1990_Book_AdvancedPracticalOrganicChemis
1990_Book_AdvancedPracticalOrganicChemis
M.CASEY
J.LEONARD
B. LYGO
Department of Chemistry and Applied Chemistry
University of Salford
G. PROCTER
George Ramage Professor of Chemistry
University of Salford
1 General Introduction 1
2.1 Introduction 3
2.2 The laboratory notebook 3
2.2.1 Why keep a lab book? 3
2.2.2 How to write a lab book 4
2.2.3 Suggested notebook format 5
2.3 Keeping records of data 8
2.3.1 What type of data should be collected 8
2.3.2 Formats for data records 9
3.1 Introduction 13
3.2 Setting up the laboratory 13
3.3 Generallaboratory equipment 14
3.4 The individual bench 20
3.4.1 Routine glassware 20
3.4.2 Personal items 21
3.4.3 Specialized personal items 21
4.1 Introduction 28
4.2 Purification of solvents 28
4.3 Drying agents 29
4.4 Drying of solvents 33
4.5 Solvent stills 39
viii Contents
5.1 Introduction 43
5.2 Classification of reagents for handling 44
5.3 Techniques for obtaining pure and dry reagents 45
5.3.1 Purification and drying ofliquids 45
5.3.2 Purifying and drying solid reagents 48
5.4 Techniques for handling and measuring reagents 50
5.4.1 Storing liquid reagents or solvents under inert atmosphere 50
5.4.2 Bulk transfer of a liquid under inert atmosphere (cannulation) 51
5.4.3 Using cannulation techniques to transfer measured
volumes of liquid under inert atmosphere 54
5.4.4 Use of syringes for the transfer of reagents or solvents 57
5.4.5 Handling and weighing solids under inert atmosphere 64
5.5 Preparationof~omethane 70
5.5.1 Safety measures 71
5.5.2 Preparation of ~omethane (a dilute ethereal solution) 71
5.5.3 General procedure for esterification of carboxylic acids 73
5.5.4 Titration of diazomethane solutions 73
6 Gases 74
6.1 Introduction 74
6.2 Use of gas cylinders 74
6.3 Handling gases 77
6.4 Measurement of gases 79
6.5 Inert gases 84
6.6 Reagent gases 84
7 Vacuum pumps 88
7.1 Introduction 88
7.2 Low vacuum pumps 88
7.2.1 Water aspirators 88
7.2.2 House vacuum systems 89
7.2.3 Electric diaphragm pumps 89
7.3 High vacuum pumps 90
Contents ix
8.1 Introduction 94
8.2 Reactions with air sensitive reagents 95
8.2.1 Introduction 95
8.2.2 Preparing to carry out areaction under inert conditions 95
8.2.3 Drying and assembling glassware 96
8.2.4 Typical reaction set-ups using a double manifold 97
8.2.5 Basic procedure for inert atrnosphere reactions 98
8.2.6 Modifications to basic procedure 101
8.2.7 Use of baUoons for holding an inert atrnosphere 106
8.2.8 The use of a 'spaghetti' tubing manifold 108
8.3 Reaction monitoring 110
8.3.1 Thin layer chromatography 110
8.3.2 High performance liquid chromatography 117
8.3.3 Gas-liquid chromatography 121
8.4 Reactions at other than room temperature 125
8.4.1 Low temperature reactions 125
8.4.2 Reactions above room temperature 128
8.5 Driving equilibria 133
8.5.1 Dean and Stark traps 133
8.5.2 High pressure reactions 134
8.6 Agitation 135
8.6.1. Magnetic stirring 135
8.6.2. Mechanical stirrers 136
8.6.3. Mechanical shakers 139
8.6.4. Sonication 139
12 Characterization 201
12.1 Introduction 201
12.2 Nmr 202
12.3 Ir 202
12.4 Uv 203
12.5 Mass spectra 203
12.6 M.p. and b.p. 204
12.7 Optical rotation 204
12.8 Microanalysis 205
12.9 Keeping the data 205
17 Safety 238
17.1 Safety is your primary responsibility 238
17.2 Safe working practice 239
17.3 Common hazards 239
17.4 Accident and emergency procedures 241
17.5 Bibliography 242
Appendices 243
Index 257
CHAPTER 1
General Introduction
2.1 Introduction
No matter how high the standard of experimental technique employed
during areaction, the results will be of litde use unless an accurate record is
kept of how that reaction was carried out and of the data obtained on the
product(s). Individuals or individual research groups will develop their
own style for recording experimental data, but no matter what format you
choose to follow, there are certain pieces ofvital information which should
always be included. In this section a format for keeping records of
experimental data will be suggested and although this need not be stricdy
adhered to, it will be used to point out the essential features which should be
included. It is suggested that records of experimental work and
experimental data be kept in two complementary forms: The lab notebook
should be a diary of experiments performed and should contain exact details
of how experiments were carried out; A data book or set of data sheets
should also be kept to record the physical data and preferred experimental
procedure for each individual compound which has been synthesized.
fonnat for polished report writing, but a daily log of work carried out in the
lab. Some of the main reasons for keeping a lab book are:
1. In order that the exact procedure followed for areaction can be referred
to later. This can be very important even if the reaction was not
successful. For instance, after several attempts to bring about a
re action have failed, it is often possible to review what has been done
then carry out a more successful experiment.
2. It should be the main index point that will enable you to find
experimental, literature and spectroscopic data on any compound which
you have synthesized.
3. It is the main source of reference when you come to write reports,
papers, theses etc.
4. It is a chronological diary of the experiments carried out and thus it
should allow you to say exactly when a particular experiment was
carried out.
5 . In order that another worker can follow your work, it is very important
to use a lab book style which is easily understood by others.
1 . General layout
It is good practice to start each new experiment on the next free right
hand page of the notebook. This makes finding any particular
experiment easier.
2 . Experiment number
The experiment number is in the top right hand corner of the page and
this is very important since it is used to reference all the compounds
which are prepared. If a notebook with numbered pages is used, it is
common for the experiment number to be the number of the page on
which the experiment starts. The way in which the notebook is indexed
is open to personal preference. In this system a researcher's first book
is book A, then B,C,D etc. Figure 2.1 therefore shows experiment 23
of book A. Compounds isolated from this experiment all carry the
number A23, prefixed with the researchers initials (in this case BB).
When more than one product is isolated from areaction a suffix, a, b, C
etc. is added to the reference number, a being the spot running highest
on tlc, b the next, and so on. Thus, for this experiment two products
were isolated and these carry reference numbers BB A23a and BB
A23b. Using this system the origin of any synthetic sampie can be
deterrnined very quickly.
3. Thedate
It is important that the date is always included.
4. Areaction scheme indicating the proposed transformation
This is always included at the top of the page so that an individual
experiment is easily found. If the reaction proceeded as desired the
scheme is left intact, but if the desired product was not obtained it can
be crossed through in red to indicate this. If other products were also
obtained they can be added, again in a different colour ink if desired.
Thus, simply flicking through the lab book, looking at the schemes,
can quickly provide a good deal of information. Some people prefer to
write only the left-hand side of the equation until the experiment is
complete.
6 Keeping Records of Laboratory Work
st(OBO
H H
OH
5+(.0&
H OBn
NaBHiceCI3
MeOHlo'C
•
HO H OBn
BB21a mw=348 mw=350
Ref. J.-L. Luche. L. Rodriguez-Hahn and P. Crabbe. J. Chem. Soc .• Chem. Commun ..
1978.601
Method:
The aldehyde (200mg) and CeCl3 solution 2ml) in MeOH (25mi). was cooled to
OOC. then treated with NaBH4 (27mg in MeOH. 8ml) .
.. .. Rxn
SM
.. ..
SM
Je Je
Rxn
After 30 min tlc shows no SM. but two products. MeOH was evaporated. CH2Cl2
(30ml) added and the mixture washedwith 10% HCI (lOml)foliowed by satd. NaHC03 (3
X10ml). dried and evaporated. (210mg crude)
it is therefore impossible to give a hard and fast rule stating the type of data
which should be obtained on every compound. Because of the rapid
expansion in the variety of techniques available there is presently much
debate about what constitutes a rigorous structure and purity analysis.
Traditionally, microanalysis was held to be the single standard test to which
any new organic molecuIe was subjected, and in some instances you may
find that it is a mandatory requirement. This is often the case for patent
applications. However, many researchers now prefer to use a range of
spectroscopic and chromatographie techniques to determine the structure and
purity of their compounds, and each researcher or research group will be
biased towards the particular types of data they find appropriate to
characterize the type of compounds that they work with.
For each individual compound you should decide first of all how
rigorously it needs to be characterized, then decide upon the appropriate
techniques to do this. For known compounds structure proof is often trivial
and simple techniques such as mixed melting point should not be
overlooked. If you are working on a synthetic sequence where the
structures of some of your intermediates are wen established, it is not
always necessary to treat each individual compound as a compIete
unknown. This is especially true if the reaction which has been carried out
is a very simple one, such as the reduction of an aldehyde to a primary
alcohol. On the other hand, when a crucial reaction has been carried out you
should be very careful to choose a characterization technique which gives
you the structural information you require, unambiguously.
flexibility which is often required to record diverse types of data which may
be used to characterize a particular compound; and it tends to encourage the
mistaken idea that every compound requires the same characterization data.
The system we prefer for data records, is one which makes use of a micro
computer word processor. A standard format blank data sheet, as shown
be1ow, is kept on disk (Fig. 2.2).
DA TA SHEET
Chemical Name:
Scheme:
Lit. Refs:
Method:
Mol. Formula:
m.p.lb.p.: ·c
Tlc : Rf (pet. ether/ethyl acetate, )
[aP25 : (e =, CHC13)
1H nmr (MHz): ~
J3C nmr (MHz): ~
V max : em- 1
A. max : nm
mlz (NH4, CI):
Found M , ealc. for
Analysis: Found C%; H%; N%; 0%,
Calc. for CHNO = C%; H%; N%; 0%
Notebook:
Spectra:
Figure 2.2
When a new compound is synthesized a blank data sheet is modified
with the data of the compound and a data record is created which is tailored
to the needs of the particular compound. Any of the data types which are
inappropriate for characterization of a particular compound can be removed,
and any additional data types can be added if required. This is especially
Keeping Records ofLaboratory Worlc 11
(±)-E xo,e xo -7,8- D ib enzyloxy -e xo -e is- bieyelo [3.3.0] -oet -2 -en -4-01-2-
m
earboxaldehyde
H OBn ..
a) Swem oxidn.
HO H OBn
Method:
To a stirred solution of oxalyl chloride (l00ml, 1.10mmol) in dry dichloromethane
(15ml) at -78°C, a solution of dimethylsulphoxide (200ml, 2.58mmol) in
dichloromethane (lml) was added dropwise. The reaction mixture was stirred for 5 min,
then epoxide (200mg, 0.55mmol) in dichloromethane (2ml) was added dropwise. The
resulting mixture was stirred for a further 20 min, then triethylamine (Iml, 7.61mmol)
was added. After 10 min at -78°C the mixture was allowed to warm to room temperature,
then poured into 2M HCI (30ml) and extracted with dichloromethane (2 x 30mI). The
extract was washed with saturated aqueous NaHC03 (4OmI), then dried, and the solvent
removed in vacuo.
The crude product was taken up in dichloromethane (15mi), 1,5-
diazabicyclo[5.4.0]undec-5-ene (167mg, 1.1 mmol) was added, and the mixture was stirred
at room temperature for 2h. The solution was then poured into 2M HCI (20mi), extracted
with dichloromethane (2 x 30mi), and washed with saturated aqueous NaHC03 (40m1),
then dried, and the solvent removed in vacuo. Tlc. examination of the cmde mixture
showed the presence of one major product. An analytical sampie of the hydroxy enal was
obtained by flash chromatography (1:1 pet. ether/ethyl acetate), yie1d, 80%.
useful for nmr data, which can be as simple as a list of 60MHz signals, or
could comprize a vast array of information from a high field FT instrument
such as, coupling constants, nOe's or 2D data. Once the data sheets have
been printed out they are kept in a ring file, to make a very flexible data
book. Similar non-computer based record systems can also be devised but
the advantage of using a computer is that the record can easily be updated at
any time. When designing our data sheet we decided upon a standard style
which we also use for experimental sections of reports and theses.
Therefore, if a researcher is conscientious about keeping data records up to
date the tedium is taken out of reporting results. Also, with only slight
modifications, the data record can become an experimental entry which fits
the style of any major journal. An example of a completed data record is
shown in Fig. 2.3. Notice that there are some quite detailed high field nmr
data which are represented in a table. The table is also a standard design and
provides a good way of presenting complex nmr data so that they can be
deciphered quickly.
CHAPTER 3
3.1 Introduction
When setting up the lab it is usual for some areas of bench space to be
set aside for communal apparatus, and other parts to be allocated as
individual benches. Clearly, the areas which are assigned as communal
bench space will depend on the amount and type of communal equipment
which is to be installed. In this chapter some pieces of equipment will be
identified as communal and others as part of the individual bench kit, but the
classification will vary from one laboratory to another. The distinction will
to some extent depend on the type of work being undertaken, but will also
depend on the budget and space available.
U nless alternative office space is provided, it is a good idea to have
some desk space in the lab where workers can read and write, away from
areas used for chemicals. Desk space may also be required for small
computers which are also becoming a common feature of the modem
organic chemistry lab. Drawers or filing cabinets are useful for the safe
storage of spectta and other paperwork, and a blackboard is an invaluable
laboratory aid.
The area which constitutes an individual'bench' will vary considerably
from one lab to another. In our view all procedures involving organic
chemicals should be carried out in an efficient fume cup board. This implies
that each full-time worker in an organic chemistry lab permanently requires
at least one mette of fume cupboard space. However, in practice it is often
the case, particularly in academic labs, that much less fume cupboard space
is actually provided, and fume cup boards may be communal. In this chapter
the term 'bench' will refer to the space occupied by an individual worker
and it will be assumed that this space incorporates an adequate area of fume
cupboard, where all reactions are carried out.
members of the lab, should be located so that they are easily accessible,
without causing a disturbance to anyone working elose by.
Rotary evaporators
Rotary evaporators are perhaps the most heavily used pieces of equipment in
an organic research lab. They should therefore be located conveniently
throughout the lab, or preferably, each worker should have his or her own
on the bench.
Fridge and/or jreezer
A fridge and/or freezer should be provided to store chemicals, and this
should never be used for food storage.
Glass drying oven(s)
Again these should be conveniently located.
Vacuumoven
This may be used infrequently, but is still a very valuable piece of
equipment.
Balances
A two place balance is indispensable and a four place balance needs to be
available for sma11 scale work. When locating these remember that they are
used frequently, by people carrying out careful work, so do not put them in
the way of others, also avoid drafts and vibrations.
Kugelrohr bulb to bulb distillation apparatus (e.g. Buchi)
This apparatus is invaluable when preparing relatively small quantities of
high-boiling liquids. It is relatively mobile and can be moved from bench to
bench and attached to various vacuum sourees. However, it is useful to
locate it near to a high vacuum pump, where it is normally used.
Vacuum pumps
Vacuum at various levels will be required around the lab. Each bench
should have a modest vacuum source available (about 20mmHg), and this
can be provided by a water aspirator, a house vacuum system or a small
diaphragm pump. This level of vacuum is sufficient for rotary evaporation
of most solvents, filtering under vacuum, distillation of relatively volatile
oils, and similar tasks. However, there are a variety of operations for which
a high vacuum pump is either preferable or necessary. These operations
16 Equipping the Laboratory and the Bench
/~ =t t..---t--,
W~
~ connected to
high vacuum
.....-,~ pump via traps
connected 10
vacuum gauge
high vacuum
stopcock
Figure 3.1
lnen gases
A constant supply of dry inert gas (usually argon or nitrogen) is now
essential in an organic chemistry lab. Some labs will have nitrogen or argon
on tap and this is ideal, provided there are enough outlets and the gas is dry.
However, in most labs gas is supplied from cylinders, which take up a lot
of space and are quite expensive to rent. It causes great inefficiency in a
modem lab if there are too few inert gas outlets, and it is therefore important
when setting up the lab to decide how many gas outlets are required. We
suggest that there should be at least one outlet for each person working in
the lab, plus one for each piece of apparatus (such as a still) which is kept
permanently under inert gas. This may suggest that the same number of
cylinders is required, but the cost of this is usually prohibitive and space can
also be a problem. One efficient way to use cylinders to service a large
number of permanent inert gas lines, is to fit them with multi-way needle
Equipping the Laboratory and the Bench 17
valves (e.g. 3-way). These are available from gas line hardware suppliers at
a very modest cost and give several gas outlets for the price of one!
Having decided on the number of gas cylinders required these should
be positioned in the laboratory so that semi-permanent PVC (Tygon) tubing
lines can easily be taken to each bench and to each piece of apparatus
requiring a supply. Some of the apparatus commonly used in conjunction
with inert gases will be described below and in Chapters 6 and 8.
Solvent stills
Two basic types of solvent still are found in the lab, one is for distillation of
solvents for routine use and the other is for distillation of ultra-dry solvents
for carrying out reactions under dry conditions. Unless very high grade
solvents are purchased, most solvents used in the lab for reactions or
chromatography should be distilled. For solvents which are used regularly
it is best to set up permanent stills and some typical designs for these are
given in Chapter 4. Routine stills are usually quite large (up to 5 litres), and
so therefore take several bottles of solvent at a time. Permanent large scale
stills should normally be available for petroleum ether (usually the largest
still), ethyl acetate, methylene chloride and any other solvents, according to
your particular needs.
Some solvents, such as tetrahydrofuran and diethyl ether (there may be
others you require also) are frequently required in very dry form, and it is
inconvenient to set up a still each time a small quantity is required for a trial
reaction. The efficiency of an organic lab is therefore greatly enhanced by
having these dry solvents available 'on-tap' from permanent stills, which
provide very effective drying, and are kept under an inert atmosphere.
Modern designs for permanent stills used for these distillations are very
compact, and do not cause inconvenience (see Chapter 4). A permanent still
should be installed for any solvent which is used regularly in the lab. It is
also worth keeping aspare distillation apparatus on hand for occasions
when a less common dry solvent is required.
General distillation equipment
Apart from standard Quickfit equipment, labs should also have some one-
piece distillation kits, which provide more effective distillation. A short-
path distillation apparatus is very useful for low-holdup, high-throughput
18 Equipping the Laboratory and the Bench
~E
ABC D E F
Size 1 5.5crn 3.5cm 6crn 2crn B14 B14
Size 2 llcrn 3.5crn 13cm 3cm B24 B14
Figure 3.2
TB
tA
Bl4or~
B24
1 Alternative take
off for direet
B14
(14/20)
eoupling with
vaeuum
A B C
Sizel 8em 3.5cm 13em
Size 2 18em 3.5cm 18cm
Figure 3.3
Chromatographie equipment
Tlc is by far the most widely used technique for rapid, routine reaction
monitoring and it is essential that the lab should have permanent facilities for
visualization of t1c plates. A range of solvent dips, an iodine tank and a
small uv lamp are usually all that is necessary (see Chapter 8).
More sophisticated chromatographie techniques are also finding much
wider application in organic chemistry labs, and both hp1c and capillary gc
instruments are now quite common-place. These techniques complement t1c
20 Equipping the Laboratory and the Bench
described here, but for full details of how they are used see the appropriate
chapter.
The double manifold
P------------ 25cm--------------~
inert gas from cylinder
~ via bubbIer
Figure 3.4
'"
2-way double
oblique tap - see Fig. 3.5
If you would like to carry out reactions under dry and/or inert
conditions as a matter of routine, the double manifold is perhaps the single
most useful item of equipment which will enable you to do this. We
recommend that this be a standard item of equipment, permanently fitted to
every individual bench and connected to the laboratory inert gas supply by
PVC (Tygon) tubing. The manifold consists of two glass barrels, one
evacuated and one filled with an inert gas. An outlet is supplied by either
barrel of the manifold via a two-way double oblique tap (see Figs. 3.4 and
3.5). Thus, at the turn of the tap, equipment connected to the manifold can
be alternately evacuated or filled with inert gas.
r
12cm
+-+- anti suck-back
valve
1
++--- oil
Figure 3.6
A rotary pump is normally connected to the vacuum barrel of the
manifold and a schematic diagram showing the complete set-up is shown in
Fig. 3.7.
PCV (Tygon) tubing
/
~===:==~==~==>=========~========~~r=~,+--inertgas
from cy linder
regulator
double
1
vacuum
tubing bubbIer
manifold
~vacuum
solvent tubing
traps
rotary pump
Figure 3.7
Although it is preferable for workers to have their own individual
manifold, in some labs this may not be possible, because of insufficient
fume cupboard space or lack of pumps. In this case, manifolds can be
treated as communal items of equipment, but discipline must be observed to
keep the manifold clean and avoid cross-contamination.
Another type of manifold which is useful for carrying out reactions
under inert atmosphere is the 'spaghetti tubing' manifold (Fig. 3.8). This is
24 Equipping the Laboratory and the Bench
a single barrel inert gas manifold, fitted with narrow bore Teflon tubing
outlets. Each outlet has a syringe needle attached which can be pushed
through a septum to provide an inert atmosphere within a flask. This type
of system is particularly useful for small scale work.
+-- stopcock
(optional)
Figure 3.8
Full descriptions of how to use manifolds are given in Chapters 5 and
8.
Three-way Quickfit gas inlet T taps
A simple piece of equipment which is applicable to a variety of tasks is a 3-
way teflon tap connected to a Quickfit cone joint (Fig. 3.9).
2=~'~1 2=f'i
3.4 rnrn bore
stopcock
~ +- ~2~~40) 1.2 rnrn bore
stopcock
PTt +- B 14
(14/20)
Figure 3.9
These are so universally useful that we suggest you have at least three
of the B 14 size and two of the B24 size as part of your personal bench set.
They are particularly useful when used in conjunction with a double
manifold. With the inert gas from the manifold connected to the horizontal
inlet and the tap in position A (Fig. 3.10), areaction flask can be kept under
a slight positive inert gas press ure. If the gas flow is increased and the tap
is turned to position B, liquids can be introduced via the vertical inlet, whilst
maintaining an inert atmosphere (see Chapters 5 and 8 for more details).
Equipping the Laboratory and the Bench 25
Another simple use is for connecting fIasks to a high vacuum system for
removal of last traces of solvent.
tap
position
position B position C
~r ~l-- ~l
flow through
tap
Figure 3.10
Filtering aids
Rapid filtration is often needed, and the speed of filtration is increased
dramatically by applying pressure or a vacuum. There are two types of one-
piece sintered funnels wh ich we find very useful, as shown in Fig. 3.11.
The parallel-sided type are constructed by fusing a circular sinter into the
appropriate diameter glass tubing, which is then joined to a cone joint and a
piece of narrow-bore tubing (about lOmm od). It is useful to have two or
three of these in the bench kit, ranging in diameters of between 1 and 10cm.
The larger ones are particularly valuable for filtering off drying agent,
leaving the the dried solution in a round-bottom fIask, from which solvent
can be evaporated directly. One or two of the smaller Hirsch-style funnels
are useful and these are more commonly used for filtering off crystals after
small-scale recrystallizations, the mother liquor being conveniently
deposited in a round-bottom fiasko These can be made starting from a
sintered funnel, but a glassblower can make them more cheaply by inserting
a circular sinter into a narrow tube, then forming the funnel from this.
Two other pieces of equipment of value as filtration aids are the Craig
tube (Fig. 3.12a, see Chapter 9), used for very small recrystallizations, and
a sintered funnel (Fig. 3.12b) for filtration under inert atmosphere (see
Chapter 5 for more details).
26 Equipping the Laboratory and the Bench
+ - - C-----+
A------+ 1D
"""""""'"
1
1·····1
E
1
A B C D E
size 1 5em 4em size 1 5em 5em 5.5em
size 2 7.5em 4em size 2 7.5em 7em 5.5em
size 3 lOem gem 5.5em
Figure 3.11
B14
i~ ~ I
(14/20)
12em
lOem '\. solid
1
glass
1
+- glass
tubing +--- porosity 3
sinter
+-- PI'FE
+-824
(25/40)
+-B24
1
(25/40)
25cm
+-500ml
+-Teflon
stopcock
Figure 3.13
CHAPTER 4
4.1 Introduction
The use of appropriately purified solvents is vital to the success of the
procedures for which they are used. It is important to note that the degree of
purity and dryness required depends on the intended application, so when
choosing a method from this chapter remember to pick one which is
appropriate for your purpose. Consult Appendix 1 for useful data about
commonly used organic solvents.
Remember that solvents are hazardous materials and
beware particularly of the flammability of the hydrocarbons
and ether, the toxicity of benzene, chloroform, and carbon
tetrachloride, and the possibility of peroxide contamination of
ethereal solvents.
are available but they are more expensive and less reliable than solvents
dried by the techniques described below.
2. Solvents used for spectroscopy, especially nmr and uv, should be of
high purity. Many suppliers provide 'spectroscopic grade' solvents
which are particularly suitable for uv spectroscopy because ultraviolet
absorbing impurities have been removed.
3. Solvents used for chromatography should always be fractionally
distilled to ensure that non-volatile impurities are removed. Solvents
for hplc should be of high purity and again many suppliers provide
special'hplc grade' solvents, wh ich have been purified and filtered to
remove contaminants which might degrade hplc columns.
4. All applications involving quantitative analysis require the use of
'analytical grade' (typically >99.5% purity) solvents. It is good
practice in general to use high quality solvents for the purification of
your products, and it is particularly important to use pure solvents
when purifying sampies for microanalysis.
In most cases purification of a solvent simply involves drying and
distilling it, so the next sections are devoted to drying agents and methods
used for drying common solvents, and the final section contains
descriptions of typical continuous stills. Apart from water the only other
commonly encountered contaminant is peroxidic material formed by aerial
oxidation of ethereal solvents. Methods for dealing with these dangerous
impurities are described in the section on the purification of diethyl ether.
must be taken to ensure the safe destruction and disposal of excess reagent
remaining in solvent residues.
When the solvent is to be distilled after standing over adesiccant, the
drying agent should be filtered off before distillation if it removes water
reversibly, e.g. by hydrate formation (MgS04, CaCI2), or by absorption
(molecular sieves). The solvent can be distilled without removal of the
desiccant in cases where water removal is irreversible (CaR2, P20s).
The recommendations in this and the following section are largely
based on the work of Burfield and Smithers et al. who have carried out
quantitative studies on the efficiency of drying agents for a wide range of
solvents. 1 Their work has supplanted earlier studies many of which are of
doubtful reliability. ether useful sources of information inc1ude Perrin and
Armareg02, and Riddick, Bunger, and Sakano.3
Alumina, Al203
Neutral or basic alumina of activity grade I is an efficient drying agent for
hydrocarbons, 5% w/v loading giving extremely dry solvents. It is also
useful for the purification of chloroform.
Barium oxide, BaO
The commercially available anhydrous product is an inexpensive drying
agent which is useful for amines and pyridines (30-50ppm after standing for
24h over 5% w/v). It is strongly basic and is ineffective for alcohols and
dipolar aprotic solvents.
Boric anhydride, Bz03
Recommended drying agent for acetone, and effective also for thorough
drying of acetonitrile.
1. (a) D.R Burfield, K.H. Lee, and RH. Smithers, J. Org. ehern., 1977,42, 3060; (b)
D.R Burfield, G.H. Gan, and RH. Smithers, J. Appl. ehern. Biotechnoi., 1978,
28, 23; (e) D.R Burfield and RH. Smithers, J. Org. ehern., 1978, 43, 3966; (d)
D.R Burfield and RH. Smithers, J. ehern. Technol. Biotechnoi., 1980, 30, 491; (e)
D.R Burfield and RH. Smithers, and A.S.C. Tan, J. Org. ehern., 1981,46,629;
(f) D.R. Burfield and RH. Smithers, J. ehern. Educ., 1982,59,703; (g) D.R
Burfield and RH. Smithers, J. Org. ehern., 1983,48,2420.
2. D.D. Perrin and W.L.F. Armarego, Purification o[Laboratory Chemieals, 3rd ed.,
Pergamon, Oxford, 1988.
3. J.A. Riddiek, W.B. Bunger, and T.K. Sakano, Organic Solvents. Physical Properties
and Methods o[ Purification, 4th ed., Wiley-Interscienee, New York, 1986.
Purification and Drying of Solvents 31
stored in a desiccator or in an oven at >lOO·e for a few weeks but they are
rapidly hydrated in the air. If you are doubtful about the effectiveness of an
old batch, place a few beads in the palm of your hand and add a drop of
water - if the sieves are active you should feel a distinct1y exothermic
reaction. In most cases extremely dry solvent can be obtained simply by
batchwise drying over sieves, i.e. aIlowing the solvent to stand over 5%
w/v of sieves for 12h, decanting, adding a second batch of sieves etc.
Sieves absorb water reversibly so a solvent should always be decanted from
the sieves prior to distillation. 4A Beads are recommended for thorough
drying of amines, DMF, DMSO, and HMPA, and almost all rigorously
dried solvents are best stored over 5% w/v of 4A sieves. However,only
the 3A form is suitable for drying acetonitrile, methanol, and ethanol, and
higher alcohols require the use of powdered 3A sieves. They are not useful
for drying acetone because they cause self-condensation. Provided that they
are not discoloured, sieves can be reused by washing weIl with a volatile
organic solvent, allowing to dry, drying at lOO·C for several hours, and
then reactivating at 300·C.
Plwsphorus pentoxide, P20S
(Causes burns) Although a rapid and efficient desiccant its use is
limited by its high chemical reactivity. It reacts with alcohols, amines,
acids, and carbonyl compounds and causes significant decomposition of
HMPA, DMSO, and acetone. It is useful for drying acetonitrile and may be
used for hydrocarbons and ethers but is less convenient than other reagents.
It is often used in desiccators. It is best decomposed by careful portionwise
addition to ice-water followed by neutralization with base (do not add water
to P20s, the mixture may become so hot that the vessel could crack). It is
extremely efficient for drying gases and is available in a convenient form,
mixed with an inert support, so that it does not become syrupy.
Potassium hydroxide, KOH
(Causes burns) Freshly powdered KOH is a good drying agent for
amines and pyridines but is inferior to calcium hydride. It should not be
used with base sensitive solvents.
Sodium, Na
Sodium is widely used to dry hydrocarbons and ethers. It may be formed
into wire using a sodium press or used as granules by cutting the bars under
Purification and Drying of Solvents 33
petroleum ether. It suffers from the disadvantage that the metal surface
rapidly becomes coated with an inert material so it should not be used unless
the solvent is pre-dried. Sodium reacts with benzophenone to give a dark
blue ketyl radical which is protonated by water to give colourless products.
Thus the sodium-benzophenone system is particularly convenient because it
is self-indicating, and it is the preferred reagent for rigorous drying of ether,
THF, DME, and other ethereal solvents. Sodium-potassium alloy has been
recommended because it is liquid and therefore its surface does not become
coated so easily, but this advantage is outweighed by the increased danger
resulting from the use of potassium. Sodium residues can be destroyed by
slow careful addition of ethanol until hydrogen evolution ceases. The
mixture should then be stirred weIl, to ensure that no coated lumps of
sodium remain, before carefully adding methanol. After leaving for several
hours the mixture should be stirred again to ensure that all of the sodium has
been consumed, and then the mixture should be added cautiously to a large
excess of water before disposal.
Sodium should never be added to chlorinated solvents
because a vigorous or explosive reaction could occur.
Sodium sulphate, Na2S04
Anhydrous sodium sulphate is a weak drying agent suitable only for drying
extracts. It is preferable to magnesium sulphate for drying very acid
sensitive compounds.
Aceticacid
(Causes bums) Acetic acid is very hygroscopic. It can be dried by
adding acetic anhydride (3% w/v) and distilling (b.p. 118°C). Reagent
grade acetic acid usually contains some acetaldehyde. If this is likely to
cause problems add chromium trioxide (2% w/v) as wen as acetic anhydride
before distilling, or use analytical grade material.
Acetonitrile
(Toxic) Preliminary drying is accomplished by stirring over potassium
carbonate for 24h. A further 24h over 3A sieve or boric anhydride gives
moderately dry solvent (- 5Oppm) but much better results are obtained by
stirring over phosphorus pentoxide (5% w/v) for 24h. and then distilling 1(a).
Drawbacks of this method are the formation of substantial quantities of
coloured residue, and the possibility that the product is contaminated with
traces of acidic impurities. If the acetonitrile is required for use with very
acid sensitive compounds it is best to redistil it from potassium carbonate.
Ammonia
Distil from the cylinder into a flask cooled to <-40°C and fitted with a dry-
ice condenser (Chapter 14). Add pieces of sodium until the dark blue colour
persists, and then distil the ammonia into your reaction vessel.
Benzene
(Carcinogenic) Benzene, like most hydrocarbons is very easy to dry.
No preliminary drying is required and several reagents will reduce the water
content to < Ippm. Alumina, calcium hydride, and 4A sieves (all 3% w/v
for 6h) are the most convenient drying agents and the benzene is then
distilled and stored over 4A sieves 1(a). Altematively benzene may be dried
over calcium hydride in a continuous still. Toluene may be dried in the
same way.
Purification and Drying of Solvents 35
tert-Butanol
Reflux over calcium hydride (5% w/v) and distil onto powdered 3A
sieves 1(g). Other low molecular weight alcohols, but not methanol, can be
dried in this way.
Carbon disulphide
(Highly flammable, toxic) Distil (using a water bath) from calcium
chloride or phosphorus pentoxide (2% w/v). Do not use sodium or
potassium.
Carbon tetrachloride
(Carcinogenic) See chloroform.
Chlorobenzene
See dichloromethane
Chloroform
(Toxic) Perhaps the simplest procedure is to pass the chloroform through
a column of basic alumina (grade I, lOg per 14mi). This removes traces of
water and acid and also removes the ethanol which is present as astabiliser.
Carbon tetrachloride may be purified in the same way. Larger volumes of
either solvent can be dried with 4A sieves, or by distillation from
phosphorus pentoxide (3% w/v). Distilled chloroform should be stored in
the dark to prevent formation of phosgene.
Addition of sodium to chloroform or carbon tetrachloride
may cause an explosion. Chloroform mayaiso react
explosively with strong bases and with acetone.
Cyclohexane
See petroleum ether.
Decalin (decahydronaphthalene)
Decalin is very easy to dry but it forms peroxides on prolonged contact with
air so it is advisable to use a drying agent which will reduce the peroxides.
Reflux over sodium for 2h and distil onto 4A sieves. Tetralin should be
treated similarly.
l,2-Dichloroethane
See dichloromethane.
36 Purification and Drying of Solvents
Dichloromethane
Reflux over calcium hydride (5% w/v) and distil onto 4A molecular sieves.
Chlorobenzene and 1,2-dichloroethane can be dried in the same way.
Dichloromethane can be dried over calcium hydride in a continuous still.
Never add sodium or powerful bases to chlorinated
solvents - an explosion may occur. Reaction of azide saUs
with dichloromethane resuUs in the formation of explosive
azides.
Diethyl ether (ether)
(Flammable) Ether itself, along with the other commonly used ethereal
solvents, THF, DME, and dioxan, can contain substantial amounts of
peroxides formed by exposure to the air. These peroxides can cause
serious explosions. Test for the presence of peroxides by adding Iml
of the solvent to Iml of a 10% solution of sodium iodide in acetic acid. A
yellow colour indicates the presence of low concentrations of peroxides
whereas a brown colour indicates high concentrations. Low concentrations
of peroxides must be removed before further purification and a number of
methods have been suggested4 • One frequently recommended procedure is
to shake the solvent with concentrated aqueous ferrous sulphate. Ethers are
usually pre-dried over calcium chloride or sodium wire, and rigorously
dried over sodium-benzophenone. Careful preliminary drying is necessary
because all of these solvents can dissolve substantial quantities of water.
The pre-dried solvent is then placed in a reflux apparatus or a continuous
still and sodium pieces (1 % w/v) and benzophenone (0.2% w/v) are added.
The mixture is refluxed under an inert atmosphere until the deep blue colour
of the ketyl radical anion persists. The ether may then be collected or
distilled onto 4A sieves. This drying method also removes the peroxides
wh ich are a serious hazard when handling ethereal solvents. If a continuous
still is used it will be necessary to add more sodium and benzophenone
occasionally. Eventually the still will become very murky as the
benzophenone reduction products accumulate. If this happens, or if the blue
colour no longer persists, it is time to distil most of the solvent - do not distil
to dryness. The sodium residues may then be destroyed as described in
4. D.R. Burfield, J. Org. ehern., 1982,47,3821.
Purification and Drying of Solvents 37
Methanol
(Toxie) To dry llitre of methanol place magnesium turnings (5g) and
iodine (0.5g) in a 2 litre flask fitted with a reflux condenser and add
methanol (50ml). Warm the mixture until the iodine disappears, and if a
stream of bubbles (hydrogen) is not observed add more iodine (0.5g).
Continue heating until all of the magnesium has been consumed and then
add the remainder of the methanol. Reflux the mixture for 3h , distil
(bumping) onto 3A sieve beads (10% w/v), and allow to stand for at least
24h 1(g).
Nitromethane
Dry by standing over calcium chloride, filtering, and distilling onto
molecular sieves. Do not use phosphorus pentoxide.
Pentane
See petroleum ether.
Petroleum ether (petrol)
(Flammable) This confusing name is used for mixtures of aliphatic
hydrocarbons containing smaller amounts of aromatic compounds. It is
generally supplied as several fractions each having a 20'C boiling range (40-
60, 60-80 etc.). Alkane mixtures which do not contain aromatic compounds
are supplied as pentane, hexane, cyc10hexane etc. All of these solvents are
readily drled by distilling, and standing over activity grade I alumina (5%
w/v), or over 4A molecular sieves.
Pyridines
(Toxie) Distil from calcium hydride onto 4A molecular sieves 1(e).
Tetrahydrojuran (THF)
See diethyl ether.
Tetralin (tetrahydronaphthalene)
See decalin.
Toluene
See benzene.
Xylene
See benzene.
Purification and Drying of Solvents 39
condenser_
collecting pot-
solvent
outlet
/
heating
'"
mantle
..A
~
Figure 4.1
40 Purification and Drying of SoIvents
in Chapter 9, and is used for the distillation of solvents that are either
required infrequently, or that can be stored without deterioration for long
periods of time. The other is a continuous still set-up that consists of a
distillation pot, collecting head, and condenser (see Fig. 4.1).
This type of still arrangement is used for solvents that are required on a
regular basis and the still system is usually left set up, although generally it
is only turned on when the solvent is required - it is not recommended that
any type of solvent still is left on unattendedfor prolonged periods oftime .
Continuous still systems typically have an upright arrangement that
takes up less space than the conventional still set-up and a collecting vessel
that is positioned between the still-pot and the condenser. The apparatus is
designed such that the distilling solvent is condensed and collects in a
collecting head. Once the collecting head is full the solvent simply
overflows back into the still pot, allowing continuous distillation without the
still boiling dry. The solvent can simply be drawn off from the collecting
head when required, or poured back into the still pot if not.
round-bottomed
flask
3-way tap
~
- collecting head
Figure 4.2
Purification and Drying of Solvents 41
(a) (b)
Figure 4.3
CHAPTER 5
5.1 Introduction
One of the key steps to becoming a successful practitioner of modern
organic chemistry is knowing how to handle and store air- and moisture-
sensitive reagents, with the certainty that they have not been contaminated.
This is a skill which takes some time to acquire and many people
unfortunately learn the hard way, after astring of failed reactions. An
efficient, successful organic chemist achieves good results rapidly, not by
cutting corners, but by rigidly observing strict working practices that allow
sensitive reagents to be used with confidence. The biggest waste of
research time sterns from employing reagents or procedures which you think
are 'OK'. It will usually take far more time to repeat areaction than it would
have taken to re-purify a suspect reagent before starting. Perhaps more
important than the time was ted by failed reactions, is the uncertainty which
is introduced by using suspect reagents. No matter how carefully the
outcome of an experimental reaction is quantified, in terms of yield,
stereoselectivity, by-products, etc., the data will be meaningless if there was
any uncertainty about the re action conditions or reagents.
Handling sensitive reagents confidently is not difficult once a few
standard techniques are learned and adhered to. In this chapter we will give
examples of simple general methods which can be employed to handle a
variety of reagents.
44 Reagents: Purification and Handling
Dimethyl formamide MgS04 before dist. 76'/39 do not use CaH2 or other
basic drying agents
BF3·Et20 dist. from CaH2 67'/43 reacts with moisture
SnCl4 dist. from tin high vac. reacts with moisture
Et2AlCI dist. from NaCI 107'/25 heated nichrome spiral in
50cm column; Reacts
with air - spontaneously
flammable!
Et3Al none 130'/55 ibid.
Benzaldehyde MgS04 before dist. 62'/10 wash with Na2C03before
dist. Store over
0.1 % hydroquinone
Benzyl bromide MgS04 before dist. 85'/12 dist. in dark. lacrymator
pressure) and the techniques are exactly as above. Some examples are given
in Table 5.3.
Table 5.3
Reagent Distil from b.p.oC Comment
Although extreme care must be taken to exclude air and moisture from
reagents kept under an inert atmosphere, once you become familiar with the
simple techniques outlined in this section, you should find them easy to use.
Bottles and flasks containing liquids or solutions sealed under an inert
atmosphere are widely used in modern organic chemistry. A knowledge of
the techniques required to manipulate such reagents, without allowing them
to come into contact with the atmosphere, is therefore essential.
When handling air sensitive reagents always think ahead, and design
the whole sequence of events you intend to follow so that air never comes
into contact with the reagent.
A method for titration of alkyllithium reagents is given in Chapter 16
and a method for titration of hydride reagents is given in Chapter 6.
the container with argon (or nitrogen). This is done using a syringe needle
connected to an inert gas manifold by a Luer adapter. Another needle is
used as a vent as shown in Fig. 5.1a. The septum should be secured with
copper wire and for extra protection another septum of the same size,
without any needle holes in it, can be pulled over the first (Fig. 5.1 band c).
The second septum. is removed before syringing liquid from the bottle. For
long term storage, it is also a good idea to wrap parafilm around the septum,
especially if the bottle is to be stored in the fridge.
inert---+
gas second
septum
"- ~-""
(b) (c)
(a)
Figure 5.1
If the septum has been fitted properly and the reagent is used carefully,
according to the techniques below, it can be kept for many months and used
many times over.
If you need to use areagent which has been stored in a fridge or
freezer, always allow it to warm up to room temperature and remove any
condensation, before unsealing it.
inert gas
from cylinder---+ c:::::::;-;::::c::::'-
restriet vent---+
with finger or
needle valve
to increase
press ure
~vent to
inert gas bubbIer or
from cylinder--+ C::::::;:-;::::::::I:t:~ to inert gas
system that
restriet vent--+ incorporates
with fmger or a bubbIer
needle valve
to increase
pressure
_._ .. __ ....... _-
-_ ....... -_.
. . . . . . . . 0' ___ •
.....
:::::.'1;-::::.
Figure 5.3
6. When all the liquid has been transferred, first remove the vent needle,
then remove the cannula at the receiver end. The remainder of the
system can then be dismantled in any order, but remember to be very
careful with the residues, particularly if the liquid is toxic or reactive to
moisture.
--+ to bubbier
(or manifold) cannula inlet
i +- septum
+-stopcock
inert---+
gas
inert --+ fli.I-ft'I\)----.,.",..J
gas
............... __ ..
-- ---..........
.. ....... _........ -
_....
::_-::_-:;'::::-
-_ .. _--------
(b)
(a)
Figure 5.4
inert---+
gas
--+ to bubbier
(or manifold)
... =.....-
-_.=
Figure 5.5
If the liquid is being measured for addition to a reaetion flask, an
alternative proeedure is to use a graduated pressure equalizing dropping
funnel attaehed to the apparatus. The required quantity of liquid ean then be
eannulated into the dropping funnel direet1y.
56 Reagents: Purification and Handling
Types oj cannula
For most purposes cannulation can be carried out using an ordinary double
ended needle, bent to a suitable shape (Fig. 5.6a). A cannula made by
joining two long syringe needles to a Luer to Luer stopcock allows the flow
of liquid to be controlled (Fig. 5.6b). For tansferring large volumes of
liquid the 'flex-needle' (available from Aldrich Chemical Co.) is useful.
This is a jacketed Teflon tube with a wide-bore needle attached to either end
(Fig. 5.6c). It provides rapid liquid transfer, and a degree of insulation for
a cold reagent. The wide-bore needles of the flex-needle should not be
inserted through a new septum without first making a hole with anormal
gauge needle.
Luer to Luer
stopcock
long --+
needles Teflon inner
insulation tubing
Figure 5.6
For small scale work all Teflon cannulas are very useful, and some
people prefer them to syringes. They are easily made by taking a piece of
narrow-bore (1.5mm) Teflon tubing, of a suitable length and cutting each
end at a shallow angle with a sharp razor blade or scalpel (Fig. 5.7).
Always be sure to make holes in the septa, using a metal needle, before
attempting to insert the Teflon cannula.
ends cut at an angle narrow-bore
using scalpel /Teflon tubing
~ ....:==:=:::::=---
Figure 5.7
Reagents: Purification and Handling 57
Types of syringe
Micro syringes
Micro syringes are extremely useful for sm all scale synthetic work,
delivering small quantities of liquid accurately. Sizes ranging from 10111 to
Iml are common, but 100111 is perhaps the most useful size to have.
There are two general types ofmicro syringe, gas-tight and liquid-tight.
The liquid-tight syringes have matched glass bores and stainless steel
plungers, which should not be mixed-up (Fig. 5.8). Certain types of
reagent will corrode the steel plunger and cause ceasure if the syringe is not
c1eaned immediately after use. These are indeed delicate and quite expensive
items which should always be treated with great care (see later for care of
syringes).
58 Reagents: Purification and Handling
ndOu-uu-uuu-u--uu-uu--u-u-u--UI======~O
wrurT-rTT-j-TTTT-j-TT u
•
------------~----o
screw-on needle
Figure S.8
Gas-tight micro syringes have Teflon tipped plungers, making them
more inert to chemical reagents (e.g. Lewis acids). The barrels and
plungers are also interchangeable and they are thus much more reliable than
the liquid-tight syringes. The drawback with gas tight syringes is that they
are more expensive (Fig. 5.9).
at------------------------------------------ ß--=======================================0
Mt --j --r -j --j --r -r -j --j ur· j UjU r· r·· U
glass barrel Teflon tipped plunger
screw-on needle
E·
Figure S.9
Both types of micro syringe are available with either fixed or removable
needles, but the type with removable needles are generally preferred.
Removable needles have the advantages that they can be replaced when
damaged or blocked and that different sizes can be fitted. Long needles for
micro syringes normally have to be ordered specially.
All-glass Luer syringes
These syringes range in size from Iml to lOOml, are generally quite cheap
and are the most commonly used type. Some glass syringes have matched
barrel and plunger and they can only be used as a pair, but most modern
syringes have interchangeable barrels and plungers. Syringes with
interchangeable parts will be marked as such and these are the preferred type
to use. They are adequate for most purposes, but should be c1eaned
immediately after use and treated carefully to avoid jamming.
~i---------.J
~ 1IIlilllllillillllllllllllllllllllllllllllllllljrD
gl ass barrel
Figure 5.11
Figure 5.12
The simplest fitting is the Luer glass tip, but these break very easily and
should be avoided. Metal Luer tips are more robust, but the needles tend to
work loose and for this reason Luer-lock syringes are preferred. These
have a slot into which the rim of the needle engages so that it can not come
loose. The most common fitting is the metal tip Luer-lock, but the best seal
of all is provided by the Teflon tip Luer-lock, normally found only on gas-
60 Reagents: Purification and Handling
tight syringes. In this case the Luer tip is Teflon and the locking sleeve is
meta!.
Syringe needles
Luer fitting syringe needles are available in a range of lengths and bore
diameters. The optimum bore diameter depends on the volume of liquid
being dispensed. As a rough guide we use 22 gauge needles for volumes of
less than Iml, 20 gauge for volumes between 1 to 5ml and 18 gauge for
larger quantities. Long needles are usually preferred for transferring liquids
by syringe, but short needles are useful as nitrogen inlets or vents.
Disposable needles with plastic shanks can also be used for this purpose.
/ " Bevel must be clean and sharp
~========================~
long all-metal needle for transferring liquid
"'===~
short disposable needle for introducing and venting inert gas
Figure 5.13
Most needles have a 12· bevel and this should be kept in good, sharp
condition so that minimal damage is caused when penetrating a septum. Flat
ended needles are sometimes useful for removing the last traces of liquid
from a container, but they should only be introduced through septa which
have previously been pierced by a sharp needle.
Cleaning and care 0/ syringes
Syringes and needles can give good service for long periods, but careful
treatment is required to prevent leakage and jamming. It is therefore very
important to clean syringes and needles immediately after use. Organic
liquids or solutions are generally quite easily removed, by simply flushing
with solvent. However, more care is required when a reactive reagent such
as butyllithium or titanium tetrachloride has been used. To c1ear such
reactive reagents the syringe should immediately be flushed several times
with the solvent in which the reagent is dissolved. Whatever the syringe has
been used for it should always be dismantled for final cleaning of the
individual components. If the syringe has been used for an alkyllithium or
any other strongly basic reagent, it should be washed with dilute Hel,
Reagents: Purification and Handling 61
water, then acetone. After using an acidic reagent or a Lewis acid, the
components should be washed with dilute sodium hydroxide, water, then
acetone.
If a syringe has jammed or become contaminated with remnants of a
stubborn reagent, handle it carefully and do not use force to free the
plunger. This will almost certainly lead to the syringe being broken and you
may weIl end up with a cut hand also. One of the most reliable methods for
removing a contaminant from a syringe, or syringe needle, is to submerge it
in a solvent which will dissolve the contaminant and place it in an ultra sonic
cleaning bath for a few minutes. Sonication will often free syringes which
are completely jammed provided you choose the correct solvent. If you do
not know what is blocking the syringe try sequentially sonicating in the
following series of solvents; 10% Hel, then water, then dilute NaOH, then
water, then acetone, then methanol and finally methylene chloride.
How to handle a syringe
The type of reagent that you will need to deli ver by syringe is very often one
that has to be measured precisely in order for the reaction to be successful.
It is also likely to be a very corrosive reagent, such as butyllithium, which
you do not want to splash around the lab or spill onto your skin. So, before
attempting to use a syringe for transferring any reagent, practise using it first
by measuring solvent until you are confident that you can handle it safely.
You will often need to hold a syringe fuH of liquid in one hand whilst
performing another operation with the other hand. One way to do this is as
follows: fill the syringe with the required volume of liquid using both
hands; then place the syringe against the palm of one hand using the 3rd, 4th
and 5th fingers of the same hand to grip the barrel; carefully place the
forefinger around the plunger and against the end of the barrel to hold the
plunger firrnly in pla('e; the liquid can now be delivered by pressing the
plunger down using the thumb, and the other hand is completely free. This
method should be practised using solvent until your technique is good
enough to confidently transfer liquid without spillage.
Preparing a syringefor use
An appropriate clean syringe and needle should be chosen, dried in an oven,
then left to cool in a desiccator. Before using a syringe it should be purged
with argon (or nitrogen) and this is conveniently done by 'syringing out' the
62 Reagents: Purification and Handling
gas via a septum inlet attached to the inert gas system. This should be
repeated several times (Fig. 5.14). After purging with argon (or nitrogen)
the syringe can be kept for a short time before use if the tip of the needle is
inserted into a rubber stopper.
/ septum
~::::::::::::::::::::::::::::::::
Figure 5.14
/ l o n g needle connection to
inert gas system
~inert
~
gas
--
measured-+
volume
_...
----- ------
._-- ..
_---------
...---_ ..... _...
_----
..... - --_ ...
------ --_ ......
---_ _------
_-_ ..
(a)
T (b) (c)
Figure 5.15
+-- inert
:H
·1:
gas
septum
glass
~
--inert
~~ g gas
Figure 5.16
'\.~dry
oil
01 pet. ether
'\.~---- .. _.. _._
.....~
.:.:.:.:.:.:.:.:.:.:.:.:.: ----+'\.
'1
•••::":":" =:
•• '!":':
••• ..
-
:.::-:>i.:
1-:-::.::-:-:. ..rz:.:.:-:-:.:.::-:-:.:.::-:-1.:
.............
........... __ .
.... _--.....
._---_ .. -...._.--- ------------
............
------------ _.. - -.. __---
.... .. _-_......
.. - .. -
...............
..---- ..---
--_. .... --- ... _--_ ........-.
.. ............. ••••••I"'l . . . .
.. ..
balance
i) clean metal ii) wash üi) weigh into a beaker
andcut of oil, then re-wash
Figure 5.17
dispersed in the oil the reagents are moderately stable and ean be weighed
out quiekly in the atmosphere.
For some experiments the dispersion ean be used without separating
out the oil. To do this simply weight it into a pre-dried reaetion flask and
place it under inert atmosphere by sequentially evacuating, then purging the
flask with an inert gas. A 3-way Quickfit tap eonneeted to a double
manifold is ideal for this purpose.
There are various teehniques for removing the oil from a metal
dispersion. The simplest method is as follows:
1. Weigh the dispersion in a flask (do not forget to take the oil into
account) and plaee it under inert atmosphere. Use a double manifold
conneeted by a 3-way Quickfit tap ifpossible (Fig. 5.18a).
2. While maintaining a rapid flow of inert gas through the bubbier of the
inert gas system, open the 3-way tap and add some dry petroleum ether
using a syringe. Swirl the flask to dissolve the oil, then let it stand until
the metal has settled at the bottom (Fig. 5.18b).
inert gas
Reaction solvents and other reagents can then be added directly to this
flask.
If you need to separate the oil from a quantity of metal dispersion
without placing it directly into areaction flask, the piece of apparatus shown
in Fig. 5.19 is very useful. A typical procedure is as follows:
1. After drying the filtration apparatus and cooling under a stream of inert
gas, quickly weigh into it slightly more than the required quantity of the
dispersion, loosely packed.
2. Add some dry petroleum ether to the apparatus and quickly connect a
Quickfit 3-way Teflon tap to the top.
3 . Open the stopcock at the bottom of the funnel and pressurize with inert
gas (Fig. 5.19a). The inert gas can be introduced from a simple line
wh ich incorporates a bubbler (see Fig. 5.2a). The vent of the bubbler
should be restricted to increase the pressure.
Pasteur
pipette
.- inert gas
.-dry
pet.ether
.- meta!
dispersion
.- porosity 3
sinter
.- stopcock
(a) (b)
Figure 5.19
4. Before the level of the solvent reaches the top of the dispersion, turn the
the 3-way tap so that argon is directed out of the vertical inlet as well as
Reagents: Purification and Handling 69
into the funnel. Then add more solvent through the vent using a Pasteur
pipette (Fig. 5.19b).
5. Repeat the washing steps until solvent flows freely, indicating that all
the oil has been washed away, then elose the top and bottom stopcocks.
This procedure will leave you with finely divided metal or hydride,
under argon, and this can be kept for a few hours in the sealed apparatus
without deterioration. The fine powder will be very reactive and therefore
great care is required in handling it. For titration of metal hydrides see
Chapter6.
Weighing reactive powdered solids
The simple procedure given below can be used for most solids, other than
those which are extremely reactive with even the slightest amount of
moisture or air. Examples of reagents which can be weighed by this method
are: sodium hydride, potassium hydride, lithium metal, lithium aluminium
hydride, finely divided metals and metal hydrides (from dispersions). A
glove bag should be used for particularly reactive solids, or in cases where
extreme accuracy is required.
As always, plan ahead when handling reactive materials and make sure
you have a dry vessel, under inert atmosphere, into which you want to
weigh the powder.
1. Remove the receiving vessel from the inert gas system and place on a
balance pan under a stream of argon, provided by the simple set-up
shown in Fig. 5.20. Caution: argon is the prejerred inert gas jor this
procedure, but because it is heavier than air, make sure the jlask
remainsfilled through-out the weighing procedure.
2. Keep the top of the container holding the powdered metal under the
argon stream whilst removing it, and quickly weigh the required
amount into the flask. When weighing any finely powdered
reactive solid avoid sifting it through the air as this can
lead to a fire.
3. Re-connect the flask to the inert gas system (preferably a double
manifold, evacuate very carefully to avoid the power being disturbed,
then refill with inert gas.
4. Carefully neutralize remaining traces of the metal or hydride on the
apparatus using an alcohol.
70 Reagents: Purification and Handling
balance
Figure 5.20
The inverted fllter funnel technique described above reduces contact of
the reactive solid with moist air and thus lowers the chances of deterioration,
and also the risk of fire. In some instances the less reactive powders, such
as sodium hydride, can be safely weighed in the atmosphere if you are quick
and careful. However, it is recommend that an inert atmosphere blanket is
always used for weighing more reactive reagents, such as potassium
hydride.
rubber ~
stopper t
8em
\------~~100ml
flask
8em
i
Figure 5.21
Method 1
Dissolve KOH (1g) in water (2mi), dilute with 95% ethanol (8ml), and add
to the reaction flask. Cool the flask in an ice bath, then slowly add a
solution of Diazald (N-methyl-N-nitroso-p-toluenesuphonamide) in diethyl
ether (15mi), using a Pasteur pipette. When all the Diazald has been added,
stopper the neck of the flask, put the ice bath under the receiver and place a
warm water bath (about 65°C) under the re action flask. The yellow
diazomethane-ether mixture will start to distil out of the flask. Carry-on
distilling until all the yellow colour has disappeared from the reaction flask,
adding a further 5ml of diethyl ether if necessary. The yellow distillate
contains about 3.3mmol of diazomethane.
Method2
To produce 16.6mmol of diazomethane use 5g Diazald, 5g KOH, 8ml
water, 10ml ethanol and 50ml of diethyl ether. The procedure is similar, but
on this larger scale, add the Diazald solution slowly to the warm KOH
solution, while distilling off the diazomethane. A separating funnel (with a
PTFE stopcock) pushed through a rubber stopper can be used for the
addition.
Reagents: Purification and Handling 73
Gases
6 .1 Introduction
Many organic reactions require the use of gases, either inert gases which are
used to protect the reaction, or reagent gases which actually take part in the
reaction. Special experimental techniques are required for handling gases
and this chapter contains a summary of methods for the preparation,
handling and measurement of the more commonly encountered gases.
It must be emphasized at the outset that many gases are
very hazardous, either beeause they are toxie or beeause they
are supplied in eylinders whieh eontain eompressed gas at very
high pressures and, as a result, partieular attention must be
paid to safe praetiee in gas manipulation.
r - fitting for
"regulator spindle
- cylindcr valv. ~
Figure 6.1
Most cylinders are fitted with a unit containing an on-off valve, and an
outlet fitting (Fig. 6.1). This unit should never be tampered with.
The valve fitting is a weak point in the cylinder and might be dislodged or
weakened if the cylinder was dropped. Apart from releasing a potentially
dangerous gas this could allow the highly pressurized gas to vent
uncontrollably converting the cylinder into an extremely dangerous missile.
For this reason cylinders and lecture bottles should never
be allowed to stand unsupported.
They should always be securely c1amped to a bench or a wall. If they
have to be moved frequently they should be supported in a sturdy metal
frame, or in a trolley designed for this purpose. Cylinders should only be
moved in purpose designed trolleys and should always be treated with great
care.
Cylinders are generally pressurized to 175-200atm and the on/off
valves provide no more control than their name suggests so cylinders must
be fitted with a regulator to allow controlled delivery of the gas. The most
popular regulators are the two stage types shown in Fig. 6.2.
The regulators provide a constant outlet pressure which can be adjusted
to suit the particular application. These are often fitted with needle valves to
give control over the outlet flow rate. The procedure for fitting a regulator
to a cylinder is as follows:
1. Find the correct regulator for the gas in question. Different gas
cylinders have different fittings (to prevent the gas from coming into
76 Gases
output cylindcr
pressure
gauge
pressure
nccdlc adjusLing
valvc scrcw
Figure 6.2
contact with incompatible materials) so the appropriate regulator, fitted
with the correct threaded connector, must be used. (Consult supplier's
data for more detailed information). For safety reasons it is best to
choose a regulator which has a fairly low delivery pressure (~ 0-50
psi), unless a high output pressure is specifically required. (1atm =
1.01bar = 14.5psi (lbf/in2) = 1.03 kgf/cm2 = 760mmHg.)
2. Remove the protective cap from the cylinder fitting and ensure that the
fitting is clean and dry.
Never use grease or Teflon tape on cylinder fittings.
3. Screw the fitting onto the cylinder and tighten firmly. A poor fit
probably means that the regulator is not of the correct type. (Note that
some cylinders have 'left hand' threads.)
Never try to force the fitting.
4. Test for leaks by applying a little dilute soapy water, or a commercial
leak detection solution, around the joint.
4. Turn the delivery pressure adjusting screw clockwise until the required
output pressure (typically lOpsi) is registered on the outlet pressure
gauge.
5. When the gas line is correcdy attached to the apparatus (see Section
6.2) open the needle valve slowly to obtain the desired flow rate.
When the gas is no longer required the flow should be shut off and the
cylinder closed by closing the regulator (rotate the pressure adjusting screw
anticlockwise) and then closing the cylinder valve. The cylinder valve
should always be shut when the gas is not in use.
Lecture botdes should also be fitted with a regulator and needle valve,
as specified by the manufacturer. The regulators are of two types, one for
corrosive and one for non-corrosive gases, and they require a Teflon or lead
washer. The procedures for using lecture botdes are the same as for larger
cylinders.
A number of gases are supplied in liquefied form at relatively low
pressures and these generally only require the use of a flow control valve.
The valves on these cylinders are generally fitted with a handwheel. Again
you should consult the supplier's technical data for information on the
correct fittings and handling procedures.
-trap
oil
"- rnercury friucd
disc reaction bubblcr
bubbIer flask
Figure 6.3
coming into contact with the cylinder or the reagents used to prepare the gas.
The gas can be bubbled into the solution via a needle or via a glass tube
fitted with a wide bore glass frit. Use of a frit gives better dispersion of the
gas but care should be taken that the frit does not become blocked by asolid
product.
It is important to pi ace an oil or mercury bubb1er between the gas
supply and the trap so that if the gas line is blocked the gas can be vented
safely. Obviously the depth of oil or mercury in the bubbier must be large
enough to ensure that the gas will pass through the reaction mixture rather
than venting from the bubbier. With the inclusion of a bubbier as a safety
device, it is advisable to put clips on all the ground glass joints, or to wrap
them securely with Teflon tape, so that the gas cannot leak from the
apparatus.
The type of tubing used in the line will be dictated by the nature of the
gas. Rubber or PVC tubing is convenient for non-corrosive, non-toxic
gases such as carbon dioxide but many reagent gases require the use of
Gases 79
~f------I~J-_I'_I'I'_"'!L'I'_I'I'_I'I'I_'I'I_'I'I-'l'1'_1'1'_I'I'_I'I'---JI'I"~0Iir';:"
gas-tight valve With inert
syringe Lue~-lock gas line
fittmgs
Figure 6.4
Just before use, the valve is opened to purge the needle and allow the gas to
reach atmospheric pressure, and then the needle is inserted into the reaction
flask. If a sm all volume of gas is involved (less than 100ml) the reaction
vesse1 can be a sealed system and the gas can be added slowly from the
syringe. For larger volumes the reaction flask should be fitted with a
mercury bubbIer and the gas should be added at such a rate that gas does not
vent from the bubbler.
Gases 81
to fume hood -
gassupply -~
-? 'f=:3!(9;~:r:::==~
t
double -
oblique tap ==
=
levelling
bulb
to reaction
flexible
- - - tubing
Figure 6.5
82 Gases
5. Open the three-way tap to connect the burette to the line with the need1e,
thus allowing the gas to come to atmospheric pressure and giving the
the needle a final flush. Rapidly read the volume in the burette and
insert the need1e into the reaction flask.
6. Raise the levelling bulb slowly to keep a slight positive pressure of the
gas as the reaction proceeds. When the required volume has been
added, close the three-way tap. If the gas is hazardous flush the
apparatus with inert gas (steps 1-4) after use.
If gas burettes are used for a variety of different gases then some safety
points should be noted. The burette should be thoroughly flushed with inert
gas after each use. The tubing may need to be changed for use with
different gases, and the liquid in the reservoir may also need to be changed
depending on the application (aqueous copper sulphate, mercury, mineral
oil, and dibutyl phthalate are often used). The gas burette method is
particularly good for reactions in which the uptake of gas is slow. Gas
burettes are also used for measuring the volume of gas absorbed, or
evolved, in areaction. A useful application is the quantitative analysis of
hydride solutions by adding the hydride to an excess of protic or acidie
solvent, and measuring the volume of hydrogen evolved.3
The procedure for the analysis ofBH3.TIIF is illustrative.
1. Place 50 m1 of a 1:1 mixture of glycerol and water, and a stirring bar, in
a flask and seal it with a rubber septum. Connect to the gas burette
using a need1e as before.
2. Open the burette to the flask and add a few millilitres of hydride
solution to the flask in order to saturate the atmosphere with hydrogen.
3. Open the burette to the atmosphere and raise the levelling bulb to give a
reading of approximately zero. Note this reading.
4. Turn the three-way tap so that the burette is connected only to the flask.
Add an accurately measured volume of hydride solution, with rapid
stirring. The volume used should be sufficient to more than half fill the
gas burette you are using.
5. When hydrogen evolution has ceased, lower the levelling bulb so that
the liquid levels in the burette and the reservoir are equal, and read the
volume. Calculate the molarity using the following equation.
3. H.C. Brown, Organic Synthesis via Boranes, Wiley, New York, 1975.
Gases 83
+
gas - - ==~;::::r::::==~I====::---
mercury
bubbier
- - - - :.-- :::::. - - -
._-.-- - --
- - - - ......
:: .. t
suck-back
cold trap
bath
Figure 6.6
84 Gases
volume or by weight. When the required quantity has been condensed the
cylinder is closed and the reaction flask is attached to the safety trap. The
cooling bath is then removed and the gas allowed to distil into the reaction
vessel. The mercury bubbler and safety trap serve the usual functions.
Figure 6.7
The solution in the pressure-equalizing dropping funnel is added
slowly and cautiously to the reagent in the fiask, with stirring and
cooling if necessary. The gas then passes through a suck-back trap and on
to the reaction fiask, and any that is not absorbed passes through a scrubber.
As usual, there are two bubblers, the first to prevent pressure build-up and
the second to monitor the amount of gas which is not being absorbed. The
rate of addition of the reagents should be adjusted so that relatively little of
the gas escapes through the second bubbier. This bubbier also provides a
gas-tight seal to the atmosphere which protects the reaction fiask from the
reagent (often water) used in the scrubber. If a toxic gas is being generated,
the apparatus should be flushed with an inert gas before dismantling.
4. D.F. Shriver and M.A. Drezdzon, The Manipulation 0/ Air-Sensitive Compounds,
2nd ed., Wiley, New York, 1986.
5. AL Wayda and M.Y. Darensbourg Eds., Experimental Organometallic Chemistry,
A Practicum in Synthesis and Characterization, American Chemical Society,
Washington, 1987.
86 Gases
gas
L
H~h~;;ro==-- to drain
- ....... ..
-....................
............
..
::.....::..::..::::::: :::
......................
(a) (b)
Figure 6.8
Methods for preparing some common gases are described below (see
ref. 3 for other methods). The properties of these and other gases are listed
in Appendix 2. Many of these gases are very harmful and they should only
be handled in an efficient fume cup board, taking care to prevent leaks and to
scrub any excess gas.
Acetylene
Prepare by addition of water to calcium carbide and dry by passing through
a column of molecular sieves. Acetylene can explode spontaneously,
especially when pressurized. For this reason the gas should be generated
very cautiously, the generator should be cooled, and the gas line should be
fitted with a bubbier so that the pressure cannot rise above 15 psi. Dispose
by venting slowly in an efficient hood.
Carbon dioxide
The simplest method is to allow solid carbon dioxide to evaporate, and the
gas can be dried by passage over mo1ecular sieves. It can also be prepared
Gases 87
Vacuum Pumps
7.1 Introduction
There are a variety of tasks in organic chemistry which require provision of
a vacuum source. However, different tasks will require different levels of
vacuum and more than one type of vacuum source will be needed to service
all the requirements of the lab. We can split vacuum supplies loosely into
two categories: low vacuum, being 1-50mmHg and high vacuum being
below ImmHg. A low vacuum source is valuable for such tasks as vacuum
filtration, operation of rotary evaporators and vacuum distillation of
relatively volatile liquids. High vacuum is required for serving inert gas line
manifolds, vacuum distillations and drying solids by removing traces of
solvent or moisture from them.
should also be incorporated, which will give you some time to disconnect
the apparatus if suck-back occurs (Fig. 7.1). A water trap will not prevent
suck-back altogether and hence it is not advisable to leave apparatus
unattended while connected to an aspirator. Clearly, this type of pump is
not suitable for connecting to an inert gas manifold.
to apparatus +-- ~
3-way tap
/
(air inlet)
have several advantages: they are quiet, there are no problems of 'suck-
back' and they do not use vast quantities of water.
For some tasks very cheap and simple fishtank-type diaphragm pumps
are useful. These will produce a vacuum of about 300mmHg and can also
be used to provide compressed air for pressurizing chromatography
columns (see Chapter 9).
+----16cm---+
cone joint
or tubing
18cm
Figure 7.2
mercury
prcssurc
gauge
mercury
(a) (b)
Figure 7.3
8.1 Introduction
This is the most important single chapter in this book. In order to be certain
of the results of a particular experiment, and to ensure reproducibility, it is
important that the appropriate precautions are taken, and that proper
preparations are made. Many common organic reactions involve the use of
air sensitive reagents, which requires the reaction to be carried out under
inert, anhydrous conditions. These conditions are also used for reactions
which are sensitive to the presence of water. Indeed the majority of
reactions, although not requiring such conditions, will usually benefit from
being carried out in the absence of air and water. Accordingly most of this
chapter is concerned with the use of air sensitive reagents, since the same
basic techniques can be used for almost all types of reactions likely to be
encountered.
First of all, before attempting the reaction it is important to ensure that
you have the necessary glassware, apparatus, reagents, and experimental
procedure (including work up and a t1c system which will allow you to
follow the reaction). Do not forget to check on the safety aspects of the
chemicals (including solvents), of the likely reaction products, and of the
procedures which you will be using. As alluded to above, most reactions
will require monitoring of some kind (dc, gc, hp1c, etc.) and it is important
to ensure that your chosen system is suitable for the starting material, and
give some consideration as to where the expected product is likely to be
found (will the polarity increase or decrease on converting starting material
to product?). Make certain that you allow yourself plenty of time to
complete the reaction, or at least to monitor it in the early stages if it is likely
Carrying Out The Reaction 95
8.2.1 lntroduction
Many eommonly used organie reagents are extremely reactive towards water
andlor oxygen. Some examples are: alkyllithium reagents, Grignard
reagents, organoboranes, metal hydrides, organoaluminium eompounds,
euprates, titanium tetraehloride, dry solvents, ete. Some of these reagents
are available eommercially and others are prepared in situ, in either ease they
are most eonveniently handled in solution. Although extreme eare must be
taken to exc1ude air and moisture when using these reagents, you should
find them easy to handle onee you beeome familiar with some simple
teehniques. General proeedures for handling air sensitive reagents are
deseribed in Chapter 5 and the appropriate parts of that ehapter should be
eonsulted in eonjunetion with this seetion.
available. Indeed, you should always have some spare dry syringes in case
one gets jammed, contaminated or broken.
- to vacuum pump
Figure 8.1
98 Carrying Out The Reaction
A fast flow is used when filling a system, but once the reaction is set up,
one bubble every 5-10s is adequate. A piece of apparatus which is very
useful when used in conjunction with a double manifold is a Quickfit 3-way
tap, which is also described in more detail in Chapter 3.
J: 1:
and heat the glassware with a heat gun (or bunsen) for 5min. Then,
~~riti~t
position A position B position C
with inert gas flowing through the bubbier at a rapid rate, turn the
manifold tap slowly to introduce the inert gas. Keep an eye on the
bubbier while you are switching to the inert gas supply and operate the
tap carefully to avoid 'suck-back'. Finally leave the apparatus to cool.
2 . Add initial reactants
Solid reactants (or heavy gum oils) can be weighed into the reaction
flask at this stage. With the inert gas flow at a reasonably high level,
remove the flask from the system and quickly weigh the reagent into it
(for reactive solids see Chapter 5). Then re-attach the flask to the 3-
way tap and switch the manifold tap to evacuate. If there are traces of
solvent on the reactant these can be removed by leaving the fIask under
vacuum for an extended period, otherwise the 2-way tap can be
switched back to argon straight away leaving the reactant in the flask
under inert atmosphere.
3. Add solvent
The best way to add solvent to the system, and avoid breaking the inert
atmosphere, is to use a syringe. It is best to syringe the solvent straight
from the collecting head of a still, which is itself under inert atmosphere
(see Chapter 4), but the solvent might also be taken from a bottle or
fIask under inert atmosphere (see Chapter 5). In either case, once you
have a syringe loaded with solvent, open the 3-way tap to position B
(Fig. 8.2b), keeping an inert gas flow which is rapid enough to
maintain bubbling, and add the solvent. Then turn 3-way tap back to
position A (Fig. 8.2a) and reduce gas pressure to maintain steady
bubbling. For large scale reactions solvents can be added by
disconnecting the fIask, adding the solvent quickly and re-assembling,
but this inevitably leads to air entering the system. With higher boiling
solvents the air can be removed by sequentially partially evacuating and
refilling with inert gas several times, using the 2-way manifold tap.
4 . Cool the jlask
At this stage the reaction fIask can be immersed in a cooling bath and
stirred with a magnetic stirrer (see Chapter 8). Never cool a flask
be/are it is under the inert atmosphere as this causes moisture to
condense.
100 Carrying Out The Reaction
Figure 8.3
5. Add reagents
It is most commonly the reagents used in the reaction which are air
and/or moisture sensitive and they should be added with great care (see
Chapter 5 for more detail). After pressurizing the reagent container
with inert gas, syringe out the required quantity (Fig. 8.3a). Increase
the inert gas flow through the manifold so that rapid bubbling is
maintained as the 3-way tap is opened to position B, then insert the
syringe through the tap to deliver the reagent (Fig. 8.3b). When all the
re agent has been added turn the 3-way tap back to position A and
reduce the gas flow so that there is a bubble every few seconds (Fig.
8.3c). Any further reagents are added in the same way and the reaction
is left to proceed. To add a cooled reagent it is advisable to use a
cannula made out of Teflon tubing (see Chapter 5, Section 5.4.3 for
more detail).
6. Reaction monitoring
The reaction can easily be monitored by de, ge, ete. Again inerease the
inert gas flow through the manifold so that rapid bubbling is maintained
as the 3-way tap is opened to position B, then insert a drawn piece of
glass tubing or Pasteur pipette through the tap to remove a drop of the
mixture.
Carrying Out The Reaction 101
7. Work-up
Once the reaction is complete it can be worked up as normal (see
Chapter 9). However, when there is the likelihood of residual reactive
reagents such as organometallics, be very careful when adding aqueous
solutions 1t is normally safer to keep the mixture under an inert
atmosphere whilst the solution is added.
..~~o gas
inert
Figure 8.4
102 Carrying Out The Reaction
condensers (Fig. 8.4) for reactions under inert conditions. Since the
cooling surfaces are inside the apparatus, the problem of outside
condensation is alleviated.
Typical set-ups for reactions which are to be heated are shown in Fig.
8.4. The procedure for carrying out the reaction is exactly as described in
the previous section, except that the reaction is heated instead of being
cooled. Reagents can be added through the 3-way tap, provided a syringe
with a long needle is used, or as an alternative a flask with a side arm fitted
with a septum can be used. Whilst the temperature of areaction is changing
the bubbIer of the system should be checked to make sure there is a constant
inert gas pressure.
2. Slow addition ofreagents and large scale work
When slow addition of an air-sensitive reagent is required it is best to
incorporate apressure equalizing addition funnel into the apparatus, and for
coiled condenser -
Figure 8.5
Carrying Out The Reaction 103
large scale work this is always best. The apparatus can be set up with or
without a condenser, can be stirred either magnetically or mechanically. A
typical arrangement is shown in Fig. 8.5. Where very careful slow addition
is required, particular on a small scale, the use of a syringe pump is strongly
advised. These are mechanical devices which drive the plunger of a syringe
very slowly into the barrel.
mechanical
stirrer ~
inert
gas-
to bubbier
via manifold ____r--.,
• (inert gas
turned off)
- ~
(b) Filling droppingfunnel using cannula
Figure 8.6
To dry this larger, more elaborate set-up, it is best to oven-dry the
equipment first, assemble it hot and leave to cool under a stream of inert
gas. This can be done by connecting to inert gas from the manifold, with a
short needle through the septum acting as a vent (Fig. 8.6a). Once the
apparatus is cool reactants can be added as described above. A syringe can
be used to add small quantities of re agent to the addition funnel. For larger
quantities (>50mI) the cannulation technique described in Chapter 5 for bulk
transfer of liquids is best, and if the funnel is graduated the reagent can be
104 Carrying Out The Reaction
added straight from the re agent bottle (Fig. 8.6b). If a mechanical stirrer is
used a Teflon sealed guide should be used for the stirrer rod. The reaction
is carried out exact1y as before.
reaction flask -
Figure 8.7
When asolid re agent is to be added to areaction set-up which is already
under inert conditions, it is best to add it in solution. To do this, weigh the
solid into a dry flask, fit the flask with a 3-way tap, connect to aseparate
outlet on the manifold, and evacuate. Then switch the manifold tap to
introduce inert gas, and with the gas flow high enough to maintain rapid
bubbling at the bubbier, open the 3-way tap to position Band syringe in the
required dry solvent. With the 3-way tap turned back to position A, the
solution is under inert atmosphere and can be kept indefinitely (Fig. 8.7).
To transfer the reagent, make sure that the inert gas flow is high enough to
maintain rapid bubbling whilst the 3-way tap on the reagent vessel is opened
to position B. Then syringe out the solution, turn the tap back to position
A, open the tap on the reaction flask to position Band add the reagent.
Alternatively, a cannula can be used to transfer the reagent (see below).
A similar procedure can be used when areagent has to be prepared
under inert conditions for use in the reaction. This is quite a common
occurrence, for example when a Grignard or alkyllithium reagent has to be
prepared, then added to areaction mixture. In this case the reaction flask
assembly is set up attached to one outlet of the manifold, while the reagent is
Carrying Out The Reaction 105
inert
gas
(a) (b)
Figure 8.8
the manifold. Then pressurize the flask with a separate inert gas supply,
turn the 3-way tap to position B, insert a cannula, and transfer the solution
either to the re action flask directly (Fig. 8.8a), or to an addition funnel (Fig.
8.8b).
For more information about cannulation and adjustment of inert gas
pressures see Chapter 5, Seetion 5.4.2.
Figure 8.9
required to add the reagent, and you should make sure that it is not added at
a rate which might lead to the reaction going out of control. Alternatively,
an inverted funnel (see Chapter 5) can be used to blanket the apparatus with
an inert atmosphere to protect the reaction mixture.
There are alternative arrangements for adding sölids, but none are
universal. One method is to attach a bent tube containing the solid to the
side arm of the fIask, and twist the tube to allow the compound to pour into
the reaction fIask (Fig. 8.9).
_ _- - tubing - - - -
~!!!EJ
Figure 8.10
Carrying Out The Reaction 107
with the inert gas from a cylinder, then attach it to the reaction system using
either a needle/septum or 3-way tap, taking care to ensure that there are no
leaks in the system.
In the case of the needle/septum attachment, any air contained in the
reaction tlask is then tlushed out by using another need1e to vent the system
(Fig. 8.11). After the reaction tlask has been thoroughly tlushed with the
inert gas, the extra needle is removed and the reaction tlask is ready for use.
The balloon keeps the whole system under a positive press ure of the inert
gas, and also allows liquid materials to be added to, or removed from, the
tlask via syringe insertion through the septum.
balloon filled
with inert gas---+-
Figure 8.11
Generally argon is preferred when using this technique because it is
more dense than air, and will fill the reaction tlask, pushing out any air more
effectively than would nitrogen. This technique can also be employed when
syringing air-sensitive materials from bottles or containers.
With the three-way tap attachment, a vacuum line can be connected
(Fig. 8.10). The system can then be purged by sequentially evacuating,
then filling with the inert gas from the balloon. This is a more effective
procedure than simply tlushing out the tlask and so both nitrogen and argon
can be used with this set-up. Liquid materials can then be added to the tlask
by syringe via the septum inlet as before.
Since balloons are perishable it is often advisable to use a 'double
balloon system'. This is simply two balloons, one inside the other,
allowing the inert atmosphere to be maintained even if one bursts.
108 Carrying Out The Reaction
c==5~ . . c> . . ~
1!
rubber tubing 3
t ta
~'''''''"' ..._ -woy p
+needle
Figure 8.12
place using a piece of wire, elastic band, or parafilm. The open end of the
tubing can then be attached to a three-way tap, or to a needle-tubing adapter
and needle as shown. It is advisable to seal all joins with parafilm in order
to ensure that there are no leaks.
Setting up areaction
Once the balloon system has been constructed, the reaction system is simply
set up as outlined in Fig. 8.10.
If the reaction is to be heated, a condenser will need to be fitted, in this
case it may be necessary to use a two necked flask so that materials can be
added to the flask. In such cases the balloon should be placed at the top of
the condenser in order to prevent volatile liquids reaching it.
All joints should be sealed with Parafilm or Teflon tape, the latter being
especially useful if the system is to be heated (Parafilm melts!).
another. This is often the case if you need to optimize the conditions for a
reaction.
+- inert gas
'-stopcock
(optional)
tubing
. - syringe needle
(Luer fitting rernoved)
Figure 8.13
The reaction is set up in exactly the same way as described in the
previous section for use of balloons. With either type of set-up a tlc sampie
can easily be removed by pushing a syringe needle through the septum then
inserting the tlc capillary through it. When using the 'spaghetti' manifold it
is often convenient to add reagents, or transfer them, by cannulation using a
Teflon cannula (Fig. 8.14, see also Chapter 5, Seetion 5.4.2).
syringe needle.-.
(vent)
Figure 8.14
110 Carrying Out The Reaction
cases arecord of the tlc should be made in your lab book (see Chapter
2).
2. TIc can be used to indicate the identity of a compound, by comparing
the unknown sampie with a known material. In general each substance
is spotted separately and also together (co-spot). Caution should be
applied as co-running on tlc is not definitive proof of identity. Of
course, substances that do not co-run are definitely not the same.
3. TIc usually gives a good indication of the purity of a substance.
Diastereoisomers can usually (but not always), be distinguished.
4. For flash chromatography (see Chapter 9), tlc is first used to determine
the solvent system and quantity of silica required, and secondly to
monitor the column fractions.
Tlcplates
There are two main types of coating for tlc plates, silica and alumina. Silica
plates are most commonly used, they are slightly acidie and are suitable for
running a broad range of compounds. Most of the information in this
section refers directly to silica plates, but the same principles apply when
using alumina plates. Alumina plates are slightly basic and are quite
commonly used when a basic compound will not run very well on silica.
The most common tlc plates have either a glass or plastic backing
coated with a thin layer of silica, which contains a binding agent to keep it
bonded to the backing. Although tlc plates can be horne-made, most people
prefer to use commercial ones as they give very consistent results. For
analytical purposes plates with a O.25mm layer of silica are normally used.
They are available in a variety of sizes, although 5 x 20cm is probably the
most convenient size. The chromatographs are run along the 5cm length,
cut to an appropriate width. This normally gives adequate resolution, with a
very short running time. Plastic backed plates are sometimes cheaper and
can easily be cut into strips with scissors, but glass plates see m to give
better resolution and can be heated more vigorously for visualization
purposes.
To cut a glass tlc plate, place it face down on a clean piece of paper,
hold a ruler firmly along the proposed cut and draw a sharp diamond glass
cutter along the line once only. Then holding the plate with the forefinger
and thumb of each hand, on either side of the score line snap the plate along
112 Carrying Out The Reaction
the line. With practice, and a good glass cutter, you should be able to cut
plates down to about 1.5cm without ever spoiling them.
The proeedure for running a tle
1. Cut a tlc plate which is 5cm long, and wide enough so that about O.5cm
can be left between each spot (obviously the width of the plate is
dependant upon the number of spots to be run on it).
2. Make a tlc spotter by drawing out a Pasteur pipette or melting point tube
to about O.5mm using a micro burner. The spotter can be used many
times provided you wash it with clean solvent in between runs.
3. If the substance to be analysed is not already in solution, take it up in a
volatile solvent as a ca. 1-2% solution. A non-polar solvent is
preferable but methylene chloride is often used as a universal solvent
for tlc sampies. Reaction mixture solutions can be diluted down if
necessary, and with experience making sampies to a reasonable
strength becomes intuitive.
cover
I
,
tlc plate
tlc tank-
l
solvent front
tlc plate "
tlc spotter -----.... !
l ,,
j - filter paper
!
t
solvent
starting reaction
material mixture Ch) Cc)
co-spot
(a)
Figure 8.15
4. U sing the spotter, spot a small amount of each solution about O.5cm
from the bottom of the plate, leaving a similar distance between each
spot. The spots should be kept as small as possible, and you should
take care to make sure that each of the spots is the same distance from
the bottom of the plate (Fig. 8.15a).
Carrying Out The Reaction 113
plates are used the spots can sometimes be seen more c1earIy from the
glass side of the plate.
Stains U selcomments
Vanillin Good general reagent, gives a range of colours.
PMA Good general reagent, gives bIue-green spots.
Anisaldehyde Good general reagent, gives a range of colours.
Ceric suiphate Fairly general, gives a range of colours.
DNJ># Mainly for aldehydes and ketones, gives orange
spots.
Permanganate# MainIy for unsaturated compounds and
alcohoIs, gives yellow spots.
the polarity of the solvent used for elution is increased the spots will move
further up the plate and the distance between the centres of the spots usually
increases, up to about half way up the plate. However, the spots also
become more defuse, the further they travel, so an Rf value of about 0.4 is
normally the optimum for analytical purposes.
The best tlc solvent system for a particular compound or mixture can
only be determined by trial and error. However it is good practice to stick to
a 'standard' solvent mixture, which can be used most of the time and you
are familiar with. The most widely used solvent mixtures are based on a
non-polar hydrocarbon, such as 40/60 petroleum ether or hexane, with a
polar constituent added in a proportion which gives a suitable polarity.
Probably the most popular 'universal' tlc system is petroleum ether - ethyl
acetate, the polarity of which is easily adjusted by changing the proportions
of the two solvents. If the compounds being analysed will not travel in
ethyl acetate mixtures, a more polar solvent such as ethanol is used as the
additive. On the other hand, if the compounds travel too far a less polar
additive such as petroleum ether is used.
The degree of separation between compounds will also vary according
to the solvent used, so if compounds do not separate or give poor separation
in one system, different systems should be tried. Where there are a number
a compounds in a mixture it may be best to use two or more different
systems, for resolving the polar and non-polar components.
Common tlc solvents fall ioto one of three categories based on polarity,
with smaller variations within each category.
Very polar solvent additives: methanol> ethanol> isopropanol
plate. This is done by cutting a square plate (ca. 5cm x 5cm) and spotting
the compound in the bottom left-hand corner (ca. O.5cm from the bottom) as
shown in Fig. 8.16a. The plate is then eluted as nonnal to give the spots in
a line up the left-hand side of the plate. The plate is then removed from the
tlc tank and the solvent allowed to evaporate. It is then placed back in the
tank with the line of spots along the bottom, and re-eluted (Fig. 8.l6b).
decomposition
~.
elute
(a) (b) (c)
Figure 8.16
The result, if no decomposition is occurring will be the tlc running
diagonally across the square plate, however if decomposition is occurring,
the spot due to the unstable compound will show decomposition products
off-diagonal (Fig. 8.16c).
It is very useful to carry out this test before any fonn of preparative
chromatography, if you suspect that a compound may be labile on silica.
injeclion valve
(sampIe being
loaded onlO loop)
pulse darnpcr
j / r······················_·_.. _··l
pump
I
= !~ '
-OOI"m~--~
filler c:::::J
a 0 a a
a a
·····················1 injcelion valve
(sam pIe ßowing
a 0
onlO column)
/
deleclOr sampIe out
i
chart recorder
Figure 8.17
and also acts as apressure regulator. In between the pulse damper and the
column there is an injection valve which allows the sampie to be introduced
into the solvent stream. In the 'load' mode the solvent by-passes a sampie
loop, into which the sampie is injected from a syringe. On switching to
'inject', the solvent stream is diverted through the load loop, introducing a
very accurately measured volume of the sample solution onto the column.
Components are separated on the hplc column in exactly the same way as
they would be on a tlc plate, the less polar compounds running faster and
coming through first. The effluent from the column passes into a detector
(usually an ultraviolet or refractive index) which produces a signal on the
chart recorder when a component is present.
The time at which the compound comes off the column is characteristic
of that particular material, and is referred to as the retention time.
The area under any peak on the chart recorder is proportional to the
quantity of that component and the method is therefore quantitative.
Uses 0/ analytical hplc
Finding a tlc system and running a sampie can be done very quickly and for
this reason tlc is the normal method of choice for routine reaction
monitoring. However, there are occasions when it is worth spending the
time to set up an hplc system for reaction monitoring, especially if, as in
many modern synthetic labs, you have a system close to hand. One reason
to use hplc is that the compounds in which you are interested do not separate
very weil on dc. The other common reason is that you require a quantitative
Carrying Out The Reaction 119
technique. This may be the case if you are trying to optimize areaction to
maximize the quantity of one produet over another, and for this type of
extended study it is well worth the time it takes to set up the system. For
most synthetie purposes it is the relative, rather than the absolute
proportions of substances which are important, and if that is the case a
simple comparison of integrated peak areas may be all that is required. If
aeeurate quantification is needed, then a ealibration is required and this ean
be done using an intern al standard (see below).
Another common use of hplc is for identification of a compound by
comparison with a known substance. Under a specific set of conditions
(solvent, flow rate and quantity applied) any compound will have a specific
retention time and this can therefore be used as a characteristic of the
compound. However, just as a mixed spot should be always be run when
comparing substanees on tle, so with hplc a single enhanced peak should be
observed when the comparison substance and the unknown are injected as a
mixture. Again caution should be used, sinee a single peak is not absolute
proof that compounds are the same.
Preparative hplc is now beeoming widely used in organie ehemistry for
separating eompounds with very similar polarity (see Chapter 9 for more
details). Before committing all your material to a preparative column it is
always best to run a small quantity of the sample on an analytical eolumn, in
order to work out the best eonditions. Indeed, eolumns are produeed in
various sizes which are directly comparable with one another.
If you are monitoring a reaetion by hplc and you want to know the
identity of one or more of the products, you can often separate a few
milligrams from a few runs on the analytical column, wh ich is enough to get
a fuH range of speetral data. On simple hple systems this ean be done
manually by collecting the effluent from the column when the peak of
interest is coming off, and repeating several times. On more sophisticated
systems a fraction collector is often incorporated and in so me cases
injections can be made automatically, so that the system can be set up to
collect a particular peak or peaks over a large number of runs.
Several methods are now available for coupling an hplc system to a
mass spectrometer, so that a mass spectrum is produced for each peak thus
providing some structural information.
120 Carrying Out The Reaction
Quantitative analysis
For any type of detector each compound will have a different response, but
for a particular compound, the area under its peak is directly proportional to
the mass of material which produced the peak. If you are using a uv
detector to analyse two compounds which have the same chromophore and
extinction coefficients, then you can compare peak areas direct1y to
determine the proportions of each compound.
Since the peak area is proportional to the mass of a compound, it is
possible to take a known mass of a compound, make a standard solution
and inject specific quantities, to work out the proportionality constant.
However, a more accurate method of calibration is to use an internal
standard. To do this follow the procedure below:
1 . Choose as a standard a readily-available stable compound, with a
retention time away from the peaks of interest.
2. Make up at least three mixtures containing known quantities of the
standard and each of the compounds which are to be analysed.
3. Run the mixtures and measure the areas of each peak -
The mass of material under any peak, y, is:
My = ky x Areay
now, comparing the area under the standard peak with that under the
unknown:
My/M(std) = ky/k:(std) x Areay/Area(std)
Using this equation we can work out ky!k(std) which is a constant ky ,
known as the correction factor for compound y. Using data from each
of the runs, the average correction factor for each compound is
calculated.
4. Now if we want to calculate the quantity of a compound in a mixture,
the mixture is spiked with a known quantity of the standard and the
following equation is used:
My = ky/M(std) x Areay/Area(std)
Peakshape
For good quantitative results from analytical hplc (or gc, see below) you
should aim to produce chromatographs with symmetrical peaks. Tailing of
the peaks is usually caused by overloading and can thus be avoided by
Carrying Out The Reaction 121
reducing the quantity of sample applied. If this does not solve the problem
and the tail of a component is long and drawn out, there may be an
incompatibility between the compound and the stationary phase, a problem
which is less easy to rectify.
Hplc can also be used for preparative separations and this is described
in Chapter 9.
~ flame ionization
injection port dctcctor
amplifier
_ ovcn
Figure 8.18
122 Carrying Out The Reaction
of the oven can be accurately controlled and can either be kept constant or
increased at a specified rate. Separation of the components in gc is not
based on the principle of adsorption, as it is in liquid chromatography, but
on partition. A gc column is rather like an extremely effective distillation
colurnn with the relative volatility of the components being the main factor
which determines how quickly they travel through the column. The
stationary phase of the colurnn is a very high molecular weight, non-volatile
oil, which has a very large surface area (see below for more detail about
columns) and the gaseous components of the mixture are partitioned
between the oil and the carrier gas at different rates. Thus the components
are separated along the length of the column and emerge as discrete bands.
The gas stream passing out of the column enters a flame ionization detector
(FID) which produces an electric current when a compound is burned in the
its flame. The electric current is amplified to produce a peak on the chart
recorder. These detectors are very sensitive and the response produced is
proportional to the quantity of material being burned, thus the peak area is
proportional to the quantity of sample. As with hplc, the time taken for a
particular substance to reach the detector is characteristic of that substance,
and referred to as the retention time.
Types 0/ gc column
A gc colurnn must contain a liquid stationary phase with a large surface area,
which must be supported, so that it stays in the colurnn and the gas can pass
through it. Packed columns are the traditional type, they are made from
metal or glass coils, and have an internal diameter of about 2 to 4mm. The
stationary liquid phase is coated on particles of solid with which the column
is packed. There are a wide variety of stationary phases available ranging
from Apiezon greases, which are very non-polar to polyethylene glycols
which are very polar. With packed columns the choice of stationary phase
is often critical for good separation and a good deal of experimentation is
usually required before the best material is found.
The development of modern capillary columns has led to improved
resolution and has also simplified the process of running gcs considerably.
The columns are normally made from fused silica capillary with an inside
diameter of between 0.2 and 0.5mm, and are polymer coated. They have no
packing, but instead the liquid stationary phase is bonded to the inside wall
of the capillary, and this allows gas to flow very easily. Because of this the
Carrying Out The Reaction 123
columns can be made much longer than packed columns (between 12 and
100m) and they are typically ten times as efficient. Capillary columns give
extremely high sensitivity and only a very small quantity of material is
required. For this reason the injector normally incorporates a 'splitter', so
that only a small portion of the sampIe injected actually enters the column.
The development of capillary columns has been largely responsible for
the increased use of gc for monitoring organic reactions and for product
analysis. The increased sensitivity of these columns is one important reason
for this, but the fact that they are very simple to set up and operate is
perhaps more significant. The type of stationary phase is not so critical as
with packed columns and the gas flow rate is essentially determined by the
column, so the instrument can be operated successfully with minimal prior
expertise. For most purposes relating to preparative organic chemistry it is
sufficient to rely on just two types of column, one non-polar (such as a
BP1) and one polar column (such as a BP20).
Uses of gc for reaction monitoring and product analysis
Capillary gc instruments are so simple to use that, provided there is one
c10se by, monitoring areaction by gc is almost as quick as running a dc. It
is common to turn to gc monitoring when dc does not provide resolution
between starting material and product or between one product and another.
Gc will usually separate components which co-run on dc. We also find that
some compounds, such as amines, which do not run very weIl on tlc, can
be analysed very easily by capillary gc.
Gc also provides quantitative analysis and is widely used for
determination of product ratios from diastereoselective reactions, down to
about 200:1. This makes it an ideal technique for optimization studies,
where a large number of small-scale reactions are carried out under different
conditions and product ratios are measured simply by syringing out a few
microlitres from each and then injecting into the gc instrument earlier.
For absolute quantitative studies the gc instrument can be calibrated in
exacdy the same way as described for the hplc instrument. Some people use
this technique to measure theoretical yields for reactions, although it is
always preferable that isolated yields are quoted.
The identity of a compound can often be determined by gc, if an
unknown has the same retention time, and co-runs with a known compound
when the two are injected as a mixture, but just as with tlc and hplc co-
124 Carrying Out The Reaction
EtO~O
~ ~l~?+
.....-y~o
E t 0 ' ( ' A :0
° l..
...... "'>
+ I
0
+
0
A B c
Figure 8.19
Having diseovered the identity of the re action produets by ge-ms,
produet ratios for aseries of reaetion were obtained using simple ge and we
were thus able to find optimum eonditions for generation of eompound A.
In this seetion we have not endeavoured to give a eomprehensive
review of methods which ean be used for reaetion monitoring, but we have
deseribed the most universal and eommonly used modern teehniques.
Various other monitoring methods ean be devised for specifie reaetions.
For instanee ultraviolet (uv) speetroscopy ean be used if one strong
chromophore is being eonverted to another, but the disadvantage of
speetroseopic methods is that they do not indieate how many produets are
being produced. Nmr speetroseopy ean also be employed and this is
mentioned further in Chapter 10.
Carrying Out The Reaction 125
]ow-tempcrature septum
thermometer
,/
inert
ga
coolant
lagged
bath ________ ~IM----r--:---t'I
~
Figure 8.20
Reactions are usually carried out below room temperature by placing the
reaction vessel in a cooling bath. In general this is done by placing a
cooling rnixture into a lagged bath, and then immersing the reaction vessel in
the cooling mixture to a depth that ensures the reaction contents are below
the level of coolant (Fig. 820). The temperature of the coolant can be
monitored by means of a low temperature thermometer' immersed in the
bath. It should be noted when carrying out reactions using cooling baths,
that the reaction itself may not be at the same temperature as the bath due to
exothermic processes taking place, so where possible the internal reaction
temperature should also be monitored. A particularly convenient way of
doing this is to use a digital low-temperature thermometer. These are
126 Canying Out The Reaction
hypodcrmic
probe -
incTl
gas
reacLion mixturc
Figure 8.21
commercially available, and most come with a hypodermic probe which can
be inserted into the reaction flask through a septum (Fig. 8.21). The
temperature can then be noted and the probe withdrawn. This is often a
much easier procedure than the more usual method of setting up the reaction
apparatus to inc1ude an internal thermometer, and is particularly useful for
small-scale set-ups where an internal thermometer cannot be used.
In general all low temperature reactions should be done under inert
atmosphere (nitrogen or argon) to avoid atmospheric moisture from being
condensed into the reaction mixture.
The three main types of cooling mixtures are described below:
i) lee-salt baths. Various salts or solvents can be mixed with crushed ice
to produce sub-zero temperatures. In practice, temperatures ranging
from o·e to -40·e can be obtained (see Table 8.1); however at the
lower temperatures the cooling mixture consists of granular ice-salt
partic1es with little or no liquid, and so this can result in poor thermal
contact with any vessel immersed in it. For the lower temperatures,
where careful temperature control is important, it is preferable to use a
liquid or slush coolant which has good thermal contact with any vessel
immersed in it.
Carrying Out Tbe Reaction 127
Water 1:1 o
NaCI 3:1 -8
Acetone 1:1 -10
CaCI 2·6Hp 4:5 -40
ii) Dry ice-solvent baths. Solid carbon dioxide (dry ice) is commercially
available as pellets or blocks, and forms very good cooling mixtures
when combined with a variety of organic solvents (see Table 8.2). In
practice the baths are prepared by adding the dry ice pellets carefully to
a bath containing the requisite solvent until the temperature required is
reached. The temperatures quoted in Table 8.2 refer to baths in which
an excess of dry ice is contained in the solvent. In this case cooling
mixtures ranging from -l5"C to -78'C can be achieved.
iii) Liquid nitrogen slush baths. Slush baths are made by adding liquid
nitrogen carefully to a solvent contained in the bath, with continuous
stirring (gi ass rod, not athermometer). The coolant should become the
consistency of ice-cream, and stirring should prevent any solidification.
Again a variety of different liquids can be used to give temperatures
ranging from 13'C to -196'C (see Table 8.3). Such cooling systems
can often be left for several hours if the cooling bath is well lagged,
however for longer periods (overnight) some form of mechanical
cooling is usually necessary. In such instances, the reaction vessel can
1. AJ. Gordon and R.A. Ford, "The Chemist's Companion", J. Wiley and Sons, New
York,1972.
2. A.M. Phillips and D.N. Hume, 1. ehern. Ed., 1968, 54, 664.
128 Carrying Out The Reaction
scalcd
lhick-walled lUbe
- -
gl ass tube
~
rcaelion
mixlure
Figure 8.22
requires a high degree of skill, since heating leads to apressure build-up
inside the tube which can result in explosion if there are any flaws in the
seal.
Tcnon _ __
scrcw-lOp
oUllct 10 allow
cvacuation
O-ring seal -
thiek walled
glass lube
reaction
mixlure
Figure 8.23
An alternative to the all-glass sealed tube is to use areaction tube,
which consists of a thick-walled glass tube with a teflon screw seal at the
top (Fig. 8.23). This serves as a re-usable sealed tube apparatus, and
commercial versions of this apparatus are available. 1t also features a useful
side-arm that allows evacuation or purging with an inert gas prior to sealing
the tube.
130 Carrying Out The Reaction
inert
~!:>F'~gäS
- waterout
condenser-
waterin-
heating
bath
hotplate--
.-Ä
~
Figure 8.24
reaction vessel protected with a condenser (Fig. 8.24). The condenser is
used to prevent the evaporation of volatile materials (usually the solvent)
from the reaction mixture.
There are many different designs of condenser available, and the type
used depends upon the nature of the re action involved. The most common
designs of condenser are the Liebig condenser (Fig. 8.25a), the coiled
condenser (Fig. 8.25b), the double-jacketed coiled condenser (Fig. 8.25c),
and the cold-finger condenser (Fig. 8.25d). Other condensers available tend
to be simple modifications of these three types.
The Liebig condenser, the coiled condenser, and the double-jacketed
coiled condenser are similar in design and function. They are water-cooled
Carrying Out The Reaction 131
via connection to a cold-water tap, in the case of the Liebig condenser the
water flows in at the bottom and flows out at the top giving a jacket of cold
water around the condenser stern and leading to a cold surface on the inside.
Any volatile materials in the reaction condense on the cold outer surface and
Figure 8.25
run back into the reaction mixture. The coiled condenser functions in a
similar way only now the cold surface is on the inside of the condenser.
This can offer an advantage in particularly humid locations because there is
less tendency for atmospheric moisture to condense on the outside of the
condenser and run down over the reaction vessel. The double-jacketed
coiled condenser is also water-cooled, again water flows in at the bottom
and out at the top. This condenser design tends to be more efficient than the
other two because it provides a greater area of cold surface. Consequently it
is preferred when low boiling materials (:S; 40°C) are involved.
The cold-finger condenser is rather different from the above three. It is
cooled by either solid carbon dioxide / acetone (-78°C) or liquid nitrogen
(-196°C). The coolant is placed in the top of the condenser and more
coolant is added as required. This results in an extremely cold surface on
the inside of the condenser. Condensers of this type are usually employed
for reactions that involve solvents or components that boil at or below room
temperature (e.g. liquid ammonia, b.p. -33°C), although they can be used
for higher boiling materials as weIl.
132 Carrying Out The Reaction
3. Heating devices
As with low-temperature reactions, a common method of increasing the
temperature of areaction is to place the reaction flask in a bath (Fig. 8.23).
In this case the contents of the bath are then heated, usually using a hotplate.
inert
- waterout
condenser -
reaction
heating mantle
~ .-1===::::L-
~
~
Figure 8.26
There are four commonly employed heating baths; water, silicone oil,
Woods metal, flaked graphite, and sand. Water baths consist of a pyrex
glass container filled with water, and are used for reactions requiring
temperatures up to lOO·C. Silicone oll baths are similar in design but can be
used for temperatures up to about 180·C, although the maximum
temperature in this case depends upon the precise type of oil employed.
Woods metal is a commercially available alloy (50% Bi, 25% Pb, 12.5%
Sn, 12.5% Cd) that melts at 70·C. It has excellent thermal properties and so
is a safe material for use at quite high temperatures (up to 300°C). It is
normally used in a steel container, and again heated via a hot-plate device.
Carrying Out The Reaction 133
condenser-
Dean-Stark
trap -
- heating bath
.. --- --. - -
Figure 8.27 A Dean and Stark trap
A Dean and Stark trap can also be used to remove volatile alcohols,
such as methanol and ethanol, which are miscible with many organic
solvents. They can nevertheless be removed by placing sA molecular sieves
(Section 4.3) in the trap, in order to absorb the alcohol. An alternative is
use a Soxhlet extractor containing molecular sieves.
On a small scale, simply placing some activated sieves in the reaction
flask is a convenient means of removing water. This method is effective in
driving equilibria and is also used to promote reactions which are adversely
affected by water.
8.6 Agitation
For homogeneous reaction systems at constant temperature, agitation is not
normally necessary. However, most reactions involve heterogeneous
reaction mixtures and so require some form of agitation to ensure efficient
mixing of the reactants. The most commonly used methods of agitation are
outlined in this section.
speed contral
l )
,I
/ )
, ,
i
~
~
{} {}
"\ I." '
\/
speed + temperature
controls
Figure 8.28
The folIower consists of a magnet coated with an inert polymer, usually
Teflon or PVC, and come in a variety of shapes and sizes (Fig. 8.29) It is
important that the polymer coating does not react with any components in
the reaction mixture, and because of this Teflon is usually the preferred
coating. The size of the folIower is also important, it should be large
136 Carrying Out The Reaction
enough to stir the reaction mixture effectively but not so large that it will not
sit flat in the bottom of the reaction flask.
c___) ( ~ )
(a) (b) (c)
Figure 8.29 Magnetic followers: (a) bar; (b) octagonal; (c) egg-shaped
Because the follower is driven by a magnetic field and has no
mechanical connection with the stirrer machine, this method of agitation
does not require special modification of areaction apparatus.
Magnetic stirrer machines are probably the most commonly employed
method of agitation for organic reactions, but can become ineffective if
particularly viscous systems are encountered. They may also be ineffective
if the reaction vessel has to be placed inside another piece of apparatus such
as a heating mantle or large cooling bath. In such cases the extra apparatus
can effectively shield the re action flask from the magnetic field created by
the stirrer machine. Magnetic stirring can also be a problem for large scale
reactions (reaction volumes over one litre), and in such cases mechanical
stirring is a valuable alternative.
speed control
!
motor_
stirrer guide-
Figure 8.30
glass rod
~
Teflon paddle glass rod
+ paddle
~+
\
slot to aJlow attachrnent
- ,/
Figure 8.31
138 Carrying Out The Reaction
Figure 8.32
Such guides are necessarily expensive, and for many uses that do not
require rigorously controlled reaction conditions a glass guide can be
employed (Fig. 8.32b). This consists of a precision-ground glass tube
attached to anormal ground glass joint. In this case it is essential to use a
ground glass rod that forms a good fit with the tube. Oil lubrication is
required to allow the rod to rotate, and so this set-up is not suitable for
reactions above room temperature that involve volatile solvents which can
leach the lubricant into the reaction mixture.
Because mechanical stirrers use an electric motor mounted above the
reaction flask, there is a serious danger of sparks from the motor igniting
volatile flammable solvents. In such instances the use of mechanical stirrers
powered by compressed air is recommended.
Carrying Out The Reaction 139
reaction flask
~ counter balance
~ flask
Figure 8.33
Mechanical shakers come in many designs, and are simply motors that will
shake an attached reaction flask. The flask is usually clamped to the shaker
(Fig. 8.33), often with a counter-weight to balance the machine. This is a
useful device for reactions that involve vigorous mixing of two immiscible
liquids for prolonged periods of time, and can also be employed when
efficient mixing of agas and liquid are required (e.g. hydrogenation);
however, for most applications other methods of agitation are preferable.
8.6.4 Sonication
septum
----
_ ultrasonic
balh
Figure 8.34
140 Carrying Out The Reaction
~
~
ultrasonie probe
Figure 8.35
CHAPTER 9
9.1 Introduction
It is important to give some thought to the work up of the reaction before
you attempt it. There are several aspects which need to be considered which
will be dealt with here. First of all do make sure that the reaction has indeed
finished (by careful analysis using your chosen monitoring system). When
using tlc analysis it is sometimes difficult to judge by spotting the reaction
mixture directly on to the tlc plate. In these cases it is often possible to get a
more accurate assessment by withdrawing a small aliquot of reaction
mixture by syringe and adding it to a small vial containing a few drops each
of ether and aqueous ammonium chloride. Agitation followed by tlc of the
organic phase will often give clean, reliable tlc information. This technique
can also be used to screen alternative work up conditions, for example
adding to water or aqueous base rather than ammonium chloride solution, or
using other organic solvents in place of ether.
Having satisfied yourself that the reaction has run to completion, or that
it is time to end the experiment, the appropriate 'quench' is added to the
reaction mixture. Choice of this reagent can be very important in
determining the yield of desired product, and it is obviously vital to use a
reagent or procedure which is safe. Given that the product is expected to be
reasonably stable, which usually is the case, then the choice of procedure
for quenching the reaction is determined by the reagent(s) used in the
reaction. Clearly we cannot cover all possibilities in this chapter, but
general procedures wh ich should cover most of the situations that are likely
to be encountered are provided below. The classification is made on the
basis of the nature of the reaction mixture which is to be worked up.
142 Working Up The Reaction
fine precipitates which are difficult to filter and which can lead to emulsions.
One simple method which often gives good results is to add a saturated
aqueous solution of sodium sulphate dropwise with stirring (and cooling!),
until a very heavy precipitate is forrned (do not add excess solution just to
'make sure'). The supernatant can then be decanted and the solid extracted a
few times with the re action solvent, combining all the organics. (If you try
this, do not dispose of the solid until you are certain that you have obtained
a good recovery of material. It might be necessary to extract further to
obtain all of the product).
With the reaction successfully quenched, then the isolation of the cmde
product can be carried out, as outlined below.
9.4 Purification
When a product has been isolated from areaction the next step is to purify
it. The degree of purity required will depend on the use for which the
sampie is intended, a synthetic intermediate might only require rough
purification, whereas a product for elemental analysis would require
rigorous purification. This section describes the most important purification
techniques, crystallization, distillation, sublimation, and chromatography. It
is assumed that the reader is familiar with the basic principles of these
methods, so the emphasis is on more demanding applications such as the
purification of air-sensitive materials, and purifications on a micro-scale.
9.4.1 Crystallization
Simple recrystallization of an impure solid is a routine operation, which
nevertheless requires care and good judgement if good results are to be
obtained. The basic procedure can be broken down into six steps, which
are listed below, together with some tips on how to overcome common
problems.
1. Select a suitable solvent
Find a suitable solvent by carrying out small scale tests. Remember that
'like dissolves like'. The most commonly used solvents in order of
increasing polarity are petroleum ether, toluene, chloroform, acetone, ethyl
acetate, ethanol, and water. Chloroform and dichloromethane are rarely
useful on their own because they are good solvents for the great majority of
organic compounds. It is preferable to use a solvent with a boiling point in
excess of 60·C, but the b.p. should be at least lO·C lower than the m.p. of
the compound to be crystallized, in order to prevent the solute from 'oiling
out' of solution. In many cases a mixed solvent must be used, and
combinations of toluene, chloroform, or ethyl acetate, with the petroleum
ether fraction of similar boiling point are particularly usefu!. Consult
Appendix 1 for boiling points, polarity (dielectrlc constant), and toxicity of
common solvents.
2. Dissolve the compound in the minimum volume 0/ hot solvent
Remember that most organic solvents are extremely flammable
and that many produce very toxic vapour.
148 Working Up The Reaction
portions, until the compound has dissolved. Remove the apparatus from the
bath, wipe the bulb to remove oll or water, and quickly fIlter the hot solution
into a clean receiver by pressurising the vessel using hand bellows or an
inert gas line. Filtering under pressure in this way avoids the problem of
unwanted crysta1lization, and reduces transfer losses.
Figure 9.1
The hot solution can be fIltered into a sma11 conical fiask, but on a scale
of l00mg or less, a Craig tube (Fig. 9.2) gives better recovery because it
allows the crystals to be recovered without another fIltration.
glass centrifuge
rod- tube ----
crystals
Craig
tube - mother
liquor ------oO~~
(a) (b)
Figure 9.2
Filter the hot solution into a suitably sized Craig tube and cover the tube
with aluminium foil while crystallization occurs. When crystallization is
complete fit the matching (a close fit is essential) glass rod into the Craig
tube and secure it tightly with a rubber band. Place the inverted assembly in
a centrifuge tube (Fig. 9.2b) and put the centrifuge tube (and a counter-
Working Up The Reaction 151
with inert gas and placed in a cooling bath. It should be cooled slowly by
gradual addition of the cooling agent to the solvent. When crystallization is
complete the bubbler is disconnected and the filter stick is connected to an
appropriately sized receiver using chemically inert tubing (Teflon). The
filter stick is then lowered into the solution and the mother liquor is forced
through into the receiver using inert gas pressure. The thermometer adapter
will allow sufficient freedom of movement of the filter stick to pack the
crystals to the bottom of the flask and drain off the mother liquor
thoroughly. The crystals can then be washed by releasing the gas pressure
and adding small amounts of precooled solvent via the three-way tap, using
a cannula. The washings can then be removed using the filter stick as
before. The cooling bath can be removed and the crystals isolated and dried
in the usual way or the low temperature recrystallization can be repeated in
the same flask.
to bubbIer
t
flltcr
stick
cooling
balh -
Figure 9.3
The sintered disc of the filter stick should be of porosity 3 or larger in
order to avoid blockage. A convenient alternative to a filter stick is to use an
ordinary glass rod with filter paper wrapped round the end (Fig.
9.4b).Wrap some Teflon tape round the end of the tube, then carefully fold
filter paper over the end and sec ure it with wire. The Teflon tape will give a
better seal because the wire will sink into it. Another alternative, which is
Working Up The Reaction 153
, / glass tube~
_ _ Teflon tape
sintered
/ disk
Figure 9.4
Small scale low temperature recrystallizations can be carried out in the
same device recommended for ordinary small scale work. The solution is
filtered into the flask and the apparatus is then purged with inert gas. The
outlet is sealed with a small septum and the flask immersed in a cold bath
almost up to the level of the sintered disk (Fig. 9.5).
inert
gas
cooling
bath - t&Ss:S:sS~~~~~
Figure 9.5
154 Working Up The Reaction
- inert gas
to
bubbier
t
Figure 9.6
9.4.2 Distillation
Distillation is the most useful method of purifying liquids. It is used
routinely for purifying solvents and reagents and, with care and the correct
apparatus, it can be used to separate liquids whose boiling points are less
than 5°C apart. We will assume that the reader is familiar with the
fundamentals of the theory and practice of distillation but it is appropriate to
begin by reiterating some basic safety rules.
1. Never heat a c10sed system.
2. Remember that most organic liquids are extremely flammable so great
care must be taken to ensure that the vapour does not come into contact
with flames, sources of sparks (electrical motors), or very hot surfaces
(hot plates).
3. Never allow a distillation pot to boil dry. The residues may ignite or
explode.
156 Working Up The Reaction
Simple distillation
The conventional apparatus for simple distillation is shown in Fig. 9.7. It is
only useful for distilling compounds from involatile residues, or for
separating liquids whose boiling points differ by at least SO·C. Moisture
can be excluded by attaching a drying tube to the vent or by connecting the
vent to an inert gas line. It is essential to add so me boiling chips, or better
to stir the liquid, in order to prevent bumping, especially if a finely divided
solid, such as a drying agent, is present.
_ ___ thermometer
/ thermometer adapter
/ stillhead
receiver
adapter
/
tirring bar
Figure 9.7
The flask should be heated in a water bath, or an oil bath, not with a
Bunsen burner or a heating mantle (Section 8.4.2). The temperature of the
bath should be increased slowly until distillation begins and then it should
be adjusted to give a steady rate of distillation. If the boiling point is high
Working Up The Reaction 157
(> 150·C) the stillhead may need to be lagged with glass wool, and an air
condenser should be used rather than a water condenser.
For routine solvent distillation it is more convenient to use a compact
distillation apparatus such as that shown in Fig. 9.8. These can be bought,
or constructed in two or three convenient sizes to fit the common ground
glass joints.
solid
glass rod
Figure 9.8
If the distillation is being used to dry areagent, the process must be
carried out under an inert atmosphere. To do this, first of all dry all glass
apparatus in an oven, or with a heat gun under vacuum, and purge with
argon whilst cooling. This is most easily accomplished by connecting the
apparatus to a double manifoldlbubbler system (see Chapter 3). Although
Quickfit assemblies can be used, we prefer one piece type distillation
apparatus, as shown in Fig. 9.8 and Fig. 9.9b. When the glassware has
cooled, increase the argon fIow, quickly disconnect the distillation fIask,
add any drying agent required, a few anti-bumping granules, and the liquid
to be distilled, and reassemble the system. Heat the distillation fIask in an
oi! bath (do not carry out distillations using a heating mantIe) and collect the
distillate which comes over at the required temperature. When the
distillation is complete remove the collector and seal it quickly with a
septum. Most reagents can simply be poured into areagent bottle before
sealing, provided you are quick. However, if the reagent is particularly
sensitive to air or moisture a cannulation technique should be used to
transfer it (see Chapter 5). Whatever type of container is used for storage it
is always preferab1e that it be fuH or nearly so.
158 Working Up The Reaction
packing
material
vacuum
jacket
_ v.acuum
Jacket
packing
support
Figure 9.9
Fractional distillation
Separation of liquids whose boiling points are between SOC and 50·C apart
requires the use of an apparatus which gives better contact between the
vapour and liquid phases in the distillation column. Columns which do this
are called fractionating columns and the most common type, a Vigreux
column, is shown in Fig. 9.9. A 50cm long vacuumjacketed Vigreux (Fig.
9.9a) should allow reasonable separation of compounds which boil30-40·C
apart. The Vigreux assembly shown in Fig. 9.9b is less efficient but is
convenient for routine distillations and gives quite good results at reduced
pressure. The keys to getting good results from a fractional distillation are
to raise the temperature very gradually, and to collect the distillate very
slowly.
Efficient separation of compounds with a boiling point difference of 10-
30·C can be achieved using a long glass tube packed with glass rings or
helices, or for high efficiency, wire mesh rings (Fig. 9.9c). A description
of the technique for operating such columns is beyond the scope of this
Working Up The Reaction 159
400-::
700-::
600-':
300""
500-':
400~
200~
300-::
100~ 100-=
0-=
Figure 9.10
-- Quickfit thermometer
stirring bar
Figure 9.11
The chief difference from the simple distillation apparatus is in the
design of the receiver adaptor. This must allow several fractions to be
Working Up The Reaction 161
collected without needing to break the vacuum. The simplest design is the
'pig' type shown in Fig. 9.12a. When using a pig, remember to grease the
joint lightly so that the receiver can be rotated while under vacuum, and to
fix the receivers securely using clips.
inert
receiver gas
(a) (b)
Figure 9.12
The procedure for carrying out the distillation is as folIows.
1. Place the sample in the distillation flask (no more than two thirds full)
and add a stirring bar. Anti-bumping granules are not effective at
reduced pressure and so an alternative must be used. A very narrow
capillary which allows a slow stream of air or nitrogen bubbles to pass
through the solution is effective, but brisk stirring using a magnetic
folIower is much more convenient.
2. Assemble the (oven-dried) apparatus, putting a litde high vacuum grease
on each joint. Ensure that the receiver adapter and the collection flasks
are secured using clips, and connect the assembly to a vacuum pump.
One convenient method of doing this is to connect it to a vacuumlinert
gas double manifold (Chapter 3). The pump must be protected with a
cold finger trap and the line should contain a gauge for monitoring the
pressure (Chapter 7).
3. Stir the liquid rapidly and carefully open the apparatus to the vacuum.
Some bumping and frothing may occur as air and volatile components
are evacuated. Adjust the pressure to the required value by allowing
inert gas into the system via a needle valve.
162 Working Up The Reaction
4. Reat the flask slowly to drive off any volatile impurities and then to
distil the product. Monitor the stillhead temperature and collect a forerun
and a main fraction, which should distil at a fairly constant temperature.
Iffractionation is required you may have to collect severaHractions and
it is very important to distil the mixture slowly and steadily.
5. Stop the distillation when the level of liquid in the pot is running low, by
removing the hearing bath.
6. Isolate the apparatus from the vacuum and carefully fill with inert gas.
The flask containing the distillate is now under a dry, inert atmosphere
and should be quickly removed and fitted with a tightly fitting septum.
7. Switch off the pump and clean the cold trap.
APerkin triangle (Fig. 9.12b) is a more convenient device for
collecting fractions on a larger scale. It is operated as follows.
1. Attach a receiver and open both the apparatus and the receiver to the
vacuum using taps A and B. Open tap C and collect the forerun in the
receiver.
2. Close tap C so that the next fraction will collect in the bulb D, while
you are removing the receiver. Use tap B to allow inert gas/air into the
receiver, and then replace the receiver with a new one.
3. Close tap A to temporarily isolate the still. Evacuate the receiver by
opening tap B to the vacuum. Wait until the pressure has steadied,
open tap A, and then open tap C to allow the distillate to drain into the
new receiver.
4. Continue to collect fractions by repeating steps 3 and 4 as necessary.
- [0 vacuum
stirring bar
Figure 9.13
Another popular solution to these problems is the Buchi Kugelrohr
apparatus (Fig. 9.14). The key features of this system are (i) a horizontal
oven (made of glass or metal) with an iris c1osure, which heats the flasks
efficiently, (ii) short path distillation between aseries ofbulbs which can be
moved horizontally in and out of the oven, and (iii) a motor to rotate the
bulbs (optional). The Kugelrohr can be used to distil very sm all quantities
« lOOmg) and can be operated at high vacuum. A simple distillation is
carried out as folIows. Place the sampie in the end bulb using a Pasteur
pipette. If necessary wash the sampie into the bulb and then evaporate the
solvent on a rotary evaporator. Add two more bulbs and a straight length of
tubing, insert the tube in the motor assembly, and slide the end bulb into the
oven. Gently c10se the iris type seal gently onto the connecting joint. Set
the bulbs rotating (to prevent bumping and speed up the distillation), apply
the vacuum and raise the oven temperature gradually. When distillation is
complete allow the flasks to cool under an inert atmosphere and then remove
164 Working Up The Reaction
the bulbs from the apparatus and recover the distillate. Fractionation of
compounds with a 20-30°C boiling point difference can be achieved by
inserting all the bulbs except one into the oven, distilling the most volatile
component into that bulb, withdrawing the next bulb from the oven and
distilling the next fraction into that, and so on.
"- motor
unit
sliding bar
control unit
Figure 9.14
9.4.3 Sublimation
Sublimation is an excellent method for purifying relatively volatile organic
solids on scales ranging from a few milligrams to tens of grams. At reduced
pressure many compounds, especially those of low polarity, have a
sufficiendy high vapour pressure that they can be sublimed, i.e. converted
directly from the solid phase into the vapour phase without melting.
Condensation of the vapour then gives purified solid product provided, as is
often the case, that the original impurities were much less volatile. Two
designs of sublimation apparatus are shown in Fig. 9.15. The larger
sublimator (Fig. 9.15a) consists of a tube with a side arm which is fitted
with a cold-finger condenser. It is used as follows. If the crude material is
asolid, powder it and place it in the bottom of the outer vessel. If it is waxy
or oily, wash it into the tube with a small amount of solvent, cover the side
arm with a septum, and remove the solvent on a rotary evaporator. Put
Working Up The Reaction 165
some vacuum grease on the joint of the cold-finger condenser and fit it into
the sublimator (there should be a gap of approximately lcm between the
solid and the condenser). Evacuate the apparatus slowly to try to prevent
any spattering of the solid. Turn on the condenser water and slowly heat the
base of the sublimator. A fine mist of sublimed material on the condenser
indicates that sublimation is beginning and the temperature should then be
held fairly constant until the process is complete. At this stage the sublimed
product may be c1inging precariously to the cold finger so proceed with
great care. Turn off the water, carejully allow air/inert gas into the
sublimator, and very carejully remove the cold-finger. Scrape off the
product with a microspatula.
to
- wateT out vacuum
oven
- tovacuum
cold -finger
condenseT
I!,......!J..,..-- sublimate
crude
- cTude
(b)
Ca)
Figure 9.15
The simple glass tube shown in Fig. 9.l5b functions in a similar way
exeept that the water eondenser is omitted. Plaee the erude sampIe in a small
sampIe vial, put a plug of glass wool in the neck, and drop the vial into the
tube. Put a plug of glass wool in the neck of the tube and attach it to a
vacuum line. Evacuate carefully and heat the base of the tube gently in an
oil bath or in a custom designed metal block as shown. Proceed as before
and isolate the sublimed product by cutting the tube and scraping out the
solid.
166 Working Up The Reaction
for running flash columns and modifications have therefore been introduced
over the years. These make the technique easier to carry out and somewhat
safer. This seetion will lead you through the basie steps for running a
modified version of flash chromatography whieh we find to be very
convenient and effective. Flash chromatography is a very powerful and
rapid technique for separation of organic compounds, but like all
chromatographie techniques you will only gain expertise by experience, and
these instructions can only be considered as guide-lines. Every separation is
different so do not be despondent if you have some failures at first. With
practice you should be able to acquire the skill and intuition required to get
good separation, even in the most difficult circumstances, every time, and
quickly!
Equipment required
The set-up originally described by Still requires the use of long columns, to
accommodate a reasonable supply of solvent, but these are diffieult to load
and need to be dismantled for addition of extra solvent during the run. We
therefore recommend the use of shorter columns plus areservoir. It is
important to have a familiar set of columns on hand and we suggest that you
have a set of about five columns, ranging in diameter from about 5mm to
50mm. A very convenient length for all columns is 25cm, except for those
with a very narrow bore, which can be shorter (Fig. 9.16). It is convenient
to standardize on a B24 (24/40) joint on the top of the column and a
glassblower can make reservoirs from a round-bottom flask and a male
joint. The 250ml reservoir is used more than any other, but it is also useful
to have 100ml, 500ml, and 1 litre sizes available.
We find that it is difficult to control the column pressure using a
Rotaflow valve as suggested in the original paper and this can be dangerous
if too much pressure is inadvertently applied. Two alternative methods are
used successfully in our labs. One of these is simply to use a rubber
bellows attached directly to a tubing inlet and apply enough pressure to give
the required flow rate. The only problem with this technique is that it is
difficult to keep the pressure constant, partieularly with large columns. An
alternative is to construct a simple pressure-release valve, such as that
shown in Fig. 9.17 and use this in conjunction with a compressed air
supply or nitrogen cylinder.
168 Working Up The Reaction
+-B24
(25/40)
25cm
1
+-500ml
+-Teflon
stopcock
Figure 9.16
The glass part of the 'flash adapter' is made from a B24 joint and a
piece of 14mm i.d. heavy wall tubing. The valve is simply a septum, with
the skirt cut off, glued to a shaped metal (brass) plate.
-
l.4mm id.
V.. UI2cm
~
~
~
~
~
~
I
7cm
a..
I I
~ ~
§~
~ ~
~ ~
~ ~ §
~
~ 0 ~
~ § 4cm
§
12.5cm
~ ~ §
~
~
~ 1
Figure 9.17
Working Up TheReaction 169
1.0
0
Rf 0.5 0
0 0
0 0
0 0 0 0 0
0.0
O~
0
Rf 0.5
0
0 0 0 0 0
0
0 0
0
o X 0
0 S ~ o X 0
0.0
Figure 9.18
ii) You should try to use as little silica as possible since it is quite
expensive. Where the component you require is well separated from
other components, a ratio of ca. 20: 1 (silica:mixture) should be
sufficient. For more difficult separations up to 100: 1 ratio can be used
(with a ratio of 100: 1 spots that are touching on tlc should be separable
in the appropriate solvent system). As a rough guide some
representations of tIes are given in Fig. 9.18 which indicate a solvent
system which could be selected for separation of the mixture. Below
the drawings, an approximation of the relative quantity of silica
(compared to the sampIe) required to bring about separation is given.
Working Up The Reaction 171
pasteur
pipette
Figure 9.19
Note:
The eolumn onee prepared ean be left for a very short time, but onee
you load the sampie the remaining steps should be earried out as swiftly
as possible, and the solvent flow should be eontinued without
interruption if possible, especially during the early stages of elution.
Working Up The Reaction 173
Leaving the column standing with the sample loaded leads to band
dispersion and loss of resolution. So, before starting the remainder of
the procedure make sure you have everything that you will need to
hand, including plenty of solvent.
3. Load the sampIe
i) Dissolve the sample in the minimum amount of solvent, preferably the
same solvent that you intend to run the colurnn in (if this is not possible
as is often the case, dissolve the sampie in a small amount of
dichloromethane)t. Keep a tlc sample of the sampie mixture to
compare with the column fractions.
ü) Load the solution onto the top of the column, very carefully, using a
Pasteur pipette to drip it around the walls of the column, just above the
sand. Caution must be taken not to disturb the layer of sand. Repeat
the procedure using the minimum quantity of solvent to rinse any
remaining sample from the flask.
üi) Once all the solution has been added allow the level of liquid to drop so
that the top of the sand is just starting to dry.
4 . Add the solvent
i) Add the solvent to the top of the column, very carefully at first, again
using a Pasteur pipette to drip solvent around the walls of the column
just above the sand, and taking care not to disturb the sand. On ce a
head of solvent is present, more can be poured in carefully from a
beaker.
ü) Attach a solvent reservoir to the top of the column, secure it with elastic
bands (Bibby clips are not strong enough), and fill with solvent.
5. Run the column
i) Connect bellows or flash adapter to the top of the reservoir and secure
with elastic bands (your set-up should then look like the diagram
shown in Fig. 9.19).t
ii) Apply pressure to give a fast solvent flow rate and collect fractions
continuously. It is very important to maintain a fast flow rate through
t For some very non-polar compounds, even sm all quantities of methylene chloride
may cause elution problems and an alternative method of loading can be used. Thus, a
solution of the sampIe is added to a small amount of silica in a round bottom flask and
the mixture is evaporated to dryness. The dry impregnated silica is then added to the top
of the pre-packed column.
174 Working Up The Reaetion
the column - the solvent should run, rather than drip! A slow flow rate
eauses redueed resolution, NOT improved separation. tt The size of
fractions will depend mainly upon the size of the column and as a rough
guide, they should be in m1 about half the weight of silica (Le. for a
30g column you should collect ca.15m1 fractions). 1t is a myth that you
get less mixtures by collecting smaller fractions - the mixture simply
appears in more tubes and leads to a good deal of extra work! You may
feel safer collecting relatively small fractions at fIrst, but as you become
more experienced you will tend to collect larger fractions, and thus
considerably lower the amount of time you spend on chromatography.
Always be eareful not to let the eolumn run dry.
iii) Monitor the column fractions by running tlcs whilst the column is
running (5 or 6 spots per tlc plate is usually a convenient amount). You
should have time to apply a spot of the previous fraction to a tlc plate
whilst the present fraction is collecting.
6. Analyse tles and eombine fraetions
When all the compounds you are interested in have eluted, and you
have identified in which fractions they are, combine the fractions as
appropriate (keeping fractions that contain mixtures separate from those
containing pure materials). Remove the solvent from the combined
fractions on a rotary evaporator, and fInally remove the last traces of
solvent under high vacuum.
-
000
o
o
_ L..-_ _- - '
Gradient elution
When the components of a mixture run c10se together, a single solvent
system which gives the spots a tlc Rf of 0.2-0.3, will be effective.
However, when the spots are a long way apart, increasing the
solvent
tt Apressure of 7psi is reeommended in Still's original paper, but we tend to rely on
flow rate, rather than press ure.
Working Up The Reaction 175
polarity as the column is running will save a good deal of time and solvent -
this is referred to as gradient elution. You should be confident with single
solvent chromatography before you attempt gradient elution, as it requires
quite a bit of experienced judgement. The procedure for doing this is:
1. Start running the column in a solvent which will give the highest
running spot an Rf of 0.2-0.3.
2. When tlc analysis indicates that this component is almost completely
off, change the solvent polarity to that which gives the second spot an
Rf of 0.3. In some cases you may need to change the solvent polarity
in steps.
3. Continue this process until all the spots that you require are off.
Figure 9.21
sample and solvent addition. The approximate funnel sizes, compared with
the quantity of sample to be applied, are given in Table 9.1.
Table 9.1
Funnel Funnel Weightof Weightof Fraction
Diameter(mm) Length(mm) Silica(g) Sample(mg) Size(ml)
30 45 15 15-500 10-15
40 50 30 500-2000 15-30
70 55 100 1000-5000 20-50
2. Pre-elution
Under vacuum, pre-elute the column with a solvent whieh will give a t1c Rf
of about 0.2 for the least polar constituent of the sampie. If the silica has
been packed correctly, the solvent should run down the column with a
horizontal front, but if it channels, the column should be sucked dry and re-
packed. Keep the surface of the silica covered with solvent while pre-
eluting until solvent starts collecting, then suck it dry.
3 . Loading the sampie
Load the sample as a solution, in the same solvent as used for pre-elution, in
an even layer onto the surface of the silica. Alternatively, if the sampie is
insoluble in the pre-elution solvent, it can be pre-adsorbed onto a small
quantity of silica (see Seetion 9.5.1, loading a flash column)) which is then
spread on the surface of the silica in the funnel.
4. Eluring the column
Sequentially add solvent fractions to the funnel according to the quantities
indieated in Table 9.1, sucking the silica dry in between each fraction, and
keeping fractions separate. For each successive fraction increase the solvent
polarity by increasing the proportion of the more polar solvent by about 5-
10% (e.g. from 50% EtOAc/50% petrol to 55% EtOAc, for a 10%
increase). Analyse the fractions by tlc to determine the locations of the
components of interest, but as a rough guide, the solvent mixture whieh
would give the compound a tlc Rf value of about 0.5 will probably elute it.
As with any chromatographie technique expertise will only come with
experience, but given that, you should be able to separate quite closely
running compounds quickly using this technique.
178 Working Up The Reaction
used many times, thus avoiding the expense of throwing away large
quantities of silica.
pump ~ r--------, /
o 0 0 0
+- main
column
fraction
/ colleclOT
dmfrJ
.:-:- :.:.:- +- solvenl bolLle
Figure 9.22
The mplc system is essentially a simplified and much cheaper version
of an hplc set-up. At the heart of the system is any type of pump which will
operate at lOOpsi, with a controllable flow rate of up to lOOml/min. There
are a variety of moderately-priced pumps on the market, including one
produced by Buchi, which is especially designed for mplc. However, we
use a Gilson hplc pump, which has interchangeable flow heads. For mplc,
a lOOml head is fitted, but the same pump can also be used for either
analytical or preparative hplc, simply by changing the head. Since glass
columns are used for mplc it is important that apressure limiting valve is
fitted in-line, to prevent high pressure building up in the column, in the
event of a blockage. The valve should normally be set to cut off the pump at
about lOOpsi, but you should consult the instructions for your columns
before setting this.
180 Working Up The Reaction
All the pipework used in the mplc system is 3mm Teflon tubing, which
is very easily tailored to your needs. The tubing is connected to the various
components of the system by plastic screw ferules, which should only be
tightened finger tight. 1t is very easy to make up the tubing to suit your
needs, but you will need a flanging tool to fit the ferules onto the tube ends
(Fig. 9.23). The tubing can be connected using threaded plastic sleeves
(Fig. 9.24).
ferule
"'"
.z=J>. I.\\\\\\\\§ ..
hot flanging
tao 1
meta] flange
~ teflon rubing
~\\\\\\\·It:::===:=::::j .\\\\\\\~
+
rn·:~::;·:sZ~~:~;·~;·;:~ :·+tJ
Figure 9.24 Connecting 3mm tubing using a threaded sleeve
The sampie is introduced into the system via an injection valve, but this
is a much simpler and less expensive valve than that found in an hplc
system. We use a Rheodyne type 50 Teflon rotary valve, which will take a
flow rate of 100mIlmin (Fig. 9.25). The pump and the column are
connected to the red and white connectors respectively, whilst a sampie load
loop (see below) is fitted to the black and yellow connectors. A Luer fitting
is attached to the blue connector, which is where the sampie is introduced,
using a syringe. The green connector is a vent and should be positioned so
that effluent from it can be collected in a beaker or flask.
Working Up The Reaetion 181
§l1I'IIIIAI:
sideview
g
pump 8 sampie
pump
loop
column
Anti-clockwise position Clockwise position -
- sampie loaded onto loop sampie flows onto column
by injection at port 6
should be removed from the sampie before you start. This can be done
either by simple flltration or by flltration through a small pad of silica in
a short column. While the mp1c column is equilibrating, make up a
concentrated solution of the sampie in either the chosen solvent or in
methylene chloride.
4. Load the sample
Switch the injection valve to the 'load' position (Fig. 9.26, solvent by-
passing the loop). Using a good syringe with a flat-ended needle, draw
up the sampie, then invert the syringe and hold it with the needle
pointing upwards. Next draw in an air bubble of about Iml and
remove the needle from the syringe. Still holding the syringe in the
inverted position, carejully connect it to the Luer fitting on the injection
valve. Now turn the syringe the right way up and inject the sample
until all the solution and a very small bubble have entered the load loop.
To avoid a sudden pressure build-up as the sampie is injected, reduce
the flow rate by half, then switch the injection valve quickly. As the
sampie begins to enter the column, gradually turn the flow rate back up
and start the fraction collector.
5. Running the column
Whilst the column is running, collect dcs or use the detector to
determine which fractions contain the required components. Unless
you are using a refractive index detector, gradient elution is simply
achieved by gradually adding more polar solvent to the reservoir.
6. Finishing the run
Before you leave the system make sure all the components that were
loaded onto the column have been eluted and if necessary, run a more
polar solvent through to remove polar material. The column can be left
with a length of Teflon tubing attached between the ends to prevent it
drying out.
7. Regenerating columns
Sometimes a column will become contaminated with stubborn polar
material. 'Back-washing' some columns is possible, but this is not
necessarily the best way to treat them. It is better to run a polar solvent
through, such as methanol ( or even water), then gradually change the
Working Up The Reaction 185
Equipment required
A preparative hplc system has exactly the same components as an analytical
system, except that several features are larger (see Chapter 8). With some
systems, such as the Gilson, the pump head can simply be changed to
provide the higher flow rates required. The injection valve must also be the
large bore type (e.g. Rheodyne 7125) fitted with a large load loop (up to
5ml), and of course the columns are much larger. It is also useful, but not
essential, to have a fraction collector, and if this is microprocessor
controlled, it can be liked to the detector so that fractions are only collected
when a peak is being detected. There are now several extremely
sophisticated, computer-controlled systems available, which can be set to
inject sampies automatically and collect selected peaks into designated
flasks. Thus, large quantities of material can be separated over multiple
injections. However, for most research purposes the prime requirement is
for one-off separations.
The choice of detector can be quite critical. Uv detectors are very
sensitive but are of litde use if molecules without chromophores are being
separated. A refractive index detector is universally applicable but has the
drawback that gradient elution is virtually impossible.
Using these equation, the scale-up factor between the 4.5mm column
and the 22.5mm column is 22.35. and between the 22.5mm and the 45mm
it is 4. If the columns are not the same length then simply multiply the
factor by the proportionate length difference (for example going from a
4.5mm x 12.25cm column to a 22.5 x 25cm column the scale-up factor
would be 44.7). To work out the conditions for running the larger column
simply multiply the flow rate and the quantity loaded on the small column by
the scale-up factor and this should produce identical results on the large
column. As a rough guide a 4.5mm x 25cm column runs at about
O.75m1/min and can be loaded with 5-35mg; a 22.5mm x 25cm column runs
at about 16m1/min and can be loaded with lOO-800mg and a 45mm x 25cm
column runs at about 64mllmin and has a capacity of about 450mg to 3.2g.
Once you have worked out a good system for prep hplc the run times are
quite short, so that large amounts of material can be separated quite
conveniently by multiple runs.
10.1 Introduction
For the purposes of this chapter we will define small scale reactions as those
involving re action mixture volumes of less than 5ml. When performing
organie reactions on this sc ale special problems arise, most notably:
1. Difficulties in measuring out small quantities of sensitive reagents.
2. Significant losses of material due to apparatus design.
3. Difficulties in excluding trace amounts of water from moisture-sensitive
reactions.
Whenever reactions are performed, material los ses are obtained as a
consequence of the above problems. Normally these los ses only account
for a few percent of the total material, however this percentage can increase
dramatically as the reaction sc ale decreases. For example if a moisture-
sensitive reaction was carried out on a one mole scale, it would take 18g of
water to completely stop the reaction occurring, however if the same
reaction is carried out on a O.1mmol sc ale, then only 1.8mg of water would
completely quench the reaction.
The problem of weighing out small quantities of sensitive reagents is
best solved by accurately weighing larger quantities, and making up
solutions in an inert solvent, ideally the reaction solvent. Since the molarity
of the solution is known, a quantitative aliquot of this solution can then be
added to the reaction mixture using a syringe. This effectively reduces the
problems of weighing out the material to those that exist in larger scale
reactions, and discussed in earlier chapters. As a general rule many of the
techniques used in setting up reactions discussed in the earlier chapters, can,
with care, be applied to small sc ale reactions. Indeed there are a wide
Small Scale Reactions 189
sampIe vial
/ extraction
-
solvent
--:::--: ~
reaction
•: •: . :. -mixture
Pasteur pipette
:/
- or
- ... _
~~~~~~~ -
extracts
drying agent
""'"
Figure 10.1
tall, thin vial rather than a short, fat one. A second vial containing drying
agent can then be used to dry the extracts.
Removal of the drying agent is normally achieved by filtration of the
solvent mixture, and on a small scale this is best achieved using a Pasteur
pipette fitted with a cotton wool plug as the filtration apparatus (Fig. 10.2).
Once the solvent has been transferred into the filtration pipette, it can be
forced through the plug by applying pressure with a pipette teat.
Evaporation of the solvent in the normal way then yields the emde reaction
product. As stated earlier, material is inevitably lost with each transfer of
apparatus. 1t is possible to cut down the number of transfers by using the
sample vial as the reaction vessel. Very small magnetic fleas are now
commonly available, and will fit most small sample vials. Consequently,
with magnetic stirring, the vials can serve as small reaction vessels (Fig.
10.3). They are conveniently attached to the top of a magnetic stirrer
machine using plasticine.
Small Scale Reactions 191
/
tcat-
---
Pasteur -:-
pipette
~
o
o
-
cotton wool -
plug -
riltered
Solulion
/
Figure 10.2
When capped, the vial is a sealed system, consequently this set-up is
only useful for small scale reactions at room temperature that do not involve
an increase or decrease in press ure inside the reaction vessel. For reactions
at low temperature, or those requiring a positive pressure of an inert gas
atmosphere, it is often more convenient to use a Pyrex test tube fitted with a
septum (Fig. 10.3). In all other respects the arrangement is the same and,
since Pyrex test tubes are available in a range of sizes, this apparatus can
inert gas
I
eap
j
!- septum
::;
-.;
sampie via1--
- test tube
reaclion
_____ mi)tlure~
=. __ magnelic
ficas - =
Figure 10.3 Use of sampie via! or test tube as areaction vessel
192 Small Scale Reactions
cope with a range of reaction volumes down to about O.2ml. Again the
reaction vessel can also be used for a subsequent extraction procedure
simply by replacing the septum with a bung.
,
a one-piece apparatus incorporating a water condenser can also be used
inert gas
,
inert gas
_ - - septum _ __ : 11 i
air
condenser _ water
condenser
~/~~
Figure 10.4 Small scale air condenser and water condenser systems.
Small Scale Reactions 193
(Fig. 10.4). In this case the condenser system is more efficient, however
construction of the apparatus is correspondingly more complex. In both
cases the reflux apparatus can be easily constructed to allow re action
volumes down to about O.5m1.
septum_ iii
reaction
mix.ture~ -.
Figure 10.5
The reaction can then be monitored by recording the spectrum of the
reaction mixture direct1y. There are however several important points to
note when using nmr tubes as reaction vessels:
194 Small Scale Reactions
105.1 Distillation
By far the most convenient method of carrying out distillation on a small
sc ale is to use a Kugelrohr apparatus (see Chapter 9). In order to cut down
on los ses through ground glass joint connections it is often necessary to use
a one piece Kugelrohr bulb set (Fig. 10.6). This can be conveniently made
to the size required from a piece of Pyrex tubing.
After distillation is complete the apparatus is left to cool, and the
purified material recovered by cutting up the apparatus into three seetions
using a glass knife.
Small Scale Reactions 195
G)()(j~
./'
erude sampIe 1
c ~
~
~
Figure 10.6
10.5.2 Crystallization
Crystallizations on a small scale are most conveniently carried out using a
Craig tube apparatus as described in Chapter 9.
10.5.3 Chromatography
All the normal chromatography techniques (see Chapter 9) can be used to
purify small quantities of material. Indeed preparative hplc and glc are often
more successful when small quantities of material are involved.
In the case of flash chromatography however, it is often impractical to
simply scale down the equipment. A useful alternative is to employ a
Pasteur pipette as the column. Such a column is easily constructed using a
pipette containing a cotton wool plug (Fig. 1O.7a). The pipette is then filled
with the required adsorbent (typically silica gel). The amount of adsorbent
used depends upon the quantity of cmde sampie to be purified, however it is
inadvisable to fill the pipette more than three-quarters full, otherwise there is
insufficient room for the solvent. Next, the eluting solvent is added to the
top of the column and allowed to run through the column under gravity.
More solvent is added to the top of the column as required. Once the
solvent starts to appear at the bottom of the column, pressure can be applied
196 Small Scale Reactions
using a pipette teat, forcing the solvent through at a faster rate. After about
lcal -
PaSleur pipette_
- solvenl
- ad,orbent
-adsorbent
(a) (b)
Figure 10.7
two column-volumes of solvent have been passed through the silica it
should be ready for use. The sampie is applied to the top of the column in
the usual way and pressure is applied using a pipette te at. The main
difference between this arrangement and the more usual flash
chromatography set-up is that the pressure applied to the top of the colurnn
is not constant. The teat is constantly being removed to a110w the addition
of more solvent to the top of the column. Consequently the solvent is not
passing through the column at a constant rate. In most instances this does
not appear to significantly affect the separations achieved. If, however, this
proves to be a problem, a miniature chromatography column with solvent
reservoir can easily be constructed from Pyrex glass tubing (Fig. 1O.7b).
With a cotton wool plug in the bottom this can be used in exactly the same
way as the larger columns described in Chapter 9.
CHAPTER 11
11.1 Introduction
This chapter deals briefly with some of the specialized techniques required
when working with larger scale reactions. For the purposes of this chapter
we will define a large scale reaction as one involving reaction volumes of
between 2 and 5 litres. Working on reaction volumes in excess of this
usually requires the use ofpilot-plant equipment and is beyond the scope of
this book.
When working on larger reaction volumes, several problems arise as a
consequence of the scale:
1. The use of syringes to add liquids to areaction becomes impractical if
you require to add volumes of more than 50m!.
2. Stirring the reaction mixtures can become a problem, because magnetic
stirrers become ineffective for volumes much above 1 litre.
3. Control of the reaction temperature becomes more difficult as the
reaction volume increases, because reaction mixtures will take much
longer to heat up or cool down.
4. Exothermic reactions can prove to be a major problem on a large scale
since they are prone to induction periods before reaction starts. After
the induction period, the reaction can heat up rapidly and, unless
extreme care is taken in these situations, the reaction can easily go out
of contro!. It is recommended that very careful monitoring of the
temperature is undertaken in such cases, and any addition of reagents
which may lead to an exothermic reaction is carried out slowly.
5. Purification of materials on a large scale is often less easily carried out.
198 Large Seale Reaetions
On the other hand, some problems that exist with smaller scale
reactions can be less troublesome when working on large scale. For
example, materiallosses that occur as a consequence of transferring the
material between pieces of equipment become insignificant. Moisture-
sensitive reactions are less of a problem because if traces of water get into
the reaction they will only affect a very small percentage of the reaction
mixture.
mechanical stirre T - o
dropping
Cunncl
inert gas - ............>===-
condcnser -
Figure 11.1
Large Seale Reaetions 199
mechanical stirrer_
incJt gas
jackclcd dropping
funnel - - f + -
=
= I~I-+I-+- coolant
=
low lcmpcrawrc
thcnnomcLcr -
Figure 11.2
200 Large SeaIe Reaetions
into the dropping funnel via a cannula by replacing the stopper with a
septum.
If the solution to be added from the dropping funnel requires cooling
prior to addition it is possible to use a jacketed dropping funnel, with the
cooling mixture (e.g. dry-ice/acetone) placed in the jacket. This is a
common set-up encountered for low temperature reactions on larger scales
(Fig. 11.2).
Characterization
12.1 Introduction
This chapter deals with the type of physical data that are required for the
proper characterization of the purified product. No theory is discussed as
this is wen covered in other sources and, given good data, it is often
possible to find a colleague (for example) who will help out if you are
unable to interpret a spectrum. With inadequate data it will be difficult to be
certain of the structure and purity of your product, and it will certainly be
more difficult to interest the colleague referred to above!
It is important to acquire as much information as possible on your
product. It might be 'obvious' from the nmr that the structure is what you
think that it should be, but it is still necessary to record (at least) the ir and
mass spectra. These might simply confirm the nmr data, or they might raise
other structural possibilities.
The full set of routine physical data which could and, ideally, should be
obtained on a pure compound is as follows; ir, uv, high field nmr (IH and
13C), and low and high resolution mass spectra, m.p. or b.p.,
microanalysis (for a new compound). If the compound is optically active
then the optical rotation must be measured. Only mass spectroscopy and
microanalysis from this list are destructive techniques, but modern
techniques mean that only a small amount of material need be 'sacrificed'.
Some general points concerning these techniques and the sampie
requirements are given below.
202 Characterization
12.2 Nmr
Nowadays the IH nmr spectrum is often the first measurement taken. The
size of sample required depends on the type of spectrometer used. A
continuous wave machine operating at 90 MHz will need at least lOmg of a
normal organic compound, probably more. Pulsed Fourier transform
spectrometers require less, 5mg being anormal amount, and good spectra
can be obtained using much smaller quantities. Most spectra are measured
in deuteriochloroform (CDCI3) although other solvents will be required
from time to time. A typical solvent volume would be ca. 0.4-0.5ml in a
5mm tube. Routine measurement of the 13C nmr spectrum usually requires
more sample (25-50mg) but good spectra can be obtained on less, it simply
takes more time.
The solution used must be free ofparamagnetic metal ions (it usually is)
and particles (it usually is not). Filtration through a small wad of cotton
wool forced into a Pasteur pipette will usually remove sufficient particles to
allow a good spectrum to be obtained.
The high field 1H nmr spectrum will show up impurities containing
protons. Given that the compound has been purified, the most common
impurity peaks observed in the 1H nmr spectrum are those from the last
solvent used (for example, solvents used in crystallization or
chromatography). This should be avoided, and it is always possible unless
the boiling point of your product is close to the solvent (which it should not
be) or your product is a crystalline solvate (not that common). Thorough
exposure to high vacuum should suffice but some very viscous oils and
gums will 'hold on' to solvent due to the very slow rate of diffusion. If
warming in high vacuum fails to remove all solvent, and a 'clean' 1H nmr
spectrum cannot be obtained, then dissolve the sampIe in a small amount of
CDCl3 and evaporate. Repeat once or twice and most of the residual solvent
should be CDC13 rather than (say) ethyl acetate. This will improve matters,
but do not forget that your sampIe will still be impure on evaporation as it
will contain residual CDC13.
12.3 Ir
The sampIe again needs to be free of impurities and solvents for infrared
spectroscopy. There are various methods for sampIe preparation and which
Characterization 203
you choose will depend largely on the type of compound. The amount
required is no more than a few milligrams. For a liquid sample the spectrum
can be obtained using a thin film obtained by compressing a small drop
between sodium chloride plates, or as a solution (usually in chloroform)
using solution cells. The spectra of solids can be recorded either as mulls
with a hydrocarbon ('Nujol', for example), or by mixing with KBr and
compressing to form a thin disko Which you use will depend upon the
facilities available, and often on the usual working practice of your
department.
12.4 Uv
Ultraviolet spectroscopy is only of use if your compound has a characteristic
chromophore. There is little point trying to measure weak bands which will
provide no information. However, it is of considerable value in several
areas of research; for example, natural product isolation, heteroaromatic
chernistry, porphyrin and related chemistry, and in the study of dyestuffs.
The amount of material required is usually very small (fractions of a
milligram) since the extinction coefficients are usually large. The sampie
must be as pure as possible and is dissolved in the solvent of choice (usually
spectroscopically pure ethanol). The concentration must be known
accurately before extinction coefficients can be calculated, and will vary
depending upon the type of chromophore. An estimate of the concentration
to use can be made if the extinction coefficients of compounds sirnilar to that
being studied are available. If this data is not available make up a solution
accurately and dilute it (accurately!) until a reasonable spectrum is obtained.
reasonably pure. If you are unable to obtain good microanalytical data the
accurate mass measurement may provide an acceptable alternative.
If your product is, or should be, optically active then the specific rotation
will need to be measured and recorded. The precise value of this property is
dependent on the wavelength of the light used, solvent, concentration, and
temperature. Moreover, great care should be taken to exclude any by-
products from the reaction since, although these might be present in small
quantities, they might have very large rotations and make your
measurements quite misleading. Clearly then, it is important to be sure of
the purity of your product, and to make up the solution carefully and
accurately. If you are unsure then make a measurement using a known
compound before you try to measure the rotation of your product (assuming
that the specific rotation of your product is not known). Usually you will
Characterization 205
use the sodium-D line and measure at ambient temperature, but be sure to
reeord the eoncentration, solvent, wavelength, and temperature along with
the actual value of the measured specific rotation. Oceasionally optieal
rotatory dispersion andlor cireular diehroism spectra will be required.
These measurements will usually made by specialists and the specifie
requirements for your particular type of compound are best discussed with
them.
12.8 Microanalysis
There are several schools of thought on this topic. Some maintain that all
new eompounds must be analysed, whereas others say that, with the
modem array of physical methods (high resolution nmr speetroseopy and
high resolution soft ionization mass speetroscopy, for example) the need for
eombustion analysis no longer exists. Many follow amiddie course and use
microanalysis for erystalline compounds which are available in sufficient
quantity, and high resolution mass spectrometrie measurements in all other
cases. The course you take will depend upon eircumstances (the
requirements of your supervisor or the department, for example).
1t goes without saying that the compound must be homogeneous and
free from dust, inorganics, ete. For solids, careful reerystallization using
pure, filtered solvents followed by equally eareful filtration will usually
suffice. For oils, distillation folllowed by sealing in an ampoule will
provide aeeeptable sampies. All the glassware involved must be clean and
dry including the ampoule. Solids must be dried in high vaeuum in a drying
pistol to remove traees of solvent, and submitted in clean, dry vials. 1t is
possible to obtain reasonable microanalytical data on oils wh ich cannot be
distilled by careful column chromatography using pure, distilled solvents
followed by thorough pumping down in high vaeuum.
this). Spectra are best kept in c1early labelled folders or binders of some
description, preferably ones which allow for removable attachment of the
spectra. The other data should be recorded in the laboratory notebook along
with the experimental write up. If data sheets are used then all the data
should be recorded on these as they are measured.
CHAPTER 13
abstract looks promising all that remains is to locate the journal and consult
the paper.
An Index Guide is published every eighteen months and contains
invaluable information on the use of the indexes and the system of
nomenclature used in CA. 1t is essential reading for serious users of
Chemical Abstracts. FinaUy, the Ring Systems Handbook and its
predecessors, the Ring Index and the Parent Compound Handbook, contain
information on ring and cage systems and gives the names under which ring
systems can be found in the Chemical Substance Index.
The great strengths of CA are that it provides comprehensive literature
coverage, and it has extensive indexes. However, the coverage of the
literature in the early years was not so rigorous, and Beilstein provides more
thorough coverage of the pre-1949 literature.
13.2.2 Beilstein
Beilstein's Handbuch der Organischen Chemie, or Beilstein for short, is a
huge (> 300 volumes) reference work which contains physical and chemical
data for over one and a half million compounds. The compounds are
organized according to a unique classification system and each volume
contains a subject (actually compound) index and a formula index.
Comprehensive literature coverage is attempted, so Beilstein contains
essentially all the compounds of a given class, which were prepared from
the beginning of organic chemistry to the date of publication of the most
recent volume covering that class of compounds. A considerable amount of
critically reviewed information, with references to the primary literature, is
provided for each compound. This data includes the molecular structure,
natural occurrence, methods of preparation, physical properties including
references to papers containing spectral data, and chemical properties.
The Handbook consists of the original series of 27 volumes (the
Hauptwork, H) and a number of supplementary series (Ergansingbande, EI,
EIl, etc.). Work on the supplements is constantly in progress. Coverage of
the literature up to 1959 has almost been completed, and some progress has
been made in bringing the coverage up to 1979. Cumulative indexes are
only available for the pre-1930 literature. 1t is relatively easy to find data for
any compound reported prior to 1930 by checking the cumulative formula
index (Volume 29 in three subvolumes) which will give the volume and
The Chemical Literature 211
page numbers for the entries in H, EI and EIl. Compounds of the same
c1ass will appear in the same volume of each series (although some of the
volumes are now divided into several subvolumes). Thus if a compound is
located using the cumulative index it is a simple matter to locate references in
the later series, using the volume indexes for the same volume or using the
Beilstein System Number. If a compound is not contained in the cumulative
index it is best to try to identify whieh volume it should be contained in,
using the c1assification system. A free booklet explaining how to use the
Beilstein system is available from the publishers (Springer-Verlag).
Beilstein is the best and most comprehensive source of data for organic
compounds prepared before 1930 and it is not partieularly difficult to use.
Its major weakness is the lack of data and cumulative indexes for more
modern work.
appeared. He would consult the annual indexes ofthe SCI from 1980
onwarc!s. Each index contains an alphabetic list ofjirst authors names
and under S. Smith's name would be a chronological list of his/her
publications. Under the entry for the paper of interest to our researcher
there would be a list of papers, published during the period covered by
that index, which cited it. It is reasonable to assume that any workers
who followed up on the results in the Smith paper would have cited it
in their own publications. Thus the list of papers which cited the
original should include most of the work subsequently carried out in
that area. Hence if you fmd an important paper you can use SCI to get
a list of all the papers which subsequently referred to it. The drawback
is that many of the references you find will not be relevant to your
interest and there is no way of knowing which are relevant except by
consulting the Source Index, which gives the titles of the papers, or by
consulting the papers themselves.
The citation index is an extremely useful tool and we recommend that
you carry out a search for every key paper you come across. You can also
use it to find out who is referring to your own work.
abstract numbers. The keyword list makes it much easier to identify which
abstracts are most likely to contain the information you require. With a little
experience you may prefer to use the Chemical Substance Index directly but
the complexities of nomenc1ature and indexing preclude any further
discussion here. Note that prior to the 9th Collective Index, the Chemical
Substance and General Subject Indexes were combined in the Subject
Index.
Inc1ude the bibliographie details and an abstract of the key results in the
paper. The pile of eards is not of much use unless it is properly filed.
Many filing systems, eaeh with its own strengths and weaknesses, can be
conceived but one possibility deserves special mention, at least as a starting
point. Annual Reports in Organic Synthesis (Seetion 13.3.4) is essentially
a eompilation of index eards in book form, and the systematic way in which
the material is organized could serve as a useful model for your system. As
time progresses you can modify this system to adapt it to your own interests
and requirements.
CHAPTER 14
Special Procedures
14.1 Introduction
This section deals with some of the specialized procedures which might be
encountered. All the topics covered here have one thing in common, namely
that the particular type of apparatus used will vary from one laboratory to
another and accordingly detailed instructions will not be given here for any
one piece of apparatus. Representative systems are shown, and common
operating procedures discussed. Often there will be someone in the
department who is responsible for, or has particular expertise in, one of
these techniques. If this is the case then always consult this person before
you intend attempting the reaction.
flask (see Section 6.3 for details of gas burettes). A typical arrangement is
shown in Fig. 14.1 and the operation is simple in principle. A volume of
hydrogen (more than the theoretical amount whenever possible) is
transferred to the burette, with the reaction vessel containing the solvent,
catalyst, and reactant, and the initial volume noted. The reaction is then
agitated (stirred or shaken) and the uptake of hydrogen monitored using the
burette. The detailed operation will depend upon the precise equipment
which is used, but a written step-by-step procedure should be available. It
is important to follow the procedure closely, and to consider the effect of
opening any tap before doing so.
to to to to
reaction gas bUTette manometer source
vessel ofhydrogen
to
waterpump
(or other source
ofvacuum)
The above procedure is sometimes inconvenient for small sc ale work (it
depends upon the type of hydrogenation system available), and we often
resort to the use of a small balloon of hydrogen connected to the reaction via
a three-way tap, which is also connected to a manifold. This technique is
illustrated in Fig. 14.2.
Fill the balloon with hydrogen (several times to remove air) and connect
it to the fiasko Connect the three-way tap to the manifold, open to vacuum
(carefully) and then to the inert gas (Fig. 14.2a). Several cyc1es will be
required. Then turn the tap so as to isolate the system from the manifold,
and allow hydrogen to enter the fIask (Fig. 14.2b). The reaction can then be
monitored in the usual way. When the reaction is over, vent the excess
hydrogen from the balloon (sa/ely) to a fume cupboard (Fig. 14.2c).
Residual hydrogen can be removed from the system by the use of several
cyc1es of evacuation followed by admission of inert gas using the manifold
as above (Fig. 14.2a). All precautions re/erred to in the use 0/ the
atmospheric hydrogenator must be observed here 0/ course.
t :::··:~
\:iI·
i I
14.3 Photolysis
coolir1g
L
.......-""\,·otc, OLll
I-t-H---i--l'
solution or
substrate
....
..
.. ..
glas. f';t
immersion weIl (not Pyrex). Most preparative reactions use the much
higher power (lOO-400W) medium pressure lamps, as these emit their
radiation over a much wider range (mainly at - 365nm with other bands at
both shorter and Ion ger wavelength). Occasionally a filter will be required
and it is important that the correct one is used (this will be specified in the
preparation being followed).
Reactions are usually run at fairly high dilution (up to - O.05M) and the
solvent should be pure and chosen with care. It must not decompose under
ultraviolet irradiation and should not absorb at the wavelength being used
for the reaction. Work up often involves no more than evaporation of the
solvent followed by purification.
14.4 Ozonolysis
_--1~~ ~
-
to vacuwn
flask for
--- ---- guage
_ .. 0 ____ - . - - _ • • • _ ..
substrate ..................
:::::::,,":::::::::
.- ........
........ -._-.-.
_.. ----,,--
__ 0 .... ___ • • - - - - - - -
cold batb
...... __ ....................... _-
. . . . . . . . . . . _ ...... _ ................ 0 __ -
----+ [0 bubblcr
low tcmperalllrc----+
thermomeu~r
cooling halh
Figure 14.5
The apparatus is set up as in Fig. 14.5 and a rapid flow of ammonia is
used to flush out the system. A small volume of acetone (or ethanol) is
poured into the condenser, and solid carbon dioxide pellets are added (very
slowly atfirst) until the condenser is nearly full. The ammonia will begin to
condense, and when the required volume is obtained, the ammonia flow is
shut off and the gas inlet replaced by a septum or stopper. If undried
impure ammonia will suffice, and often it will, then the reaction is carried
out as normal. A cooling bath can be added if a long reaction time is
anticipated, or if a temperature below -33°C is required (see Chapter 8).
If dry liquid ammonia is needed this is usually obtained by distillation
off sodium. The appropriate volume of ammonia is condensed as above and
small pieces of sodium are added to produce a blue solution. The ammonia
can then be distilled using anormal distillation apparatus (Chapter 9) except
that the receiver (usually the reaction flask) is cooled in asolid carbon
226 Special Proccdurcs
'Trouble Shooting':
What to do when things do not work
is to allow more time at reflux before collecting the solvent, if this does not
work the drying agents might need recharging.
Another source of water could be the inert gas which you are using;
make sure that the drying agent (if used) used is still working renew it if
necessary. Failing this, try the reaction under argon, which usually contains
much less moisture than nitrogen. Check the manifold and renew any
suspect tubing.
With pure starting materials, and anhydrous solvent and atmosphere,
the reagent(s) must be suspected. If it is possible to purify them, do so, and
make sure that you are handling them correct1y (see Chapter 5). Itwill not
be practical to purify some reagents, for example alkyllithiums, and in this
case the quality should be checked by titration where possible. It is unwise
to purchase areagent (from any source) and to take for granted the quoted
molarity and purity; even the most reputable suppliers are fallible and
occasionally make mistakes.
If starting material, reagents, solvent, and inert gas are all as they
should be then it might be you! You might be inadvertently carrying out the
re action in such a way that it will not work. For example, is the temperature
correct, is the concentration of reactant and reagent correct, is too much or
too little time being allowed at a particular stage? There are many
possibilities. To test this it is advisable to carry out the same reaction but
use a substrate which is known from the literature to react properly. If this
is successful, and your desired reaction is not then you have found a
reaction which does not work on your substrate, and alternative conditions
(different metal ions, different Lewis acid etc.) might be required. Above
all, if the reaction is an important one, do not give up; perhaps you could
take the opportunity to develop a new re agent or procedure which will
work!
CHAPTER 16
Example Reactions
Li
~CI
inert gas
~
septum -
condenser-
reaction flask
ultrasonic
- bath
Figure 16.1
~Li
Red-<>range
A dry, 2ml, round-bottomed flask or small Pyrex test tube fitted with a
magnetic stirrer bar and septum is flushed with argon, and then placed under
a positive pressure of argon (Fig. 16.2).
Example Reactions 231
inert gas
~
septum - t:t1 (
GoQ):====~l 0
reaction mixture- 7:
e
( )
Figure 16.2
1,3-Diphenyl-2-propanone p-toluenesulphonylhydrazone (113 mg,
0.5mmol)2 is placed in the flask, and dry tetrahydrofuran (2ml) added via
syringe. The reaction mixture is stirred rapidly to dissolve the hydrazone,
and the n-butyllithium solution is added dropwise to the colourless solution
until the orange-red end-point is observed. At this point the volume of n-
butyllithium solution added is noted, and from this the molarity of the
solution is calculated using the following equation:
o o
xV'o
OH
~
1) LDA
2) Benzaldehyde
A dry, 100ml, round-bottomed flask fitted with septum and magnetic stirrer
2. For a discussion of the varoious titration methods see, J. Suffert, J. Org. Chern.,
1989, 54, 509.
3. H.O. House, D.S. Crumrine, AY. Teranishi, and H.D. OImstead, J. Am. Chern.
Soc., 1973, 95, 3310.
232 Example Reactions
septum
low tempcrature
lbcnnomctcr -
/
~~I!!}---.5 - argon
lagged balb -
~
~
Figure 16.3
bar is flushed with argon, then placed under a positive pressure of argon.
Anhydrous diisopropylamine (2.9ml, 20.8mmol) is added to the flask via
syringe, followed by dry diethyl ether (20mi). The reaction mixture is
cooled to _78°C using a dry ice-acetone cooling bath (Fig. 16.3). n-
butyllithium (11.6ml of a 1.8M solution in hexanes, 20.8mmol) added
dropwise via syringe to the stirred mixture. After addition of the n-butyl-
lithium is complete the reaction mixture is stirred for lOmin at -78°C
allowing complete formation of lithium diisopropylamide then a solution of
3,3-dimethylbutan-2-one (2Aml, 19.2mmol) in dry diethyl ether (5ml) is
added dropwise via syringe to the reaction mixture. After stirring at -78°C
for 10min, benzaldehyde (2Ag, 19.2mmol) is added and the mixture stirred
for a further 30min. The reaction is quenched by careful addition of 1M
hydrochloric acid (50ml), and the resulting mixture allowed to warm to
room temperature before being transferred to a 250ml separating funnel.
The mixture is extracted with diethyl ether (3x50ml) and the combined
etheral extracts washed with saturated aqueous sodium chloride solution
(50ml). The organic extracts are then dried over magnesium sulphate, and
the solvent removed on a rotary evaporator to give the crude product as an
oil. Kugelrohr distillation gives 5-hydroxy-2,2-dimethyl-5-phenylpentan-3-
Example Reactions 233
2. NaOH
nitrogen
mcchanical stilTCF--
dropping funncl
Ihermomctef-
rcaction mixture
Figure 16.4
4. R.W. Lang and H.-I. Hansen, Org. Synth .• 1984, 62, 202.
234 Example Reactions
is stirred at room temperature for 2h, then the colourless phosphonium salt
precipitate is filtered off, and washed first with cold toluene (250ml) and
then petroleum ether (l50ml). The cmde phosphonium salt is dissolved in
water (2 litres) , placed in aseparatory funnel, and further organic impurities
removed by washing with diethyl ether (2x400ml). The aqueous solution is
transferred to a 5 litre beaker, and 2% alcoholic phenolphthalein (10 drops)
added. The solution is cooled in an ice-water bath and stirred vigorously by
means of a glass rod as 2N aqueous sodium hydroxide is added slowly until
the pink end-point is reached. At this stage as much water as possible is
decanted from the precipitated phosphorane, then ethyl acetate (1.5litres) is
added. The resulting two phase mixture is transferred to a separating funnel
and the aqueous layer removed. The ethyl acetate solution is dried over
magnesium sulphate and evaporated on a rotary evaporator, to give ethyl
(triphenylphosphoranylidene)acetate (l22g, 88%) as a cream solid, m.p.
124-126°C.
~OH 1) (EtOhCCH3, H+
~ 2) KOH
~~f.LJ-nitrogcn
condenser-
Dean-Stark
trap -
slirrer-hotplate- ~
~
Figure 16.5
is heated under reflux for 45min under nitrogen (Fig. 16.6). After the
alkaline solution has cooled to room temperature, water (3Oml) is added, the
mixture washed with diethyl ether (20ml) and then acidified with
concentrated hydrochloric acid. The acidic aqueous solution is extracted
with diethyl ether (3x20ml), and the extracts dried over magnesium
sulphate. Filtration and removal of the solvent on a rotary evaporator then
gives the crude product. Flash chromatography on silica gel (20% diethyl
ether - 80% petroleum ether) gives the purified 3-phenyl-4-pentenoic acid as
a pale yellow solid which can be recrystallized from petroleum ether (m.p.
44-46°C).
236 Example Reactions
~~EJ - nitrogen
eondenser -
slirrer-hotplate -
~
~
Figure 16.6
H2, Ethanol
Pt02
balloon filled
with hydrogen-
reaction flask
A
'\;:;:;:I
Figure 16.7
positive pressure of hydrogen maintained by the balloon.
The reaction mixture is stirred at room temperature for 1h, the hydrogen
balloon removed, and the mixture filtered through a pad of silica gel (3g) to
remove the catalyst. The solvent is the removed on a rotary evaporator to
give a white solid which can be crystallized from about 2ml of boiling
water, to give succinic acid (2g, 84%), m.p. 187-189°C.
CHAPTER 17
Safety
It has been emphasized already that you should be familiar with the
regulations and codes of practice pertaining in your laboratory. We will not
discuss safety legislation here but some fundamental rules should be
stressed. Never work alone in a laboratory. Always wear suitable safety
spectacles and a non-flammable lab coat, and use other protection such as
gloves, face masks, or safety shields if there is a particular hazard. Never
eat, drink or smoke in a laboratory. Work at a safe steady pace, and keep
your bench and your lab clean and tidy. Familiarity breeds contempt, do not
allow yourself to get careless with everyday dan gers such as solvent
flammability. Familiarize yourself with the location and operation of the
safety equipment in your laboratory.
As regards specific hazards the chief rule is to carry out an assessment
of the dan gers involved before using an unfamiliar chemical or piece of
apparatus. Some of the commonest hazards are described in the next
section. Once you are aware of the possible dangers take all the necessary
precautions, and ensure that you know what to do if an accident does occur.
Store your chemicals in clearly labelled containers, and abide by the
regulations concerning storage of solvents and other hazardous materials.
Dispose of waste chemicals safely, according to the approved procedures
for your laboratory. Never pour organic compounds down the sink.
Source Hazard
Glassware Danger of cuts, leaks of harmful compounds
Solvents Most are extreme1y flammable
Benzene, halogenated solvents are toxic
Vacuum apparatus May implode violently
Pressure apparatus May explode violently
Gas cylinders May leak harmful gases or discharge violently
(Chapter6)
Strong acids Extremely corrosive
React violently with water, bases
May produce harmful vapours
Strong bases Extremely corrosive
React violently with acids, protic solvents
Strong oxidizing agents React violently with easily oxidizable compounds
such as organic solvents
Alkali metals React violently with water, protic solvents and
chlorinated solvents
Strong alkylating agents Extremely toxic
Table 17.2
Explosion hazards
Table 17.3
Toxic and carcinogenic compounds
17.5 Bibliography
Acetic acid 118 17 6.19 1.049 2.08 s, 10-13 br s var 20.7, 177.6 Acetic anhydride Acetic anhydride 10
Acetone 56 -94 20.7 0.790 2.13 s 30.7, 206.5 3Asieve 3A sieve 1000
Acetonitrile 82 -46 36.2 0.777 1.97 s 1.7, 116.2 Potassium carbonate Phosphorus pentoxide 40
Benzene 80 5.5 2.28 0.879 7.37 s 128.3 Not necessary Calcium hydride 10
t-Butanol 82 25 3.49 0.850 1.24 s, 1.35 br s var 31.2, 69.2 Calcium hydride Calcium hydride
Carbon tetrachloride 76 -23 2.23 1.460 - 96.2 Alumina Phosphorus pentoxide 10
Chlorobenzene 132 -46 5.62 1.106 7.28 brm 126, 129, 130, 134 Not necessary Calci um hydride 75
Chloroform 62 -63 4.70 1.480 7.24 s 77.0 Alumina Phosphorus pentoxide 10
Dichloroethane 83 -35 10.4 1.235 3.71 s 44.4 Not necessary Calcium hydride 10
Dichloromethane 40 -97 8.9 1.325 5.28 s 53.4 Not necessary Calcium hydride 100
Diethyl ether 35 -116 4.34 0.714 1.18 t, 3.45 q 15.2, 65.8 Calcium chloride; Na SodiunVbenzophenone 400
Dimethoxyethane 83 -58 0.850 3.36 s, 3.51 s 59.0, 71.8 Calcium chloride; Na SodiunVbenzophenone g~
Dimethylformarnide 152 -61 36.7 0.945 2.81 s, 2.89 s, 7.94 s 31.4, 36.4, 162.4 Calcium hydride Phosphous pentoxide 10 ~
Dimethyl sulphoxide 189 18 49.0 1.101 2.62 s 40.6 Distillation 4A sieve
Dioxan 102 12 2.21 1.034 3.66 s 67.0 Calcium chloride: Na SodiunVbenzophenone 50
Ethanol 78 -114 24.3 0.785 1.18 t, 2.05 br s, 3.65q 18.3, 58.2 Magnesium 3A sieve 1000
Ethyl acetate 77 -84 6.02 0.900 1.22 t, 2.01 s, 4.09 q 14.1,20.9,60.3,171.0 Potassium carbonate 4A sieve 400
HMPA 235 7 1.030 2.59 d 36.8 Calcium hydride Calcium hydride
Methanol 64 -97 32.6 0.791 1.94 br s var, 3.42 s 50.5 Magnesium 3A sieve 200
Nitromethane 101 -28 38.6 1.137 4.33 s 62.4 Calcium chloride 4A sieve 100
Pyridine 116 -42 12.3 0.982 7.24 m, 7.63 m, 8.58 m 123.6, 135.8, 149.8 Calcium hydride Calcium hydride 5
Tetrahydrofuran 66 -65 18.5 0.805 1.82 m, 3.72 m 25.6, 67.9 Calcium chloride; Na SodiunVbenzophenone 200
Toluene 111 -95 2.38 0.867 2.32 s, 7.17 br s 21,125,128,129,138 Not necessary Calcium hydride ~OO
------- - -
U+ -tBu - --
>50
*These figures are very approximate. A pKa difference of >4 between base and reagent will cause complete deprotonation.
~
U1
Appendix 4 Lewis acids
~
Lewis acid Compatlble solvents Comments 0\
Mercuric chloride Many solvents Weak, useful for cleaving C-S bonds
Sil ver chloride Many solvents Moderate, used to generate carbonium ions
Sil ver triflate Many solvents Moderate, used to generate carbonium ions
Trimethylsilyl iodide Hydrocarbons, halogenated, MeCN Strong, used to cleave ethers, acetals, esters
Trimethylsilyl triflate Hydrocarbons, halogenated Strong, used with silylated reagents
N
VI
tv
Ut
Appendix 6 Common oxidizing reagents tv
sulphide~ sulphone
I
Ru04 Carbon tetrachloride/ RT cleaves a1kenes ~
acetonitrile/water carboxylic acids
f
E
N
VI
VI
Index
Accidents 241 safety 242
Aceticacid 34 synthetic chemistry 216
Acetic anhydride Boric anhydride, drying agent 30,34
drying agent 34 Boron trifluoride etherate 47
purification 48 lßutanol, drying 35
Acetone 33 nButyllithium
reaction with chloroform 35 preparation 229
Acetonitrile 34 titration 230
Acetyl chloride 48 use 231
Acetylene, preparation 86 Calcium chloride, drying agent 31,34,35
Air sensitive reagents Calcium hydride, drying agent 30-35
cannulation 51 Calcium sulphate, drying agent 31
setting up reactions 97 Cannula, cannulation 51,56,103
syringing 62,100 Capillary gc columns 122
Aldol reaction 231 Carbon dioxide, preparation 86-87
Alumina, drying agent 30,34 Carbon disulphide, drying 35
Ammonia 34,224 Carbon monoxide, preparation 87
Azeotrope 133 Carbon tetrachloride, drying 35
Azides, hazards 36 Celite 145
a,a'Azobis(isobuyronitrile), (AIBN) 49 Chemical abstracts 213-217
Balances 15 CA Selects 217
Balloons CAS OnIine 216,217
attachment 108 CAS REACT 213
for inert reactions 106 Chlorine, preparation 87
for hydrogenation 236 Chlorobenzene, drying 35
Barium oxide 30 Chloroform, drying 35
Bases, table 245 Chromatographic equipment 19
Beilstein 210,214 Chromatography
Benchkit 14,20 drycolumn 176
Benzene, drying 34 flash chromatography 27,166
Benzyl bromide 47 195,235
Bibby clips 97 gc 121-124
Boiling point 204 hplc 117,185
Books mplc 178
functional group chemistry 215 tlc 110
reference handbooks 215,216 Circular dichroism 205
258 Index
Records flammability 28
of experimental data 8-11 for chromatography 29,169-170
of laboratory work 3-8 for crystallization 147
Refractive index detector 182,185 for spectroscopy 29
Retention time inert atmosphere distillation 41,46,
gc 122 157
hplc 118 peroxide contamination 29
Safety 238-242 properties 243
Sand bath 132 purification 28,29
Science Citation Index 211 stills 39-42
Sealed tuhe 128-129,194 toxicity 28
Septum 50-51 Solvent reservoir, for flash 167,168
Shaker 139 Solvent traps, for vacuum system 91
Sieves, molecular 31-32 Sonication 139,194,229
Silica Soxhlet extractor 49,134
for flash chromatography 166 Spaghetti manifold 108
quantity for flash 170 Spaltrohr columns 159,162
recycling 175-176 Specific rotation 204
safety 169 Spinning band columns 159
Silica/sample ratio, for flash 170 Stills
Sodium 33 one-piece 18,19,157,158
Sodium/benzophenone ketyl 33 solvent 39-42
Sodium chloride plates, ir 203 Stirring 135-138,190,199
Sodium-D line 205 Strong bases 142,246
Sodium/potassium a1loy 33 Structure determination 201-205
Sodium sulphate, drying agent 33 Sublimation apparatus 164,165
Solids Syringe fittings 59
air sensitive 66,69,154 Syringes 57-65
crystallization 61,147,154 care and c1eaning 60
handling and weighing 69 glass 58
purification and drying 48 gas-tight 59,80
Solvent handling 61
analytical grade 29 micro 57
anhydrous 28 plastic, disposable 59
distillation 39-42 preparation for use 61
drying 29,33 transfer of liquid under inert
264 Index