12
8/2/2016
DBT
Quality Control
Hologic 2016.
Medical Physics
Medical Equipment Evaluation (MEE) Testing
12 tests to be performed by
the medical physicist
Hologic 2016.
Medical Physics
QC Tests
12 tests to be performed by
the medical physicist
Hologic 2016.
13
8/2/2016
Medical Physics
QC Tests
Hologic 2016.
Technologist QC
Tests
12 tests to be performed by
the technologist
Hologic 2016.
Technologist QC
Tests
Hologic 2016.
14
8/2/2016
Alternative
Standard 9
• Category A
• Immediate corrective action is required
• Issue needs to be resolved before resuming imaging
• Category B
• Affects interpretation devices
• Interpretation device stays out of use until issue is resolved
• Alternative interpretation device may be used
• Imaging may continue
• Category C
• Issue needs to be resolved within a predefined time
window
• Imaging may continue up to the end of that time window
Hologic 2016.
Alternative
Standard 6
• Additional evaluations must be conducted after major
component change or repair
• Software changes are considered as major repairs
• Alternative Standard 6 allows manufacturers to:
• Evaluate software update for QC requirements
• Define a series of tests to be performed by the
technologist
Hologic 2016.
Hologic 2016.
15
8/2/2016
Hologic QC Manuals
MAN-01965 MAN-01476
R008 R001
MAN-03706
R002
• http://www.hologic.com/support/dimensions-3d-breast-tomosynthesis-
dimensions-2d-full-field-digital-mammography
• http://www.hologic.com/support/selenia-digital-mammography
Hologic 2016.
MAN-01965
• Applies to software version prior to v1.8.x
MAN-03706
• Applies to software version starting at v1.8.x
Important differences
• Appendix D: CNR Correction Tables
Hologic 2016.
Tomosynthesis Option
Tomosynthesis specific
tests are marked with an
icon
Icon indicates that a
special action is required
under tomosynthesis
NOTE: When testing FFDM only,
these instructions are ignored
Hologic 2016.
16
8/2/2016
Diagnostic Option
Hologic 2016.
Contrast Option
Hologic 2016.
• Tomo Geometry
Calibration
• Usually performed by
technologist
Hologic 2016. 51
17
8/2/2016
Quality Control
Tests
Hologic 2016.
CAUTION
Direct x-ray exposure of the image receptor may damage the receptor
The image receptor should be covered with lead or copper during testing for exposures
other than those required to qualify image quality
Following the test procedures in the Hologic QC Manual will ensure the safety of the
image receptor
Hologic 2016.
CAUTION
Hologic 2016.
18
8/2/2016
Hologic 2016.
2. COLLIMATOR ASSESSMENT
Hologic 2016.
Hologic 2016.
19
8/2/2016
Compression paddles
Manufactured as single pieces
3. ARTIFACT EVALUATION
Hologic 2016.
20
8/2/2016
Procedure Highlights
DICOM printer
Send an artificial flat field image to the printer
FFDM testing
Test all focus/filter combinations
(LFS/Rh; LFS/Ag; SFS/Rh; SFS/Ag; LFS Cu)
Preview image in full resolution
DBT testing
Test using middle projection
Preview image in full resolution
Hologic 2016.
Hologic 2016.
Procedure Highlights
21
8/2/2016
Hologic 2016.
Procedure Highlights
Hologic 2016.
22
8/2/2016
Procedure Highlights
Place the line pair phantom on
top of the 4 cm acrylic block
Rotate the line pair phantom 45°
Apply 15-20 lb of compression to
avoid vibration during DBT
Use the Flat Field view (no image
processing)
Resolution guidelines:
FFDM: > 7 lp/mm @ 45°
DBT: > 3 lp/mm @ 45°
Hologic 2016.
Procedure Highlights
Hologic 2016.
Hologic 2016.
23
8/2/2016
AEC positions
• Auto AEC: two ~1cm2 floating sensors in 5 x 14cm 2 area
• One of seven manual positions (marked on compression paddle)
AEC function
• kVp and filter parameters are determined by compression thickness and AEC technique tables
• kVp can be adjusted upwards if the exposure time will be too long
• Starting mAs is determined from short pre-exposure targeting a specific exposure index (EI)
• Final mAs is adjusted by CNR correction factor
Hologic 2016.
Procedure Highlights
Compression
Thickness
Compression thickness must be
set using the compression
display
FFDM testing
• Range of phantom thickness
• Different operating modes (i.e. mag)
• Exposure compensation steps
DBT testing
• Range of phantom thickness
Hologic 2016.
Hologic 2016.
24
8/2/2016
LFS
Hologic 2016.
SFS
Hologic 2016.
LFS
Hologic 2016.
25
8/2/2016
Calculation Example
Hologic 2016.
Hologic 2016.
Procedure Highlights
Hologic 2016.
26
8/2/2016
Procedure Highlights
Test AGD in all three modes
• FFDM
• Tomo
• Combo
Hologic AGD recommended
dose for ACR phantom
• FFDM: 1.2 mGy
• Tomo: 1.45 mGy
Performance criteria
• AGD < 3 mGy
Hologic 2016.
Hologic 2016.
Procedure Highlights
Hologic 2016.
27
8/2/2016
Hologic 2016.
Procedure Highlights
Wait until the image receptor
goes from Warming to
Ready status
Use ACR Phantom view to
overwrite compression
thickness to 4.2 cm
Hologic 2016.
Phantom Scoring
Score phantom on AWS display
Review image in full resolution
FFDM scoring
• 5 fibers, 4 specs, 4 masses
• Due to phantom variations a score of 4.5/4.0/3.5 is
acceptable providing SNR and high contrast
resolution tests pass
DBT scoring
• Scroll to the slice that puts the different elements
in focus
• 4 fibers, 3 specs, 3 masses
Hologic 2016.
28
8/2/2016
Hologic 2016.
Procedure Highlights
Hologic 2016.
Automatic Computation
Hologic 2016.
29
8/2/2016
Hologic 2016.
Procedure Highlights
Hologic 2016.
Hologic 2016.
30
8/2/2016
Procedure Highlights
Hologic 2016.
Procedure Highlights
3 2 1
Hologic 2016.
Computation of Ghosting
𝑚𝑒𝑎𝑛𝑅3 − 𝑚𝑒𝑎𝑛𝑅2
𝐺ℎ𝑜𝑠𝑡 =
𝑚𝑒𝑎𝑛𝑅1 − 𝑚𝑒𝑎𝑛𝑅2
Hologic 2016.
31
8/2/2016
Tomo-guided Biopsy
Hologic 2016.
2D
Tomo
Hologic 2016. 95
Affirm™
Interventional Add-On Device Utilizing Tomosynthesis Localization
Hologic 2016.
32
8/2/2016
Stereotactic Biopsy
• Take 2D scout exposure
• Acquire ±30° stereo pairs
• Use triangulation to determine x,y,z
lesion location
Tomo-guided Biopsy
• Take Tomo exposure
• Use tomo slices to determine x,y,z
lesion location
Hologic 2016. 97
Tomosynthesis procedure –
Target
Hologic 2016. 98
Tomosynthesis procedure -
Target
Hologic 2016.
33
8/2/2016
• 70 μm pixel size
• 20-49 kVp
• No anti-scatter grid
• 15° sweep tomo, 30° stereo
34
8/2/2016
Digital Mammography
QC Program:
TRAINING
Hologic 2016.
Required Training
Hologic 2016.
35
8/2/2016
Hologic 2016.
PRE-00023
PRE-00023
Hologic 2016.
PRE-00023
PRE-00023
Hologic 2016.
PRE-00023
PRE-00023
36
8/2/2016
Facility Certification
37
8/2/2016
Thank you!
Email:
[email protected]
38