You are on page 1/ 90

A3Plus

A5
A6Plus
A7Plus
series 2019
Installation manual
code 97070068 - rev. 05 - 16/06/2021

After Sales Service


Plant Via Bicocca 14/c 40026 Imola (BO) Italy
Tel. +39 0542 653866- Fax +39 0542 653596
E-mail: [email protected]
A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

CONTENTS

INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT................................................... 3


1 - HANDLING AND SHIPMENT OF THE DENTAL CHAIR ......................................................................... 3
1.1 - GENERAL PROCEDURE...................................................................................................................3
1.2 - INSTRUCTIONS FOR USE.................................................................................................................3
1.3 - DENTAL CHAIR LIFT RODS..............................................................................................................3
2 - MOUNTING KIT........................................................................................................................................... 4
2.1 - MOUNTING KIT (DENTAL UNIT)....................................................................................................... 4
2.2 - MOUNTING KIT (SINGLE CART)...................................................................................................... 5
3 - REMOVING THE DENTAL CHAIR CARTERS.......................................................................................... 6
4 - PLACING THE DENTAL CHAIR................................................................................................................ 6
5 - INSTALLING THE FLOOR BOX................................................................................................................. 7
6 - FIXING THE FOOT CONTROL................................................................................................................... 8
6.1 - FIXING THE FOOT CONTROL (DENTAL UNIT).............................................................................. 8
6.2 - FIXING THE FOOT CONTROL (SURGICAL CART)........................................................................ 9
7 - BACKREST INSTALLATION (OPTIONAL - NOT IN A3PLUS)............................................................. 10
8 - INSTALLING THE DENTAL UNIT............................................................................................................ 12
9 - DENTAL UNIT / DENTAL CHAIR LEVELLING....................................................................................... 18
10 - DENTAL CHAIR ELECTRICAL CONNECTION.................................................................................... 19
11 - ANCHORING HOLES.............................................................................................................................. 20
11.1 - DENTAL CHAIR ANCHORING HOLES....................................................................................... 20
11.2 - FLOOR BOX CASTING ANCHORING HOLES (SINGLE CART).............................................. 24
12 - ATTACHING THE TUBES / WIRING TO THE DENTAL CHAIR.......................................................... 25
13 - INSTALLING THE DENTAL CHAIR CARTERS.................................................................................... 27
14 - ELECTRICAL CONNECTIONS.............................................................................................................. 28
15 - AIR CONNECTIONS................................................................................................................................ 30
16 - SET THE LIMIT SWITCHES FOR DENTAL CHAIR MOVEMENTS (AUTOMATIC SETUP).............33
17 - ASSEMBLING THE EMERGENCY PATIENT CHAIR BASE CONTROL........................................... 34
18 - INSTALLING THE SUCTION TUBES.................................................................................................... 36
19 - TRAY-HOLDING TABLE ON ASSISTANT’S BOARD.......................................................................... 36
20 - INSTALLING THE WATER SUPPLIERS TO THE CUP AND THE BOWL / FILTER AND FILTER
COVER...................................................................................................................................................... 36
21 - ASSEMBLING THE 6TH INSTRUMENT CORD SUPPORT (CONTINENTAL).................................. 37
22 - ASSEMBLING THE TRAY HOLDER TABLE ON THE DENTIST’S MODULE................................... 38
23 - ASSEMBLING THE MANUAL HEADREST.......................................................................................... 40
24 - PNEUMATIC HEADREST INSTALLATION........................................................................................... 41
25 - OPERATING LAMP ................................................................................................................................ 43
26 - ASSEMBLING THE RIGHT ARMREST................................................................................................. 43

INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT 1


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

27 - DENTAL CHAIR PADDING INSTALLATION ........................................................................................ 44


27.1 - INSTALLING THE HEADREST PADDING....................................................................................44
27.2 - BACKREST PADDING INSTALLATION.......................................................................................44
27.3 - INSTALLING THE SEAT PADDING...............................................................................................49
28 - CHECK DURING THE ASSEMBLING AND/OR PERIODIC MAINTENANCE (ANNUAL)................ 50
29 - INSTALLING THE HYDROGROUP DOOR........................................................................................... 52
30 - INSTALLING THE COVER OF THE FLOOR BOX............................................................................... 54
31 - LEVELING AND ADJUSTING THE ARM’S CLUTCH.......................................................................... 56
31.1 - ADJUSTING THE FIXED ARM’S CLUTCH...................................................................................56
31.2 - ADJUSTING THE PANTOGRAPH ARM’S CLUTCH.................................................................... 57
31.3 - CALIBRATING THE PANTOGRAPH ARM SPRING / LEVELING THE INSTRUMENT
BOARD.............................................................................................................................................58
31.4 - ADJUSTING THE DENTIST’S INSTRUMENT BOARD / DENTAL ASSISTANT’S
INSTRUMENT BOARD....................................................................................................................59
32 - ADJUSTING THE OSCILLATING ARM SPRING................................................................................. 60
33 - CLEANING THE TURBINE RETURN AIR FILTER............................................................................... 61
34 - INSTRUMENT’S TABLE ARMS EXTRA-ROTATION........................................................................... 62
35 - SET THE LIMIT SWITCHES FOR DENTAL CHAIR MOVEMENTS (MANUAL SETUP)...................65
36 - CONNECTION TO CLOUD..................................................................................................................... 67
37 - CHECK OF THE CORRECT CONNECTION TO THE CLOUD............................................................ 70

2 INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT

CAUTION:
Once the dental unit has been installed in the dental surgery and the water has filled the ducting, the apparatus is expo-
sed to the risk of contamination by aquatic microorganisms in the mains water supply; these can cause biofilm to form
in the ducting. For this reason it is recommended that you only install the dental unit when you are ready to use it on a
daily basis and that you carry out the decontamination procedures as per the schedules provided from the very first day
of installation.

1 - HANDLING AND SHIPMENT OF THE DENTAL CHAIR

WARNING! Wear the proper protection:

1.1 - General procedure


An ID code and warning symbol are stamped on each rod.
The rods have been treated with rust-preventative and must only be used to lift and handle the dental chairs indicated
in the instructions manual.
The rods must not be used for any other purpose than that specified in the instructions.
The rods must be kept in a suitable sheltered place to prevent the material from rusting or deteriorating.
The rods must be used together to avoid placing too great a load on only two of them causing them to bend or be
damaged.

1.2 - Instructions for use


- do not change the set-up of the dental chair (it should remain as it was in its packing) or apply any other weight.
- when the chair is being handled make sure the load remains perfectly level to prevent all the weight from bearing down
on two levers and just one operator.
- each time the rods are used make sure that the tooth of the rod is correctly inserted in the rectangular slot provided.
- periodically check the condition of the rods.
- these rods can only be used for handling ANTHOS dental chairs.
- the overall weight of the dental chair is 115 Kg whereas the dental unit weighs 90 Kg.

1.3 - Dental chair lift rods

99918926 99918926

99918932

INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT 3


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

2 - MOUNTING KIT
2.1 - MOUNTING KIT (DENTAL UNIT)

4 INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

2.2 - MOUNTING KIT (SINGLE CART)

INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT 5


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

3 - REMOVING THE DENTAL CHAIR CARTERS

4 - PLACING THE DENTAL CHAIR


- Grip the metal parts of the patient chair, position it in line with the electro-pneumatic connections and position according
to the surgery layout.

WARNING! the plastic parts are not strong enough to be used as hand-holds when moving the chair.

6 INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

5 - INSTALLING THE FLOOR BOX

INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT 7


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

6 - FIXING THE FOOT CONTROL


6.1 - FIXING THE FOOT CONTROL (DENTAL UNIT)

NOTE: Fully charge the foot control battery before


using it for the first time.

NOTE: When the dental unit is switched on wait


about 20 sec. and act on the control lever (B) to acti-
vate operation of the wireless foot control.

8 INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

6.2 - FIXING THE FOOT CONTROL (SURGICAL CART)


- Extract the column upper cover using the nylon rod (a) supplied (see fig. 1).

NOTE: Fully charge the foot control battery before


using it for the first time.

NOTE: When the dental unit is switched on wait


about 20 sec. and act on the control lever (B) to acti-
vate operation of the wireless foot control.

- Fit the removed items by repeating the procedure in reverse order.


INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT 9
A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

7 - BACKREST INSTALLATION (OPTIONAL - not in A3Plus)

10 INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT 11


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

8 - INSTALLING THE DENTAL UNIT

(A3 Plus)

12 INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

(A5 / A6 Plus / A7 Plus)

INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT 13


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

14 INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT 15


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

WARNING! Keep the dentist’s module arms (a - b) together.

16 INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT 17


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

9 - DENTAL UNIT / DENTAL CHAIR LEVELLING


CAUTION: Lubricate the unpainted parts of poles / arms with silicon grease type S1.

18 INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

10 - DENTAL CHAIR ELECTRICAL CONNECTION

WARNING! before the connection of the main supply (CN2: 230/115/100 Vac), check the main switch of the
dental surgery is OFF.

- Connect the main supply on CN2 (boad 97661871): LINE on “L” - NEUTRAL on “N”.

INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT 19


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

11 - ANCHORING HOLES
11.1 - DENTAL CHAIR ANCHORING HOLES
WARNING!
- Installation must always be permanent with anchoring to the floor (Standard CEI EN 60601).
- The floor conditions (continuous type) must meet the load-bearing capacity as set forth in standard “DIN
1055”. The total weight of the dental unit with a 190 Kg. Patient is approximately 395 Kg.

- Fix the chair to the floor in positions (F) by using the supllied nogs (C).

20 INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

(With JOYSTICK)

INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT 21


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

A3 Plus
(Without JOYSTICK)

22 INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

WARNING! The characteristics of the floor must assure the breaking strength of the anchor bolts is not
less than 1200 daN each (considering Rck cement strength 20 MPa).

NOTE:
Wherever concrete floors are not present, it is advisable to:
- have qualified personnel prepare an area fir anchoring the bolts provided with the equipment;
- or, have qualified personnel prepare an area with embedded threaded bars. In this case, the dimensional specifica-
tions for securing (not provided) must be:
- no. 2 threaded bars M10;
- length min. 40 mm, length max. 60 mm;
- bar strength class 8.8;
- medium nut M10 UNI 5588 strength class 8.

NOTE:
Wherever brick floors are present, chemical anchors should be used (not supplied):
FISCHER chemical anchor consisting of:
- anchor FIS HK H20x85 (item 41906);
- steel bush with internal threading FIP 14x80 M10 i (item 43633);
- T-BOND chemical anchor.
Drilling: Ø20 mm; depth: 95 mm.
Thoroughly clean the hole before installation. Any type of silicon gun can be used to apply T-BOND resin.
When the cartridge is new, discard the initial resin until it is an even color before installing it. Inject resin regularly starting
from the bottom of the hole (fill at least 2/3). Put in the bush by turning it to assure even distribution and adherence of
the resin.

Observe the hardening and definitive bonding times:

Support temperature °C Hardening time Application time


+0 ÷ +5 - 360 min
+5 ÷ +10 15 min 180 min
+10 ÷ +20 8 min 120 min
+20 ÷ +30 5 min 60 min
+30 ÷ +40 3 min 30 min

ìHardening time refers to the time within which perpendicularity and the position of the bush can be slightly adjusted.
Application time refers to the minimum time required for the resin to completely harden for definitive bonding.

WARNING!
In any case, as effective anchoring is strictly dependant on the material on which it is done and given the va-
riety of ways in which floors are constructed, the installation worker is responsible for assuring the support is
suitable.
The fastening characteristics must assure the extraction/breaking strength is not less than 1200 daN each.
Cefla s.c. shall not be held responsible for any personal injury or property damage deriving from failure to fol-
low the instructions given above.

INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT 23


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

11.2 - FLOOR BOX CASTING ANCHORING HOLES (SINGLE CART)

24 INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

12 - ATTACHING THE TUBES / WIRING TO THE DENTAL CHAIR

INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT 25


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

5x370

26 INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

13 - INSTALLING THE DENTAL CHAIR CARTERS

INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT 27


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

14 - ELECTRICAL CONNECTIONS
- Connect up the electrical leads coming from the dental unit in the positions indicated in fig. 1 (module
code 97661871).
B
- Connect the YELLOW/GREEN EARTH wire ( ) coming from the dental unit to the connections
box.

WARNING! before the connection of the main supply (CN2: 230/115/100Vac +/- 10V),
check the main switch of the dental surgery is OFF.
- Connect the main supply on CN2 (boad 97661871): LINE on “L” - NEUTRAL on “N”.

(Code 97661871)

NOTE: Earth connection (A) to be


executed where 115/100 Vac power
supplies are in use.

28 INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

Suction electrical connection

(SINGLE SUCTION)

LINE (230 VAC)

Differential
safety switch

(Optional)

TO SUCTION
MOTOR

(CENTRALIZED SUCTION)

CAUTION:
The relay (K4) can
support a maximum
current of 5A.

TO SUCTION CONTROL PANEL (SINGLE CONTACT: 24VAC-5A)

INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT 29


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

15 - AIR CONNECTIONS
- Remove the caps fitted into the water drainpipe and the air suction hose to prevent the remaining water from gushing
out.
- Connect the water pipes and the air hoses as illustrated in fig. 1/2. In order to avoid errors, each fitting comes with a
piece of the corresponding pipe/hose. Remove it before connecting the water/air lines.
- To connect the drain and suction system, use the fittings and the seal supplied with the dental unit.

CAUTION: Drain and bleed the air and water lines before connecting them.

CAUTION: (8) White hose Ø 6x8 (Water inlet): Bending radius ≥ 60 mm.

NOTE:
For installations compliant with DVGW/SVGW (GERMANY / SWITZERLAND): the water delivery pipe to the dental
unit (8) and all components for connection to the water mains must be DVGW- / SVGW-certified.

- After the hydro-pneumatic connections, check the air and water pressure with the special pressure gauge code
91104498 (See “SPECIAL TOOLS”).

(Without W.H.E. / With BY-PASS: S.H.S.-S)

(1) Clear hose Ø 3x1.7 (Pneumatic headrest air)

(3) White tube Ø 6x4 (Bowl water)

(4) Blue tube Ø 3x5 (Air outlet 3 Bar)

(5) Blue tube Ø 6x8 (Air outlet 6 Bar)

(6) Blue tube Ø 6x8 (Air inlet)

(7) Green tube Ø 6x4 (To dental unit)


(DENTAL CHAIR)
(8) White tube Ø 6x8 (Water inlet)
(UNIT BODY)
(9) Green tube Ø 6x4 (Water 2.5 Bar_fast connectors) - Optional

(BY-PASS)

30 INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

(With W.H.E. + BIOSTER)

(1) Clear hose Ø 3x1.7 (Pneumatic headrest air)


(BIOSTER)
(2) Clear tube Ø 3x5 (H2O2 tank)

(3) White tube Ø 6x4 (Bowl water) (W.H.E.)

(4) Blue tube Ø 3x5 (Air outlet 3 Bar)

(5) Blue tube Ø 6x8 (Air outlet 6 Bar)

(6) Blue tube Ø 6x8 (Air inlet)


(DENTAL
CHAIR)
(7) Green tube Ø 6x4 (To dental unit)
(UNIT
BODY)
(8) White tube Ø 6x8 (Water inlet)

(9) Green tube Ø 6x4 (Water 2.5 Bar_fast connectors) - Optional

(*) W.H.E. system overflow tube (Clear hose Ø 9x12):


Leave the tube disconnected in the connection space (A).
The tube must not be connected to the drainage union (B). Otherwise, the DVGW - EN1717 certification will imme-
diately expire.

Suction system

D
Water outflow

INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT 31


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

(With CART)

(1) Clear hose Ø 3x1.7 (Pneumatic headrest air)

(3) White tube Ø 6x4 (Bowl water)

(4) Blue tube Ø 3x5 (Air outlet 3 Bar)

(5) Blue tube Ø 6x8 (Air outlet 6 Bar)

(6) Blue tube Ø 6x8 (Air inlet)

(7) Green tube Ø 3x1,7 (To dental unit)


(DENTAL CHAIR)
(8) White tube Ø 6x8 (Water inlet)
(UNIT BODY)
(9) Blue tube Ø 3x1,7 (To dental unit)

(CART)

32 INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

16 - SET THE LIMIT SWITCHES FOR DENTAL CHAIR MOVEMENTS (AUTOMATIC SE-
TUP)
NOTE: The dental chair’s backrest and seat must not be at the end of their stroke but in an intermediate posi-
tion.

INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT 33


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

17 - ASSEMBLING THE EMERGENCY PATIENT CHAIR BASE CONTROL

34 INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT 35


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

18 - INSTALLING THE SUCTION TUBES

19 - TRAY-HOLDING TABLE ON ASSISTANT’S BOARD

20 - INSTALLING THE WATER SUPPLIERS TO THE CUP AND THE BOWL / FILTER AND
FILTER COVER

36 INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

21 - ASSEMBLING THE 6th INSTRUMENT CORD SUPPORT (CONTINENTAL)

INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT 37


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

22 - ASSEMBLING THE TRAY HOLDER TABLE ON THE DENTIST’S MODULE

13 mm

13 mm

38 INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

(INTERNATIONAL)

INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT 39


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

23 - ASSEMBLING THE MANUAL HEADREST

(*) NOTE:
If you wish to make the headrest
rod slide more smoothly remove
a spring.

40 INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

24 - PNEUMATIC HEADREST INSTALLATION

INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT 41


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

PNEUMATIC HEADREST INSTALLATION (OPTIONAL)

42 INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

25 - OPERATING LAMP
NOTE: Lubricate the pin (c) of the light arm with silicone grease type S1.

26 - ASSEMBLING THE RIGHT ARMREST

INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT 43


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

27 - DENTAL CHAIR PADDING INSTALLATION

27.1 - Installing the headrest padding


(*) NOTE: Position of quick-connect coupler before making adjustments (if necessary).

27.2 - Backrest padding installation


(*) NOTE: Position of quick-connect coupler before making adjustments (if necessary).

(OPTIONAL)

44 INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

(STANDARD)

INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT 45


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

46 INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

(*) NOTE: Position of quick-connect coupler before making adjustments (if necessary).

INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT 47


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

Should it be necessary to remove backrest cover, operate as follows:

48 INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

27.3 - Installing the seat padding

INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT 49


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

28 - CHECK DURING THE ASSEMBLING AND/OR PERIODIC MAINTENANCE (ANNUAL)

PRECAUTION: After installation, carry out a decontamination cycle of the dental unit water ducts using
the suitable device (see sect. “Tools for the technician” - website: extranetmedical.cefla.it).

PREINSTALLATION CHECK
Intact external package 
Check compliance of delivery documents 
Check ID label/package labels match 
Check place of installation and supplies (*) 
POWER SUPPLY CHECK
Connection of power and earth leads 
Power supply voltage (Vac +\- 10%) 
Water feed connection seals 
Compressed air pressure 
Operation of main switch 
Check presence of power mains protection 
Check water supply pressure 
Check water hardness (°f) 
ANTI-ENTRAPMENT SAFETY DEVICE CHECKS
The dental unit has been correctly fixed to the floor, as indicated in the installation manual 
Patient chair base safety stop / suction stop 
Backrest safety stop 
Assistant’s module safety stop 
MECHANICAL CHECKS
Balance of dentist’s module pantograph arm 
Balance of operating light pantograph arm 
Rotation / braking of dental unit / monitor and light arms 
Rotation / braking of cuspidor bowl 
Headrest sliding and articulation 
Proper attachment and good condition of dental unit casing 
Proper attachment and good condition of upholstery 
Programmed patient chair positions 
DEVICE CHECK AND CALIBRATION
Software installation and configuration 
Control panel functions 
Foot controls and rheostat function 
Suction system and (where applicable) amalgam separator system 
Dental unit water supply tank sanitising system (*): operation and emptying 
Separate feed system (*) 
Quantity of water fed to cup 
Quantity of water fed to cuspidor bowl 
Heater function for water 
Function check/replacement of spray air sanitisation filter (*) 
INSTRUMENT CHECK AND CALIBRATION
Syringe feed air 
50 INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT
A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

Syringe feed water 


Syringe spray heating function 
Turbine rotation air 
Micromotor cooling air 
Check Micromotor operating mode (*) 
Air / spray water to instruments functions – head nozzles cleaning 
Air / spray water to syringes functions - tip nozzles cleaning 
Spray water heating function 
Fibre optic ON function (*) 
Speed / power adjustment (all instruments) 
Curing lamp check (*) 
Apex locator check (*) 
Out drop no retraction 
Instrument recall arms functions 
Peristaltic pump functions (*) 
Camera functions and image quality check (*) 
OPERATING LIGHT CHECK
Power supply voltage 
On-off and brightness adjustment (*) 
Cooling fan functions (*) 
MULTIMEDIA
Power on and monitor functions (*) 
COMPRESSOR
Working pressure check (via compressor pressure gauge - 6÷8 bar) 
Dryer efficiency check 
Condensate discharge from tank 
disinfecting system 
Suction tube washing system 
Water duct disinfection system (*) 
Water duct washing system (*) 

INSTALLATION CARRIED OUT IN ACCORDANCE WITH LOCAL LAW 


Check presence of Quick Guide 
(*) if applicable

NOTE:
For reference values use the data indicated in the dental unit technical manual.

WARNING!
Maintenance work on the suction/drainage system must be carried out using gloves, face mask and eye goggles
to prevent contact with infected material.

INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT 51


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

29 - INSTALLING THE HYDROGROUP DOOR

(A3 Plus)

52 INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

(A5 / A6 Plus / A7 Plus)

INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT 53


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

30 - INSTALLING THE COVER OF THE FLOOR BOX

54 INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

Z
Z

(SINGLE CART)

INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT 55


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

31 - LEVELING AND ADJUSTING THE ARM’S CLUTCH

31.1 - Adjusting the fixed arm’s clutch

(A6 Plus / A7 Plus)

(A3 Plus / A5)

56 INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

31.2 - Adjusting the pantograph arm’s clutch

(A6 Plus / A7 Plus)

(A3 Plus / A5)

(A) PRECAUTION: Position the slit of the braking element, visible from the central hole of the braking grub
screw, in a vertical position using a screwdriver.
INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT 57
A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

31.3 - Calibrating the pantograph arm spring / Leveling the instrument board

Calibrating the pantograph arm spring

Leveling the instrument board

58 INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

31.4 - Adjusting the dentist’s instrument board / dental assistant’s instrument board

(A3 PLUS INTERNATIONAL)

INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT 59


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

32 - ADJUSTING THE OSCILLATING ARM SPRING

(STANDARD)

[Scaler - Syringe - Telecamera - T LED light]

(STANDARD)

[Turbines (all) - Micromotor]

60 INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

(SIDE FLEX)

33 - CLEANING THE TURBINE RETURN AIR FILTER

INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT 61


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

34 - INSTRUMENT’S TABLE ARMS EXTRA-ROTATION


WARNING!
Installation must always be permanent with anchoring to the floor (see par. 9 “DENTAL CHAIR ANCHORING
HOLES”).
CEFLA s.c. will not be liable for any damage due to not following the guidelines detailed in the installation ma-
nuals.

62 INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT 63


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

64 INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

35 - SET THE LIMIT SWITCHES FOR DENTAL CHAIR MOVEMENTS (MANUAL SETUP)

INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT 65


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

66 INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

36 - CONNECTION TO CLOUD

USER (same used to access Extranet Medical)

PASSWORD (same used to access Extranet Medical)

INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT 67


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

Available device
serial number

station_SSID:
Network name

station_PWD:
Network Password

In the event that a FIREWALL is present, for the correct connection of the device to the Cloud it is obligatory
the opening in exit towards the addresses:
- 52.236.129.90
- 13.80.27.52
- Door 8883

68 INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

Enter the serial number or


scan the QR code on the
inside of the hydrogroup

station_SSID:
Network name

station_PWD:
Network Password

In the event that a FIREWALL is present, for the correct connection of the device to the Cloud it is obligatory
the opening in exit towards the addresses:
- 52.236.129.90
- 13.80.27.52
- Door 8883

INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT 69


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

37 - CHECK OF THE CORRECT CONNECTION TO THE CLOUD

The connection must be verified through the portal that can be used via smartphone, tablet
and PC.

https://easycheck.cefla.it

USER (same used to access Extranet Medical)

PASSWORD (same used to access Extranet Medical)

Select the menu


“Devices”

Search for the dental


unit by entering the serial
number in the “Serial
number” field
and then selecting
the search button.

Successful connection is
indicated by the icon ( )
that appears in the “Cloud”
column.

70 INSTALLATION AND MECHANICAL ADJUSTMENT OF DENTAL UNIT


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

CONTENTS

ACCESSORIES AND MULTIMEDIA INSTALLATION................................................................................... 2

1 - ASSEMBLING THE INTEGRATED RX DC CONTROL UNIT (NOT IN A3PLUS).................................. 2

2 - ASSEMBLING THE INTEGRATED RX DC (NOT IN A3PLUS)............................................................... 5

3 - ASSEMBLING THE “BARCO’’ MONITOR................................................................................................ 9

4 - CONNECTING THE MONITOR ONLY..................................................................................................... 11

5 - DOUBLE ARM MONITOR LEVELLING................................................................................................... 15

6 - MONITOR / RX DC..................................................................................................................................... 17

ACCESSORIES AND MULTIMEDIA INSTALLATION 1


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

ACCESSORIES AND MULTIMEDIA INSTALLATION

1 - ASSEMBLING THE INTEGRATED RX DC CONTROL UNIT (not in A3Plus)

WARNING! The chair-integrated x-ray unit must be installed only on dental chairs fitted with stabilizer
plate, or on dental chairs that are firmly attached to the floor with plugs/anchors or threaded bars, as specified
in the installation manual.
CEFLA s.c. will not be liable for any damage due to not following the guidelines detailed in the installation ma-
nuals.

2 ACCESSORIES AND MULTIMEDIA INSTALLATION


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

ACCESSORIES AND MULTIMEDIA INSTALLATION 3


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

4 ACCESSORIES AND MULTIMEDIA INSTALLATION


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

2 - ASSEMBLING THE INTEGRATED RX DC (not in A3Plus)

CAUTION: Lubricate the unpainted parts of poles/arms with silicon grease type S1.

(*) NOTE: Screw in the rotation stop grub screw of the extension arm: screw it all the way in then unscrew it by ½ a
turn.

ACCESSORIES AND MULTIMEDIA INSTALLATION 5


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

(A) PRECAUTION: Position the slit of the braking element, visible from the central hole of the braking grub
screw, in a vertical position using a screwdriver.

(**) NOTE: Screw in the rotation grub screw of the pantograph arm: screw it all the way in then unscrew it by ½ a
turn.

6 ACCESSORIES AND MULTIMEDIA INSTALLATION


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

ACCESSORIES AND MULTIMEDIA INSTALLATION 7


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

- Carry out the following electrical connections.

X-RAY UNIT

UNIT BODY

RX DC CONTROL UNIT

LINE

LINE

CONNECTIONS BOX

8 ACCESSORIES AND MULTIMEDIA INSTALLATION


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

3 - ASSEMBLING THE “BARCO’’ MONITOR


CAUTION: Lubricate the unpainted parts of poles/arms with silicon grease type S1.

ACCESSORIES AND MULTIMEDIA INSTALLATION 9


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

MONITOR “BARCO” (TOUCHSCREEN) MONITOR “BARCO” (NO TOUCHSCREEN)

10 ACCESSORIES AND MULTIMEDIA INSTALLATION


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

4 - CONNECTING THE MONITOR ONLY

ACCESSORIES AND MULTIMEDIA INSTALLATION 11


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

NOTE:
Place one of the slots at the base of the post in the rotation stop grub screw located in its seat. The proper slot
will allow for rotation as shown in fig. 5 / 6.

12 ACCESSORIES AND MULTIMEDIA INSTALLATION


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

- Carry out the following electrical connections.

CAUTION: Disconnect the power supply


connector (12 VDC) from the monitor and recon-
nect it only after checking for correct polarity:
- internal connector contact = 12 VDC
- external connector contact = 0 VDC

Red wire (12 VDC)

Black wire (0 VDC)

UNIT
BODY

ACCESSORIES AND MULTIMEDIA INSTALLATION 13


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

NOTE: After installing the monitor it is recommended that you carry out:
- self-adjustment of display and volume (see user’s manual);
- adjustment of the touchscreen functions of the TOUCHSCREEN monitor (see user’s manual).
14 ACCESSORIES AND MULTIMEDIA INSTALLATION
A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

5 - DOUBLE ARM MONITOR LEVELLING

- Undo the screw (A) pictured in figure 3 by 1/2 turn.

ACCESSORIES AND MULTIMEDIA INSTALLATION 15


A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

- Undo the three screws (B) pictured in figure 4 by 1 turn (to increase the adjustment range, undo the three screws by
more than 1 turn).

- Raise and hold up the monitor and tighten the grub screw (C), pictured in figure 5, until it is supported.

- Tighten the grub screws (D), pictured in figure 6, until they are supported.

- Tighten the 4 fixing screws (A ; B).


- Place the plug back in its special holder (see fig. 2).
16 ACCESSORIES AND MULTIMEDIA INSTALLATION
A3 Plus - A5 - A6 Plus - A7 Plus (Series 2019) - TECHNICAL MANUAL

6 - MONITOR / RX DC

(2)

(1)

(1) = Single arm screen fixing hole.


(2) = Double arm screen fixing hole.

ACCESSORIES AND MULTIMEDIA INSTALLATION 17

You might also like