Specifications: AX-700
SHENZHEN COMEN MEDICAL INSTRUMENTS CO.,LTD
Floor 10, Floor 11 and Section C of Floor 12 of Building 1A & Floor 1 to
Floor 5 of Building 2, FIYTA Timepiece Building, Nanhuan Avenue,
Matian Sub-district, Guangming District, Shenzhen, Guangdong,
518106,P.R. China
Tel: +86-755-26408879
Fax: +86-755-26431232
Email:
[email protected] Web: www.comen.com
2023 V 1.0
No. 046-001577-04
Anesthesia Machine
AX-700
Physical Characteristics Maximum bearing 1kg
weight
Complete Machine
Caster Wheels
Dimension 689mm*800mm*1400mm
Size 5 inch
Weight 125kg (standard configuration)
LED Indication
Maximum load 210kg
bearing AC indicator lamp Green LED
Top Plate Battery indicator Green LED
lamp
Operational 508mm*313mm
Dimension Working state Green LED
indicator lamp
Additional accessory 508mm*313mm*380mm
Dimensions Alarm indicator lamp Yellow/Red LED
Maximum load 20kg Audible Indication
bearing Speaker Alarm tone, key-stoke tones
Workbench Buzzer Fault alarm tone
Operational 465mm*275mm Interface
Dimension
Power supply 1 AC power interface
Additional accessory 472mm*248mm*380mm
4 auxiliary output power interfaces
Dimension
Equal potential 1 equal-potential grounding terminal
Maximum load 20kg
bearing Communication 1 RJ45 interface
connectors 1 USB interface
Dovetail rail
1 DB9 interface
Dimension 500mm*750mm*500mm
Display Screen
Handrail
Size 12.1’’ TFT touch screen
Length 750mm
Resolution 800*600 pixels
Drawers
Brightness Adjustable
Dimensions Top: 462mm*287mm*141mm
Bottom: 437mm*287mm*245mm Waveforms Pressure-time
Flow rate-time
Quantity Two drawers
Volume-time
Maximum bearing Top: 1kg
EtCO2
weight Bottom: 3kg
EEG
Gas-bag Sway Brace
Loops Pressure-volume Loops
Size Length: 425mm Flow-volume Loops
Height: 240mm Pressure-flow Loops
Loop hook
Operation Environment
Working Temp 10~40°C Type Float flowmeter
Relative humidity ≤93% Range 0~15L/min
Atmospheric pressure 70~106 kPa Auxiliary Oxygen Supply Flowmeter
Machine noise ≤65dB Type Float flowmeter
Power Supply Range 0~15L/min
External AC Power Supply O2-N20 Linked System
Input voltage 100~240V Type Mechanical type proportional control
device
Input frequency 50/60Hz
Range O2 concentration shall not be lower
Input current 3.5
than 25%
Fuse T10 AL 250V
ACGO
Power Cord 5m
Connector Outside: 22 mm taper coaxial fitting
Auxiliary Output Supply
Inside: 15 mm taper coaxial fitting
Output voltage 100~240V
Flush O2
Output frequency 50/60Hz
Range 25~75L/min
Output current 1.0A
Breathing Circuit
Internal Battery
Performance
Type Li-ion battery
Driving pressure 280~600kPa
Rated battery voltage 11.1VDC
Inspiratory flow 120L/min (280kPa)
Capacity 4400mAh
Flow valve 1~120L/min
Shutdown delay ≥30min
Pressure limit Mechanical relief valve
Minimum power-on 120min controlling means Electronic relief valve
time
Leakage and Compliance
Charging time 4h
Leak in breathing ≤65ml/min (3kPa)
Gas Supply
system and its cycle
Pipeline gasses O2-N2O-AIR absorption assembly
Backup gas-cylinder O2, O2-AIR, O2-N2O Compliance of Adult mode: ≤4ml/100Pa
gasses breathing system and Pediatric mode: ≤3ml/100Pa
Pipeline gas NIST its cycle absorption
connection assembly
Backup cylinder YOKE-CGA Leak in CO2 canister ≤50ml/min (3kPa)
connection APL valve leak ≤50ml/min (3kPa)
Inlet pressure range 280~600kPa Ports and Connectors
Filter 60~80um Expiratory end Outside: 22 mm taper coaxial fitting
Driving gas Standard: O2 Inside: 15 mm taper coaxial fitting
Optional: AIR Inspiratory end Outside: 22 mm taper coaxial fitting
Flowmeter Inside: 15 mm taper coaxial fitting
Electronic Flowmeter manual bag end Outside: 22 mm taper coaxial fitting
Air range 0~10L/min, the flow can be adjusted to Inside: 15 mm taper coaxial fitting
50ml/min Airway Pressure Gauge
O2 range 0~10L/min, the flow can be adjusted to Range -20~100cmH2O
50ml/min
Accuracy ± (4% of full scale reading + 4% of
N2O range 0~10L/min, the flow can be adjusted to actual reading)
50ml/min
APL Valve
Summation Flowmeter
Range 1~75cmH2O
Minimum opening Dry: 0.3cmH2O Sevoflurane: 0.2~8%
pressurett Humid: 0.5cmH2O Operation 15~35°C
Touch indication ≥30cmH2O temperature
Heater Volume 300ml
Heater Standard Ventilation Modes
CO2 Canister Standard Manual/Spont (optional:CPB)
VCV, PCV
Volume 2000ml
SIMV-VC, SIMV-PC
Bypass Standard
CPAP/PSV
AGSS (optional) Optional PRVC
Applicable laws ISO 80601-2-13 and YY 0635-2 SIMV-PRVC
Dimension 535mm*120mm*155mm PSVPro
Weight 2.2kg Ventilator Parameters
Applies ISO 80601-2-13 and YY 0635-2 Controlled Parameters Range
Connector ISO9170-2 or BS6834 standard Plimit (Pressure limit) 10~100cmH2O
connector
Pinsp (Inspiratory 5~70cmH2O
Pressure relief device Atmospheric pressure compensation pressure)
port
ΔPps (Support 3~60cmH2O, OFF
Filter Stainless steel mesh, with pore size of Pressure)
60~100 μm
Apnea Pressure 3~60cmH2O
Suction flow AGSS-L: 25~50L/min
PEEP 3~30cmH2O, OFF
AGSS-H: 50~80L/min,
VT 15~1500ml
Vacuum Suction System (optional) In the PCV mode, tidal volume can be
The performance High vacuum/High flow detected to 5ml
category Rate (Respiratory 4~100bpm
Air source pressure 280-550kPa Rate)
Recommended 75kPa I:E 4:1~1:10
maximum negative Apnea.IE 4:1~1:8
pressure
Tpause (Inspiratory 5%~60% of inspiratory time, OFF,
suction tubing Φ8 mm Pause)
Vaporizer (optional) Trig window(Trigger 5%~90%
Quantity Two window)
Vapor 2000/D-Vapor Rate (SIMV 4~60bpm
Frequency)
Range Halothane: 0.2%~6%
Enflurane: 0.2%~8% Apnea Time 10~30s
Isoflurane: 0.2%~6% Exp% (Inspiratory 5%~80%
Sevoflurane: 0.2%8% stop level)
Desflurane: 2%~18% Minrate 2~60bpm
Operation Vapor 2000: 10~40°C Tinsp (Inspiratory 0.2~5.0s
temperature D-Vapor: 18~30°C time)
Volume Vapor 2000: 360ml Trigger (Inspiratory Pressure triggering: -20~-1cmH2O
triggering) Flow rate triggering: 0.2~15L/min
D-Vapor: 300ml
Ibis200 Tslope (Pressure 0~2.0s
slope)
Range Halothane: 0.2%~5%
Monitoring Parameters Range
Enflurane: 0.2%~5%
Isoflurane: 0.2%~5% MV (Per-minute 0~100L/min
ventilation amount) value
VT (Inspiratory and 0~3000ml Other: not defined
expiratory tidal Monitoring Parameters Accuracy
volume)
VTexp (expiratory tidal 0~60ml: ±10ml
FiO2 (Oxygen 18~100% volume) 60ml~3000ml: ±20ml or ±10% of the
concentration) actual reading, whichever is greater
Paw (Airway -20~120cmH2O VTins(Inspiratory tidal ±20ml or ±10% of the actual reading,
pressure) volume) whichever is greater
PEEP 0~70cmH2O Paw (Airway ±2.0cmH2O or ±4% of setting value,
Pmean (Mean -20~120cmH2O pressure) whichever is greater
pressure) PEEP ±2.0cmH2O or ±4% of setting value,
Pplat (Platform 0~120cmH2O whichever is greater
pressure) Pplat (Platform ±2.0cmH2O or ±4% of setting value,
I:E 4:1~1:12 pressure) whichever is greater
Freq (Respiratory 0~120bpm Pmean (Mean ±2.0cmH2O or ±4% of setting value,
frequency) pressure) whichever is greater
Compl (Compliance) 0~300ml/cmH2O Rate (Respiratory ±1bpm or ±5% of the setting value,
Rate) whichever is greater
Resistance 0~600cmH2O/(s/L)
I:E 4:1~2:1: ±25% of actual reading
O2 concentration of 18%~100%
oxygen sensor 2:1~1:4: ±10% of actual reading
1:4~1:12: ±25% of actual reading
Flow control system 0~10L/min
Other: not defined.
Controlled Parameters Accuracy
MV (Per-minute 0~30L/min: ±1L/min or ±15% of the
VT 15~60ml: ±10ml ventilation amount) setting value, whichever is greater
60~210ml: ±15ml >30 L/min: not defined
210~1500ml: ±7% of set value
Compl 0~250ml/cmH2O: ±0.5ml/cmH2O or ±
Pinsp (Inspiratory ±2.5cmH2O or ±7% of setting value, 15% of the actual reading, whichever is
pressure) whichever is greater greater
Plimit (Pressure limit) ±2.5cmH2O or ±7% of setting value, Other: not defined
whichever is greater Resistance 0~20cmH2O/(L/s): ±10cmH2O/(L/s)
PEEP 3~30 cmH2O: ±2.0cmH2O or ±8% of 20~500cmH2O/(L/s): ±50% of the
setting value, whichever is greater actual reading
ΔPps (Support ±2.5cmH2O or ± 7% of setting value, Other: not defined
Pressure) whichever is greater Flow control system 10%~100%: ±10% of the indicated
Apnea Pressure ±2.5cmH2O or ± 7% of setting value, value
whichever is greater Other: not defined
Trigger pressure ±2cmH2O O2 concentration ±3%
Rate (Respiratory ±1bpm or ±5% of the setting value, Alarm
Rate) whichever is greater
VT High limit: 5~1600ml
I:E and Apnea.IE 2:1~1:4: ±10% of actual reading Low limit: 0~(High limit-5)ml
Other: ±25% of actual reading High limit: 2~100L/min
MV
Tpause (Inspiratory 20%~60%: ±15% of the set value or Low limit: 0~(High limit-2)L/min
Pause) ±0.1s High limit: 20~105%
FiO2
Other: not defined Low limit: 18~(High limit-2)%
Trigger window ±10% EtCO2 (Respironics High limit: (Low limit+2)~150mmHg
Trigger flow rate ±1L/min CO2) Low limit: 0~(High limit-2)mmHg
Inspiratory stop level ±10% FiCO2 (Respironics High limit: (Low limit+1)~76mmHg
CO2) Low limit: 0~74mmHg
Flow control system 10%~100%: ±10% of the indicated
EtCO2 (Masimo CO2) High limit: (Low limit+2)~190mmHg 9.3%~13.2%: ±8%
Low limit: 0~(High limit-2)mmHg 13.3%~19.7%: ±10%
FiCO2 (Masimo CO2) High limit: (Low limit+1)~99mmHg Total system response Mainstream: <1s
Low limit: 0~97mmHg time Sidestream: <4s
Ppeak High limit: 2~100cmH2O
AG (optional)
Low limit: 0~(High Limit-2)cmH2O
Masimo AG/AG+O2 (Sidestream)
Asphyxiation alarm 20 s, with error of ±3 s
time CO2 0~15%: ± (0.2%+2%of reading)
High limit: 2~100 15~25%: Not specified
BIS
Low limit: 0~98 N2O 0~100%: ± (2%+2%of reading)
Alarm pause 120s HAL, ENF, ISO 0~8%: ± (0.15%+5%of reading)
Alarm sound pressure 45~85dB 8~25%: Not specified
Peak sound pressure High-priority alarm: 82dB SEV 0~10%: ± (0.15%+5%of reading)
Medium-priority alarm: 80dB 10~25%: Not specified
High-priority alarm: 79dB DES 0~22%: ± (0.15%+5%of reading)
Air source pressure 190~220kPa 22~25%: Not specified
alarm O2 0~100%: ± (1%+2%of reading)
Oxygen Sensors Artema AG
6 CO2 0~1%: ±0.1%
Expected operation 20°C: 0.94x10 % O2 measurement time
life 40°C: 0.6x106% O2 measurement time 1~5%: ±0.2%
Response time <15s 5~7%: ±0.3%
7~10%: ±0.5%
Operating -20~+50°C
temperature N2O 0~20%: ±2%
Pressure 50~200KPa 20~100%: ±3%
Relative humidity 0 to 99% HAL, ENF, ISO 0~1%: ±0.15%
1~5%: ±0.2%
Output wandering at Typical value <5% (over 1 year)
oxygen concentration SEV 0~1%: ±0.15%
100% 1~5%: ±0.2%
Material White ABS 5~8%: ±0.4%
Period of Validity Term of validity shall not exceed 13 DES 0~1%: ±0.15%
months after the package is unpacked 1~5%: ±0.2%
CO2 (optional) 5~10%: ±0.4%
Masimo (Mainstream and Sidestream) 10~15%: ±0.6%
15~18%: ±1%
Range 0~190 mmHg, 0~25% (at 760 mmHg),
O2 0~25%: ±1%
Resolution 1mmHg or 0.1kPa or 0.1%
25~80%: ±2%
Accuracy ± (0.3%+4% of reading)
80~100%: ±3%
Total system response Mainstream: <1s
BIS (optional)
time Sidestream: <3s
Covidien (BIS)
Respironics (Mainstream and Sidestream)
BIS 0~100, accuracy: ±1%
Range 0~150mmHg
SQI 0~100%, accuracy: ±1%
0%~19.7%
0~20.0kPa (at760 mmHg) EMG 0~100dB, accuracy: ±1%
Resolution 1mmHg or 0.1kPa or 0.1% ESR 0~100%, accuracy: ±1%
Accuracy 0%~5.3%: ±0.3%
5.4%~9.2%: ±5%
*Notice: Specifications subject to changes without prior
notice. All rights reserved by Comen