Technical Specification-ECG-3010(V2.0)

The document outlines the specifications for the Biocare CardioPro ECG analysis program, detailing its signal processing capabilities, display features, recorder technology, software functionalities, and power supply options. It includes information on environmental requirements, external input/output specifications, and standard accessories. The ECG device supports simultaneous 12-lead acquisition, offers various measurement modes, and is designed for both adult and pediatric use.

Uploaded by

Poushi Adhikary
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
58 views8 pages

Technical Specification-ECG-3010(V2.0)

The document outlines the specifications for the Biocare CardioPro ECG analysis program, detailing its signal processing capabilities, display features, recorder technology, software functionalities, and power supply options. It includes information on environmental requirements, external input/output specifications, and standard accessories. The ECG device supports simultaneous 12-lead acquisition, offers various measurement modes, and is designed for both adult and pediatric use.

Uploaded by

Poushi Adhikary
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 8

Signal processing specification:

Biocare CardioPro ECG analysis program for adults and


ECG interpretation
pediatrics
Acquisition mode Simultaneous standard 12-lead acquisition
A/D conversion 12 bit
Sampling rate 1,000 samples/s
Dynamic range ± 9 mV
Resolution 0.3μV
Frequency response 0.05 Hz~150 Hz
Common mode
>89 dB
rejection
Input circuit Floating circuit input
Input impedance >50 M
Input CIR current ≤0.1 μA
Patient leakage
<10 μA
current
Time constant ≥3.2 s
Noise level <15 μVp-p
Sensitivity threshold ≤20 μV
Calibration voltage 1 mV±3 %
Sensitivity Four levels as (2.5,5,10,20) mm/mV,Auto Gain
Four levels as H50 d, H50 d 75Hz, H50 d 35Hz,
Filter setting
0.05-150Hz
AC filter 50Hz, 60Hz
Heart rate range 30~300 bpm
Special acquisition
Lead-off detection, lead overflow
functions
Acquisition time 3~12 seconds
Display and keyboard specifications:

Display type 3.8 inch LCD with backlight


Display resolution 320*240
Display data Waveforms, heart rate, clock, printing mode, filters,
sensitivity, paper speed, system prompt, alarming
messages, battery, power indicator, tone level
Waveform display 12-lead simultaneous display: 6*2 display mode;
mode
Keyboard type Silicon quick-access function keys;

Recorder specifications:

Recorder technology Thermal dot matrix word printing system


Recorder resolution 8 points/mm (perpendicular)
40 points/mm (horizontal, 25 mm/s)

Recorder speed (6.25, 12.5, 25, 50)mm/s

Recording paper 63 mm×30 m or 63 mm×20 m,roll paper


63 mm×75 mm×20 m,Z- fold paper

Standard software specifications:

Measurement and Supports measurement and interpretation with Biocare


interpretation CardioPro ECG analysis program for adults and
pediatrics
Measurement values HR, PR interval, QRS duration, QT/QTC interval,
P/QRS/T axis, RV5/SV1 amplitude, RV5+SV1
amplitude
Working mode Automatic mode, Manual mode, Rhythm mode
DEMO mode No

Rhythm analysis Single rhythm mode: 30-180 seconds

Arrhythmia mode No

Waveform frozen No
Report auto-saved No
Interpretation printout Selectable interpretation printing
Measurement display Selectable measurement display
Printing preview Preview the measurement before printing
Patient information No
setting
Language Chinese, English, Russian, Spanish
Local memory 20 files
capacity
Printing lead Printing out the lead connection diagram during the
connection diagram printing test

Report specifications:

Report type Only waveforms, Simple report, MVB report(Media


beat report)
Report format 3*4;
1*12, 1*12+1R;

External peripheral use:

SD card port 2 GB SD card for data transfer, standard


USB port Communication with ECG-1000 workstation software
on PC
LAN port NO
External signal input To input an signal from the oscilloscope to test the
waveform display quality on the ECG screen
External signal output To export an signal from ECG to oscilloscope to test the
waveform display quality

Power supply

Power supply AC / battery operation


AC power supply 100 V~240 V,50 Hz /60 Hz,50 VA
Battery power supply Rechargeable lithium battery,
14.8 V,1550 mAh
Battery capacity Supports more than 0.5 hours for continuous operation
Battery charge time Approximately 4 hours from total discharge
Fuse specification T1.6 A/250V (built-in the instrument)
DC time-lag fuse Self-recovery fuse of 3. 1 A/250V (built-in the
instrument)

Standard accessories
ECG cable 10-lead, TPU, Φ4mm banana plug, defi, IEC
Chest electrode Adult chest electrode,Φ4mm,6pcs/set
Adult limb electrode, Φ4mm, (2pcs big, 2pcs small,
Limb electrode
4pcs/set), Biocare LOGO
Power cord European standard
Grounding cable L=4m
Battery HYLB-947 1550mAh/14.8V
Thermal recording
Roll, 63mm*30m , NO LOGO
Paper
Paper roller ABS UL94V-0
Dimensions and weight

Length × width × height 268 mm×206 mm×67.5 mm


Weight: About 2.2 kg (with battery)
About 2 kg (without battery)

Environment requirements

1 Transportation
Environment temperature -20 ℃~+55 ℃
Relative humidity ≤95% (No condensation)
Air pressure 70 kPa~106 kPa
In accordance with the requirements stipulated in the contract order, the
transport process to prevent rain and sun.
2 Storage
Environment temperature -20 ℃~+55℃
Relative humidity: ≤95%(No condensation)
Air pressure 70 kPa~106 kPa
The packaging of ECG stored in the non-corrosive gases and
well-ventilated room.
3 Using
Environment temperature +5 ℃~+40 ℃
Relative humidity: ≤95%(No condensation)
Air pressure 86 kPa~106 kPa

External input and external output specifications

1 External input ( EXT)

Input impedance ≥100 kΩ


Sensitivity: 10 mm/V±5%

Input voltage range ±2.5 V

Input with the single ended mode

2 External output (CRO)

Output impedance ≤100 Ω

Sensitivity 1 V/mV±5%(at 10 mm/mV)

Output with the single ended mode

 All rights reserved by Biocare.

 Based on the need of product technical improvement or the file updates, we reserve the right

to modify the contents contained in this manual; if the change does not involve safety issues,

the contents are subject to amend without notification.