The HG series hydrogen generators produce ultra-high purity hydrogen (99.99999%) through water electrolysis, eliminating the need for gas cylinders and offering significant cost savings. They feature a sophisticated control and safety system, simple installation, and meet all regulatory standards. Various models are available with different flow rates and power consumption, requiring minimal maintenance and using only deionized water.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
24 views1 page

HG200 CMC

The HG series hydrogen generators produce ultra-high purity hydrogen (99.99999%) through water electrolysis, eliminating the need for gas cylinders and offering significant cost savings. They feature a sophisticated control and safety system, simple installation, and meet all regulatory standards. Various models are available with different flow rates and power consumption, requiring minimal maintenance and using only deionized water.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 1

Hydrogen-Generators

• Eliminates costly dangerous gas cylinders


• Significant savings over cylinders
• Simple installation, no special requirements
• Constant purity with no fall off in performance
• Sophisticated control and safety system
• with automatic changeover switch
• Meets all standards and regulations

Hydrogen generating in ultra high purity

The HG series hydrogen generator produces pure hydrogen (and


oxygen as a byproduct) by the electrolysis of water. The hydrogen
(uhp quality 99,99999%) output flow is dependand on the model
purchased, and it will generate the exact amount of hydrogen flow
at a preset pressure (0…4 barg) as required by the attached
application.
Technical data
TECHNICAL DETAILS
Type HG 200
Output Pressure 0…4 bar (optional 0…6 bars)
The key element of the generator is an electrochemical cell Maximum Flow Rate 200 ml/min
assembly which contains a solid polymer electrolyte. The solid Electrical Requirements 230 V / 50 Hz
electrolyte material is in the form of a sheet of plastic. There are no optional 110 V
free acids or alkalines used. Deionized, destilled water is the only Power consumption 250 W
liquid contained in the unit; this must be periodically replenished as it Deionized Water < 10 µS / cm
is consumed. The generated hydrogen gas is accumulated in the (quality)
hydrogen/water reservoir and the desiccant housing. The pressure Ambient temperature +5°…+35°C
is controlled through the use of a pressure transducer and adjusted Outlet Port 1/8 "
at the front of the unit by service menu. The LC-display indicates Dimensions W x H x D 45 x 26 x 50 cm
the output pressure, and the hydrogen flow. The gas is Shipping Weight approx. 45 kg
dried by passing it through a drying nafion-tube and a desiccant
tower. The water flows from the internal water tanks to the Shipping weights:
electrochemical cell assembly. Inside the cell assembly, electrolysis
occurs, and positively charged hydrogen ions are transported HG 300 HG 600 HG 1200
across the solid polymer electrolyte along with excess water 50 kg 55 kg 65 kg
molecules. Negatively charged oxygen ions recombine to form
molecular oxygen. The oxygen saturated water is fed into a Sizes of housings
water/oxygen separator vent, where the water is returned to the
water tank, and the oxygen is vented to the atmosphere. Desktop version
The positively charged hydrogen ions recombine to form molecular 19”-rack-version
hydrogen. The generated hydrogen is made available at the output
port of the system. Residual hydrogen is vented to the atmosphere, Options
and excess water is returned to the water tank.
– Failure indication and alarm by gas- and watersupply failure
The quantity of hydrogen generated is directly proportional to the – Failure indication and alarm by powersupply failure
flow of current across the cell module. This current is carefully – Automatic gasline switch to cylinders by system failure
modulated by the electronic control circuitry that is an important part
of the generator. The electrical power supplied to the cell is Maintenance and spare parts
regulated to provide just enough power to create the required flow
rate of hydrogen. The hydrogen generators can be installed in The primary routine maintenance required for the hydrogen
the laboratory and do not classify the surrounding area as generators are:
dangerous.
– Replacing or regenerating the silicagel desiccant when
H2 Generator Models for large volume of H2 indicated by colour change
– Replacing the deionizer bags inside the generator (for your
Flow Power consumption own safety, we recommend the change once a year)
HG 300 300 ml/min 350 W
HG 600 600 ml/min 750 W Article Part No.
HG 1200 1200 ml/min 1250 W Deionizer bag HPSA 020

All Hydrogen Generators in 19”-housing available. Water consumption

cmc Instruments GmbH It is possible to produce 1,2 m³ hydrogen out of 1 ltr. deionized
Meß-, Regel- und Analysentechnik
Hauptstraße 388
water (55,6 mol H2O).
D-65760 Eschborn, Germany
fon: + 49 6173 / 32 00 78 To supply a GC-FID a volume of 30…40ml/min. of carrier gas is
fax: + 49 6173 / 6 50 50 needed. The supply of 6 FID’s with 2 ltrs. of water is for one week
e-mail: [email protected]
by 24-hours-working possible.
www.cmc-instruments.de