Control Serum Preparation
Control Serum Preparation
437
438 LOGAN & ALLEN Clinical Chemistry
Stability
The stability of the serums was checked by assaying several lots of
each product at approximately (3-month intervals. Values were 01)-
tamed for the following constituents: urea nitrogen, glucose, total
protein, cholesterol,calcium, and chloride.
Results
Precision of Analytic Procedures
The day-to-day precision of the methods employed in our laboratory
for the determination of each of the 9 serum constituents is shown iii
Table 1. It is expressed in each case as percentage confidence limits
which are calculated as ± 3 standard deviations from the mean, divided
by the mean, and multiplied by 100. The values have been established by
Glucose (mg./l00 ml.) Automated 84.4 ± 4.4 206.2 ± 10.8 296.6 ± 5.2
ferricyanide
Urea nitrogen (mg./I00 nIl.) Automated 13.0 ± 14.0 25.9 ± 6.1 47.7 ± 8.4
diacetyl
monoxi me
Total protein (gm./lOO ml.) Automated .‘i.8 ± 2.6 7.1 ± 3.8 7.6 ± 3.95
biuret
Total cholesterol (mg/100 ml.) Manual, Abell 83.6 ± 9.0 189.0 ± 3.2 328.0 ± 1.8
ci al.
Chloride (mEq./L.) Automated 50.3 ± 4.9 97. 1 ± 4. S 130.6 ± 7.6
Inercummic
lIio(vatmate
Sodium (mEq./L.) Automated 119.1 ± 7.8 141.5 ± 6.8 157.5 ± 2.5
flame
Potassium (mEq./L.) Automated 4.46 ± 12.1 5.13 +4.8 5.76 ± 10.8
CaCa - 1’- - -
C’)
0
a
a
a CI a
-
a
a x
I -
Ca CC-
4-
a
‘ m-
a
a
w I I I I I’
a4-
cI
4a a
a
CI ‘
z a
0 C)
‘-4 CC- C)
0 a
-- Ca
0
C
z z 41
1.1
0
0 1.1 I-
a a a
0
0 C)
Ca C-I - a Ca a
z
0
z -
z
I I I I
a
C) a
z a -u
0 a
a
z C)
Ca
C)
Ca - ‘ Ca - a CaCa
I., I I
a
4-
c-I a
-a .a
4)
.a - a a
p
. I- 01 Ca C’l C-I CC-Ca
Cl - Ca Ca Ca CI - a
C)
C)
a
a
a
a
a
.0
a a
‘a
ha
-H CaH.C’E ‘a E
1:
N#{176}
t - -H -H 2 -HH -112
a
I m. a’i
0
4..
a -u i a
2 C-I C-I
-
a
a
0 a
4-
aj 2 bf
a - ‘
a
:
Vol. 14, No. 5, 196$ CONTROL SERUM PREPARATIONS 441
TOTAL CHOLESTEROL
Warner-Chilcott (±5%) 17 11 3 1
Dade (±5%)t 5 1 2 2
Hyland (±5%) 14 7 5 2
Clinton (±4% to ±8%) 5 1 2 2
Michael Reese (±5%)t 3 1 2 -
SOI)IUM
1)ade (±3%) 10 6 4 -
POTASSI (TM
Warner-Chilcott (±5%) 14 3 5 3
Hyland (±2.5%; ±4%) 5 4 1
Clinton (±1.7%) 1 - 1 -
(AmAmcM
Warner-Chilcott S 4 :i 1
Dade 4 1 2 1
Hyland 2 1 1 -
Clinton 2 1 - 1
URI( A( ID
Dade (±5%)f :1 3 0 -
Clinton (±3%) 1 1 0 -
(Table 3). Actually, most of our results were obtaillt’d using the
AutoAnalyzer method of Kessler and Wolfinan (8). With 1 t’xceptioii
(the Hyland products), there was considerable disagreenlent, the
AutoAllalyzer results tending to be higher than the di If erent maiiu-
facturers’ assay values. Aryan (12) has reported values by the Kessler
automated procedure averaging 3% lower than those by the Clark-
Collip method. We had performed the Clark-Collip niethod Ofl some
commercial products. Since at least 3 manufacturers used this method,
it seemed indicated to present only Clark-Collip results. It will be lloted
that discrepancies occurred in 3 of 16 lots exannne(l by this method.
uric acid determinations on the commercial control sei’ums were
made using tile automated method of Crowley (9), the results of which
are shown ui Table 3. The agreement was good in tile limited number
of products that we have tested for this constituent.
Stability
As mentioned above, we have studied tile stability of several con-
stituents in commercial control serums for 2 years and sometimes 3 or
4 years, when stored according to tile manufacturer’s directions. Sev-
eral bottles or vials of a given lot were purchased and a new vial opened
at each interval for analysis. For each constituent the same procedure
was used throughout. The day-to-day precision of the procedures as
shown in Table 1 was used to determine the significance of change in
coiiceiitration from the initial value. The serums were checked foi’
stability of urea nitrogen, glucose, total protein, total cholesterol,
chloride, and calcium values. Data were obtained up to 2 years on each
of tile constituents; for glucose amid urea nitrogen, some lots were
checked up to 4 years. Only in the total protein analyses were any sig-
nificant changes noted. With I lot of Chemtrol, we obtained a value of
7.73 gm./100 ml. initially and 7.2 gm./i00 ml. after 2 years. On the other
hland, 1 lot of Versatol-A Alternate increased from an initial value of
4.5 gm./i00 ml. to 4.9 gm./100 ml. after 2 years. These are not large
changes but are greater than expected on the basis of the precision of
the automated biuret method which was used for the total protein
determinations.
Discussion
Our results have shown that in analyzing a control serum for a given
constituent, the values obtained will vary frequently according to the
method or modification of the method used. This is particularly true
for glucose and cholesterol. When the same method was used as that
of the manufacturer in assigning values, tile agreement was often
better than when a different method was used. In general, when the
446 LOGAN & ALLEN Clinical Chemistry
same method was used, the value we obtained fell within tue acceptable
range recommended by the manufacturer or quite closely to it. More-
over, the various constituents we have analyzed are (1iute stable for at
least 2 years.
Our studies have led us to conclude that commercial control serums
should not be used in place of a primary standard when the constituent
to be assayed can be obtained in a relatively pure form and a stable
solution can be prepared. This opinion has been reached by a number
of other workers (13-17) and was emphasized by the participants of
the symposium on standards and reference samples at the 18th National
Meeting of the American Association of Clinical Chemists in 1966.
Perhaps more important is the fact that at least 1 manufacturer has
accepted this limitation for the use of their assayed serums. If used
as a primary standard, the assay values of the control serum iiiust be
absolutely accurate and the serum itself must be absolutely stable for
an indefinite period. Only Hyland Laboratories and Burroughs Well-
come & Co. place expiration dates on their packages of commercial
control serums. With 1 exception, the manufacturers derive the values
for their serums’ constituents from tile mean of a number of analyses,
and thus these results cannot be expected to have the accuracy required
of a primary standard. In addition, if these products are to be used as
standards the laboratory should follow exactly the same technic as that
used by the manufacturer in establishing the assay value. One other
manufacturer claims that their products caii be considered as “stand-
ards-in-serum” since they are prepared by weigiiing the constituents
into serums which have first been dialyzed either to remove them com-
pletely from the serum base or to reduce them to fixed low levels. Since
dialysis would not remove protein, it follows that values given for total
protein, total nitrogen, and protein-bound iodine are not based on a
weighing system. For those constituents which are “completely”
removed and then weighed back in, the claim as a standard may have a
slightly better foundation. However, to be considered a primary stand-
ard, the standard-in-serum must be absolutely correct as to concentra-
tion and identity. As Radin (17) points out, as long as the analyst meas-
ures constituents in the presence of interfering substances, the back-
ground absorbance of the manufacturer’s serum base would have to
be the same as that of the unknown to avoid introduction of error. The
manufacturer would seem to be assuming a difficult task to ensure that
his product can be used as a standard for the following reasons: (1)
he is working with human serum or defibrinated plasma which, because
of its inherent biologic variatiQn, can never be exactly duplicated from
batch to batch; (2) he must be certain of the absolute stability of the
Vol. 14, No. 5, 1968 CONTROL SERUM PREPARATIONS 447
Urea
#{176}. nitrogen. Technicon Laboratory Method File No. N-la. Technicon IllstrU!I14’lItS
Corporation, Chauncoy, N. Y.
3. Glucose. Technicon Laboratory Method F’ile No. X-2n. Techinico,r Instruinei*ts (Torporatioli,
Chauncey, N. Y.
4. Total protein. TeeIinion Lal)oratorv M(tIio,I I’ih No. N-I 4a. Technicon Instruments
Corporation, Ghauiicey, N. Y.
5. Aboll, IL. L., Levy, B. B., Brodie, B. B., and Kendall, F. E., Simplified method for thi
estimation of total cholesterol in serum and demonstration of its specificity. .1. Biol.
Chem. 195, 357 (1952).
6. Chloride. Technicon Laboratory \lethiod File No. N-5a. Techinicoii Instruments Corpora-
tion, Chauncey, N. Y.
7. Sodium and potassiuiit. ‘l’cchii,ieoit Laboratory Method File No. N-20a. Teelinicon Tnstru-
nieiits Corporation, Chauiicey, N. V.
8. Calcium. Technicon Laboratory Method File No. N-3. Technicon Instruments Corporation,
Chauncey, N. Y.
9. Crowley, L. V., I)eterlninatioll of uric acid--all automated analysis based on a carbonate
method. Gun. Chem. 10, 838 (1964).
10. Barnett, R. N., A scheme for the comparison of quantitative methods. Am. J. GUn. Pat hol.
43, 562 (1965).
11. Manual of Clinical Chemistry (rev. ed.) Laboratory of Hygiene, Department of National
Health & Welfare, Ottawa, Canada, 1966.
12. Aryan, D. A., Observations oti an automated method for determination of serum and urine
calcium. Am. J. Gun. Pat/ioi. 45, 358 (1966).
13. Klugerman, M. R., and lloutwehl, J. H., Commercial control sera in the clinical chemistry
laboratory. GUn. Chem. 7, 185 (1961).
14. Henry, R. J., Clismhai Chemistry: Principles and Technics. Hoeber, New York, 1964.
15. Wootton, I. D. P., Micro-analysis in Medical Bio-chemistry (ed. 4). Churchill, London,
1964.
16. Caraway, W. T., “Sources of Error in Clinical Chemistry.” In Standard Metl,od. of
Clinical Chemistry (Vol. 5), Meites, S., Ed. Acad. Press, New York, 1965, p. 19.
17. Badin, N., What is a standarM GUn. Chem. 13, 55 (1967).