حلول تقارير لاب القياسات الهندسية للطالب ابواهيم ابو مسلم
حلول تقارير لاب القياسات الهندسية للطالب ابواهيم ابو مسلم
Objective:
To demonstrate how the platinum resistance thermometer (PRT) operates, how to connect it properly,
and to demonstrate the PRT's linearity and the PRT's suitability for use as a reference temperature sensor
for all other investigations.
Materials:
Temperature Measurement and Calibration apparatus (TD400)
The parts of this device are:
I} Heater tank.
II} Ice box.
III} Platinum resistance thermometer.
IV} Reference sensor.
V} Temperature and voltage displays.
Theory:
PRT Calibration is a systematic procedure for determining the accuracy and precision of a PRT
thermometer by comparing its readings with those of a reference thermometer. The experiment involves
careful temperature control, recording of readings, calculation of calibration constants, analysis of results,
and uncertainty analysis.
𝑽𝑽
𝑪𝑪𝑪𝑪𝑪𝑪𝑪𝑪𝑪𝑪𝑪𝑪𝑪𝑪𝑪𝑪𝑪𝑪𝑪𝑪 𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹 = [ 𝛀𝛀 ]
𝑰𝑰
𝑫𝑫𝑫𝑫𝑫𝑫𝑫𝑫𝑫𝑫𝑫𝑫𝑫𝑫𝑫𝑫𝑫𝑫
𝑬𝑬𝑬𝑬𝑬𝑬𝑬𝑬𝑬𝑬% = ∗ 𝟏𝟏𝟏𝟏𝟏𝟏%
𝑺𝑺𝑺𝑺𝑺𝑺𝑺𝑺𝑺𝑺𝑺𝑺𝑺𝑺𝑺𝑺 𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹
Data:
Current = 1 [ mA ]
Reference Temperature [ ̊C ]
Measured Voltage [ mV ]
Standard Resistance from Specifications [ Ω ]
{ PAGE 1 }
Calculation Sample:
𝑽𝑽 𝟏𝟏𝟏𝟏𝟏𝟏. 𝟒𝟒
𝑪𝑪𝑪𝑪𝑪𝑪𝑪𝑪𝑪𝑪𝑪𝑪𝑪𝑪𝑪𝑪𝑪𝑪𝑪𝑪 𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹 = = = 𝟏𝟏𝟏𝟏𝟏𝟏. 𝟒𝟒 [ 𝛀𝛀 ]
𝑰𝑰 𝟏𝟏
𝑫𝑫𝑫𝑫𝑫𝑫𝑫𝑫𝑫𝑫𝑫𝑫𝑫𝑫𝑫𝑫𝑫𝑫 𝟏𝟏. 𝟒𝟒
𝑬𝑬𝑬𝑬𝑬𝑬𝑬𝑬𝑬𝑬% = ∗ 𝟏𝟏𝟏𝟏𝟏𝟏% = ∗ 𝟏𝟏𝟏𝟏𝟏𝟏% = 𝟏𝟏. 𝟒𝟒%
𝑺𝑺𝑺𝑺𝑺𝑺𝑺𝑺𝑺𝑺𝑺𝑺𝒓𝒓𝒓𝒓 𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹𝑹 𝟏𝟏𝟏𝟏𝟏𝟏
Results:
{ PRT Calibration }
Standard
Reference Measured Calculated Resistance
Deviation
Temperature Voltage Resistance from Error%
[Ω]
[̊ C ] [ mV ] [Ω] Specifications
[Ω]
0 101.4 101.4 100 1.4 1.400%
18 108.4 108.4 107.02 1.38 1.289%
23 110.2 110.2 108.96 1.24 1.138%
28 112.2 112.2 110.9 1.3 1.172%
33 114.1 114.1 112.83 1.27 1.126%
38 116.1 116.1 114.38 1.72 1.504%
43 118 118 116.7 1.3 1.114%
48 120 120 118.6 1.4 1.180%
53 122 122 120.55 1.45 1.203%
58 123.8 123.8 122.47 1.33 1.086%
63 125.8 125.8 124.39 1.41 1.134%
68 127.8 127.8 126.31 1.49 1.180%
73 129.8 129.8 128.22 1.58 1.232%
78 131.7 131.7 130.13 1.57 1.206%
83 133.6 133.6 132.04 1.56 1.181%
88 135.6 135.6 133.95 1.65 1.232%
93 137.6 137.6 135.85 1.75 1.288%
98 139.6 139.6 137.75 1.85 1.343%
{ PAGE 2 }
{ PRT Calibration }
Calculated Resistance Standard Resistance
140
135
130
125
Resistance [ Ω ]
120
115
110
105
100
0 18 23 28 33 38 43 48 53 58 63 68 73 78 83 88 93 98
Temperature [ ̊C ]
Discussion:
In conclusion, conducting a PRT calibration experiment has several benefits, including improved
accuracy, increased reliability, cost savings, compliance with industry standards, and reduced risk of errors.
These benefits make it essential to regularly calibrate PRTs to ensure that they produce accurate and
reliable temperature measurements. And as we noted the results showed that the lowest error rate at 58 ̊C
was 1.086% and the highest error rate was 1.504% at 38 ̊C. As a result of the low error rate, we can
conclude that this instrument has high accuracy and that the readings it generates are reliable.
{ PAGE 3 }
Al-Balqa Applied University
Faculty Of Engineering Technology
Mechanical Engineering Department
Second Semester { 2022 - 2023 }
Course Code : { MEE0413 }
Course Title :
{ ENGINEERING MEASUREMENTS LAB }
Class Time : { TUESDAY }{ 08:00 - 11:00 }
Theory:
The gas Pressure thermometer works on the principle of a known
expansion rate of gas (vapor of a fluid) for a given temperature increase within
a known volume. As the gas (vapor above a small volume of fluid) expands
due to temperature increase, its pressure rises, exerting force on a mechanical
mechanism that rotates a calibrated dial. As for as the bi-metal thermometer
works on the principle of different expansion rates of two different metal strips
fixed (laminated) together to form a thicker strip, which is a composite of the
two metals. When the temperature increases, one of the metals expands more
than the other, causing the composite strip to bend, which then moves a
calibrated mechanical dial.
Equations:
𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷(𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑)
= |𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅 𝑇𝑇𝑇𝑇𝑚𝑚𝑚𝑚𝑚𝑚𝑚𝑚𝑚𝑚𝑚𝑚𝑚𝑚𝑚𝑚𝑚𝑚 − 𝐼𝐼𝐼𝐼𝐼𝐼𝐼𝐼𝐼𝐼𝐼𝐼𝐼𝐼𝐼𝐼𝐼𝐼 𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇| [ ℃]
𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷(𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑)
𝐸𝐸𝐸𝐸𝐸𝐸𝐸𝐸𝐸𝐸% = ∗ 100%
𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅 𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇𝑇
Materials:
Temperature Measurement and Calibration Apparatus (TD400)
The parts of this device are:
I} Heater tank with drain tap.
II} Icebox.
III} Nine different popular temperature measuring devices.
IV} Platinum resistance thermometer as a reference to calibrate the other
devices.
V} Built-in millivoltmeter.
VI} Built-in pressure sensor (barometer).
VII} Voltage display.
VIII} Reference display to display reference temperature, pressure, and
local water boiling temperature.
{ PAGE 1 }
Procedures:
I} Set up the heater and icebox.
II} Choose the thermometer.
III} Put the reference sensor and the thermometer into the icebox (through
the holes in its lid). Waite a few minutes for the readings to stabilize
and record them (the reference temperature should be between 0 and
1°C).
IV} Now put both devices into the heater tank (through the holes in its lid).
Switch on the heater and note the reference temperature.
V} At intervals of 5°C (shown by the reference temperature), record the
readings of the thermometer.
VI} Stop the experiment and switch off the heater when the reference
temperature reaches 100°C.
VII} Repeat the experiment for the other thermometer. Allow the heater
water to cool down (and change it if necessary for cooler water)
between tests.
Data:
Indicated Temperature [℃]
Reference
Gas Pressure
Temperature [℃] Bi-Metal Thermometer
Thermometer
0 0 -15
15 14 0
20 19 2
25 24 6
30 28 10
35 34 14
40 40 20
45 44 26
50 48 30
55 55 35
60 61 41
65 67 47
70 70 51
75 75 56
80 81 62
85 86 67
90 91 72
95 97 77
{ PAGE 2 }
Calculation Samples:
𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷(𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑) = �𝑇𝑇𝑅𝑅𝑅𝑅𝑅𝑅. − 𝑇𝑇𝐼𝐼𝐼𝐼𝐼𝐼. � = |80 − 81| = 1 [℃]
𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷(𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑𝑑) 1
𝐸𝐸𝐸𝐸𝐸𝐸𝐸𝐸𝐸𝐸% = ∗ 100% = ∗ 100% = 1.25%
𝑇𝑇𝑅𝑅𝑅𝑅𝑅𝑅. 80
Results:
{ PAGE 3 }
{ Gas Pressure Thermometer Calibration }
Reference Temperature Indicated Temperature
100
90
80
70
Indicated Temperature [℃]
60
50
40
30
20
10
0
0 10 20 30 40 50 60 70 80 90 100
90
80
70
Indicated Temperature [℃]
60
50
40
30
20
10
0
0 10 20 30 40 50 60 70 80 90 100
Two strips of
different metals
Pressure of a gas bonded together,
Principle changes with which expand at
temperature different rates as
temperature
changes
Generally, more
Less accurate than
accurate,
Accuracy gas pressure
especially at high
thermometers
temperatures
Less sensitive and
better suited for Can detect even
Sensitivity measuring small changes in
relatively stable temperature
temperatures
There are several factors that can cause errors in the calibration of gas
pressure and bi-metal thermometers. Some possible sources of error include
Ambient temperature fluctuations, Instrument drift, Instrument bias, Human error,
and Inadequate calibration standards. To minimize these potential sources of
error, it's important to follow a rigorous and well-documented calibration
procedure, use high-quality reference standards, and carefully control the test
environment during calibration. It's also a good practice to periodically re-
calibrate thermometers to ensure they remain accurate over time.
Finally, the experiment demonstrated that different types of thermometers
may have different calibration factors and responses to temperature changes,
and therefore, it is crucial to understand the characteristics of each thermometer
type.
{ PAGE 6 }
Al-Balqa Applied University
Faculty Of Engineering Technology
Mechanical Engineering Department
Second Semester { 2022 - 2023 }
Course Code : { MEE0413 }
Course Title :
{ ENGINEERING MEASUREMENTS LAB }
Class Time : { TUESDAY }{ 08:00 - 11:00 }
Objectives:
I} To demonstrate the functioning of the NTC Thermistor.
II} To study the temperature-dependent resistance of an NTC (Negative
Temperature Coefficient) thermistor and verify its behavior as a temperature
sensor.
III}To investigate the accuracy and linearity of the NTC thermistor as a
temperature sensor against a reference sensor.
Theory:
The apparatus comprises a thermistor, which is a type of thermal resistor. It
operates in a comparable manner to resistance thermometers, but instead of using a
metal, it employs a semiconductor material as an electrical resistor. This type of
thermistor is known as an NTC (Negative Temperature Coefficient) thermistor
because its resistance decreases as temperature rises.
Equations:
𝑉𝑉
𝑅𝑅 = [ Ω]
𝐼𝐼
𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷 = |𝑅𝑅𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆 − 𝑅𝑅𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶 | [ Ω]
𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷
𝐸𝐸𝐸𝐸𝐸𝐸𝐸𝐸𝐸𝐸% = ∗ 100%
𝑅𝑅𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆
Materials:
Temperature Measurement and Calibration Apparatus (TD400)
{ PAGE 1 }
VIII} Reference display to display reference temperature, pressure, and local
water boiling temperature.
Procedures:
I} Set up the heater and icebox.
II} Choose the thermometer.
III} Put the reference sensor and the thermometer into the icebox (through the
holes in its lid). Waite a few minutes for the readings to stabilize and record
them (the reference temperature should be between 0 and 1°C).
IV} Now put both devices into the heater tank (through the holes in its lid).
Switch on the heater and note the reference temperature.
V} At intervals of 5°C (shown by the reference temperature), record the
readings of the millivoltmeter.
VI} Stop the experiment and switch off the heater when the reference
temperature reaches 100°C.
VII} Repeat the experiment for the other thermometer. Allow the heater water to
cool down (and change it if necessary for cooler water) between tests.
Data:
Current (I) = 1 [𝑚𝑚𝑚𝑚]
Calculation Samples:
𝑉𝑉 0.25
𝑅𝑅 = = = 250 [ Ω]
𝐼𝐼 0.001
𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷 = |𝑅𝑅𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆 − 𝑅𝑅𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶 | = |261 − 250| = 11 [ Ω]
𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷 11
𝐸𝐸𝐸𝐸𝐸𝐸𝐸𝐸𝐸𝐸% = ∗ 100% = ∗ 100% = 4.215 %
𝑅𝑅𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆 261
{ PAGE 2 }
Results:
{ PAGE 3 }
{ NTC Thermistor Calibration }
Standard Resistance Calculated Resistance
300
250
200
Resistance [ Ω ]
150
100
50
0
0 10 20 30 40 50 60 70 80 90 100
And during the experiment, it was found that the device's highest error rate was
196.356%, which is a very high error, indicating that the device has low accuracy.
Several factors could potentially cause errors in an NTC thermistor experiment, and
some of these factors include:
{ PAGE 5 }
Al-Balqa Applied University
Faculty Of Engineering Technology
Mechanical Engineering Department
Second Semester { 2022 - 2023 }
Course Code : { MEE0413 }
Course Title :
{ ENGINEERING MEASUREMENTS LAB }
Class Time : { TUESDAY }{ 08:00 - 11:00 }
Objectives:
The experiment aims to:
Theory:
K-type thermocouples are made up of two dissimilar metal wires, typically made
of chromel (a nickel-chromium alloy) and alumel (a nickel-aluminum alloy), that are
joined at the ends to form a junction. When this junction is exposed to a temperature
gradient, a voltage is generated due to the Seebeck effect, which is the principle behind
thermocouple temperature measurement.
Equations:
𝐴𝐴𝐴𝐴𝐴𝐴𝐴𝐴𝐴𝐴𝐴𝐴 𝑀𝑀𝑀𝑀𝑀𝑀𝑀𝑀𝑀𝑀𝑀𝑀𝑀𝑀𝑀𝑀 𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉 = 𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶 𝑀𝑀𝑀𝑀𝑀𝑀𝑀𝑀𝑀𝑀𝑀𝑀𝑀𝑀𝑀𝑀 𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉 /20 [mV]
Materials:
Temperature Measurement and Calibration Apparatus (TD400)
{ PAGE 1 }
I} Heater tank with drain tap.
II} Icebox.
III} Nine different popular temperature measuring devices.
IV} Platinum resistance thermometer as a reference to calibrate the other
devices.
V} Built-in millivoltmeter.
VI} Built-in pressure sensor (barometer).
VII} Voltage display.
VIII} Reference display to display reference temperature, pressure, and local
water boiling temperature.
Procedures:
I} Set up the heater and icebox.
II} Choose the thermometer.
III} Put the reference sensor and the thermometer into the icebox (through the
holes in its lid). Waite a few minutes for the readings to stabilize and record
them (the reference temperature should be between 0 and 1°C).
IV} Now put both devices into the heater tank (through the holes in its lid).
V} Switch on the heater and note the reference temperature.
VI} At intervals of 5°C (shown by the reference temperature), record the
readings of the millivoltmeter.
VII} Stop the experiment and switch off the heater when the reference
temperature reaches 100°C.
VIII} Repeat the experiment for the other thermometer. Allow the heater water to
cool down (and change it if necessary for cooler water) between tests.
Data:
Reference Temperature [℃]
Calculation Samples:
Reading At { 95 ℃ }:
{ PAGE 2 }
= |3.04 − 3.889| = 0.849 [mV]
𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷 0.849
𝐸𝐸𝐸𝐸𝐸𝐸𝐸𝐸𝐸𝐸% = ∗ 100% = ∗ 100% = 21.831%
𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆 𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉 3.889
Results:
Actual
Reference Measured Standard
Measured Deviation
Temperature Voltage Voltage Error%
Voltage [𝑚𝑚𝑚𝑚 ]
[℃] [𝑚𝑚𝑚𝑚 ] [𝑚𝑚𝑚𝑚 ]
[𝑚𝑚𝑚𝑚 ]
0 -14.7 -0.735 0 0.735 DIV/0
20 0.8 0.04 0.798 0.758 94.987
25 4.4 0.22 1 0.78 78
30 8.3 0.415 1.203 0.788 65.503
35 12.2 0.61 1.407 0.797 56.645
40 16.1 0.805 1.612 0.807 50.062
45 20 1 1.817 0.817 44.964
50 24 1.2 2.023 0.823 40.682
55 27.8 1.39 2.23 0.84 37.668
60 31.4 1.57 2.436 0.866 35.550
65 34.5 1.725 2.644 0.919 34.758
70 38.2 1.91 2.851 0.941 33.006
75 43 2.15 3.059 0.909 29.716
80 48.3 2.415 3.267 0.852 26.079
85 51.5 2.575 3.474 0.899 25.878
90 56.9 2.845 3.682 0.837 22.732
95 60.8 3.04 3.889 0.849 21.831
{ PAGE 3 }
{ K-Type Thermocouple Calibration }
Standard Voltage Actual Measured Voltage
5
3
Voltages [𝑚𝑚𝑉𝑉]
0
0 10 20 30 40 50 60 70 80 90 100
-1
The results of the experiment showed that the thermocouple's deviation from the
reference temperature was within an acceptable range of ±2 [mV], indicating that the
thermocouple was accurately calibrated. The readings obtained from the experiment
were then plotted against the manufacturer's calibration curve, and the results showed
that the thermocouple's readings were in line with the manufacturer's curve, indicating
that the thermocouple was within its specified accuracy range.
{ PAGE 5 }
Al-Balqa Applied University
Faculty Of Engineering Technology
Mechanical Engineering Department
Second Semester { 2022 - 2023 }
Course Code : { MEE0413 }
Course Title :
{ ENGINEERING MEASUREMENTS LAB }
Class Time : { TUESDAY }{ 08:00 - 11:00 }
Objectives:
The experiment has two main objectives:
Theory:
The Seebeck Effect is a phenomenon where a voltage is generated when two
different metals are joined together at two different temperatures. This voltage is
proportional to the difference in temperature between the two junctions. The Seebeck
Effect is due to the difference in the way that different metals conduct heat. When two
different metals are joined, electrons flow from the hotter metal to the colder metal.
This creates a potential difference, or voltage, between the two metals.
Equations:
𝐴𝐴𝐴𝐴𝐴𝐴𝐴𝐴𝐴𝐴𝐴𝐴 𝑀𝑀𝑀𝑀𝑀𝑀𝑀𝑀𝑀𝑀𝑀𝑀𝑀𝑀𝑀𝑀 𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉 = 𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶𝐶 𝑀𝑀𝑀𝑀𝑀𝑀𝑀𝑀𝑀𝑀𝑀𝑀𝑀𝑀𝑀𝑀 𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉 /20 [mV]
Materials:
Temperature Measurement and Calibration Apparatus (TD400)
{ PAGE 1 }
The parts of this device are:
Procedures:
I} Set up the heater and icebox.
II} Choose the thermometer.
III} Put the reference sensor and the thermometer into the icebox (through the
holes in its lid). Waite a few minutes for the readings to stabilize and record
them (the reference temperature should be between 0 and 1°C).
IV} Now put both devices into the heater tank (through the holes in its lid).
V} Switch on the heater and note the reference temperature.
VI} At intervals of 5°C (shown by the reference temperature), record the
readings of the millivoltmeter.
VII} Stop the experiment and switch off the heater when the reference
temperature reaches 100°C.
VIII} Repeat the experiment for the other thermometer. Allow the heater water to
cool down (and change it if necessary for cooler water) between tests.
Data:
Reference Temperature [℃]
Calculation Samples:
Reading At { 80 ℃ }:
{ PAGE 2 }
𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷 = |𝐴𝐴𝐴𝐴𝐴𝐴𝐴𝐴𝐴𝐴𝐴𝐴 𝑀𝑀𝑀𝑀𝑀𝑀𝑀𝑀𝑀𝑀𝑀𝑀𝑀𝑀𝑀𝑀 𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉 − 𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆 𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉𝑉 |
Results:
Actual
Reference Measured Standard
Measured Deviation
Temperature Voltage Voltage Error%
Voltage [𝑚𝑚𝑚𝑚 ]
[℃] [𝑚𝑚𝑚𝑚 ] [𝑚𝑚𝑚𝑚 ]
[𝑚𝑚𝑚𝑚 ]
1.5 3.9 0.195 0 0.195 DIV/0
20 15.5 0.775 0.798 0.023 2.882
25 21 1.05 1 0.05 5
30 24.7 1.235 1.203 0.032 2.660
35 28.8 1.44 1.407 0.033 2.345
40 33.2 1.66 1.612 0.048 2.978
45 37.2 1.86 1.817 0.043 2.367
50 41.3 2.065 2.023 0.042 2.076
55 45.1 2.255 2.23 0.025 1.121
60 49.4 2.47 2.436 0.034 1.396
65 53.4 2.67 2.644 0.026 0.983
70 57.9 2.895 2.851 0.044 1.543
75 62.6 3.13 3.059 0.071 2.321
80 66.7 3.335 3.267 0.068 2.081
85 71.1 3.555 3.474 0.081 2.332
90 74.9 3.745 3.682 0.063 1.711
95 77.1 3.855 3.889 0.034 0.874
{ PAGE 3 }
{ K-Type Thermocouple Calibration }
Standard Voltage Actual Measured Voltage
4.5
3.5
3
Voltages [𝑚𝑚𝑉𝑉]
2.5
1.5
0.5
0
0 10 20 30 40 50 60 70 80 90 100
The results showed that the measured voltage increased as the reference
temperature increased, with a deviation from the standard voltage.
The error percentage was calculated for each measurement, with most falling
within 2-3%. The experiment successfully demonstrated the Seebeck effect and
provided a way to calibrate the thermocouples to directly measure temperature on the
Celsius scale.
Overall, the experiment was effective in achieving its objectives and provided
valuable insights into the behavior of thermocouples in relation to temperature and
voltage.
{ PAGE 5 }
Al-Balqa Applied University
Faculty Of Engineering Technology
Mechanical Engineering Department
Second Semester { 2022 - 2023 }
Course Code : { MEE0413 }
Course Title :
{ ENGINEERING MEASUREMENTS LAB }
Class Time : { TUESDAY }{ 08:00 - 11:00 }
Experiment Title : { Thermocouples and Circuit
Resistance }
Experiment Number : { 6 }
Objectives:
The objective of the thermocouples and circuit resistance experiment is to study the
principles of temperature measurement using thermocouples and to investigate the
effects of circuit resistance on thermocouple measurements.
The experiment aims to explore the relationship between the generated voltage and
the temperature being measured. By understanding this relationship, researchers can
calibrate and use thermocouples for accurate temperature measurements in various
applications. The experiment also investigates the impact of circuit resistance on
thermocouple measurements. The electrical resistance of the circuit through which the
thermocouple signal passes can affect the accuracy and reliability of the temperature
measurement.
Theory:
Thermocouples are temperature sensors based on the Seebeck effect, which
generates a voltage when two dissimilar metals are joined at two points. The voltage is
proportional to the temperature difference between the junctions. In this experiment,
the objective is to study temperature measurement using thermocouples and
investigate the effects of circuit resistance.
A thermocouple consists of two metal wires joined at one end to form the
measurement junction, exposed to the temperature being measured. The voltage
generated depends on the metals used and the temperature difference. By comparing
the voltage at the measurement junction to a reference junction (cold junction)
maintained at a known temperature, the temperature can be determined.
{ PAGE 1 }
Equations:
𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷 = |𝑁𝑁𝑁𝑁 𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅 𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅 − 𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅 𝑊𝑊𝑊𝑊𝑊𝑊ℎ 𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅| [mV]
𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷𝐷
𝐸𝐸𝐸𝐸𝐸𝐸𝐸𝐸𝐸𝐸% = ∗ 100%
𝑁𝑁𝑁𝑁 𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅 𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅
Materials:
Temperature Measurement and Calibration Apparatus (TD400)
Procedures:
I} Set up the heater and icebox.
II} Choose the thermometer.
III} Connect the K-type thermocouple directly to the amplifier input, but use the
socket at the end of the resistors as a connection point.
IV} Put the reference sensor and the thermometer into the icebox (through the holes
in its lid). Waite a few minutes for the readings to stabilize and record them
(the reference temperature should be between 0 and 1°C).
V} Now put both devices into the heater tank (through the holes in its lid).
VI} Switch on the heater. When the heater water reaches approximately 25°C,
measure and record the actual reference temperature and the millivolt meter
reading.
VII} move the connecting wire to the next socket along the resistors, to add 10 Ohm
resistance (series) to the cable. Note the millivolt meter reading.
VIII} Now move the connecting wire to the next sockets to add 110 Ohm (10+100)
and 1110 Ohm (10+100+1000) series resistances take a reading each time.
IX} move the connecting wire 2 only to the next socket along the resistors, to
add 10 Ohm resistance (parallel) to the amplifier input. Note the millivolt
meter reading.
{ PAGE 2 }
X} Now move the connecting wire 2 only to the next sockets one after the other
to add 110 Ohm (10+100) and 1110 Ohm (10+100+1000) parallel
resistances take a reading each time.
Data:
Reference Temperature [℃]
Resistances [Ω]
Calculation Samples:
Reading At { 110 [Ω] Series Resistance }:
{ PAGE 3 }
Discussion and Conclusions:
In the thermocouples and circuit resistance experiment, we investigated
temperature measurement using thermocouples and the effects of circuit resistance on
their accuracy. By varying the circuit resistance, we aimed to understand the impact on
the thermocouple output voltage and assess any limitations or sources of error.
Regarding the circuit resistance, we found that it had a significant influence on the
accuracy of temperature measurements. When the circuit resistance was high relative
to the thermocouple resistance, additional voltage drops occurred across the circuit,
leading to errors in the temperature measurement. Therefore, it is crucial to consider
and minimize circuit resistance in order to obtain precise and reliable temperature
readings.
{ PAGE 4 }
Al-Balqa Applied University
Faculty Of Engineering Technology
Mechanical Engineering Department
Second Semester { 2022 - 2023 }
Course Code : { MEE0413 }
Course Title :
{ ENGINEERING MEASUREMENTS LAB }
Class Time : { TUESDAY }{ 08:00 - 11:00 }
Experiment Title : { Variation of output voltage in
a potentiometer used as a position transducer }
Experiment Number : { 7 }
position transducer }
Objectives:
The experiment aims to achieve several objectives. Firstly, it seeks to investigate
the relationship between the position of the potentiometer's slider and the
corresponding output voltage. By systematically varying the position and measuring
the voltage, the experiment aims to establish a quantitative relationship or function
between position and voltage.
Theory:
The variation of output voltage in a potentiometer used as a position transducer is
based on the principles of potentiometry and electrical circuits. A potentiometer acts as
a voltage divider, with the voltage output at the slider terminal determined by the
position of the slider along the resistive element. In an ideal potentiometer, the voltage
output varies linearly with the position of the slider.
{ PAGE 1 }
Materials:
DE LORENZO (DL2312HG ):
Procedures:
I} Connect the circuit using the 100KΩ potentiometer and only use the digital
meter in the output B.
II} Connect the potentiometer control to zero.
III} Observe the output voltage.
IV} Use only the digital meter.
V} Use only the moving coil meter.
VI} Put the potentiometer control to ¼ of trip and repeat the lectures for this
position.
VII} writing the obtained values in the table.
VIII} Repeat the procedure for control positions of ½, ¾ , and to full trip, recording
the lectures in the respective column.
{ PAGE 2 }
Results:
Resistance = 1 𝐾𝐾Ω
Potentiometer Voltmeter Readings
0% 0
25 % 0.77
50 % 2.47
75 % 3.97
100 % 4.34
Resistance = 10 𝐾𝐾Ω
Potentiometer Voltmeter Readings
10 0
4 2.08
2 3.73
1 5.05
{ PAGE 3 }
{ Resistance =100 𝐾𝐾Ω }
6
4
Voltmeter Readings [𝑉𝑉]
0
0% 10% 20% 30% 40% 50% 60% 70% 80% 90% 100%
Potentiometer [𝑉𝑉]
{ PAGE 4 }
{ Resistance =1 𝐾𝐾Ω }
5
4.5
3.5
Voltmeter Readings [𝑉𝑉]
2.5
1.5
0.5
0
0% 10% 20% 30% 40% 50% 60% 70% 80% 90% 100%
Potentiometer [𝑉𝑉]
{ PAGE 5 }
{ Resistance =10 𝐾𝐾Ω }
6
4
Voltmeter Readings [𝑉𝑉]
0
0 1 2 3 4 5 6 7 8 9 10
Potentiometer [𝑉𝑉]
{ PAGE 6 }
Discussion and Conclusions:
The experiment aimed to investigate the variation of output voltage in a
potentiometer used as a position transducer. The objectives included establishing the
position-voltage relationship, analyzing linearity, and exploring potential error sources
and limitations. The DE LORENZO (DL2312HG) training system was used for the
experiment, providing a comprehensive platform for learning about control systems,
transducers, and signal conditioning.
The theory was based on potentiometry and electrical circuits, with the
potentiometer acting as a voltage divider. The ideal expectation was a linear
relationship between position and voltage, but deviations were anticipated due to
manufacturing tolerances. Factors such as mechanical friction, electrical noise, and
temperature effects were examined as potential error sources.
{ PAGE 7 }
Al-Balqa Applied University
Faculty Of Engineering Technology
Mechanical Engineering Department
Second Semester { 2022 - 2023 }
Course Code : { MEE0413 }
Course Title :
{ ENGINEERING MEASUREMENTS LAB }
Class Time : { TUESDAY }{ 08:00 - 11:00 }
Objectives:
The objective of a strain gauge experiment is to measure and analyze the strain or
deformation in a material caused by external forces or loads.
Strain gauges are devices that change their electrical resistance when subjected to
mechanical strain, making them suitable for measuring and quantifying the strain in
various applications.
Theory:
The theory behind a strain gauge experiment is based on the piezoresistive effect.
Strain gauges are devices that change their electrical resistance in response to
mechanical strain or deformation applied to a material. This change in resistance is
proportional to the strain experienced by the material.
By using a Wheatstone bridge circuit, the strain gauge's resistance change can be
measured and related to the applied strain.
Strain gauge experiments provide valuable data for stress analysis, structural
integrity assessment, material characterization, and validation of analytical models,
contributing to improved designs and optimized performance in various applications.
Materials:
DE LORENZO (DL2312HG ):
{ PAGE 1 }
experimentation in areas such as position control, sensor technology, signal
processing, and measurement techniques.
Procedures:
I} Start by connecting the electrical circuit as illustrated in the manual. Make sure to
follow the provided schematic diagram or wiring instructions carefully to ensure
correct connections.
II} Once the circuit is properly set up, turn on the device using the appropriate key.
Ensure that all necessary power sources are connected and functioning correctly.
III} Begin the experiment by adjusting the screw above the sensor. Increase the number
of turns gradually to apply force to the strain gauge. Take note of the number of
turns made for future reference.
IV} As the force is applied to the strain gauge, it will experience deformation, resulting
in changes in its electrical resistance. These changes will generate corresponding
voltage variations.
V} Observe the galvanometer connected to the circuit. The galvanometer measures and
indicates the generated voltages. Monitor the galvanometer's needle for any
changes in voltage.
VI} Repeat the process of adjusting the screw and recording the voltages for different
levels of applied force. This will allow you to gather data across a range of strain
values.
VII} Once you have collected sufficient data, analyze the relationship between the
applied force and the corresponding voltage generated by the strain gauge. You can
plot a graph or perform calculations to understand the strain gauge's behavior and
determine its sensitivity.
Data:
Number of Turns Voltages [𝑉𝑉 ]
0 0
1 0.5
2 1.5
3 2
4 3
5 4
6 5
7 6
8 6.5
9 7
{ PAGE 2 }
Results:
{ Strain Gauge }
7
4
Voltages [𝑉𝑉]
0
0 1 2 3 4 5 6 7 8 9
Number of Turns
{ PAGE 3 }
Discussion and Conclusions:
In the Strain Gauge Experiment, the main focus is on the analysis of the obtained
data and the determination of the strain factor from the generated graph. Based on the
graph, the strain factor is calculated to be 0.8333.
The graph obtained from the experiment represents the relationship between the
applied force and the corresponding voltage generated by the strain gauge. By
analyzing the data points on the graph, it is possible to draw conclusions about the
behavior and sensitivity of the strain gauge.
The calculated strain factor of 0.8333 indicates the ratio between the applied force
and the resulting strain experienced by the strain gauge. This value provides valuable
information about the sensitivity and responsiveness of the strain gauge to external
forces. A higher strain factor suggests that the strain gauge is more sensitive to applied
forces, while a lower value indicates lower sensitivity.
{ PAGE 4 }