AMI_Phosphate-II_01
AMI_Phosphate-II_01
Operator’s Manual
A-96.250.641 / 290312
Customer Support
SWAN and its representatives maintain a fully trained staff of technical specialists
around the world. For any technical question, contact your nearest
SWAN representative, or the manufacturer:
SWAN ANALYTISCHE INSTRUMENTE AG
Studbachstrasse 13
8340 Hinwil
Switzerland
Internet: www.swan.ch
E-mail: [email protected]
Document Status
ID: A-96.250.641
Revision Issue
Table of Contents
1. Safety Instructions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.1. Warning Notices. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.2. General Safety Regulations . . . . . . . . . . . . . . . . . . . . . . . 6
1.3. Restrictions for use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2. Product Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.1. Instrument Specification. . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.2. Instrument Overview AMI Phosphate-II . . . . . . . . . . . . . . 15
2.3. Instrument Overview AMI Phosphate-II B . . . . . . . . . . . . 16
3. Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.1. Installation Check List . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.2. Mounting of Instrument Panel . . . . . . . . . . . . . . . . . . . . . 18
3.3. Connecting Sample and Waste . . . . . . . . . . . . . . . . . . . . 18
3.4. Electrical Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.5. Connection Diagram Phosphate-II . . . . . . . . . . . . . . . . . . 21
3.6. Connection Diagram Phosphate-II B . . . . . . . . . . . . . . . . 22
3.7. Install the Cleaning Module . . . . . . . . . . . . . . . . . . . . . . . 23
3.8. Install 2nd Sample Stream . . . . . . . . . . . . . . . . . . . . . . . . 24
3.9. Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
3.10. Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
3.11. Relay Contacts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
3.11.1 Alarm Relay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
3.11.2 Relay Contacts 1 and 2 . . . . . . . . . . . . . . . . . . . . . . . . . . 27
3.12. Signal Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
3.12.1 Signal Output 1 and 2 (current outputs) . . . . . . . . . . . . . . 29
3.12.2 Signal Output 3 (optional) . . . . . . . . . . . . . . . . . . . . . . . . 29
3.13. Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
4. Instrument Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
4.1. Establish Sample Flow. . . . . . . . . . . . . . . . . . . . . . . . . . . 32
4.2. Fill / Flush Reagent System . . . . . . . . . . . . . . . . . . . . . . . 33
4.3. Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
5. Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
5.1. Function of the Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
5.2. Measured Values and Symbols on the Display . . . . . . . . 35
5.3. Software Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
5.4. Changing Parameters and Values . . . . . . . . . . . . . . . . . . 38
A-96.250.641 / 230312 1
AMI Phosphate II
6. Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
6.1. Maintenance Schedule. . . . . . . . . . . . . . . . . . . . . . . . . . . 39
6.2. Stop of Operation for Maintenance . . . . . . . . . . . . . . . . . 40
6.3. Refill / replace Reagents . . . . . . . . . . . . . . . . . . . . . . . . . 41
6.3.1 Reagents for Measuring Phosphate. . . . . . . . . . . . . . . . . 41
6.4. Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
6.5. Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
6.6. Cleaning the protective Filter . . . . . . . . . . . . . . . . . . . . . . 45
6.7. Cleaning the Photometer . . . . . . . . . . . . . . . . . . . . . . . . . 46
6.8. Cleaning the Flow Cell . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
6.8.1 Disassemble the Flow Cell . . . . . . . . . . . . . . . . . . . . . . . . 47
6.8.2 Assemble the Flow Cell . . . . . . . . . . . . . . . . . . . . . . . . . . 49
6.9. Cleaning the solenoid valve . . . . . . . . . . . . . . . . . . . . . . . 50
6.10. Tube Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
6.10.1 Changing Pump Tubes. . . . . . . . . . . . . . . . . . . . . . . . . . . 52
6.10.2 Tube Numbering. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
6.11. Replacing Fuses. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
6.12. Longer Stop of Operation . . . . . . . . . . . . . . . . . . . . . . . . . 56
7. Error List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
8. Program Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
8.1. Messages (Main Menu 1). . . . . . . . . . . . . . . . . . . . . . . . . 61
8.2. Diagnostics (Main Menu 2). . . . . . . . . . . . . . . . . . . . . . . . 62
8.3. Maintenance (Main Menu 3). . . . . . . . . . . . . . . . . . . . . . . 63
8.4. Operation (Main Menu 4) . . . . . . . . . . . . . . . . . . . . . . . . . 64
8.5. Installation (Main Menu 5) . . . . . . . . . . . . . . . . . . . . . . . . 65
9. Program List and Explanations . . . . . . . . . . . . . . . . . . 67
1 Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
2 Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
3 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
4 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
5 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
10. Material Safety Data sheets. . . . . . . . . . . . . . . . . . . . . . 82
10.1. OXYCON ON-LINE phosphate reagent 1 . . . . . . . . . . . . 82
10.2. OXYCON ON-LINE phosphate reagent 2 . . . . . . . . . . . . 86
11. Default Values. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
12. Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
13. Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
2 A-96.250.641 / 230312
AMI Phosphate-II
Safety Instructions
1. Safety Instructions
General The instructions included in this section explain the potential risks
associated with instrument operation and provide important safety
practices designed to minimize these risks.
If you carefully follow the information contained in this section, you
can protect yourself from hazards and create a safer work environ-
ment.
More safety instructions are given throughout this manual, at the
respective locations where observation is most important.
Strictly follow all safety instructions in this publication.
Target Operator: Qualified person who uses the equipment
audience for its intended purpose.
Instrument operation requires thorough knowledge of applications,
instrument functions and software program as well as all applicable
safety rules and regulations.
OM Location The AMI Operator’s Manual shall be kept in proximity of the instru-
ment.
Qualification, To be qualified for instrument installation and operation, you must:
Training read and understand the instructions in this manual as well
as the Material Safety Data Sheets.
know the relevant safety rules and regulations.
A-96.250.641 / 290312 3
AMI Phosphate-II
Safety Instructions
Corrosive
Harmful to health
Flammable
Warning general
4 A-96.250.641 / 290312
AMI Phosphate-II
Safety Instructions
ATTENTION ATTENTION
Symbols
The attention symbol indicates the possibility of equipment
damage, malfunctions or incorrect process results if instructions
are not followed.
Attention general
Safety goggles
Safety gloves
A-96.250.641 / 290312 5
AMI Phosphate-II
Safety Instructions
WARNING
WARNING
6 A-96.250.641 / 290312
AMI Phosphate-II
Safety Instructions
WARNING
ATTENTION
Never loosen pump tubes when flow is on. Otherwise you pol-
lute the reagents.
A-96.250.641 / 290312 7
AMI Phosphate-II
Product Description
2. Product Description
Application This manual describes the functions of the analyzers:
AMI Phosphate-II
AMI Phosphate-II B
Both analyzers are complete monitoring systems for the automatic
continuous measurement of ortho-Phosphate. ortho-Phosphate
can be found in many applications like corrosion protection in sani-
tary systems and boilers or as additive to detergents. On-line mea-
surement is used in quality control of drinking water, waste water
plants and thermal power plants.
Measuring The measurement is based on the molybdenum blue colorimetric
principle method according to APHA 4500-P E. and on EN ISO 6878. When
the reaction product of o-phosphate with ammonium-molybdate is
reduced with ascorbic acid, the intensely colored molybdenum blue
is formed. The color intensity is proportional to the o-phosphate
concentration of the sample and is measured photometrically at
815 nm.
Programmable The duration of a measuring interval can be set to:
Measuring 10 min (not available if “2 Channels” is selected)
Intervals 15 min
20 min
30 min
Independent of the programmed measuring interval, the measure-
ment time of a sample takes 7 minutes.
Signal Two signal outputs programmable for measured values (freely scal-
Outputs able, linear or bilinear) or as continuous control output (control pa-
rameters programmable).
Current loop: 0/4–20 mA
Maximal burden: 510
Third signal output with the same specifications as option.
(Only possible if no communication interface is used.)
Relay Two potential-free contact programmable as limit switches for mea-
suring values, controllers or timer for system cleaning with auto-
matic hold function.
Maximum load: 1 A/250 VAC
8 A-96.250.641 / 290312
AMI Phosphate-II
Product Description
WARNING
A-96.250.641 / 290312 9
AMI Phosphate-II
Product Description
On-line The sample flows through the sample inlet [K] and the filter vessel
operation [J] into the constant head [A]. Adjust the flow regulating valve so
that always a small part of the sample flows through the overflow
tube [B] into the drain [R]. This adjustment ensures a sufficient
sample flow through the measuring chamber of the photometer [M].
If no measurement takes place, the sample flows through the outlet
of the photometer where it will be aerated through air inlet [H] to
generate bubbles. Then the sample flows through the bubble coun-
ter [P] into the waste [Q].
If a measuring cycle starts:
1 A zero measurement with the sample is performed before the
reagents are added.
2 The peristaltic pump [C] pumps the reagents [N] and [O] into the
mixing chamber [G] where they are mixed together with the
sample and then flowing through the photometer [M].
3 The solenoid valve [I] will be activated to close the inlet of the
photometer.
4 The sample remains in the photometer for 7 minutes. During
this time a reaction with the reagents takes place.
5 After the 7 min have elapsed a second measurement is carried
out and the o-phosphate concentration is calculated.
6 If the measurement has been finished, the solenoid valve will be
deactivated to open the inlet of the photometer.
7 The sample flows through the outlet of the photometer where it
will be aerated to generate bubbles.
8 The sample flows through the bubble detector [P] and into the
waste [Q].
Calibration Close the flow regulating valve to drain the constant head. Prepare
a standard solution with deionized water and fill it into the constant
head. A normal measuring cycle is then initialized, the measured
value is used to calculate the photometer's slope. For details see
Calibration, S. 44.
10 A-96.250.641 / 290312
AMI Phosphate-II
Product Description
Fluidics
C D
3 1 E F
4 2 G
H I J K
3 4
L L M
N O P
Q R
A-96.250.641 / 290312 11
AMI Phosphate-II
Product Description
ATTENTION
12 A-96.250.641 / 290312
AMI Phosphate-II
Product Description
13 mm / ½”
AMI Phosphate-II
824 mm / 32 7/16”
850 mm / 33½”
4 x dia. 6.5 mm / ¼”
30 mm / 1 3 /16”
A-96.250.641 / 290312 13
AMI Phosphate-II
Product Description
400 mm / 15¾”
13 mm / ½”
374 mm / 14¾”
13 mm / ½”
AMI Phosphate-II
824 mm / 32 7/16”
850 mm / 33½”
4 x dia. 6.5 mm / ¼”
14 A-96.250.641 / 290312
AMI Phosphate-II
Product Description
F
G
O
H
I
N
J
M
K
L
A-96.250.641 / 290312 15
AMI Phosphate-II
Product Description
F
O G
H
I
J
M
K
L
16 A-96.250.641 / 290312
AMI Phosphate-II
Installation
3. Installation
A-96.250.641 / 290312 17
AMI Phosphate-II
Installation
Waste Connect the 1/2” tubes to the nozzle of the waste funnels and place
them into an atmospheric drain of sufficient capacity.
A Tube from photometer
A
C B Waste (photometer)
C Tube from constant head
D Drain (constant head)
B D
WARNING
the sample which flows after the measurement through the out-
let of the photometer contains Hexammonium heptamolybdate
4-hydrate.
At no means recirculate it into the water system.
18 A-96.250.641 / 290312
AMI Phosphate-II
Installation
WARNING
A B C
WARNING
A-96.250.641 / 290312 19
AMI Phosphate-II
Installation
WARNING
WARNING
20 A-96.250.641 / 290312
AMI Phosphate-II
Installation
WARNING
Use only the terminals shown in this diagram, and only for the
mentioned purpose. Use of any other terminals will cause short
circuits with possible corresponding consequences to material
and personnel.
A-96.250.641 / 290312 21
AMI Phosphate-II
Installation
WARNING
Use only the terminals shown in this diagram, and only for the
mentioned purpose. Use of any other terminals will cause short
circuits with possible corresponding consequences to material
and personnel.
22 A-96.250.641 / 290312
AMI Phosphate-II
Installation
A-96.250.641 / 290312 23
AMI Phosphate-II
Installation
A
B
D
E
G 2 1
H
I
J
K
L
24 A-96.250.641 / 290312
AMI Phosphate-II
Installation
Electrical
Connection
WARNING
Use one of the PG7 cable glands to feed the cable of the solenoid
valve into the AMI transmitter housing.
A B
A-96.250.641 / 290312 25
AMI Phosphate-II
Installation
B
C
ATTENTION
26 A-96.250.641 / 290312
AMI Phosphate-II
Installation
3.10. Input
Note: Use only potential-free (dry) contacts.
• Terminals 30/31 at AMI Phosphate-II transmitter
• Terminals 9/10 at AMI Phosphate-II B peristaltic pump
If signal output is set to hold, measurement is interrupted if input is
active.
For programming see menu 5.3.4, S. 82
Terminals Description
a) usual use
A-96.250.641 / 290312 27
AMI Phosphate-II
Installation
ATTENTION
28 A-96.250.641 / 290312
AMI Phosphate-II
Installation
A-96.250.641 / 290312 29
AMI Phosphate-II
Installation
3.13. Interface
Interface Terminal 50, 52, 53
RS232 The AMI Interface RS232 PCB is used for Logger down load and
Firmware up load. For detailed information see the corresponding
manual “AMI RS232 Interface”.
ON
PB-DPV1 V5.5
Profibus
81.420.020
30 A-96.250.641 / 290312
AMI Phosphate-II
Instrument Setup
4. Instrument Setup
After installation according to the checklist proceed as following:
C
D
4 Turn on sample flow and wait until the flow cell is completely
filled.See Establish Sample Flow, S. 32
5 Switch on power.
6 Fill system. See Fill / Flush Reagent System, S. 33
7 Let the instrument run continuously for 1 hour.
A-96.250.641 / 290312 31
AMI Phosphate-II
Instrument Setup
A
A Cover
B Outer tube
B C Flow cell block
D Flow regulating valve
Level E Filter
F Filter vessel
C
D
WARNING
32 A-96.250.641 / 290312
AMI Phosphate-II
Instrument Setup
Service 3.2.2
Verification
Fill System Enter Fill System 3.2.2.5
Cleaning
Progress
Progress
Done
Exit
Exit
Exit
Exit
4.3. Programming
Programming Program all parameters for external devices (interface, recorders,
etc.) Program all parameters for instrument operation (Phosphate,
limits, alarms). SeeProgram Overview, S. 64 and for explanations,
see Program List and Explanations, S. 70
A-96.250.641 / 290312 33
AMI Phosphate-II
Operation
5. Operation
Exit Enter
Program
RUN 14:10:45 Main Menu 1
Access, Exit Enter
R1 Messages
2.42 ppm
Diagnostics
R2
Maintenance
Exit Operation
52 B/s 25.4°C Installation
34 A-96.250.641 / 290312
AMI Phosphate-II
Operation
R1
F
R2 21.5 ppm
H 23 b/s G
A-96.250.641 / 290312 35
AMI Phosphate-II
Operation
Display when Only applicable with AMI Phosphate-II B and installed 2nd sample
Operating with stream option.
two Sample A B C D E
Streams
RUN 15:20:18
R1
PO4 1 21.5 ppm F
R2
PO4 2 21.5 ppm ~
H 23 b/s G
36 A-96.250.641 / 290312
AMI Phosphate-II
Operation
Main Menu 1
Messages
Diagnostics
Maintenance
Operation
Installation
Messages 1.1
Menu 1: Messages
Pending Errors Reveals pending errors as well as an event history
Maintenance List (time and state of events that have occurred at an ear-
Message List lier point of time) and maintenance requests.
It contains user relevant data.
Diagnostics 2.1
Menu 2: Diagnostics
Identification Provides user relevant instrument and sample data.
Sensors
Sample
I/O State
Interface
Maintenance 3.1
Menu 3: Maintenance
Calibration For instrument calibration, service, relay and signal
Service output simulation and to set the instrument time.
Simulation It is used by the service personnel.
Set Time 01.01.05 16:30:00
Cleaning
Operation 4.1
Menu 4: Operation
Sensors Subset of menu 5 - installation, but process-related.
Relay Contacts User relevant parameters that might need to be modi-
Logger fied during daily routine. Normally password protected
and used by the process-operator.
Installation 5.1
Menu 5: Installation
Sensors For initial instrument set up by SWAN authorized per-
Signal Outputs sons, to set all instrument parameters. Can be pro-
Relay Contacts tected by means of a password.
Miscellaneous
Interface
A-96.250.641 / 290312 37
AMI Phosphate-II
Operation
Sensors 4.1.3
Log interval Yes highlighted, press Enter to save
Save ? event driven
Clear logger no the new parameter. The system
Yes reboots, the new parameter is set.
No
Changing
value Alarm Phosphate. 5.3.1.1.1
Alarm High 5.00 ppm Select parameter and press Enter
Alarm Low 0.00 ppm
Hysteresis 0.10 ppm
Delay 5 Sec to set required value.
38 A-96.250.641 / 290312
AMI Phosphate-II
Maintenance
6. Maintenance
WARNING
A-96.250.641 / 290312 39
AMI Phosphate-II
Maintenance
40 A-96.250.641 / 290312
AMI Phosphate-II
Maintenance
ATTENTION
Before refilling the reagents, rinse the container with demin. wa-
ter.
A-96.250.641 / 290312 41
AMI Phosphate-II
Maintenance
WARNING
General Fill the containers with 2 liters of demineralized water. Add the pre-
scribed reagents and mix to complete dissolution.
Reagent 1 First dissolve content of one bucket reagent 1. Only then measure
100 ml sulfuric acid and add carefully while gently mixing.
Reagent 2 Dissolve content of one bucket reagent 2. Some foam forms on the
surface, this is normal and does not disturb the measurement
42 A-96.250.641 / 290312
AMI Phosphate-II
Maintenance
6.4. Verification
The “Verification kit for AMI Photometer” is available as an
accessory. An optical window with a precisely determined
absorbance value is placed into the light beam of the pho-
tometer. The actual measured absorbance will be compared
to the reference value labeled on each kit.
Set reference Prior to performing a verification the reference value, e.g. 0.255,
value: needs to be set in menu 5.1.1 (Installation\Sensors\Ref. Verifica-
tion).
Verification Navigate to the menu 3.2.1 (Maintenance\Service\Verification) and
procedure: follow the dialog on the screen.
A-96.250.641 / 290312 43
AMI Phosphate-II
Maintenance
6.5. Calibration
Select Menu 3.1 (Maintenance/Calibration) and follow the instruc-
tions on the display.
Calibration 3.1.1
Wait till calibration is
finished
Progress
Calibration 3.1.1
Calibration completed
Wait till flow cell is empty
Calibration 3.1.1
Operation completed
Open sample tap
Slope 1.006
<Enter> to save
Press <Enter> to save the value or leave the menu with <Exit>.
44 A-96.250.641 / 290312
AMI Phosphate-II
Maintenance
Normally the filter in your sample supply line will retain most debris.
If the protective filter shows deposits, proceed as follows:
A-96.250.641 / 290312 45
AMI Phosphate-II
Maintenance
46 A-96.250.641 / 290312
AMI Phosphate-II
Maintenance
ATTENTION
Level
D
E
A-96.250.641 / 290312 47
AMI Phosphate-II
Maintenance
WARNING
48 A-96.250.641 / 290312
AMI Phosphate-II
Maintenance
D
E
F
A-96.250.641 / 290312 49
AMI Phosphate-II
Maintenance
50 A-96.250.641 / 290312
AMI Phosphate-II
Maintenance
ATTENTION
Clean the white plate (G) and membrane (H) with clean water
only.
A-96.250.641 / 290312 51
AMI Phosphate-II
Maintenance
The long tubes, connected to the left (suction) side, leads to the re-
agent container, the short tubes, connected to the right (delivery)
side, leads to the flow cell. Tube Numbering, S. 54
ATTENTION
WARNING
Never let the pump run during any manipulation, e.g replacing
reagent tubes, removing or mounting of the occlusion frame.
52 A-96.250.641 / 290312
AMI Phosphate-II
Maintenance
WARNING
2 Relax the pump tubes by turning the occlusion frames (B) coun-
ter-clockwise.
3 Remove the occlusion frames (B) from the rotor (C) by pulling
the complete occlusion frame out of the holder.
4 Dismount all FEP tubes from the occlusion frame.
5 Exchange the occlusion frames.
6 Install all FEP tubes. See Connecting Sample and Waste, S. 18
A-96.250.641 / 290312 53
AMI Phosphate-II
Maintenance
C
D
3 1
4 2
3 4
N O
Tube
Nr. from to
1 Pump (C): rear frame, delivery side Flow cell (D), input 1
2 Pump (C): front frame, delivery side Flow cell (D), input 2
54 A-96.250.641 / 290312
AMI Phosphate-II
Maintenance
WARNING
Find and repair the cause for the short circuit before replacing the
fuse.
Use tweezers or needle-nosed pliers to remove the defective fuse.
Use original fuses provided by SWAN only.
A B C D
A-96.250.641 / 290312 55
AMI Phosphate-II
Maintenance
56 A-96.250.641 / 290312
AMI Phosphate-II
Error List
7. Error List
Error
Non-fatal Error. Indicates an alarm if a programmed value is ex-
ceeded.
Such Errors are marked E0xx (bold and black).
<Enter> to Acknowledge
A-96.250.641 / 290312 57
AMI Phosphate-II
Error List
58 A-96.250.641 / 290312
AMI Phosphate-II
Error List
A-96.250.641 / 290312 59
AMI Phosphate-II
Error List
60 A-96.250.641 / 290312
AMI Phosphate-II
Program Overview
8. Program Overview
For explanations about each parameter of the menus see Program
List and Explanations, S. 70
A-96.250.641 / 290312 61
AMI Phosphate-II
Program Overview
62 A-96.250.641 / 290312
AMI Phosphate-II
Program Overview
A-96.250.641 / 290312 63
AMI Phosphate-II
Program Overview
64 A-96.250.641 / 290312
AMI Phosphate-II
Program Overview
A-96.250.641 / 290312 65
AMI Phosphate-II
Program Overview
66 A-96.250.641 / 290312
AMI Phosphate-II
Program List and Explanations
1 Messages
1.1 Pending Errors
1.1.5 Provides the list of active errors with their status (active, acknowl-
edged). If an active error is acknowledged, the alarm relay opens
again. Cleared errors are moved to the Message list.
1.2 Maintenance List
1.2.5 Demands necessary maintenance, e.g. preparing new reagents.
1.3 Message List
1.3.1 Shows the error history: Error code, date / time of issue and status
(active, acknowledged, cleared). 65 errors are memorized. Then
the oldest error is cleared to save the newest error (circular buffer).
2 Diagnostics
In diagnostics mode, the values can only be viewed, not modified.
2.1 Identification
Designation: View the Designation of instrument.
Version: Firmware of instrument (e.g. V5.01 – 03/12)
2.1.3 Peripherals: PeriClip: Firmware of peristaltic pump (e.g. 1.03)
2.1.4 Factory Test: Test date of the Instrument -, Motherboard - and
Frontend - QC factory test.
2.1.5 Operating Time: Years / Days / Hours / Minutes / Seconds
2.2 Sensors
2.2.1 FOME Sensor:
Current value: Shows the actual photometer signal [ppm].
Raw value: Shows the actual photometer signal [Hz].
Absorbance: Process value, depends on sample.
2.2.1.4 Cal. History: Review diagnostic values of the last calibrations.
2.2.1.4.1 Number: Calibration counter
Date, Time: Date and time of calibration
Slope: Slope photometer: 0.8–1.2
A-96.250.641 / 290312 67
AMI Phosphate-II
Program List and Explanations
2.5 Interface
Only available if optional interface is installed.
Review programmed communication settings.
68 A-96.250.641 / 290312
AMI Phosphate-II
Program List and Explanations
3 Maintenance
3.1 Calibration
3.1.5 Calibration: Performs a calibration using the standard soluti-
non.Follow dialog on the screen.See Calibration, S. 44
3.2 Service
3.2.1 Verification: Performs a verification using the reference kit. Follow
dialog. See Verification, S. 43
3.2.2 Fill System: Activates the reagent pump.Can be used to fill, flush
or empty the system.
3.3 Simulation
Test alarm relay, relay 1 / 2 or signal output 1/2 by selecting with the
<Enter> key and the up/down keys to change the value. After con-
firming the setting with the <Enter> key, the value is simulated by
the relay/signal output.
At the absence of any key activities, the instrument will switch back
to normal mode after 20 min. If you quit the menu, all simulated val-
ues will be reset.
3.4 Set Time
Adjust date and time.
A-96.250.641 / 290312 69
AMI Phosphate-II
Program List and Explanations
3.5 Cleaning
Automatic cleaning process using Cleaning module-II (optional)
controlled by transmitter of AMI Phosphate-II.
– Measurement is interrupted.
– Flow error interrupts cleaning.
– Cleaning does not start: E022 Reagent empty, E023
Cleaning solution, no flow.
3.5.1 Parameter:
3.5.1.1 Mode: Select daily, weekly, interval or off.
3.5.1.xx Start time / Calendar: Time of day respectively weekday(s) for the
automatic start of the cleaning process.
3.5.1.3 Delay: During cleaning plus the delay time, the signal- and control
outputs are held in the operation mode programmed as shown be-
low.
Range: 0–6’000 sec
3.5.1.4 Signal Outputs: Select the operation mode of the signal outputs
during cleaning:
3.5.1.5
Output/Control: Relay or signal output:
Continuous: Limit/Controller continues normally.
Hold: Limit/Controller continues on the last valid value.
Off: Limit/Controller is switched off.
3.5.2 Fill Channel 11: Activates the cleaning pump and switches the
valve to the cleaning solution 1 (right container)
3.5.3 Fill Channel 12: Activates the cleaning pump and switches the
valve to the cleaning solution 2 (left container)
70 A-96.250.641 / 290312
AMI Phosphate-II
Program List and Explanations
4 Operation
4.1 Sensors
4.1.1 Filter Time Constant: Used to damp noisy signals. The higher the
filter time constant, the slower the system reacts to changes of the
measured value.
Range: 5–300 sec
4.1.2 Hold after Cal: Delay permitting the instrument to stabilize again af-
ter calibration. During calibration- plus hold-time, the signal outputs
are frozen (held on last valid value), alarm values, limits are not ac-
tive.
Range: 0–6‘000 sec
4.2 Relay Contacts
See 5.3 Relay Contacts, S. 75
4.3 Logger
The instrument is equipped with an internal logger. The data can be
downloaded to a PC by Hyperterminal if option “HyperTerminal in-
terface” is installed.
The logger can save approx. 1500 data records. Records consists
of: Date, time, alarms, measured value, measured value uncom-
pensated, temperature, flow.
Range: 1 second to 1 hour
4.3.1 Log Interval: Select a convenient log interval. Consult the table be-
low to estimate the max logging time. When the loggin buffer is full,
the oldest data record is erased to make room for the newest one.
(circular buffer)
4.3.2 Clear Logger: If confirmed with yes, the complete logger data is de-
leted. A new data series is started.
A-96.250.641 / 290312 71
AMI Phosphate-II
Program List and Explanations
5 Installation
5.1 Sensors
5.1.1 Ref. Verification: Set absorbance value of verification kit according
to label.
Range: 0.200–0.600
5.1.2 Phosphate as: Default setting is PO4. If the setting is changed from
PO4 to P, the measured value is expressed in P.
Available values: P, PO4
5.1.3 Standard PO4: Usual concentration range:
Range: 0.10 ppm to 8.00 ppm
5.1.4 Meas. Interval: Set the measuring interval:
Range: 10 min to 30 min
5.1.5 Channels: for AMI Phosphate-II B only. If the 2nd sample stream
option is installed, you can choose 1 or 2 channels.
5.1.6 Channel Selection: The following 3 operating modes can be set:
Internal
External
Fieldbus
Mode Internal: In this mode, the AMI Phosphate-II B sequentially
measures each sample stream of the second stream option.
Mode External: In this mode, the AMI Phosphate-II B can be con-
trolled via signal input.
Mode Fieldbus: The AMI Phosphate-II B is controlled via fieldbus
5.1.7 Cleaning: Only available for AMI Phosphate-II and only visible if a
cleaning module is connected.
Program if one or two solutions are used on cleaning module.
Range: 1 Solution or 2 Solutions
5.2 Signal Outputs
5.2.1 - 5.2.2 Signal Output 1/2: Assign process value, the current loop range
and a function to each signal output.
5.2.1.1-5.2.2.1 Parameter: Assign one of the process values to the signal output.
Available values: Phosphate, Sample Flow.
5.2.1.2-5.2.2.2 Current Loop: Select the current range of the signal output.
Make sure the connected device works with the same current
range.
Available ranges: 0–20 mA or 4–20 mA
72 A-96.250.641 / 290312
AMI Phosphate-II
Program List and Explanations
20
(0 - 20 [mA])
(4 - 20 [mA])
B A
10 12
0/4 X
0.0 0.1 0.2 0.3 0.4 0.5
[mA]
20
(0 - 20 [mA])
(4 - 20 [mA])
10 12
0/4 X
2 4 6
1 10 100 1’000 10’000
0 1 2 3 4
5.2.x.40 Scaling: Enter beginning and end point (Range low & high) of the
linear or logarithmic scale. In addition, the midpoint for the bilinear
scale.
Range: 0–10 ppm
A-96.250.641 / 290312 73
AMI Phosphate-II
Program List and Explanations
As control out- Signal outputs can be used for driving control units. We distinguish
put different kinds of controls:
P-controller: The controller action is proportional to the devi-
ation from the setpoint. The controller is characterized by
the P-Band. In the steady-state, the setpoint will never be
reached. The deviation is called steady-state error.
Parameters: setpoint, P-Band
PI-controller: The combination of a P-controller with an
I-controller will minimize the steady-state error. If the reset
time is set to zero, the I-controller is switched off.
Parameters: setpoint, P-Band, reset time.
PD-controller: The combination of a P-controller with a
D-controller will minimize the response time to a fast change
of the process value. If the derivative time is set to zero, the
D-controller is switched off.
Parameters: setpoint, P-Band, derivative time.
PID-controller: The combination of a P-, an I - and a D-con-
troller allows a proper control of the process.
Parameters: setpoint, P-Band, reset time, derivative time.
Ziegler-Nichols method for the optimization of a PID controller:
Parameters: Setpoint, P-Band, Reset time, Derivative time
Y
B A
X
a
74 A-96.250.641 / 290312
AMI Phosphate-II
Program List and Explanations
A-96.250.641 / 290312 75
AMI Phosphate-II
Program List and Explanations
5.3.1.1.x.1 Alarm High: If the measured value rises above the alarm high val-
ue, the alarm relay is activated and E001 is displayed in the mes-
sage list.
Range: 0.00–10.00 ppm
5.3.1.1.x.2x Alarm Low: If the measured value falls below the alarm low value,
the alarm relay is activated and E002 is displayed in the message
list.
Range: 0.00–10.00 ppm
5.3.1.1.x.3x Hysteresis: Within the hyst. range, the relay does not switch. This
prevents damage of relays contacts when the measured value fluc-
tuates around the alarm value.
Range: 0.00–10.00 ppm
5.3.1.1.x.4x Delay: Duration, the activation of the alarm relay is retarded after
the messuring value has risen above/fallen below the programmed
alarm.
Range: 0–28‘800 sec
5.3.1.2 Sample Flow: Define at which sample flow a flow alarm should be
issued.
5.3.1.2.1 Flow Alarm: Program if the alarm relay should be activated if there
is a flow alarm. Choose between yes or no. The flow alarm will al-
ways be indicated in the display, pending error list, saved in the
message list and the logger.
Available values: Yes or no
ATTENTION
76 A-96.250.641 / 290312
AMI Phosphate-II
Program List and Explanations
5.3.2/3 Relay 1/2: The function of relay contacts 1 or 2 are defined by the
user
A-96.250.641 / 290312 77
AMI Phosphate-II
Program List and Explanations
5.3.2.32.30 Response time: Minimal time the metering device needs to react.
Range: 0–240 sec.
Frequency Examples of metering devices that are puls frequency driven are
the classic membrane pumps with a potential free triggering input.
Dosing is controlled by the repetition speed of dosing shots.
5.3.2.32.21 Pulse frequency: Max. pulses per minute the device is able to re-
spond to. Range: 20–300/min.
Motor valve Dosing is controlled by the position of a motor driven mixing valve.
5.3.2.32.22 Run time: Time needed to open a completely closed valve
Range: 5–300 sec.
5.3.2.32.32 Neutral zone: Minimal response time in % of the runtime. If the re-
quested dosing output is smaller than the response time, no
change will take place.
Range: 1–20 %
Timer The relay will be closed repetitively depending on the programmed
time scheme.
ATTENTION
78 A-96.250.641 / 290312
AMI Phosphate-II
Program List and Explanations
5.3.4 Input: The functions of the relays and signal outputs can be de-
fined depending on the position of the input contact, i.e. no function,
closed or open.
5.3.4.1 Active: Define when the input should be active:
5.3.4.2 Signal Outputs: Select the operation mode of the signal outputs
when the relay is active:
5.3.4.4 Fault:
5.3.4.5 Delay: Time which the instrument waits, after the input is deactivat-
ed, before returning to normal operation.
Range: 0–6‘000 sec
A-96.250.641 / 290312 79
AMI Phosphate-II
Program List and Explanations
5.4 Miscellaneous
5.4.1 Language: Set the desired language.
Available settings: German / English / French / Spanish / Italian
5.4.2 Set defaults: Reset the instrument to factory default values in three
different ways:
Calibration: Sets calibration values back to default. All oth-
er values are kept in memory.
In parts: Communication parameters are kept in memory.
All other values are set back to default values.
Completely: Sets back all values including communication
parameters.
5.4.3 Load Firmware: Firmware updates should be done by instructed
service personnel only.
5.4.4 Password: Select a password different from 0000 to prevent unau-
thorized access to the menus “Messages”, “Maintenance”, “Opera-
tion” and “Installation”.
Each menu may be protected by a different password.
If you forgot the passwords, contact the closest SWAN representa-
tive.
5.4.5 Sample ID: Identify the process value with any meaning full text,
such as KKS number.
80 A-96.250.641 / 290312
AMI Phosphate-II
Program List and Explanations
5.5 Interface
Select one of the following communication protocols. Depending on
your selection, different parameters must be defined.
5.5.1 Protocol: Profibus
5.5.20 Device address: Range: 0–126
5.5.30 ID No.: Range: Analyzer; Manufacturer; Multivariable
5.5.40 Local operation: Range: Enabled, Disabled
5.5.1 Protocol: Modbus RTU
5.5.21 Device address: Range: 0–126
5.5.31 Baud Rate: Range: 2400–115’200 Baud
5.5.41 Parity: Range: none, even, odd
5.5.1 Protocol: Hyper Terminal
5.5.23 Baud Rate: Range: 2400–115’200 Baud
see separate RS232 Operation Manual for Logger download
5.5.1 Protocol: Webserver
5.5.24 Device Address: Range: 0–126
see separate Webserver Manual.
A-96.250.641 / 290312 81
AMI Phosphate-II
Material Safety Data sheets
Use of substance/preparation
Reagent for analysis.
Manufacturer/supplier identification
Company: SWAN Analytische Instrumente AG, CH-8340 Hinwil
Tel: +41 44 943 63 00 / Fax: +41 44 943 63 01
2 Hazard identification
3 Composition/information on ingredients
Synonyms
Ammonium molybdate; Hexammonium heptamolybdate 4-hydrate.
CAS-No.: 12054-85-2
Molecular formula: H24Mo7N6O24 * 4H2O
EINECS-No.: 234-722-4
Hazardous ingredients
Name according to EC Directives: None
5 Fire-fighting measures
82 A-96.250.641 / 290312
AMI Phosphate-II
Material Safety Data sheets
Procedures for cleaning / absorption: Take up dry. Forward for disposal. Clean up affected
area
Industrial hygiene
Change contaminated clothing. Wash hands after working with substance.
Form: Solid
Color: Off-White
Odor: Of ammonia
A-96.250.641 / 290312 83
AMI Phosphate-II
Material Safety Data sheets
11 Toxicological information
Quantitative data on the toxicity of this product are not available. Symptoms of an acute molyb-
denum (VI) intoxication: diarrhea, anemia (decreased hemoglobin concentration in the blood),
fatigue. Toxic effect on liver and kidneys after high doses. The product should be handled with care
usual when dealing with chemical.
12 Ecological information
The following applies to molybdenum compounds in general: Biological effects: Fish: Lethal from
25mg/l up; algea: Sc. quadricauda toxic from 54mg/l up; protozoa: Microregma toxic from 27mg/l
up.
13 Disposal considerations
Product
There are no uniform EC regulations for the disposal of chemicals or residues. Chemical residues
generally count as special waste. The disposal of the latter is regulated in the EC member
countries through corresponding laws and regulations, and in Switzerland also by the cantons. We
recommend that you contact either the authorities in charge or approved waste disposal
companies which will advise you on how to dispose of special waste.
Packaging
Disposal in compliance with official regulations. Handle contaminated packaging in the same way
as the substance itself. If not officially specified differently, non-contaminated packaging may be
treated like household waste or recycled.
14 Transport information
84 A-96.250.641 / 290312
AMI Phosphate-II
Material Safety Data sheets
15 Regulatory information
Symbol: --
R-phrases: --
S-phrases: --
16 Other information
A-96.250.641 / 290312 85
AMI Phosphate-II
Material Safety Data sheets
Use of substance/preparation
Reagent for analysis.
Manufacturer/supplier identification
Company: SWAN Analytische Instrumente AG, CH-8340 Hinwil
Tel: +41 44 943 63 00 / Fax: +41 44 943 63 01
2 Hazard identification
3 Composition/information on ingredients
Synonyms
Preparation of Ascorbic acid
CAS-No: 50-81-7
Hazardous ingredients
86 A-96.250.641 / 290312
AMI Phosphate-II
Material Safety Data sheets
5 Fire-fighting measures
Industrial hygiene
Change contaminated clothing. Wash hands after working with substance.
Form: Solid
Color: Colorless
Odor: Odorless
A-96.250.641 / 290312 87
AMI Phosphate-II
Material Safety Data sheets
11 Toxicological information
Acute toxicity
LD50 (oral, rat): 115mg/ kg (pure hazardous ingredients)
Further data
The product should be handled with the care usual when dealing with chemicals.
12 Ecological information
Ecotoxic effects
Quantitative data on the ecological effect of this product are not available.
13 Disposal considerations
Product
There are no uniform EC regulations for the disposal of chemicals or residues. Chemical residues
generally count as special waste. The disposal of the latter is regulated in the EC member
countries through corresponding laws and regulations, and in Switzerland also by the cantons. We
recommend that you contact either the authorities in charge or approved waste disposal
companies which will advise you on how to dispose of special waste.
Packaging
Disposal in compliance with official regulations. Handle contaminated packaging in the same way
as the substance itself. If not officially specified differently, non-contaminated packaging may be
treated like household waste or recycled.
14 Transport information
88 A-96.250.641 / 290312
AMI Phosphate-II
Material Safety Data sheets
15 Regulatory information
Symbol: --
R-phrases: 20/22 Harmful by inhalation and if swallowed
S-phrases: --
16 Other information
A-96.250.641 / 290312 89
AMI Phosphate-II
Default Values
90 A-96.250.641 / 290312
AMI Phosphate-II
Default Values
A-96.250.641 / 290312 91
AMI Phosphate II
Index
12. Index
A O
Alarm Relay ......... 27 occlusion frames. . . . . . . 31
On site requirements . . . . 17
C option
Cable thicknesses . ..... 19 cleaning module . . . . 23
Cleaning module . . ...... 7 second sample stream 24
cleaning module . . ...... 9
P
D Peristaltic pump . . . . . . . 52
Default Values . . . . . . . . 90 Power Supply . . . . . . 12, 26
Dimensions Phosphate-II . . 13 Prepare reagents . . . . . . 31
Dimensions Phosphate-II B 14 Program Access . . . . . . . 34
Protective filter . . . . . . . . 45
Pump tubes . . . . . . . . . 52
E
Electrical wiring . . . . . . . 17
Electrodes, Installation of . . 30 R
Error List . . . . . . . . . . . 57 Reagent 1 . . . . . . . . .. 42
Reagent 2 . . . . . . . . .. 42
Reagent consumption. . .. 41
F
Relay Box. . . . . . . . . .. 28
Fill System .......... 33
Relay Contacts 1 and 2 . .. 27
Fluidics . . .......... 11
Rinse the system . . . . .. 40
I
S
Input . . . . . . . . . . . . . 9, 27 Sample Flow . . ....... 32
Instrument Setup . . . . . . 17 Signal Outputs . ....... 29
Interface . . . . . . . . . . . 30 Solenoid valve . ....... 50
Interface RS232 . . . . . . . 30
Interface RS485 . . . . . . . 30
T
Terminals . . . . 21–22, 27, 30
M
two sample streams . . . . . . 9
Measuring principle ...... 8
W
Wire . ............. 19
A-96.250.641 / 230312 92
AMI Phosphate-II
Notes
13. Notes
A-96.250.641 / 290312 93
AMI Phosphate-II
SWAN
is represented worldwide by subsidiary companies
and distributors.
SWAN Products
Analytical Instruments for:
Potable Water
Cooling Water
Made in Switzerland
94 A-96.250.641 / 290312