LABORATORY 2a
INTRODUCTION TO KEYSIGHT NETWORK ANALYZER AND BENCHVUE
SOFTWARE
Objective
i) To perform TRM calibration using the user-defined calibration kits
1. Connect both Port 1 and Port 2 with SMA RF coaxial m-m cable (1 m) each.
Note:
In this situation, for open measurement, the correction will be only up until the SMA
cable and not to the TRM PCB path.
3. Press Port x Open to start the calibration measurement (x denotes either Port 1 or Port
2).
4. Disconnect the OPEN calibration standard and replace it with a SHORT calibration
standard.
5. Press Port x Short to start the calibration measurement. (SHORT refer to REFLECT
in TRM PCB calibration kit as shown in Figure 1.)
6. Disconnect the SHORT calibration standard and replace it with a LOAD calibration
standard.
7. Press Port x Load to start the calibration measurement (LOAD refer to Match in TRM
PCB calibration kit as shown in Figure 1.).
9. Press Return.
12. Press Port 1-2 Thru to start the calibration measurement. (THRU refer to Through in
TRM PCB calibration kit as shown in Figure 1).
14. Press Done to terminate the full two-port calibration process. Upon pressing the key,
the calibration coefficients will be calculated and saved. The error correction function
will also be automatically enabled.
2. One-Port Measurement
Note: Go back to [Preset] > [OK]
a. Observe the trace on the Smith Chart in Figure 4 for (i) with the load removed
(open), (ii) the cable with matched load, and (iii) with a short (zero-ohm)
termination added to the cable, as shown in 2, ensure that there is marker at
the ends of each trace.
Note: Placing a marker on the trace will help you identify the frequency as
you move it along the trace.
1m RF
coaxial (iii) Short
cable (TRM Reflect)
2. Remove all terminations, leaving only the RF coaxial cable with the open end. Double
the Stop Frequency to 20MHz, and then double the frequency again to 40MHz.
NMK 31503 RF & MICROWAVE TECHNOLOGY SEM 1(2023/2024)
3. Two-Port Measurement
P1 P2
1m RF coaxial
In some steps cable
DUT
2-port
the DUT is
inserted at P1
1. Connect the VNA as shown in Figure 3, with the same 1m RF coaxial cable that was
used in the previous One-Port Measurement. Initially connect the RF coaxial cable to
both P1 and P2, without any DUT attached.
b. Compare this Insertion Loss (IL= −S21dB) with the Return Loss (RL=
−S11dB) plot.
NMK 31503 RF & MICROWAVE TECHNOLOGY SEM 1(2023/2024)
2. Double click on Keysight BenchVue Lab – Client icon to open the Benchvue
software. The window as Figure 5 will be displayed.
3. Click on the Network Analyzer icon on the right panel of the BenchVue
software. The window as Figure 6 will be displayed.
NMK 31503 RF & MICROWAVE TECHNOLOGY SEM 1(2023/2024)
4. On the left panel of BenchVue software show the Network Analyzer setting. Do the
same setting as in “2. One-Port Measurement” and “3. Two-Port Measurement”, then
observe the results. Perform 2 traces of S11 and S21 in one window.