233-Elemental Imp. Procedures
233-Elemental Imp. Procedures
Speciation
The determination of the oxidation state, organic complex, or combination is termed speciation. Analytical procedures for speciation are not included in this
chapter, but examples may be found elsewhere in USP–NF and in the literature.
PROCEDURES
• COMPENDIAL PROCEDURES 1 AND 2
System standardization and suitability evaluation using applicable reference materials should be performed on the day of analysis.
Procedure and detection technique: Procedure 1 can be used for elemental impurities generally amenable to detection by inductively coupled plasma–atomic
(optical) emission spectroscopy (ICP–AES or ICP–OES). Procedure 2 can be used for elemental impurities generally amenable to detection by ICP–MS. Before
initial use, the analyst should verify that the procedure is appropriate for the instrument and sample used (procedural veri cation) by meeting the alternative
procedure validation requirements below.
Sample preparation: Forms of sample preparation include Neat, Direct aqueous solution, Direct organic solution, and Indirect solution. The selection of the
appropriate sample preparation depends on the material under test and is the responsibility of the analyst. When a sample preparation is not indicated in the
monograph, an analyst may use any of the following appropriately validated preparation procedures. In cases where spiking of a material under test is necessary
to provide an acceptable signal intensity, the blank should be spiked with the same Target elements, and where possible, using the same spiking solution.
Standard solutions may contain multiple Target elements. [NOTE—All liquid samples should be weighed.]
Neat: Used for liquids or alternative procedures that allow the examination of unsolvated samples.
Direct aqueous solution: Used when the sample is soluble in an aqueous solvent.
Direct organic solution: Used where the sample is soluble in an organic solvent.
Indirect solution: Used when a material is not directly soluble in aqueous or organic solvents. Total metal extraction is the preferred sample preparation
approach to obtain an Indirect solution. Digest the sample using the Closed vessel digestion procedure provided below or one similar to it. The sample
preparation scheme should yield su cient sample to allow quanti cation of each element at the limit speci ed in the corresponding monograph or chapter.
Closed vessel digestion: This sample preparation procedure is designed for samples that must be digested in a Concentrated acid using a closed vessel
digestion apparatus. Closed vessel digestion minimizes the loss of volatile impurities. The choice of a Concentrated acid depends on the sample matrix. The
use of any of the Concentrated acids may be appropriate, but each introduces inherent safety risks. Therefore, appropriate safety precautions should be used at
all times. [NOTE—Weights and volumes provided may be adjusted to meet the requirements of the digestion apparatus used.]
An example procedure that has been shown to have broad applicability is the following. Dehydrate and predigest 0.5 g of primary sample in 5 mL of freshly
prepared Concentrated acid. Allow to sit loosely covered for 30 min in a fume hood. Add an additional 10 mL of Concentrated acid, and digest, using a closed
vessel technique, until digestion or extraction is complete. Repeat, if necessary, by adding an additional 5 mL of Concentrated acid. [NOTE—Where closed
vessel digestion is necessary, follow the manufacturer’s recommended procedures to ensure safe use.]
Alternatively, leachate extraction may be appropriate with justi cation following scienti cally validated metal disposition studies, which may include animal
studies, speciation, or other means of studying disposition of the speci c metal in the drug product.
Reagents: All reagents used for the preparation of sample and standard solutions should be free of elemental impurities, in accordance with Plasma
Spectrochemistry 〈730〉.
• PROCEDURE 1: ICP–OES
Standardization solution 1: 1.5J of the Target element(s) in a Matched matrix
Standardization solution 2: 0.5J of the Target element(s) in a Matched matrix
Sample stock solution: Proceed as directed in Sample preparation above. Allow the sample to cool, if necessary. For mercury determination, add an appropriate
stabilizer.
Sample solution: Dilute the Sample stock solution with an appropriate solvent to obtain a nal concentration of the Target elements at NMT 1.5J.
Blank: Matched matrix
Elemental spectrometric system
(See Plasma Spectrochemistry 〈730〉.)
Mode: ICP
Detector: Optical detection system
Rinse: Diluent used
Standardization: Standardization solution 1, Standardization solution 2, and Blank
System suitability
Sample: Standardization solution 1
Suitability requirements
Drift: Compare results obtained from Standardization solution 1 before and after the analysis of the Sample solution.
Suitability criteria: NMT 20% for each Target element. [NOTE—If samples are high in mineral content, rinse system well before introducing the Sample in
order to minimize carryover.]
Analysis: Analyze according to the manufacturer's suggestions for program and wavelength. Calculate and report results on the basis of the original sample size.
[NOTE—Appropriate measures must be taken to correct for matrix-induced interferences (e.g., wavelength overlaps).]
• PROCEDURE 2: ICP–MS
Standardization solution 1: 1.5J of the Target element(s) in a Matched matrix
Standardization solution 2: 0.5J of the Target element(s) in a Matched matrix
https://online.uspnf.com/uspnf/document/GUID-3B140F3B-B8D0-4E2A-ACB5-C91968146674_2_en-US 1/3
2/22/2019 USP-NF
Sample stock solution: Proceed as directed for Sample preparation above. Allow the sample to cool, if necessary. For mercury determination, add an appropriate
stabilizer.
Sample solution: Dilute the Sample stock solution with an appropriate solvent to obtain a nal concentration of the Target elements at NMT 1.5J.
Blank: Matched matrix
Elemental spectrometric system
(See Plasma Spectrochemistry 〈730〉.)
Mode: ICP. [NOTE—An instrument with a cooled spray chamber is recommended. (A collision cell or reaction cell may also be bene cial.)]
Detector: Mass spectrometer
Rinse: Diluent used
Standardization: Standardization solution 1, Standardization solution 2, and Blank
System suitability
Sample: Standardization solution 1
Suitability requirements
Drift: Compare results obtained from Standardization solution 1 before and after the analysis of the Sample solution.
Suitability criteria: Drift NMT 20% for each Target element. [NOTE—If samples are high in mineral content, rinse system well before introducing the Sample in
order to minimize carryover.]
Analysis: Analyze according to the manufacturer's suggestions for program and m/z. Calculate and report results based on the original sample size. [NOTE—
Appropriate measures must be taken to correct for matrix-induced interferences (e.g., argon chloride interference with arsenic determinations).]
QUANTITATIVE PROCEDURES
The following section de nes the validation parameters for the acceptability of alternative quantitative procedures. Meeting these requirements must be
demonstrated experimentally, using an appropriate system suitability procedure and reference materials. Meeting these requirements demonstrates that the
procedure is equivalent to the compendial procedure for the purpose of quantifying the Target elements.
• ACCURACY
Standard solutions: Prepare solutions containing the Target elements at concentrations ranging from 50% to 150% of J, using appropriate reference materials.
Test samples: Prepare samples of the material under test spiked with appropriate reference materials before any sample preparation steps (digestion or
solubilization) at concentrations ranging from 50% to 150% of J for each Target element.
Acceptance criteria
Spike recovery: 70%–150% for the mean of three replicate preparations at each concentration
• PRECISION
Repeatability
Test samples: Six independent samples of material under test (taken from the same lot) spiked with appropriate reference materials for the Target element(s)
at the indicated level
Acceptance criteria
Relative standard deviation: NMT 20% (N = 6) for each Target element
Intermediate precision (ruggedness)
Perform the Repeatability analysis again either on a different day, with a different instrumentation, with a different analyst, or a combination thereof.
Combine the results of this analysis with the Repeatability analysis so the total number of analyses is 12.
Acceptance criteria
Relative standard deviation: NMT 25% (N = 12) for each Target element
https://online.uspnf.com/uspnf/document/GUID-3B140F3B-B8D0-4E2A-ACB5-C91968146674_2_en-US 2/3
2/22/2019 USP-NF
• SPECIFICITY
The procedure must be able to unequivocally assess (see 〈1225〉) each Target element in the presence of components that may be expected to be present,
including other Target elements, and matrix components.
• LIMIT OF QUANTITATION, RANGE, AND LINEARITY
Demonstrated by meeting the Accuracy requirement.
GLOSSARY
Concentrated acid: Concentrated ultra-pure nitric, sulfuric, hydrochloric, or hydro uoric acids or Aqua regia
Aqua regia: Aqua regia is a mixture of concentrated hydrochloric and nitric acids, typically at ratios of 3:1 or 4:1, respectively.
Matched matrix: Solutions having the same solvent composition as the Sample solution. In the case of an aqueous solution, Matched matrix would indicate that the
same acids, acid concentrations, and mercury stabilizer are used in both preparations.
Target elements: Elements with the potential of being present in the material under test. Include arsenic (As), cadmium (Cd), lead (Pb), and mercury (Hg) in the target
element evaluation when testing is done to demonstrate compliance. Target elements should also include any elements that may be added through material processing
or storage.
Target limit or Target concentration: The acceptance value for the elemental impurity being evaluated. Exceeding the Target limit indicates that a material under test
exceeds the acceptable value. The determination of compliance is addressed in other chapters. [NOTE—When applying this chapter to Elemental Impurities—Limits 〈232〉
and Elemental Contaminants in Dietary Supplements 〈2232〉, Target limits can be approximated by dividing the Daily Dose PDEs by the maximum daily dose for the Drug
Product Analysis Option in 〈232〉 or the Daily Serving PDE divided by the maximum daily serving size in 〈2232〉.]
J: The concentration (w/w) of the element(s) of interest at the Target limit, appropriately diluted to the working range of the instrument. For example, if the target
elements are lead and arsenic for an analysis of an oral solid drug product with a daily dose of 10 g/day using inductively coupled plasma–mass spectrometry (ICP–
MS), the target limit for these elements would be 0.5 µg/g and 1.5 µg/g (see Table 2 in 〈232〉). However, in this case, the linear dynamic range of the ICP–MS is known
to extend from 0.01 ng/mL to 0.1 µg/mL for these elements. Therefore, a dilution factor of at least 1:100 is required to ensure that the analysis occurs in the linear
dynamic range of the instrument. J would thus equal 5 ng and 15 ng/mL for lead and arsenic, respectively, when the dilution factor is added.
Appropriate reference materials: Where Appropriate reference materials are speci ed in the chapter, certi ed reference materials (CRM) from a national metrology
institute (NMI), or reference materials that are traceable to the CRM of an NMI should be used. An example of an NMI in the United States is the National Institute of
Standards and Technology.
Auxiliary Information- Please check for your question in the FAQs before contacting USP.
<233> ELEMENTAL IMPURITIES - PROCEDURES Kahkashan Zaidi GCCA2015 General Chapters-Chemical Analysis
Principal Scienti c Liaison 2015
+1 (301) 816-8269
Page Information:
USP42-NF37 - 6572
USP41-NF36 - 6151
USP40-NF35 - 298
https://online.uspnf.com/uspnf/document/GUID-3B140F3B-B8D0-4E2A-ACB5-C91968146674_2_en-US 3/3