DI Manual-1
DI Manual-1
1
`
List of Experiments
2
`
Aim:
To observe ECG wave form of subject (human Body) using lead I, II, III of standard bipolar
lead configuration and find the heart rate from it.
Apparatus Required:
ECG ring electrode, Trainer kit, connecting cables and different modules.
Theory:
As the heart undergoes depolarization, the electrical currents that are generated spread not
only within the heart, but also throughout the body. This electrical activity generated by the heart
can be measured by an array of electrodes placed on the body surface. The recorded tracing is
called an Electrocardiogram (ECG or EKG). A typical ECG tracing is shown below. The different
waves that comprise the ECG represent the sequence of depolarization and repolarization of the
atria and ventricles.
The P wave represents the wave of depolarization that spreads from the SA node
throughout the atria, and is usually 0.08 to 0.1 seconds (80 – 100ms) in duration.
The QRS complex represents ventricular depolarization. The duration of the QRS complex
is normally 0.06 to 0.1 seconds. This relatively short duration indicates that ventricular
depolarization normally occurs very rapidly.
3
`
Diagram:
Einthoven Triangle
4
`
BLOCK DIAGRAM:
Brady Amp
5
`
Waveforms:
Observation:
Lead –I
Result:
Thus ECG waveform of a subject (Human Body) has been observed using standard Bipolar Limb
Leads configuration
6
`
To stimulate ECG waveform using modules and to measure the amplitude and frequency
of the ECG signal.
APPARATUS REQUIRED:
Battery
ECG Amplifier
ECG Sensor
DSO Probe
Theory:
As the heart undergoes depolarization, the electrical currents that are generated spread
not only within the heart, but also throughout the body. This electrical activity generated by the
heart can be measured by an array of electrodes placed on the body surface. The recorded tracing
is called an Electrocardiogram (ECG or EKG). A typical ECG tracing is shown below. The
different waves that comprise the ECG represent the sequence of depolarization and repolarization
of the atria and ventricles
Waveform:
7
`
Block Diagram:
Ring
electrode
Observations:
P wave
QRS wave
ST Wave
Procedure:
Result:
The ECG waveform has been observed and the real time of the various parameters has been
measured and tabulated.
8
`
Apparatus Required:
Battery, peripheral pulse Amplifier, peripheral pulse sensor, DSO.
Theory:
The pulse sensor has an LED that transmit IR light. The absorption of the light passing
through the tissue is measured with the photo sensor The AC signal is obtained is due to the
pulsing of the arterial blood
Block Diagram:
peripheral
pulse
Sensor
9
`
Waveform:
Observation:
Amplitude
Time
Frequency
Procedure:
Result:
The pulse waveform was studied and the various parameter were tabulated.
10
`
AIM:
Apparatus Required:
Battery, EMG Amplifier, High Pass filter, And DSO, Surface electrode, EMG sensor
Theory:
EMG is an instrument used for recording the electrical activity of the muscles to determine
where the muscle is contracting or not of a displacing the action potential generated by the motor
units.
The electrical activity of the underlying muscles are measured using Ag-Agcl surface
electrodes. The EMG appears like a random noise waveform.
BLOCK DIAGRAM:
EMG
Electrode
WAVEFORM:
11
`
Observation:
Amplitude
Time
Frequency
PROCEDURD:
Connect the `modules according to the block diagram place the electrodes at
appropriate site on the subject.
RESULT:
12
`
Apparatus Required:
Theory:
EEG deals with recording and study of the electrical activity of the brain. The electrodes
are attached to the Skull of the patient and the brain waves are picked up and recorded. The
brainwaves are the summation if the action potentials of the neurons in the brain. The electrodes
in EEG are placed in standard position on the skull in an arrangement 10-20 electrodes system.
The EEG helps to diagnose the level of consciousness, sleep disorder, brain tumor and epilepsy
13
`
BLOCK DIAGRAM:
Amplifier
Electrode
Montage
selector
Filter Sensitivity
Chart Inkjet
Jack Box recorder
Oscilloscope Amplifier drive
Patient
15
14
`
Observation:
Delta Wave
Theta wave
Alpha Wave
Beta Wave
Procedure:
Connect the DSO cable in the appropriate slot to Obtain the brain wave
Result:
15
`
6. PCG Monitoring
Aim:
To build the PCG monitor and observe the heart sound.
Apparatus Required:
PCG amplifier, Battery, Low pass Filter, PCG Sensor, DSO
Theory:
The heart sounds are vibrations on sound due to vibrations, acceleration and declaration of
blood murmurs are sound or due to blood flow. There are four heart sound related to electric and
mechanical events of cardiac cycle. The first heart sound is regulated with the movements of blood
during ventricular systole. The second heart sound is a low frequency vibration associated with
declaration and reverse of flow in aorta with pulmonary artery and with the closure of semi lunar
valves. The second heart sound is coincident with completion of T- wave of ECG’s, the 3rd heart
sound attributed to sudden termination of rapid filling phase of ventricles, from atria and associated
vibration of ventricular muscle walls. The fourth of atrial heart sound which is not audible but can
be recorded by phonocardiogram and occurs when the atria contract and propel blood into
ventricles.
BLOCK DIAGRAM:
PCG
Sensor
16
`
Waveform:
Observations:
Sound 1
Sound 2
Procedure:
Result:
The heart sound is obtained and waveform representing the same is plotted.
17
`
7. AUDIOMETER
AIM:
To detect the hearing loss of the individual and measure the output dB level.
Apparatus required:
Audiometer, headphone.
Theory:
An audiometer is used for the identification of hearing loss is individual and quantitative
determination of degree and nature of such losses. It is an oscillation driving a pair of headphone
and its calibrate in terms of frequency and acoustic output . The frequency and output are
adjustable over the audio range. Audiometer are divided into pure tone audiometer and speech
audiometer.
Pure tone audiometer is used to obtain our conduction and one conduction threshold of
hearing loss.
BLOCK DIAGRAM:
Headphone
18
`
Observation:
Left hearing: dB
Right hearing: dB
Procedure:
3. Adjust masking level to a suitable level so that it does not cause discomfort to the
subject.
5. Adjust output dB level till the subject can detect the sound and indicate by response
switch.
Result:
To study the operation biokit, physiographic to capture and analysis the biomedical signals
like EMG,EEG and pulse rate.
Apparatus Required:
Data acquisition kit, software, to view graph various amplifier, pc, power supply.
Theory:
These systems are designed to monitor different parameter at a time. Hardware responsible
for physiological signal analysis and information display are a set of modules implemented interns
of microcomputer program. The various parameters that can be measured are pulse ECG, BP,
Temperature, respiration and EMG.
ECG:
ECG simulation produces the fundamentals of standard limb lead interpretation and
rhythm recognition in a easy to use state it also demonstrates ECG observation compression
PQRST wave.
Pulse Rate:
HB conc of deoxygenated HB
20
`
Block Diagram:
Pulse Sensor
10 Channel DAC Pulse Amplifier
Observation:
ECG
T(s) =
dV =
T =
V =
EMG
T(s) =
dV =
T =
V =
Pulse rate
T(s) =
dV =
T =
V =
Procedure:
Connect the physiographic kit with the system using RS232 cable
Connect ECG setup, EMG setup and pulse rate setup and connect the filter output to
each channel of the data acquisition kit.
21
`
Now you can view the output of ECG, EMG and pulse rate in the physiograph main
screen.
Result:
Thus the operation of biokit physiograph and various biomedical signals have been
analyzed and studied.
22
`
Apparatus Required:
Theory:
The primary function of the respiratory system are to supply oxygen and remove Co2 from
the tissues. The action of breathing is controlled by a molecular action a causing the volume of
the lung to increase and decrease and to control the Co2 content in the arterial blood
This rhythmic action provides a fairly good respiration rate. The respiration cycle is
accompanied by changes in the thoracle volume. These changes can be sensed by means a
displacement transducer interpolating a strain gauge (or) a variable resistive element. The
respiratory moments results in resistance changes in the gauge element.
Block Diagram:
Respiration
Sensor
23
`
Observations:
Waveform
Procedure:
Result:
The respiratory rate waveform is plotted. The respiration rate of the subject is breathes/min.
24
`
10.Telemetry System
AIM:
To study medical telemetry system for transmitting and receiving bio signals.
Apparatus Required:
Theory:
Medical telemetry is used to remotely track the vital science of ambulatory patient using
wireless telemetry physiological signal can be transported during transportation within the
hospital or outside. The system consists of several components including sensor, transmitters
and receiver
The 2 parts of ECG telemetry system and ECG amplifier and oscillator or modulator and an
ultra high frequency transmitter and the telemetry receiver with the demodulator to which a
recording device is connected.
The ECG Telemetry system operates in 450-470 Mhz ban which is well suitable for transmission
within a hospital.
Block Diagram:
FM receiver
Battery Demodulator LPF DSO
25
`
Procedure:
Result:
Medical telemetry system was studied .Waveforms in transmitter and receiver are
compared and verified and the output Waveforms is generated.
26
`
11. Electroocculogram
AIM:
To study the movement of the electrical activity associated with eye movements. It is
useful for monitoring eyeball movement in REM and Non-REM sleep.
Apparatus Required:
Data Acquisition unit, software to view physiography, EOG Amplifier, EOG Sensor,
LPF, PC, Power Supply
Theory:
BLOCK DIAGRAM:
EOG
Amplifier
27
`
Waveform:
Procedure:
Result:
28
`
12.Study of spirometer
AIM:
To measure the lung capacity and volume and to record the spirogram obtained from the
device
Apparatus Required:
Theory:
To measure the lung capacity & volume a instrument called a spirometer is used.
Basically the recorder obtained from the device is called a spirogram. Spirometer are calibrated
that can collect gas & make measurement of lung volume & capacity that can be expired by
adding a flow base dependent quantity can be measured. The addition of gas analyzer makes the
laboratory.
A numbers of force breathing test are carried out that uses the muscle power associated
with breathing & resistance of the airway among these are:
This is the total amount of air that can be forcibly expired as quickly as possible breath
29
`
This is the total amount of air to measure the maximum amount of air that can be exhaled &
inhaled in one min, measure in lit/min
Procedure:
Connect the spirometer device to the system with the help of es232 connector cable.
Now place the mouth piece in your mouth & proceed the various test like
Result:
Thus the entire three took has been performed & the functioning of the lung is analysed
30